LED measurement radiant power measurement
application note
2
Contents 1 Introduction ...................................................................................................................................... 3
2 Setup ............................................................................................................................................... 4
2.1 Introduction .............................................................................................................................. 4
2.2 Measurement setup ................................................................................................................ 4
2.3 Integrating spheres ................................................................................................................. 5
3 LED measurement procedure ......................................................................................................... 6
4 CIE 127 & Admesy calibration procedure ....................................................................................... 7
4.1 Overview ................................................................................................................................. 7
4.2 Spectral and spatial distribution .............................................................................................. 7
5 Remarks for measurement procedure ............................................................................................ 8
6 References ...................................................................................................................................... 9
3
1 Introduction This document is intended to clarify the procedure for accurate measurements of LED and SSL (Solid
State Lighting) products by means of an integrating sphere. In perspective of CIE 127: 2007
recommendations, the entire procedure including calibration, absorption correction and led
measurements is listed. Special attention is given to the enhancements of the calibration procedure
with an Admesy tungsten broadband reference standard compared to the conventional calibration
procedure. At last a number of general remarks concerning LED measurements are provided to
ensure accurate and repeatable LED measurements.
These measurements can be carried out using Admesy’s Iliad software application. Latest Iliad Version can be downloaded from the Admesy website for free. www.admesy.com/download/software/iliad.zip
4
2 Setup
2.1 Introduction A typical LED measurement procedure with integrating sphere consists of three steps: calibrating the
test setup, absorption correction for device under test (DUT) and actual measurements. Special
attention is given to enhancements of the sphere setup calibration procedure with Admesy products
compared to conventional calibration procedures as defined by CIE 127. Details are discussed in
chapter CIE 127 & Admesy calibration procedure.
2.2 Measurement setup In order to conduct fast and accurate LED and SSL measurements, an integrating sphere setup as
shown in figure 1 is recommended (CIE 127: 2007)]. An integrating sphere (1) connected to a
spectrometer (2) by a fiber-optic flex cable (3) to measure a LED (4). For absorption correction an
additional correction connecting an auxiliary lamp (5) may be necessary. Various reference standard
lamps (not shown in the scheme below) are needed for spectral and luminous flux calibration as
defined by CIE 127. This Admesy test setup is developed for luminous flux (lm) measurements and
provides information regarding X, Y, Z parameters as well as x,y, u’,v’, CCT, dominant and peak
wavelength, CRI and spectral distribution.
Fig 1 Integrating sphere test setup for LED measurements.
1
2
3
4
5
5
2.3 Integrating spheres An integrating sphere is used to spatially distribute the unidirectional radiation emitted by the LED into
homogeneous radiation. Depending on properties of the LED, an appropriate sphere size has to be
chosen. Large spheres typically have better light distribution compared to small spheres due to the
relative small geometrical errors objects in the sphere cause. While small spheres have the
advantage of a relatively high throughput, resulting in an increased signal-noise ratio. Table 1
provides an overview of Admesy spheres and typical applications.
Model Diameter Auxiliary lamp Application Type of DUT
1
75mm
No
· Production
· Standard single LEDs
2
150mm
Yes
· Production · Laboratory
· Standard single LEDs · High power LEDs
3
250mm
Yes
· Laboratory, CIE 127 compliant for 2𝜋 measurements
· Standard single LEDs · High power LEDs · Small LED modules
Table 1 Admesy integrating spheres and applications.
6
3 LED measurement procedure The scheme below shows three steps of a typical LED measurement procedure, starting with sphere
calibration procedure according to CIE 127 (left column) in two steps. The Admesy procedure with
only one step is shown in the right column. The proposed adjustments in calibration procedures are
described in detail in chapter 4. Followed by DUT and fixture absorption by means of an auxiliary
lamp according to CIE 127 procedure. At last the actual LED measurement procedure.
Fig 2 Overview of complete LED measurement procedure including calibration procedure.
Fig 2 Overview of complete LED measurement procedure including calibration procedure.
Procedure overview
CIE 127: 2007 procedure Admesy procedure
Admesy calibration
Spectral wavelength calibration with tungsten broadband reference standard.
Absorption correction: step 1
Applicable to sphere with auxiliary port only (150 and 250 mm).
Measure spectral radiant distribution without objects attached to sphere.
Absorption correction: step 2
Applicable to sphere with auxiliary port only (150 and 250 mm).
Measure spectral radiant distribution with test LED
(turned off) attached to sphere.
