Anaesthesia Breathing Circuits (EORCAPS 2009)

Embed Size (px)

Citation preview

  • 7/24/2019 Anaesthesia Breathing Circuits (EORCAPS 2009)

    1/9

    Dr. A K Sethis EORCAPS-20

    Dr. A K SethisEORCAPS-2009

    Dr. Pradeep Jain

    Anaesthesia Breathing Circuits

    Dr. A K Sethis EORCAPS-2009

    Breathing System

    A breathing system is defined as an assembly of

    components which connects the patients airway to the

    anaesthetic machine creating an artificial atmosphere, from

    and into which the patient breathes

    Link machine to the patient

    Eliminate CO2

    Dr. A K SethisEORCAPS-2009

    Component of Breathing system A fresh gas entry port/delivery tube through which the gases

    are delivered from the machine to the systems

    A port to connect it to the patients airway

    A reservoir for gas, in the form of a bag or a corrugated tube

    to meet the peak inspiratory flow requirements

    An expiratory port/valve through which the expired gas is

    vented to the atmosphere

    A CO2 absorber if total rebreathing is to be allowed

    Corrugated tubes for connecting these components

    Dr. A K Sethis EORCAPS-2009

    Requirements of

    Breathing SystemEssential:The breathing system must

    Have minimal apparatus dead space

    Have low resistance

    Deliver the gases from the machine to the alveoli in the

    same concentration as set and in the shortest possible time

    Effectively eliminate CO2

    Dr. A K SethisEORCAPS-2009

    Requirements of

    A Breathing SystemDesirable:The desirable requirements are

    Economy of fresh gas

    Conservation of heat

    Adequate humidification of inspired gas

    Light weight

    Convenience during use

    Efficiency during spontaneous as well as controlled ventilation

    Adaptability for adults, children and mechanical ventilators

    Provision to reduce theatre pollution

    Dr. A K Sethis EORCAPS-2009

    History

    McMohan 1951 classified them as open, semi closed &

    closed - criteria rebreathing

    Dripps 1984 - Insufflation, Open, Semi open, Semi closed

    & Closed - criteria presence or absence of Reservoir,

    Rebreathing, CO2 absorption & Directional valves

    Conway 1985 suggested functional classification according

    to the method used for CO2 elimination

    Breathing systems with CO2 absorber

    Breathing systems without CO2 absorber

    Miller 1995 new breathing system called Maxima

  • 7/24/2019 Anaesthesia Breathing Circuits (EORCAPS 2009)

    2/9

    Dr. A K Sethis EORCAPS-20

    Dr. A K SethisEORCAPS-2009

    Classification of Breathing Systems

    Bi-directional flowTo and Fro system.

    Bi-directional flow:a) Afferent reservoir systems.

    Mapleson A

    Mapleson B

    Mapleson C

    Lacks system.

    B) Enclosed afferent reservoir systems

    Millers (1988)

    c) Efferent reservoir systems

    Mapleson D

    Mapleson E

    Mapleson F

    Bains system

    d) Combined systems

    Humphrey ADE

    Unidirectional flow

    Circle system with absorber.

    Unidirectional flow:

    a. Non rebreathing systems.

    Breathing Systems With Co2Absorption.Breathing Systems Without Co

    2Absorption.

    Dr. A K Sethis EORCAPS-2009

    Non Rebreathing System

    Non Rebreathing Valves - Unidirectional Flow

    FGF has to be constantly adjusted, not economical

    No humidification of inspired gas

    No conservation of heat

    Inconvenient as the bulk of the valve near the patient

    Valve malfunction due to condensation of moisture

    Eg. Ruben valve, Ambu valve & Laerdal valve

    Dr. A K SethisEORCAPS-2009

    Four Basic Circuits

    Open

    Semi-open

    Semi-closed

    Closed

    Dr. A K Sethis EORCAPS-2009

    Open System

    Insufflation

    Blow anesthetic gas over face

    No direct contact

    No rebreathing of gases

    Ventilation cannot be controlled

    Unknown amount delivered

    Dr. A K SethisEORCAPS-2009

    Open Systems

    Open drop anesthesia

    Gauze covered wire mask

    Anesthetic dripped

    Inhaled air passes through gauze & picks up anesthetic

    Concentration varies

    Environmental pollution

    Dr. A K Sethis EORCAPS-2009

    Semi-open Systems

    Reservoir

    No rebreathing

  • 7/24/2019 Anaesthesia Breathing Circuits (EORCAPS 2009)

