Upload
trananh
View
224
Download
6
Embed Size (px)
Citation preview
TO DOWNLOAD A COPY OF THIS POSTER, VISIT WWW.WATERS.COM/POSTERS ©2014 Waters Corporation
INTRODUCTION
Fourier Transform (FT) mass spectrometers
have ultimate resolutions exceeding those of
conventional Time-of-flight (TOF) mass
spectrometers. However, in order to achieve
this, long acquisition times are required (of up
to several seconds in some cases). In contrast,
high-performance TOF mass spectrometers have
typical flight times of tens of microseconds. This
allows them, for example, to profile mobility
separations lasting only a few milliseconds.
In this poster we present a theoretical
treatment of a novel orthogonal acceleration
(oa)-TOF geometry, utilising the quadro-
logarithmic potential familiar from Orbitrap
mass spectrometry, which could combine the
high resolution of FT instruments with the speed
of TOF analysis.
AN INVESTIGATION OF THE SPACE OF TRAJECTORIES IN A NOVEL OA-TOF GEOMETRY
Keith Richardson and John Hoyes Waters Corporation, Wilmslow, UK
METHODS The cylindrically symmetric quadro-logarithmic potential takes
the form:
This potential produces independent motion in the radial
(trapping) and axial (mass analysis) directions. The axial
component is quadratic, leading to simple harmonic motion
with a frequency that is m/z dependent but energy
independent. In the Orbitrap1 mass spectrometer, the chosen
shape of the electrodes fixes the ratio between the logarithmic
(b0) and quadratic (k) parts of the potential. We propose
instead to use a segmented construction allowing freedom of
choice of this ratio. The method of construction (Figures 1 and
2) consists of two concentric sets of N rings to which a set of
voltages Vo(n) and Vi(n) are applied. A typical device geometry
would be a length of around 1m with an outer radius of 10cm
and an inner radius of 2cm.
References
1. A. Makarov, Anal. Chem 72 (6) pp 1156-1162 (2000).
2. J.M. Brown, A.J. Gilbert, J.B. Hoyes, D.J. Langridge, and J.L. Wildgoose, WO 2011/154731 A1.
3. See, for example, H. Goldstein, Classical Mechanics, Addison Wesley, (1980).
The parameter space (R0,b0) of trajectories for a device with
k=8x104Vm-2 and ΔU=1000V is shown in Figure 4 along with
examples of effective potentials corresponding to the points
labeled a)-f). To obtain a stable bound orbit, the starting
(injection radius) must lie in the range RE < R0 < RS+ where RE
is the only non-trivial solution of U(RE)=U(RS+ ). This
corresponds to the blue shaded region on the large plot.
Above the straight red line the effective potential has a
minimum, while trajectories lying along the brown quadratic
curve are circular: R=RS- (stable) or R=RS+ (unstable). Above
this curve, the trajectory starts at a radial maximum R0=Rmax.
In Figure 5, further constraints on this space are explored.
1) Rmin<RD<Rmax so that ions can repeatedly miss then
ultimately hit the detector.
2) Rmin>Rin so that ions do not hit the inner electrodes.
CONCLUSION Following axial beam expansion, the quadro-
logarithmic potential can be utilized for multi
pass oa-TOF mass analysis.
The parameter space has been classified for a
chosen geometry, and a solution with an effective
TOF path length of over 10m has been identified.
Many technical challenges remain to be
addressed before such a device could be
constructed.
DISCUSSION In this poster we have briefly examined the theoretical
properties of a novel multi pass orthogonal time of flight mass
analyser utilizing a quadro-logarithmic potential. Many
practical problems (both electrical and mechanical) would have
to be overcome before such a device could be constructed:
each of the ring electrodes in the assembly must be carefully
machined and positioned, to achieve orthogonal acceleration
voltages must be pulsed accurately an rapidly from their initial
values to their final values, the detector must be placed within
the analyser field without destroying isochronicity and the
number of passes that can be utilized is limited by the initial
radial spread of the ions. Nevertheless with appropriate beam
conditioning, it is anticipated that mass resolutions of several
hundred thousand should be achievable in a relatively compact
geometry.
