Upload
alejandra
View
1.261
Download
137
Embed Size (px)
Citation preview
8/15/2019 Accumate 2300 Operation Manual-F
1/44
Log
o
Accumate® 2300Volumetric Infusion Pump
Operation Manualwww.wooyoungmed.com
WMA-370210-Rev.0
8/15/2019 Accumate 2300 Operation Manual-F
2/44
2
Important
This Operation Manual is for clinician
use only.
Do not permit patients to have access
to this Operation Manual. Please read the entire Operation
Manual before operating the Accumate®
2300 volumetric infusion pump.
Failure to properly follow warnings,
cautions and instructions could result in
death or serious injury to the patient.
8/15/2019 Accumate 2300 Operation Manual-F
3/44
3
Warnings
If the pump is used to deliver life-sustainingmedication, an additional pump must be available.
Prior to use, always check the audible and visualalarms.
The pump is not to be used for delivery of blood orcellular blood products.
Do not use this pump on mains power if the integrityof the protective earth system is suspect.
To prevent a possible explosion hazard do not usethis pump in presence of inflammable anestheticgases.
Do not use this pump in close proximity to equipmentsuch as surgical diathermy with generates high levels
of RFI. Do not permit the use of mobile telephones in wards
where this pump is in use.
If the pump is dropped or hit, inspect the pump fordamage. Do not use a pump that is damaged or notfunctioning properly. Contact Customer Service toreturn a pump for service.
Do not permit this pump to use if standing in a pool ofliquid.
System delivery inaccuracies may occur as a resultof back pressure or fluid resistance, which dependsupon drug viscosity, catheter size, and extension settubing.
8/15/2019 Accumate 2300 Operation Manual-F
4/44
8/15/2019 Accumate 2300 Operation Manual-F
5/44
5
Cautions
Do not operate the pump at temperatures below+10◦C(50 ◦F) or above 40◦C(104◦F) and humiditylevels below 30% or above 85%.
Do not store the pump at temperatures below+5◦C(41 ◦F) or above 40◦C(104◦F) and humidity levelsbelow 10% or above 90%. Do not store the pump
with the Administration Set attached. Do not store the pump for prolonged periods with the
battery installed (above 3months).
Do not immerse the pump in cleaning fluid or wateror allow solution to soak into the pump, accumulateon the key pad, or enter the battery compartment.
Do not clean the pump with acetone, other plastic
solvents, or abrasive cleaners. Do not use the pump to therapeutic levels of ionizing
radiation.
Do not use the pump directly to ultrasound.
Do not use the pump in the vicinity of magneticresonance imaging (MRI) equipment.
Do not use the pump near ECG equipment.
Do not sterilize the pump. All service and repair of the Accumate® 2300 pumps
must be performed by Woo Young Medical or itsauthorized agents.
8/15/2019 Accumate 2300 Operation Manual-F
6/44
6
Table of Contents
Introduction .........................................................7
Features of the Accumate® 2300 – Front View & Key Pad .................................................................8
– Controls and Indicators .................................................................9
– Rear View ...........................................................10
Directions for Operation – Preparation of the IV-set ..............................................................12
– Turning ON the Power - Rear ............................................................13
–
Turning ON the Power - Front ...........................................................14 – Insertion of the IV-set ................................................................15
– Attach drop sensor ............... ..................................................................17
– Setting the Flow Rate and Delivery Limit ...........................................20
– Clearing the Volume Delivered ..........................................................21
– Operating the Pump .............................................................21
– Purge Infusion (stop mode) ...............................................................22
– Bolus Infusion (start mode) ...............................................................22
– Titration ..................................................................23
– Data Lock ...................................................................24
– Adjust Sound ..................................................................25 – Occlusion Level ................................................................26
– Near End of Infusion ............................................................27
– End of Infusion ...................................................................28
Alarms and Troubleshooting .....................................................29
Specifications ..........................................................32
Maintenance – Routine Maintenance Procedures ...................................................34
– Cleaning and Storage ................................................................35
– Replacing the AC fuses .................................................................36
– Battery operation ...............................................................37
Occlusion Pressure Limits ........................................................38
Start-up & Trumpet Curves ......................................................39
Warranty ................................................................42
Service and Packaging .............................................43
Contact ................................................................44
8/15/2019 Accumate 2300 Operation Manual-F
7/44
7
Introduction
The Accumate® 2300 volumetric infusion pump is a fully
featured infusion pump suitable for critical care and
generation applications.
