A Platform to Identify Endogenous Metabolites Using a Novel High Performance Orbitrap MS and the mzCloud Library

Junhua Wang,1 David A. Peake,1 Robert Mistrik,2 Yingying Huang1

1Thermo Fisher Scientific Inc, San Jose, CA, USA; 2HighChem, Ltd., Bratislava, Slovakia

Page 2: A Platform to Identify Endogenous Metabolites …...A Platform to Identify Endogenous Metabolites Using a Novel High Performance Orbitrap MS and the mzCloud Library Junhua Wang, 1

2 A Platform to Identify Endogenous Metabolites Using a Novel High Performance Orbitrap MS and the mzCloud Library

60 80 100 120m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

96.0556

69.0447

RT: 3.1775 - 3.1907

3.178 3.180 3.182 3.184 3.186 3.188 3.190Time (min)

0

10

20

0

10

20

Rel

ativ

e A

bund

ance 0

10

20NL: 3.38E6Base Peak m/z= 181.0703-181.0739 MS b667JWUrine_HCDIT_CIDOTtop2_500ms

NL: 1.59E5Base Peak F: ms2 MS b667JWUrine_HCDIT_CIDOTtop2_500ms

NL: 2.93E4Base Peak F: ms3 MS b667JWUrine_HCDIT_CIDOTtop2_500ms

3.1791 3.1891Full Scan MS

MS2

MS3

0.6 sec

HCDIT-MS2

60 80 100 120 140 160 180m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

124.0153

105.0077

180.9644141.9145

95.077369.0120

CIDOT-MS3

60 80 100 120 140 160 180m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

124.0153

105.0077

180.9644141.9145

95.077369.0120

An Improved Metabolomics WorkflowA traditional workflow for metabolite profiling and identification is based on a full scan LC/MS experiment followed by targeted MS/MS confirmation.

The major problems associated with this methodology are:1) Instrumentation limitations

• Does not allow simultaneous profiling and MSn in a single run, the scan speed su ers.

• Identification is usually based on accurate mass and MS2, but a large number of LC peaks remain not identifiable .

2) MS/MS library limitations• MS/MS spectral search is usually manually performed and time-

consuming; • Automated scoring for spectral matching is usually not available;• The fragmentation needs to be manually interpreted, which is not

only time-consuming but also prone to errors.

3) There is no good solution for unknowns.

The Orbitrap Fusion MS enables fast parallel data acquisition on a UHPLC timescale with a variety of dissociation choices The Orbitrap Fusion MS o ers ultimate flexibility in fragmentation methods at all stages of MSn analysis, with fragment ions detected by either mass analyzer. CID-HCD, HCD-CID, and HCD-HCD can all be set up as simple as the conventional CID-CID. This complete flexibility facilitates the examination of fragmentation pathways that leads to the most comprehensive structural information.

mzCloud MSn library enables MSn search for both high and low resolution mzCloud is a multistage MS library. Based on the spectral tree similarity, an algorithm called precursor ion fingerprinting (PIF) enables the identification of compounds even if they are not in the library, and the fragmental peaks can be annotated [2]. Powered by the novel MS and the mzCloud library, an improved, more connected workflow is shown below:

Major benefits of the improved workflow: 1) HCD MS/MS can be performed simultaneously with full scan MS on the

UHPLC timescale with ion trap detection at >22 Hz.2) Simultaneous ddMSn ion tree with flexible combinations of CID and/or HCD

fragmentation with ion trap and/or OrbitrapTM detection. 2) Both the MS2 and MSn will be automatically searched in mzCloud, the

matching entry will be scored based on PIF, and the fragments will be annotated.

4) De novo structural identification is enabled by using this workflow.

OverviewPurpose: To demonstrate the capability of the Thermo Scientific™ Orbitrap Fusion™ Tribrid™ Mass Spectrometer in combination with MSn library for metabolite structural identification.

Methods: UHPLC coupled to the Orbitrap Fusion MS to simultaneously perform global profiling and structural elucidation with MSn.

Results: Within UHPLC timescale (2-3 second FWHM), >10 cycles can be finished with fast ion trap HCD MS2 for top 11 precursors, each with ion trees of MS2-MS3(top2)-MS4(top2).

IntroductionThe structural identification of metabolites represents a significant challenge in metabolomics study. Multistage mass spectrometry (MSn) is a powerful tool for compound identification and structural elucidation that goes beyond identifying the exact same compounds, to discovering additional compounds with structural homology to those in the library. Combining with high resolution accurate mass (HRAM) measurement, MSn is highly e ective in identifying the unknown but biologically relevant compounds in metabolomics studies[1]. However, the speed on current platforms is yet to be e ectively compatible with UHPLC front end, which limited such application in biological samples.

Presented here is a new platform to identify endogenous metabolites using a novel high performance Orbitrap hybrid instrument in conjunction with UHPLC and an MSn library.

Methods Sample Preparation

Urine samples from an adult male were analyzed. The urine samples were diluted 1 to 10 with water and processed with a 3kD MWCO filter.

For plasma samples, proteins were removed with cold methanol and centrifugation, followed by centrifugal evaporation at 35 oC. The residue was reconstituted in MeOH/Water, 1/9, and also processed with a 3kD MWCO filter.

Liquid ChromatographyUHPLC separation was implemented on a Dionex Ultimate 3000 HPG (high-pressure gradient) pump using Hypersil GOLD RP C18 column at 450 L/min, column temperature at 55 oC. LC solvents were 0.1% FA (A) and 0.1% FA in MeOH (B). A linear gradient was applied from 0.5-50% B for 5.5 min, followed by increasing to 98% at 6 min, held 98% B for 6 min, then decreased to 0.5% at 13min, and equilibrated for another 2 min.

FIGURE 1. Schematic of Orbitrap Fusion Tribrid Mass Spectrometer

A Platform to Identify Endogenous Metabolites Using a Novel High Performance Orbitrap MS and the mzCloud Library Junhua Wang1, David A. Peake1, Robert Mistrik2, Yingying Huang11Thermo Fisher Scientific Inc, San Jose, CA; 2HighChem, Ltd., Bratislava, Slovakia

Results

S-Lens

Active beam guide (ABG)

Ultra-high-field Orbitrap mass analyzer

Large-surface-Area detector (LSAD)

Ion-routing multipole

Dual-pressure linear ion trap

Quadrupole mass filter

C-Trap

EASY-Internal Calibration ion source

FIGURE 2. Ultrahigh resolution resolves the isotopes

PositiveMode20130518092831 #656 RT: 2.64 AV: 1 NL: 1.48E8T: FTMS + p ESI Full ms [100.00-212.08]

150.0 150.5 151.0 151.5 152.0m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

150.0584R=627006

151.0618R=621502

152.0542R=601802

PositiveMode20130518092831 #654 RT: 2.60 AV: 1 NL:T: FTMS + p ESI Full ms [100.00-212.08]

151.05 151.06m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

PositiveMode20130518092831 #539 RT: 1.44 AV: 1 NL:T: FTMS + p ESI Full ms [100.00-212.08]

151.05 151.06m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

PositiveMode20130518092831 #511 RT: 1.33 AV: 1 NL:T: FTMS + p ESI Full ms [100.00-212.08]

151.05 151.06m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

PositiveMode20130518092831 #615 RT: 1.97 AV: 1 NL:T: FTMS + p ESI Full ms [100.00-212.08]

151.05 151.06m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

PositiveMode20130518092831 #237 RT: 0.51 AV: 1 NL:T: FTMS + p ESI Full ms [100.00-212.08]

151.05 151.06m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

R=30k

R=60k

R=120k! =240kR=500k

L-Methionine C5H11NO2S

0.7ppm

Fine isotope pattern

C5H11NO2S +H: C5 H12 N1 O2 S1 p(gss, s/p:40) Chrg 1...

150.0 150.5 151.0 151.5 152.0m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

150.0583R=499553

151.0617R=499368

152.0541R=499443

PositiveSimulated

R=500,000

0.0ppm

C5H11NO2S +H: C5 H12 N1 O2 S1 p(gss, s/p:40) Chrg 1...

151.050 151.055 151.060 151.065m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

151.0617C4

13CH12O2NS

151.0625C5H12O17ONS

151.0646C5

13C14CH918O17O

151.0577C5H12O2N33S

151.55414C4H10

18O233S

PositiveObserved

Negative20130518111518 #1449 RT: 8.07 AV: 1 NL: 1.00E6T: FTMS - p ESI Full ms [140.00-220.00]

179 180 181 182m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

179.0576R=550402

180.0611R=330000

NegativeFull Scan

0.55 ppm

Negative20130518111518 #158 RT: 1.66 AV: 1 NL: 1.24E6T: FTMS - p ESI Full ms [140.00-220.00]

179.050 179.055 179.060 179.065m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

179.0573R=67606

Negative20130518111518 #110 RT: 1.34 AV: 1 NL: 8.48E5T: FTMS - p ESI Full ms [140.00-220.00]

179.050 179.055 179.060 179.065m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

179.0574R=148204

179.0558R=130000

Negative20130518111518 #54 RT: 0.83 AV: 1 NL: 9.54E5T: FTMS - p ESI Full ms [140.00-220.00]

179.050 179.055 179.060 179.065m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

179.0574R=304604

179.0560R=275000

Negative20130518111518 #21 RT: 0.40 AV: 1 NL: 7.38E5T: FTMS - p ESI Full ms [140.00-220.00]

179.050 179.055 179.060 179.065m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

179.0574R=607806

179.0560R=562802

R=60,000 R=120,000

R=240,000 R=500,000200 M

1 M

1.1 ppm

0.56 ppm

0.55 ppm

0.55 ppm

0.56 ppm

1.7 ppm

0.55 ppm

FIGURE 3. Ultrahigh resolution resolves isobaric metabolites

RT: 2.75160 - 2.86730

2.76 2.77 2.78 2.79 2.80 2.81 2.82 2.83 2.84 2.85 2.86Time (min)

