Apparatus The following items will be required for this procedure. Ensure each item is available before proceeding to the next step of this procedure: 1 Micro-Separometer – the unit is completely portable and self-contained, capable of operating on an internal rechargeable battery pack or being connected to an A-C power resource. Connection to an A-C power source will provide power to the unit and effect battery recharge. See Figure 1. 2 Syringe Plug – a plastic plug used to stopper the syringe during the CLEAN and EMULSION cycles. See Figure 2. 3 Syringe with Plunger – a disposable plastic syringe. 4 Vials – 25 mm outside diameter, Premark for proper alignment in the turbidimeter well. 5 Alumicel Coalescer – an expendable, precalibrated cell with a tapered end to fit the syringe. 6 Pipette with Plastic Tip – an automatic hand held pipette with a disposable plastic tip. A Pipet is supplied with each Micro-Separometer. 7 Wire Aid – a piece of wire with a loop on one end, used during the test to release the air trapped in the barrel of the syringe when the plunger is being inserted. A wire aid is supplied with each Micro-Separometer. 8 Water Container – a clean container for distilled water. 9 Beaker, Catch Pan, or Plastic Container – used to receive the waste fuel during the coalescence period of the test. One is supplied with the Micro-Separometer. 10 Micro-Separometer Six Pack – designed to fit inside the top lid of the Micro- Separometer. It contains 6 each of a new syringe, Pipet tip, test vials, syringe plug and Alumicel coalescer cell. See Figure 3. PPE Use Standard (SSP-0011) Y Full Face Shield / Goggles Hood Clothing—Chemical Protection Gloves—Chemical Personal Fall Protection Respirator / Filter Mask Job Hazards Yes Job Hazards Yes Access and Egress Moving Parts Activities of Others Noise Adverse Weather Rail Movement Chemical Exposure Sharp Edges Electricity Slips, Trips, and Falls Fatigue Static Electricity Fire and Explosion Trip Injuries Ground Contamination Trapped Pressure Manual Handling Working at Height Other: References / Documents MSDS on Kerosene Unit Safe Upper and Lower Limits Are a Unit Operating Procedure # Certified by Revisio n # Date Page 3 Lab OPS-LB-NM-011 B. Gereighty 2 9/10/07 1 of 5 Standard Test Method for Determining Water Characteristic of Aviation Turbine Fuels by Portable Separometer All samples must be properly labeled. All unused samples must be dumped into the Slop Tank on west side of Octane Quality Assurance Proper sampling and analytical analysis are critical for maintaining proper product quality. Special Instructions This procedure is used to determine the Water Characteristic of Aviation Turbine Fuels by Portable Separometer. Any problems, notify Shift Supervisor or Duane Chaisson. W A R N I N G C A U T I O N q Procedure Completed Date: Prepared By: Approved By:

Apparatus The following items will be required for this procedure. Ensure each item is available before proceeding to the next step of this procedure:

Embed Size (px)

Citation preview

Page 1: Apparatus The following items will be required for this procedure. Ensure each item is available before proceeding to the next step of this procedure:

Apparatus

The following items will be required for this procedure. Ensure each item is available before proceeding to the next step of this procedure:

1Micro-Separometer – the unit is completely portable and self-contained, capable of operating on an internal rechargeable battery pack or being connected to an A-C power resource. Connection to an A-C power source will provide power to the unit and effect battery recharge. See Figure 1.

2 Syringe Plug – a plastic plug used to stopper the syringe during the CLEAN and EMULSION cycles. See Figure 2.

3 Syringe with Plunger – a disposable plastic syringe.

4 Vials – 25 mm outside diameter, Premark for proper alignment in the turbidimeter well.

5 Alumicel Coalescer – an expendable, precalibrated cell with a tapered end to fit the syringe.

6 Pipette with Plastic Tip – an automatic hand held pipette with a disposable plastic tip. A Pipet is supplied with each Micro-Separometer.

7 Wire Aid – a piece of wire with a loop on one end, used during the test to release the air trapped in the barrel of the syringe when the plunger is being inserted. A wire aid is supplied with each Micro-Separometer.

8 Water Container – a clean container for distilled water.

9 Beaker, Catch Pan, or Plastic Container – used to receive the waste fuel during the coalescence period of the test. One is supplied with the Micro-Separometer.

