19
Agilent InfiniiVision 5000 Series Portable Oscilloscopes Data Sheet The Next Generation of Portable Oscilloscopes 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.TestEquipmentDepot.com Back to the Agilent DSO5032A Info Page

Agilent InfiniiVision 5000 Series Portable … InfiniiVision 5000 Series Portable Oscilloscopes ... If you haven’t purchased an Agilent oscilloscope lately, ... to training and consulting,

Embed Size (px)

Citation preview

Page 1: Agilent InfiniiVision 5000 Series Portable … InfiniiVision 5000 Series Portable Oscilloscopes ... If you haven’t purchased an Agilent oscilloscope lately, ... to training and consulting,

Agilent InfiniiVision 5000 Series Portable OscilloscopesData Sheet

The Next Generation of Portable Oscilloscopes

99 Washington Street Melrose, MA 02176 Phone 781-665-1400Toll Free 1-800-517-8431

Visit us at www.TestEquipmentDepot.com

Back to the Agilent DSO5032A Info Page

Page 2: Agilent InfiniiVision 5000 Series Portable … InfiniiVision 5000 Series Portable Oscilloscopes ... If you haven’t purchased an Agilent oscilloscope lately, ... to training and consulting,

2

• 100, 300, and 500 MHz bandwidths

• 2 or 4 channels

• MegaZoom III memory and display technology• Up to 1 Mpts acquisition memory

(page 4)• Up to 100,000 waveforms per

second real-time update rate (page 5)

• High-definition XGA (1024 x 768) display with 256 levels of intensity grading

• Triggering and hardware accelerated decode for I2C, SPI, CAN, LIN and RS-232/UART(page 6)

• Up to 12 bits of vertical resolution, even in single-shot acquisitions (page 7)

• Complete connectivity – standard (page 6)• USB (2 host, 1 device), LAN,

GPIB, XGA display out• Full remote control, including

web browser • LXI class C compliant

• Manuals and in-scope help available in 11 languages

The new standard for everyday scopes

Traditional bench scopes are great for charac-terizing things that you know about. Agilent's MegaZoom III deep memory and fast update rates help you find the bugs you don’t know about.

Page 3: Agilent InfiniiVision 5000 Series Portable … InfiniiVision 5000 Series Portable Oscilloscopes ... If you haven’t purchased an Agilent oscilloscope lately, ... to training and consulting,

3

If you haven’t purchased an Agilent oscilloscope lately, why should you consider one now?

Leading-edge technology for all scope users

The InfiniiVision 5000 Series oscilloscopes leverage the same third-generation MegaZoom III technology blocks used in our higher performance bench and lab oscilloscopes – responsive deep memory, fast update rates with minimal “dead time,” analog-like display systems, and integrated serial analysis – and deliver them in a compact package, at a price similar to oscilloscopes with older technology blocks.

Industry-leading customer support

As the world’s leading test and measurement vendor, Agilent maintains the largest network of sales engineers, application engineers, support engineers and technicians. From pre-sales collaboration, to calibration, to training and consulting, to repair and servicing, Agilent stands with you throughout the life of your product. It’s no accident that Agilent has such loyal customers.

Don’t take our word for it

Compare the InfiniiVision 5000 Series with your current bench scope. Or compare it to one of our competitors’ newest scopes. You’ll see why Agilent has been the fastest-growing oscilloscope supplier since 1997 (CAGR, source: Prime Data 2006 Test Instrument Industry Service Market Share Analysis).

Model BW (MHz) Channels Maximum sample rate Memory Update rate

DSO5012A 100 2 2 GSa/s

DSO5014A 100 4 2 GSa/s

DSO5032A 300 2 2 GSa/s Up to 1 Mpts1 (page 4) Up to 100,000 waveforms

DSO5034A 300 4 2 GSa/s per second (page 5)

DSO5052A 500 2 4 GSa/s1

DSO5054A 500 4 4 GSa/s1

1 Maximum sample rate and memory are achieved when two channels are interleaved

Page 4: Agilent InfiniiVision 5000 Series Portable … InfiniiVision 5000 Series Portable Oscilloscopes ... If you haven’t purchased an Agilent oscilloscope lately, ... to training and consulting,

4

Why does deep memory matter?

See more time

Seeing more time is the most easily understood use of deep memory. The more samples you acquire, the more time you can see at a particular sample rate.

Long capture times give you better visibility into cause-effect relationships in your design, which greatly simplifies root-cause debugging. It also allows you to capture start-up events (like the start-up sequence in Figure 1) in a single acquisition.

You don’t have to stitch together multiple acquisitions or set precise triggering conditions. Spend less time finding events, and more time analyzing them.

See more details

The relationship between memory depth and acquisition rate isn’t as obvious. All scopes have a “banner” maximum sample rate specification, but many can only sustain these rates at a few timebase settings.