0

10

20

30

40

50

60

70

80

90

1000

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

205.09721

205.09724

205.09721

205.09723

205.09717

205.09717205.09717205.09715 205.09718 205.09720

205.09749

205.09747

205.09741205.09746

205.09740205.09770

205.09772205.09746 205.09735 205.09738

NL:3.23E7Base Peak m/z= 205.09525-205.09935 F: ms MS JWUrine_FS240k_LM_Re

NL:2.85E7Base Peak m/z= 205.09525-205.09935 MS jwurine_fs240k_nolm_re

PBUrine_MS2_120k_600ms_bothLM #824 RT: 2.82 AV: 1 NL: 8.68E6F: FTMS + p ESI d Full ms2 [email protected] [50.00-216.00]

80 100 120 140 160 180 200m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

146.06033

118.06544

132.08105 188.07079170.06030without Int. Cal.Avg =1.5 ppm

Full Scan MS HCD MS/MSJWUrine_MS2_120k_600ms_bothLM #896 RT: 2.84 AV: 1 NL: 1.08E7T: FTMS + p ESI d Full ms2 [email protected] [50.00-216.00]

80 100 120 140 160 180 200m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

146.06020

118.06530

132.08090 170.06015 188.0706391.05422

0.2 ppmO

NH

NH

OHO

NH

0.9 ppm 0.7 ppmNH •

0.4 ppm

0.8 ppm

N2H

1 ppm

NH

N 2HNH

-0.1 ppm

with Int. Cal.Avg = 0.4 ppm

1.1 ppm

1.8 ppm

2.0 ppm

1.7 ppm

1.9ppm

2.9 ppm

1.1 ppm

with Int. Cal.

without Int. Cal.

Metabolite Identification Requires High Resolution and Isotope Ratio Measurements Structure elucidation of unknown small molecules by mass spectrometry is a great challenge. The first crucial step is to obtain correct elemental compositions.

Natural occurring elements can be monoisotopic (F, Na, P, I) or polyisotopic (H, C, N, O, S, Cl, Br) [1]. The molecule L-Methionine C5H11NO2S contains multiple polyisotopic elements, especially the S, which result in multiple isotopic peaks needing resolving power above 240,000 at m/z 200 (FWHM). Figure 2A shows the better isotope pattern of molecular ion [M+1+H]+. With increasing resolution up to 450,000 (FWHM), all five isotope peaks are clearly resolved. Figure 2B shows the simulated isotope pattern.

A

B

Metabolite Identification and Quantitation in Complex Samples Requires High ResolutionD-Glucose C6H12O6 and Paraxanthine C7H8N4O2 co-exist in human blood at distinct concentration levels: 5300 M (D-Glucose) vs. 10 M (Paraxanthine) [3]. Although D-Glucose is hundreds of fold more concentrated, it is less readily detectable because of its low ionization e ciency. Their masses di erentiate by 0.0013 amu (179.0561 vs. 179.0574), making the detection of these individual components very challenging. Figure 3 shows that, with >300,000 resolving power, the 200 M (D-Glucose) and 1 M (Paraxanthine) can be readily separated and quantified.

FIGURE 4. Easy Internal Calibration

Metabolite Identification Requires High Accuracy of Mass Measurement in Both MS Full Scan Level and MS/MS LevelFor an unknown metabolite, the number of calculated chemically possible formulae could be hundreds (20 ppm tolerance window), but the possibilities significantly reduce to tens (3 ppm) and several if the tolerance window is narrowed to sub-ppm. However, due to the large amount of known isomeric structures, one single formula may correspond to hundreds of structural isomers, thus structure identification of metabolite isomer using only MS full scan is not feasible. MS/MS and MSn techniques can help solve this analytical problem.

The Orbitrap Fusion MS has a built-in internal calibration functionality (Easy IC). Figure 4 shows LC-MS/MS data of L-Tryptophan from an urine sample with and without internal calibration. Sub-ppm mass accuracy was observed for MS and MS2 fragments with IC activated.

FIGURE 5. Rapid Ion Tree Acquisition on UHPLC Timescale from a Urine Sample (1 MS, 2 MS2 and 4 MS3 in 0.6sec/cycle).(A) Ion trap HCD MS2 followed by Orbitrap CID MS3 were acquired for the precursor ion m/z 181.0721. (B) Spectral tree search in mzCloud matches the MS3 with MS3 from theophylline with the highest score, indicating the unknown must have the same substructure as theophylline.

ConclusionThe Orbitrap Fusion MS o ers ultra high resolution and high fidelity in isotope ratio measurements of metabolites. This allows metabolites in complex mixture with a wide dynamic range be resolved and detected. The Easy-Internal Calibration allows compound assignment be readily done with confidence. The Orbitrap Fusion MS provides ultimate flexibility that facilitates the examination of multiple structural pathways for more structural information at higher speed.

The mzCloud library allows metabolite identification with MS2 and/or MSn data in an automated fashion. The identification based on multistage MS spectral trees and PIF algorithm opens the possibility to identify unknown metabolites and isomers. An improved metabolomics workflow is enabled.

References1. Budzikiewicz H, Grigsby RD, Mass Spectrom. Rev. 2006, 25(1):146-157.

2. Sheldon MT, Mistrik R, Croley TR, J. Am. Soc. Mass Spectrom. 2009, 20(3):370-376

3. http://www.hmdb.ca/metabolites/HMDB00122 & HMDB01860

A

BTheophylline

MS3

isomers

Query (down) and library spectra (up) matchScore

N

NH NH

N

NH

1.5 ppm 1.5 ppm

OH

NH

1 ppm

2.1 ppm

FIGURE 1. The Orbitrap Fusion MS and the Drag and Drop Instrument Method Editor.

O

N

NH NH

Page 3: A Platform to Identify Endogenous Metabolites …...A Platform to Identify Endogenous Metabolites Using a Novel High Performance Orbitrap MS and the mzCloud Library Junhua Wang, 1

3Thermo Scientific Poster Note • PN ASMS13_MP 045_JWang e 06/13S

60 80 100 120m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

96.0556

69.0447

RT: 3.1775 - 3.1907

3.178 3.180 3.182 3.184 3.186 3.188 3.190Time (min)

0

10

20

0

10

20

Rel

ativ

e A

bund

ance 0

10

20NL: 3.38E6Base Peak m/z= 181.0703-181.0739 MS b667JWUrine_HCDIT_CIDOTtop2_500ms

NL: 1.59E5Base Peak F: ms2 MS b667JWUrine_HCDIT_CIDOTtop2_500ms

NL: 2.93E4Base Peak F: ms3 MS b667JWUrine_HCDIT_CIDOTtop2_500ms

3.1791 3.1891Full Scan MS

MS2

MS3

0.6 sec

HCDIT-MS2

60 80 100 120 140 160 180m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

124.0153

105.0077

180.9644141.9145

95.077369.0120

CIDOT-MS3

60 80 100 120 140 160 180m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

124.0153

105.0077

180.9644141.9145

95.077369.0120

An Improved Metabolomics WorkflowA traditional workflow for metabolite profiling and identification is based on a full scan LC/MS experiment followed by targeted MS/MS confirmation.

The major problems associated with this methodology are:1) Instrumentation limitations

• Does not allow simultaneous profiling and MSn in a single run, the scan speed su ers.

• Identification is usually based on accurate mass and MS2, but a large number of LC peaks remain not identifiable .

2) MS/MS library limitations• MS/MS spectral search is usually manually performed and time-

consuming; • Automated scoring for spectral matching is usually not available;• The fragmentation needs to be manually interpreted, which is not

only time-consuming but also prone to errors.

3) There is no good solution for unknowns.

The Orbitrap Fusion MS enables fast parallel data acquisition on a UHPLC timescale with a variety of dissociation choices The Orbitrap Fusion MS o ers ultimate flexibility in fragmentation methods at all stages of MSn analysis, with fragment ions detected by either mass analyzer. CID-HCD, HCD-CID, and HCD-HCD can all be set up as simple as the conventional CID-CID. This complete flexibility facilitates the examination of fragmentation pathways that leads to the most comprehensive structural information.

mzCloud MSn library enables MSn search for both high and low resolution mzCloud is a multistage MS library. Based on the spectral tree similarity, an algorithm called precursor ion fingerprinting (PIF) enables the identification of compounds even if they are not in the library, and the fragmental peaks can be annotated [2]. Powered by the novel MS and the mzCloud library, an improved, more connected workflow is shown below:

Major benefits of the improved workflow: 1) HCD MS/MS can be performed simultaneously with full scan MS on the

UHPLC timescale with ion trap detection at >22 Hz.2) Simultaneous ddMSn ion tree with flexible combinations of CID and/or HCD

fragmentation with ion trap and/or OrbitrapTM detection. 2) Both the MS2 and MSn will be automatically searched in mzCloud, the

matching entry will be scored based on PIF, and the fragments will be annotated.

4) De novo structural identification is enabled by using this workflow.

OverviewPurpose: To demonstrate the capability of the Thermo Scientific™ Orbitrap Fusion™ Tribrid™ Mass Spectrometer in combination with MSn library for metabolite structural identification.

Methods: UHPLC coupled to the Orbitrap Fusion MS to simultaneously perform global profiling and structural elucidation with MSn.

Results: Within UHPLC timescale (2-3 second FWHM), >10 cycles can be finished with fast ion trap HCD MS2 for top 11 precursors, each with ion trees of MS2-MS3(top2)-MS4(top2).