The Advantages
High precision. Purge, automatic & manual bolus capability.
Easy to operate in only three steps.
– Power On > Select (rate, total, VTBI) > Start
Easy to read, large display.
History function.
– Historical data of infusion records.
(warnings, settings, starts and stops of infusion,etc.)
Micro-infusion for neonatal and high risk application.
None dedicated sets, available from variousmanufacturers.
Programmable, ultrasonic air bubble detector.
Multi-position clamp for vertical or horizontal fixationto poles.
SAFETY STANDARD
The Accumate® 2300 meets all safety standards formedical electrical devices, corresponding to IEC60601-1 and IEC 60601-2-24.
8/15/2019 Accumate 2300 Operation Manual-F
8/44
8
Features of the Accumate® 2300
-Front View & Key Pad
a
b
c
d
e
fg h
ij
k
l
m
n
op
qr
s
t u v
w
x
y
8/15/2019 Accumate 2300 Operation Manual-F
9/44
9
Controls and Indicators
NO. Name Function
a Handle To handle for carrying
b Name Plate To display the pump model and name
c Display All essential information is displayed (flow rate,
delivery limit, volume delivered, alarms
information)d Locker To lock pump door
e Key Pad To set infusion data
f Finger To squeeze the tube with the spring housing by
peristaltic motion
g Occlusion
Sensor
To detect the downstream occlusion of the tube
h Clamp Lever To prevent free flow when the pump door is
opened
i Spring
Housing
To squeeze the tube with the peristaltic fingers
j Door Handle To lock pump door with locker
k Door To protect the tubing set when running
l Door Magnet To detect whether the door is open or not
m Bubble Sensor To detect air in the tube
n LED To indicates pump is infusing or lights up during
an alarm (normal : green, malfunction : red)
8/15/2019 Accumate 2300 Operation Manual-F
10/44
10
Controls and Indicators
NO. Name Function
o Volume
Delivered
Displays Volume Delivered (VTBI:Volume To
Be Infused)
p Delivery Limit Displays delivery limit or alarms information
q Flow Rate Displays flow rate (infusion rate)
r SELECT
/SILENCE
To select setting mode (flow rate, delivery limit,
volume delivered)
/ To silence alarm for 2minutes
s BATTERY
LEVEL
To display the charging status of internal
rechargeable battery
t EXTERNALPOWER
To inform whether the external power isconnected or not
u PURGE To access PURGE or BOLUS key
v POWER
ON/OFF
Press once to switch the pump ON. Press and
hold down for 2 seconds to switch the pump
OFF
w START To start the infusion
x STOP
/CLEAR
To stop the infusion
/To clear the value when setting data
y UP/DOWN To increase or decrease of value shown on
display
8/15/2019 Accumate 2300 Operation Manual-F
11/44
11
Features of the Accumate® 2300
-Rear View
a
b
c
d
e
f g h i
j
k
l
NO. Name Function
a Multi-position Clamp To fix the pump to vertical pole stand or horizontal bar of bed
b Rear Support To support to pole stand
c Drop Sensor Port To connect the drop sensor with the pump
d Nurse Call/ Comm. Port To connect the nurse call jack or communication cable
e Front Panel Lock Switch To lock the front key pad
f 12V DC Jack To supply the power of 12V in ambulance car
g Protective Ground To connect the protective earth line
h Spec Plate To inform the product information
i AC Outlet/Fuse Socket To connect the AC 90~245V main power cord where the 1.25A fuse is
j Rechargeable Battery To hold the NIMH rechargeable battery
k Power Switch To switch the main power
l Clamp Knob To fix the clamp
8/15/2019 Accumate 2300 Operation Manual-F
12/44
12
Direction for Operation
Preparation of the IV-set
– Check for cleanliness, completeness, and
damage. If unit appears damaged, do not use it.
