Dr.Syed Muzzammil Masaud Mphill.Pharmaceutical Chemistry Atomic Absorption & Atomic Emission Spectroscopy

Atomic Absorption & Atomic Emission Spectroscopy

Embed Size (px)

DESCRIPTION

Atomic Absorption & Atomic Emission Spectroscopy. Dr.Syed Muzzammil Masaud Mphill.Pharmaceutical Chemistry. BASIC PRINCIPLE. ATOMIC ABSORPTION SPECTROSCOPY (AAS) is an analytical technique that measures the concentrations of - PowerPoint PPT Presentation

Citation preview

Page 1: Atomic Absorption & Atomic Emission Spectroscopy

Dr.Syed Muzzammil Masaud

Mphill.Pharmaceutical Chemistry

Atomic Absorption& Atomic Emission

Spectroscopy

Page 2: Atomic Absorption & Atomic Emission Spectroscopy

BASIC PRINCIPLE

ATOMIC ABSORPTION SPECTROSCOPY (AAS) is an

analytical technique that measures the concentrations of

elements. It makes use of the absorption of light

by these elements in order to measure their concentration .

Page 3: Atomic Absorption & Atomic Emission Spectroscopy

- Atomic-absorption spectroscopy quantifies the absorption of ground state atoms in the gaseous state .

- The atoms absorb ultraviolet or visible light and make transitions to higher electronic energy levels . The analyte concentration is determined from the amount of

absorption.

Page 4: Atomic Absorption & Atomic Emission Spectroscopy

- Concentration measurements are usually determined from a working curve after calibrating the instrument with standards of known concentration.

- Atomic absorption is a very common technique for detecting metals and metalloids in environmental samples.

Page 5: Atomic Absorption & Atomic Emission Spectroscopy

Elements detectable by atomic absorption are highlighted in pink in this periodic table

                                                                   

                          

Page 6: Atomic Absorption & Atomic Emission Spectroscopy

The Atomic Absorption SpectrometerAtomic absorption spectrometers have 4

principal components1 - A light source ( usually a hollow cathode

lamp )2 – An atom cell ( atomizer )3 - A monochromator4 - A detector , and read out device .

Page 7: Atomic Absorption & Atomic Emission Spectroscopy

Schematic Diagram of an Atomic Absorption Spectrometer

Light source(hollow cathode Lamp )

atomizer monochromat

or

Detector and readout device

Page 8: Atomic Absorption & Atomic Emission Spectroscopy

Atomic Absorption Spectrophotometer

Page 9: Atomic Absorption & Atomic Emission Spectroscopy

1 – Light Source

The light source is usually a hollow cathode lamp of the element that is being

measured . It contains a tungsten anode and a hollow cylindrical cathode made of the element to be determined. These are sealed in a glass tube filled with an inert gas (neon or argon ) . Each element has its own unique lamp which must be used for

that analysis .

Page 10: Atomic Absorption & Atomic Emission Spectroscopy

Hollow Cathode Lamp

cathode

Anode

Quartz window

Pyrex body

Anode

Cathode

Page 11: Atomic Absorption & Atomic Emission Spectroscopy

How it works

Applying a potential difference between the anode and the cathode leads to the ionization of some gas atoms .

These gaseous ions bombard the cathode and eject metal atoms from the cathode in a process called sputtering. Some sputtered atoms are in excited states and emit radiation characteristic of the metal as they fall back to the ground state .

Page 12: Atomic Absorption & Atomic Emission Spectroscopy

Scheme of a hollow cathode lamp

Page 13: Atomic Absorption & Atomic Emission Spectroscopy

The shape of the cathode which is hollow cylindrical concentrates the emitted radiation into a beam which passes through a quartz window

all the way to the vaporized sample.

Since atoms of different elements absorb characteristic wavelengths of light.

Analyzing a sample to see if it contains a particular

element means using light from that element .

Page 14: Atomic Absorption & Atomic Emission Spectroscopy

For example with lead, a lamp containing lead emits light from excited lead atoms that produce the right mix of wavelengths to be absorbed by any lead atoms from the sample .

A beam of the electromagnetic radiation emitted from excited lead atoms is passed through the vaporized sample. Some of the radiation is absorbed by the lead atoms in the sample. The greater the number of atoms there is in the vapor , the more radiation is absorbed .