In the case of an oscilloscope with a 5 GSa/s acquisition rate and 10-Kpts of memory (Figure 2), those 10,000 points can only fill 2 µs of time. Since scopes have 10 time divisions, the sample rate drops at any time/div setting below 200 ns/div.

As a result, if you look at “slow/fast” events like a modulated signal, you run the risk of aliasing your signal. Or you may simply miss out on important signal details when you zoom in on it.

Deep memory oscilloscopes let you sustain a high sample rate over longer periods of time.

Always fast, always on

MegaZoom III is the third generation of the fast and deep memory architecture that Agilent introduced in 1996. It’s not a special mode, unlike other deep-memory oscilloscopes. You have access to your MegaZoom memory at all times. And the display will respond instantly to your commands as you scale the +/div setting or pan and zoom in the Delayed (or “zoom”) window.

You can learn more about MegaZoom III deep memory in Application Note 1446 - Deep Memory Oscilloscopes: The New Tools of Choice.

5 GSa/s, 10k memory 4 GSa/s, 1M memory

Inflection occurs at200 ns/div

Inflection occurs at10 µs/div (20 s with single-shot trigger)

Sam

ple

Rate

10 GSa/s

1 kSa/s

1 GSa/s

100 MSa/s

10 MSa/s

1 MSa/s

100 kSa/s

10 kSa/s

Time Base

1ns/div

10ns/div

100ns/div

1µs/div

10µs/div

100µs/div

1ms/div

10ms/div

100ms/div

1s/div

Figure 2.

Figure 1.

Page 5: Agilent InfiniiVision 5000 Series Portable … InfiniiVision 5000 Series Portable Oscilloscopes ... If you haven’t purchased an Agilent oscilloscope lately, ... to training and consulting,

5

Why does a fast update rate matter?

If the human eye has trouble discerning above 30-50 frames per second, is there really a difference between 3,600 and 100,000 waveforms per second? If you know what you’re looking for, the answer is probably “no.” However, if you are hunting for unknown signal anomalies or characterizing jitter, the answer is “yes.”

Acquisition time Dead time Dead timeAcquisition time

Figure 3. Reducing the dead time between acquisitions …

Figure 4. … improves your chances of finding random events like glitches

5000 Series TDS3000B WaveSurfer 400

Upda

te R

ate

(wav

efor

ms p

er se

cond

)

100,000

10,000

1,000

100

10

Time Base

1ns/div

10ns/div

100ns/div

1µs/div

10µs/div

100µs/div

Figure 5. Update rates of popular oscilloscopes using their default real-time acquisition mode

Catch problems sooner and cover more of your debug checklist – our 100,000 waveforms per second update rate helps you find intermittent problems up to 100x faster than competitive scopes

If you know that there is a glitch in your system, it’s easy to capture it using a pulse-width trigger. However, if you are just browsing through your design, your chances of finding a glitch increase as the oscilloscope’s update rate increases. If a glitch occurs during the "dead time" between samples, you miss it (Figure 3). With MegaZoom III technology, the dead time is much smaller (Figure 5). A scope with a slower update rate will capture the glitch eventually (if it recurs), but most engineers and technicians don’t have the time or patience to wait for their tools to catch up.

If you are characterizing signal jitter, a fast update rate gives you accurate results sooner. And when the fast update rate is combined with the 5000 Series’ XGA high-definition display (1024 x 768, 256 intensity levels), subtle differences in these acquisitions become obvious.

And like all other aspects of MegaZoom III technology, this update rate is delivered as a default real-time acquisition mode. It’s always fast, always on.

Learn more about the benefits of a fast update rate by reading Application Note 1551 - Improve Your Ability to Capture Elusive Events: Why Oscilloscope Waveform Update Rates Are Important.

Page 6: Agilent InfiniiVision 5000 Series Portable … InfiniiVision 5000 Series Portable Oscilloscopes ... If you haven’t purchased an Agilent oscilloscope lately, ... to training and consulting,

6

Software options

RS-232/UART serial decode and trigger (N5457A or Option 232 on new scope purchases)The application lets you easily view the information sent over a RS-232 or other UART serial bus. Display real-time time-aligned decode of transmit and receive lines. The application also enables triggering on RS-232/UART conditions.

Figure 8. Trigger on and decode RS-232/UART transmission

Figure 6. On-screen serial decode of an I2C packet

I2C/SPI serial trigger and decode (N5423A or Option LSS on new scope purchases)This application displays real-time time-aligned decode of I2C and SPI serial buses. Hardware-assisted triggering and decode provide the industries fastest throughput and update rate. Hardware-assisted triggering and decode guarantees

you will never miss a trigger event or anomaly – unlike other scopes that have triggering dead time between acquisitions.

This application requires a 4-channel DSO.