– Check the IV administration sets. If other IV
administration sets are used, the operating safety
of the pump can no longer be guaranteed.
Patient safety may, as a result, be compromised.
• Recommended IV-set
– BD(BECTON DICKINSON) :A122
– KV(KOREA VACCINE) :143T/243T
– Carefully purge the infusion set, without allowingany air bubbles to enter, until the drip chamber is
to 1/3 full. If air has entered, repeat purge-
procedure.
– Close the tubing roller clamp.
1/3
chamber
close
8/15/2019 Accumate 2300 Operation Manual-F
13/44
13
Direction for Operation
Turning ON the Power - Rear
Attach the pump to an IV
pole by turning the knob
on pole clamp.
Plug the pump's power
cord into an electrical
outlet.
Turn ON the MAIN POWER
switch. MAIN POWER switch
must be ON to charge the
battery.
8/15/2019 Accumate 2300 Operation Manual-F
14/44
14
Direction for Operation
Turning ON the Power - Front
Check whether the AC
power LED is turned on.
External Power LED disp lay
wi l l l ight .
Press the POWER ON key
on the front panel.When turnin g the pump OFF, the
pum p must be at STOP mode in
order for the POWER key to respon d.
Wait the self test
procedure.
Check the previously
used data.
8/15/2019 Accumate 2300 Operation Manual-F
15/44
15
Direction for Operation
Insertion of the IV-set
Completely insert the IV-set
into the BUBBLE SENSOR.
Insert the IV-set chamber
into the IV bottle/bag. Fill 1/3
of the chamber.Take caut ions to avoid any air bubb les.
Open the door by pulling
the DOOR HANDLE.
Push the CLAMP LEVER to right
to open. Place the IV-set into
the clamp. This is a safety feature thatprevents free f low w hen the doo r is opened
dur ing infus ion.
8/15/2019 Accumate 2300 Operation Manual-F
16/44
16
Direction for Operation
Insert the IV-set in the occlusion
guide.
Check if the IV-set is
straightforward.
Close the door by pushing
the tightly DOOR HANDLE.
The roller clamp should be below
the pump. Open the roller clamp
to be ready to infuse.Check that no dro ps are fal l ing in th e drip
chamber.
CAUTION : Make sure that IV tube is NOT loosely fastened into the internal
chamber of the pump. It shou ld be straightforw ard.
8/15/2019 Accumate 2300 Operation Manual-F
17/44
17
Direction for Operation
Attach Drop Sensor (When supplied with the
pump - optional)
The drop sensor detects :
– Drops falling in the drip chamber.
–Empty fluid container.
– Upstream occlusion.
Correct drop sensor positioning is very important. The
drip chamber should be about one-third full, allowing
air space for sufficient sensor range. The sensor
should rest halfway between the drop orifice and the
fluid surface so that the sensor beam "sees" only
passing drops.
8/15/2019 Accumate 2300 Operation Manual-F
18/44
8/15/2019 Accumate 2300 Operation Manual-F
19/44
19
If the drop sensor is not positioned correctly, thepump activates false alarms. For example :
If the drop sensor rests
too close to the drop orifice,
the beam may pass throughdrops slowly forming on
the orifice.
The pump activates a false
'drop' alarm.
Positioning the drop sensor
near the bottom of the drip
chamber is incorrect. When
the administration set is
stable, the fluid is still.
The beam passes through
the fluid as it passes through
air and does not detect drops
falling.
The pump activates a false 'drop' alarm.
Direction for Operation
INCORRECT
INCORRECT
8/15/2019 Accumate 2300 Operation Manual-F
20/44
20
Direction for Operation
Setting the Flow Rate and Delivery Limit
Press the SELECT key.
The digits in the LED digital
display will blink in 7seconds.
Use the UP(▲) DOWN(▼) keys to
set the desired rate of infusion.The rate of infusion can be set
from 0.1 to 999.9m l/hr.
