Are the light sources of light cabinets temporarily stable? Almuñécar, September 24th 2008 R. Roa, L. Gómez-Robledo, R. Huertas, M. Melgosa Workshop on “Colorimetry and Color Imaging” Departamento de Óptica, Facultad de Ciencias, Universidad de Granada. 18071 Granada (SPAIN). email: [email protected]

Are the light sources of light cabinets temporarily stable?

  • Upload
    levia

  • View
    16

  • Download
    2

Embed Size (px)

DESCRIPTION

Workshop on “Colorimetry and Color Imaging”. Almuñécar , September 24th 2008. Are the light sources of light cabinets temporarily stable?. R. Roa, L. Gómez-Robledo, R. Huertas, M. Melgosa. Departamento de Óptica , Facultad de Ciencias , Universidad de Granada. - PowerPoint PPT Presentation

Citation preview

Page 1: Are the light sources of light cabinets temporarily stable?

Are the light sources oflight cabinets temporarily stable?

Almuñécar, September 24th 2008

R. Roa, L. Gómez-Robledo, R. Huertas, M. Melgosa

Workshop on “Colorimetry and Color Imaging”

Departamento de Óptica, Facultad de Ciencias, Universidad de Granada.18071 Granada (SPAIN). email: [email protected]

Page 2: Are the light sources of light cabinets temporarily stable?

Introduction

Method

Results and discussion

Conclusions

References

2

Contents Are the light sources of light cabinets…?

Page 3: Are the light sources of light cabinets temporarily stable?

Light source

Object

ObserverIn a first approach, color specification depends on :

Light source Object Observer

X E R x

Y E R y

Z E R z

To standardize color coordinates:

CIEInternational Commission

on Illumination

Standards for the observer: Standard Observer 1931 and Standard Observer 1964

Standards for the light source: illuminants (relative spectral distribution)

3

CIE 15:2004, 3rd Edition, Colorimetry (Technical report), CIE Central Bureau, Vienna (2004).

Introduction Are the light sources of light cabinets…?

Page 4: Are the light sources of light cabinets temporarily stable?

CIE proposes the following illuminants:

A (tungsten), C (average daylight), D50, D55, D65 y D75 (daylight simulators)

Typical fluorescents lamps (FL1 to FL12)

New fluorescents lamps (FL3.1 a FL3.15) and high pressure discharge lamps (HP1 to HP5)

On visual experiments:

Cabinets with source simulators which represents properly the considered illuminant.

Daylight Tungsten Fluorescent

4

CIE 15:2004, 3rd Edition, Colorimetry (Technical report), CIE Central Bureau, Vienna (2004).

Introduction Are the light sources of light cabinets…?

Page 5: Are the light sources of light cabinets temporarily stable?

From previous works:

There is an important color variability in commercial light cabinets

There are relevant differences between light sources and illuminants

R. Huertas, M. J. Rivas, M. Melgosa, M. Sánchez-Marañón, S. Bhosle, J. J. Damelincourt. “Uniformirty of lighting in color assessment cabinets”, 19th Congress of the International Commission for Optics (ICO XIX), SPIE Proceedings, Volume 4829, 879-880, Firenze, Italy (2002).

R. Roa, R. Huertas, M. A. López-Álvarez, L. Gómez-Robledo, M. Melgosa. “Comparación entre iluminantes y fuentes simuladoras”, Opt. Pura Apl. 41 (3) 291-300 (2008).

5

Objective:Study the temporal evolution of some light sources of commercial light cabinets

Introduction Are the light sources of light cabinets…?

It is important to know the state of the light we are using (temporarily dependent)R. Roa, L. Gómez-Robledo, R. Huertas, M. Melgosa. “Temporal evolution of light sources available in commercial

color cabinets”, 4th Balkan Conference on Lighting - Balkan Light 2008, Ljubljana, Slovenia (2008) (in press)

Page 6: Are the light sources of light cabinets temporarily stable?

EXPERIMENTAL SETUP

2 light cabinetsGretagMacbeth Spectralight III

VeriVide CAC 60

Photo Research SpectraScan PR-704 spectroradiometer

PTFE reference white

3 sources studied in each light cabinet

D65 (illuminant D65, daylight, 6500K)

A or F (illuminant A, tungsten, 2856K)

TL84 (illuminant F11, fluorescent, 4000K)

We measured the spectral radiance of the reference white at the following hours of use of the light sources:

0, 0.5, 1, 2, 3, 4, 5, 10, 20, 30, 40, 50, 75 and 100 hours

Total: 14 hours x 3 light sources x 2 light booths = 84 measurements

6

Method Are the light sources of light cabinets…?

Page 7: Are the light sources of light cabinets temporarily stable?

WAYS OF COMPARISON

Normalized spectral distributions of the light sources with their respective illuminants.