10 Micro-Separometer Six Pack – designed to fit inside the top lid of the Micro-Separometer. It contains 6 each of a new syringe, Pipet tip, test vials, syringe plug and Alumicel coalescer cell. See Figure 3.

PPE Use

Standard (SSP-0011) Y

Full Face Shield / Goggles

Hood

Clothing—Chemical Protection

Gloves—Chemical

Personal Fall Protection

Respirator / Filter Mask

Supplied Air

Job Hazards Yes Job Hazards Yes

Access and Egress Moving Parts

Activities of Others Noise

Adverse Weather Rail Movement

Chemical Exposure Sharp Edges

Electricity Slips, Trips, and Falls

Fatigue Static Electricity

Fire and Explosion Trip Injuries

Ground Contamination Trapped Pressure

Manual Handling Working at Height

Other:

References / Documents

MSDS on Kerosene Unit Safe Upper and Lower Limits Spreadsheet

Area Unit Operating Procedure # Certified by Revision # Date Page

3 Lab OPS-LB-NM-011 B. Gereighty 2 9/10/07 1 of 5

Standard Test Method for Determining Water Characteristic of Aviation Turbine Fuels by Portable Separometer

Environmental

All samples must be properly labeled. All unused samples must be dumped into the Slop Tank on west side of Octane room.

Quality Assurance

Proper sampling and analytical analysis are critical for maintaining proper product quality.

Special Instructions

This procedure is used to determine the Water Characteristic of Aviation Turbine Fuels by Portable Separometer.

Any problems, notify Shift Supervisor or Duane Chaisson.

W A R N I N G

C A U T I O N

q Procedure Completed

Date:

Prepared By:

Approved By:

Page 2: Apparatus The following items will be required for this procedure. Ensure each item is available before proceeding to the next step of this procedure:

Area Unit Operating Procedure # Certified by Revision # Date Page

3 Lab OPS-LB-NM-011 B. Gereighty 2 9/10/07 2 of 5

Reagents

The following reagents will be required for this procedure. Ensure all reagents are available in the proper quantities before proceeding to the next step.

1 Aerosol OT, solid (100% dry) bis-2-ethylhexyl sodium sulfosuccinate.

2 Toluene, ASS reagent grade.

3 Dispersing Agent-Toluene Solution containing 1 mg of Aerosol OT per ml of toluene.

4 Reference Fluid Base - a surfactant-free, clean hydrocarbon material which is used to verify proper operation.

5

Reference Fluids – for checking the operational performance of the Micro-Separomater instrument. It consists of increasing concentrations (0 to 1.2 ml/L) of dispersing agent added to the reference fluid. The MSEP-A rating for this range of concentrations appears in Table 4. If the results do not fall within the range of limits in Table 4, the reference fluid shall be discarded and a fresh quantity of reference fluid prepared and the check repeated. Repeated out of tolerance test results are cause for returning the instrument to the factory for adjustment and calibration.

6 Water-clean, distilled, and surfactant free.

Standard Test Method for Determining Water Characteristic of Aviation Turbine Fuels by Portable Separometer (continued)

Preparation of Appartus

The following steps must be completed prior to performing the actual procedure:

1 Place the Micro-Separometer on a clean workbench in an area where the temperature is between 65 F and 85 F, and does not vary more than +/- 3 F.

2 Open the case and remove the six pack from the lid.

3 Raise the right panel until it is completely vertical and locked in place.

4 If A-C power is available, connect the power cord to the instrument.

5 If the internal battery power is used, assure that the batteries are charged sufficiently to perform the desired number of tests.

6 Turn the instrument on by depressing the ON switch. The ON power indicator light will alternately pulse on and off when the instrument is connected to an A-C power source and will stay on continuously when operated by the battery pack.

7 Let the instrument warm up for 15 minutes.

8 Have your test supplies ready: syringe, syringe plug, cell, pipette, pipette tip, wire aid, water and waste container.

q Procedure Completed

Date:

Prepared By:

Approved By:

Page 3: Apparatus The following items will be required for this procedure. Ensure each item is available before proceeding to the next step of this procedure:

Area Unit Operating Procedure # Certified by Revision # Date Page

3 Lab OPS-LB-NM-011 B. Gereighty 2 9/10/07 3 of 5

Sample Preparation

The test sample must be prepared as described below before performing the test:

1 Extreme care and cleanliness are required in taking samples. The test fuel should be caught in a clean container.

2 Do not filter the sample. If the test fuel is contaminated with particulate matter, allow such material to settle out of the fuel before starting the test.

