Albumin Analysis by Potentiometric Method
Albumin concentration
(ppm)
Epa Epc Ipa Ipc
(V vs. Ag/AgCl(0.5MKCl)) (A)
0 0.291 0.070 53.87 -40.42
0 0.292 0.070 49.96 -47.43
1 0.445 -0.017 31.69 -34.26
2 - - - -
4 - - - -
6 - - - -
8 - - - -
10 - - - -
Table : The relationship of peak current and albumin concentration
Fig 2. The cyclic voltammogram of the selectivity
WE:Au wire, WE area: 0.0113 cm2, RE: Ag/AgCl(0.5M), CE: Pt, Electrolyte: 0.5 M KCl 50 ml(20 mM K4Fe(CN)6+20 mM K3Fe(CN)6), Potentiostat: CHI-405, Scan range: -0.2~0.6 V, scan rate: 0.05 V/sec
-0.2 0 0.2 0.4 0.6Potential (V vs. Ag/AgCl(0.5M KCl))
- 8 0
- 4 0
0
4 0
8 0
Cu
rren
t (
A)
nonealbuminEPHADANEP
Epa Epc Ipa Ipc
(V vs. Ag/AgCl(0.5MKCl)) (A)
None 0.275 0.109 54.51 -48.12
5ppm Albumin 0.246 0.153 17.89 -16.89
0.0025ppm EP 0.246 0.139 88.08 -64.24
3.45ppm HA 0.259 0.131 60.88 -55.65
1.125ppm DA 0.246 0.124 74.18 -55.79
0.015ppm NEP 0.252 0.093 70.79 -43.68
Table Ⅱ : The relationship of peak current and albumin concentration
Fig 3. The cyclic voltammogram of Au wire electrode
WE: Au wire, WE area: 0.0113 cm2, RE: Ag/AgCl(0.5M), CE: Pt, Electrolyte: 0.5 M KCl 50 ml(20 mM K4Fe(CN)6+20 mM K3Fe(CN)6), Potentiostat: EG&G, Scan range: -0.2~0.6 V, scan rate: 0.02 V/sec,albumin 在電解液中加入
-0.2 0 0.2 0.4 0.6Potential (V vs. Ag/AgCl(0.5M KCl))
- 4 0
- 2 0
0
2 0
4 0
Cu
rren
t (
A)
album in concentration0 ppm1 ppm2 ppm
Albumin concentration
(ppm)
Epa Epc Ipa Ipc
(V vs. Ag/AgCl(0.5M)) (A)
0 ppm 0.248 0.142 36.70 -36.90
1 ppm 0.236 0.126 37.30 -32.70
2 ppm 0.234 0.130 39.90 -32.70
Table Ⅲ : The relationship of peak current and albumin concentration