© Siemens AG The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Print No.: Siemens Medical Solutions Ultrasound Division n.a. lne Antares Peripherals Installation Instructions US 2004 US03-201.812.01.02.02 Replaces: US03-201.812.01.01.02 English Doc. Gen. Date: 07.05 n.a. Part No.: n.a. sd Printer and VCR Configurations 05936518 08653771 10032746 10032747

Antares Peripheral Instruction

Embed Size (px)

DESCRIPTION

ANTARES TECHNICAL MANUAL

Citation preview

  • PeripheralsInstallation Instructions

    Printer and VCR Configurations Siemens AGe reproduction, tranthis document or itrmitted without hority. Offenders wmages. All rights, ated by patent graa utility model erved.

    Print No.:

    Siemens Medical SolutionsUltrasound Divisionn.a.lne

    US03-201.812.0Replaces: US03-201.812.0

    glishc. Gen. Da

    Part No.: n.a.sd

    05936518086537711003274610032747

    smission or uses contents is notexpress writtenill be liable forincluding rights

    nt or registrationor design, are

    2004

    te: 07.05n.a.Thof peautdacreof res

    1.02.021.01.02

    EnDoAntaresUS

  • 2 Revision / Disclaimer

    1Revision / Disclaimer

    CopyrightCopyright 2004 by Siemens Corporation. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, stored in retrieval systems, or translated into any language or computer language, in any form or by any means, electronic, mechanical, magnetic, optical, chemical, manual, or otherwise, without the prior written permission of Siemens Corporation. Siemens Corporation reserves the right to change its products and services at any time. In addition, this manual is subject to change without notice. Siemens Corporation welcomes customer input on corrections and suggestions for improvements to this manual.Although Siemens Corporation has attempted to ensure accuracy through-out this manual, Siemens Corporation assumes no liability for any errors or omissions, nor for any damages resulting from the application or use of this information. TrademarksACUSON, Sequoia, Cypress, AEGIS, Aspen, 128XP, XP, AcuNav, CV70, SONOLINE, Adara, Antares, G20, G40, G50, G60S, Omnia and KinetDx are trademarks of Siemens Corporation registered in the U.S. Patent and Trademark Office. CWS3000, DS3000, WS3000, and WebPro are trademarks of Siemens Corporation.Windows XP, 2000 and NT are registered trademarks of Microsoft Corporation. Internet Explorer is a trademark of Microsoft Corporation. Netscape and Netscape Navigator are registered trademarks of Netscape Communications Corporation. Java and the JavaScript name are registered trademarks of Sun Microsystems, Inc. Adobe Acrobat is a registered trademark of Adobe Systems, Inc.Other products and brand names are trademarks of their respective owners.DisclaimerThe service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by Sie-mens or one of its affiliates to provide such services.Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro-cedures.License AgreementAll computer programs copyright 1990-2004 by Siemens Corporation or its suppliers. Pro-grams are licensed under the following agreement:Siemens or its suppliers retain(s) ownership of and title to any computer program supplied with the equipment and to the trade secrets embodied in such computer programs. Sub-ject to the Buyers acceptance and fulfillment of the obligations in this paragraph, Siemens grants the Buyer a personal, non-transferable, perpetual, non-exclusive license to use any computer program supplied with the Equipment that is necessary to operate the Equip-ment solely on the medium in which such program is delivered for the purpose of operat-ing the equipment in accordance with the instructions set forth in the operators manuals supplied with the Equipment and for no other purpose whatsoever. Buyer may not reverse-assemble, reverse-compile or otherwise reverse-engineer such computer pro-grams nor may Buyer make a copy of such program or apply any techniques to derive the trade secrets embodied therein. In the event of a failure by Buyer to comply with the terms Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 2 of 150Medical Solutions USA, Inc.

  • Revision / Disclaimer 3of this license, the license granted by this paragraph shall terminate. Further, because unauthorized use of such computer programs will leave Siemens without an adequate remedy at law, Buyer agrees that injunctive or other equitable relief will be appropriate to restrain such use, threatened or actual. Buyer further agrees that (i) any of Siemens"s suppliers of software is a direct and intended beneficiary of this end-user sublicense and may enforce it directly against Buyer with respect to software supplied by such supplier, and (ii) No supplier of Siemens shall be liable to buyer for any general, special, direct, indi-rect, consequential, incidental or other damages arising out of the sublicense of the com-puter programs supplied with the equipment.Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 3 of 150Medical Solutions USA, Inc.

  • 4 Table of Contents

    1- 0Table of Contents

    1 _______ Overview_______________________________________________________ 6

    Report Printers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Mitigating the Risk of Connecting a Report Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Patient Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

    2 _______ Hardware Configuration _________________________________________ 14

    Single Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14CP770 or CP800 Color Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14UP-D23MD Color Printer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17UP-D55MD Color Printer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20P91 D or P93 D B/W Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23HS-MD3000 VCR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26SVO9500 VCR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

    Double Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34B/W Printer and Color Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34B/W Printer and VCR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47Color Printer and VCR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62

    Peripheral Shelf Mounting Hole Locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95I/O Module Peripheral Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97

    Printer I/O Module Cable Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97VCR I/O Module Cable Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98

    3 _______ Printer Setup and Configuration __________________________________ 99

    SW v2.1 and V3.0. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100CP800 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100CP770DW, P91DW, P93DW, UPD23MD, and UPD55MD Installation . . . . . . . . . . 107HP1150, HP2500n, HP4050, HP4100, HP4200, and HP6122 Installation . . . . . . . 108

    SW v4.0 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110CP800 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110CP770DW, P91DW, P93DW, UPD23MD, and UPD55MD Installation . . . . . . . . . . 118HP4050, HP4100 and HP4200 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120HP1150, HP2500n and HP6122 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121

    Reboot System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122Print/Store Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123

    Restore Factory Settings and Factory Printer Settings . . . . . . . . . . . . . . . . . . . . . . 123Configure Routing to Print and Store Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125Configure Print . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128Configure Store . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129

    Verify Printer Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130

    4 _______ VCR Setup and Configuration ___________________________________ 131Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 4 of 150Medical Solutions USA, Inc.

  • Table of Contents 5HS-MD3000 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132Front Panel Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133

    SVO9500 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138Front Panel Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138Main Menu Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139Hidden Menu Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140

    Verify the Operation of the VCR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142

    5________ Printer Troubleshooting _________________________________________143

    Printer Installation Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143Restoring Plug-and-Play Resources . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145Re-Install A Printer After Troubleshooting or Restoring Plug-and-Play. . . . . . . . . . . . . 147

    Report Printers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147Re-install Plug-and-Play Printers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 148Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 5 of 150Medical Solutions USA, Inc.

  • 6 Overview

    2- 1Overview

    The Peripheral Installation procedure lists the latest peripheral configurations for installa-tion and describes how to install each configuration for the the following Antares systems: ACUSON Antares ultrasound system ACUSON Antarespe ultrasound system SONOLINE Antares ultrasound systemInstall an onboard or offboard single configuration or double configuration on the Antares peripheral shelf. The peripherals consist of color printers, black and white (B/W) printers, and VCRs. The following table lists the onboard peripheral image printers and VCR and their software version installation types.

    Tab. 1 ONboard Peripheral Image Printers and VCRs - Software Installation Types:

    The following table lists the offboard peripheral report printers and their software version installation types.

    Peripheral SONOLINE Antares ACUSON Antares ACUSON AntaresPre-mier Edition

    Software Versions

    v2.1 v3.0 v4.0 v4.0 v4.0

    P91 D X X X NA NA

    P93 D X X X X X

    CP770DW X X X X X

    CP800DW X X X1

    1. CP800 is not supported on systems configured with the RM-165 PCB.

    NA NA

    UP-D23MD X X X X X

    UP-D55MD NA X X NA X

    HS-MD3000 X2

    2. Valid with combined code installed.

    X X X X

    SVO9500 X X X NA NAAntares US03-201.812.01.02.02 Siemens07.05 sd

    Page 6 of 150Medical Solutions USA, Inc.

  • Overview 7Tab. 2 Report Printers:

    Before installing report printers, comply with the guidellines in the following sections: (Report Printers / p. 8) (Mitigating the Risk of Connecting a Report Printer / p. 10) (Patient Environment / p. 11)

    Peripheral SONOLINE Antares ACUSON Antares ACUSON AntaresPre-mier Edition

    Software Versions

    v2.1 v3.0 v4.0 v4.0 v4.0

    HP1150 X X X X X

    HP2500n X X X X X

    HP4050 X X X X X

    HP4100 X X X X X

    HP4200 X X X X X

    HP6122 DeskJet

    X X X X XSiemens US03-201.812.01.02.02 Antares07.05 sd

    Page 7 of 150Medical Solutions USA, Inc.

  • 8 OverviewReport Printers 0

    WARNING Accessory equipment connected to the USB port must be certifiedaccording to the respective EN and IEC standards (for example, EN60950 and IEC 60950 for data processing equipment and EN60601-1 and IEC 60601-1 for medical equipment). Furthermore, allconfigurations shall comply with the system standard EN60601-1-1 and IEC 60601-1-1.

    Anyone who connects additional equipment to the signal input or signal output port configures a medical system and is therefore responsible that the system complies with the requirements of the system standard EN 60601-1-1 and IEC 60601-1-1. Siemens can only guaran-tee the performance and safety of the devices listed in the Accessories and Options chapter of the System Ref-erence (shipped with the ultrasound imaging system). If in doubt, consult the Siemens Uptime Service Center.