    3/9

    Dr. A K Sethis EORCAPS-20

    Dr. A K SethisEORCAPS-2009

    Semi Closed

    Mapleson Breathing System Absence of unidirectional value to direct gases to and fro

    from the patient No device for absorbing CO2

    Flow controlled breathing systems

    FGF must wash CO2 out of the circuit

    CO2 without circuit

    No clear separation of inspired and expired gases -rebreathing will occur when inspiratory flow exceeds FGF

    Composition of inspired mixture will depend uponrebreathing

    Dr. A K Sethis EORCAPS-2009

    Bi-Directional Flow

    Systems extensively used

    Depend on the FGF for effective CO2 elimination

    Functions can be manipulated by changing

    parameters like

    Fresh Gas Supply

    Elimination of CO2

    Apparatus Dead Space

    Dr. A K SethisEORCAPS-2009

    Bi-Directional Flow

    Fresh Gas Supply; Low do not eliminate CO2 effectively

    High wastage

    If the system has to deliver a set concentration in the shortest possible timeto the alveoli, the FGF should be delivered as near the patients airway aspossible

    Elimination of CO2 Normal production of CO2 in a 70 kg patient 200 ml/min

    End-tidal concentration of CO2 is 5%

    For elimination 200ml of of CO2 as a 5% gas mixture, the alveolarventilation has to be: 200 x 100 = 4,000 ml.

    5

    4 litres is the normal alveolar ventilation.

    Apparatus Dead Space

    It is the volume of the breathing system from the patient-end to the point upto which, to and fro movement of expired gas takes place

    Dr. A K Sethis EORCAPS-2009

    Classification

    Five Basic types

    A through E

    F added by willis in 1975

    Dr. A K SethisEORCAPS-2009

    Mapleson Classification

    Reservoir bag

    Corrugated tubing

    APL valve

    FGI

    Patient connection

    Dr. A K Sethis EORCAPS-2009

    Mapleson A

    Magill Circuit

    Popularized by Sir Ivan Magill in 1920s

    It differs from the other Mapleson systems in that fresh gas does notenter the system near the patient connection but enters at the other

    end of the system near the reservoir bag.

    A corrugated tubing connects the bag to the adjustable pressure

    limiting (APL) valve at the patient end of the system

    Length of corrugated breathing tube 110 cm with an internal volume

    of 550 ml

  • 7/24/2019 Anaesthesia Breathing Circuits (EORCAPS 2009)

    4/9

    Dr. A K Sethis EORCAPS-20

    Dr. A K SethisEORCAPS-2009

    Mapleson A

    Most efficient for spontaneous respiration

    Resistance 2.5cm of H2O when expiratory valve is fully open

    No rebreathing if FGF equal to minute volume

    Rebreathing occurs if FGF less than alveolar ventilation

    Dr. A K Sethis EORCAPS-2009

    Functional Analysis

    Mapleson A with Spontaneous Breathing

    Dr. A K SethisEORCAPS-2009

    Functional Analysis

    Mapleson A with Control Ventilation

    a = at the end of inspiration

    b = at the end of expiration

    c = during subsequent inspiration

    d = at the end of subsequent inspiration

    Dr. A K Sethis EORCAPS-2009

    Mapleson B & C systems

    In order to reduce the rebreathing of alveolar gas and to

    improve the utilization of FG during controlled

    ventilation, the FG entry was shifted near the patient.

    This allows a complete mixing of FG and expired gas.

    The end result is that these systems are neither efficient

    during spontaneous nor during controlled ventilation.

    Dr. A K SethisEORCAPS-2009

    Modifications of Mapleson A Circuit

    Lack Circuit Added expiratory limb which runs from pt end to APL

    valve at machine end

    This limb is coaxially placed inside the affarent limb

    Length 1.5 m, resistance 1.63 cm H2O Advantage

    Facilitates easy scavenging

    Easy adjustment of APL valve

    Disadvantage

    Increased resistance to breathing

    Dr. A K Sethis EORCAPS-2009

    Enclosed Afferent Reservoir System

    Miller & Miller 1988

    Mapleson A system enclosed within a non

    distensible structure

    It may also be constructed by enclosing the

    reservoir bag alone in a bottle and connecting

    the expiratory port to the bottle with a

    corrugated tube and a one way valve

    To the bottle is also attached a reservoir bag

    and a variable orifice for providing positive

    pressure ventilation.