Press the SELECT key.
The digits in the LED digital
display will blink in 7seconds.
Use the UP(▲) DOWN(▼) keys to
set the desired volume to be
infused. The volume to be infus ed can be set from 0.1 to 9,999ml.
8/15/2019 Accumate 2300 Operation Manual-F
21/44
21
Direction for Operation
Clearing the Volume Delivered
Operating the Pump
Press the SELECT key.
The digits in the LED digital
display will blink in 7seconds.
Press the STOP/CLR key.
Press the START key.
Check the three LED turned
on sequentially.
8/15/2019 Accumate 2300 Operation Manual-F
22/44
22
Direction for Operation
Purge Infusion (stop mode)
Bolus Infusion (start mode)
Press and hold the PURGE key
when the pump is not infusion.Press and hold the PURGE key unt i l f lu id
f lows and the purg ing o f the IV-set is complete.
CAUTION : PURGE will infuse at the rate of 999 9ml/hr
The volume used dur ing purg ing wi l l be disp layed, but i t is
not added to the volume del ivered.
The pump wi l l not p urge i f the lock h as been enabled.
Press the PURGE key when the
pump is infusion.PURGE wil l in fuse at the rate of 999.9ml/hr.
The feature al lows a bo lus to be given du ring
an infus ion.
Enter the amount of bolus to
be given in ml and Press the
START key. The amo unt of bo lus can be set from 1.0 to 99.9ml.
8/15/2019 Accumate 2300 Operation Manual-F
23/44
23
Direction for Operation
Titration (To change the flow rate while infusing)
Press the SELECT key.
The digits in the LED digital
display will blink in 7seconds.
Use the UP(▲) DOWN(▼) keys to
set the desired rate of infusion.
The rate of infusion can be setfrom 0.1 to 999.9m l/hr.
Press the START key.
CAUTION : The pump wil l no t t i t ration i f th e lock has been enabled.
8/15/2019 Accumate 2300 Operation Manual-F
24/44
24
Direction for Operation
Data Lock (This features prevents unauthorized
use by locking out all key except SILENCE key.)
Turn ON [KEY LOCK] button
to lock the key pad on thefront panel.
Press the any key and
check the displayed in 'HOLd'.
Check the set up and pump
status again.
HOLd
8/15/2019 Accumate 2300 Operation Manual-F
25/44
25
Direction for Operation
Adjust Sound (This feature allows the volume of
the alarm tone to be changed)
Turn ON [KEY LOCK] button
to lock the key pad on thefront panel.
Press and hold the SELECT
key and UP(▲) /DOWN(▼)
keys to increase or decreasethe sound level.
Select the volume of the alarm
tone to be changed.
caut ion : The volum e of the alarm to ch ange can be set from
1(45dB/1m) to 5(65dB/1m)
8/15/2019 Accumate 2300 Operation Manual-F
26/44
26
Direction for Operation
Occlusion Level (The pressure at which a
downstream occlusion alarm will be triggered
varied. The higher the pressure limit, the less
sensitive the pump is changes in fluid
resistance.)
Turn ON [KEY LOCK] button
to lock the key pad on the
front panel.
Press and hold the SELECT
key and UP(▲)/DOWN(▼) keys
to increase or decrease the
occlusion level.
Select the occlusion level
to be changed.
caut ion : The occlus ion level to change can be set low (7 2psi),
center(11 3psi), or h igh (15 3psi).
8/15/2019 Accumate 2300 Operation Manual-F
27/44
27
Direction for Operation
Near End of Infusion (Sets the near end of
infusion warning time, as time left to End of
Infusion.)
Turn ON [KEY LOCK] buttonto lock the key pad on the
front panel.
Press and hold the SELECT
key and UP(▲)/DOWN(▼) keys
to increase or decrease.
Select the near end of infusion
to be changed.
caut ion : The end of infusio n to change can be set from 0(no set)
to 15m inutes.