Page 15: Atomic Absorption & Atomic Emission Spectroscopy

2 – Atomizer

Elements to be analyzed needs to be in atomic sate

Atomization is separation of particles into

individual molecules and breaking molecules into atoms .This is done by exposing the analyte to high temperatures in a flame or graphite furnace .

Page 16: Atomic Absorption & Atomic Emission Spectroscopy

The role of the atom cell is to primarily dissolvate a liquid sample and then the solid particles are vaporized into their free gaseous ground state form . In this form atoms will be available to absorb radiation emitted from the light source and thus generate a measurable signal proportional to concentration .

There are two types of atomization : Flame and Graphite furnace atomization .

Page 17: Atomic Absorption & Atomic Emission Spectroscopy
Page 18: Atomic Absorption & Atomic Emission Spectroscopy

Flame Flame AA can only analyze solutions , where it uses a slot type burner to increase the path length, and therefore to increase the total absorbance . Sample solutions are usually introduced into a nebuliser by being sucked up

a capillary tube .In the nebuliser the sample is dispersed into tiny droplets , which can be readily broken down in the flame.

Page 19: Atomic Absorption & Atomic Emission Spectroscopy

FLAME ATOMIZERS

Used in all Atomic Spectroscopic techniques

Converts analyte into free atoms in the form of vapor phase free atoms

Heat is required

Routes for sample introduction

Page 20: Atomic Absorption & Atomic Emission Spectroscopy

Various flame atomization techniques

Page 21: Atomic Absorption & Atomic Emission Spectroscopy

Types of Flames Used in Atomic Spectroscopy

Page 22: Atomic Absorption & Atomic Emission Spectroscopy

Processes that take place in flame

Page 23: Atomic Absorption & Atomic Emission Spectroscopy

Effect of flame temperature on excited state population

# atoms in Ground state

# atoms in Excited state

Energydifference

Boltzmann constant

Temperature

Statistical factor

Page 24: Atomic Absorption & Atomic Emission Spectroscopy

For Zn: N*/No = 10-15%

Page 25: Atomic Absorption & Atomic Emission Spectroscopy

Thus 99.998% of Na atoms are in the ground stateAtomic emission uses Excited atoms

Atomic absorption uses Ground state atoms

Page 26: Atomic Absorption & Atomic Emission Spectroscopy

Effect of flame temperature on excited state population

Page 27: Atomic Absorption & Atomic Emission Spectroscopy

ATOMIZATION DEVICES

ATOMIZATION : A process of forming free atoms by heatAtomizers are devices that carry out atomization:

ContinuousNon-continuous

Continuous: (Constant temperature with time)FlamePlasma

Non-Continuous: (temperature varies with time)Electrothermal Spark discharge

Page 28: Atomic Absorption & Atomic Emission Spectroscopy

SAMPLE INTRODUCTION SYSTEMS

In continuous atomizers sample is constantly introduced in form of droplets, dry aerosol, vapor

Nebulizer : A device for converting the solution into fine spray or droplets

Continuous sample introduction is used with continuous nebulizers in which a steady state atomic population is produced. Sample is introduced in fixed or discrete amounts.

Discontinuous samplers are used with continuous atomizers

Page 29: Atomic Absorption & Atomic Emission Spectroscopy

1- Discrete samples are introduced into atomizers in many ways:

Electrothermal atomizersa syringe is used

a transient signal is produced as temperature changes

with time and sample is consumed

2- Indirect insertion (Probe) sample is introduced into a probe (carbon rod) and mechanically moved into the atomization region vapor cloud is transient because sample introduced is

limited

Page 30: Atomic Absorption & Atomic Emission Spectroscopy

3-Flow Injection

The analyte is introduced into the carrier stream into a nebulizer as mist

4-Hydride Generation

the volatile sample is stripped from the analyte solution and carried out by a gas into the atomizer. This strip is followed by chemically converting the analyte to hydride vapor form.