For more information: www.agilent.com/find/I2C-SPI

Figure 7. Real-time totalize functions provide CAN bus efficiency and quality measurement statistics

CAN/LIN triggering and decode (N5424A or Option AMS on new scope purchases)Trigger on and decode serially transmitted data based on CAN and LIN protocols. This application not only provides triggering on complex serial signals, but also provides unique hardware-accelerated decode capabilities. Hardware-assisted triggering and decode guarantees

you will never miss a trigger event or anomaly – unlike other scopes that have triggering dead time between acquisitions.

This application requires a 4-channel DSO.

For more information: www.agilent.com/find/CAN-LIN

Segmented memory (N5454A or Option SGM on new scope purchases)Segmented memory optimizes available memory for data streams that have long dead times between activity. The application is most useful for analyzing signal activity associated with laser pulses, serial buses, and bursty signals such as radar. View an

overlay of all signal segments, including serial decode, while highlighting the current segment. Quickly move between segments to view signal detail associated with a specific segment.

For more information:www.agilent.com/find/segmented

This application requires a 4-channel DSO or 4-channel MSO and can use any combination of the scope or logic acquisition channels.

For more information: www.agilent.com/find/RS-232

Figure 9. Use segmented memory to optimize available memory

Page 7: Agilent InfiniiVision 5000 Series Portable … InfiniiVision 5000 Series Portable Oscilloscopes ... If you haven’t purchased an Agilent oscilloscope lately, ... to training and consulting,

7

Other nice features

High resolution mode Offers up to 12 bits of vertical resolution in real-time, single-shot mode. This is accomplished by serially filtering sequential data points and mapping the filtered results to the display when operating at base time settings greater than 10 µs/div.

Help is at your fingertipsAn embedded help system – available in 11 languages – gives you quick answers if you don’t understand a feature. Simply press and hold the corresponding front-panel key, and a screen pops up to explain its function (Figure 10).

Waveform math with FFTAnalysis functions include subtract, multiply, integrate, and differentiate, as well as Fast Fourier Transforms (FFT).

Peak detect 250 ps peak detect on 500-MHz models, 500 ps on 300-MHz models and 1 ns on 100-MHz models helps you find narrow glitches.

AutoProbe interfaceAutomatically sets probe attenuation factors and provides power for selected Infiniium active probes, including the award-winning 1130A 1.5-GHz InfiniiMax differential active probe and 1156A 1.5-GHz single-ended active probe systems.

5-digit hardware counterMeasures frequency up to the bandwidth of the scope.

Trig OutProvides an easy way to synchronize your scope to other instruments. Use the Trig Out port to connect your scope to a frequency counter for more accurate frequency measurements or to cross trigger other instruments.

Autoscale Displays all active signals, and automatically sets the vertical, horizontal and trigger controls.

23 automatic measurements with QuickMeasPressing [QuickMeas] brings up the last four measurements selected. Cursors automatically track the most recently selected measurement.

HDTV triggerThe 5000 Series supports analog HDTV/EDTV triggering for standards like 1080i, 1080p, 720p and 480p as well as standard video triggering on any line within a field, all lines, all fields, and odd or even fields for NTSC, SECAM, PAL and PAL-M video signals.

Easy software upgrades System software is stored in Flash ROM that can be upgraded from the scope’s built-in USB port or LAN. You can find the latest system and IntuiLink Data Capture software at www.agilent.com/find/DSO5000sw.

Oscilloscope toolsE2690B Oscilloscope tools package, licensed by Agilent Technologies from Amherst Systems Associates (ASA), makes it easy to perform in-depth analysis of captured signals. More information can be found in the Oscilloscope Tools data sheet (Agilent publication number 5989-3525EN).

Offline viewing and analysis (B4610A)Save your scope data to a USB or network drive and import the data into a PC-based offline viewer. Pan and zoom. Use searching and filtering to gain insight on analog and digital buses. Email the data to team members who can use the same tool at their PCs. (www.agilent.com/find/InfiniiVisionOffline)

Figure 10. Press and hold a key for instant help

Figure 11. FFT allows you to view the spectral content of this unfiltered sine wave

Figure 12. View and analyze previously acquired scope data on a PC-based offline tool

Page 8: Agilent InfiniiVision 5000 Series Portable … InfiniiVision 5000 Series Portable Oscilloscopes ... If you haven’t purchased an Agilent oscilloscope lately, ... to training and consulting,

8

Probing

To get the most out of your scope, you need the right probes and accessories for your application. That’s why Agilent Technologies offers a complete family of innovative passive and active probes for the 5000 Series scopes to help you get your job done easily. For more

comprehensive information, refer to the Agilent 6000 and 5000 Series Oscilloscope Probes and Accessories Data Sheet (Agilent publication number 5968-8153EN/ENUS) or visit www.agilent.com/find/scope_probes.

