AOE 3054 Digital Measurements: Data Acquisition with LabView Credits to Borgoltz, Devenport, and Edwards for some content 1

AOE 3054 Digital Measurements: Data Acquisition with LabView Credits to Borgoltz, Devenport, and Edwards for some content 1

Embed Size (px)

Citation preview

  • Slide 1
  • Slide 2
  • AOE 3054 Digital Measurements: Data Acquisition with LabView Credits to Borgoltz, Devenport, and Edwards for some content 1
  • Slide 3
  • Goals of the session Understand the basics of making the NI myDAQ work for controlling an experiment Build the data acquisition and control techniques needed to digitally run Experiment 6 in LabVIEW Basic scope Storing scope measurements Calibrating and ultimately controlling the function generator 2
  • Slide 4
  • Digital Measurements Lab Agenda 1.Experiment 6 Digital introduction 2.Measuring function generator input and beam response using myDAQ 3.Calibrating function generators 4.Find natural frequency of beam using myDAQ 5.myDAQ Resolution Example 6.Modify Homework codes 3
  • Slide 5
  • Experiment 6 Digital Introduction In the second Instrumentation Lab (Experiment 6a), you manually controlled a function generator to excite a beam and used an oscilloscope to measure the response of that beam. Week 5s Instrumentation Lab is essentially a redo of the first Experiment 6, but will incorporate new digital measurement techniques to automate most of the data taking. 4
  • Slide 6
  • Experiment 6 Digital Introduction Specifically, you will be using the myDAQ to output a voltage signal that will control the function generator. The myDAQ will also measure the function generator output as well as the output from the proximeter. All operations will be controlled via Labview, using a code that builds off of the homework assignments and will be completed in Instrumentation Lab 4. Further details of Experiment 6 Digital can be found on the course website. 5
  • Slide 7
  • Digital Measurements Lab Agenda 1.Experiment 6 Digital introduction 2.Measuring function generator input and beam response using myDAQ 3.Calibrating function generators 4.Find natural frequency of beam using myDAQ 5.myDAQ Resolution Example 6.Modify Homework codes 6
  • Slide 8
  • Homework 3 VI Your Homework 3 VIs utilized the 2 analog input channels on the myDAQ to measure two signals. We will use this code again to measure and display two signals- an excitation signal (function generator) and response signal (proximeter) 7
  • Slide 9
  • Connect Experiment 6 Components Connect the Function Generator to the power amplifier, using a BNC T-connector. Connect the Power Supply to the proximeter and set it up for the correct output voltage. Remember to take into account whether you are using the NEW or OLD proximeter. Refer to previous lecture slides for instructions on properly connecting the devices. 8
  • Slide 10
  • myDAQ Connections Connect your myDAQ to your computer using the USB port. Open LabView and your functional Homework 3 VI 9
  • Slide 11
  • myDAQ Connections BEFORE turning any of the equipment on, make sure that you do not supply more than 20V to the myDAQ through its analog inputs (myDAQ Overvoltage protection: +/-30V, 20 Vrms) This will require connecting the excitation and response signals to the oscilloscope and measuring amplitudes before connecting the myDAQ. A good first step is to turn the function generator amplitude control all the way down. 10
  • Slide 12
  • myDAQ Connections: Excitation Attach a BNC-to-BNC-probe connector to the T-connector on the function generator 11 To Amplifier BNC to BNC probe (to oscilloscope)
  • Slide 13
  • myDAQ Connections: Excitation Attach the BNC-to-BNC connection from the function generator to a T-connector on CH1 of the oscilloscope, and a BNC-to-clipping-probe connector to the T-connector. 12 From function generator BNC to clipping probe (to myDAQ)
  • Slide 14
  • myDAQ Connections: Excitation Clip the two ends of the probe to wires and attach to the myDAQ AI0 channel. Make sure the red clip goes to the 0+ channel, and black to the 0-. 13 AI0+ AI0-
  • Slide 15