Metrics for comparison of spectral values:

RMSE (Root Mean Square Error) 2

1

( ) '( )n

i ii

RMSE S S

WRMSE (Weighted Root Mean Square Error) 2

1

( ) ( ) '( )n

i ii

WRMSE R i S S

SCI (Spectral Comparison Index)1

( ) ( )n

i ii

SCI w

CSCM (Colorimetric and Spectral Combined Metric)

*1 1000 1 (%)abCSCM Ln GFC E IRE

7

F. H. Imai, M. R. Rosen, R. S. Berns, “Comparative study of metrics for spectral match quality”, Proceedings of the First European Conference on Colour in Graphics, Imaging and Vision, 492-496 (2002).

J. A. S. Viggiano, “A perception-referenced method for comparison of radiance ratio spectra and its application as an index of metamerism”, Proceedings of AIC Color 01, 701-704 (2004).

M. A. López-Álvarez, J. Hernández-Andrés, J. Romero, R. L. Lee Jr., “Designing a practical system for spectral imaging of skylight”, Appl. Opt. 44, 5688-5695 (2005).

Method Are the light sources of light cabinets…?

Page 8: Are the light sources of light cabinets temporarily stable?

Correlated color temperature (CCT)

CIE Color Rendering Index (CRI)

8

CIEDE2000 color differences

CIEDE2000 color differences between GretagMacbeth ColorChecker chips illuminated by the source simulator or its respective illuminant

1/222 2' ' ' ' '

00 TL L C C H H C C H H

L C H C HE R

k S k S k S k S k S

J. Hernández-Andrés, R. L. Lee Jr., J. Romero, “Calculating correlated color temperatures across the entire gamut of daylight and skylight chromaticities”, Appl. Opt. 38, 5703-5709 (1999).

CIE Publication 13.3-1995, Method of Measuring and Specifying Colour Rendering Properties of Light Sources (Technical report), CIE Central Bureau, Vienna (1995).

CIE Publication 142-2001, Improvement to Industrial Colour-Difference Evaluation (Technical report), CIE Central Bureau, Vienna (2001).

Method Are the light sources of light cabinets…?

Page 9: Are the light sources of light cabinets temporarily stable?

SPECTRAL DISTRIBUTIONS

There are some differences between light sources and their respective illuminants.

Sources are stabilized after half an hour of use, except sources simulating the A illuminant (approximately stable from the first measurement).

9

R. Roa, R. Huertas, M. A. López-Álvarez, L. Gómez-Robledo, M. Melgosa. “Comparación entre iluminantes y fuentes simuladoras”, Opt. Pura Apl. 41 (3) 291-300 (2008).

Spectralight III D65 Spectralight III A

Spectralight III TL84

CAC 60 D65CAC 60 F

CAC 60 TL84

Results and discussion Are the light sources of light cabinets…?

Page 10: Are the light sources of light cabinets temporarily stable?

METRICS

RMSE WRMSE

Cabinet Source Average Std. Dev. (CV) Max Min Average Std. Dev. (CV) Max Min

Spectralight III

D65 0.1315 0.0029 (2.2%) 0.1364 (10h) 0.1210 (100h) 0.0121 0.0005 (4.1%) 0.0135 (0h) 0.0110 (100h)

A 0.0419 0.0012 (2.9%) 0.0440 (75h) 0.0397 (2h) 0.0040 0.0001 (2.8%) 0.0042 (75h) 0.0037 (2h)

TL84 0.1712 0.0103 (6.0%) 0.2042 (0h) 0.1642 (0.5h) 0.0306 0.0005 (1.8%) 0.0314 (100h) 0.0296 (0h)

CAC 60

D65 0.3796 0.0087 (2.3%) 0.4078 (0h) 0.3738 (1h) 0.0450 0.0030 (6.6%) 0.0549 (0h) 0.0434 (1h)

F 0.0841 0.0037 (4.4%) 0.0909 (100h) 0.0768 (0h) 0.0113 0.0005 (4.8%) 0.0122 (100h) 0.0102 (0h)

TL84 0.1843 0.0071 (3.8%) 0.1905 (100h) 0.1615 (0h) 0.0325 0.0016 (4.8%) 0.0338 (100h) 0.0275 (0h)

Values in green are those with the lowest coefficient of variation. The red ones have the largest coefficient of variation.

10

Results and discussion Are the light sources of light cabinets…?

Page 11: Are the light sources of light cabinets temporarily stable?

Values slightly rise over the time except for the source D65 from Spectralight III cabinet.

The results obtained with the 4 metrics are not the same.