Selection guide

N2863A (shipped with 10073C 100 MHz and (shipped with 10076A N2771A 10070C 300 MHz models) 500 MHz models) high-voltage probe high-voltage probe

Probe bandwidth 20 MHz 300 MHz 500 MHz 250 MHz 50 MHz

Probe rise time (calculated) < 17.5 ns < 1.16 ns < 700 ps < 1.4 ns < 7 ns

Attenuation ratio 1:1 10:1 10:1 100:1 1000:1

Input resistance (when 1 MΩ 10 MΩ 2.2 MΩ 66.7 MΩ 100 MΩterminated into 1 MΩ)

Input capacitance Approx. 70 pF Approx. 12 pF Approx. 12 pF Approx. 3 pF Approx. 1 pF

Maximum input 400 Vpk CAT I 300 Vrms 500 Vpk CAT I 4000 Vpk 15 kV dc, 10 kVrms, (dc+peak ac) (mains isolated) (mains isolated) 30 kV dc + peak ac 400 Vpk CAT II 400 Vpk CAT II (post receptacle (post receptacle mains) mains)

Compensation range None 5-30 pF 6-15 pF 6-20 pF 7-25 pF

Probe sense No Yes Yes Yes No

Page 9: Agilent InfiniiVision 5000 Series Portable … InfiniiVision 5000 Series Portable Oscilloscopes ... If you haven’t purchased an Agilent oscilloscope lately, ... to training and consulting,

9

Selection guide (continued)

Current probes Description

1147A 50-MHz 115 A current probe, AC/DC

1146A 100-kHz current probe, AC/DC

N2780A 2-MHz/500 A current probe, AC/DC

N2781A 10-MHz/150 A current probe, AC/DC

N2782A 50-MHz/30 A current probe, AC/DC

N2783A 100-MHz/30 A current probe, AC/DC

N2779A Power supply for N278xA

Active single-ended probes Description

1141A 200-MHz differential probe

1144A 800-MHz active probe

1145A 2-channel 750-MHz active probe

1142A Power supply for 1144A/1145A

1156A 1.5-GHz active probe

Active differential probes Description

N2772A 20-MHz differential probe

N2773A Probe power supply for N2772A

1130A 1.5-GHz InfiniiMax differential probe amplifier (Order one or more InfiniiMax probe heads or connectivity kits per amplifier.)

Probing (continued)

Page 10: Agilent InfiniiVision 5000 Series Portable … InfiniiVision 5000 Series Portable Oscilloscopes ... If you haven’t purchased an Agilent oscilloscope lately, ... to training and consulting,

10

Connectivity

Our customers tell us that oscilloscope connectivity is an increasingly important feature of their test instruments. That’s why the 5000 Series scopes come with the most comprehensive hardware and software connectivity tools in their class.

Hardware connectivity

Standard ports include:

• 2 x USB host ports (for external storage and printing devices), one on the front and one on the rear

• 1 x USB device port for high-speed PC connectivity

• 10/100 Mbit LAN for Internet/intranet connectivity

• GPIB to allow easy migration into existing test systems

• XGA out for external monitors and projectors

LXI class C

LAN eXtensions for Instrumentation (LXI) is a standards-based architecture for test systems. By specifying the interaction of system components, LXI enables fast and efficient test system creation and reconfiguration. The 5000 Series oscilloscopes follow specified LAN protocols and adhere to LXI requirements such as a built-in Web control server, IVI-COM driver, and easy-to-use SCPI commands. The standard Agilent I/O Library Suite makes it easy to configure and integrate instruments in your system.

IntuiLink toolbars

IntuiLink gives you a quick way to move screenshots and data into Microsoft® Word and Excel. These toolbars can be installed from www.agilent.com/find/intuilink.

View Scope logic analyzer and oscilloscope correlation

View Scope enables simple time-correlated measurements between an InfiniiVision 5000 Series oscilloscope and an Agilent 1680/90 or 16800/900 logic analyzer. Scope and logic waveforms are integrated into a single logic analyzer waveform display for easy analysis – all with a simple point-to-point LAN connection. You can also cross-trigger the instruments, automatically de-skew the waveforms, and maintain marker tracking between the instruments.

Figure 13. Agilent Remote Front Panel running in a Web browser

Figure 14. The 5000 Series has just about any connection you might need – standard. Figure 15. Combine best-in-class instruments with a simple connection.

Page 11: Agilent InfiniiVision 5000 Series Portable … InfiniiVision 5000 Series Portable Oscilloscopes ... If you haven’t purchased an Agilent oscilloscope lately, ... to training and consulting,

11

Performance characteristics

Acquisition

Sample rate DSO501xA/503xA: 2 GSa/a each channel DSO505xA: 4 GSa/s half channel1, 2 GSa/s each channel

Equivalent-time sample rate 400 GSa/s (when real-time mode is turned off)

Memory depth 1 Mpts half channel1, 500 kpts each channel

Vertical resolution 8 bits, up to 12 in high-resolution or averaging modes

Peak detection DSO501xA: 1 ns peak detect DSO503xA: 500 ps peak detect DSO505xA: 250 ps peak detect

Averaging Selectable from 2, 4, 8, 16, 32, 64 … to 65536

High resolution mode Average mode with #avg = 1 12 bits of resolution when ≥10 µs/div, at 4 GSa/s or ≥20 µs/div, at 2 GSa/s

