The active element of an accelerometer is a piezoelectric material. Figure 1 illustrates the piezoelectric effect with the
help of a compression disk. A compression disk looks like a capacitor with the piezoceramic material sandwiched be-
tween two electrodes. A force applied perpendicular to the disk causes a charge production and a voltage at the elec-trodes.
electrode areathicknessforcechargevoltage
piezo constants
AdFqu
d33, e33
piezo disk
q = d33 F
u = Fd33 d
e33 A
A F
q
u
F
d
Figure 1: Piezoelectric effect, basic calculations
The sensing element of a piezoelectric accelerometer consists of two major parts:
• Piezoceramic material
• Seismic mass
One side of the piezoelectric material is connected to a rigid post at the sensor base. The so-called seismic mass is
attached to the other side. When the accelerometer is subjected to vibration, a force is generated which acts on the pie-
zoelectric element (compare Figure 2). According to Newton’s Law this force is equal to the product of the acceleration
and the seismic mass. By the piezoelectric effect a charge output proportional to the applied force is generated. Since
the seismic mass is constant the charge output signal is proportional to the acceleration of the mass.
F = m a
charge sensitivity:
Bqa =qa
Bua =ua
voltage sensitivity:Acceleration a
m
Seismic mass
Piezoceramics
q
u
.
Figure 2: Principle of a piezoelectric accelerometer
Over a wide frequency range both sensor base and seismic mass have the same acceleration magnitude. Hence, the
sensor measures the acceleration of the test object.
The piezoelectric element is connected to the sensor socket via a pair of electrodes. Some accelerometers feature an
integrated electronic circuit which converts the high impedance charge output into a low impedance voltage signal (see
Application Note AN4E).
8/8/2019 An2e Piezoelectric Principle
http://slidepdf.com/reader/full/an2e-piezoelectric-principle 2/2
AN2EApplication Note
Piezoelectric Accelerometers
Piezoelectric Principle
Metra Mess- und Frequenztechnik Radebeul Page 2
Meissner Str. 58 P.O. Box 01 01 13 Oct. 03
D-01445 Radebeul D-01435 Radebeul Internet: www.MMF.de
Phone +49-351-836 21 91 Fax +49-351-836 29 40 Email: [email protected]
Within the useful operating frequency range the sensitivity is independent of frequency.
A piezoelectric accelerometer can be regarded as a mechanical low-pass with resonance peak. The seismic mass and the
piezoceramics (plus other “flexible” components) form a spring mass system. It shows the typical resonance behavior
and defines the upper frequency limit of an accelerometer. In order to achieve a wider operating frequency range the
resonance frequency should be increased. This is usually done by reducing the seismic mass. However, the lower the
seismic mass, the lower the sensitivity. Therefore, an accelerometer with high resonance frequency, for example a
shock accelerometer, will be less sensitive whereas a seismic accelerometer with high sensitivity has a low resonance
frequency.
Figure 3 shows a typical frequency response curve of an accelerometer when it is excited by a constant acceleration.
f0 fr0.5fr0.3fr
0.2fr ffL 2fL3fL
0.900.951.001.051.10
1.30
0.71
fLf
0
fr
lower frequency limitcalibration frequency
resonance frequency
Figure 3: Frequency response curve
Several useful frequency ranges can be derived from this curve:
•At approximately 1/5 the resonance frequency the response of the sensor is 1.05. This means that the measured error
compared to lower frequencies is 5 %.
•At approximately 1/3 the resonance frequency the error is 10 %. For this reason the “linear” frequency range should
be considered limited to 1/3 the resonance frequency.
•The 3 dB limit with approximately 30 % error is obtained at approximately one half times the resonance frequency.
The lower frequency limit mainly depends on the chosen preamplifier. Often it can be adjusted. With voltage amplifiers
the low frequency limit is a function of the RC time constant formed by accelerometer, cable, and amplifier input ca-
pacitance together with the amplifier input resistance (see Application Note AN6E).