30
Ambulatory ECG Monitoring Gill Bouncir Blackpool Victoria Hospital

Ambulatory ECG Monitoring€¦ · •Holter – templates with manual editing – time series with manual adjustments. STORAGE DATA • Analogue - on magnetic tapes – Disadvantage

  • Upload
    others

  • View
    6

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Ambulatory ECG Monitoring€¦ · •Holter – templates with manual editing – time series with manual adjustments. STORAGE DATA • Analogue - on magnetic tapes – Disadvantage

Ambulatory ECG Monitoring

Gill Bouncir

Blackpool Victoria Hospital

Page 2: Ambulatory ECG Monitoring€¦ · •Holter – templates with manual editing – time series with manual adjustments. STORAGE DATA • Analogue - on magnetic tapes – Disadvantage

History

• Developed in late 1950’s as a result of research of Dr Norman Holter

Page 3: Ambulatory ECG Monitoring€¦ · •Holter – templates with manual editing – time series with manual adjustments. STORAGE DATA • Analogue - on magnetic tapes – Disadvantage

TYPES

• Holter - continuous recording.

• Event recorder- patient activated.

- loop / non loop

• Implantable loop recorders.

Page 4: Ambulatory ECG Monitoring€¦ · •Holter – templates with manual editing – time series with manual adjustments. STORAGE DATA • Analogue - on magnetic tapes – Disadvantage

ANALYSIS.

• Event recordings– short duration

– automated unnecessary

• Holter– templates with manual editing

– time series with manual adjustments

Page 5: Ambulatory ECG Monitoring€¦ · •Holter – templates with manual editing – time series with manual adjustments. STORAGE DATA • Analogue - on magnetic tapes – Disadvantage

STORAGE DATA

• Analogue - on magnetic tapes– Disadvantage : tape and capstan speed

fluctuation

• Digital - on solid memory cards

Page 6: Ambulatory ECG Monitoring€¦ · •Holter – templates with manual editing – time series with manual adjustments. STORAGE DATA • Analogue - on magnetic tapes – Disadvantage

INDICATIONS

• Investigation symptoms

• CAD therapy

• Investigation known arrhythmias and arrhythmogenic conditions

• post open heart surgery

• Evaluation therapy

• Research

Page 7: Ambulatory ECG Monitoring€¦ · •Holter – templates with manual editing – time series with manual adjustments. STORAGE DATA • Analogue - on magnetic tapes – Disadvantage

CORRELATE ECG FINDINGS WITH SYMPTOMS

Page 8: Ambulatory ECG Monitoring€¦ · •Holter – templates with manual editing – time series with manual adjustments. STORAGE DATA • Analogue - on magnetic tapes – Disadvantage

ARRHYTHMIAS CAUSE DIFFERENT SYMPTOMS

• Palpitations

• Dizziness LOC– Hypotension and poor cerebral flow

• Chest pain– poor coronary flow

Page 9: Ambulatory ECG Monitoring€¦ · •Holter – templates with manual editing – time series with manual adjustments. STORAGE DATA • Analogue - on magnetic tapes – Disadvantage

PALPITATIONS

• Most frequent cardiac symptom

• 40% due to arrhythmia

• Daily occurrence– 24 or 48 hour holter

Page 10: Ambulatory ECG Monitoring€¦ · •Holter – templates with manual editing – time series with manual adjustments. STORAGE DATA • Analogue - on magnetic tapes – Disadvantage

PALPITATIONS - ASSOCIATED RHYTHMS

• Sinus tachycardia

• Atrial tachycardias

• Reentrant tachycardias– AVNRT, AVRT

• Ventricular tachycardia

• Ectopic beats

Page 11: Ambulatory ECG Monitoring€¦ · •Holter – templates with manual editing – time series with manual adjustments. STORAGE DATA • Analogue - on magnetic tapes – Disadvantage

SYNCOPE

• Holter– low diagnostic rate

• large range of causes– List as many causes as possible?

• Infrequent symptoms

Page 12: Ambulatory ECG Monitoring€¦ · •Holter – templates with manual editing – time series with manual adjustments. STORAGE DATA • Analogue - on magnetic tapes – Disadvantage

CAD THERAPY

• ST analysis– facilitate treatment

– show silent ischaemia

Page 13: Ambulatory ECG Monitoring€¦ · •Holter – templates with manual editing – time series with manual adjustments. STORAGE DATA • Analogue - on magnetic tapes – Disadvantage

INVESTIGATION OF KNOWN ARRHYTHMIAS AND ARRHYTHMOGENIC CONDITIONS

• Cardiomyopathies

• WPW syndrome

• Prolonged Q-T syndrome

• Intermittent conduction defects

• Abnormal resting ECG’s

• VE’s - risk stratification

Page 14: Ambulatory ECG Monitoring€¦ · •Holter – templates with manual editing – time series with manual adjustments. STORAGE DATA • Analogue - on magnetic tapes – Disadvantage

POST OPEN HEART SUGERY

• To exclude conduction system damage

• To assess post op chest pain

Page 15: Ambulatory ECG Monitoring€¦ · •Holter – templates with manual editing – time series with manual adjustments. STORAGE DATA • Analogue - on magnetic tapes – Disadvantage

EVALUATION THERAPY• Antiarrhythmic drugs• Antianginal drugs• AF rate control

– to assess baseline daytime rate and exercise response

• Pacemakers– to document functions and malfunctions – to evaluate underlying rhythm– to evaluate symptoms

Page 16: Ambulatory ECG Monitoring€¦ · •Holter – templates with manual editing – time series with manual adjustments. STORAGE DATA • Analogue - on magnetic tapes – Disadvantage