Figure 1. Construction of the quadro-logarithmic oa-TOF with equipotential lines for an Orbitrap superimposed for purposes of illustration only. The segmentation of the inner electrodes is not shown.
Figure 2. Ions are introduced through an aperture in the outer electrodes and a switched electric sector helps to accomplish stable circular trajecto-ries. The annular detector is shown in green.
Figure 5. Constraints: k=8x104Vm-2, ΔU=1000V, RD=3.9cm, Rin=2cm. The white region corresponds to potentially useful trajectories with
Rmin<RD<Rmax. This roughly corresponds to 2.6<ρ<5.2.
The sequence of events for a single injection is as follows:
1. Axial (z) beam expansion2 prior to injection to reduce turn-around time in low extraction field ~20 kV/m.
2. Ion injection into stable circular orbits with r=R0 and k=0
(Figure 2).
3. Orthogonal and radial acceleration (k > 0 and increased b0). 4. Unique path, multipass time of flight in eccentric orbits be-
tween radii r=Rmax and r=Rmin with simple harmonic axial
motion (Figure 3). Ions repeatedly miss an annular detec-
tor situated at the isochronous plane (z=0). 5. Ions strike the detector on the final pass.
In order to produce this behaviour, it is important to control
the interplay between axial and orbital motion. The equations of motion are derived below starting from the Lagrangian3
written in cylindrical polar coordinates:
The three Euler-Lagrange equations are
where qi stands in turn for the coordinates r,ϕ and z:
The second equation simply expresses conservation of the z-
component of angular momentum Lz, and the first equation
may therefore be rewritten in the form of motion in an effec-tive potential containing a “centrifugal” term:
For ions injected into circular orbits radius R0 through a poten-
tial drop ΔU, Lz2 = 2mq R0
2 ΔU and we find
(1)
(2)
This potential has extrema at RS– and RS+ satisfying
RESULTS The initial stable circular orbits are obtained by setting k=0
and b0=2ΔU. Following orthogonal acceleration, Eq. (1) pro-
duces simple harmonic axial motion with period Tz=2π / ωz
where ωz= √(qk/m).
From Eq. (2) the period of the radial motion is Tr=2√(m/q)τ where
Rmax
Rmin
3) Ions will hit the detector when r=Rmin and z=0 simultane-
ously. This occurs for pairs of positive integers (i,j) satisfying
We therefore look for points near which condition 3) is satisfied
for j=N but away from where it is satisfied for 1≤j<N.
Conditions 1) and 2) correspond roughly to 2.6<ρ<5.2, but
most useful solutions lie towards the lower end of this range.
The example in Figure 3 corresponds to the solution (i,j)=
(19,7) or ρ=2.85 for which ions hit the detector after 18½ ra-
dial periods and 6 detector passes. The time of flight for an
ion of m/z 500 is just under 165μs and the effective TOF path
length is just over 10m. Figure 3. Trajectory with k=8x104Vm-2, ΔU=1000V , R0=0.065m, b0=3700V. The dot indicates the start of the trajectory. The detector extends from r=0.02m to RD=0.0385m.
r =R0=Rmax
r =Rmin
RD
RD
a)
Ions escape
b)
f) e) d)
c)
Ions escape
Bound orbit
Stable circular orbit Bound orbit Ions escape
Figure 4. Trajectory classification and examples of effective potentials for k=8x104Vm-2, ΔU=1000V: a) R0=0.1m, b0=1300V, b) R0=0.1m, b0=1900V c) R0=0.1m, b0=1990V, d) R0=0.1m, b0=2400V, e) R0=0.1m, b0=3590V, f) R0=0.1m, b0=3590V. The shaded region corresponds to bound trajectories. Orbits satisfying the requirements of the analyser discussed here are of types c) and e).