IntroductionThe structural identification of metabolites represents a significant challenge in metabolomics study. Multistage mass spectrometry (MSn) is a powerful tool for compound identification and structural elucidation that goes beyond identifying the exact same compounds, to discovering additional compounds with structural homology to those in the library. Combining with high resolution accurate mass (HRAM) measurement, MSn is highly e ective in identifying the unknown but biologically relevant compounds in metabolomics studies[1]. However, the speed on current platforms is yet to be e ectively compatible with UHPLC front end, which limited such application in biological samples.

Presented here is a new platform to identify endogenous metabolites using a novel high performance Orbitrap hybrid instrument in conjunction with UHPLC and an MSn library.

Methods Sample Preparation

Urine samples from an adult male were analyzed. The urine samples were diluted 1 to 10 with water and processed with a 3kD MWCO filter.

For plasma samples, proteins were removed with cold methanol and centrifugation, followed by centrifugal evaporation at 35 oC. The residue was reconstituted in MeOH/Water, 1/9, and also processed with a 3kD MWCO filter.

Liquid ChromatographyUHPLC separation was implemented on a Dionex Ultimate 3000 HPG (high-pressure gradient) pump using Hypersil GOLD RP C18 column at 450 L/min, column temperature at 55 oC. LC solvents were 0.1% FA (A) and 0.1% FA in MeOH (B). A linear gradient was applied from 0.5-50% B for 5.5 min, followed by increasing to 98% at 6 min, held 98% B for 6 min, then decreased to 0.5% at 13min, and equilibrated for another 2 min.

FIGURE 1. Schematic of Orbitrap Fusion Tribrid Mass Spectrometer

A Platform to Identify Endogenous Metabolites Using a Novel High Performance Orbitrap MS and the mzCloud Library Junhua Wang1, David A. Peake1, Robert Mistrik2, Yingying Huang11Thermo Fisher Scientific Inc, San Jose, CA; 2HighChem, Ltd., Bratislava, Slovakia

Results

S-Lens

Active beam guide (ABG)

Ultra-high-field Orbitrap mass analyzer

Large-surface-Area detector (LSAD)

Ion-routing multipole

Dual-pressure linear ion trap

Quadrupole mass filter

C-Trap

EASY-Internal Calibration ion source

FIGURE 2. Ultrahigh resolution resolves the isotopes

PositiveMode20130518092831 #656 RT: 2.64 AV: 1 NL: 1.48E8T: FTMS + p ESI Full ms [100.00-212.08]

150.0 150.5 151.0 151.5 152.0m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

150.0584R=627006

151.0618R=621502

152.0542R=601802

PositiveMode20130518092831 #654 RT: 2.60 AV: 1 NL:T: FTMS + p ESI Full ms [100.00-212.08]

151.05 151.06m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

PositiveMode20130518092831 #539 RT: 1.44 AV: 1 NL:T: FTMS + p ESI Full ms [100.00-212.08]

151.05 151.06m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

PositiveMode20130518092831 #511 RT: 1.33 AV: 1 NL:T: FTMS + p ESI Full ms [100.00-212.08]

151.05 151.06m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

PositiveMode20130518092831 #615 RT: 1.97 AV: 1 NL:T: FTMS + p ESI Full ms [100.00-212.08]

151.05 151.06m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

PositiveMode20130518092831 #237 RT: 0.51 AV: 1 NL:T: FTMS + p ESI Full ms [100.00-212.08]

151.05 151.06m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

R=30k

R=60k

R=120k! =240kR=500k

L-Methionine C5H11NO2S

0.7ppm

Fine isotope pattern

C5H11NO2S +H: C5 H12 N1 O2 S1 p(gss, s/p:40) Chrg 1...

150.0 150.5 151.0 151.5 152.0m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

150.0583R=499553

151.0617R=499368

152.0541R=499443

PositiveSimulated

R=500,000

0.0ppm

C5H11NO2S +H: C5 H12 N1 O2 S1 p(gss, s/p:40) Chrg 1...

151.050 151.055 151.060 151.065m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

151.0617C4

13CH12O2NS

151.0625C5H12O17ONS

151.0646C5

13C14CH918O17O

151.0577C5H12O2N33S

151.55414C4H10

18O233S

PositiveObserved

Negative20130518111518 #1449 RT: 8.07 AV: 1 NL: 1.00E6T: FTMS - p ESI Full ms [140.00-220.00]

179 180 181 182m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

179.0576R=550402

180.0611R=330000

NegativeFull Scan

0.55 ppm

Negative20130518111518 #158 RT: 1.66 AV: 1 NL: 1.24E6T: FTMS - p ESI Full ms [140.00-220.00]

179.050 179.055 179.060 179.065m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

179.0573R=67606

Negative20130518111518 #110 RT: 1.34 AV: 1 NL: 8.48E5T: FTMS - p ESI Full ms [140.00-220.00]

179.050 179.055 179.060 179.065m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

179.0574R=148204

179.0558R=130000

Negative20130518111518 #54 RT: 0.83 AV: 1 NL: 9.54E5T: FTMS - p ESI Full ms [140.00-220.00]

179.050 179.055 179.060 179.065m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

179.0574R=304604

179.0560R=275000

Negative20130518111518 #21 RT: 0.40 AV: 1 NL: 7.38E5T: FTMS - p ESI Full ms [140.00-220.00]

179.050 179.055 179.060 179.065m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

179.0574R=607806

179.0560R=562802

R=60,000 R=120,000

R=240,000 R=500,000200 M

1 M

1.1 ppm

0.56 ppm

0.55 ppm

0.55 ppm

0.56 ppm

1.7 ppm

0.55 ppm

FIGURE 3. Ultrahigh resolution resolves isobaric metabolites

RT: 2.75160 - 2.86730

2.76 2.77 2.78 2.79 2.80 2.81 2.82 2.83 2.84 2.85 2.86Time (min)

0

10

20

30

40

50

60

70

80

90

1000

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

205.09721

205.09724

205.09721

205.09723

205.09717

205.09717205.09717205.09715 205.09718 205.09720

205.09749

205.09747

205.09741205.09746

205.09740205.09770

205.09772205.09746 205.09735 205.09738

NL:3.23E7Base Peak m/z= 205.09525-205.09935 F: ms MS JWUrine_FS240k_LM_Re

NL:2.85E7Base Peak m/z= 205.09525-205.09935 MS jwurine_fs240k_nolm_re

PBUrine_MS2_120k_600ms_bothLM #824 RT: 2.82 AV: 1 NL: 8.68E6F: FTMS + p ESI d Full ms2 [email protected] [50.00-216.00]

80 100 120 140 160 180 200m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

146.06033

118.06544

132.08105 188.07079170.06030without Int. Cal.Avg =1.5 ppm

Full Scan MS HCD MS/MSJWUrine_MS2_120k_600ms_bothLM #896 RT: 2.84 AV: 1 NL: 1.08E7T: FTMS + p ESI d Full ms2 [email protected] [50.00-216.00]

80 100 120 140 160 180 200m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

146.06020

118.06530

132.08090 170.06015 188.0706391.05422

0.2 ppmO

NH

NH

OHO

NH

0.9 ppm 0.7 ppmNH •

0.4 ppm

0.8 ppm

N2H

1 ppm

NH

N 2HNH

-0.1 ppm

with Int. Cal.Avg = 0.4 ppm

1.1 ppm

1.8 ppm

2.0 ppm

1.7 ppm

1.9ppm

2.9 ppm

1.1 ppm

with Int. Cal.

without Int. Cal.

Metabolite Identification Requires High Resolution and Isotope Ratio Measurements Structure elucidation of unknown small molecules by mass spectrometry is a great challenge. The first crucial step is to obtain correct elemental compositions.

Natural occurring elements can be monoisotopic (F, Na, P, I) or polyisotopic (H, C, N, O, S, Cl, Br) [1]. The molecule L-Methionine C5H11NO2S contains multiple polyisotopic elements, especially the S, which result in multiple isotopic peaks needing resolving power above 240,000 at m/z 200 (FWHM). Figure 2A shows the better isotope pattern of molecular ion [M+1+H]+. With increasing resolution up to 450,000 (FWHM), all five isotope peaks are clearly resolved. Figure 2B shows the simulated isotope pattern.

A

B

Metabolite Identification and Quantitation in Complex Samples Requires High ResolutionD-Glucose C6H12O6 and Paraxanthine C7H8N4O2 co-exist in human blood at distinct concentration levels: 5300 M (D-Glucose) vs. 10 M (Paraxanthine) [3]. Although D-Glucose is hundreds of fold more concentrated, it is less readily detectable because of its low ionization e ciency. Their masses di erentiate by 0.0013 amu (179.0561 vs. 179.0574), making the detection of these individual components very challenging. Figure 3 shows that, with >300,000 resolving power, the 200 M (D-Glucose) and 1 M (Paraxanthine) can be readily separated and quantified.

FIGURE 4. Easy Internal Calibration

Metabolite Identification Requires High Accuracy of Mass Measurement in Both MS Full Scan Level and MS/MS LevelFor an unknown metabolite, the number of calculated chemically possible formulae could be hundreds (20 ppm tolerance window), but the possibilities significantly reduce to tens (3 ppm) and several if the tolerance window is narrowed to sub-ppm. However, due to the large amount of known isomeric structures, one single formula may correspond to hundreds of structural isomers, thus structure identification of metabolite isomer using only MS full scan is not feasible. MS/MS and MSn techniques can help solve this analytical problem.

The Orbitrap Fusion MS has a built-in internal calibration functionality (Easy IC). Figure 4 shows LC-MS/MS data of L-Tryptophan from an urine sample with and without internal calibration. Sub-ppm mass accuracy was observed for MS and MS2 fragments with IC activated.