RESEARCH

• Identifcation of sudden death at risk groups

• Evaluation of new drugs

• RR variability

Page 17: Ambulatory ECG Monitoring€¦ · •Holter – templates with manual editing – time series with manual adjustments. STORAGE DATA • Analogue - on magnetic tapes – Disadvantage

PAEDIATRIC INDICATIONS

• Any rhythm abnormality

• WPW

• Non specific chest pain and palpitations

• Post open heart surgery

• Familial cot deaths

• Assessment treatment drugs

• Familial Conditions– Long Q-T syndrome

Page 18: Ambulatory ECG Monitoring€¦ · •Holter – templates with manual editing – time series with manual adjustments. STORAGE DATA • Analogue - on magnetic tapes – Disadvantage

RANGE OF NORMALITY -GENERALLY ACCEPTED AS NORMAL

• Sinus arrhythmia

• SVE’s

• Incomplete RBBB

• High Take off ST segment

• T wave inversion in lead III but not in aVF

• T wave inversion in aVR and V1

Page 19: Ambulatory ECG Monitoring€¦ · •Holter – templates with manual editing – time series with manual adjustments. STORAGE DATA • Analogue - on magnetic tapes – Disadvantage

RANGE OF NORMALITY -NOT NECESSARILY INDICATIVE OF HEART DISEASE

• VE’s

• RBBB

• T wave inversion in other leads

• Non- specific ST changes

Page 20: Ambulatory ECG Monitoring€¦ · •Holter – templates with manual editing – time series with manual adjustments. STORAGE DATA • Analogue - on magnetic tapes – Disadvantage

RANGE OF NORMALITY -NOCTURNAL OR DURING SLEEP,IN ATHLETES OR FIT YOUNG PEOPLE.

• Wenckebach

• Sinus bradycardia

• Sinus arrest

• Sinus node exit block

Page 21: Ambulatory ECG Monitoring€¦ · •Holter – templates with manual editing – time series with manual adjustments. STORAGE DATA • Analogue - on magnetic tapes – Disadvantage

AMBULATORY MONITORING -PATIENT PREPARATION• Accuracy of analysis depends upon quality

of data recorded

• Good preparation and patient instruction are paramount

• Good electrical contact must be obtained and secured

Page 22: Ambulatory ECG Monitoring€¦ · •Holter – templates with manual editing – time series with manual adjustments. STORAGE DATA • Analogue - on magnetic tapes – Disadvantage

SKIN PREPARATION

• Determines the conductivity at electrode and skin contact point

• Reduces amount artefact and lost data

Page 23: Ambulatory ECG Monitoring€¦ · •Holter – templates with manual editing – time series with manual adjustments. STORAGE DATA • Analogue - on magnetic tapes – Disadvantage

SKIN PREPARATION• To make area clean and dry

• Shave all body hair in area

• Use abrasive material to remove layer of dead skin cells, without causing irritation or bleeding

• Wipe with alcohol to remove dead skin and skin oils

• Dry area with gauze pad

Page 24: Ambulatory ECG Monitoring€¦ · •Holter – templates with manual editing – time series with manual adjustments. STORAGE DATA • Analogue - on magnetic tapes – Disadvantage

ELECTRODE ATTACHMENT, AND SECURING• Attach lead wires to electrodes

• Place electrodes in position

• Secure by pressing around the OUTER adhesive edge

• DO NOT press wet gel electrodes over gelled area to cause gel expulsion under adhesive

Page 25: Ambulatory ECG Monitoring€¦ · •Holter – templates with manual editing – time series with manual adjustments. STORAGE DATA • Analogue - on magnetic tapes – Disadvantage

ELECTRODE PLACEMENT - to obtain most significant data.• Precordial modified leads - bipolar

• Modified V1– identification of origin of ventricular ectopic

activity

– easier SVE with aberration vs. VE identification

– more visible f or FL waves

Page 26: Ambulatory ECG Monitoring€¦ · •Holter – templates with manual editing – time series with manual adjustments. STORAGE DATA • Analogue - on magnetic tapes – Disadvantage

• Modified V5– displays identifiable P waves and large QRS deflections

– mainly see lateral ST changes

• Modified V3– LAD associated

– ST changes

• Seven lead cable– identification localise areas of ischaemic episodes,

including silent ischaemia

Page 27: Ambulatory ECG Monitoring€¦ · •Holter – templates with manual editing – time series with manual adjustments. STORAGE DATA • Analogue - on magnetic tapes – Disadvantage

E.C.G. SIGNAL TEST• To check quality of ECG signal and recorder

performance

• Check calibration– 1mv=10mm

• Check signal– R wave amplitude MUST BE >9mm V5 and >4mm V1

– to increase amplitude - reposition electrodes closer to sternum

• Eliminate artefact as will intensify during recording

– Re-prepare or re-position electrodes

Page 28: Ambulatory ECG Monitoring€¦ · •Holter – templates with manual editing – time series with manual adjustments. STORAGE DATA • Analogue - on magnetic tapes – Disadvantage

HOOK UP COMPLETION

• Secure electrodes with tape to maintain electrical contact

• Stress loops should be one to two inches diameter, and two inches from the electrode

Page 29: Ambulatory ECG Monitoring€¦ · •Holter – templates with manual editing – time series with manual adjustments. STORAGE DATA • Analogue - on magnetic tapes – Disadvantage

PATIENT DIARY AND INSTRUCTION

• Patient co-operation with documentation of symptoms, events and activities is very important

Page 30: Ambulatory ECG Monitoring€¦ · •Holter – templates with manual editing – time series with manual adjustments. STORAGE DATA • Analogue - on magnetic tapes – Disadvantage

DEFINITIONS

• Wet Gel electrodes– gel that is semi liquid at room temperature

• Dry Gel electrodes– gel that is usually solid at room temperature