A Step-By-Step Guide to A Step-By-Step Guide to Field Calibration Field Calibration for Gilian Air Sampling Pump

A Step-By-Step Guide to Field Calibration

Embed Size (px)

DESCRIPTION

A Step-By-Step Guide to Field Calibration. for Gilian Air Sampling Pump. Items required. Sampling Pump Collection Media Gilibrator Calibrator. Field Calibration. A Field Calibration Consists of Two Parts The Pre-Calibration (Pre-Cal) is done before the sample is taken - PowerPoint PPT Presentation

Citation preview

Page 1: A Step-By-Step Guide to  Field Calibration

A Step-By-Step Guide to A Step-By-Step Guide to Field CalibrationField Calibrationfor Gilian Air Sampling Pump

Page 2: A Step-By-Step Guide to  Field Calibration

Items requiredItems required

• Sampling Pump• Collection Media• Gilibrator Calibrator

Page 3: A Step-By-Step Guide to  Field Calibration

Field CalibrationField Calibration

• A Field Calibration Consists of Two Parts – The Pre-Calibration (Pre-Cal) is done before the sample is taken

– The Post-Calibration (Post Cal) is done after the sample is taken

Page 4: A Step-By-Step Guide to  Field Calibration

The Pre-CalThe Pre-Cal

• The Pre-Cal is used to set the desired sample flow rate, and to verify it on an air flow calibrator.

Page 5: A Step-By-Step Guide to  Field Calibration

Conducting the Pre-CalibrationConducting the Pre-Calibration

Page 6: A Step-By-Step Guide to  Field Calibration

The Pre-CalThe Pre-Cal

• Set the Flow Rate• Observe the Flow

Rate• Record the Flow Rate• An average of 6 to 10

Gilibrator readings is recommended

Page 7: A Step-By-Step Guide to  Field Calibration

Run The SampleRun The Sample

• Pump is placed on the belt

• Sample media is placed on the lapel (Breathing Zone) and pointed downward

• Record the Start Time• Record the End Time

Page 8: A Step-By-Step Guide to  Field Calibration

The Post-CalThe Post-Cal

• The Post-Cal is used to obtain the final flow rate and to verify that the flow rate stayed within +/- 5% of the set flow.

Page 9: A Step-By-Step Guide to  Field Calibration

The Post-CalThe Post-Cal

• Observe the Final Flow Rate

• Record the Final Flow Rate

• An average of 6 to 10 Gilibrator readings is recommended

Page 10: A Step-By-Step Guide to  Field Calibration

Average Flow RateAverage Flow Rate

• The Pre-Cal flow rate and the Post-Cal flow rate may be averaged to obtain an average flow rate for the sample.

Page 11: A Step-By-Step Guide to  Field Calibration

Determining the Average Flow RateDetermining the Average Flow Rate

• Pre-Cal Flow Rate + Post Cal Flow Rate = Avg Flow Rate 2

Page 12: A Step-By-Step Guide to  Field Calibration

Cyclone CalibrationCyclone Calibration

• The Gilian 10mm nylon cyclone has only an exit fitting and must be placed inside a calibration jar to accept the Gilibrator.

Page 13: A Step-By-Step Guide to  Field Calibration

Low Flow ModulesLow Flow Modules

• Calibrating with low flow modules

Page 14: A Step-By-Step Guide to  Field Calibration

Constant Flow ModulesConstant Flow Modules

• Set pump flow to 1500cc/min

• Turn module switch on• Connect sample

tubing to pump• Adjust desired flow

rate at module• Read flow rate at

Gilibrator

Page 15: A Step-By-Step Guide to  Field Calibration

Multi-Flow ModulesMulti-Flow Modules

• Set pump to 1500 cc/min

• Turn module switch on• Connect sample

tubing to pump• Adjust flow rate at

holder• Adjust each tube twice• Read the flow rates

Page 16: A Step-By-Step Guide to  Field Calibration

Further InformationFurther Information

• For detailed information on any of the Gilian products, please consult the product operation manuals.

• Manuals may be accessed through our web site at www.Sensidyne.com