    WARNING Equipment connected to the ultrasound system and in the patientenvironment must be powered from a medically isolated powersource or must be a medically isolated device.

    Equipment powered from a non-isolated source can result in chassis leakage currents exceeding safe levels. Chassis leakage current created by an accessory or device connected to a non-isolated outlet may add to the chassis leakage current of the ultrasound system.

    WARNING Non-medical grade report printers cannot be used within a patientenvironment.

    During use of a non-medical grade report printer or when a non-medical grade report printer is connected to the ultrasound system, the ultrasound system cannot in any way be in contact with a patient.

    NOTE Siemens cannot support or service any external devices connect-ed to the USB port. Siemens cannot assume responsibility forfunctionality beyond the scope of this specification.

    NOTE Siemens does not specifically recommend any particular commu-nication or analysis software; please contact a local vendor for as-sistance with interpretation and analysis of the data transmitted bythe ultrasound imaging system.Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 8 of 150Medical Solutions USA, Inc.

  • Overview 9Relevant standards for some non-medical electrical equipment may have limits for enclo-sure leakage currents higher than required by medical standards. These higher standards are acceptable only outside the patient environment. It is essential to reduce enclosure leakage currents when non-medical electrical equipment is to be used within the patient environment. Measures for reducing leakage current include use of a medically-approved isolation transformer. The ultrasound system has printer drivers compatible with non-medical grade report print-ers. Siemens does not guarantee the performance and safety of any non-medical grade report printer. Non-medical grade report printers do not fulfill the following safety require-ments: EN 60601-1-1 and IEC 60601-1-1 (Medical Electrical Equipment, Part 1: General Re-

    quirements for Safety). EN 60601-1-2 and IEC 60601-1-2 (Electromagnetic Compatibility of Medical Devices).If a non-medical grade report printer is to be used with the ultrasound system, then ensure mitigation is provided to meet all safety requirements. It is the responsibility of the user to ensure that the ultrasound system in combination with the non-medical grade report printer complies with safety requirements.Always use a medically-approved isolation transformer with a non-medical grade report printer. Mitigate risk with regard to the Electromagnetic Compatibility of Medical Devices require-ment. A non-medical grade device must be designed, manufactured, and certified to meet the same EMC (electromagnetic compatibility) requirements as the ultrasound system, or other means must ensure that the overall EMC requirements are met.

    Tab. 3 Manufacturers of Medically Approved Isolation Transformers

    Manufacturer Web Site

    Tripp Lite www.tripplite.com

    Toroid Corporation of Maryland www.toroid.com

    Dale Technology www.daletech.comSiemens US03-201.812.01.02.02 Antares07.05 sd

    Page 9 of 150Medical Solutions USA, Inc.

  • 10 OverviewMitigating the Risk of Connecting a Report Printer 0To fulfill EN 60601-1-1 and IEC 60601-1-1 (Medical Electrical Equipment, Part 1: General Requirements for Safety) requirements for non-medical peripheral equipment:The non-medical peripheral equipment must be approved according to any other EN or IEC standard (EN XXXXX or IEC XXXXX, e.g., equipment complying with EN 60348 and IEC 60348, EN 60950 and IEC 60950, etc.)The connection of non-medical peripheral equipment to the ultrasound system must adhere to the following conditions:

    1. Connect the ultrasound system to an AC power outlet within a medically used room within the patient environment. A patient environment is defined as an area in which medical examination, monitoring, or treatment of the patient takes place. The patient environment is located 1.5 meters (or, in Canada and the U.S.A., 1.8 meters [6 feet]) around the patient location.

    2. Connect the peripheral equipment AC power cable to a medically approved isolation transformer.

    3. Connect the medically approved isolation transformer to a main AC outlet either inside or outside the patient environment. The peripheral device and medically approved iso-lation transformer combination can be located:- Within the same room as the ultrasound system, or- In a non-medically used room.

    For additional information and other possible combinations, please refer to the Medical Electrical Equipment Standard EN 60601-1-1 or IEC 60601-1-1, Annex BBB.7, Scenario 3c.

    NOTE The above information is based on current EN 60601-1-1 and IEC60601-1-1 standards, dated 2000-12. If a country"s regulatory stan-dards for medical equipment do not correspond to EN 60601-1 andIEC 60601-1, as well as EN 60601-1-1 and IEC 60601-1-1, then thelocal requirements may differ.Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 10 of 150Medical Solutions USA, Inc.

  • Overview 11Patient Environment 0

    Fig. 1: Patient Environment (Sheet 1)Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 11 of 150Medical Solutions USA, Inc.

  • 12 OverviewFig. 2: Patient Enviroment (Sheet 2)Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 12 of 150Medical Solutions USA, Inc.

  • Overview 13Requirements 0Time required: Software setup and configuration: 10 minutes per device Single hardware configuration: 30 minutes Double hardware configuration: 60 minutesTools: Pen Exacto knife or 1/4-inch hollow punch Phillips head screw driver Flat head screw driverSiemens US03-201.812.01.02.02 Antares07.05 sd

    Page 13 of 150Medical Solutions USA, Inc.

  • 14 Hardware Configuration

    3- 2Hardware ConfigurationSingle Configuration 0

    Tab. 4 Single Configurations

    CP770 or CP800 Color Printer 0

    1. Cut the fastener holes into the rubber mat of the peripheral shelf.See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.

    2. Return the mat to the peripheral shelf.3. Align the fastener holes of the mounting bracket to the holes in the peripheral mat.4. Remove the feet and the shipment locking screw from the bottom of the printer.

    a) Point the bottom side of the printer face up.b) Remove the four screws that secure the feet to the printer and remove the feet.c) Remove the shipment locking screw.

    Fig. 3: CP770 or CP800 Printer - Shipping screw removal

    Color Printers B/W Printers VCRs

    (CP770 or CP800 Color Printer / p. 14)(UP-D23MD Color Printer / p. 17)(UP-D55MD Color Printer / p. 20)

    (P91 D or P93 D B/W Printer / p. 23)

    (HS-MD3000 VCR / p. 26)(SVO9500 VCR / p. 31)Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 14 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 155. Secure the mounting bracket to the color printer.a) Put the mounting bracket on the printer.b) Align the mounting holes.c) Secure the bracket to the bottom of the color printer.

    Fig. 4: CP770 or CP800 - Mounting Bracket

    WARNING Use fasteners of the correct length. Using fasteners that are too long may cause the hard-

    ware to contact other components, leading to product damage or a potential electrical safety hazard. Using fas-teners that are too short may not secure the component to its mount, presenting a performance and safety haz-ard.

    6. Put the color printer assembly on the peripheral shelf and align the mounting holes of the color printer with the mounting holes in the mat.Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 15 of 150Medical Solutions USA, Inc.

  • 16 Hardware Configuration7. Set the color printer dip switches as follows:

    8. Connect the color printer cables.a) Connect the power cable to the printer AC power inlet and to the peripheral AC pow-

    er outlet. b) Connect the USB cable to the back of the printer.

    Fig. 7: CP770 or CP800 - Cable connectionsPos. 1 USB Cable (CP800 may have a parrallel cable)Pos. 2 Power cord

    c) Route the other end of the USB cable through the peripheral shelf cable passage to the I/O module.

    9. Install the cover.a) Put the cover over the printer.b) Align the four captive screws in the cover with the mounting holes.c) Tighten the four captive screws.

    Fig. 5: CP770 Dip Switch Settings to OFF1-OFF, 2-OFF, 3-OFF, 4-OFF,5-OFF, 6-OFF, 7-OFF, and 8-OFF.

    Fig. 6: CP800 Dip Switch Settings1-ON, 2-OFF, 3-OFF, 4-OFF,5-OFF, 6-OFF, 7-OFF, and 8-ON.Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 16 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 1710. See the Mitsubishi User Instruction to perform the following tasks: - Load the ink cartridge- Load the paper tray- Slide the paper tray into the printer

    11. Complete the printer cable connections to the I/O module, see (I/O Module Peripheral Connections / p. 97).

    12. To complete the color printer configuration, see (Printer Setup and Configuration / p. 99).

    UP-D23MD Color Printer 0

    1. Cut the fastener holes into the rubber mat of the peripheral shelf.See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.

    2. Return the mat to the peripheral shelf.3. Align the fastener holes of the mounting bracket to the holes in the peripheral mat.Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 17 of 150Medical Solutions USA, Inc.

  • 18 Hardware Configuration4. Put the printer feet into the mounting bracket cut-outs.

    Fig. 8: Printer Foot Placement LocationsPos. 1 Rear mounting plate cutout for placement of the printers rear feetPos. 2 Front mounting plate cutout for placement of the printers front feet

    Fig. 9: Printer Placement on Mounting Plate Pos. 1 Cable routing passage through the mat and panel to the I/O moduleAntares US03-201.812.01.02.02 Siemens07.05 sd

    Page 18 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 195. Connect the printer cables.

    Fig. 10: Cable Connections - AC Power and USBPos. 1 Antares OEM AC power outletPos. 2 Printer AC power inletPos. 3 Printer USB connection

    a) Connect the AC power cable to the Antares peripheral AC power outlet and to the printer AC power inlet.

    b) Connect the USB cable to the back of the printer.c) Route the other end of the USB cable through the I/O module cable routing pas-

    sage. (See (1/Fig. 9 / p. 18) for the routing passage location.)

    NOTE Do not connect the USB cable to the I/O module at this time. Use cable ties to neatly bundle excess cable length. Excess cablewill be stored under the printer cover.Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 19 of 150Medical Solutions USA, Inc.

  • 2 0 Hardware Configuration6. Install the printer cover.

    Fig. 11: Printer Cover - InstalledPos. 1 Captive screw Pos. 2 Captive screw locations at base of cover (not in view)

    a) Put the printer cover over the printer.b) Align the three captive screws with the three fastener holes cut into the mat.c) Tighten the captive screws

    7. See the Sony User Instruction to perform the following tasks: - Load the ink cartridge

    NOTE Power on the system prior to loading the ink cartridge. This resetsthe gears to their default location, permitting the ink cartridge to beinserted.

    - Load the paper tray- Slide the paper tray into the printer

    8. Complete the printer cable connections to the I/O module, see (I/O Module Peripheral Connections / p. 97).

    9. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).

    UP-D55MD Color Printer 0

    1. Cut the fastener holes into the rubber mat of the peripheral shelf.See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.

    2. Return the mat to the peripheral shelf.Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 20 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 213. Secure the printer bracket to the bottom of the color printer.

    Fig. 12: UP-D55M Color Printer Bracket

    NOTE In (Fig. 12 / p. 21), the flat side of the bracket is oriented on the leftside of the picture.

    a) Point the bottom side of the printer face up.b) Put the bracket on the bottom of the printer. c) Tighten the bracket captive screw.

    4. Align the mounting holes of the mounting bracket to the holes in the peripheral mat. Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 21 of 150Medical Solutions USA, Inc.

  • 22 Hardware Configuration5. Connect the printer cables.

    Fig. 13: UP-D55M Cable ConnectionsPos. 1 Power cord between the OEM AC power outlet and the printer AC power inletPos. 2 USB cable

    a) Connect the AC power cable to the Antares peripheral AC power outlet and to the printer AC power inlet.

    b) Connect the USB cable to the back of the printer.c) Route the other end of the USB cable through the I/O module cable routing pas-

    sage.

    NOTE Do not connect the USB cable to the I/O module at this time. Use cable ties to neatly bundle excess cable length. Excess cablewill be stored under the printer cover.Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 22 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 236. Install the printer cover.

    Fig. 14: UP-D55 M Cover - Installed Pos. 1 Captive screw locationPos. 2 Captive screw locations at base of cover (not in veiw)

    a) Put the printer cover over the printer.b) Align the three captive screws with the three fastener holes cut into the mat.c) Tighten the captive screws

    7. See the Sony User Instruction to perform the following tasks: - Load the ink cartridge- Load the paper tray- Slide the paper tray into the printer

    8. Complete the printer cable connections to the I/O module, see (I/O Module Peripheral Connections / p. 97).

    9. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).

    P91 D or P93 D B/W Printer 0

    1. Cut the fastener holes into the rubber mat of the peripheral shelf.See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.

    2. Return the mat to the peripheral shelf.3. Put the B/W printer bracket on the B/W printer mounting plate. Use the applicable in-

    structions that follow.Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 23 of 150Medical Solutions USA, Inc.

  • 24 Hardware ConfigurationWARNING Use fasteners of the correct length. Using fasteners that are too long may cause the hard-

    ware to contact other components, leading to product damage or a potential electrical safety hazard. Using fas-teners that are too short may not secure the component to its mount, presenting a performance and safety haz-ard.

    P91 D Instructions: P93 D Instructions:

    Put the flat side of the bracket face down on the peripheral mat.

    Align the bracket mounting holes to the peripheral mat holes.

    Put the P91 B/W printer on top of the P91 mounting bracket.

    Align the B/W printer feet against the mounting bracket.

    Fig. 15: P91 D Positioned on the Bracket

    Point the bottom side of the P93 D print-er up.

    Put the P93 D mounting bracket on the bottom of the P93 D printer.

    Fig. 16: P93 D Mounting Bracket Secure the P93 D printer mounting

    bracket to the P93 D printer. Point the bottom side of the printer fac-

    ing down. Put the Printer assembly on the periph-

    eral mat.Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 24 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 254. Set the printer B/W dip switches: - P91 D settings: 1-ON, 2-OFF, 3-OFF, and 4-OFF- P93 D settings: 1-ON, 2-OFF, 3-ON, and 4-OFF

    Fig. 17: P91 D and P93 D Printer Cables and Dip Switches (P91 D shown)Pos. 1 Dip SwitchesPos. 2 USB CablePos. 3 AC Power Cord

    5. Connect the printer cables.a) Connect the AC power cable to the Antares peripheral AC power outlet and to the

    printer AC power inlet.b) Connect the USB cable to the back of the printer.c) Route the other end of the USB cable through the I/O module cable routing pas-

    sage.

    NOTE Do not connect the USB cable to the I/O module at this time. Use cable ties to neatly bundle excess cable length. Excess cablewill be stored under the printer cover.Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 25 of 150Medical Solutions USA, Inc.

  • 26 Hardware Configuration6. Install the printer cover.

    Fig. 18: P91 D and P93 D Printer Cover Installeda) Put the cover over the printer.b) Tighten the two captive screws.c) Make sure the back of the printer touches the printer cover stop.

    7. Complete the printer cable connections to the I/O module, see (I/O Module Peripheral Connections / p. 97).

    8. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).

    HS-MD3000 VCR 0

    Fig. 19: HS-MD3000 VRC - Single ConfigurationPos. 1 VCR bracketPos. 2 VCRPos. 3 CoverAntares US03-201.812.01.02.02 Siemens07.05 sd

    Page 26 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 271. Cut the fastener holes into the rubber mat of the peripheral shelf.See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.

    2. Return the mat to the peripheral shelf.3. Remove the feet from the bottom of the VCR.

    Fig. 20: VCR Foot Locationsa) Point the bottom side of the VCR facing up.b) Remove the four screws that secure the feet to the VCR. c) Remove the feet.Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 27 of 150Medical Solutions USA, Inc.

  • 28 Hardware Configuration4. Secure the mounting bracket to the VCR.

    Fig. 21: HS-MD3000 Secured to the Mounting Bracket

    WARNING Use fasteners of the correct length. Using fasteners that are too long may cause the hard-

    ware to contact other components, leading to product damage or a potential electrical safety hazard. Using fas-teners that are too short may not secure the component to its mount, presenting a performance and safety haz-ard.

    NOTE In (Fig. 21 / p. 28), the front of the VCR is facing the left side of thepicture.

    a) Put the mounting bracket on the VCR.b) Align the mounting holes.c) Secure the bracket to the bottom of the VCR.Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 28 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 29WARNING Use fasteners of the correct length. Using fasteners that are too long may cause the hard-

    ware to contact other components, leading to product damage or a potential electrical safety hazard. Using fas-teners that are too short may not secure the component to its mount, presenting a performance and safety haz-ard.

    5. Put the VCR assembly on the peripheral shelf by aligning the mounting holes of the VCR with the mounting holes in the mat.

    Fig. 22: HS-MD3000 Positioned on the OEM BaySiemens US03-201.812.01.02.02 Antares07.05 sd

    Page 29 of 150Medical Solutions USA, Inc.

  • 30 Hardware Configuration6. Connect the VCR power cord to the peripheral AC power outlet and to the VCR AC power inlet.

    Fig. 23: HS-MD3000 Power Cord Connection Pos. 1 OEM power outletsPos. 2 VCR power inlet

    7. Connect the VCR cables. See (VCR I/O Module Cable Connections / p. 98) for details.

    Fig. 24: HS-MD3000 VCR Connections from the I/O ModulePos. 1 Connect P128 to RS-232CPos. 2 Connect VIDEO INPos. 3 Connect VIDEO OUTPos. 4 Connect AUDIO INPos. 5 Connect AUDIO OUTPos. 6 Set the 75 VIDEO and the 75 S-VIDEO switches to ON

    8. Route the cables through the I/O panel cable routing passage.Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 30 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 31NOTE Do not connect the cables to the I/O module at this time. Use cable ties to neatly bundle excess cable length. Excess cablewill be stored under the VCR cover.

    9. Install the cover.

    Fig. 25: VCR Cover Captive Screw Locationsa) Put the cover over the VCR.b) Align the three captive screws in the cover with the mounting holes.c) Tighten the three captive screws.

    10. Complete the VCR cable connections to the I/O module, see (I/O Module Peripheral Connections / p. 97).

    11. Complete the VCR configuration, see (VCR Setup and Configuration / p. 131).

    SVO9500 VCR 0

    1. Cut the fastener holes into the rubber mat of the peripheral shelf.See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.

    2. Secure the mounting bracket to the peripheral shelf.3. Put the VCR assembly on the peripheral shelf by aligning the mounting holes of the

    VCR with the mounting holes in the mat.Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 31 of 150Medical Solutions USA, Inc.

  • 32 Hardware Configuration4. Set the VCR switches: 1-OFF, 2-OFF, 3-OFF, 4-OFF, 5- ON, 6-ON

    Fig. 26: SVO9500 Dip Switch Settings- Set both 75 ohm switches to ON.- For PAL model only, set the voltage to 240V.

    NOTE Only the PAL model includes the 240V switch.