  • 7/24/2019 Anaesthesia Breathing Circuits (EORCAPS 2009)

    5/9

    Dr. A K Sethis EORCAPS-20

    Dr. A K SethisEORCAPS-2009

    Enclosed Afferent Reservoir System

    Functional analysis:

    During spontaneous ventilation it is identical to the Mapleson A system

    In this mode the variable orificeis kept widely open to allow free communicationto the atmosphere

    In controlled ventilation the reservoir bag B is squeezed intermittently & the

    variable orificeis partly closed to allow building up of pressure in the bottle

    The pressure thus developed closes the expiratory valve squeezes the enclosed

    afferent reservoir and the patient gets ventilated

    Hence this system should function efficiently during spontaneous and

    controlled ventilation with a FGF equivalent to alveolar ventilation

    Dr. A K Sethis EORCAPS-2009

    Efferent Reservoir Systems

    Mapleson DEF All have a T-piece near the patient and

    function similarly

    The T-piece is a three-way tubular

    connector with a patient connection port,

    a FGF port & a port for connection to a

    corrugated tubing

    Light metal tube 1cm in diameter ,5cm in

    length with a side arm

    6 mm tube as the affarent limb that supplies

    the FG from the machine

    Dr. A K SethisEORCAPS-2009

    Mapleson D

    A length of tubing connects the T-piece at the patient end

    to the APL valve and the reservoir bag adjacent to it

    The length of the tubing determines the distance the user

    can be from the patient but has minimal effects on

    ventilation

    Popular because excess gas scavenging is relatively easy

    Most efficient of the Mapleson systems during controlled

    ventilation

    Dr. A K Sethis EORCAPS-2009

    Mapleson D Functional Analysis

    Spontaneous Ventilation

    Dr. A K SethisEORCAPS-2009

    Mapleson D Functional Analysis

    Controlled Ventilation

    Dr. A K Sethis EORCAPS-2009

    Bain Breathing Circuit

    Modified Mapleson D

    Bain 1972

    The fresh gas supply tube runs coaxially inside the corrugated tubing

    and ends at the point where the fresh gas would enter if the classicMapleson D form were used

    The outer tube is clear so that the inner tube can be inspected

    The outer tubing of most commercially available versions of the

    Bain system is narrower than conventional corrugated tubing

    A long version of the Bain system may be used for remote

    anesthesia in locations such as MRI

  • 7/24/2019 Anaesthesia Breathing Circuits (EORCAPS 2009)

    6/9

    Dr. A K Sethis EORCAPS-20

    Dr. A K SethisEORCAPS-2009

    Bain System

    Depends on fresh gas flow to flush out CO2

    Spontaneous ventilation

    200 - 300 ml / kg / min

    Controlled ventilation

    infants 60 kg 70 ml/kg/min

    Dr. A K Sethis EORCAPS-2009

    Bain Circuit

    Coaxial length 1.8m

    Inner inspiratory tubing - 7 mm diameter

    Outer expiratory tubing - 22 mm diameter

    Resistance - < 0.7 cm H2O

    Dr. A K SethisEORCAPS-2009

    Relation Between Alveolar

    Ventilation and FGF

    Dr. A K Sethis EORCAPS-2009

    Bain System Advantages

    Light-weight, Compact, easy to handle

    Warming of inspired gases

    Partial rebreathing improves humidification

    Ability of scavenging

    Can be used with ventilator

    Dr. A K SethisEORCAPS-2009

    Disadvantage

    Accidental disconnection of inner tube, kinking & twisting of

    inner tube converts entire exhalation limb into dead space

    FGF may be connected to outer tube

    Lower FGF may result in re-breathing

    Tests-visual inspection for damage, disconnection

    Pethicks test, occlusion test

    Dr. A K Sethis EORCAPS-2009

    Mapleson E

    Ayres T Piece, 1937

    Low resistance, low dead space

    Simple piece of metal with 3 limbs & a port at the end of

    each limb Exhalation tube is a reservoir

    No bag

    FGF near mask

  • 7/24/2019 Anaesthesia Breathing Circuits (EORCAPS 2009)