8/15/2019 Accumate 2300 Operation Manual-F
28/44
28
Direction for Operation
End of Infusion (The pump has reached the end
of the infusion)
In certain conditions the Accumate® 2300
automatically switches to the KVO (Keep-Vein-
Open) rate. The KVO-rate is set to 1.0ml/hr.However, if the flow-rate set by the user is lower
than 1.0ml/hr, the KVO-rate is equal to flow-rate.
This alarm occurs 5 minutes
before infusion is completed.Press the SILENCE key.
This alarm occurs when infusion
is completed. (KVO : Keep Vein
Open)Press the STOP key to reset.
8/15/2019 Accumate 2300 Operation Manual-F
29/44
29
Alarms and Troubleshooting
Alarms are indicated by a combination of an
audible alarm, flashing alarm indicator and
abbreviated message in the display.
First press the SILENCE key to silence the
alarm for a maximum of 2 minutes, then checkthe display for an alarm message. Press the
STOP key to cancel the alarm message.
If the infusion has stopped, rectify the cause of
the alarm then press the START key resume
the infusion.
8/15/2019 Accumate 2300 Operation Manual-F
30/44
30
Alarms and Troubleshooting
Alarm
Indication &
Sound
Possible Cause Action
Air-in-line
Air
bb-bb-bb
Air bubble air-in-line sensor
IV tube not properly inserted
into the Bubble Sensor
Remove the air from the tube
Completely insert the IV
tube into the Bubble Sensor
Door Open
OPEn
bb-bb-bb
Door opened during infusion Close the door
Occlusion
OccL
bb-bb-bb
Downstream tubing kinked
Filter blocked
Closed roller clamp
IV positional or infiltratedCatheter or vein too small
Occlusion limit too low
Correct kinks
Correct blocked filter
Open roller clamp
Check IV site
Increase occlusion limit
Drive Error
Err
bb-bb-bb
Motor malfunction
Mechanical error
Request A/S
Request A/S
Low battery
BAtt
b-b-b
30 minutes until battery is fully
discharged
Connect pump to AC
power
End of infusion
End, conP
b-b-b
Programmed infusion is
complete
Stop infusion and reset
No Good
ng-?
silence
Self-Test NG
Internal Malfunction
Request A/S
8/15/2019 Accumate 2300 Operation Manual-F
31/44
31
Alarms and Troubleshooting
Alarm
Indication
& Sound
Possible Cause Action
Upstream
occlusion
drop
bb-bb-bb
Closed roller clamp Open roller clamp
No flow
drop
bb-bb-bb
No fluid remains in solution
chamber.
Sensor is mounted under fluid
surface on drip chamber.
Drop sensor is misplaced.
Replace container.
Place sensor halfway
between drop orifice and
fluid surface.
LOW-FLOW
FLO.L
bb-bb-bb
Air vent in the administration
is closed while using a bottle.
The pump calibration is out of
range.
Open the air vent in the
administration set.
Checkout and
configuration or request
A/S.
HIGH-FLOW
FLO.H
bb-bb-bb
Damaged administration set.
The pump calibration is out of
range.
Replace administration set.
Checkout and
configuration or request
A/S.
8/15/2019 Accumate 2300 Operation Manual-F
32/44
32
Specifications
Name Volumetric Infusion Pump
Model Accumate® 2300
Pumping System Linear peristaltic
Flow Rate Range 0.1-999.9ml/hr (0.1ml/hr increments)
Delivery Limit Range0.1-999.9ml (0.1 ml increments), 1,000-9,999ml(1ml
increments)
Total Volume Equals to the upper limit of the delivery limit range
Purge Rate Range Equals to the upper limit of the flow-rate range
Accuracy +/- 5% nominal
KVO( Keep-Vein-
Open)0.1-5.0ml/hr(0.1ml/hr increments)
Adjustable occlusion[5.8-17.4psi(3 level : low, center, high)
Adjustable alarm volume (level up or down)
Infusion completed
Alarms Low battery(30 min. before infusion stop)
Battery depleted(3 min. before power off)
Air-in-line
Door open
Drive error
Near End of Infusion
Others
Titration while running
Self testing (System check)
One key bolus
Front panel lock
8/15/2019 Accumate 2300 Operation Manual-F
33/44
33
Specifications
EMS DC 12V
Electrical Safety Class I, Type BF
Dimensions 140(W)*225(H)*140(D)mm
Weight Approximately 2.5Kg
Power Requirements AC 90-245V, 50/60Hz
Battery Built-in rechargeable dc 12V(NI-MH 2.3Ah)
Battery Operation Ca. 4hr @ 100ml/hr
Battery Charging
Time7hrs
Operation Conditions 10-40℃(Relative humidity 30-85%)
Storage Conditions 5-40℃(Relative humidity 10-90%)
Accessories AC power cord for 220V, Operating manual
Warranty 2 years from the date of delivery
Options RS-485, nurse call, drop sensor
CAUTION : Specifications are subject to change without
notice.
