Upload
bijay-kc
View
368
Download
17
Embed Size (px)
Citation preview
7/24/2019 Anaesthesia Breathing Circuits (EORCAPS 2009)
1/9
Dr. A K Sethis EORCAPS-20
Dr. A K SethisEORCAPS-2009
Dr. Pradeep Jain
Anaesthesia Breathing Circuits
Dr. A K Sethis EORCAPS-2009
Breathing System
A breathing system is defined as an assembly of
components which connects the patients airway to the
anaesthetic machine creating an artificial atmosphere, from
and into which the patient breathes
Link machine to the patient
Eliminate CO2
Dr. A K SethisEORCAPS-2009
Component of Breathing system A fresh gas entry port/delivery tube through which the gases
are delivered from the machine to the systems
A port to connect it to the patients airway
A reservoir for gas, in the form of a bag or a corrugated tube
to meet the peak inspiratory flow requirements
An expiratory port/valve through which the expired gas is
vented to the atmosphere
A CO2 absorber if total rebreathing is to be allowed
Corrugated tubes for connecting these components
Dr. A K Sethis EORCAPS-2009
Requirements of
Breathing SystemEssential:The breathing system must
Have minimal apparatus dead space
Have low resistance
Deliver the gases from the machine to the alveoli in the
same concentration as set and in the shortest possible time
Effectively eliminate CO2
Dr. A K SethisEORCAPS-2009
Requirements of
A Breathing SystemDesirable:The desirable requirements are
Economy of fresh gas
Conservation of heat
Adequate humidification of inspired gas
Light weight
Convenience during use
Efficiency during spontaneous as well as controlled ventilation
Adaptability for adults, children and mechanical ventilators
Provision to reduce theatre pollution
Dr. A K Sethis EORCAPS-2009
History
McMohan 1951 classified them as open, semi closed &
closed - criteria rebreathing
Dripps 1984 - Insufflation, Open, Semi open, Semi closed
& Closed - criteria presence or absence of Reservoir,
Rebreathing, CO2 absorption & Directional valves
Conway 1985 suggested functional classification according
to the method used for CO2 elimination
Breathing systems with CO2 absorber
Breathing systems without CO2 absorber
Miller 1995 new breathing system called Maxima
7/24/2019 Anaesthesia Breathing Circuits (EORCAPS 2009)
2/9
Dr. A K Sethis EORCAPS-20
Dr. A K SethisEORCAPS-2009
Classification of Breathing Systems
Bi-directional flowTo and Fro system.
Bi-directional flow:a) Afferent reservoir systems.
Mapleson A
Mapleson B
Mapleson C
Lacks system.
B) Enclosed afferent reservoir systems
Millers (1988)
c) Efferent reservoir systems
Mapleson D
Mapleson E
Mapleson F
Bains system
d) Combined systems
Humphrey ADE
Unidirectional flow
Circle system with absorber.
Unidirectional flow:
a. Non rebreathing systems.
Breathing Systems With Co2Absorption.Breathing Systems Without Co
2Absorption.