    5. Connect the VCR power cord to the peripheral AC power outlet and to the VCR AC power inlet.

    6. Connect the VCR cables. See (VCR I/O Module Cable Connections / p. 98) for details.

    Fig. 27: SVC 9500 VCR Connections from I/O ModulePos. 1 Audio In (left and right) Pos. 2 Video In (use either connector)Pos. 3 Audio Out (left and right)Pos. 4 Video OutPos. 5 Dip Switches Pos. 6 RS-232C (connect to P128)Pos. 7 75 (set both switches to ON)

    7. Route the cables through the I/O panel cable routing passage.Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 32 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 33NOTE Do not connect the cables to the I/O module at this time. Use cable ties to neatly bundle excess cable length. Excess cablewill be stored under the VCR cover.

    8. Install the cover.a) Put the cover over the VCR.b) Align and tighten the two captive screws in the cover with the mounting holes.

    9. Complete the VCR cable connections to the I/O module, see (VCR I/O Module Cable Connections / p. 98).

    10. Complete the VCR configuration, see (VCR Setup and Configuration / p. 131).Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 33 of 150Medical Solutions USA, Inc.

  • 34 Hardware ConfigurationDouble Configuration 0Tab. 5 Double Configuration

    B/W Printer and Color Printer 0

    P91 D or P93 D / CP770 or CP800Installation Summary1. Install the color printer on the peripheral shelf.2. Install the B/W printer on the color printer.3. Complete the cable I/O module connections.4. Complete the printer configuration.

    NOTE When performing this upgrade on a system that already has a B/Wprinter or a color printer installed, remove the printer from the pe-ripheral shelf.Do not disconnect the USB cable from the I/O module. The systemhas previously configured the USB port for the printer connected.Therefore, the same USB port must be used for that printer whenthe USB cable is reconnected.Label the USB cable for the correct reconnection to the printer.

    1. Cut the fastener holes into the rubber mat of the peripheral shelf.See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.

    2. Return the mat to the peripheral shelf.3. Align the fastener holes of the mounting bracket to the holes in the peripheral mat.

    B/W Printer/Color PrinterDouble Configuration

    B/W Printer/VCR Double Configuration

    Color Printer/VCRDouble Configuration

    (P91 D or P93 D / CP770 or CP800 / p. 34)

    (P91 D or P93 D / HS-MD3000 / p. 47)

    (CP770 or CP800 / HS-MD3000 / p. 62)

    (P91 D or P93 D / UP-D23MD / p. 37)

    (P91 D or P93 D / SVO9500 / p. 56)

    (UP-D23MD / HS-MD 3000 / p. 75)

    (P91 D or P93 D / UP-D55M / p. 42)

    (UP-D55MD / HS-MD3000 / p. 86)(CP770 or CP800 / SVO9500 / p. 67)Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 34 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 354. Remove the feet and the shipment locking screw from the bottom of the printer.a) Point the bottom side of the printer face up.b) Remove the four screws that secure the feet to the printer and remove the feet.

    Fig. 28: CP770 or CP800 Secured to the Mounting Bracketc) Remove the shipment locking screw.

    5. Secure the mounting bracket to the color printer.a) Put the mounting bracket on the printer.b) Align the mounting holes.c) Secure the bracket to the bottom of the color printer.

    WARNING Use fasteners of the correct length. Using fasteners that are too long may cause the hard-

    ware to contact other components, leading to product damage or a potential electrical safety hazard. Using fas-teners that are too short may not secure the component to its mount, presenting a performance and safety haz-ard.

    6. Put the color printer assembly on the peripheral shelf and align the mounting holes of the color printer with the mounting holes in the mat.Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 35 of 150Medical Solutions USA, Inc.

  • 36 Hardware Configuration7. Set the color printer dip switches as follows:

    8. Connect the color printer cables.a) Connect the color printer AC power cable to the color printer AC power inlet and to

    the lower peripheral AC power outlet. b) Connect the B/W printer AC power cable to the upper peripheral AC power outlet.c) Connect the color printer USB cable to the back of the color printer. d) Route the other end of the color printer USB cable through the peripheral shelf ca-

    ble passage to the I/O module.e) Route the I/O module end of the B/W printer USB cable through the peripheral shelf

    cable passage to the I/O module.

    NOTE Do not connect the USB cables to the I/O module at this time, un-less the system previously had one of the printers installed on theperipheral bay. Then, connect the previously installed printer tothe same port it was connected to before.

    9. Install the color printer cover/black and white printer mounting plate assembly.a) Route the B/W printer USB cable and power cable up through the hole in the B/W

    printer mounting plate.b) Put the printer cover over the printer.c) Align the four captive screws with the fasener holes cut into the mat.d) Tighten the captive screws.

    10. Set the B/W printer dip switches to: - P91 D settings: 1-ON, 2-OFF, 3-OFF, and 4-OFF- P93 D settings: 1-ON, 2-OFF, 3-ON, and 4-OFF

    11. Put the B/W printer bracket on the B/W printer mounting plate. Use the applicable in-structions that follow.

    Fig. 29: CP770 Dip Switch Settings to OFF1-OFF, 2-OFF, 3-OFF, 4-OFF,5-OFF, 6-OFF, 7-OFF, and 8-OFF.

    Fig. 30: CP800 Dip Switch Settings1-ON, 2-OFF, 3-OFF, 4-OFF,5-OFF, 6-OFF, 7-OFF, and 8-ON.Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 36 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 3712. Install the B/W printer.a) Put the B/W printer on the bracket and mounting bracket. b) Connect the AC power cable to the back of the B/W printer.c) Connect the USB cable to the back of the B/W printer.

    NOTE Do not connect the USB cable to the I/O module at this time.

    13. Install the B/W printer cover.a) Put the B/W cover over the B/W printer.b) Tighten the captive screws (of the B/W printer cover) through the mounting bracket,

    to the mounting plate.c) Make sure the back of the B/W printer touches the B/W printer cover stop.

    14. See the Sony User Instruction to perform the following tasks for the color printer: - Load the ink cartridge

    NOTE Power on the system prior to loading the ink cartridge. This resetsthe gears to their default location, permitting the ink cartridge to beinserted.

    - Load the paper tray- Slide the paper tray into the printer

    15. Complete the printer cable connections to the I/O module, see (I/O Module Peripheral Connections / p. 97).

    16. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).

    P91 D or P93 D / UP-D23MDInstallation Summary1. Install the color printer on the peripheral shelf2. Install the B/W printer on the color printer.3. Complete the cable I/O module connections.4. Complete the printer configuration.Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 37 of 150Medical Solutions USA, Inc.

  • 38 Hardware ConfigurationNOTE When performing this upgrade on a system that already has a B/Wprinter or a color printer installed, remove the printer from the pe-ripheral shelf.Do not disconnect the USB cable from the I/O module. The systemhas previously configured the USB port for the printer connected.Therefore, the same USB port must be used for that printer whenthe USB cable is reconnected.Label the USB cable for the correct reconnection to the printer.

    1. Cut the fastener holes into the rubber mat of the peripheral shelf.See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.

    2. Return the mat to the peripheral shelf.3. Align the fastener holes of the mounting bracket to the holes in the peripheral mat.4. Put the color printer feet into the mounting bracket cut-outs.

    Fig. 31: Printer Foot Placement LocationsPos. 1 Rear mounting plate cutout for placement of the printers rear feetPos. 2 Front mounting plate cutout for placement of the printers front feetAntares US03-201.812.01.02.02 Siemens07.05 sd

    Page 38 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 395. Connect the color and B/W power cables to the Antares AC outlet.

    Fig. 32: UP-D23MD Printer Placement and Cable ConnectionsPos. 1 OEM Bay to I/O Module cable routing passagePos. 2 AC power outlet UP-D23MD Pos. 3 AC power outlet P91 D/P93 DPos. 4 Power inlet UP-D23MDPos. 5 USB cable connection UP-D23MD

    a) Connect the AC power cable to the Antares peripheral AC power outlet and to the color printer AC power inlet.

    b) Connect the B/W printer AC power cable to the Antares peripheral AC power outlet and route the cable through the color printer top cover.

    c) Connect the USB cable to the back of the color printer.d) Route both of the USB cables through the I/O module cable routing passage.

    NOTE Do not connect the USB cable to the I/O module at this time, unlessthe system previously had this printer installed. Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 39 of 150Medical Solutions USA, Inc.

  • 40 Hardware Configuration6. Install the printer cover.

    Fig. 33: UP-D23MD Printer Placement and Cable ConnectionsPos. 1 Mounting plate B/W printerPos. 2 Cable routing holesPos. 3 Captive screw locations

    a) Put the printer cover over the printer.b) Align the three captive screws with the three fastener holes cut into the mat.c) Tighten the captive screws

    NOTE Use cable ties to neatly bundle excess cable length. Excess cablewill be stored under the printer cover.

    7. Set the B/W printer dip switches to: - P91 D settings: 1-ON, 2-OFF, 3-OFF, and 4-OFF- P93 D settings: 1-ON, 2-OFF, 3-ON, and 4-OFF

    8. Put the B/W printer bracket on the B/W printer mounting plate. Use the applicable in-structions that follow.

    WARNING Use fasteners of the correct length. Using fasteners that are too long may cause the hard-

    ware to contact other components, leading to product damage or a potential electrical safety hazard. Using fas-teners that are too short may not secure the component to its mount, presenting a performance and safety haz-ard.Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 40 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 419. Install the B/W printer.a) Put the B/W printer on the bracket and mounting bracket. b) Connect the AC power cable to the back of the B/W printer.c) Connect the USB cable to the back of the B/W printer.

    NOTE Do not connect the USB cable to the I/O module at this time.

    10. Install the B/W printer cover.a) Put the B/W cover over the B/W printer.b) Tighten the captive screws (of the B/W printer cover) through the mounting bracket,

    to the mounting plate.c) Make sure the back of the B/W printer touches the B/W printer cover stop.

    P91 D Instructions: P93 D Instructions:

    Put the flat side of the bracket face down on the B/W printer mounting plate of the color printer cover.

    Align the bracket mounting holes to the plate mounting holes.

    Put the P91 B/W printer on top of the P91 mounting bracket.

    Align the B/W printer feet against the mounting bracket.

    Fig. 34: P91 D Positioned on the Bracket

    Point the bottom side of the P93 D print-er up.

    Put the P93 D mounting bracket on the bottom of the P93 D printer.

    Fig. 35: P93 D Mounting Bracket Secure the P93 D printer mounting

    bracket to the P93 D printer. Point the bottom side of the printer fac-

    ing down. Put the Printer assembly on the color

    printer mounting plate.Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 41 of 150Medical Solutions USA, Inc.

  • 42 Hardware Configuration11. See the Sony User Instruction to perform the following tasks for the color printer: - Load the ink cartridge

    NOTE Power on the system prior to loading the ink cartridge. This resetsthe gears to their default location, permitting the ink cartridge to beinserted.

    - Load the paper tray- Slide the paper tray into the printer

    12. Complete the printer cable connections to the I/O module, see (I/O Module Peripheral Connections / p. 97).

    13. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).

    P91 D or P93 D / UP-D55MInstallation Summary1. Install the color printer on the peripheral shelf2. Install the B/W printer on the color printer.3. Complete the cable I/O module connections.4. Complete the printer configuration.

    NOTE When performing this upgrade on a system that already has a B/Wprinter or a color printer installed, remove the printer from the pe-ripheral shelf.Do not disconnect the USB cable from the I/O module. The systemhas previously configured the USB port for the printer connected.Therefore, the same USB port must be used for that printer whenthe USB cable is reconnected.Label the USB cable for the correct reconnection to the printer.

    1. Cut the fastener holes into the rubber mat of the peripheral shelf.See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.

    2. Return the mat to the peripheral shelf.Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 42 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 433. Tighten the printer bracket to the bottom of the color printer.

    Fig. 36: UP-D55M Color Printer Bracketa) Point the bottom side of the printer face up.b) Put the bracket on the bottom of the printer. c) Tighten the bracket captive screw.

    4. Connect the printer cables.

    Fig. 37: UP-D55M Cable ConnectionsPos. 1 Power cord between the OEM AC power outlet and the printer AC power inletPos. 2 USB cable

    a) Connect the AC power cable to the Antares peripheral AC power outlet and to the printer AC power inlet.Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 43 of 150Medical Solutions USA, Inc.

  • 44 Hardware Configurationb) Connect the USB cable to the back of the printer.c) Route the other end of the USB cable through the I/O module cable routing pas-

    sage.

    NOTE Do not connect the USB cable to the I/O module at this time, unlessthe system previously had this printer connected. Use cable ties to neatly bundle excess cable length. Excess cablewill be stored under the printer cover.

    5. Connect the B/W printer AC power cable into the Antares peripheral AC power outlet. Route the other end of the power cable through the color printer cover.

    6. Route the B/W printer USB cable through the I/O panel cable routing passage.

    NOTE Do not connect the USB cable to the I/O module at this time. Use cable ties to neatly bundle excess cable length. Excess cablewill be stored under the printer cover.

    7. Install the printer cover.

    Fig. 38: UP-D55MD Printer Cover with B/W Mounting Platea) Put the printer cover over the printer.b) Align the three captive screws with the three fastener holes cut into the mat.c) Tighten the captive screws

    NOTE Use cable ties to neatly bundle excess cable length. Excess cablewill be stored under the printer cover.Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 44 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 458. Put the B/W printer bracket on the B/W printer mounting plate. Use the applicable in-structions that follow.

    WARNING Use fasteners of the correct length. Using fasteners that are too long may cause the hard-

    ware to contact other components, leading to product damage or a potential electrical safety hazard. Using fas-teners that are too short may not secure the component to its mount, presenting a performance and safety haz-ard.

    9. Set the B/W printer dip switches to: - P91 D settings: 1-ON, 2-OFF, 3-OFF, and 4-OFF- P93 D settings: 1-ON, 2-OFF, 3-ON, and 4-OFF

    P91 D Instructions: P93 D Instructions:

    Put the flat side of the bracket face down on the B/W printer mounting plate of the color printer cover.

    Align the bracket mounting holes to the plate mounting holes.

    Put the P91 B/W printer on top of the P91 mounting bracket.

    Align the B/W printer feet against the mounting bracket.

    Fig. 39: P91 D Positioned on the Bracket

    Point the bottom side of the P93 D print-er up.

    Put the P93 D mounting bracket on the bottom of the P93 D printer.

    Fig. 40: P93 D Mounting Bracket Secure the P93 D printer mounting

    bracket to the P93 D printer. Point the bottom side of the printer fac-

    ing down. Put the Printer assembly on the color

    printer mounting plate.Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 45 of 150Medical Solutions USA, Inc.

  • 46 Hardware Configuration10. Connect the B/W printer cables.

    Fig. 41: Back of Black and White PrinterPos. 1 Dip switchPos. 2 USB cablePos. 3 Power cord

    a) Connect the AC power cable to the back of the P91 D/P93 D printer. b) Connect the USB cable to the back of the P91 D/P93 D printer.

    NOTE Do not connect the USB cable to the I/O module at this time.

    11. Install the B/W printer cover.

    Fig. 42: UP-D55M Color Printer and P91 D or P93 D B/W Printer - Double Configurationa) Put the B/W cover over the P91 D/P93 D printer.b) Tighten the captive screws (of the B/W printer cover) through the mounting bracket, Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 46 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 47to the mounting plate.c) Make sure the back of the P91 D/P93 D printer touches the B/W printer cover stop.

    12. See the Sony User Instruction to perform the following tasks: - Load the ink cartridge- Load the paper tray- Slide the paper tray into the printer

    13. Complete the printer cable connections to the I/O module, see (I/O Module Peripheral Connections / p. 97).

    14. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).

    B/W Printer and VCR 0

    P91 D or P93 D / HS-MD3000Installation Summary1. Install the VCR on the peripheral shelf2. Install the B/W printer on the VCR.3. Complete the cable I/O module connections.4. Complete the VCR configuration.

    NOTE When performing this upgrade on a system that already has a B/Wprinter installed, remove the B/W printer from the peripheral shelf.Do not disconnect the USB cable from the I/O module. The systemhas previously configured the USB port for the B/W printer. There-fore, the same USB port must be used for the B/W printer when theUSB cable is reconnected.Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 47 of 150Medical Solutions USA, Inc.

  • 48 Hardware ConfigurationFig. 43: HS-MD3000 VRC/P91 Dor P93 D B/W Printer - Double ConfigurationPos. 1 VCR bracketPos. 2 VCRPos. 3 VCR Cover with B/W printer platePos. 4 B/W printer with bracket attached to the bottom of the printer (Bracket not shown)Pos. 5 B/W printer cover

    1. Cut the fastener holes into the rubber mat of the peripheral shelf.See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.

    2. Return the mat to the peripheral shelf.3. Align the fastener holes of the mounting bracket to the holes in the peripheral mat.Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 48 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 494. Remove the feet from the bottom of the VCR.

    Fig. 44: VCR Foot Locationsa) Point the bottom side of the VCR facing up. b) Remove the four screws that secure the feet to the VCR.c) Remove the feet.Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 49 of 150Medical Solutions USA, Inc.

  • 50 Hardware Configuration5. Secure the mounting bracket to the VCR.

    Fig. 45: HS-MD3000 Secured to the Mounting Bracketa) Put the mounting bracket on the VCR.b) Align the mounting holes.c) Secure the bracket to the bottom of the VCR.

    WARNING Use fasteners of the correct length. Using fasteners that are too long may cause the hard-

    ware to contact other components, leading to product damage or a potential electrical safety hazard. Using fas-teners that are too short may not secure the component to its mount, presenting a performance and safety haz-ard.Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 50 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 516. Put the VCR assembly on the peripheral shelf and align the mounting holes of the VCR bracket with the mounting holes in the mat.

    Fig. 46: HS-MD3000 Positioned on the OEM Bay7. Connect the power cables to the peripheral AC power outlet.

    Fig. 47: HS-MD3000 Power Cord Connection Pos. 1 OEM power outletsPos. 2 VCR power inlet

    a) Connect the power cable to the VCR AC power inlet and to the lower peripheral AC power outlet.

    b) Connect the printer power cable to the upper peripheral AC outlet.Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 51 of 150Medical Solutions USA, Inc.

  • 52 Hardware Configuration8. Connect the VCR cables. See (VCR I/O Module Cable Connections / p. 98) for details.

    Fig. 48: HS-MD3000 VCR Connections from the I/O Module

    Pos. 1 Connect P128 to RS-232CPos. 2 Connect VIDEO INPos. 3 Connect VIDEO OUTPos. 4 Connect AUDIO INPos. 5 Connect AUDIO OUTPos. 6 Set the 75 VIDEO and the 75 S-VIDEO switches to ON

    NOTE Do not connect the cables to the I/O module at this time. Use cable ties to neatly bundle excess cable length. Excess cablewill be stored under the VCR cover.

    9. Route the B/W printer power cord and USB cable through the cable channel in the VCR cover.

    Fig. 49: Cable ChannelsPos. 1 VCR cover to black and white printerPos. 2 To I/O moduleAntares US03-201.812.01.02.02 Siemens07.05 sd

    Page 52 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 5310. Route the other end of the VCR cables through the I/O module cable channel.11. Route the other end of the B/W printer USB cable through the I/O module cable chan-

    nel.12. Put the B/W printer bracket on the B/W printer mounting plate. Use the applicable in-

    structions that follow.

    WARNING Use fasteners of the correct length. Using fasteners that are too long may cause the hard-

    ware to contact other components, leading to product damage or a potential electrical safety hazard. Using fas-teners that are too short may not secure the component to its mount, presenting a performance and safety haz-ard.Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 53 of 150Medical Solutions USA, Inc.

  • 54 Hardware ConfigurationP91 D Instructions: P93 D Instructions:

    Put the flat side of the bracket face down on the B/W printer mounting plate of the VCR cover.

    Align the bracket mounting holes to the plate mounting holes.

    Put the P91 B/W printer on top of the P91 mounting bracket.

    Align the B/W printer feet against the mounting bracket.

    Fig. 50: P91 D Positioned on the Bracket

    Point the bottom side of the P93 D print-er up.

    Put the P93 D mounting bracket on the bottom of the P93 D printer.

    Fig. 51: P93 D Mounting Bracket Secure the P93 D printer mounting

    bracket to the P93 D printer. Point the bottom side of the printer fac-

    ing down. Put the Printer assembly on the VCR

    mounting plate.Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 54 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 55Fig. 52: VCR Cover with Black and White Mounting PlatePos. 1 VCR CoverPos. 2 Black and white printer mounting plate

    13. Set the B/W printer dip switches: - P91 D settings: 1-ON, 2-OFF, 3-OFF, and 4-OFF- P93 D settings: 1-ON, 2-OFF, 3-ON, and 4-OFF

    Fig. 53: Back of Black and White PrinterPos. 1 Dip switchPos. 2 USB cablePos. 3 Power cordSiemens US03-201.812.01.02.02 Antares07.05 sd

    Page 55 of 150Medical Solutions USA, Inc.

  • 56 Hardware Configuration14. Install the B/W printer cover.

    Fig. 54: Black and White Printer and VCR Installeda) Put the printer cover on the B/W printer.b) Tighten the two printer cover captive screws.

    15. Complete the VCR and printer cable connections to the I/O module, see (I/O Module Peripheral Connections / p. 97).

    16. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).17. Complete the VCR configuration, see (VCR Setup and Configuration / p. 131).

    P91 D or P93 D / SVO9500Installation Summary1. Install the VCR on the peripheral shelf2. Install the B/W printer on the VCR.3. Complete the cable I/O module connections.4. Complete the VCR configuration.

    NOTE When performing this upgrade on a system that already has a B/Wprinter installed, remove the B/W printer from the peripheral shelf.Do not disconnect the USB cable from the I/O module. The systemhas previously configured the USB port for the B/W printer. There-fore, the same USB port must be used for the B/W printer when theUSB cable is reconnected.Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 56 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 571. Cut the fastener holes into the rubber mat of the peripheral shelf.See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.

    2. Secure the mounting bracket to the peripheral shelf.3. Put the VCR assembly on the peripheral shelf by aligning the mounting holes of the

    VCR with the mounting holes in the mat.4. Set the VCR switches:

    - Set the dip switches: 1-OFF, 2-OFF, 3-OFF, 4-OFF, 5- ON, 6-ON.

    Fig. 55: SVO9500 Dip Switch Settings- Set both 75 ohm switches to ON.- For PAL model only, set the voltage to 240V.

    NOTE Only the PAL model includes the 240V switch.

    5. Connect the VCR power cord to the peripheral AC power outlet and to the VCR AC power inlet.Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 57 of 150Medical Solutions USA, Inc.

  • 58 Hardware Configuration6. Connect the VCR cables. See (VCR I/O Module Cable Connections / p. 98) for details.

    Fig. 56: SVC 9500 VCR Connections from I/O ModulePos. 1 Audio In (left and right) Pos. 2 Video In (use either connector)Pos. 3 Audio Out (left and right)Pos. 4 Video OutPos. 5 Dip Switches Pos. 6 RS-232C (connect to P128)Pos. 7 75 (set both switches to ON)

    7. Route the cables through the I/O panel cable routing passage.

    NOTE Do not connect the cables to the I/O module at this time. Use cable ties to neatly bundle excess cable length. Excess cablewill be stored under the VCR cover.

    8. Install the cover.a) Route the B/W printer power cord and USB cable through the cable channel in the

    VCR cover.b) Route the other end of the VCR cables through the I/O module cable channel.c) Route the other end of the B/W printer USB cable through the I/O module cable Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 58 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 59channel.d) Put the cover over the VCR.

    Fig. 57: Cover Installation - Cables Routed Through the Covere) Align and tighten the two captive screws in the cover with the mounting holes.

    Fig. 58: Cover Captive Screw9. Put the B/W printer bracket on the B/W printer mounting plate. Use the applicable in-

    structions that follow.

    WARNING Use fasteners of the correct length. Using fasteners that are too long may cause the hard-

    ware to contact other components, leading to product damage or a potential electrical safety hazard. Using fas-teners that are too short may not secure the component to its mount, presenting a performance and safety haz-ard.Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 59 of 150Medical Solutions USA, Inc.

  • 60 Hardware ConfigurationP91 D Instructions: P93 D Instructions:

    Put the flat side of the bracket face down on the B/W printer mounting plate of the VCR cover.

    Align the bracket mounting holes to the plate mounting holes.

    Put the P91 B/W printer on top of the P91 mounting bracket.

    Align the B/W printer feet against the mounting bracket.

    Fig. 59: P91 D Positioned on the Bracket

    Point the bottom side of the P93 D print-er up.

    Put the P93 D mounting bracket on the bottom of the P93 D printer.

    Fig. 60: P93 D Mounting Bracket Secure the P93 D printer mounting

    bracket to the P93 D printer. Point the bottom side of the printer fac-

    ing down. Put the Printer assembly on the VCR

    mounting plate.Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 60 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 6110. Set the B/W printer dip switches: - P91 D settings: 1-ON, 2-OFF, 3-OFF, and 4-OFF- P93 D settings: 1-ON, 2-OFF, 3-ON, and 4-OFF

    Fig. 61: Back of Black and White PrinterPos. 1 Dip switchPos. 2 USB cablePos. 3 Power cord

    11. Install the B/W printer cover.a) Put the printer cover on the B/W printer.

    Fig. 62: B/W Printer Cover InstallationSiemens US03-201.812.01.02.02 Antares07.05 sd

    Page 61 of 150Medical Solutions USA, Inc.

  • 62 Hardware Configurationb) Tighten the two printer cover captive screws.

    Fig. 63: B/W and VCR Covers Installed12. Complete the VCR and printer cable connections to the I/O module, see (I/O Module

    Peripheral Connections / p. 97).13. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).14. Complete the VCR configuration, see (VCR Setup and Configuration / p. 131).

    Color Printer and VCR 0

    CP770 or CP800 / HS-MD3000Installation Summary1. Adjust the height of the monitor/control panel.2. Install the color printer on the peripheral shelf3. Install the VCR printer on the color printer.4. Complete the cable I/O module connections.5. Complete the VCR configuration.

    NOTE When performing this upgrade on a system that already has a col-or printer installed, remove the color printer from the peripheralshelf.Do not disconnect the USB cable from the I/O module. The systemhas previously configured the USB port for the color printer.Therefore, the same USB port must be used for the color printerwhen the USB cable is reconnected.Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 62 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 631. Adjust the height of the monitor/control panel.Press the height adjustment knob to set the monitor/control panel to the highest position.

    2. Remove the upper, right-front panel.a) Loosen the 1/4-turn captive screw.

    Fig. 64: Locking Bolt in the Storage Loctionb) Grip the transducer port insets of the panel and pull to disengage the ball snaps

    3. Remove the left-front panel.a) Loosen the 1/4-turn captive screw.

    Fig. 65: Upper, Left-Front Panel Captive Screwb) Grip the air flow vent insets of the panel and pull the panel to disengage the ball

    snaps.4. Remove the bottom post cover.

    a) Loosen the four screws located on either side of the post.Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 63 of 150Medical Solutions USA, Inc.

  • 64 Hardware Configurationb) Remove the post cover.5. Install the locking bolt.

    a) Remove the locking bolt from the locking bolt storage location.b) Tighten the locking bolt to the post to lock the height adjustment on the highest posi-

    tion6. Attach the bottom post cover.

    a) Place the cover with the painted, chamfered side facing up.b) Tighten the four post cover screw that secure the cover to the system. Locate two

    screws on either side of the post.7. Install the right-front panel.

    a) Align the panel ball snap receptacles with the system ball snaps and press the pan-el to engage the ball snaps.

    b) Tighten the 1/4-turn captive screw.8. Install the left-front panel.

    a) Align the panel ball snap receptacles with the system ball snaps and press the pan-el to engage the ball snaps.

    b) Tighten the 1/4-turn captive screw.9. Cut the fastener holes into the rubber mat of the peripheral shelf.