FIGURE 5. Rapid Ion Tree Acquisition on UHPLC Timescale from a Urine Sample (1 MS, 2 MS2 and 4 MS3 in 0.6sec/cycle).(A) Ion trap HCD MS2 followed by Orbitrap CID MS3 were acquired for the precursor ion m/z 181.0721. (B) Spectral tree search in mzCloud matches the MS3 with MS3 from theophylline with the highest score, indicating the unknown must have the same substructure as theophylline.

ConclusionThe Orbitrap Fusion MS o ers ultra high resolution and high fidelity in isotope ratio measurements of metabolites. This allows metabolites in complex mixture with a wide dynamic range be resolved and detected. The Easy-Internal Calibration allows compound assignment be readily done with confidence. The Orbitrap Fusion MS provides ultimate flexibility that facilitates the examination of multiple structural pathways for more structural information at higher speed.

The mzCloud library allows metabolite identification with MS2 and/or MSn data in an automated fashion. The identification based on multistage MS spectral trees and PIF algorithm opens the possibility to identify unknown metabolites and isomers. An improved metabolomics workflow is enabled.

References1. Budzikiewicz H, Grigsby RD, Mass Spectrom. Rev. 2006, 25(1):146-157.

2. Sheldon MT, Mistrik R, Croley TR, J. Am. Soc. Mass Spectrom. 2009, 20(3):370-376

3. http://www.hmdb.ca/metabolites/HMDB00122 & HMDB01860

A

BTheophylline

MS3

isomers

Query (down) and library spectra (up) matchScore

N

NH NH

N

NH

1.5 ppm 1.5 ppm

OH

NH

1 ppm

2.1 ppm

FIGURE 1. The Orbitrap Fusion MS and the Drag and Drop Instrument Method Editor.

O

N

NH NH

Page 4: A Platform to Identify Endogenous Metabolites …...A Platform to Identify Endogenous Metabolites Using a Novel High Performance Orbitrap MS and the mzCloud Library Junhua Wang, 1

4 A Platform to Identify Endogenous Metabolites Using a Novel High Performance Orbitrap MS and the mzCloud Library

60 80 100 120m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

96.0556

69.0447

RT: 3.1775 - 3.1907

3.178 3.180 3.182 3.184 3.186 3.188 3.190Time (min)

0

10

20

0

10

20

Rel

ativ

e A

bund

ance 0

10

20NL: 3.38E6Base Peak m/z= 181.0703-181.0739 MS b667JWUrine_HCDIT_CIDOTtop2_500ms

NL: 1.59E5Base Peak F: ms2 MS b667JWUrine_HCDIT_CIDOTtop2_500ms

NL: 2.93E4Base Peak F: ms3 MS b667JWUrine_HCDIT_CIDOTtop2_500ms

3.1791 3.1891Full Scan MS

MS2

MS3

0.6 sec

HCDIT-MS2

60 80 100 120 140 160 180m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

124.0153

105.0077

180.9644141.9145

95.077369.0120

CIDOT-MS3

60 80 100 120 140 160 180m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

124.0153

105.0077

180.9644141.9145

95.077369.0120

An Improved Metabolomics WorkflowA traditional workflow for metabolite profiling and identification is based on a full scan LC/MS experiment followed by targeted MS/MS confirmation.

The major problems associated with this methodology are:1) Instrumentation limitations

• Does not allow simultaneous profiling and MSn in a single run, the scan speed su ers.

• Identification is usually based on accurate mass and MS2, but a large number of LC peaks remain not identifiable .

2) MS/MS library limitations• MS/MS spectral search is usually manually performed and time-

consuming; • Automated scoring for spectral matching is usually not available;• The fragmentation needs to be manually interpreted, which is not

only time-consuming but also prone to errors.

3) There is no good solution for unknowns.

The Orbitrap Fusion MS enables fast parallel data acquisition on a UHPLC timescale with a variety of dissociation choices The Orbitrap Fusion MS o ers ultimate flexibility in fragmentation methods at all stages of MSn analysis, with fragment ions detected by either mass analyzer. CID-HCD, HCD-CID, and HCD-HCD can all be set up as simple as the conventional CID-CID. This complete flexibility facilitates the examination of fragmentation pathways that leads to the most comprehensive structural information.

mzCloud MSn library enables MSn search for both high and low resolution mzCloud is a multistage MS library. Based on the spectral tree similarity, an algorithm called precursor ion fingerprinting (PIF) enables the identification of compounds even if they are not in the library, and the fragmental peaks can be annotated [2]. Powered by the novel MS and the mzCloud library, an improved, more connected workflow is shown below:

Major benefits of the improved workflow: 1) HCD MS/MS can be performed simultaneously with full scan MS on the

UHPLC timescale with ion trap detection at >22 Hz.2) Simultaneous ddMSn ion tree with flexible combinations of CID and/or HCD

fragmentation with ion trap and/or OrbitrapTM detection. 2) Both the MS2 and MSn will be automatically searched in mzCloud, the

matching entry will be scored based on PIF, and the fragments will be annotated.

4) De novo structural identification is enabled by using this workflow.

OverviewPurpose: To demonstrate the capability of the Thermo Scientific™ Orbitrap Fusion™ Tribrid™ Mass Spectrometer in combination with MSn library for metabolite structural identification.

Methods: UHPLC coupled to the Orbitrap Fusion MS to simultaneously perform global profiling and structural elucidation with MSn.

Results: Within UHPLC timescale (2-3 second FWHM), >10 cycles can be finished with fast ion trap HCD MS2 for top 11 precursors, each with ion trees of MS2-MS3(top2)-MS4(top2).

IntroductionThe structural identification of metabolites represents a significant challenge in metabolomics study. Multistage mass spectrometry (MSn) is a powerful tool for compound identification and structural elucidation that goes beyond identifying the exact same compounds, to discovering additional compounds with structural homology to those in the library. Combining with high resolution accurate mass (HRAM) measurement, MSn is highly e ective in identifying the unknown but biologically relevant compounds in metabolomics studies[1]. However, the speed on current platforms is yet to be e ectively compatible with UHPLC front end, which limited such application in biological samples.

Presented here is a new platform to identify endogenous metabolites using a novel high performance Orbitrap hybrid instrument in conjunction with UHPLC and an MSn library.

Methods Sample Preparation

Urine samples from an adult male were analyzed. The urine samples were diluted 1 to 10 with water and processed with a 3kD MWCO filter.

For plasma samples, proteins were removed with cold methanol and centrifugation, followed by centrifugal evaporation at 35 oC. The residue was reconstituted in MeOH/Water, 1/9, and also processed with a 3kD MWCO filter.

Liquid ChromatographyUHPLC separation was implemented on a Dionex Ultimate 3000 HPG (high-pressure gradient) pump using Hypersil GOLD RP C18 column at 450 L/min, column temperature at 55 oC. LC solvents were 0.1% FA (A) and 0.1% FA in MeOH (B). A linear gradient was applied from 0.5-50% B for 5.5 min, followed by increasing to 98% at 6 min, held 98% B for 6 min, then decreased to 0.5% at 13min, and equilibrated for another 2 min.

FIGURE 1. Schematic of Orbitrap Fusion Tribrid Mass Spectrometer

A Platform to Identify Endogenous Metabolites Using a Novel High Performance Orbitrap MS and the mzCloud Library Junhua Wang1, David A. Peake1, Robert Mistrik2, Yingying Huang11Thermo Fisher Scientific Inc, San Jose, CA; 2HighChem, Ltd., Bratislava, Slovakia

Results

S-Lens

Active beam guide (ABG)

Ultra-high-field Orbitrap mass analyzer

Large-surface-Area detector (LSAD)

Ion-routing multipole

Dual-pressure linear ion trap

Quadrupole mass filter

C-Trap

EASY-Internal Calibration ion source

FIGURE 2. Ultrahigh resolution resolves the isotopes

PositiveMode20130518092831 #656 RT: 2.64 AV: 1 NL: 1.48E8T: FTMS + p ESI Full ms [100.00-212.08]

150.0 150.5 151.0 151.5 152.0m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

150.0584R=627006

151.0618R=621502

152.0542R=601802

PositiveMode20130518092831 #654 RT: 2.60 AV: 1 NL:T: FTMS + p ESI Full ms [100.00-212.08]

151.05 151.06m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

PositiveMode20130518092831 #539 RT: 1.44 AV: 1 NL:T: FTMS + p ESI Full ms [100.00-212.08]

151.05 151.06m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

PositiveMode20130518092831 #511 RT: 1.33 AV: 1 NL:T: FTMS + p ESI Full ms [100.00-212.08]

151.05 151.06m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

PositiveMode20130518092831 #615 RT: 1.97 AV: 1 NL:T: FTMS + p ESI Full ms [100.00-212.08]

151.05 151.06m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

PositiveMode20130518092831 #237 RT: 0.51 AV: 1 NL:T: FTMS + p ESI Full ms [100.00-212.08]

151.05 151.06m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

R=30k

R=60k

R=120k! =240kR=500k

L-Methionine C5H11NO2S

0.7ppm

Fine isotope pattern

C5H11NO2S +H: C5 H12 N1 O2 S1 p(gss, s/p:40) Chrg 1...

150.0 150.5 151.0 151.5 152.0m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

150.0583R=499553

151.0617R=499368

152.0541R=499443

PositiveSimulated

R=500,000

0.0ppm

C5H11NO2S +H: C5 H12 N1 O2 S1 p(gss, s/p:40) Chrg 1...