11

SCI CSCM

Cabinet Source Average Std. Dev. (CV) Max Min Average Std. Dev. (CV) Max Min

Spectralight III

D65 0.0473 0.0016 (3.3%) 0.0519 (0h) 0.0439 (100h) 8.81 0.51 (5.8%) 9.78 (0h) 7.36 (100h)

A 0.0095 0.0003 (3.1%) 0.0101 (75h) 0.0090 (2h) 9.34 0.40 (4.3%) 9.88 (2h) 8.51 (0h)

TL84 0.1297 0.0039 (3.0%) 0.1349 (100h) 0.1220 (1h) 24.84 1.90 (7.7%) 26.16 (5h) 18.84 (0h)

CAC 60

D65 0.1795 0.0114 (6.4%) 0.2169 (0h) 0.1728 (1h) 21.16 0.82 (3.9%) 22.42 (100h) 19.40 (0h)

F 0.0262 0.0012 (4.5%) 0.0283 (100h) 0.0239 (0h) 30.82 1.01 (3.3%) 32.65 (100h) 28.88 (0h)

TL84 0.1373 0.0051 (3.7%) 0.1434 (100h) 0.1210 (0h) 24.17 2.79 (11.6%) 26.02 (100h) 15.18 (0h)

Results and discussion Are the light sources of light cabinets…?

Page 12: Are the light sources of light cabinets temporarily stable?

CORRELATED COLOR TEMPERATURE

CCT (K)

Cabinet Source Average Std. Dev. (CV) Max Min

Spectralight III

D65 6129 (6500) 36 (0.6%) 6263 (100h) 6084 (2h)

A 2823 (2856) 6 (0.2%) 2833 (0h) 2814 (2h)

TL84 3698 (4000) 30 (0.8%) 3761 (0.5h) 3663 (100h)

CAC 60

D65 6117 (6500) 133 (2.2%) 6539 (0h) 5992 (100h)

F 2549 (2856) 13 (0.5%) 2574 (0h) 2527 (100h)

TL84 3747 (4000) 47 (1.2%) 3883 (0h) 3705 (100h) 12

Source A from Spectralight III

Lowest variation over time

Source D65 from Spectralight III

Shows an increase over time

Source D65 from CAC 60

Largest decrease over time

Results and discussion Are the light sources of light cabinets…?

Page 13: Are the light sources of light cabinets temporarily stable?

COLOR RENDERING INDEX

CRI

Cabinet Source Average Std. Dev. (CV) Max Min

Spectralight III

D65 95 0.44 (0.5%) 95 (20h) 94 (0h)

A 98 0.05 (0.1%) 98 (2h) 97 (0h)

TL84 97 0.27 (0.3%) 98 (0h) 96 (1h)

CAC 60

D65 95 0.44 (0.5%) 96 (0h) 94 (100h)

F 91 0.44 (0.5%) 92 (0h) 91 (100h)

TL84 97 0.09 (0.1%) 97 (0h) 97 (75h)

Sources A from Spectralight III and TL84 from CAC 60

Variation on CRI

Source F from CAC 60

13

Lowest variation over time

Suffers (insignificant) oscillations while evolves

Almost insignificant in all the light sources

Results and discussion Are the light sources of light cabinets…?

Page 14: Are the light sources of light cabinets temporarily stable?

CIEDE2000 COLOR DIFFERENCES

14

Results and discussion Are the light sources of light cabinets…?

Page 15: Are the light sources of light cabinets temporarily stable?

Chips and standard deviations in CIEDE2000 units

Cabinet Source Average Max Min

Spectralight III

D65 0.055 0.158 (13) 0.003 (24)

A 0.012 0.030 (15) 0.000 (23)

TL84 0.056 0.192 (13) 0.001 (24)

CAC 60

D65 0.091 0.409 (13) 0.002 (20)

F 0.039 0.118 (15) 0.002 (24)

TL84 0.120 0.437 (13) 0.000 (24)

Bluish chips 8 and 13

Source A from Spectralight III

Achromatic samples (chips 19 to 24)

15

largest standard deviations in color differences

smallest variations when the sources age

Source TL84 from CAC 60

smallest color-difference variations over time

greatest standards deviations

1 2 3 4 5 6

7 8 9 10 11 12

13 14 15 16 17 18

19 20 21 22 23 24

Results and discussion Are the light sources of light cabinets…?

Page 16: Are the light sources of light cabinets temporarily stable?

Source A from Spectralight III cabinet is the source which suffers the lowest variations over time, attending to the most of indices.

When high precision color measurements (e.g. some threshold visual experiments) are needed, it should be necessary to know the state of the light source we are using, because it is temporally dependent, and it cannot be assumed that it is identical to a CIE illuminant.

To complete this analysis, it would be interesting to study the evolution of the light sources at more hours of use, increase the number of light sources and use other indices for the comparison.

Future work:

Source D65 from Spectralight III cabinet gets closer to its illuminant with age. On the other side, source D65 from CAC 60 cabinet gets older faster than others.

16

Conclusions Are the light sources of light cabinets…?

This work is related to current TC1-44 “Practical Sources for Daylight Colorimetry”.

Final conclusion:

Page 17: Are the light sources of light cabinets temporarily stable?

Thanks for your attention

Acknowledgements:Research Project FIS2007-64266, Ministerio de Educación y Ciencia (Spain), with FEDER support.