    See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.10. Return the mat to the peripheral shelf.11. Align the fastener holes of the mounting bracket to the holes in the peripheral mat.12. Remove the feet and the shipment locking screw from the bottom of the printer.

    a) Point the bottom side of the printer face up.b) Remove the four screws that secure the feet to the printer and remove the feet.

    Fig. 66: UP-D23M Color Printer ConnectionsAntares US03-201.812.01.02.02 Siemens07.05 sd

    Page 64 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 65c) Remove the shipment locking screw.13. Secure the mounting bracket to the color printer.

    a) Put the mounting bracket on the printer.b) Align the mounting holes.c) Secure the bracket to the bottom of the color printer.

    WARNING Use fasteners of the correct length. Using fasteners that are too long may cause the hard-

    ware to contact other components, leading to product damage or a potential electrical safety hazard. Using fas-teners that are too short may not secure the component to its mount, presenting a performance and safety haz-ard.

    14. Put the color printer assembly on the peripheral shelf and align the mounting holes of the color printer with the mounting holes in the mat.

    15. Set the color printer dip switches as follows:

    16. Connect the color printer cables.a) Connect the color printer AC power cable to the color printer AC power inlet and to

    the lower peripheral AC power outlet. b) Connect the VCR AC power cable to the upper peripheral AC power outlet.c) Connect the color printer USB cable to the back of the color printer. d) Route the other end of the color printer USB cable through the peripheral shelf ca-

    ble passage to the I/O module.e) Route the I/O module end of the VCR USB cable through the peripheral shelf cable

    passage to the I/O module.

    Fig. 67: CP770 Dip Switch Settings to OFF1-OFF, 2-OFF, 3-OFF, 4-OFF,5-OFF, 6-OFF, 7-OFF, and 8-OFF.

    Fig. 68: CP800 Dip Switch Settings1-ON, 2-OFF, 3-OFF, 4-OFF,5-OFF, 6-OFF, 7-OFF, and 8-ON.Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 65 of 150Medical Solutions USA, Inc.

  • 66 Hardware ConfigurationNOTE Do not connect the USB cables to the I/O module at this time, un-less the system previously had one of the printers installed on theperipheral bay. Then, connect the previously installed printer tothe same port it was connected to before.

    17. Remove the feet and the shipment locking screw from the bottom of the VCR.a) Point the bottom side of the VCR facing up.b) Remove the four screws that secure the feet to the VCR.c) Remove the feet.

    18. Secure the mounting bracket to the VCR.a) Put the mounting bracket on the VCR.b) Align the mounting holes.c) Secure the bracket to the bottom of the VCR.

    WARNING Use fasteners of the correct length. Using fasteners that are too long may cause the hard-

    ware to contact other components, leading to product damage or a potential electrical safety hazard. Using fas-teners that are too short may not secure the component to its mount, presenting a performance and safety haz-ard.

    19. Put the VCR assembly on the color printer.20. Connect the VCR cables.

    a) Connect the power cable to the VCR and to the Antares peripheral AC power outlet.b) Connect the AUDIO IN left and right channel cables to the back of the VCR.c) Connect the AUDIO OUT left and right channel cables to the back of the VCR.d) Set the VIDEO and S VIDEO switches to the on position.e) Connect the VIDEO in/out cables to the back of the VCR.f) Connect the S VIDEO in/out cables to the back of the VCR.g) Connect the USB cable to the back of the VCR.h) Connect the RS232 C cable to the back of the VCR.

    21. Install the cover.a) Put the cover over the VCR.b) Align the four captive screws in the cover with the mounting holes.c) Tighten the four captive screws.

    22. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).23. Complete the VCR configuration, see (VCR Setup and Configuration / p. 131).Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 66 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 6724. Complete the VCR and printer cable connections to the I/O module, see (VCR I/O Mod-ule Cable Connections / p. 98).

    CP770 or CP800 / SVO9500Installation Summary1. Adjust the height of the monitor/control panel.2. Install the color printer on the peripheral shelf3. Install the VCR printer on the color printer.4. Complete the cable I/O module connections.5. Complete the VCR configuration.

    NOTE When performing this upgrade on a system that already has a col-or printer installed, remove the color printer from the peripheralshelf.Do not disconnect the USB cable from the I/O module. The systemhas previously configured the USB port for the color printer.Therefore, the same USB port must be used for the color printerwhen the USB cable is reconnected.

    1. Adjust the height of the monitor/control panel.Press the height adjustment knob to set the monitor/control panel to the highest position.

    2. Remove the upper, right-front panel.a) Loosen the 1/4-turn captive screw.

    Fig. 69: Locking Bolt in the Storage Loctionb) Grip the transducer port insets of the panel and pull to disengage the ball snapsSiemens US03-201.812.01.02.02 Antares07.05 sd

    Page 67 of 150Medical Solutions USA, Inc.

  • 68 Hardware Configuration3. Remove the left-front panel.a) Loosen the 1/4-turn captive screw.

    Fig. 70: Upper, Left-Front Panel Captive Screwb) Grip the air flow vent insets of the panel and pull the panel to disengage the ball

    snaps.4. Remove the bottom post cover.

    a) Loosen the four screws located on either side of the post.b) Remove the post cover.

    5. Install the locking bolt.a) Remove the locking bolt from the locking bolt storage location.b) Tighten the locking bolt to the post to lock the height adjustment on the highest posi-

    tion6. Attach the bottom post cover.

    a) Place the cover with the painted, chamfered side facing up.b) Tighten the four post cover screw that secure the cover to the system. Locate two

    screws on either side of the post.7. Install the right-front panel.

    a) Align the panel ball snap receptacles with the system ball snaps and press the pan-el to engage the ball snaps.

    b) Tighten the 1/4-turn captive screw.8. Install the left-front panel.

    a) Align the panel ball snap receptacles with the system ball snaps and press the pan-el to engage the ball snaps.

    b) Tighten the 1/4-turn captive screw.Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 68 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 699. Cut the fastener holes into the rubber mat of the peripheral shelf.See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.

    10. Return the mat to the peripheral shelf.11. Align the fastener holes of the mounting bracket to the holes in the peripheral mat.12. Put the plastic paper tray on the bracket, aligning the dashed line with the front edge of

    the bracket.13. Position the printer on the peripheral shelf.

    a) Put the color printer on the plastic paper tray and bracket.b) Align the printer so the back panel contacts the two sheet-metal tabs. The front

    sheet-metal tabs should sit in the grooves on the bottom of the printer, a few centi-meters from the front panel.

    Fig. 71: Color Printer Foot Placement on Bracket14. Set the color printer dip switches as follows:

    Fig. 72: CP770 Dip Switch Settings to OFF1-OFF, 2-OFF, 3-OFF, 4-OFF,5-OFF, 6-OFF, 7-OFF, and 8-OFF.

    Fig. 73: CP800 Dip Switch Settings1-ON, 2-OFF, 3-OFF, 4-OFF,5-OFF, 6-OFF, 7-OFF, and 8-ON.Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 69 of 150Medical Solutions USA, Inc.

  • 70 Hardware Configuration15. Connect the color printer cables.

    Fig. 74: CP770 and CP800 Cable Connections on Bacl of Printera) Connect the color printer AC power cable to the color printer AC power inlet and to

    the lower peripheral AC power outlet. b) Connect the VCR printer AC power cable to the upper peripheral AC power outlet.c) Connect the parallel end of the parallel-to-USB cable to the back of the color printer. d) Route the USB end of the parallel-to-USB cable through the peripheral shelf cable

    passage to the I/O module.e) Route the I/O module end of the VCR USB cable through the peripheral shelf cable

    passage to the I/O module.

    NOTE Do not connect the USB cables to the I/O module at this time, un-less the system previously had one of the printers installed on theperipheral bay. Then, connect the previously installed printer tothe same port it was connected to before.Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 70 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 7116. Put the mounting bracket on the printer.

    Fig. 75: SVOC9500 Barcket Placement on the CP770 or CP800Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 71 of 150Medical Solutions USA, Inc.

  • 72 Hardware Configuration17. Position the VCR on the VCR bracket and color printer. Align the rear feet of the VCR against the tabs of the retaining bracket.

    Fig. 76: SVO9500 Positioned on the CP770 or CP80018. Install the cover.

    a) Put the cover over the VCR.Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 72 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 73b) Align the four captive screws in the cover with the mounting holes.c) Do not tighten the four captive screws.

    Fig. 77: Install the Cover - SVO9500/CP770 or CP80019. Connect the VCR cables.

    Fig. 78: SVC 9500 VCR Connections from I/O ModulePos. 1 Audio In (left and right) Pos. 2 Video In (use either connector)Pos. 3 Audio Out (left and right)Pos. 4 Video OutPos. 5 Dip Switches Pos. 6 RS-232C (connect to P128)Pos. 7 75 (set both switches to ON)Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 73 of 150Medical Solutions USA, Inc.

  • 74 Hardware Configurationa) Connect the power cable to the VCR and to the Antares peripheral AC power outlet.b) Connect the AUDIO IN left and right channel cables to the back of the VCR.c) Connect the AUDIO OUT left and right channel cables to the back of the VCR.d) Set the VIDEO and S VIDEO switches to the on position.e) Connect the VIDEO in/out cables to the back of the VCR.f) Connect the S VIDEO in/out cables to the back of the VCR.g) Connect the USB cable to the back of the VCR.h) Connect the RS232 C cable to the back of the VCR.