151.050 151.055 151.060 151.065m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

151.0617C4

13CH12O2NS

151.0625C5H12O17ONS

151.0646C5

13C14CH918O17O

151.0577C5H12O2N33S

151.55414C4H10

18O233S

PositiveObserved

Negative20130518111518 #1449 RT: 8.07 AV: 1 NL: 1.00E6T: FTMS - p ESI Full ms [140.00-220.00]

179 180 181 182m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

179.0576R=550402

180.0611R=330000

NegativeFull Scan

0.55 ppm

Negative20130518111518 #158 RT: 1.66 AV: 1 NL: 1.24E6T: FTMS - p ESI Full ms [140.00-220.00]

179.050 179.055 179.060 179.065m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

179.0573R=67606

Negative20130518111518 #110 RT: 1.34 AV: 1 NL: 8.48E5T: FTMS - p ESI Full ms [140.00-220.00]

179.050 179.055 179.060 179.065m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

179.0574R=148204

179.0558R=130000

Negative20130518111518 #54 RT: 0.83 AV: 1 NL: 9.54E5T: FTMS - p ESI Full ms [140.00-220.00]

179.050 179.055 179.060 179.065m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

179.0574R=304604

179.0560R=275000

Negative20130518111518 #21 RT: 0.40 AV: 1 NL: 7.38E5T: FTMS - p ESI Full ms [140.00-220.00]

179.050 179.055 179.060 179.065m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

179.0574R=607806

179.0560R=562802

R=60,000 R=120,000

R=240,000 R=500,000200 M

1 M

1.1 ppm

0.56 ppm

0.55 ppm

0.55 ppm

0.56 ppm

1.7 ppm

0.55 ppm

FIGURE 3. Ultrahigh resolution resolves isobaric metabolites

RT: 2.75160 - 2.86730

2.76 2.77 2.78 2.79 2.80 2.81 2.82 2.83 2.84 2.85 2.86Time (min)

0

10

20

30

40

50

60

70

80

90

1000

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

205.09721

205.09724

205.09721

205.09723

205.09717

205.09717205.09717205.09715 205.09718 205.09720

205.09749

205.09747

205.09741205.09746

205.09740205.09770

205.09772205.09746 205.09735 205.09738

NL:3.23E7Base Peak m/z= 205.09525-205.09935 F: ms MS JWUrine_FS240k_LM_Re

NL:2.85E7Base Peak m/z= 205.09525-205.09935 MS jwurine_fs240k_nolm_re

PBUrine_MS2_120k_600ms_bothLM #824 RT: 2.82 AV: 1 NL: 8.68E6F: FTMS + p ESI d Full ms2 [email protected] [50.00-216.00]

80 100 120 140 160 180 200m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

146.06033

118.06544

132.08105 188.07079170.06030without Int. Cal.Avg =1.5 ppm

Full Scan MS HCD MS/MSJWUrine_MS2_120k_600ms_bothLM #896 RT: 2.84 AV: 1 NL: 1.08E7T: FTMS + p ESI d Full ms2 [email protected] [50.00-216.00]

80 100 120 140 160 180 200m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

146.06020

118.06530

132.08090 170.06015 188.0706391.05422

0.2 ppmO

NH

NH

OHO

NH

0.9 ppm 0.7 ppmNH •

0.4 ppm

0.8 ppm

N2H

1 ppm

NH

N 2HNH

-0.1 ppm

with Int. Cal.Avg = 0.4 ppm

1.1 ppm

1.8 ppm

2.0 ppm

1.7 ppm

1.9ppm

2.9 ppm

1.1 ppm

with Int. Cal.

without Int. Cal.

Metabolite Identification Requires High Resolution and Isotope Ratio Measurements Structure elucidation of unknown small molecules by mass spectrometry is a great challenge. The first crucial step is to obtain correct elemental compositions.

Natural occurring elements can be monoisotopic (F, Na, P, I) or polyisotopic (H, C, N, O, S, Cl, Br) [1]. The molecule L-Methionine C5H11NO2S contains multiple polyisotopic elements, especially the S, which result in multiple isotopic peaks needing resolving power above 240,000 at m/z 200 (FWHM). Figure 2A shows the better isotope pattern of molecular ion [M+1+H]+. With increasing resolution up to 450,000 (FWHM), all five isotope peaks are clearly resolved. Figure 2B shows the simulated isotope pattern.

A

B

Metabolite Identification and Quantitation in Complex Samples Requires High ResolutionD-Glucose C6H12O6 and Paraxanthine C7H8N4O2 co-exist in human blood at distinct concentration levels: 5300 M (D-Glucose) vs. 10 M (Paraxanthine) [3]. Although D-Glucose is hundreds of fold more concentrated, it is less readily detectable because of its low ionization e ciency. Their masses di erentiate by 0.0013 amu (179.0561 vs. 179.0574), making the detection of these individual components very challenging. Figure 3 shows that, with >300,000 resolving power, the 200 M (D-Glucose) and 1 M (Paraxanthine) can be readily separated and quantified.

FIGURE 4. Easy Internal Calibration

Metabolite Identification Requires High Accuracy of Mass Measurement in Both MS Full Scan Level and MS/MS LevelFor an unknown metabolite, the number of calculated chemically possible formulae could be hundreds (20 ppm tolerance window), but the possibilities significantly reduce to tens (3 ppm) and several if the tolerance window is narrowed to sub-ppm. However, due to the large amount of known isomeric structures, one single formula may correspond to hundreds of structural isomers, thus structure identification of metabolite isomer using only MS full scan is not feasible. MS/MS and MSn techniques can help solve this analytical problem.

The Orbitrap Fusion MS has a built-in internal calibration functionality (Easy IC). Figure 4 shows LC-MS/MS data of L-Tryptophan from an urine sample with and without internal calibration. Sub-ppm mass accuracy was observed for MS and MS2 fragments with IC activated.

FIGURE 5. Rapid Ion Tree Acquisition on UHPLC Timescale from a Urine Sample (1 MS, 2 MS2 and 4 MS3 in 0.6sec/cycle).(A) Ion trap HCD MS2 followed by Orbitrap CID MS3 were acquired for the precursor ion m/z 181.0721. (B) Spectral tree search in mzCloud matches the MS3 with MS3 from theophylline with the highest score, indicating the unknown must have the same substructure as theophylline.

ConclusionThe Orbitrap Fusion MS o ers ultra high resolution and high fidelity in isotope ratio measurements of metabolites. This allows metabolites in complex mixture with a wide dynamic range be resolved and detected. The Easy-Internal Calibration allows compound assignment be readily done with confidence. The Orbitrap Fusion MS provides ultimate flexibility that facilitates the examination of multiple structural pathways for more structural information at higher speed.

The mzCloud library allows metabolite identification with MS2 and/or MSn data in an automated fashion. The identification based on multistage MS spectral trees and PIF algorithm opens the possibility to identify unknown metabolites and isomers. An improved metabolomics workflow is enabled.

References1. Budzikiewicz H, Grigsby RD, Mass Spectrom. Rev. 2006, 25(1):146-157.

2. Sheldon MT, Mistrik R, Croley TR, J. Am. Soc. Mass Spectrom. 2009, 20(3):370-376

3. http://www.hmdb.ca/metabolites/HMDB00122 & HMDB01860

A

BTheophylline

MS3

isomers

Query (down) and library spectra (up) matchScore

N

NH NH

N

NH

1.5 ppm 1.5 ppm

OH

NH

1 ppm

2.1 ppm

FIGURE 1. The Orbitrap Fusion MS and the Drag and Drop Instrument Method Editor.

O

N

NH NH

Page 5: A Platform to Identify Endogenous Metabolites …...A Platform to Identify Endogenous Metabolites Using a Novel High Performance Orbitrap MS and the mzCloud Library Junhua Wang, 1

5Thermo Scientific Poster Note • PN ASMS13_MP 045_JWang e 06/13S

60 80 100 120m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

96.0556

69.0447

RT: 3.1775 - 3.1907

3.178 3.180 3.182 3.184 3.186 3.188 3.190Time (min)

0

10

20

0

10

20

Rel

ativ

e A

bund

ance 0

10

20NL: 3.38E6Base Peak m/z= 181.0703-181.0739 MS b667JWUrine_HCDIT_CIDOTtop2_500ms

NL: 1.59E5Base Peak F: ms2 MS b667JWUrine_HCDIT_CIDOTtop2_500ms

NL: 2.93E4Base Peak F: ms3 MS b667JWUrine_HCDIT_CIDOTtop2_500ms

3.1791 3.1891Full Scan MS

MS2

MS3

0.6 sec

HCDIT-MS2

60 80 100 120 140 160 180m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

124.0153

105.0077

180.9644141.9145

95.077369.0120

CIDOT-MS3

60 80 100 120 140 160 180m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

124.0153

105.0077

180.9644141.9145

95.077369.0120

An Improved Metabolomics WorkflowA traditional workflow for metabolite profiling and identification is based on a full scan LC/MS experiment followed by targeted MS/MS confirmation.

The major problems associated with this methodology are:1) Instrumentation limitations

• Does not allow simultaneous profiling and MSn in a single run, the scan speed su ers.

• Identification is usually based on accurate mass and MS2, but a large number of LC peaks remain not identifiable .

2) MS/MS library limitations• MS/MS spectral search is usually manually performed and time-

consuming; • Automated scoring for spectral matching is usually not available;• The fragmentation needs to be manually interpreted, which is not

only time-consuming but also prone to errors.

3) There is no good solution for unknowns.