  • myDAQ Connections: Response Attach the output BNC connector of the proximeter to a BNC-to-clipping- probe connector using a T-connector Then connect the T-connector to Channel 2 on the oscilloscope 14 Proximeter Output BNC to clipping probe (to myDAQ) To Ch2 on Oscilloscope
  • Slide 16
  • myDAQ Connections: Response Clip the two ends of the probe coming from the proximeter to wires and attach to the myDAQ AI1 channel. Make sure the red clip goes to the 1+ channel, and black to the 1-. 15
  • Slide 17
  • Final myDAQ Connections: Excitation+Response 16 Function Generator Proximeter
  • Slide 18
  • Verify Connections Before turning equipment on, verify all of your connections are correct and set to the correct voltage. Ask your TA if you have any questions. Then disconnect the four clip-ons connecting the myDAQ. Verify that the function generator is set to output no more than 2 V. 17
  • Slide 19
  • Turn on Equipment Once the setup is verified, turn on the function generator, amplifier, multimeter, and power supply. Verify that the signals look good and within range on the oscilloscope. Once you established that both excitation and response are under 20V amplitude, turn all the equipment off. Reconnect the myDAQ. You can now turn the equipment back on, your myDAQ is ready for acquisition! 18
  • Slide 20
  • Using Homework 3 VI The Homework 3 VI and subVI should already be configured to read the correct channels on the myDAQ. Verify this by opening your subVI block diagram. 19
  • Slide 21
  • Verify myDAQ Channels Double click the DAQ Assistant Express VI. The following window then appears: 20
  • Slide 22
  • Verify myDAQ Channels Click the Details button under Configuration 21
  • Slide 23
  • Verify myDAQ Channels 22 As expected, the myDAQ is reading the Excitation signal from channel AI0 and Response signal from AI1.
  • Slide 24
  • Set Test Conditions 23 Close out of the DAQ Assistant screen and return to the Main VI front panel. Set the amplitude of the function generator signal up to about 2 V using the AMPL knob. Make sure it is pulled out (because when it is pushed in, it supplies a voltage up to 20 V). Set the frequency of the function generator to about 12 Hz.
  • Slide 25
  • Introduction to Aliasing Set your VI to take 10 samples at a rate of 10 Samples/s. The output should look similar to this: 24
  • Slide 26
  • Introduction to Aliasing Note that even though the excitation signal is set to 12 Hz, the sampling rate in LabView is too low to accurately resolve this signal. Likewise, the response cannot be accurately characterized either. This is known as aliasing, and was introduced in the previous homework. It will be discussed extensively in the next lecture and the 4th Instrumentation Lab, and is a major concern in signal analysis.
  • Slide 27
  • Grounding the myDAQ Some computers have grounding issues when using the myDAQs to measure a voltage, including older Fujitsu models. This causes noisy looking signals, such as that seen below:
  • Slide 28
  • Grounding the myDAQ To fix this problem, connect a banana to clipping probe cable from the GND of the Power Supply to the AGND port of the Analog Input section of the myDAQ. This leads to a much cleaner signal. A picture of an example connection is found on the next slide
  • Slide 29
  • Grounding the myDAQ Power Supply GND
  • Slide 30
  • Increase Sampling Rate Stop the program and increase the sampling rate to 1000 Samples/S, and increase the number of samples to 1000. Now, at this higher sampling rate, the correct signals can be determined (see front panel screenshot on next slide). 29
  • Slide 31
  • Increased Sampling Rate 30
  • Slide 32
  • Find Approximate Natural Frequency Adjust the function generator frequency until it reaches near the natural frequency. Note: It will be challenging to settle on the exact frequency. The function generators are sensitive. In this scenario, it is easiest to spot the natural frequency using the Lissajous plot. At the natural frequency, the Lissajous plot forms an ellipse with vertical and horizontal axes. Write down the frequency you settled on; we will come back to this later. 31
  • Slide 33
  • Front Panel at Natural Frequency 32 Phase difference of 90 degrees Vertical ellipse.
  • Slide 34
  • Turn Off Equipment Shut off all of the equipment to prepare for the next phase of todays lab. 33
  • Slide 35
  • Applications to Dynamic Beam Two inputs allow force and displacement voltages to be measured. Voltages converted to force and distance. Dynamic flexibility and spring constants measured at low frequencies and the result displayed. Data file saved and analyzed. 34
  • Slide 36
  • Digital Measurements Lab Agenda 1.Experiment 6 Digital introduction 2.Measuring function generator input and beam response using myDAQ 3.Calibrating function generators 4.Find natural frequency of beam using myDAQ 5.myDAQ Resolution Example 6.Modify Homework codes 35
  • Slide 37
  • What else is needed for Exp 6? We need a method for controlling the Function Generator! This will allow us to sweep through a range of frequencies and automate the experiment. To start understanding how to control the frequency generator, we will begin by calibrating it. 36
  • Slide 38
  • Varying Function Generator Frequency The function generator frequency can be varied by sending an analog output to the BNC connector on the front of the function generator. The DC voltage signal will be produced by the Analog Output of the DAQ. Varying the DAQ voltage will vary the output frequency of the function generator. Such set-up could be particularly useful to determine the natural frequency. Calibrating the Frequency Generator will produce the relationship between myDAQ voltage and output frequency. Now we need to set up the DAQ Assistant to Output an analog signal. 37
  • Slide 39
  • Rename Labview VI We will now build off of the code used so far in the lab. Save your current VI-the main VI from Homework 3-as a new file (i.e. Name_InstrumentationLab3, etc.) 38
  • Slide 40
  • Configuring the DAQ for Analog Output From the Block Diagram of your VI, Right click and select Input, DAQ Assistant. 39
  • Slide 41
  • Place VI on the Diagram 40
  • Slide 42
  • Configure the DAQ Assistant Select generate signals, this is the output portion of the DAQ 41
  • Slide 43
  • Configure the DAQ Assistant Select Analog output 42
  • Slide 44
  • Configure the DAQ Assistant Select Voltage 43
  • Slide 45
  • Select physical output channel Select an output port 44
  • Slide 46
  • Configure the DAQ Assistant Select 1 Sample from the generation mode, finished! If the window below does not show up, change the window size and the window should appear! 45
  • Slide 47
  • DAQ Assistant for Analog Output Now we need to create to input DC signal for the DAQ. This can be done by attaching a control to the data input 46
  • Slide 48
  • DAQ Assistant for DC output Add a while loop around the DAQ Assistant for continuous output and then connect a stop button to the stop port on the DAQ and the stop button on the while loop so that the while loop will smoothly shut down the DAQ when you press stop. You are now ready to connect the myDAQ to the function generator. 47
  • Slide 49
  • Analog Output Terminals Connect bare wires to the A0 0 and AGND terminals on the DAQ by inserting the bare ends of the leads into the appropriate terminals and tightening the screw with the included screwdriver. Connect a BNC-to-clipping-probe connector to the DAQ (RED clip to the lead for A0 0 and a BLACK clip to the lead for AGND). Note that for all of these connections discussed today, either a BNC to alligator clip or BNC-to-clipping-probe connector will work 48
  • Slide 50
  • Process: How to control the function generator? 1.You will use the DC DAQ output to obtain the relationship between the voltage supplied to the function generator and the frequency output by the generator. This relationship is called the gain or calibration. 2.To do so, you will connect the DAQ analog output to the function generator input (VCF port) and use the calibration to control the generator with the DAQ. 49
