1
Agilent Mobile Laboratory for Air Toxics Monitoring in Pollution Emergencies First European Mobile Lab based on TDGC/MSD 5975T for Ambient Air online Monitoring Sampling Analysis Data processing and Report Transport Conventional Laboratory Sampling Analysis Data processing and Report Mobile Lab + + Automatically In-situ On-line Reduces the time to obtain reliable analytical results, compared to conventional laboratory testing, helping to make quick decisions on site in a Pollution Emergency Mobile Lab Specifications for Environmental Applications Vehicle with enough interior space: 2.3 meters interior height. Enough space in laboratory bench for two operators Vehicle drivable with a ordinary license (B1 type). No need of an expert driver for moving around External electrical connection, Integrated Power Generator and UPS for full autonomy and protection. Air conditioning in the roof and ventilation systems. Thermal insulation in walls and floors, with nonslip plywood Gas supply system: helium cylinder or hydrogen generator. Automated exchange of bottle for maximum autonomy Furniture with antivibration systems for analytical instruments (not required for MSD 5975T). Security system with locks for different equipment / furniture. Integrated weather station to measure : wind speed, direction, humidity, temperature, pressure, GPS data, etc., installed on top al a pneumatic telescopic mast very easy to deploy from the vehicle interior.. No need to jump on the roof anymore. Online monitoring of Volatile Organic Compounds (VOCs) by TDGC/MSD 5975T. Possibility to rum alternatively canisters and bags as well as up to 100 TD tubes automatically. Calibration with liquid sampler, pressurized gas cylinders or spiked TD tubes (loading ring) Integrated continuous multiparameter analyser for analysing Ozone (O3), Nitrogen Dioxide (NO2), Nitrogen Oxides (NOX), Carbon Monoxide (CO), Carbon Dioxide (CO2), Sulphur Dioxide (SO2), Hydrogen Sulphide (H2S), Particles (PM10 & PM2.5) Optimized control software for remote operation via PC or iPad Time between each analysis 18 min First quantitative result from arrival in < 70 minutes Amount (ug/m3) AMDIS NIST R.T. Compound Name Chem station AMDIS Match R.T. Diff sec. Reverse Match Hit Num. 2.633 Methylcyclopentane 0.71 93 88 1 2.9391 Tetrachloromethane 0.18 93 94 1 2.9582 Benzene 4.99 95 95 1 2.9857 3-Methylhexane 0.63 93 93 1 3.1628 n-Heptane 1.52 98 86 1 3.4006 Methylcyclohexane 1.44 96 89 1 3.5283 4-Methyl-2-pentanone (MIBK) 1.38 97 85 2 3.7323 3-Methylheptane 0.38 91 89 1 3.7908 Toluene 42.52 100 93 2 3.9157 n-Octane 0.96 94 84 2 3.9454 n-Hexanal 1.67 97 91 1 4.0209 n-Butyl acetate 1.02 96 91 1 4.0921 Tetrachloroethene 0.19 96 92 1 4.4323 3-Methyloctane 0.35 93 88 1 4.4640 Ethylbenzene 54.37 99 95 1 4.5466 m,p-Xylene 190.92 100 98 1 4.6087 n-Nonane 0.74 94 85 1 4.6919 Styrene 1.34 93 92 1 4.724 o-Xylene 31.68 95 90 1 4.8825 Isopropylbenzene (cumene) 2.93 90 85 1 4.9488 alpha-Pinene 26.52 97 94 2 5.1344 m-Ethyltoluene 3.73 98 93 1 5.3919 1,3,5-Trimethylbenzene 4.6 99 94 2 5.4103 2-Ethyl-1-hexanol 0.69 90 84 2 5.5324 Limonene 1.39 97 93 1 6.3445 n-Dodecane 2.68 97 92 1 6.5442 Naphthalene 0.84 91 88 1 Zero Air Standard Gas ¾Extended list of compounds (Example: more than 170 VOCs) ¾Online Sampling ¾Short analysis cycles (~20min) that permits nearly continuous analysis sequences ¾Low detection