Filter Sinx/x interpolation (single shot BW = sample rate/4 or bandwidth of oscilloscope, whichever is less) with vectors on and in real-time mode

Vertical system

Scope channels DSO50x2A: 2 Ch 1 and 2 simultaneous acquisition DSO50x4A: Ch 1, 2, 3, and 4 simultaneous acquisition

Bandwidth (–3dB)2 DSO501xA: DC to 100 MHz DSO503xA: DC to 300 MHz DSO505xA: DC to 500 MHz

AC coupled DSO501xA: 3.5 Hz to 100 MHz DSO503xA: 3.5 Hz to 300 MHz DSO505xA: 3.5 Hz to 500 MHz

Calculated rise time DSO501xA: 3.5 nsec(= 0.35/bandwidth) DSO503xA: 1.17 nsec DSO505xA: 700 psec

Single-shot bandwidth DSO501xA: 100 MHz DSO503xA: 300 MHz DSO505xA: 500 MHz

1 Half channel is when only one channel of channel pair 1-2 is turned on, or one channel of channel pair 3-4 is turned on.2 Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10° C from firmware calibration temperature.

Page 12: Agilent InfiniiVision 5000 Series Portable … InfiniiVision 5000 Series Portable Oscilloscopes ... If you haven’t purchased an Agilent oscilloscope lately, ... to training and consulting,

12

Performance characteristics (continued)

Vertical system (continued)

Range1 2 mV/div to 5 V/div (1 MΩ or 50 Ω)

Maximum input Maximum input voltage for analog inputs: CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk CAT II 100 Vrms, 400 Vpk with N2863A 10:1 probe: CAT I 600 V, CAT II 300 V (DC + peak AC) with 10073C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk with 50 Ω input: 5 Vrms, CAT I

Offset range ±5 V on ranges < 10 mV/div; ±20 V on ranges 10 mV/div to 200 mV/div; ±75 V on ranges >200 mV/div

Dynamic range ±8 div

Input impedance 1 MΩ ± 1% || 12 pF or 50 Ω ± 1.0%, selectable

Coupling AC, DC

BW limit 25 MHz selectable

Channel-to-channel isolation DC to max bandwidth > 40 dB

Standard probes DSO501xA: 10:1 N2863A shipped standard for each oscilloscope channel DSO503xA: 10:1 N2863A shipped standard for each oscilloscope channel DSO505xA: 10:1 10073C shipped standard for each oscilloscope channel

Probe ID Auto probe sense and AutoProbe interface Agilent- and Tektronix-compatible passive probe sense

ESD tolerance ±2 kV

Noise, RMS, input shorted DSO501xA: 0.5% full scale or 250 µV, whichever is greater DSO503xA: 0.5% full scale or 300 µV, whichever is greater DSO505xA: 0.5% full scale or 360 µV, whichever is greater

DC vertical gain accuracy2 ±2.0% full scale

DC vertical offset accuracy ≤ 200 mV/div: ±0.1 div ±2.0 mV ±0.5% offset value; > 200 mV/div: ±0.1 div ±2.0 mV ±1.5% offset value

Single cursor accuracy1 ±DC vertical gain accuracy + DC vertical offset accuracy + 0.2% full scale (~1/2 LSB) Example: for 50 mV signal, oscilloscope set to 10 mV/div (80 mV full scale), 5 mV offset, accuracy = ±2.0% (80 mV) + 0.1 (10 mV) + 2.0 mV + 0.5% (5 mV) + 0.2%(80 mV) = ± 4.785 mV

Dual cursor accuracy1 ±DC vertical gain accuracy + 0.4% full scale (~1 LSB) Example: for 50 mV signal, oscilloscope set to 10 mV/div (80 mV full scale), 5 mV offset, accuracy = ±2.0% (80 mV) + 0.4% (80 mV) = ±1.92 mV

1 2 mV/div is a magnification of 4 mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for 2 mV/div sensitivity setting.2 Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.

Page 13: Agilent InfiniiVision 5000 Series Portable … InfiniiVision 5000 Series Portable Oscilloscopes ... If you haven’t purchased an Agilent oscilloscope lately, ... to training and consulting,

13

Performance characteristics (continued)

Horizontal

Range DSO501xA: 5 nsec/div to 50 sec/div DSO503xA: 2 nsec/div to 50 sec/div DSO505xA: 1 nsec/div to 50 sec/div

Resolution 2.5 psec

Time scale accuracy 25 ppm (0.0025%) to 40° C, plus 1 ppm per ° C from 40° C, to 55° C

Vernier 1-2-5 increments when off, ~25 minor increments between major settings when on

Delay range Pre-trigger (negative delay): Greater of 1 screen width or 125 µs Post-trigger (positive delay): 1 s to 500 seconds