  • Slide 51
  • Setup: How to control the function generator? 1.Place a BNC T-connector on the VCF input of the function generator. 2.Connect the analog output port you selected to the VCF input of the function generator using a BNC cable/alligator clip and the two leads in the myDAQ. 3.On the other side of the T-connector, connect a BNC to BNC cable. 50 BNC to BNC connector From myDAQ
  • Slide 52
  • Setup: How to control the function generator? 4.You will want to connect your voltmeter to the DAQ as well. Disconnect the banana plug currently in the multimeter, which connected the power supply voltage to the proximeter. Insert a new banana plug to BNC connector and connect the BNC coming from the function generator VCF input to the multimeter. 5.Using the LabView VI, vary the DC voltage output and see how the function generator frequency responds. Verify that the multimeter voltage matches the output voltage specified in LabView, and also specify that the frequency displayed on the function generators matches the frequency measured by the AI0 channel in the myDAQ (displayed as the Excitation Frequency on the VI front panel). 51 From function generator VCF input Power supply to proximeter connection
  • Slide 53
  • Exercise: Calibrating the function generator The frequency of the function generator can be commanded based upon voltage You will open an Excel file to store information on the calibration (i.e. record voltage input and frequency output) Use the DAQ to control the function generator 52
  • Slide 54
  • Calibrating the function generator To determine this relationship, you need to do a calibration. You need to record the frequency output for different voltage inputs: Make sure your function generator is set to sine wave, with the RANGE knob set to a 10 Hz order of magnitude. Adjust the FREQUENCY knob to output a 14 Hz sine wave before beginning calibration. Run the DAQ VI and record the frequency from the DAQ for ~10-20 voltages. Adjust the input voltages between +/-2 V. This should lead to function generator frequencies from about.5 Hz up to about 27.5 Hz You will have to adjust your sampling scheme to obtain accurate readings. Record and plot the resulting data in excel and obtain the linear relationship, i.e. the calibration. 53
  • Slide 55
  • Calibrations The calibration curve should look something like this. The offset will be different depending on your station. This gives you a relationship between in the input voltage (x) and the frequency (y) and allows you to write a conversion between frequency and voltage. Thus in LabVIEW, the user can enter a frequency y and the VI will convert this value to voltage using (y-b)/m (if the calibration is y=mx+b) and then send this to the DAQ to control the function generator. 54 y-intercept will be equal to the frequency the function generator was set to before calibrating. In this image, the function generator started at 10 Hz. Note that you will be starting at 14 Hz.
  • Slide 56
  • How to control the function generator with LabVIEW At this point, your code is set to provide a voltage from the DAQ to the function generator, which in turns produces a frequency output that is function of its calibration. 55 Since you know the calibration equation, you will be able to change the code so that the user inputs a frequency that the code will convert to a voltage value to be fed to the function generator. You can therefore change the Offset control seen above using the arithmetic operations you learned for Homework 2 and the coefficients of the calibration you just measured.
