Design of a New Laboratoryfor Chemical Challenge Testing of

Respirator Cartridges, Canisters, and Filters

C.R. (Gus) Manning, PhD, CIH

Page 2: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

Miller-Nelson, a leading independent Lab for 3 decades Founded by Gary O. Nelson (Fellow AIHA)

Miller-Nelson Lab acquired by Assay Technology (2004)

Incorporated into existing Analytical Lab facility tasks performed are different from an Analytical Lab

Eventually * we would need to design a more appropriate Lab

___________________________* 2009

Chemical Challenge TestingBackground

Page 3: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

AT / Miller-Nelson LabTypical Services

Respirator & Filter Testing (Chemical Challenge)Pre-Certification Test Data for APR Mfr

– Submitted to NIOSH

3rd Party QC Test– for Manufactured Respirator

Special Challenge Testing– Challenge Agents additional to those included in

NIOSH certification testing

Special Projects– Filters and Adsorbents

___________________________APR = Air Purifying Respirator

Page 4: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

Chemical ChallengeRespirator & Filter Testing

Characteristics

Flowing Toxic Stream with plumbing issues– Fume removal by hood

Control Flow, Concentration, Humidity, Temp– Complex Work

Continuous Monitoring with multiple Instruments– “Instrument Cluster”

Lots of Utilities required– Power, Water, Nitrogen, Air, Compressed Gases

Page 5: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

Lab Design ConsiderationsPreliminary

SpacePerform 3 - 4 tests at one time

Health & SafetyPrevent Exposures, Spills, and Explosions

Quick Set-Up and Take -DownQuick Changeover for Short Runs

TestingAccommodate Instrument Cluster

– 5 or more Instruments used in 1 test

StorageSafe gas storage in or near Lab

Page 6: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

Assay Technology Facility(Livermore, California)

10,550 sq. ft.

Page 7: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

EntireFacility

1382 Stealth StLivermore, CA

10,550 Sq. ft. 340ft2

960ft2

Page 8: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

Components ofthe

Instrument Cluster

Page 9: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

MNR Flow GeneratorHCS-401

25” x 25” x 12”

Page 10: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

IR Spectrometer(long path vapor cell)

Page 11: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

Electrochemical Monitors

Page 12: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

Mass Flow Controller

Page 13: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

RH Vessel Cal Vessel

Page 14: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

Schematic Set-Up for Test

IR

TapeMonitorGlass Tube

NH3Flow

Switch

Page 15: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

Lab Bench Unit(criteria)

Within Fume Hood environment– May leak toxic vapors

Can Contain “Instrument Cluster”– Provides power, water, gas, & air

Rugged & Robust – Physically & chemically

Semi-Portable – Re-arrange for new experiments

Page 16: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

Lab Bench Layout(Side View)

IRIR

Recorder

FID

RHVessel

CalVessel

TestCell

MFCSyringePump

6'

Compressed Air5'

MNR FlowGenerator

E ChemMonitor

Nitrogen

Page 17: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

Lab BenchLayout(Top View)

Page 18: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

WorkStation

(model)

Page 19: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

Work Station(example)

Page 20: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

Lab BenchLayout(Top View)

Page 21: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

Lab Bench Layout(Side View)

IRIR

Recorder

FID

RHVessel

CalVessel

TestCell

MFCSyringePump

6'

Compressed Air5'

MNR FlowGenerator

E ChemMonitor

Nitrogen

Page 22: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

Lab Bench(except for Hood)

E ChemMonitor

Recorder MFCSyringePump

Test Cell

IRSpec

IRSpec

MNR FlowGenerator

RHVessel Cal

Vessel

GasCyl

Page 23: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

Lab Bench(complete view)

Page 24: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

Lab Bench(close-up)

A

E CD

BF

G

I

H

Page 25: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

Overall Lab Arrangement(Integration of Bench Units)

Need 4-5 separate Test Benches– 2-5 tests simultaneously

Comply with Air Handling Constraints – Share Hood Ventilation Fans

Spaced as close as practical– Limited to ca. 1,000 sq.ft

Open, Portable Layout – Re-arrange for new experiments

Page 26: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

Lab Facility1382 Stealth

Livermore, CA

AT MNR LabRev 12-08-091,250 sq.ft.