The Orbitrap Fusion MS enables fast parallel data acquisition on a UHPLC timescale with a variety of dissociation choices The Orbitrap Fusion MS o ers ultimate flexibility in fragmentation methods at all stages of MSn analysis, with fragment ions detected by either mass analyzer. CID-HCD, HCD-CID, and HCD-HCD can all be set up as simple as the conventional CID-CID. This complete flexibility facilitates the examination of fragmentation pathways that leads to the most comprehensive structural information.

mzCloud MSn library enables MSn search for both high and low resolution mzCloud is a multistage MS library. Based on the spectral tree similarity, an algorithm called precursor ion fingerprinting (PIF) enables the identification of compounds even if they are not in the library, and the fragmental peaks can be annotated [2]. Powered by the novel MS and the mzCloud library, an improved, more connected workflow is shown below:

Major benefits of the improved workflow: 1) HCD MS/MS can be performed simultaneously with full scan MS on the

UHPLC timescale with ion trap detection at >22 Hz.2) Simultaneous ddMSn ion tree with flexible combinations of CID and/or HCD

fragmentation with ion trap and/or OrbitrapTM detection. 2) Both the MS2 and MSn will be automatically searched in mzCloud, the

matching entry will be scored based on PIF, and the fragments will be annotated.

4) De novo structural identification is enabled by using this workflow.

OverviewPurpose: To demonstrate the capability of the Thermo Scientific™ Orbitrap Fusion™ Tribrid™ Mass Spectrometer in combination with MSn library for metabolite structural identification.

Methods: UHPLC coupled to the Orbitrap Fusion MS to simultaneously perform global profiling and structural elucidation with MSn.

Results: Within UHPLC timescale (2-3 second FWHM), >10 cycles can be finished with fast ion trap HCD MS2 for top 11 precursors, each with ion trees of MS2-MS3(top2)-MS4(top2).

IntroductionThe structural identification of metabolites represents a significant challenge in metabolomics study. Multistage mass spectrometry (MSn) is a powerful tool for compound identification and structural elucidation that goes beyond identifying the exact same compounds, to discovering additional compounds with structural homology to those in the library. Combining with high resolution accurate mass (HRAM) measurement, MSn is highly e ective in identifying the unknown but biologically relevant compounds in metabolomics studies[1]. However, the speed on current platforms is yet to be e ectively compatible with UHPLC front end, which limited such application in biological samples.

Presented here is a new platform to identify endogenous metabolites using a novel high performance Orbitrap hybrid instrument in conjunction with UHPLC and an MSn library.

Methods Sample Preparation

Urine samples from an adult male were analyzed. The urine samples were diluted 1 to 10 with water and processed with a 3kD MWCO filter.

For plasma samples, proteins were removed with cold methanol and centrifugation, followed by centrifugal evaporation at 35 oC. The residue was reconstituted in MeOH/Water, 1/9, and also processed with a 3kD MWCO filter.

Liquid ChromatographyUHPLC separation was implemented on a Dionex Ultimate 3000 HPG (high-pressure gradient) pump using Hypersil GOLD RP C18 column at 450 L/min, column temperature at 55 oC. LC solvents were 0.1% FA (A) and 0.1% FA in MeOH (B). A linear gradient was applied from 0.5-50% B for 5.5 min, followed by increasing to 98% at 6 min, held 98% B for 6 min, then decreased to 0.5% at 13min, and equilibrated for another 2 min.

FIGURE 1. Schematic of Orbitrap Fusion Tribrid Mass Spectrometer

A Platform to Identify Endogenous Metabolites Using a Novel High Performance Orbitrap MS and the mzCloud Library Junhua Wang1, David A. Peake1, Robert Mistrik2, Yingying Huang11Thermo Fisher Scientific Inc, San Jose, CA; 2HighChem, Ltd., Bratislava, Slovakia

Results

S-Lens

Active beam guide (ABG)

Ultra-high-field Orbitrap mass analyzer

Large-surface-Area detector (LSAD)

Ion-routing multipole

Dual-pressure linear ion trap

Quadrupole mass filter

C-Trap

EASY-Internal Calibration ion source

FIGURE 2. Ultrahigh resolution resolves the isotopes

PositiveMode20130518092831 #656 RT: 2.64 AV: 1 NL: 1.48E8T: FTMS + p ESI Full ms [100.00-212.08]

150.0 150.5 151.0 151.5 152.0m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

150.0584R=627006

151.0618R=621502

152.0542R=601802

PositiveMode20130518092831 #654 RT: 2.60 AV: 1 NL:T: FTMS + p ESI Full ms [100.00-212.08]

151.05 151.06m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

PositiveMode20130518092831 #539 RT: 1.44 AV: 1 NL:T: FTMS + p ESI Full ms [100.00-212.08]

151.05 151.06m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

PositiveMode20130518092831 #511 RT: 1.33 AV: 1 NL:T: FTMS + p ESI Full ms [100.00-212.08]

151.05 151.06m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

PositiveMode20130518092831 #615 RT: 1.97 AV: 1 NL:T: FTMS + p ESI Full ms [100.00-212.08]

151.05 151.06m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

PositiveMode20130518092831 #237 RT: 0.51 AV: 1 NL:T: FTMS + p ESI Full ms [100.00-212.08]

151.05 151.06m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

R=30k

R=60k

R=120k! =240kR=500k

L-Methionine C5H11NO2S

0.7ppm

Fine isotope pattern

C5H11NO2S +H: C5 H12 N1 O2 S1 p(gss, s/p:40) Chrg 1...

150.0 150.5 151.0 151.5 152.0m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

150.0583R=499553

151.0617R=499368

152.0541R=499443

PositiveSimulated

R=500,000

0.0ppm

C5H11NO2S +H: C5 H12 N1 O2 S1 p(gss, s/p:40) Chrg 1...

151.050 151.055 151.060 151.065m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

151.0617C4

13CH12O2NS

151.0625C5H12O17ONS

151.0646C5

13C14CH918O17O

151.0577C5H12O2N33S

151.55414C4H10

18O233S

PositiveObserved

Negative20130518111518 #1449 RT: 8.07 AV: 1 NL: 1.00E6T: FTMS - p ESI Full ms [140.00-220.00]

179 180 181 182m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

179.0576R=550402

180.0611R=330000

NegativeFull Scan

0.55 ppm

Negative20130518111518 #158 RT: 1.66 AV: 1 NL: 1.24E6T: FTMS - p ESI Full ms [140.00-220.00]

179.050 179.055 179.060 179.065m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

179.0573R=67606

Negative20130518111518 #110 RT: 1.34 AV: 1 NL: 8.48E5T: FTMS - p ESI Full ms [140.00-220.00]

179.050 179.055 179.060 179.065m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

179.0574R=148204

179.0558R=130000

Negative20130518111518 #54 RT: 0.83 AV: 1 NL: 9.54E5T: FTMS - p ESI Full ms [140.00-220.00]

179.050 179.055 179.060 179.065m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

179.0574R=304604

179.0560R=275000

Negative20130518111518 #21 RT: 0.40 AV: 1 NL: 7.38E5T: FTMS - p ESI Full ms [140.00-220.00]

179.050 179.055 179.060 179.065m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

179.0574R=607806

179.0560R=562802

R=60,000 R=120,000

R=240,000 R=500,000200 M

1 M

1.1 ppm

0.56 ppm

0.55 ppm

0.55 ppm

0.56 ppm

1.7 ppm

0.55 ppm

FIGURE 3. Ultrahigh resolution resolves isobaric metabolites

RT: 2.75160 - 2.86730

2.76 2.77 2.78 2.79 2.80 2.81 2.82 2.83 2.84 2.85 2.86Time (min)

0

10

20

30

40

50

60

70

80

90

1000

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

205.09721

205.09724

205.09721

205.09723

205.09717

205.09717205.09717205.09715 205.09718 205.09720

205.09749

205.09747

205.09741205.09746

205.09740205.09770

205.09772205.09746 205.09735 205.09738

NL:3.23E7Base Peak m/z= 205.09525-205.09935 F: ms MS JWUrine_FS240k_LM_Re

NL:2.85E7Base Peak m/z= 205.09525-205.09935 MS jwurine_fs240k_nolm_re

PBUrine_MS2_120k_600ms_bothLM #824 RT: 2.82 AV: 1 NL: 8.68E6F: FTMS + p ESI d Full ms2 [email protected] [50.00-216.00]

80 100 120 140 160 180 200m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

146.06033

118.06544

132.08105 188.07079170.06030without Int. Cal.Avg =1.5 ppm

Full Scan MS HCD MS/MSJWUrine_MS2_120k_600ms_bothLM #896 RT: 2.84 AV: 1 NL: 1.08E7T: FTMS + p ESI d Full ms2 [email protected] [50.00-216.00]

80 100 120 140 160 180 200m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

146.06020

118.06530

132.08090 170.06015 188.0706391.05422

0.2 ppmO

NH

NH

OHO

NH

0.9 ppm 0.7 ppmNH •

0.4 ppm

0.8 ppm

N2H

1 ppm

NH

N 2HNH

-0.1 ppm

with Int. Cal.Avg = 0.4 ppm

1.1 ppm

1.8 ppm

2.0 ppm

1.7 ppm

1.9ppm

2.9 ppm

1.1 ppm

with Int. Cal.

without Int. Cal.

Metabolite Identification Requires High Resolution and Isotope Ratio Measurements Structure elucidation of unknown small molecules by mass spectrometry is a great challenge. The first crucial step is to obtain correct elemental compositions.

Natural occurring elements can be monoisotopic (F, Na, P, I) or polyisotopic (H, C, N, O, S, Cl, Br) [1]. The molecule L-Methionine C5H11NO2S contains multiple polyisotopic elements, especially the S, which result in multiple isotopic peaks needing resolving power above 240,000 at m/z 200 (FWHM). Figure 2A shows the better isotope pattern of molecular ion [M+1+H]+. With increasing resolution up to 450,000 (FWHM), all five isotope peaks are clearly resolved. Figure 2B shows the simulated isotope pattern.