Channel delta-t accuracy Same channel: ±0.0025% reading ±0.1% screen width ±20 ps Channel-to-channel: ±0.0025% reading ±0.1% screen width ±40 ps Same channel example (DSO505xA): For signal with pulse width of 10 µs, oscilloscope set to 5 µs/div (50 µs screen width), delta-t accuracy = ±0.0025% (10 µs) + 0.1% (50 µs) + 20 ps = 50.27 ns

Modes Main, delayed, roll, XY

XY Bandwidth: Max bandwidth Phase error at 1 MHz: < 0.5 degrees Z Blanking: 1.4 V blanks trace (use external trigger on DSO50x2A, channel 4 on DSO50x4A)

Reference positions Left, center, right

Trigger system

Sources DSO50x2A: Ch 1, 2, line, ext DSO50x4A: Ch 1, 2, 3, 4, line, ext

Modes Auto, normal (triggered), single

Holdoff time ~60 ns to 10 seconds

Trigger jitter 15 ps rms

Page 14: Agilent InfiniiVision 5000 Series Portable … InfiniiVision 5000 Series Portable Oscilloscopes ... If you haven’t purchased an Agilent oscilloscope lately, ... to training and consulting,

14

Performance characteristics (continued)

Trigger system (continued)

Selections Edge, pulse width, pattern, TV, duration

Edge Trigger on a rising, falling, or alternating edge of any source

Pulse width Trigger when a positive- or negative-going pulse is less than, greater than, or within a specified range on any of the source channels. Minimum pulse width setting: 5 ns (DSO501xA) 2 ns (DSO503xA, DSO505xA) Maximum pulse width setting: 10 s

Pattern Trigger at the beginning of a pattern of high, low, and don't care levels and/or a rising or falling edge established across any of the channels, but only after a pattern has been established for a minimum of 2 nsec. The channel’s high or low level is defined by that channel’s trigger level.

TV Trigger using any oscilloscope channel on most analog progressive and interlaced video standards including HDTV/EDTV, NTSC, PAL, PAL-M or SECAM broadcast standards. Select either positive or negative sync pulse polarity. Modes supported include Field 1, Field 2, all fields, all lines, or any line within a field. TV trigger sensitivity: 0.5 division of sync signal. Trigger holdoff time can be adjusted in half field increments.

Duration Trigger on a multi-channel pattern whose time duration is less than a value, greater than a value, greater than a time value with a timeout, or inside or outside of a set of time values. Minimum duration setting: 2 ns Maximum duration setting: 10 s

CAN Trigger on CAN (Controller Area Network) version 2.0A and 2.0B signals. Trigger on the start of frame (SOF) bit (standard). N5424A option supports triggering on remote frame ID (RTR), data frame ID (~RTR), remote or data frame ID, data frame ID and data, error frame, all errors, acknowledge error and overload frame.

LIN Trigger on LIN (Local Interconnect Network) sync break at beginning of message frame (standard). N5424A option supports triggering on frame ID. I2C Trigger on I2C (Inter-IC bus) serial protocol at a start/stop condition or user defined frame with address and/or data values. Also trigger on missing acknowledge, address with no acq, restart, EEPROM read, and 10-bit write.

SPI Trigger on SPI (Serial Protocol Interface) data pattern during a specific framing period. Supports positive and negative Chip Select framing as well as clock Idle framing and user-specified number of bits per frame.

RS-232/UART View the information sent over a RS-232 serial bus. Display real-time time-aligned decode of transmit and receive lines. The application also enables triggering on RS-232/UART conditions.

AutoScale Finds and displays all active channels, sets edge trigger mode on highest-numbered channel, sets vertical sensitivity on channels, time base to display ~1.8 periods. Requires minimum voltage > 10 mVpp, 0.5% duty cycle and minimum frequency > 50 Hz.

Page 15: Agilent InfiniiVision 5000 Series Portable … InfiniiVision 5000 Series Portable Oscilloscopes ... If you haven’t purchased an Agilent oscilloscope lately, ... to training and consulting,

15

Performance characteristics (continued)

Channel triggering

Range (internal) ±6 div from center screen

Sensitivity1 < 10 mV/div: greater of 1 div or 5mV; ≥ 10 mV/div: 0.6 div

Coupling AC (~10 Hz), DC, noise reject, HF reject and LF reject (~50 kHz)

1 Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10° C from firmware calibration temperature.

External (EXT) triggering DSO50x2A DSO50x4A

Input impedance 1 MΩ ±3% || 12 pF or 50 Ω ±1% 1.015 kΩ ±5%

Maximum input CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk ±15 V CAT II 100 Vrms, 400 Vpk with N2863A 10:1 probe: CAT I 600 V, CAT II 300 V (DC + peak AC) with 10073C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk 5 Vrms with 50-Ω input, CAT I

Range DC coupling: trigger level ± 1 V and ± 8 V ±5 V

Sensitivity For ± 1V range setting: DC to 100 MHz, 500 mV DC to 100 MHz, 100 mV, >100 MHz to bandwidth of the oscilloscope, 200 mV For ±8 V range setting: DC to 100 MHz, 250 mV; >100 MHz to bandwidth of the oscilloscope, 500 mV

Coupling AC (~10 Hz), DC, noise reject, HF reject and LF reject (~50 kHz)

Probe ID Auto probe sense and AutoProbe interface Agilent- and Tektronix-compatible passive probe sense

Display system

Display 6.3-inch (161 mm) diagonal color TFT LCD

Display update rate Up to 100,000 waveforms/sec in real-time mode

Resolution XGA: 768 vertical by 1024 horizontal points (screen area); 640 vertical by 1000 horizontal points (waveform area) 256 levels of intensity scale

Controls Waveform intensity on front panel. Vectors on/off; infinite persistence on/off, 8 x 10 grid with intensity control

Built-in help system Key-specific help displayed by pressing and holding key or softkey of interest

Real-time clock Time and date (user adjustable)

Page 16: Agilent InfiniiVision 5000 Series Portable … InfiniiVision 5000 Series Portable Oscilloscopes ... If you haven’t purchased an Agilent oscilloscope lately, ... to training and consulting,

16

Performance characteristics (continued)

Measurement features

Automatic measurements Measurements are continuously updated. Cursors track last selected measurement.

Voltage Peak-to-peak, maximum, minimum, average, amplitude, top, base, overshoot, preshoot, RMS, standard deviation

Time Frequency, period, + width, – width and duty cycle on any channel. Rise time, fall time, X at max Y (time at max volts), X at min Y (time at min volts), delay, and phase on oscilloscope channels only.

Counter Built-in 5-digit frequency counter on any channel. Counts up to the oscilloscope’s bandwidth.

Threshold definition Variable by percent and absolute value; 10%, 50%, 90% default for time measurements

Cursors Manually or automatically placed readout of Horizontal (X, ∆X, 1/∆X) and Vertical (Y, ∆Y). Additionally oscilloscope channels can be displayed as binary or hex values.

Waveform math One function of 1-2, 1x2, FFT, differentiate, integrate. Source of FFT, differentiate, integrate: oscilloscope channels 1 or 2, 1-2, 1+2, 1x2.

FFT

Points Fixed at 1000 points

Source of FFT Scope channels 1 or 2 (or 3 or 4 on DSO50x4A only), 1+2, 1-2, 1*2

Window Rectangular, flattop, hanning

Noise floor –50 to –90 dB depending on averaging

Amplitude Display in dBV, dBm at 50 Ω

Frequency resolution 0.05/time per div

Maximum frequency 50/time per div

Storage

Save/recall 10 setups and traces can be saved and recalled using internal non-volatile memory. Optional secure environment mode ensures setups and traces are stored to internal volatile memory so data is erased when power is removed. Compliant to NISPOM Chapter 8 requirements.

Storage type and format USB 1.1 host ports on front and rear panels Image formats: BMP (8-bit), BMP (24-bit), PNG (24-bit) Data formats: X and Y (time/voltage) values in CSV format, ASCII XY format, BIN format Trace/setup formats: Recalled

I/O

Standard ports USB 2.0 high speed device, two USB 1.1 host ports, 10/100-BaseT LAN, IEEE488.2 GPIB, XGA video output

Max transfer rate IEEE488.2 GPIB: 500 kbytes/sec USB (USBTMC-USB488): 3.5 Mbytes/sec 100 Mbps LAN (TCP/IP): 1 Mbytes/sec

Printer compatibility Current printer support can be found at http://agilent.com/find/InfiniiVision-printers

Page 17: Agilent InfiniiVision 5000 Series Portable … InfiniiVision 5000 Series Portable Oscilloscopes ... If you haven’t purchased an Agilent oscilloscope lately, ... to training and consulting,

17

Performance characteristics (continued)

General characteristics

Physical size 35.4 cm wide x 18.8 cm high x 17.4 cm deep (without handle) 38.5 cm wide x 18.8 cm high x 17.4 cm deep (with handle)

Weight Net: 4.1 kg (9 lbs) Shipping: approximately 9 kgs (20 lbs)

Probe comp output frequency ~ 1.2 kHz, amplitude ~2.5 V

Trigger out 0 to 5 V into open circuit (~23 ns delay) 0 to 2.5 V into 50 Ω

Kensington lock Connection on rear panel for security

Power requirements

Line rating range 100-120 V, 50/60/400 Hz; 100-240V, 50/60 Hz auto ranging

Line frequency 50/60 Hz, 100-240 VAC; 440 Hz, 100-132 VAC

Power usage 110 W max

Environmental characteristics

Ambient temperature Operating –10° C to +55° C; non-operating –51° C to +71° C

Humidity operating 95% RH at 40° C for 24 hour; non-operating 90% RH at 65° C for 24 hour

Altitude Operating to 4,570 m (15,000 ft); non-operating to 15,244 m (50,000 ft)

Vibration Agilent class GP and MIL-PRF-28800F; Class 3 random

Shock Agilent class GP and MIL-PRF-28800F (operating 30 g, 1/2 sine, 11-ms duration, 3 shocks/axis along major axis. Total of 18 shocks)

Pollution degree Normally only dry non-conductive pollution occurs. Occasionally a temporary conductivity caused by condensation must be expected.

Indoor use Rated for indoor use only

Other

Measurement categories CAT I: Mains isolated CAT II: Line voltage in appliance and to wall outlet

Regulatory information Safety IEC 61010-1:2001 / EN 61010-1:2001 Canada: CSA C22.2 No. 61010-1:2004 USA: UL 61010-1:2004

Supplementary information The product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC and the EMC Directive 89/336/EEC, and carries the CE-marking accordingly. The product was tested in a typical configuration with HP/Agilent test systems. Product specifications, characteristics, and descriptions in this document are subject to change without notice.

Page 18: Agilent InfiniiVision 5000 Series Portable … InfiniiVision 5000 Series Portable Oscilloscopes ... If you haven’t purchased an Agilent oscilloscope lately, ... to training and consulting,

18

Ordering information

Available models

Product number Description

DSO5012A 100 MHz, 2-channel portable oscilloscope

DSO5014A 100 MHz, 4-channel portable oscilloscope

DSO5032A 300 MHz, 2-channel portable oscilloscope

DSO5034A 300 MHz, 4-channel portable oscilloscope

DSO5052A 500 MHz, 2-channel portable oscilloscope

DSO5054A 500 MHz, 4-channel portable oscilloscope

Standard features

Product number Description

Warranty 3-year return-to-Agilent warranty

Hardcopy user’s guide ABA Printed user's guide in English ABD Printed user's guide in German ABF Printed user's guide in French ABZ Printed user's guide in Italian ABJ Printed user's guide in Japanese AC6 Printed user's guide in Korean AB9 Printed user's guide in Portuguese AKT Printed user's guide in Russian AB2 Printed user's guide in simplified Chinese ABE Printed user's guide in Spanish AB0 Printed user's guide in traditional ChinesePower cord 900 United Kingdom 901 Australia/New Zealand 902 Continental Europe 903 United States/Canada 906 Switzerland 912 Denmark 917 South Africa/India 918 Japan 919 Israel 920 Argentina 921 Chile 922 China 927 Brazil/Thailand

Probes One probe per channel DSO501x, DSO503x: N2863A DSO505x: 10073C

Software Agilent IO Libraries

Other documentation Service guide (CD-ROM, English), Certificate of Calibration, Declaration of Conformance

Miscellaneous Protective front cover

Page 19: Agilent InfiniiVision 5000 Series Portable … InfiniiVision 5000 Series Portable Oscilloscopes ... If you haven’t purchased an Agilent oscilloscope lately, ... to training and consulting,

19

Options

Product number Description

SEC Secure environment mode – Provides compliance with National Industrial Security Program Operating Manual (NISPOM) Chapter 8 requirements (factory installed option only for new purchase, not retrofittable for existing 5000 Series scope).

A6J ANSI Z540-compliant calibration

Serial data analysis applications

Option number – Option number – user installed factory installed Description

N5424A AMS CAN/LIN automotive triggering and decode (4-channel models only)

N5423A LSS I2C/SPI serial decode option (for 4-channel models only)

N5457A 232 RS-232/UART triggering and decode (4 channel models only)

N5454A SGM Segmented memory

Available software

Product number Description

Oscilloscope Tools E2690B Oscilloscope Tools software (U.S. and Canada)

N5385B Oscilloscope Tools software (International)

E2693B 1 year update subscription for oscilloscope tools (U.S. and Canada)

N5388B 1 year update subscription for oscilloscope tools (International)

(You will order Option 005 – Scope Guide. High-bandwidth Agilent oscilloscopes support more comprehensive versions of Oscilloscope Tools. Please refer to Agilent Publication 5989-3525EN: E2690B Oscilloscope Tools for more details.)

Accessories

Product number Description

N2916B Rackmount kit for 6000 and 5000 Series oscilloscopes

N2917B Transit case for 6000 and 5000 Series oscilloscopes

N2760A Soft carrying case for 5000 Series oscilloscopes

Cable

Product number Description

10833A GPIB cable, 1 m long

Probing

Common probes are listed on page 9. For a more comprehensive list, please refer to the Agilent 6000 and 5000 Series Oscilloscope Probes and Accessories Data Sheet (Agilent publication number 5968-8153EN/ENUS), or visit www.agilent.com/find/scope_probes

Ordering information (continued)

Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176

FAX 781.665.0780 - TestEquipmentDepot.com

99 Washington Street Melrose, MA 02176 Phone 781-665-1400Toll Free 1-800-517-8431

Visit us at www.TestEquipmentDepot.com

Back to the Agilent DSO5032A Info Page