Dr. A K Sethis EORCAPS-2009
Non Rebreathing System
Non Rebreathing Valves - Unidirectional Flow
FGF has to be constantly adjusted, not economical
No humidification of inspired gas
No conservation of heat
Inconvenient as the bulk of the valve near the patient
Valve malfunction due to condensation of moisture
Eg. Ruben valve, Ambu valve & Laerdal valve
Dr. A K SethisEORCAPS-2009
Four Basic Circuits
Open
Semi-open
Semi-closed
Closed
Dr. A K Sethis EORCAPS-2009
Open System
Insufflation
Blow anesthetic gas over face
No direct contact
No rebreathing of gases
Ventilation cannot be controlled
Unknown amount delivered
Dr. A K SethisEORCAPS-2009
Open Systems
Open drop anesthesia
Gauze covered wire mask
Anesthetic dripped
Inhaled air passes through gauze & picks up anesthetic
Concentration varies
Environmental pollution
Dr. A K Sethis EORCAPS-2009
Semi-open Systems
Reservoir
No rebreathing
7/24/2019 Anaesthesia Breathing Circuits (EORCAPS 2009)
3/9
Dr. A K Sethis EORCAPS-20
Dr. A K SethisEORCAPS-2009
Semi Closed
Mapleson Breathing System Absence of unidirectional value to direct gases to and fro
from the patient No device for absorbing CO2
Flow controlled breathing systems
FGF must wash CO2 out of the circuit
CO2 without circuit
No clear separation of inspired and expired gases -rebreathing will occur when inspiratory flow exceeds FGF
Composition of inspired mixture will depend uponrebreathing
Dr. A K Sethis EORCAPS-2009
Bi-Directional Flow
Systems extensively used
Depend on the FGF for effective CO2 elimination
Functions can be manipulated by changing
parameters like
Fresh Gas Supply
Elimination of CO2
Apparatus Dead Space
Dr. A K SethisEORCAPS-2009
Bi-Directional Flow
Fresh Gas Supply; Low do not eliminate CO2 effectively
High wastage
If the system has to deliver a set concentration in the shortest possible timeto the alveoli, the FGF should be delivered as near the patients airway aspossible
Elimination of CO2 Normal production of CO2 in a 70 kg patient 200 ml/min
End-tidal concentration of CO2 is 5%
For elimination 200ml of of CO2 as a 5% gas mixture, the alveolarventilation has to be: 200 x 100 = 4,000 ml.
5
4 litres is the normal alveolar ventilation.
Apparatus Dead Space
It is the volume of the breathing system from the patient-end to the point upto which, to and fro movement of expired gas takes place
Dr. A K Sethis EORCAPS-2009
Classification
Five Basic types
A through E
F added by willis in 1975
Dr. A K SethisEORCAPS-2009
Mapleson Classification
Reservoir bag
Corrugated tubing
APL valve
FGI
Patient connection
Dr. A K Sethis EORCAPS-2009
Mapleson A
Magill Circuit
Popularized by Sir Ivan Magill in 1920s
It differs from the other Mapleson systems in that fresh gas does notenter the system near the patient connection but enters at the other
end of the system near the reservoir bag.
A corrugated tubing connects the bag to the adjustable pressure
limiting (APL) valve at the patient end of the system
Length of corrugated breathing tube 110 cm with an internal volume
of 550 ml
7/24/2019 Anaesthesia Breathing Circuits (EORCAPS 2009)
4/9
Dr. A K Sethis EORCAPS-20
Dr. A K SethisEORCAPS-2009
Mapleson A
Most efficient for spontaneous respiration
Resistance 2.5cm of H2O when expiratory valve is fully open
No rebreathing if FGF equal to minute volume
Rebreathing occurs if FGF less than alveolar ventilation
Dr. A K Sethis EORCAPS-2009
Functional Analysis
Mapleson A with Spontaneous Breathing
Dr. A K SethisEORCAPS-2009
Functional Analysis
Mapleson A with Control Ventilation
a = at the end of inspiration
b = at the end of expiration
c = during subsequent inspiration
d = at the end of subsequent inspiration
Dr. A K Sethis EORCAPS-2009
Mapleson B & C systems
In order to reduce the rebreathing of alveolar gas and to
improve the utilization of FG during controlled
ventilation, the FG entry was shifted near the patient.
This allows a complete mixing of FG and expired gas.
The end result is that these systems are neither efficient
during spontaneous nor during controlled ventilation.
Dr. A K SethisEORCAPS-2009
Modifications of Mapleson A Circuit
Lack Circuit Added expiratory limb which runs from pt end to APL
valve at machine end
This limb is coaxially placed inside the affarent limb
Length 1.5 m, resistance 1.63 cm H2O Advantage
Facilitates easy scavenging
Easy adjustment of APL valve
Disadvantage
Increased resistance to breathing
Dr. A K Sethis EORCAPS-2009
Enclosed Afferent Reservoir System
Miller & Miller 1988
Mapleson A system enclosed within a non
distensible structure
It may also be constructed by enclosing the
reservoir bag alone in a bottle and connecting
the expiratory port to the bottle with a
corrugated tube and a one way valve
To the bottle is also attached a reservoir bag
and a variable orifice for providing positive
pressure ventilation.
7/24/2019 Anaesthesia Breathing Circuits (EORCAPS 2009)
5/9
Dr. A K Sethis EORCAPS-20
Dr. A K SethisEORCAPS-2009
Enclosed Afferent Reservoir System
Functional analysis:
During spontaneous ventilation it is identical to the Mapleson A system
In this mode the variable orificeis kept widely open to allow free communicationto the atmosphere
In controlled ventilation the reservoir bag B is squeezed intermittently & the
variable orificeis partly closed to allow building up of pressure in the bottle
The pressure thus developed closes the expiratory valve squeezes the enclosed
afferent reservoir and the patient gets ventilated
Hence this system should function efficiently during spontaneous and
controlled ventilation with a FGF equivalent to alveolar ventilation
Dr. A K Sethis EORCAPS-2009
Efferent Reservoir Systems
Mapleson DEF All have a T-piece near the patient and
function similarly
The T-piece is a three-way tubular
connector with a patient connection port,
a FGF port & a port for connection to a
corrugated tubing
Light metal tube 1cm in diameter ,5cm in
length with a side arm
6 mm tube as the affarent limb that supplies
the FG from the machine
Dr. A K SethisEORCAPS-2009
Mapleson D
A length of tubing connects the T-piece at the patient end
to the APL valve and the reservoir bag adjacent to it
The length of the tubing determines the distance the user
can be from the patient but has minimal effects on
ventilation
Popular because excess gas scavenging is relatively easy
Most efficient of the Mapleson systems during controlled
ventilation
Dr. A K Sethis EORCAPS-2009
Mapleson D Functional Analysis
Spontaneous Ventilation
Dr. A K SethisEORCAPS-2009
Mapleson D Functional Analysis
Controlled Ventilation
Dr. A K Sethis EORCAPS-2009
Bain Breathing Circuit
Modified Mapleson D
Bain 1972
The fresh gas supply tube runs coaxially inside the corrugated tubing
and ends at the point where the fresh gas would enter if the classicMapleson D form were used
The outer tube is clear so that the inner tube can be inspected
The outer tubing of most commercially available versions of the
Bain system is narrower than conventional corrugated tubing
A long version of the Bain system may be used for remote
anesthesia in locations such as MRI
7/24/2019 Anaesthesia Breathing Circuits (EORCAPS 2009)
6/9
Dr. A K Sethis EORCAPS-20
Dr. A K SethisEORCAPS-2009
Bain System
Depends on fresh gas flow to flush out CO2
Spontaneous ventilation
200 - 300 ml / kg / min
Controlled ventilation
infants 60 kg 70 ml/kg/min
Dr. A K Sethis EORCAPS-2009
Bain Circuit
Coaxial length 1.8m
Inner inspiratory tubing - 7 mm diameter
Outer expiratory tubing - 22 mm diameter
Resistance - < 0.7 cm H2O
Dr. A K SethisEORCAPS-2009
Relation Between Alveolar
Ventilation and FGF
Dr. A K Sethis EORCAPS-2009
Bain System Advantages
Light-weight, Compact, easy to handle
Warming of inspired gases
Partial rebreathing improves humidification
Ability of scavenging
Can be used with ventilator
Dr. A K SethisEORCAPS-2009
Disadvantage
Accidental disconnection of inner tube, kinking & twisting of
inner tube converts entire exhalation limb into dead space
FGF may be connected to outer tube
Lower FGF may result in re-breathing
Tests-visual inspection for damage, disconnection
Pethicks test, occlusion test
Dr. A K Sethis EORCAPS-2009
Mapleson E
Ayres T Piece, 1937
Low resistance, low dead space
Simple piece of metal with 3 limbs & a port at the end of
each limb Exhalation tube is a reservoir
No bag
FGF near mask
7/24/2019 Anaesthesia Breathing Circuits (EORCAPS 2009)
7/9
Dr. A K Sethis EORCAPS-20
Dr. A K SethisEORCAPS-2009
Mapleson E
Hazards
Over inflation and barotrauma by prolonged occlusion
and no feel of inflation
Also pressure buffering effect of bag absent