A

B

Metabolite Identification and Quantitation in Complex Samples Requires High ResolutionD-Glucose C6H12O6 and Paraxanthine C7H8N4O2 co-exist in human blood at distinct concentration levels: 5300 M (D-Glucose) vs. 10 M (Paraxanthine) [3]. Although D-Glucose is hundreds of fold more concentrated, it is less readily detectable because of its low ionization e ciency. Their masses di erentiate by 0.0013 amu (179.0561 vs. 179.0574), making the detection of these individual components very challenging. Figure 3 shows that, with >300,000 resolving power, the 200 M (D-Glucose) and 1 M (Paraxanthine) can be readily separated and quantified.

FIGURE 4. Easy Internal Calibration

Metabolite Identification Requires High Accuracy of Mass Measurement in Both MS Full Scan Level and MS/MS LevelFor an unknown metabolite, the number of calculated chemically possible formulae could be hundreds (20 ppm tolerance window), but the possibilities significantly reduce to tens (3 ppm) and several if the tolerance window is narrowed to sub-ppm. However, due to the large amount of known isomeric structures, one single formula may correspond to hundreds of structural isomers, thus structure identification of metabolite isomer using only MS full scan is not feasible. MS/MS and MSn techniques can help solve this analytical problem.

The Orbitrap Fusion MS has a built-in internal calibration functionality (Easy IC). Figure 4 shows LC-MS/MS data of L-Tryptophan from an urine sample with and without internal calibration. Sub-ppm mass accuracy was observed for MS and MS2 fragments with IC activated.

FIGURE 5. Rapid Ion Tree Acquisition on UHPLC Timescale from a Urine Sample (1 MS, 2 MS2 and 4 MS3 in 0.6sec/cycle).(A) Ion trap HCD MS2 followed by Orbitrap CID MS3 were acquired for the precursor ion m/z 181.0721. (B) Spectral tree search in mzCloud matches the MS3 with MS3 from theophylline with the highest score, indicating the unknown must have the same substructure as theophylline.

ConclusionThe Orbitrap Fusion MS o ers ultra high resolution and high fidelity in isotope ratio measurements of metabolites. This allows metabolites in complex mixture with a wide dynamic range be resolved and detected. The Easy-Internal Calibration allows compound assignment be readily done with confidence. The Orbitrap Fusion MS provides ultimate flexibility that facilitates the examination of multiple structural pathways for more structural information at higher speed.

The mzCloud library allows metabolite identification with MS2 and/or MSn data in an automated fashion. The identification based on multistage MS spectral trees and PIF algorithm opens the possibility to identify unknown metabolites and isomers. An improved metabolomics workflow is enabled.

References1. Budzikiewicz H, Grigsby RD, Mass Spectrom. Rev. 2006, 25(1):146-157.

2. Sheldon MT, Mistrik R, Croley TR, J. Am. Soc. Mass Spectrom. 2009, 20(3):370-376

3. http://www.hmdb.ca/metabolites/HMDB00122 & HMDB01860

A

BTheophylline

MS3

isomers

Query (down) and library spectra (up) matchScore

N

NH NH

N

NH

1.5 ppm 1.5 ppm

OH

NH

1 ppm

2.1 ppm

FIGURE 1. The Orbitrap Fusion MS and the Drag and Drop Instrument Method Editor.

O

N

NH NH

Page 6: A Platform to Identify Endogenous Metabolites …...A Platform to Identify Endogenous Metabolites Using a Novel High Performance Orbitrap MS and the mzCloud Library Junhua Wang, 1

6 A Platform to Identify Endogenous Metabolites Using a Novel High Performance Orbitrap MS and the mzCloud Library

60 80 100 120m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

96.0556

69.0447

RT: 3.1775 - 3.1907

3.178 3.180 3.182 3.184 3.186 3.188 3.190Time (min)

0

10

20

0

10

20

Rel

ativ

e A

bund

ance 0

10

20NL: 3.38E6Base Peak m/z= 181.0703-181.0739 MS b667JWUrine_HCDIT_CIDOTtop2_500ms

NL: 1.59E5Base Peak F: ms2 MS b667JWUrine_HCDIT_CIDOTtop2_500ms

NL: 2.93E4Base Peak F: ms3 MS b667JWUrine_HCDIT_CIDOTtop2_500ms

3.1791 3.1891Full Scan MS

MS2

MS3

0.6 sec

HCDIT-MS2

60 80 100 120 140 160 180m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

124.0153

105.0077

180.9644141.9145

95.077369.0120

CIDOT-MS3

60 80 100 120 140 160 180m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

124.0153

105.0077

180.9644141.9145

95.077369.0120

An Improved Metabolomics WorkflowA traditional workflow for metabolite profiling and identification is based on a full scan LC/MS experiment followed by targeted MS/MS confirmation.

The major problems associated with this methodology are:1) Instrumentation limitations

• Does not allow simultaneous profiling and MSn in a single run, the scan speed su ers.

• Identification is usually based on accurate mass and MS2, but a large number of LC peaks remain not identifiable .

2) MS/MS library limitations• MS/MS spectral search is usually manually performed and time-

consuming; • Automated scoring for spectral matching is usually not available;• The fragmentation needs to be manually interpreted, which is not

only time-consuming but also prone to errors.

3) There is no good solution for unknowns.

The Orbitrap Fusion MS enables fast parallel data acquisition on a UHPLC timescale with a variety of dissociation choices The Orbitrap Fusion MS o ers ultimate flexibility in fragmentation methods at all stages of MSn analysis, with fragment ions detected by either mass analyzer. CID-HCD, HCD-CID, and HCD-HCD can all be set up as simple as the conventional CID-CID. This complete flexibility facilitates the examination of fragmentation pathways that leads to the most comprehensive structural information.

mzCloud MSn library enables MSn search for both high and low resolution mzCloud is a multistage MS library. Based on the spectral tree similarity, an algorithm called precursor ion fingerprinting (PIF) enables the identification of compounds even if they are not in the library, and the fragmental peaks can be annotated [2]. Powered by the novel MS and the mzCloud library, an improved, more connected workflow is shown below:

Major benefits of the improved workflow: 1) HCD MS/MS can be performed simultaneously with full scan MS on the

UHPLC timescale with ion trap detection at >22 Hz.2) Simultaneous ddMSn ion tree with flexible combinations of CID and/or HCD

fragmentation with ion trap and/or OrbitrapTM detection. 2) Both the MS2 and MSn will be automatically searched in mzCloud, the

matching entry will be scored based on PIF, and the fragments will be annotated.

4) De novo structural identification is enabled by using this workflow.

OverviewPurpose: To demonstrate the capability of the Thermo Scientific™ Orbitrap Fusion™ Tribrid™ Mass Spectrometer in combination with MSn library for metabolite structural identification.

Methods: UHPLC coupled to the Orbitrap Fusion MS to simultaneously perform global profiling and structural elucidation with MSn.

Results: Within UHPLC timescale (2-3 second FWHM), >10 cycles can be finished with fast ion trap HCD MS2 for top 11 precursors, each with ion trees of MS2-MS3(top2)-MS4(top2).

IntroductionThe structural identification of metabolites represents a significant challenge in metabolomics study. Multistage mass spectrometry (MSn) is a powerful tool for compound identification and structural elucidation that goes beyond identifying the exact same compounds, to discovering additional compounds with structural homology to those in the library. Combining with high resolution accurate mass (HRAM) measurement, MSn is highly e ective in identifying the unknown but biologically relevant compounds in metabolomics studies[1]. However, the speed on current platforms is yet to be e ectively compatible with UHPLC front end, which limited such application in biological samples.

Presented here is a new platform to identify endogenous metabolites using a novel high performance Orbitrap hybrid instrument in conjunction with UHPLC and an MSn library.

Methods Sample Preparation

Urine samples from an adult male were analyzed. The urine samples were diluted 1 to 10 with water and processed with a 3kD MWCO filter.

For plasma samples, proteins were removed with cold methanol and centrifugation, followed by centrifugal evaporation at 35 oC. The residue was reconstituted in MeOH/Water, 1/9, and also processed with a 3kD MWCO filter.

Liquid ChromatographyUHPLC separation was implemented on a Dionex Ultimate 3000 HPG (high-pressure gradient) pump using Hypersil GOLD RP C18 column at 450 L/min, column temperature at 55 oC. LC solvents were 0.1% FA (A) and 0.1% FA in MeOH (B). A linear gradient was applied from 0.5-50% B for 5.5 min, followed by increasing to 98% at 6 min, held 98% B for 6 min, then decreased to 0.5% at 13min, and equilibrated for another 2 min.

FIGURE 1. Schematic of Orbitrap Fusion Tribrid Mass Spectrometer

A Platform to Identify Endogenous Metabolites Using a Novel High Performance Orbitrap MS and the mzCloud Library Junhua Wang1, David A. Peake1, Robert Mistrik2, Yingying Huang11Thermo Fisher Scientific Inc, San Jose, CA; 2HighChem, Ltd., Bratislava, Slovakia

Results

S-Lens

Active beam guide (ABG)

Ultra-high-field Orbitrap mass analyzer

Large-surface-Area detector (LSAD)

Ion-routing multipole

Dual-pressure linear ion trap

Quadrupole mass filter

C-Trap

EASY-Internal Calibration ion source

FIGURE 2. Ultrahigh resolution resolves the isotopes

PositiveMode20130518092831 #656 RT: 2.64 AV: 1 NL: 1.48E8T: FTMS + p ESI Full ms [100.00-212.08]

150.0 150.5 151.0 151.5 152.0m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

150.0584R=627006

151.0618R=621502

152.0542R=601802

PositiveMode20130518092831 #654 RT: 2.60 AV: 1 NL:T: FTMS + p ESI Full ms [100.00-212.08]

151.05 151.06m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

PositiveMode20130518092831 #539 RT: 1.44 AV: 1 NL:T: FTMS + p ESI Full ms [100.00-212.08]

151.05 151.06m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

PositiveMode20130518092831 #511 RT: 1.33 AV: 1 NL:T: FTMS + p ESI Full ms [100.00-212.08]

151.05 151.06m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

PositiveMode20130518092831 #615 RT: 1.97 AV: 1 NL:T: FTMS + p ESI Full ms [100.00-212.08]

151.05 151.06m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

PositiveMode20130518092831 #237 RT: 0.51 AV: 1 NL:T: FTMS + p ESI Full ms [100.00-212.08]

151.05 151.06m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

R=30k

R=60k

R=120k! =240kR=500k

L-Methionine C5H11NO2S

0.7ppm

Fine isotope pattern

C5H11NO2S +H: C5 H12 N1 O2 S1 p(gss, s/p:40) Chrg 1...