  • Slide 57
  • How to control the function generator with LabVIEW You can use the sub-VI from Homework 2 to modify the Offset control in the current code. To confirm you have written your code and performed your calibration correctly, set a frequency in LabVIEW and measure the excitation signal on the scope. If you have done everything correctly, the two should match. 56
  • Slide 58
  • Digital Measurements Lab Agenda 1.Experiment 6 Digital introduction 2.Measuring function generator input and beam response using myDAQ 3.Calibrating function generators 4.Find natural frequency of beam using myDAQ 5.myDAQ Resolution Example 6.Modify Homework codes 57
  • Slide 59
  • Reconnect Equipment The myDAQ output voltage can control the output frequency of the function generator much more precisely than manually adjusting the knob. We will now try to more accurately measure the natural beam frequency. Disconnect the banana plug and BNC to BNC cable in the multimeter from the function generator calibration, and reconnect the banana plug linking the power supply to the proximeter input. Return the BNC to BNC cable to the wall rack. 58
  • Slide 60
  • Turn on Equipment Turn on the power supply and amplifier Using the LabView VI, adjust the desired frequency of the function generator to find the beams natural frequency. See how much closer to a phase of 90 degrees you can get, compared to manually adjusting the frequency knob. 59
  • Slide 61
  • Power Down Equipment Next, turn off all equipment, disconnect all cables and wires, and return all equipment to their respective storage locations. Leave two wires, a BNC-to-clipping-probe connector, and a BNC to BNC connector at your table. 60
  • Slide 62
  • Digital Measurements Lab Agenda 1.Experiment 6 Digital introduction 2.Measuring function generator input and beam response using myDAQ 3.Calibrating function generators 4.Find natural frequency of beam using myDAQ 5.myDAQ Resolution Example 6.Modify Homework codes 61
  • Slide 63
  • myDAQ Generate Sine Wave Next we will go through an example showing the limitations of a digital signal. You have seen that the myDAQ is capable of producing an analog signal through its analog output. So what if we were to try to replace the function generator altogether with the myDAQ? We will have the myDAQ generate a sine wave, and compare that to a sine wave generated by the function generator 62
  • Slide 64
  • myDAQ Resolution 63
  • Slide 65
  • Download Code Download the OutputSineWave.vi from the Scholar site, under Resources-> Instrumentation->Lab 3 64
  • Slide 66
  • Connect Devices Use a BNC to BNC connector from the MAIN output port on the function generator to CH 1 on the oscilloscope. Connect two wires into the myDAQ for analog output, in the same manner as described on Slide 48. Use a BNC-to-clipping-probe to connect the wire terminals of the myDAQ to CH 2 on the oscilloscope. Make sure the red clip corresponds to the AO0 wire, and black clip to the AGND. 65
  • Slide 67
  • Device Connections 66
  • Slide 68
  • Run VI Turn on the oscilloscope and function generator. Set the function generator to an output frequency of about 4 Hz with a very low amplitude (always keep in mind the voltage limit on your DAQ). In the VI front panel, set the frequency at 4 Hz and amplitude to 0.02 V. Run the code, and adjust the oscilloscope screen to accommodate the signals. 67
  • Slide 69
  • Oscilloscope Display 68 Function Generator Signal myDAQ Signal
  • Slide 70
  • Digital Resolution Limitations As you can see, the blue myDAQ signal is much choppier and noisier than the function generator signal. While the function generator can output a smooth sine wave, the myDAQ can only output voltages in increments of 0.3 mV. For a low amplitude signal such as this, the 0.3 mV resolution of the myDAQ can be a significant limitation. Consequently, the myDAQ is used to control the function generator (and automate the acquisition), rather than to replace it. 69
  • Slide 71
  • Digital Measurements Lab Agenda 1.Experiment 6 Digital introduction 2.Measuring function generator input and beam response using myDAQ 3.Calibrating function generators 4.Find natural frequency of beam using myDAQ 5.myDAQ Resolution Example 6.Modify Homework codes 70
  • Slide 72
  • Modify Homework/Lab 3 Code Modify todays code to calculate the spring stiffness as well, using beam theory. Divide the forcing amplitude (in Newtons) by the response amplitude (in meters). Note that this is only valid at low frequencies. See slide 73 for block diagram screenshot. 71
  • Slide 73
  • Note: for those using the station with the -24VDC supply to the proximeter: 1)There is a voltage divider on the proximeter drive that divides the output by 2 before you get to measure it. It is therefore required to multiply the response signal by 2 to recover the true displacement voltage. 2)The proximeter calibration is 200mV/mils (as opposed to 106mV/mils for the older proximeters). 72
  • Slide 74
  • Final Modifications 73
  • Slide 75
  • Wrapping Up You can run your code for various frequencies and find if the values of k you obtain are consistent with Experiment 6a. We now have most of the building blocks for digitally controlling and measuring the beam response. Next lab, we will write code to fully automate the process and find the resonant frequency of the beam. To do this, well have to learn about for loops and LabView data storage through arrays. 74