Page 31: Atomic Absorption & Atomic Emission Spectroscopy

5- With Arc SparkSolids are employed

6- Laser Microbe TechniqueA beam of laser is directed onto a small solid sample, gets vaporized, atomized by relative heating. Either sample is probed by encoding system or vapor produced is swept into a second absorption or fluorescence

Page 32: Atomic Absorption & Atomic Emission Spectroscopy

Nebulization gas is always compressed, usually acts as the oxidant; it is oxygen (O2) in flame and argon (Ar) in plasma

Nebulization chambers produce smaller droplets and remove or drain larger droplets called aerosol modifiers

Aspiration rate is proportional to compressed gas pressure. The pressure drops through capillary, here 1/4 capillary diameters are recommended. This is inversely proportional to viscocity of the solution

Peristaltic and/or syringe pumps could be used

Page 33: Atomic Absorption & Atomic Emission Spectroscopy

Oxidant and fuel are usually brought into the nebulization chamber through a separate port. They mix and pass the burner head called premixed burner system.

Add organic solvents to reduce the size of the drop

Page 34: Atomic Absorption & Atomic Emission Spectroscopy

The Atomic Absorption Spectrometer Sample Introduction System

Nebuliser

Capillary

Solution

Page 35: Atomic Absorption & Atomic Emission Spectroscopy

The fine mist of droplets is mixed with fuel ( acetylene ) , and oxidant ( nitrous oxide) and burned.

The flame temperature is important because it influences the distribution of atoms. It can be manipulated by oxidant and fuel ratio.

Page 36: Atomic Absorption & Atomic Emission Spectroscopy

Graphite Furnace The graphite furnace has several advantages over a

flame. First it accept solutions, slurries, or solid samples.

Second it is a much more efficient atomizer than a flame and it can directly accept very small absolute quantities of sample. It also provides a reducing environment for easily oxidized elements. Samples are placed directly in the graphite furnace and the furnace is electrically heated in several steps to dry the sample, ash organic matter, and vaporize the analyte atoms.

It accommodates smaller samples but it’s a difficult operation, because the high energy that is provided to atomize the sample particles into ground state atoms might excite the atomized particles into a higher energy level and thus lowering the precision .

Page 37: Atomic Absorption & Atomic Emission Spectroscopy

3- Monochromators

This is a very important part in an AA spectrometer. It is used to separate out all of the thousands of lines. Without a good monochromator, detection limits are severely compromised.

A monochromator is used to select the specific wavelength of light which is absorbed by the sample, and to exclude other wavelengths. The selection of the specific light allows the determination of the selected element in the presence of others.

Page 38: Atomic Absorption & Atomic Emission Spectroscopy

4 - Detector and Read out Device The light selected by the monochromator

is directed onto a detector that is typically a photomultiplier tube , whose function is to convert the light signal into an electrical signal proportional to the light intensity.

The processing of electrical signal is fulfilled by a signal amplifier . The signal could be displayed for readout , or further fed into a data station for printout by the requested format.

Page 39: Atomic Absorption & Atomic Emission Spectroscopy

Calibration Curve A calibration curve is used to determine

the unknown concentration of an element in a solution. The instrument is calibrated using several solutions of known concentrations. The absorbance of each known solution is measured and then a calibration curve of concentration vs absorbance is plotted.

The sample solution is fed into the instrument, and the absorbance of the element in this solution is measured .The unknown concentration of the element is then calculated from the calibration curve

Page 40: Atomic Absorption & Atomic Emission Spectroscopy

Calibration Curve A 1.0 -b 0.9 -S 0.8 - .o 0.7 - .r 0.6 - . b 0.5 - . .a 0.4 - .n 0.3 - . c 0.2 -e 0.1 -

10 20 30 40 50 60 70 80 90 100

Concentration ( g/ml )

Page 41: Atomic Absorption & Atomic Emission Spectroscopy

Determining concentration from

Calibration Curve A 1.0 - absorbance measured

b 0.9 - S 0.8 - .o 0.7 - .r 0.6 - . b 0.5 - . .a 0.4 - .n 0.3 - . concentration

calculated c 0.2 - e 0.1 -

10 20 30 40 50 60 70 80 90 100

Concentration ( mg/l )

Page 42: Atomic Absorption & Atomic Emission Spectroscopy

Interferences The concentration of the analyte element

is considered to be proportional to the ground state atom population in the flame ,any factor that affects the ground state atom population can be classified as an interference .

Factors that may affect the ability of the instrument to read this parameter can also be classified as an interference .

Page 43: Atomic Absorption & Atomic Emission Spectroscopy

The different interferences that are encountered in atomic absorption spectroscopy are :

- Absorption of Source Radiation : Element other than the one of interest may absorb the wavelength being used.

- Ionization Interference : the formation of ions rather than

atoms causes lower absorption of radiation .This problem is overcome by adding ionization suppressors.