    7/9

    Dr. A K Sethis EORCAPS-20

    Dr. A K SethisEORCAPS-2009

    Mapleson E

    Hazards

    Over inflation and barotrauma by prolonged occlusion

    and no feel of inflation

    Also pressure buffering effect of bag absent

    Dr. A K Sethis EORCAPS-2009

    Mapleson F

    Jackson Rees Modification

    Addition of a bag with an opening to exhalation limb of T- piece

    Corrugated hose 15cm X 11 mm

    Reservoir bag 0.5 1 L

    Green tube for FGF 1.5m in length

    Advantage

    Early to assist / Controlled ventilation

    Easy to monitor ventilation

    Dr. A K SethisEORCAPS-2009

    Advantage of the Mapleson Systems

    Simple, inexpensive, light weight. With the exception of the APL

    valve, there are no moving parts

    In coaxial systems (Lack, Bain), the inspiratory limb is heated by the

    warm exhaled gas in the coaxial expiratory tubing

    Resistance is usually low at flows likely to be experienced in practice

    Easy to position conveniently. A long Mapleson D system may be

    used to ventilate a patient in the MRI unit

    Compression & compliance volume losses are less than with the circle

    system

    Changes in fresh gas concentrations result in rapid changes in

    inspiratory gas composition

    Dr. A K Sethis EORCAPS-2009

    Disadvantage of the Mapleson Systems

    Require high gas flows - results in higher costs, increased

    atmospheric pollution, and difficulty assessing spontaneous

    ventilation

    Inspired heat and humidity tend to be low because of high FGF

    In the Mapleson A, B, and C systems the APL valve is located close

    to the patient, where it may be inaccessible to the user.

    The Mapleson E and F systems are difficult to scavenge

    Air dilution can occur with the Mapleson E system.

    Mapleson systems are not suitable for patients with malignant

    hyperthermia

    Dr. A K SethisEORCAPS-2009

    Combined System

    Humphrey ADE To over come the difficulties of changing the breathing system for

    different modes of ventilation

    Humphery designed a system called Humphery ADE in 1983

    Two reservoirs, one in the afferent limb and the other in the

    efferent limb

    Only one reservoir will be in operation and the system can be

    changed from ARS to ERS by changing the position of a lever

    Can be used for adults as well as children

    Functional analysis is the same as Mapleson A in ARS mode and

    as Bain in ERS mode

    Not yet widely used

    Dr. A K Sethis EORCAPS-2009

    Paediatric Circuit

    work of breathing

    Prevent rebreathing

    Volume of the non rebreathing circuit is small as

    compared with that of circuit system, the compression

    and compliance volume will be significantly less. This

    improves the ability to observe respiratory effect as

    reflected by movement of the anaesthesia bag

    Mapleson D recommended for use in children < 10 kg wt

  • 7/24/2019 Anaesthesia Breathing Circuits (EORCAPS 2009)

    8/9

    Dr. A K Sethis EORCAPS-20

    Dr. A K SethisEORCAPS-2009

    Questions ?

    Dr. A K Sethis EORCAPS-2009

    Which system have the FGF

    near the reservoir bag

    A

    Dr. A K SethisEORCAPS-2009

    Apparatus Dead Space extend

    upto

    In ARS Expiratory valve

    In ERS Fresh Gas entry

    Dr. A K Sethis EORCAPS-2009

    F circuit was introduced by

    Whom & When?

    Willis 1975

    Dr. A K SethisEORCAPS-2009

    Bain Circuit

    Coaxial length 1.8 meter

    Inner inspiratory tubing diameter 7 mm

    Outer expiratory tubing diameter 22 mm

    Resistance

  • 7/24/2019 Anaesthesia Breathing Circuits (EORCAPS 2009)

    9/9

    Dr. A K Sethis EORCAPS-20

    Dr. A K SethisEORCAPS-2009

    The efficiency of the circuit

    is determined in terms of

    CO2 elimination

    FGF utilization

    Dr. A K Sethis EORCAPS-2009

    Which system is most efficient with

    respect to prevention of rebreathing

    during Spontaneous Respiratory

    A > DEF > BC

    Dr. A K SethisEORCAPS-2009

    Which system is most efficient with

    respect to prevention of rebreathing

    during Controlled ventilation

    DEF > BC > A

    Dr. A K Sethis EORCAPS-2009

    To prevent rebreathing of CO2

    system DEF require a FGF of

    2.5 times the minute volume

    Dr. A K SethisEORCAPS-2009

    In spontaneous respiration to

    prevent rebreathing the

    Mapleson A requires flow rate

    Equal to minute volume