8/15/2019 Accumate 2300 Operation Manual-F
34/44
34
Maintenance
Routine Maintenance Procedures
To ensure that pump remains in good operating
condition, it is important to keep it clean and carry out
routine maintenance procedures described below. All
servicing should only be performed by a qualified
service engineer with reference to the TechnicalService Manual for this product.
– As required
Thoroughly clean external surfaces of the pump
before and after prolonged period of storage.
–
At least once per year Inspect AC power supply plug and cable for
damage. Perform functional tests as outlined in
the Technical Service Manual. Operate the pump
on battery power until the battery low alarm then
charge the battery to confirm battery operation
and charging.
8/15/2019 Accumate 2300 Operation Manual-F
35/44
35
Maintenance
Cleaning and Storage
Before the transfer of the pump to a new patient andperiodically during the use, clean the pump by wipingover with a lint-free cloth lightly dampened with warmwater and a standard disinfectant/detergent solution.
Disinfectants which are known to be corrosive tometals must not be used. Concentrated Isopropylalcohol based cleaners will degrade plastic parts.
The IV-set and extension lines are disposable singleuse items and should be discarded after useaccording to their manufacturers' instructions.
If the pump is to be stored for an extended period itshould be first cleaned and the internal battery fullycharged. Store in a clean, dry atmosphere at roomtemperature and, if available, employ the originalpacking for protection.
Once every 3 months during storage, carry outfunctional tests as described in the Technical Service
Manual and ensure that the internal battery is fullycharged.Caut ion : B efore cleaning always switch OFF disconn ect from the
AC power sup ply . Never al low l iquid to enter the casing and
excess f lu id bu i ld up o n the pum p. Do no t steam autoc lave,
ethy lene oxide ster i l ize or imm erse this pump in any f lu id. Keep
the pump plu gged into electr ical power dur in g storage.
8/15/2019 Accumate 2300 Operation Manual-F
36/44
36
Maintenance
Replacing the AC fuses
If the pump continually illuminates the battery
indicator and the AC power indicator light does not
illuminate when the pump is connected to the AC
power supply and switched ON, suspect that either
the power supply fuse in the AC plug or the internal
fuses have blown.
First check the power supply fuse in the AC mains
plug. If the AC power indicator light does not
illuminate remove the pump from service.
It is recommended that only a qualified service
engineer replaces the AC fuses. For further
information regarding the replacement of internal AC
fuses refer to the Technical Service Manual.
8/15/2019 Accumate 2300 Operation Manual-F
37/44
37
Maintenance
Battery Operation
The internal rechargeable battery allows continued
operation when the AC power is unavailable, for
example during patient transfer or AC power failure.
A fully charged battery will provide over 4 hours of
operation at typical infusion rates.
A low battery alert is activated approximately 30
minutes before the battery depleted alarm.
The battery is maintenance free, sealed Nickel Metal
Hydride and requires no routine servicing. However,
to achieve optimum operation, ensure that the battery
is fully recharged after full discharge, before storage,
and at regular 3 month intervals during storage.
Charge retention will eventually degrade. Where
retention is critical the internal battery should be
replaced every 3 years.
It is recommended that only a qualified service
engineer replaces the Battery. For further information
regarding the replacement of batteries refer to the
Technical Service Manual.