150.0 150.5 151.0 151.5 152.0m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

150.0583R=499553

151.0617R=499368

152.0541R=499443

PositiveSimulated

R=500,000

0.0ppm

C5H11NO2S +H: C5 H12 N1 O2 S1 p(gss, s/p:40) Chrg 1...

151.050 151.055 151.060 151.065m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

151.0617C4

13CH12O2NS

151.0625C5H12O17ONS

151.0646C5

13C14CH918O17O

151.0577C5H12O2N33S

151.55414C4H10

18O233S

PositiveObserved

Negative20130518111518 #1449 RT: 8.07 AV: 1 NL: 1.00E6T: FTMS - p ESI Full ms [140.00-220.00]

179 180 181 182m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

179.0576R=550402

180.0611R=330000

NegativeFull Scan

0.55 ppm

Negative20130518111518 #158 RT: 1.66 AV: 1 NL: 1.24E6T: FTMS - p ESI Full ms [140.00-220.00]

179.050 179.055 179.060 179.065m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

179.0573R=67606

Negative20130518111518 #110 RT: 1.34 AV: 1 NL: 8.48E5T: FTMS - p ESI Full ms [140.00-220.00]

179.050 179.055 179.060 179.065m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

179.0574R=148204

179.0558R=130000

Negative20130518111518 #54 RT: 0.83 AV: 1 NL: 9.54E5T: FTMS - p ESI Full ms [140.00-220.00]

179.050 179.055 179.060 179.065m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

179.0574R=304604

179.0560R=275000

Negative20130518111518 #21 RT: 0.40 AV: 1 NL: 7.38E5T: FTMS - p ESI Full ms [140.00-220.00]

179.050 179.055 179.060 179.065m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

179.0574R=607806

179.0560R=562802

R=60,000 R=120,000

R=240,000 R=500,000200 M

1 M

1.1 ppm

0.56 ppm

0.55 ppm

0.55 ppm

0.56 ppm

1.7 ppm

0.55 ppm

FIGURE 3. Ultrahigh resolution resolves isobaric metabolites

RT: 2.75160 - 2.86730

2.76 2.77 2.78 2.79 2.80 2.81 2.82 2.83 2.84 2.85 2.86Time (min)

0

10

20

30

40

50

60

70

80

90

1000

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

205.09721

205.09724

205.09721

205.09723

205.09717

205.09717205.09717205.09715 205.09718 205.09720

205.09749

205.09747

205.09741205.09746

205.09740205.09770

205.09772205.09746 205.09735 205.09738

NL:3.23E7Base Peak m/z= 205.09525-205.09935 F: ms MS JWUrine_FS240k_LM_Re

NL:2.85E7Base Peak m/z= 205.09525-205.09935 MS jwurine_fs240k_nolm_re

PBUrine_MS2_120k_600ms_bothLM #824 RT: 2.82 AV: 1 NL: 8.68E6F: FTMS + p ESI d Full ms2 [email protected] [50.00-216.00]

80 100 120 140 160 180 200m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

146.06033

118.06544

132.08105 188.07079170.06030without Int. Cal.Avg =1.5 ppm

Full Scan MS HCD MS/MSJWUrine_MS2_120k_600ms_bothLM #896 RT: 2.84 AV: 1 NL: 1.08E7T: FTMS + p ESI d Full ms2 [email protected] [50.00-216.00]

80 100 120 140 160 180 200m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

146.06020

118.06530

132.08090 170.06015 188.0706391.05422

0.2 ppmO

NH

NH

OHO

NH

0.9 ppm 0.7 ppmNH •

0.4 ppm

0.8 ppm

N2H

1 ppm

NH

N 2HNH

-0.1 ppm

with Int. Cal.Avg = 0.4 ppm

1.1 ppm

1.8 ppm

2.0 ppm

1.7 ppm

1.9ppm

2.9 ppm

1.1 ppm

with Int. Cal.

without Int. Cal.

Metabolite Identification Requires High Resolution and Isotope Ratio Measurements Structure elucidation of unknown small molecules by mass spectrometry is a great challenge. The first crucial step is to obtain correct elemental compositions.

Natural occurring elements can be monoisotopic (F, Na, P, I) or polyisotopic (H, C, N, O, S, Cl, Br) [1]. The molecule L-Methionine C5H11NO2S contains multiple polyisotopic elements, especially the S, which result in multiple isotopic peaks needing resolving power above 240,000 at m/z 200 (FWHM). Figure 2A shows the better isotope pattern of molecular ion [M+1+H]+. With increasing resolution up to 450,000 (FWHM), all five isotope peaks are clearly resolved. Figure 2B shows the simulated isotope pattern.

A

B

Metabolite Identification and Quantitation in Complex Samples Requires High ResolutionD-Glucose C6H12O6 and Paraxanthine C7H8N4O2 co-exist in human blood at distinct concentration levels: 5300 M (D-Glucose) vs. 10 M (Paraxanthine) [3]. Although D-Glucose is hundreds of fold more concentrated, it is less readily detectable because of its low ionization e ciency. Their masses di erentiate by 0.0013 amu (179.0561 vs. 179.0574), making the detection of these individual components very challenging. Figure 3 shows that, with >300,000 resolving power, the 200 M (D-Glucose) and 1 M (Paraxanthine) can be readily separated and quantified.

FIGURE 4. Easy Internal Calibration

Metabolite Identification Requires High Accuracy of Mass Measurement in Both MS Full Scan Level and MS/MS LevelFor an unknown metabolite, the number of calculated chemically possible formulae could be hundreds (20 ppm tolerance window), but the possibilities significantly reduce to tens (3 ppm) and several if the tolerance window is narrowed to sub-ppm. However, due to the large amount of known isomeric structures, one single formula may correspond to hundreds of structural isomers, thus structure identification of metabolite isomer using only MS full scan is not feasible. MS/MS and MSn techniques can help solve this analytical problem.

The Orbitrap Fusion MS has a built-in internal calibration functionality (Easy IC). Figure 4 shows LC-MS/MS data of L-Tryptophan from an urine sample with and without internal calibration. Sub-ppm mass accuracy was observed for MS and MS2 fragments with IC activated.

FIGURE 5. Rapid Ion Tree Acquisition on UHPLC Timescale from a Urine Sample (1 MS, 2 MS2 and 4 MS3 in 0.6sec/cycle).(A) Ion trap HCD MS2 followed by Orbitrap CID MS3 were acquired for the precursor ion m/z 181.0721. (B) Spectral tree search in mzCloud matches the MS3 with MS3 from theophylline with the highest score, indicating the unknown must have the same substructure as theophylline.

ConclusionThe Orbitrap Fusion MS o ers ultra high resolution and high fidelity in isotope ratio measurements of metabolites. This allows metabolites in complex mixture with a wide dynamic range be resolved and detected. The Easy-Internal Calibration allows compound assignment be readily done with confidence. The Orbitrap Fusion MS provides ultimate flexibility that facilitates the examination of multiple structural pathways for more structural information at higher speed.

The mzCloud library allows metabolite identification with MS2 and/or MSn data in an automated fashion. The identification based on multistage MS spectral trees and PIF algorithm opens the possibility to identify unknown metabolites and isomers. An improved metabolomics workflow is enabled.

References1. Budzikiewicz H, Grigsby RD, Mass Spectrom. Rev. 2006, 25(1):146-157.

2. Sheldon MT, Mistrik R, Croley TR, J. Am. Soc. Mass Spectrom. 2009, 20(3):370-376

3. http://www.hmdb.ca/metabolites/HMDB00122 & HMDB01860

A

BTheophylline

MS3

isomers

Query (down) and library spectra (up) matchScore

N

NH NH

N

NH

1.5 ppm 1.5 ppm

OH

NH

1 ppm

2.1 ppm

FIGURE 1. The Orbitrap Fusion MS and the Drag and Drop Instrument Method Editor.

O

N

NH NH

Page 7: A Platform to Identify Endogenous Metabolites …...A Platform to Identify Endogenous Metabolites Using a Novel High Performance Orbitrap MS and the mzCloud Library Junhua Wang, 1

ASMS13_MP 045_JWang_E 07/16S

Africa-Other +27 11 570 1840Australia +61 3 9757 4300Austria +43 1 333 50 34 0Belgium +32 53 73 42 41Canada +1 800 530 8447China +86 10 8419 3588Denmark +45 70 23 62 60

Europe-Other +43 1 333 50 34 0Finland/Norway/Sweden

+46 8 556 468 00France +33 1 60 92 48 00Germany +49 6103 408 1014India +91 22 6742 9434Italy +39 02 950 591

Japan +81 45 453 9100Latin America +1 561 688 8700Middle East +43 1 333 50 34 0Netherlands +31 76 579 55 55New Zealand +64 9 980 6700Russia/CIS +43 1 333 50 34 0South Africa +27 11 570 1840

Spain +34 914 845 965Switzerland +41 61 716 77 00UK +44 1442 233555USA +1 800 532 4752

www.thermofisher.com©2016 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific, Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.