3 The sample should be within the test temperature limits, 65 F to 85 F, +/- 3 F.

Standard Test Method for Determining Water Characteristic of Aviation Turbine Fuels by Portable Separometer (continued)

Test Procedure

1 Depress OPERATING MODE A switch.

2 Remove the plunger from the syringe barrel.

3 Insert a plug into the bottom of the syringe.

4 Add 50 ml of the test fuel to the syringe.

5 Place the syringe on the emulsifier.

6 Twist the syringe to lock in place.

7 Add 15 – 20 ml of the test fuel to a clean vial (1/2 full).

8 Insert the vial in the turbidimeter well.

9 Align the black mark on the vial with the white mark on the front panel.

10 Depress the START switch to initiate the CLEANING cycle. This will also initiate the Automatic Program. See Table 3.

11 When the mixer stops, remove the syringe from the emulsifier.

12 Discard the fuel and drain the syringe thoroughly.

13 Add 50 ml of fresh test fuel to the syringe.

14 Attach a fresh plastic tip to the pipette.

15 Push in the plunger on the pipette.

16 Immerse the tip of the pipette just below the water surface and withdraw 50 microliters of water.

17 Withdraw the pipette from the water slowly in order to avoid water drips adhering to the outside of the tip.

q Procedure Completed

Date:

Prepared By:

Approved By:

Page 4: Apparatus The following items will be required for this procedure. Ensure each item is available before proceeding to the next step of this procedure:

Area Unit Operating Procedure # Certified by Revision # Date Page

3 Lab OPS-LB-NM-011 B. Gereighty 2 9/10/07 4 of 5

Test Procedure (continued)

18 Immerse the tip of the pipette just below the test fuel surface in the center of the syringe.

19 Inject the water into the test fuel.

20 Slowly withdraw the pipette out of the syringe.

21 Place the syringe barrel on the emulsifier.

22 Twist the syringe to lock in place.

23 Depress the START switch to initiate emulsification.

24Adjust the meter to read 100 by using the UP (↑) or DOWN (↓) Arrow switch. The meter adjustment period lasts 10 seconds. If the adjustment cannot be completed at this time, a final adjustment can be made during the second meter adjustment period. When the mixer stops, you have 30 seconds to prepare for coalescence.

25 When the mixer stops, remove the syringe from the emulsifier.

26 Partially insert the plunger in the syringe. With or without the help of the wire aid.

27 Turn the syringe upside down.

28 Remove the plug from the bottom of the syringe.

29 Push the plunger up to the 50 ml mark.

30 Attach an Alumicel coalescer cell to the bottom of the syringe.

31 Place the syringe in the syringe drive.

32 Place a drip pan under the syringe to collect the unwanted portion of the fuel.

33 You will hear 4 short tones. If necessary, adjust the meter to read 100 by using the UP (↑) or DOWN (↓) ARROW switch.

34 Remove the vial from the turbidimeter well.

35 Discard the fuel and drain the vial.

36 A pulsed tone will sound indicating when to collect the last 15 ml of test fuel. Tilt the vial at a slight angle and allow the fuel to flow down the inner surface of the vial.

37 Place the vial in the turbidimeter well.

38 Align the black mark on the vial with the white mark on the front panel.

Standard Test Method for Determining Water Characteristic of Aviation Turbine Fuels by Portable Separometer (continued)

q Procedure Completed

Date:

Prepared By:

Approved By:

Page 5: Apparatus The following items will be required for this procedure. Ensure each item is available before proceeding to the next step of this procedure:

Area Unit Operating Procedure # Certified by Revision # Date Page

3 Lab OPS-LB-NM-011 B. Gereighty 2 9/10/07 5 of 5

Test Procedure (continued)

39 The setting time is 56 seconds. You will hear a steady tone for 4 seconds to let you know that the meter is about to activate.

40 The meter will activate for 10 seconds. Take your meter reading.

Standard Test Method for Determining Water Characteristic of Aviation Turbine Fuels by Portable Separometer (continued)

q Procedure Completed

Date:

Prepared By:

Approved By: