585 Charest Blvd. East, Suite 300 Québec, Québec G1K 9H4, Canada Phone (418) 877–2944 Fax (418) 877-2834 or (418) 266–1422 www.abb.com/analytical PROPRIETARY The information contained herein is proprietary to and should be considered a trade secret of ABB Bomem Inc. It shall not be reproduced in whole or in part without the written authorization of ABB Bomem Inc. FO-2100-10, rev 0523 Document Ingénierie ARIES ARIES Airborne Radiometer Evaluation System ARIES Electrical ICD Document Number: BOM-ARIES-0003 Issue Date: 4 July 2006 Issue: Draft Revision: Distribution: See page v Classification: N/A Function Name Signature Date Prepared by Electronics Technician Jean-Yves Trottier Checked by Electronics Lead Mario Lehoux Approved by Project Manager Robert Poulin

ARIES - ensembles-eu.metoffice.comensembles-eu.metoffice.com/met-res/aries/technical/... · ARIES modules interconnections are shown in the following diagram: Figure 2–1 ARIES Cable

  • Upload
    others

  • View
    8

  • Download
    0

Embed Size (px)

Citation preview

Page 1: ARIES - ensembles-eu.metoffice.comensembles-eu.metoffice.com/met-res/aries/technical/... · ARIES modules interconnections are shown in the following diagram: Figure 2–1 ARIES Cable

585 Charest Blvd. East, Suite 300 Québec, Québec G1K 9H4, Canada

Phone (418) 877–2944 Fax (418) 877-2834 or (418) 266–1422

www.abb.com/analytical

PROPRIETARY The information contained herein is proprietary to and should be considered a trade secret of ABB Bomem Inc.

It shall not be reproduced in whole or in part without the written authorization of ABB Bomem Inc.

FO-2100-10, rev 0523 Document Ingénierie

ARIES

ARIES Airborne Radiometer Evaluation System

ARIES Electrical ICD

Document Number: BOM-ARIES-0003 Issue Date: 4 July 2006

Issue: Draft Revision: –

Distribution: See page v Classification: N/A

Function Name Signature Date

Prepared by Electronics Technician Jean-Yves Trottier

Checked by Electronics Lead Mario Lehoux

Approved by Project Manager Robert Poulin

Page 2: ARIES - ensembles-eu.metoffice.comensembles-eu.metoffice.com/met-res/aries/technical/... · ARIES modules interconnections are shown in the following diagram: Figure 2–1 ARIES Cable

ABB ARIES

ARIES Electrical ICD

Document No: BOM-ARIES-0003

Issue: Draft Rev: – Page

Date: 4 July 2006 i

TABLE OF CONTENTS

1. INTRODUCTION ........................................................................................................................................1

1.1 Purpose of Document ............................................................................................................................1

2. ELECTRICAL INTERFACE........................................................................................................................2

2.1 Cable interconnections ..........................................................................................................................2 2.2 Placement of new connectors.................................................................................................................2 2.3 System cabling evolution.......................................................................................................................3 2.4 Power 1 cable wire function reassignment .............................................................................................4 2.5 Power 2 cable wire function reassignment .............................................................................................4 2.6 Heater & fan cable wire function reassignment ......................................................................................5 2.7 Ethernet cable description......................................................................................................................5 2.8 RS-422 cable description.......................................................................................................................6 2.9 RS-485 cable description.......................................................................................................................7

Page 3: ARIES - ensembles-eu.metoffice.comensembles-eu.metoffice.com/met-res/aries/technical/... · ARIES modules interconnections are shown in the following diagram: Figure 2–1 ARIES Cable

ABB ARIES

ARIES Electrical ICD

Document No: BOM-ARIES-0003

Issue: Draft Rev: – Page

Date: 4 July 2006 ii

LIST OF FIGURES

Figure 2–1 ARIES Cable Interconnections ...................................................................................................... 2 Figure 2–2 ARIES Optical Head Connectors Placement .................................................................................. 3 Figure 2–3 Amphenol RJF Series RJ45 to Field Converter .............................................................................. 6 Figure 2–4 Crossover Ethernet Cable Wiring Diagram .................................................................................... 6 Figure 2–5 RS-422 Cable Wiring Diagram ...................................................................................................... 7 Figure 2–6 RS-485 Cable Wiring Diagram ...................................................................................................... 8

Page 4: ARIES - ensembles-eu.metoffice.comensembles-eu.metoffice.com/met-res/aries/technical/... · ARIES modules interconnections are shown in the following diagram: Figure 2–1 ARIES Cable

ABB ARIES

ARIES Electrical ICD

Document No: BOM-ARIES-0003

Issue: Draft Rev: – Page

Date: 4 July 2006 iii

LIST OF TABLES

Table 2–1 Cable List with Description ............................................................................................................. 3 Table 2–2 Power 1 Cable Wire Function Reassignment.................................................................................... 4 Table 2–3 Power 2 Cable Wire Function Reassignment.................................................................................... 4 Table 2–4 Heater & Fan Cable Wire Function Reassignment ........................................................................... 5 Table 2–5 Ethernet Cable Parts ........................................................................................................................ 6 Table 2–6 RS-422 Cable Parts ......................................................................................................................... 7 Table 2–7 RS-485 Cable Parts ......................................................................................................................... 8

Page 5: ARIES - ensembles-eu.metoffice.comensembles-eu.metoffice.com/met-res/aries/technical/... · ARIES modules interconnections are shown in the following diagram: Figure 2–1 ARIES Cable

ABB ARIES

ARIES Electrical ICD

Document No: BOM-ARIES-0003

Issue: Draft Rev: – Page

Date: 4 July 2006 iv

DOCUMENT CHANGE RECORD

Issue Rev. Date Change Description

Draft – 4 July 2006 First draft.

Prepared by Jean-Yves Trottier. Checked by Mario Lehoux.

Approved by Robert Poulin.

Page 6: ARIES - ensembles-eu.metoffice.comensembles-eu.metoffice.com/met-res/aries/technical/... · ARIES modules interconnections are shown in the following diagram: Figure 2–1 ARIES Cable

ABB ARIES

ARIES Electrical ICD

Document No: BOM-ARIES-0003

Issue: Draft Rev: – Page

Date: 4 July 2006 v

DISTRIBUTION LIST

The following persons (or company) shall receive this document:

Project Team

Client

Other

Page 7: ARIES - ensembles-eu.metoffice.comensembles-eu.metoffice.com/met-res/aries/technical/... · ARIES modules interconnections are shown in the following diagram: Figure 2–1 ARIES Cable

ABB ARIES

ARIES Electrical ICD

Document No: BOM-ARIES-0003

Issue: Draft Rev: – Page

Date: 4 July 2006 1

1. INTRODUCTION

1.1 PURPOSE OF DOCUMENT

This document describes the electrical interface for the ARIES Optical Head and sub-modules.

Page 8: ARIES - ensembles-eu.metoffice.comensembles-eu.metoffice.com/met-res/aries/technical/... · ARIES modules interconnections are shown in the following diagram: Figure 2–1 ARIES Cable

ABB ARIES

ARIES Electrical ICD

Document No: BOM-ARIES-0003

Issue: Draft Rev: – Page

Date: 4 July 2006 2

2. ELECTRICAL INTERFACE

2.1 CABLE INTERCONNECTIONS

ARIES modules interconnections are shown in the following diagram:

Figure 2–1 ARIES Cable Interconnections

2.2 PLACEMENT OF NEW CONNECTORS

Ethernet connector J125 and RS-422 connector J124 will be added to the Optical Head assembly. Detailed

design about connector placement is still open but Figure 2–2 below illustrates the baseline for the proposed

location. A bracket will be added at the top right corner of the Optical Head. It will hold J125 on its upper side

and J124 on its lower side.

Page 9: ARIES - ensembles-eu.metoffice.comensembles-eu.metoffice.com/met-res/aries/technical/... · ARIES modules interconnections are shown in the following diagram: Figure 2–1 ARIES Cable

ABB ARIES

ARIES Electrical ICD

Document No: BOM-ARIES-0003

Issue: Draft Rev: – Page

Date: 4 July 2006 3

RS422

New Bracket

(Side Closed

for rigidity)

Ethernet RS485

Figure 2–2 ARIES Optical Head Connectors Placement

2.3 SYSTEM CABLING EVOLUTION

Effort was made to minimize cabling modifications. Most of the cables will keep their actual configuration and

function, some will see their wires reassigned to other functions without modification. This is mainly the case for

signal and power cables.

The situation is not the same for communication cables. Optical fibers are replaced with an Ethernet

communication link, which is a completely new cable design. RS-422 link will connect between Scan

Electronics and optical head, requiring, once again, a new cable design. RS-485 communication link will no

longer exist between the PC and the Optical Head module. As RS-485 communication is only used during

temperature controller configuration, a dedicated connector will be mounted on the optical head and a special

interface cable will allow for device reconfiguration. A cap will close and protect the connector when not in use.

Table 2–1 Cable List with Description

From - To Description Status

X17 – J163 Warm Black Body No modification.

X15 – J164 Cold Black Body No modification.

X14 – J162 Resolver No modification.

X16 – J165 Heaters No modification.

X18 – J161 Motor Drive No modification.

J105 – J102 Power 1 Same cable wiring but with wires function reassigned. See Power 1 cable wires function reassignment.

J106 – J103 Power 2 Same cable wiring but with wires function reassigned. See Power 2 cable wires function reassignment.

