Artifact Recognition and Troubleshooting

2017 Focus Fall Super Session

The Best of the Best For Respiratory Therapists and Sleep Technologists

The Doubletree Hilton Hotel

Pittsburgh, PA

Thursday Sept. 28, 2017

LAURA A. LINLEY CRTT/RPSGT

Page 2: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

Conflict of Interest Disclosures for Speakers

1. I do not have any potential conflicts to disclose.

2. I wish to disclose the following potential conflicts of interest:

Type of Potential Conflict Details of Potential Conflict

Grant/Research Support

Consultant

Speakers’ Bureaus

Financial support

Other

3. The material presented in this lecture has no relationship with any of these potential conflicts, OR

4. This talk presents material that is related to one or more of these potential conflicts, and the following

objective references are provided as support for this lecture:

1.

2.

3.

X

Page 3: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

OBJECTIVES

1. Physiology, Instrument and environmental artifacts

2. Discuss and demonstrate the appropriateness of filter changes vs. referencing

3. Recognition of high frequency artifacts

4. Recognition of low frequency artifacts

5. Troubleshooting

Page 4: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

4

WHAT'S KEY TO AN ARTIFACT FREE RECORDING

• Good, clean prep

• Balanced impedances

• Good hook-up, neatly bundled electrodes

• Place jack-box close to patients head

• Keep patient cool, not cold

• Unplug all electrical items close to patient, i.e. bed, radio, fan, etc.

Page 5: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

PHYSIOLOGIC VS. NON-PHYSIOLOGIC

• Heart

• Muscles

• Eyes

• Sweat

• Respiratory

• Electrical

• Impedance

• Noise

• Electrode Popping

Page 6: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

NOT ALL ARTIFACT IS BAD

• SNORING ARTIFACT: helps identify sleep onset

• MUSCLE ARTIFACT: helps identify movement in sleep

Page 7: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

GENERAL GUIDELINES

• DOCUMENT and CORRECT

• If artifact is localized to just one channel, it typically indicates the problem lies with one electrode.

• If problem is detected during the impedence/biocalibrations, every effort should be made to resolve prior to lights out.

• If problems presents after patient is asleep, reference or change derivation first with documentation and replace when patient awakes

• CLINICAL JUDGEMENT

Page 8: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

Eye Movement

The eye has a significant polar electrical field known as the corneoretinal potential (CRP).

The front (cornea) of the eye is

positive with to the back (retina).

8

.

AAST Technologist Guide to Sleep Studies Second Edition. Fig 35

Look Left: LEOG Positive– Downward Deflection

Page 9: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

Verify Eye Movement

Look Left ( 1) . Look Right (2) Blink (3)

NOTE: positive part of the eye (the cornea) rotates closer to the right EOG electrode, resulting in a downward deflection. When the eyes open they rotate down, resulting in an upward deflection in the right EOG channel. Voluntary blink artifact in the frontal channels is slower and much higher in amplitude than the spontaneous blinking

Page 10: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

10

EOG Artifact in EEG seen in biocals.

Page 11: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

11

Eye Movement Facts

• Drowsiness and Stage 1 are noted to have slow rolling eye movements.

• REM (rapid eye movements)

• The CRP is often diminished or absent in blind individuals.

• This artifact cannot be eliminated.

Page 12: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

12

Eye movement in the EEG, during REMNormal phasic twitching of legs duringNormal variation in respirations associated with REM.

Page 13: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

13

ECG Artifact

• This artifact is prominent in patients who are obese or have short, thick necks.

• Common Mode Rejection will eliminate if impedance levels among the electrodes are low and equal

• This occurs most often when using a referential montage, particularly if the input impedances are unbalanced.

Page 14: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

14

ECG artifact in EOG, EEG, & EMG.

Page 15: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

15

ECG ARTIFACT FIX

If the offending electrode is one of the mastoid reference electrodes (M1 or M2)

• Solution #1: Re-apply these leads by raising the placement higher up on the mastoid (behind the point of the ear) to minimize the artifact

• Solution #2: Double-reference M1 and M2 to minimize the ECG artifact• Any M1/M2 artifacts will now contaminate all linked channels

Page 16: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

16

DOUBLE REFERENCE

Page 17: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

17

Cardiac Oscillation

• Usually seen when using respiratory channels.

• This artifact often occurs during prolonged central apneas and it is often incorrectly considered to indicate a patent airway.

Page 18: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

18

Cardiac-oscillation in the flow channel

Page 19: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

19

Glossokinetic Potentials

• The tongue, like the eye, is electrically charged. The tip of the tongue is more negative than the root.

• The movement will change the electrical field around the face; this artifact in the EEG is call a glossokinetic potential.

• Commonly seen in very young patients and is due to sucking or chewing.

• Document what the patient is doing.

• Elimination of the artifact is difficult.

Page 20: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

20

INFANT SUCKING ON PACIFIER-ARTIFACT EEG

Page 21: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

21

Infant Sucking-EEG clean

Page 22: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

22

Muscle Artifact• Caused by localized muscle activity in the vicinity of an exploring or

reference electrode. Common in tense or anxious patients.

• There is no correction of the artifact in a sleeping patient

• Muscle artifact may appear in any channel that records a physiological signal. It can also be used to identify bruxism, movement arousals, vocalization, etc.

• Depending on the particular channel filter settings, such as with EEG channels, this artifact can mimic cortical spikes.

Page 23: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

23

Movement artifact, associated with bruxism

Page 24: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have
Page 25: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

MOVEMENT IN SLOW WAVE SLEEP

Page 26: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

26

Muscle artifact in EOG & EEG

Page 27: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

27

Electrode “Pop”

• May simulate a spike or sharp wave.

• Caused by an electrically unstable electrode, by a drying electrode, or by a slight mechanical instability that changes the area of electrode surface in contact with the skin.

Page 28: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

28

Bad Ground-popping

Movement artifact

Chin popping

Page 29: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

29

M1 electrode popping

Leg lead popping/off

Page 30: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

No.

Page 31: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

ELECTRODE “POP” FIX

• Bad O2 electrode (no common referential lead)

• Check lead security as well as pin input in head box

• Solution: Re-apply and/or replace the lead

Page 32: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

32

60 Hz Artifact

1. Appears as a regular, sinusoidal wave and it obscures the physiologic signal.

2. Common Mode Rejection will eliminate if impedance levels among the electrodes are low and equal.

3. Patient Ground Electrode helps dissipate 50/60Hz interference

Page 33: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

Causes of 60 Hz

• High Electrode Impedances

• Excessive Current Leakage: • Ancillary Equipment

• Extension Cords

• TV/Fans

• Long Leg EMG Leads

Page 34: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

34

Intermittent 60 Hz

Page 35: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

35

Vagal nerve stimulator, cycling on for 7

seconds and off for 15 seconds.

Page 36: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

36

60 Hz Artifact-Fix

1. NOTCH Filter (Leg EMG)

2. Replace: Normalize Impedances

3. Electromagnetic-may be reduced by minimizing loops in electrode wires (ponytail).

4. Multiple grounds must be avoided. Unplug anything that could be the cause.

Page 37: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

37

Movement Artifact

• This artifact occurs most often during a position change but can also be produced by limb movement or the characteristic head, neck and mandible movements seen in OSA.

• Patient movement may cause electrode popping and amplifier blocking artifact, particularly if an electrode becomes loose during the study.

• In many cases, movement artifact can be useful when scoring events because it helps to identify arousals when accompanied by an EEG frequency shift.

Page 38: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

38

Movement Artifact Seen in EEG

Page 39: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

39

Awakening-movement artifact

Page 40: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

40

Movement Artifact-Fix

• Usually disappears when patient falls asleep.

• Identify problem electrode and change derivation to back up configuration.

• Bundle the electrodes together to form a neat ponytail. Try to avoid suspending the jack-box over the headboard. It is better if the electrode bundle rests on the bed and the patient does not lay on the wires.

Page 41: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

41

Respiration Artifact

• Respiration artifact in the EEG and EKG appears as a low frequency baseline sway that occurs in phase with the respirations. It is mostly positional.

• It frequently arises in patients with severe OSA & COPD.

Page 42: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

42

Respiratory artifact primarily in M1 & ECG

Leg electrode popping.

Page 43: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

43

Respiratory Artifact in M1

60 Hz in legs

Page 44: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

44

Respiration Artifact-Fix

• The LRR in the EEG/EPG can be raised from 0.3 to 1Hz to decrease the slow wave artifact—document and address when able.

• Reposition the electrode wires or the patient’s head by either raising or lowering it by taking or giving an extra pillow.

• Wrap the electrodes together in a neat bundle to help prevent this.

Page 45: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

45

Sweat Artifact

• Sweat artifact is best described as a low frequency baseline sway that often affects one or more channels.

• Sweat is an electrolyte that increases the conductivity of the skin surface creating a short circuit between two electrodes.

• Profuse perspiration can loosen electrodes making them susceptible to electrode popping severe enough to often cause amplifier blocking artifact.

Page 46: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

46

Sweat Artifact in the EOG & EEG

Page 47: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

47

Sweat Artifact-Fix

• Try to cool the patient, either by reducing the room temperature or removing a blanket.

• Placing a fan in the patient’s room helps but can also be a potential source of 60 Hz artifact.

• Wipe electrodes sites with alcohol

• As a last resort, reduce the low-frequency filter to minimize the effect of sweat artifact. .

Page 48: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

IDENTIFY PROBLEM

Page 49: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

PREVIOUS SLIDE DISCUSSION

• Previous slide shows a run of quadrigemenal PVCs

• This can be also seen on the EMG channels

• Solution: Document your findings

Page 50: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

IDENTIFY PROBLEM

Page 51: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

PREVIOUS SLIDE DISCUSSION• The C3 electrode was off; there was no common

reference electrode (M1, M2)• Solution: Re-apply C3 electrode

• ECG artifact was seen in the EEG and EOG leads• Solution: Double-reference M1 + M2 to attenuate the EKG

artifact

Page 52: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

IDENTIFY PROBLEM

Page 53: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

PREVIOUS SLIDE DISCUSSION

• Respiratory artifact was seen in the occipital leads (O1 and O2)

• Solution: Increase your LFF from 0.3 Hz to either 0.5 Hz, 0.8 Hz or no higher than 1.0 Hz to attenuate the respiratory artifact

Page 54: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

QUESTION 1

If the positive side of the eye is the cornea, and the negative side of the eye is the retina, when the patient moves his eyes to the left, the left EOG electrode becomes more:

a. Positive

b. Negative

c. Stable

d. Upward

Page 55: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

QUESTION 1 ANSWERLook Left….LEOG Positive---Downward Deflectiona. Positive

Page 56: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

QUESTION 2:What’s occurring in this picture?

a. Sweat Artifact EOG

b. ECG Artifact

c. Paradoxical Breathing

Page 57: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

QUESTION 2:What’s occurring in this picture?

B. ECG Artifact

Seen in M1 and M2

Page 58: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

QUESTION 3 : WHAT IS OCCURING?

a. Electrode popping

b. No nasal/oral airflow sensor

c. Paradoxical breathing

d. All the above

Page 59: Artifact Recognition and Troubleshooting - foocus.com · Artifact Recognition and Troubleshooting ... ECG Artifact •This artifact is prominent in patients who are obese or have

Question 3 Answer

ALL THE ABOVE