Start LED measurement
Measurement of LEDs
CIE 127 calibration: step 2
Absolute luminous flux calibration with LED reference standard (compensate for spectral and spatial distribution of tungsten).
Spectral wavelength calibration with Admesy tungsten broadband reference standard. Compensated for enhanced blue region spectral response (blue filter) and spatial distribution (cosine corrector).
CIE 127 calibration: step 1
7
4 CIE 127 & Admesy calibration procedure
4.1 Overview CIE 127 LED sphere setup calibration procedure describes two steps:
Spectral wavelength calibration for spectral response of sphere.
Absolute luminous flux calibration to compensate for poor blue spectral response of tungsten reference standard and typical spatial distribution of LED.
Admesy enhanced calibration procedure in one step:
Spectral calibration with blue enhanced broadband reference standard (poor blue region compensation) and cosine corrector (spatial LED distribution).
4.2 Spectral and spatial distribution CIE 127 requires the second step with a LED reference standard to calibrate the setup for LEDs
based on their typical spectral and spatial distribution. Due to the wide range of variations in LEDs,
CIE 127 states that a large amount of LED reference standards should be maintained. For each type
of LED to be measured, a corresponding LED reference standard is required for calibration.
Concerning the spectral deviations, the spectral distribution of the tungsten reference standard (fig 3)
shows relatively low spectral power in the blue region compared to the longer wavelengths. In the
shorter wavelength region, the signal noise ratio therefore is non optimal. Measuring white and blue
LEDs which have typically high luminous flux in this region (fig 4) with a poor signal noise ratio may
result in inaccurate measurement results. Second reason for calibrating with a LED reference
standard is because of typical spatial distribution of LEDs.
Fig 3 Spectral distribution of broadband tungsten reference standard used for CIE 127 calibration step 1.
Fig 4 Spectral distribution of a (white) LED reference standard used for CIE 127 calibration step 2.
Fig 5 Spectral distribution of Admesy enhanced broadband tungsten reference standard.
To overcome all inconveniences in both calibration procedure as well as time and cost efficient LED
measurements, Admesy has developed an enhanced halogen reference standard with blue filter and
cosine corrector. This reference standard allows integrating both calibration steps of the CIE 127
recommendation. Because of the integrated blue light balancing filter, a more balanced spectral
distribution is created (fig 5) causing a better signal noise ratio in the short wavelength region. Thus,
allowing accurate measurements of white and blue LEDs. In addition, a cosine corrector ensures
optimal Lambertian distribution to match the spatial distribution of a wide range of LEDs. Instead of
maintaining multiple reference standards, one Admesy reference standard is sufficient for calibrating
integrating sphere setups.
8
5 Remarks for measurement procedure High speed measurements (i.e. production settings) are technically speaking not as sophisticated
compared to laboratory measurements. To ensure accuracy and repeatability of production
measurements as recommended by CIE 127: 2007, a number of remarks are listed in this chapter
with a brief explanation.
Control sphere for dust and dirt.
Control calibration (date) of spectrometer.
Control calibration (date) of reference standard: due to maximum lifetime.
Control stabilization time of reference standard before calibration.
Control power supplies (current, voltage, AC/ DC, waveform).
Control measurement parameters: i.e. measurement integration time, resolution.
Control ambient temperature and humidity of test environment.
CIE 127: 2007 states an ambient temperature of 25ºC (77F) unless otherwise specified.
Control ambient lighting (no ambient light should enter the sphere).
Control LED position during measurement (see examples below)
Fig 6 Undesired situation (left) light beam is affected by obstructions. Desired situation (right) light beam is not affected by any obstructions.
Fig 7 Generally undesired situation (left) where light from LED leaks outside the sphere. Desired situation (right) where all light emitted from LED enters the sphere.
Fig 8 Components of LED and/or fixture may absorb light
9
6 References CIE, Publication No. 127. Measurements of LEDs, Central Bureau of the CIE, Vienna, Austria 2007 http://www.cie.co.at/index.php?i_ca_id=402
10
Admesy B.V. Sleestraat 3 6014 CA Ittervoort The Netherlands T +31 (0)475 600 232 F +31 (0)475 600 316 www.admesy.com [email protected]
The material in this document is subject to change. No rights can be derived from the content of this document. All rights reserved. No part of this document may be reproduced, stored in a database or retrieval system, or published in any form or way, electronically, mechanically, by print, photo print, microfilm or any other means without prior written permission from the publisher.