J107 – J104 Heater & Fan Same cable wiring but with wires function reassigned. See

Page 10: ARIES - ensembles-eu.metoffice.comensembles-eu.metoffice.com/met-res/aries/technical/... · ARIES modules interconnections are shown in the following diagram: Figure 2–1 ARIES Cable

ABB ARIES

ARIES Electrical ICD

Document No: BOM-ARIES-0003

Issue: Draft Rev: – Page

Date: 4 July 2006 4

From - To Description Status

Heater/Fan cable wires function reassignment.

J108 - PC RS-485 Link New design. See RS-485 cable description.

J124 – J159 RS-422 Link New design. See RS-422 cable description.

J125 - PC Ethernet Link New design. See Ethernet cable description.

2.4 POWER 1 CABLE WIRE FUNCTION REASSIGNMENT

The Power 1 cable linking J105 on the interferometer to J102 on the power tray requires no modification. The

cable assembly schedule ARAMRF048006 meets the requirements of the upgraded instrument.

Wire function will change according to the following table:

Table 2–2 Power 1 Cable Wire Function Reassignment

J105 J102 Actual Function Upgraded Function

A A +8VA1 Not used.

B B +8VA1RTN Not used.

C C +18VA1 Not used.

D D 18VA1RTN Not used.

E E -18VA1 Not used.

F F +8VA2 Not used.

G G 8VA2RTN Not used.

H H +18VA2 Not used.

I I 18VA2RTN Not used.

J J -18VA2 Not used.

K K +15VS +15VS

L L 15VSRTN 15VSRTN

M M -15VS -15VS

N N SHIELD SHIELD

2.5 POWER 2 CABLE WIRE FUNCTION REASSIGNMENT

The Power 2 cable linking J106 on the interferometer to J103 on the power tray requires no modification. The

cable assembly schedule ARAMRF048007 meets the requirements of the upgraded instrument.

Wire function will change according to the following table:

Table 2–3 Power 2 Cable Wire Function Reassignment

J106 J103 Actual Function Upgraded Function

A A +5VS Not used.

B B 5VSRTN Not used.

C C +5SENSE Not used.

D D -5SENSE Not used.

Page 11: ARIES - ensembles-eu.metoffice.comensembles-eu.metoffice.com/met-res/aries/technical/... · ARIES modules interconnections are shown in the following diagram: Figure 2–1 ARIES Cable

ABB ARIES

ARIES Electrical ICD

Document No: BOM-ARIES-0003

Issue: Draft Rev: – Page

Date: 4 July 2006 5

E E +28VL +28VL

F F 28VLRTN 28VLRTN

G G +24VM1 +24VM1

H H 24VM1RTN 24VM1RTN

I I 24VM2RTN 24VM2RTN

J J +24VM2 +24VM2

K K +12VT +12VT

L L 12VTRTN 12VTRTN

M M SHIELD SHIELD

N N OUTPUT1(-) TEMP/CONT OUTPUT1(-) TEMP/CONT

2.6 HEATER & FAN CABLE WIRE FUNCTION REASSIGNMENT

The Heater & Fan cable linking J107 on the interferometer to J104 on the power tray requires no modification.

The cable assembly schedule ARAMRF048008 meets the requirements of the upgraded instrument.

Wire function will change according to the following table:

Table 2–4 Heater & Fan Cable Wire Function Reassignment

J107 J104 Actual Function Upgraded Function

A A HA HA

B B HB HB

C C HC HC

D D N N

E E SYS/HEAT SYS/HEAT

F F FAN FAN

G G SHIELD SHIELD

H H LINEMON Not used.

I I NEUTRAL Not used.

J J OUTPUT1(+) TEMP/CONT OUTPUT1(+) TEMP/CONT

2.7 ETHERNET CABLE DESCRIPTION

The Ethernet cable will serve as communication link between the laptop PC and the ARIES system. All

command and data exchanges will travel through it, this will be the only link between the laptop PC and the

ARIES system.

The Ethernet connection will be made using a standard category 5e shielded crossover Ethernet cable, using

T568A to T568B wiring standard. The proposed PIC P/N E50824 cable proposed by the customer meets the

requirement. A RJ45 connector that provides shielded termination and connection must be used to allow for

proper cable shielding. This is generally the case for all metal body RJ45 connector.

The military style connector identified to be used on the Optical Head for J125 is of Amphenol RJF series. It

accepts simple RJ45 connector as well as compatible assembly of RJ45 and RFJ series plug. When mated

together, these connectors are sealed against fluids and dusts according to IP67 rating.

Page 12: ARIES - ensembles-eu.metoffice.comensembles-eu.metoffice.com/met-res/aries/technical/... · ARIES modules interconnections are shown in the following diagram: Figure 2–1 ARIES Cable

ABB ARIES

ARIES Electrical ICD

Document No: BOM-ARIES-0003

Issue: Draft Rev: – Page

Date: 4 July 2006 6

On laptop computer side no special connector assembly is required.