    20. Tighten the four captive screws to secure the cover.

    Fig. 79: Tighten the Cover -Captive Screws - SVO9500/CP770 or CP800Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 74 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 7521. Install the back peripheral cover and tighten the four screws.

    Fig. 80: Install the Back Cover - SVO9500/CP770 or CP80022. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).23. Complete the VCR configuration, see (VCR Setup and Configuration / p. 131).24. Complete the VCR and printer cable connections to the I/O module, see (I/O Module

    Peripheral Connections / p. 97).

    UP-D23MD / HS-MD 3000Installation Summary1. Adjust the height of the monitor/control panel.

    Press the height adjustment knob to set the monitor/control panel to the highest position.

    2. Install the VCR on the peripheral shelf3. Install the color printer on the VCR.4. Complete the cable I/O module connections.5. Complete the VCR configuration.Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 75 of 150Medical Solutions USA, Inc.

  • 76 Hardware ConfigurationNOTE When performing this upgrade on a system that already has a col-or printer installed, remove the color printer from the peripheralshelf.Do not disconnect the USB cable from the I/O module. The systemhas previously configured the USB port for the color printer.Therefore, the same USB port must be used for the color printerwhen the USB cable is reconnected.

    Fig. 81: HS-MD3000 VRC/UPD-23MD Color Printer - Double ConfigurationPos. 1 VCR bracketPos. 2 VCRPos. 3 Color printer bracketPos. 4 Color printerPos. 5 Cover

    1. Adjust the height of the monitor/control panel.Press the height adjustment knob to set the monitor/control panel to the highest position.Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 76 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 772. Remove the upper, right-front panel.a) Loosen the 1/4-turn capptive screw.

    Fig. 82: Upper, Right-Front Panel Captive Screwb) Grip the transducer port insets of the panel and pull to disengage the ball snaps

    3. Remove the left-front panel.a) Loosen the 1/4-turn capptive screw.

    Fig. 83: Upper, Left-Front Panel Captive Screwb) Grip the air flow vent insets of the panel and pull the panel to disengage the ball

    snaps.4. Remove the bottom post cover.

    a) Loosen the four screws located on either side of the post.Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 77 of 150Medical Solutions USA, Inc.

  • 78 Hardware Configurationb) Remove the post cover.

    Fig. 84: Bottom Post Cover Screws5. Install the locking bolt.

    a) Remove the locking bolt from the locking bolt storage location.

    Fig. 85: Locking Bolt in the Storage Loctionb) Tighten the locking bolt to the post to lock the height adjustment on the highest posi-Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 78 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 79tion.

    Fig. 86: Locking Height Adjustment6. Attach the bottom post cover.

    a) Place the cover with the painted, chamfered side facing up.b) Tighten the four post cover screw that secure the cover to the system. Locate two

    screws on either side of the post.7. Install the right-front panel.

    a) Align the panel ball snap recepticals with the system ball snaps and press the panel to engage the ball snaps.

    b) Tighten the 1/4-turn captive screw.8. Install the left-front panel.

    a) Align the panel ball snap recepticals with the system ball snaps and press the panel to engage the ball snaps.

    b) Tighten the 1/4-turn captive screw.9. Cut the fastener holes into the rubber mat of the peripheral shelf.

    See (Peripheral Shelf Mounting Hole Locations / p. 95) for the mounting hole locations.10. Return the mat to the peripheral shelf.11. Align the fastener holes of the mounting bracket to the holes in the peripheral mat.Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 79 of 150Medical Solutions USA, Inc.

  • 80 Hardware Configuration12. Remove the feet from the bottom of the VCR.

    Fig. 87: VCR Foot Locationsa) Point the bottom side of the VCR facing up.b) Remove the four screws that secure the feet to the VCR.c) Remove the feet.Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 80 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 8113. Secure the mounting bracket to the VCR.

    Fig. 88: HS-MD3000 Secured to the Mounting Bracketa) Put the mounting bracket on the VCR.b) Align the mounting holes.c) Secure the bracket to the bottom of the VCR.

    WARNING Use fasteners of the correct length. Using fasteners that are too long may cause the hard-

    ware to contact other components, leading to product damage or a potential electrical safety hazard. Using fas-teners that are too short may not secure the component to its mount, presenting a performance and safety haz-ard.Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 81 of 150Medical Solutions USA, Inc.

  • 82 Hardware Configuration14. Put the VCR assembly on the peripheral shelf and align the mounting holes of the VCR bracket with the mounting holes in the mat.

    Fig. 89: HS-MD3000 Positioned on the OEM Bay15. Connect the VCR power cable to the VCR and to the lower peripheral AC power outlet.

    Fig. 90: HS-MD3000 Power Cord Connection Pos. 1 OEM power outletsPos. 2 VCR power inletAntares US03-201.812.01.02.02 Siemens07.05 sd

    Page 82 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 8316. Connect the VCR cables. See (VCR I/O Module Cable Connections / p. 98) for details.

    Fig. 91: HS-MD3000 VCR Connections from the I/O ModulePos. 1 Connect P128 to RS-232CPos. 2 Connect VIDEO INPos. 3 Connect VIDEO OUTPos. 4 Connect AUDIO INPos. 5 Connect AUDIO OUTPos. 6 Set the 75 VIDEO and the 75 S-VIDEO switches to ON

    17. Route the other end of the VCR cables through the I/O module cable channel.

    NOTE Do not connect the VCR cables to the I/O module at this time. Use cable ties to neatly bundle excess cable length. Excess cablewill be stored under the VCR cover.Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 83 of 150Medical Solutions USA, Inc.

  • 84 Hardware Configuration18. Secure the color printer to the printer mounting bracket.

    Fig. 92: UP-D23M Color Printer Positioned on the MD3000 VCRa) Point the bottom side of the printer face up.b) Put the printer mounting bracket on the printer.c) Tighten the bracket captive screws to secure the printer to the bracket.

    19. Put the printer assembly on the VCR.

    Fig. 93: UP-D23M Color Printer Positioned on the MD3000 VCRAntares US03-201.812.01.02.02 Siemens07.05 sd

    Page 84 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 8520. Connect the color printer cables.

    Fig. 94: UP-D23M Color Printer Connections

    NOTE (Fig. 94 / p. 85) does not show the VCR with the VCR cables con-nected. However, the VCR cables were connected in a previoussteps, see (VCR I/O Module Cable Connections / p. 98).

    a) Connect the printer power cable into the upper peripheral outletb) Connect the printer USB cable to the back of the printer

    21. Route the other end of the printer USB cable through the I/O module cable channel.

    NOTE Do not connect the USB cable to the I/O module at this time. Use cable ties to neatly bundle excess cable length. Excess cablewill be stored under the VCR cover.Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 85 of 150Medical Solutions USA, Inc.

  • 86 Hardware Configuration22. Install the cover.

    Fig. 95: UP-D23M Color Printer and MD3000 VCR Installeda) Put the printer cover over the color printer/VCR assembly.b) Verify the front lip of the cover is against the front of the printer/VCR assembly.c) Tighten the three printer cover captive screws.

    23. See the Sony User Instruction to perform the following tasks: - Load the ink cartridge

    NOTE Power on the system prior to loading the ink cartridge. This resetsthe gears to their default location, permitting the ink cartridge to beinserted.

    - Load the paper tray- Slide the paper tray into the printer

    24. Complete the VCR and printer cable connections to the I/O module, see (I/O Module Peripheral Connections / p. 97).

    25. Complete the VCR configuration, see (VCR Setup and Configuration / p. 131).26. Complete the printer configuration, see (Printer Setup and Configuration / p. 99).

    UP-D55MD / HS-MD3000Installation Summary1. Adjust the height of the monitor/control panel.

    Press the height adjustment knob to set the monitor/control panel to the highest position.Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 86 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 872. Install the color printer on the peripheral shelf3. Install the VCR on the color printer.4. Complete the cable I/O module connections.5. Complete the VCR configuration.

    NOTE When performing this upgrade on a system that already has a col-or printer installed, remove the color printer from the peripheralshelf.Do not disconnect the USB cable from the I/O module. The systemhas previously configured the USB port for the color printer.Therefore, the same USB port must be used for the color printerwhen the USB cable is reconnected.

    Fig. 96: UP-D55M Color Printer/HS-MD3000 VRC - Double ConfigurationPos. 1 Color printer bracketPos. 2 Color printerPos. 3 VCR bracketPos. 4 VCRPos. 5 Cover

    1. Adjust the height of the monitor/control panel.Press the height adjustment knob to set the monitor/control panel to the highest position.Siemens US03-201.812.01.02.02 Antares07.05 sd

    Page 87 of 150Medical Solutions USA, Inc.

  • 88 Hardware Configuration2. Remove the upper, right-front panel.a) Loosen the 1/4-turn capptive screw.

    Fig. 97: Upper, Right-Front Panel Captive Screwb) Grip the transducer port insets of the panel and pull to disengage the ball snaps

    3. Remove the left-front panel.a) Loosen the 1/4-turn capptive screw.

    Fig. 98: Upper, Left-Front Panel Captive Screwb) Grip the air flow vent insets of the panel and pull the panel to disengage the ball

    snaps.4. Remove the bottom post cover.

    a) Loosen the four screws located on either side of the post.Antares US03-201.812.01.02.02 Siemens07.05 sd

    Page 88 of 150Medical Solutions USA, Inc.

  • Hardware Configuration 89b) Remov