1382 Stealth StLivermore, CA

AnalyticalLab

MNRLAB E1

E1

Diluted Gas Cylinders2% Hydrogen Cyanide1.5% Cyanogen Chloride0.9% Phosphine0.2% Phosgene7% Chlorine30% Carbon Monoxide7% Hydrogen Sulfide

Solvents(1, 2, 4 L. bottles)AcetonitrileDichloromethaneCyclohexaneMethanolEthanolGlyceroletc.

T

E2

E3

E3

5'

2.5'

5'

5'

5.5'

E4

E 4

13'

8.5'

E2

2.5'

Ovn

6'

13'

E5

4'

10'

Ovn

GC GC

GCGC

40'

24'

Page 27: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

MNR Lab(close-up)

AT MNR LabRev 12-08-09

1,250 sq.ft.1382 Stealth StLivermore, CA

AnalyticalLab

MNRLAB

Ho

od

Hoo

d

E1

E1

A

Diluted Gas Cylinders2% Hydrogen Cyanide1.5% Cyanogen Chloride0.9% Phosphine0.2% Phosgene7% Chlorine30% Carbon Monoxide7% Hydrogen Sulfide T

E2

E3

E3

5'

2.5'

5'

5'

5.5'

E4

E 4

13'

8.5'

E2

2.5'

10'

GC GC

GCGC

40'

24'

Stor

age

Lab

Ben

ch Test A

Test B

Test C

Test D

Test EGas Cabinets

Page 28: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

AT Respirator and FilterChemical Challenge Test Lab

Page 29: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

Chemical Challenge Testing Benches(+ Sink)

Page 30: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

Lab Center Aisle

Page 31: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

Set-Up Bench & Storage

Page 32: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

AT Respirator and FilterChemical Challenge Test Lab

Page 33: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

Finis

Page 34: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

Ammonia Testing

Challenge Level at 1000 mg/M3– Same as some NIOSH tests– Monitor by IR

Breakthrough Monitoring at 17 mg/M3– Monitor by Colorimetric Tape Method

Beyond Breakthrough (up to 170 mg/M3) – Monitor by IR

Page 35: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

Dimethylmethylphosponate (DMMP) Testing

Challenge Level at 3000 mg/M3– There is no NIOSH test– Verify Gravimetrically & by FID

Breakthrough Monitoring (0.04 mg/M3)– Monitor by Flame Photometric Detector *

Beyond Breakthrough (0.4 mg/M3) – Monitor by FID

___________________________________________________(*) We have only found FPD available as part of a GC-FPD

(Need to purchase soon to complete tests on schedule.)

Page 36: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

Nitrogen Dioxide Testing

Challenge Level at 376 & 188 mg/M3– 100 and 200 ppm– Similar to NIOSH tests– Verify by Titration or Monitor by IR

Breakthrough Monitoring– 9 mg/M3 (as NO2) or 30 mg/M3 (as NO)– Monitor by Electrochemical Monitors

Beyond Breakthrough (9/30 mg/M3) – Interrupt Challenge and Monitor by FID

Page 37: AIHce 2010 - Design of a Test Laboratory for Chemical ...€¦ · Design of a New Laboratory for Chemical Challenge Testing of Respirator Cartridges, Canisters, and Filters C.R. (Gus)

Hydrogen Cyanide Testing

Challenge Level at 4000 mg/M3 – Higher than NIOSH tests– Verify by Titration

Breakthrough Monitoring– 5 mg/M3 HCN and 5 mg/M3 (CN)2– Monitor by Colorimetric Tape & FID

Beyond Breakthrough ( up to 50 mg/M3) – Monitor by FID