The following picture shows the Amphenol RJF series cable conversion approach:

Figure 2–3 Amphenol RJF Series RJ45 to Field Converter

Table 2–5 Ethernet Cable Parts

Reference Part Number Description Location

J125 Amphenol RJF2PEM1G RJ45 field receptacle Optical Head module

P125 Amphenol RJF6MG RJ45 field plug Ethernet cable – Instrument side

RJ45 Tyco 5-569552-X Shielded RJ45, stranded 24 AWG Both ends, suggested part only

Cable PIC E50824 Category 5e shielded Ethernet cable N/A

Figure 2–4 below shows standard crossover Ethernet cable wiring, T568A to T568B standard. Shielded

connection is omitted for clarity.

Figure 2–4 Crossover Ethernet Cable Wiring Diagram

2.8 RS-422 CABLE DESCRIPTION

The RS-422 link will serve as communication means between the Digital Control Module (new electronics,

located in the Optical Head module) and the Scan Electronics Box. All commands and status related to the

pointing mirror and the black bodies will travel through it.

Page 13: ARIES - ensembles-eu.metoffice.comensembles-eu.metoffice.com/met-res/aries/technical/... · ARIES modules interconnections are shown in the following diagram: Figure 2–1 ARIES Cable

ABB ARIES

ARIES Electrical ICD

Document No: BOM-ARIES-0003

Issue: Draft Rev: – Page

Date: 4 July 2006 7

The RS-422 communication will take place over two, individually shielded, 100Ω twisted pair cables with

overall sheath. The cable assembly will link J124 to J159. Military type connectors will be used.

Table 2–6 RS-422 Cable Parts

Reference Part Number Description Location

J124 ITT KPT02E10-6S Box mounting receptacle, female Optical Head Module

P124 ITT KPT06F10-6P Straight plug, male RS-442 cable – Optical Head Module side

J159 Souriau 6020-10-6SN Box mounting receptacle, female Scan Electronics Box

P159 Souriau 6026-10-6PN Straight plug, male RS-422 cable – Scan Electronics Box side

Cable Cable USA M27500-22TG2T14

22 AWG single twisted pair shielded N/A, suggested part only

The following figure shows the wiring diagram of the RS-422 cable to be connected between J124 and J159.

A

F

E

D

C

B

A

F

E

D

C

B

P124 P159

SHIELD

SHIELD SHIELD

WHT/BLU

WHT

SHIELD

WHT/BLU

WHT

WHT

WHT/BLU

WHT

WHT/BLU

DATH

DATL

AGND

COML

COMH

AGND

RX+

RX-

GND

TX-

TX+

GND

Figure 2–5 RS-422 Cable Wiring Diagram

2.9 RS-485 CABLE DESCRIPTION

The RS-485 link will serve to verify and/or modify the set point of the temperature controller located in the

Optical Box. The cable described in this section is not intended to be used during normal system operation. A

cap will close and protect J108 on the Optical Head module during flight conditions.

When required, a communication link will be established with the interferometer internal temperature controller

using a special cable. This cable will include a RS-232 to RS-485 converter. A special software, already included

in the actual version of the instrument, is required to communicate with this controller.

The RS-485 communication will take place over a shielded 100Ω twisted pair cable. The cable assembly will

link J108 on the Optical Head module to a laptop PC’s serial port. Military type connector will be used on the

Optical Head side, while the RS-232 to RS-485 converter and its power supply will be installed on the other end

of the cable.

Page 14: ARIES - ensembles-eu.metoffice.comensembles-eu.metoffice.com/met-res/aries/technical/... · ARIES modules interconnections are shown in the following diagram: Figure 2–1 ARIES Cable

ABB ARIES

ARIES Electrical ICD

Document No: BOM-ARIES-0003

Issue: Draft Rev: – Page

Date: 4 July 2006 8

Table 2–7 RS-485 Cable Parts

Reference Part Number Description Location

J108 ITT KPT02E8-4S Box mounting receptacle, female Optical Head Module

J108 CAP ITT KPT818C Cap for receptacle Optical Head Module

P108 ITT KPT06F8-4P Straight plug, male RS-485 cable – Optical Head Module side

Converter B&B 485OI9TB Isolated RS-232 to RS-485 converter RS-485 cable – Laptop PC side

Power Supply

B&B 485PS2 120VAC to 12VDC power supply RS-485 cable – Laptop PC side

Cable Cable USA M27500-22TG2T14

22 AWG single twisted pair shielded N/A, suggested part only

The following figure shows the wiring diagram of the RS-485 cable to be connected between J108 and the laptop

computer’s RS-232 (DB9 serial) port:

Figure 2–6 RS-485 Cable Wiring Diagram

— End of document —