Dr. A K Sethis EORCAPS-2009
Mapleson F
Jackson Rees Modification
Addition of a bag with an opening to exhalation limb of T- piece
Corrugated hose 15cm X 11 mm
Reservoir bag 0.5 1 L
Green tube for FGF 1.5m in length
Advantage
Early to assist / Controlled ventilation
Easy to monitor ventilation
Dr. A K SethisEORCAPS-2009
Advantage of the Mapleson Systems
Simple, inexpensive, light weight. With the exception of the APL
valve, there are no moving parts
In coaxial systems (Lack, Bain), the inspiratory limb is heated by the
warm exhaled gas in the coaxial expiratory tubing
Resistance is usually low at flows likely to be experienced in practice
Easy to position conveniently. A long Mapleson D system may be
used to ventilate a patient in the MRI unit
Compression & compliance volume losses are less than with the circle
system
Changes in fresh gas concentrations result in rapid changes in
inspiratory gas composition
Dr. A K Sethis EORCAPS-2009
Disadvantage of the Mapleson Systems
Require high gas flows - results in higher costs, increased
atmospheric pollution, and difficulty assessing spontaneous
ventilation
Inspired heat and humidity tend to be low because of high FGF
In the Mapleson A, B, and C systems the APL valve is located close
to the patient, where it may be inaccessible to the user.
The Mapleson E and F systems are difficult to scavenge
Air dilution can occur with the Mapleson E system.
Mapleson systems are not suitable for patients with malignant
hyperthermia
Dr. A K SethisEORCAPS-2009
Combined System
Humphrey ADE To over come the difficulties of changing the breathing system for
different modes of ventilation
Humphery designed a system called Humphery ADE in 1983
Two reservoirs, one in the afferent limb and the other in the
efferent limb
Only one reservoir will be in operation and the system can be
changed from ARS to ERS by changing the position of a lever
Can be used for adults as well as children
Functional analysis is the same as Mapleson A in ARS mode and
as Bain in ERS mode
Not yet widely used
Dr. A K Sethis EORCAPS-2009
Paediatric Circuit
work of breathing
Prevent rebreathing
Volume of the non rebreathing circuit is small as
compared with that of circuit system, the compression
and compliance volume will be significantly less. This
improves the ability to observe respiratory effect as
reflected by movement of the anaesthesia bag
Mapleson D recommended for use in children < 10 kg wt
7/24/2019 Anaesthesia Breathing Circuits (EORCAPS 2009)
8/9
Dr. A K Sethis EORCAPS-20
Dr. A K SethisEORCAPS-2009
Questions ?
Dr. A K Sethis EORCAPS-2009
Which system have the FGF
near the reservoir bag
A
Dr. A K SethisEORCAPS-2009
Apparatus Dead Space extend
upto
In ARS Expiratory valve
In ERS Fresh Gas entry
Dr. A K Sethis EORCAPS-2009
F circuit was introduced by
Whom & When?
Willis 1975
Dr. A K SethisEORCAPS-2009
Bain Circuit
Coaxial length 1.8 meter
Inner inspiratory tubing diameter 7 mm
Outer expiratory tubing diameter 22 mm
Resistance
7/24/2019 Anaesthesia Breathing Circuits (EORCAPS 2009)
9/9
Dr. A K Sethis EORCAPS-20
Dr. A K SethisEORCAPS-2009
The efficiency of the circuit
is determined in terms of
CO2 elimination
FGF utilization
Dr. A K Sethis EORCAPS-2009
Which system is most efficient with
respect to prevention of rebreathing
during Spontaneous Respiratory
A > DEF > BC
Dr. A K SethisEORCAPS-2009
Which system is most efficient with
respect to prevention of rebreathing
during Controlled ventilation
DEF > BC > A
Dr. A K Sethis EORCAPS-2009
To prevent rebreathing of CO2
system DEF require a FGF of
2.5 times the minute volume
Dr. A K SethisEORCAPS-2009
In spontaneous respiration to
prevent rebreathing the
Mapleson A requires flow rate
Equal to minute volume