8/15/2019 Accumate 2300 Operation Manual-F
38/44
38
Occlusion Pressure Limits
Time to alarm following occlusion is achieved in less
than 5minutes at rates of 25ml/hr and higher by the
appropriate selection of occlusion levels.
The following graphs show the typical values for time
to alarm and bolus volume that can be expected in
the event of an occlusion when the IV-set (KV 243T)is selected.
Time to Alarm - 25ml/hr
0
2
4
6
8
10
12
14
16
0 1:00 2:00 3:00 4:00
min : sec
O c c l u s i o n L e v e l ( p s i
Bolus Volume
0
0.2
0.4
0.6
0.8
1
1.2
1.4
1.6
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
Occlusion Level(psi)
m l
8/15/2019 Accumate 2300 Operation Manual-F
39/44
39
Start-up Curves
& Trumpet Curves
In this pump, as with all infusion systems, the action
of the pumping mechanism and variations in
individual IV-sets cause short-term fluctuations in
rate accuracy.
The following curves show typical performance of the
system in two ways : 1) the delay in onset of fluidflow when infusion commences (start-up curves)...
and 2) the accuracy of fluid delivery over various time
periods is measured (trumpet curves).
The start-up curves represent continuous flow versus
operating time from the start of the infusion. They
exhibit the delay in onset of delivery due tomechanical compliance and provide a visual
representation of uniformity. Trumpet curves are
derived from the second hour of this data. Tests
performed per IEC60601-2-24 standard.
8/15/2019 Accumate 2300 Operation Manual-F
40/44
40
Start-up Curves
& Trumpet Curves
Trumpet curves are named for their characteristic
shape. They display discrete data averaged over
particular time periods or 'observation windows', not
continuous data versus operating time. Over long
observation windows, short term fluctuations have ㅣ
little effect on accuracy as represented by the flatpart of the curve. As the observation window is
reduced, short term fluctuations have greater effects
as represented by the "mouth" of the trumpet.
Knowledge of system accuracy over various
observation windows may be of interest when certain
drugs are being administered. Short term fluctuationsin rate accuracy may have impact depending on the
half-life of the drug being infused, therefore the
clinical effect cannot be determined from the trumpet
curves alone.
Caution : Start-up and trumpet curves may not be
indicative of operation under negative pressure.
8/15/2019 Accumate 2300 Operation Manual-F
41/44
41
Start-up Curves
& Trumpet Curves
Start-up (25ml/hr)
0
5
10
15
20
25
30
1 7 s 2
m 5 m
9 m
1 3 m
1 7 m
3 1 m
Time
R a t e ( m l / h r
Trumpet Curve (25ml/hr)
-15.00%
-10.00%
-5.00%
0.00%
5.00%
10.00%
15.00%
17s 1m 5m 11m 19m 31m
Observation Window(mins)
E r r o r (
8/15/2019 Accumate 2300 Operation Manual-F
42/44
8/15/2019 Accumate 2300 Operation Manual-F
43/44
43
Service & Packaging
Service
– If the pump fails to respond to the operating or
troubleshooting procedures listed in this manual
and the cause cannot be determined, discontinue
use and forward it to an authorized Woo Young
Medical's agency.
– With each complaint, please include pump serial
number and full description of difficulty, including
all setting, types of fluids, time, and alarm
messages.
– Return IV-set used if possible.
– Woo Young Medical cannot assume
responsibility for loss or damage to returned
pumps while they are in transit.
Packaging
– The pump packaging is reusable.
– Package Contents : Accumate® 2300 InfusionPump, power cable, operation manual
8/15/2019 Accumate 2300 Operation Manual-F
44/44
Contact
Manufactured by: WOO YOUNG MEDICAL CO.,LTD.
374-3, Sangsin-Ri,Jincheon-Eup,Jincheon-Gun,
Chungcheongbuk-Do,Korea
Tel: +82 -43-536-0291 Fax: +82-43-536-0290
Homepage : www.wooyoungmed.com
http://www.wooyoungmed.com/http://www.wooyoungmed.com/