limits (pptv~ppbv) ¾Meets criteria ¾Integrated LTM (Low Thermal Mass), enables very fast separation for 170 VOC’s ¾Automatic & Rapid DRS report generation for calibrated and unknown compounds ¾Continuous monitoring and logging of weather parameters allows correlation between atmospheric conditions and analytical results Advantages of a TDGC/MSD based Mobile Laboratory for Air Toxics Analysis Mobility in case of emergency Instruments with antivibration design and test for transportability Ready to start to work immediately without hardware modifications Versatility of applications (air, water & soil samples) Extended list of compounds in a very fast single analysis ( 8 minutes) Hyphenated Analytical Techniques for a high range of sample matrices Multiple possible sample introduction techniques (Liquid sampler, TSP, TD, P&T, HS, SPME…) Minimum response time Easy to start up of all systems (Quick start up protocols) First analytical result generated in < 100min. since arrival with the Mobile Laboratory HTML Results table automatically generated after each analysis Real Case of Urban Outdoor Pollution Emergency 1- EMERGENCY CALL AND MOBILE LAB PREPARATION Time 0:00 to 0:10 (10 min) 2- MOVING THE MOBILE LAB TO THE CRITICAL ZONE Time 0:10 to 0:38 (28 min) 3- START-UP ANALYTICAL SYSTEMS & WEATHER STATION Time 0:38 to 0:48 (10 min) 4- CHECKING SYSTEMS AND TUNE MSD Time 0:48 to 1:14 (26 min) 5- RUN SYSTEM BLANK AND 1 ST SAMPLE ANALYSED Time 1:14 to 1:48 (34 min) Amount (ug/m3) AMDIS NIST R.T. Compound Name Chem station AMDIS Match R.T. Diff sec. Reverse Match Hit Num. 1.7790 Acetone 1.77 97 86 1 2.968 Benzene 0.06 90 95 1 3.168 n-Heptane 0.23 91 89 1 3.8000 Toluene 4.39 100 92 1 4.100 Tetrachloroethene 0.18 97 94 1 4.4725 Ethylbenzene 5.32 98 95 1 4.5564 m,p-Xylene 31.21 99 96 1 4.615 n-Nonane 0.27 93 91 1 4.7004 Styrene 0.64 97 90 1 4.732 o-Xylene 4.57 95 90 1 4.9579 alpha-Pinene 0.97 94 90 2 5.1414 m-Ethyltoluene 0.51 97 90 2 5.2550 n-Decane 0.76 95 94 1 5.3990 1,3,5-Trimethylbenzene 2.22 97 92 1 5.847 n-Nonanal 0.58 92 86 1 6.3523 n-Dodecane 1.24 96 96 1 6.5513 Naphthalene 0.87 93 88 2 ANALYTICAL RESULTS GENERATED WITHIN 70 MINUTES FROM ARRIVAL Ph.D. Johannes L. González, BSc. Marc Gibert, Ph.D. Mariano Martínez, BSc. Angel Sánchez Ingenieria Analitica S.L. Avda. Cerdanyola, 73 E08172 Sant Cugat del Valles BarcelonaSPAIN 1 2 3 5 4 Ingeniería Analítica S.L. Avda. Cerdanyola, 73 E08172 Sant Cugat del Valles BarcelonaSPAIN Ph. +34935902850 Fax. +34936750516 Email: inf@ingenieriaanalitica.com www.ingenieriaanalitica.com Instrument Configuration : Thermo Desorber coupled to a GC/MSD: Air ServerUnityUltra5975T MSD ¾ Mobility and Rapid Response in emergency situations obtaining in situ analytical accurate data ¾ Fast analysis and automatic results generation ¾ Easytouse. Non expertise user to start the measurements. ¾ Remote access, control and supervision. ¾ Multipurpose: Discrete and Online continuous analysis. Indoor and outdoor sampling. ¾ Pain relief for emergency bodies, governmental agencies, police departments and military. ¾ GC/MSD Lab performance in the field for Environmental, Toxicology and Food applications ¾ Other Mobile Solutions from Agilent: HPLC, FTIR, Microwave PlasmaAES and microGC ¾ Integration of nonAgilent instruments in the Mobile Lab: Multiparametric, Weather Station,… Conclusions Weather Station Log

Agilent Mobile Laboratory for Air Toxics Monitoring in Pollution … · Agilent Mobile Laboratory for Air Toxics Monitoring in Pollution Emergencies First European Mobile Lab based

  • Upload
    others

  • View
    4

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Agilent Mobile Laboratory for Air Toxics Monitoring in Pollution … · Agilent Mobile Laboratory for Air Toxics Monitoring in Pollution Emergencies First European Mobile Lab based

Agilent Mobile Laboratory for Air Toxics Monitoring in Pollution Emergencies

FirstEuropeanMobileLabbasedonTD‐GC/MSD5975TforAmbientAiron‐lineMonitoring

Sampling AnalysisData processing

and ReportTransportConventional

Laboratory

Sampling AnalysisData processing

and ReportMobile Lab + +

AutomaticallyIn-situOn-line

Reduces the time to obtain reliable analytical results, comparedto conventional laboratory testing, helping to make quickdecisions on site in a Pollution Emergency

Mobile Lab Specifications for Environmental ApplicationsVehicle with enough interior space: 2.3 meters interior height. Enough space in laboratory bench for two operatorsVehicle drivable with a ordinary license (B1 type). No need of an expert driver for moving aroundExternal electrical connection, Integrated Power Generator and UPS for full autonomy and protection.Air conditioning in the roof and ventilation systems. Thermal insulation in walls and floors, with non‐slip plywoodGas supply system: helium cylinder or hydrogen generator.  Automated exchange of bottle for maximum autonomy Furniture with anti‐vibration systems for analytical instruments (not required for MSD 5975T). Security system with locks for different equipment / furniture.Integrated weather station to measure : wind speed, direction, humidity, temperature, pressure, GPS data, etc., installed on top al a pneumatic telescopic mast very easy to deploy from the vehicle interior.. No need to jump on the roof anymore.On‐line monitoring of Volatile Organic Compounds (VOCs) by TD‐GC/MSD 5975T. Possibility to rum alternatively  canisters and bags as well as up to 100  TD tubes automatically.Calibration with liquid sampler, pressurized gas cylinders or spiked TD tubes (loading ring)Integrated continuous multiparameter analyser for analysing Ozone (O3), Nitrogen Dioxide (NO2), Nitrogen Oxides (NOX), Carbon Monoxide (CO), Carbon Dioxide (CO2), Sulphur Dioxide (SO2), Hydrogen Sulphide (H2S), Particles (PM10 & PM2.5)Optimized control software for remote operation via PC or iPad

Time between each analysis18 min

First quantitative result from arrival in < 70 minutes Amount (ug/m3) AMDIS NIST

R.T. Compound Name Chem station

AMDIS Match R.T. Diff sec.

Reverse Match

Hit Num.

2.633 Methylcyclopentane 0.71 93 88 1 2.9391 Tetrachloromethane 0.18 93 94 1 2.9582 Benzene 4.99 95 95 1 2.9857 3-Methylhexane 0.63 93 93 1 3.1628 n-Heptane 1.52 98 86 1 3.4006 Methylcyclohexane 1.44 96 89 1 3.5283 4-Methyl-2-pentanone (MIBK) 1.38 97 85 2 3.7323 3-Methylheptane 0.38 91 89 1 3.7908 Toluene 42.52 100 93 2 3.9157 n-Octane 0.96 94 84 2 3.9454 n-Hexanal 1.67 97 91 1 4.0209 n-Butyl acetate 1.02 96 91 1 4.0921 Tetrachloroethene 0.19 96 92 1 4.4323 3-Methyloctane 0.35 93 88 1 4.4640 Ethylbenzene 54.37 99 95 1 4.5466 m,p-Xylene 190.92 100 98 1 4.6087 n-Nonane 0.74 94 85 1 4.6919 Styrene 1.34 93 92 1 4.724 o-Xylene 31.68 95 90 1 4.8825 Isopropylbenzene (cumene) 2.93 90 85 1 4.9488 alpha-Pinene 26.52 97 94 2 5.1344 m-Ethyltoluene 3.73 98 93 1 5.3919 1,3,5-Trimethylbenzene 4.6 99 94 2 5.4103 2-Ethyl-1-hexanol 0.69 90 84 2 5.5324 Limonene 1.39 97 93 1 6.3445 n-Dodecane 2.68 97 92 1 6.5442 Naphthalene 0.84 91 88 1  