- Self Absorption : the atoms of the same kind that are absorbing radiation will absorb more at the center of the line than at the wings ,and thus resulting in the change of shape of the line as well as its intensity .

Page 44: Atomic Absorption & Atomic Emission Spectroscopy

- Back ground Absorption of Source Radiation :

This is caused by the presence of a particle from incomplete atomization .This problem is overcome by increasing the flame temperature .

- Transport Interference : Rate of aspiration, nebulization, or transport of

the sample ( e g viscosity, surface tension, vapor pressure , and density ) .

Page 45: Atomic Absorption & Atomic Emission Spectroscopy

2Atomic Emission Spectroscopy

Atomic emission spectroscopy is also an analytical technique that is used to measure the concentrations of elements in samples .

It uses quantitative measurement of the emission from excited atoms to determine analyte concentration .

Page 46: Atomic Absorption & Atomic Emission Spectroscopy

The analyte atoms are promoted to a higher energy level by the sufficient energy that is provided by the high temperature of the atomization sources .

The excited atoms decay back to lower levels by emitting light . Emissions are passed through monochromators or filters prior to detection by photomultiplier tubes.

Page 47: Atomic Absorption & Atomic Emission Spectroscopy

The instrumentation of atomic emission spectroscopy is the same as that of atomic absorption ,but without the presence of a radiation source .

In atomic Emission the sample is atomized and the analyte atoms are excited to higher energy levels all in the atomizer .

Page 48: Atomic Absorption & Atomic Emission Spectroscopy

Schematic Diagram of an Atomic Emission spectrometer

Page 49: Atomic Absorption & Atomic Emission Spectroscopy

Introduction to AES Atomization Emission Sources

Flame – still used for metal atoms Electric Spark and Arc Direct current Plasmas Microwave Induced Plasma Inductively Coupled Plasma – the most

important technique Advantages of plasma

Simultaneous multi-element Analysis – saves sample amount

Some non-metal determination (Cl, Br, I, and S)

Concentration range of several decades (105 – 106)

Disadvantages of plasma very complex Spectra - hundreds to

thousands of lines High resolution and expensive optical

components Expensive instruments, highly trained

personnel required

Page 50: Atomic Absorption & Atomic Emission Spectroscopy

10A Plasam Source AESPlasma

an electrically conducting gaseous mixture containing significant concentrations of cations and electrons.

Three main typesInductively Coupled Plasma (ICP)Direct Current Plasma (DCP)Microwave Induced Plasma (MIP)

Page 51: Atomic Absorption & Atomic Emission Spectroscopy

ICP Inductively Coupled Plasma (ICP)

Plasma generated in a device called a Torch

Torch up to 1" diameterAr cools outer tube, defines plasma

shapeRapid tangential flow of argon cools

outer quartz and centers plasmaRate of Argon Consumption 5 - 20 L/MinRadio frequency (RF) generator 27 or

41 MHz up to 2 kWTelsa coil produces initiation spark

Ions and e- interact with magnetic field and begin to flow in a circular motion.

Resistance to movement (collisions of e- and cations with ambient gas) leads to ohmic heating.

Sample introduction is analogous to atomic absorption.

Page 52: Atomic Absorption & Atomic Emission Spectroscopy

Sample introductionNebulizerElectrothermal vaporizer

Table 8-2 methods of sample introducton

Page 53: Atomic Absorption & Atomic Emission Spectroscopy

Nebulizer

convert solution to fine spray or aerosol

Ultrasonic nebulizer uses ultrasound

waves to "boil" solution flowing across disc

Pneumatic nebulizer uses high pressure

gas to entrain solution

Page 54: Atomic Absorption & Atomic Emission Spectroscopy

Electro-thermal vaporizer ETV

Electrothermal vaporizer (ETV)electric current

rapidly heats crucible containing sample

sample carried to atomizer by gas (Ar, He)

only for introduction, not atomization

Page 55: Atomic Absorption & Atomic Emission Spectroscopy

Plasma structure

Brilliant white coreAr continuum and lines

Flame-like tail up to 2 cm

Transparent region where measurements are

made (no continuum)

Page 56: Atomic Absorption & Atomic Emission Spectroscopy

Plasma characteristics Hotter than flame (10,000 K) -

more complete atomization/ excitation

Atomized in "inert" atmosphere

Ionization interference small due to high density of e-

Sample atoms reside in plasma for ~2 msec and

Plasma chemically inert, little oxide formation

Temperature profile quite stable and uniform.