Zero AirStandard Gas

Extended list of compounds (Example: more than 170 VOCs)On‐line SamplingShort analysis cycles (~20min) that permits nearly continuous analysis sequencesLow detection limits (pptv~ppbv)Meets criteriaIntegrated LTM (Low Thermal Mass), enables very fast separation for 170 VOC’sAutomatic & Rapid DRS report generation for calibrated and unknown compoundsContinuous  monitoring and logging of weather parameters allows  correlation 

between atmospheric conditions and analytical results

Advantages of a TD‐GC/MSD based Mobile Laboratory for Air Toxics Analysis

Mobility in case of emergency• Instruments with anti‐vibration design and test for transportability• Ready to start to work immediately without hardware modifications

Versatility of applications (air, water & soil samples)• Extended list of compounds in a very fast single analysis ( 8 minutes)• Hyphenated Analytical Techniques for a high range of sample matrices• Multiple possible sample introduction techniques (Liquid sampler, TSP, TD,

P&T, HS, SPME…)

Minimum response time• Easy to start up of all systems (Quick start up protocols)• First analytical result generated in < 100min. since arrival with the Mobile Laboratory• HTML Results table automatically generated after each analysis

Real Case of Urban Outdoor Pollution Emergency

1- EMERGENCY CALL AND MOBILE LAB PREPARATIONTime 0:00 to 0:10 (10 min)

2- MOVING THE MOBILE LAB TO THE CRITICAL ZONETime 0:10 to 0:38 (28 min)

3- START-UP ANALYTICAL SYSTEMS & WEATHER STATIONTime 0:38 to 0:48 (10 min)

4- CHECKING SYSTEMS AND TUNE MSD Time 0:48 to 1:14 (26 min)

5- RUN SYSTEM BLANK AND 1ST SAMPLE ANALYSEDTime 1:14 to 1:48 (34 min)

Amount (ug/m3) AMDIS NIST R.T. Compound Name Chem

station AMDIS Match R.T. Diff

sec. Reverse Match

Hit Num.

1.7790 Acetone 1.77 97 86 1 2.968 Benzene 0.06 90 95 1 3.168 n-Heptane 0.23 91 89 1 3.8000 Toluene 4.39 100 92 1 4.100 Tetrachloroethene 0.18 97 94 1 4.4725 Ethylbenzene 5.32 98 95 1 4.5564 m,p-Xylene 31.21 99 96 1 4.615 n-Nonane 0.27 93 91 1 4.7004 Styrene 0.64 97 90 1 4.732 o-Xylene 4.57 95 90 1 4.9579 alpha-Pinene 0.97 94 90 2 5.1414 m-Ethyltoluene 0.51 97 90 2 5.2550 n-Decane 0.76 95 94 1 5.3990 1,3,5-Trimethylbenzene 2.22 97 92 1 5.847 n-Nonanal 0.58 92 86 1 6.3523 n-Dodecane 1.24 96 96 1 6.5513 Naphthalene 0.87 93 88 2  

ANALYTICAL RESULTS GENERATED WITHIN 70 MINUTES FROM ARRIVAL

Ph.D. Johannes L. González, BSc. Marc Gibert, Ph.D. Mariano Martínez, BSc. Angel SánchezIngenieria Analitica S.L.  ‐ Avda. Cerdanyola, 73 ‐ E‐08172 Sant Cugat del Valles ‐ Barcelona‐ SPAIN 

1 2 3 54

Ingeniería Analítica S.L.  ‐ Avda. Cerdanyola, 73 ‐ E‐08172 Sant Cugat del Valles ‐ Barcelona‐ SPAIN Ph. +34‐935902850 ‐ Fax. +34‐936750516 ‐ Email: inf@ingenieria‐analitica.com ‐ www.ingenieria‐analitica.com

Instrument Configuration : Thermo Desorber coupled to a GC/MSD:  Air Server‐Unity‐Ultra‐5975T MSD

Mobility and Rapid Response in emergency situations obtaining in situ analytical accurate dataFast analysis and automatic results generationEasy‐to‐use. Non expertise user to start the measurements.Remote access, control and supervision.Multipurpose: Discrete and On‐line continuous analysis. Indoor and outdoor sampling.Pain relief for emergency bodies, governmental agencies, police departments and military.GC/MSD Lab performance in the field for Environmental, Toxicology and Food applicationsOther Mobile Solutions from Agilent: HPLC, FTIR, Microwave Plasma‐AES and microGCIntegration of non‐Agilent instruments in the Mobile Lab: Multiparametric, Weather Station,…

Conclusions

Weather Station Log