Page 57: Atomic Absorption & Atomic Emission Spectroscopy

DC plasmaFirst reported in 1920sDC current (10-15 A)

flows between C anodes and W cathode

Plasma core at 10,000 K, viewing region at ~5,000 K

Simpler, less Ar than ICP - less expensive

Less sensitive than ICPShould replace the

carbon anodes in several hours

Page 58: Atomic Absorption & Atomic Emission Spectroscopy

Atomic Emission SpectrometerMay be >1,000 visible lines (<1 Å) on continuumNeed

higher resolution (<0.1 Å)higher throughputlow stray lightwide dynamic range (>1,000,000)precise and accurate wavelength calibration/intensitiesstabilitycomputer controlled

Three instrument types:sequential (scanning and slew-scanning)Multichannel - Measure intensities of a large number of

elements (50-60) simultaneouslyFourier transform FT-AES

Page 59: Atomic Absorption & Atomic Emission Spectroscopy

Desirable properties of an AE spectrometer

Page 60: Atomic Absorption & Atomic Emission Spectroscopy

Sequential vs. multichannelSequential instrument

PMT moved behind aperture plate, or grating + prism moved to focus new l on exit slitPre-configured exit slits to detect up to 20 lines, slew

scancharacteristics

CheaperSlower

Multichannel instrument Polychromators (not monochromator) - multiple PMT'sArray-based system

charge-injection device/charge coupled devicecharacteristics

Expensive ( > $80,000)Faster

Page 61: Atomic Absorption & Atomic Emission Spectroscopy

Sequential vs. multichannel

Page 62: Atomic Absorption & Atomic Emission Spectroscopy

Sequential monochromatorSlew-scan spectrometers

even with many lines, much spectrum contains no information

rapidly scanned (slewed) across blank regions (between atomic emission lines)From 165 nm to 800 nm in 20 msec

slowly scanned across lines0.01 to 0.001 nm increment

computer control/pre-selected lines to scan

Page 63: Atomic Absorption & Atomic Emission Spectroscopy

Slew scan spectrometer

Two slew-scan gratings

Two PMTs for VIS and UV

Most use holographic grating

Page 64: Atomic Absorption & Atomic Emission Spectroscopy

Scanning echelle spectrometerPMT is moved to monitor signal from slotted aperture.

About 300 photo-etched slits1 second for moving one slit

Can be used as multi channel spectrometerMostly with DC plasma source

Page 65: Atomic Absorption & Atomic Emission Spectroscopy

AES instrument typesThree instrument types:

sequential (scanning and slew-scanning)Multichannel - Measure intensities of a large

number of elements (50-60) simultaneouslyFourier transform FT-AES

Page 66: Atomic Absorption & Atomic Emission Spectroscopy

Multichannel polychromator AES

• Rowland circle• Quantitative det. 20 more elementsWithin 5 minutes

In 10 minutes

Page 67: Atomic Absorption & Atomic Emission Spectroscopy

Applications of AESAES relatively insensitive

small excited state population at moderate temperature

AAS still used more than AESless expensive/less complex instrumentationlower operating costsgreater precision

In practice ~60 elements detectable10 ppb range most metalsLi, K, Rb, Cs strongest lines in IRLarge # of lines, increase chance of overlap

Page 68: Atomic Absorption & Atomic Emission Spectroscopy

Detection power of ICP-AES

Page 69: Atomic Absorption & Atomic Emission Spectroscopy

ICP/OES INTERFERENCES Spectral interferences:

caused by background emission from continuous or recombination phenomena,

stray light from the line emission of high concentration elements, overlap of a spectral line from another element, or unresolved overlap of molecular band spectra.

CorrectionsBackground emission and stray light compensated for by

subtracting background emission determined by measurements adjacent to the analyte wavelength peak.

Correction factors can be applied if interference is well characterized

Inter-element corrections will vary for the same emission line among instruments because of differences in resolution, as determined by the grating, the entrance and exit slit widths, and by the order of dispersion.

Page 70: Atomic Absorption & Atomic Emission Spectroscopy

Physical interferences of ICPcause

effects associated with the sample nebulization and transport processes.

Changes in viscosity and surface tension can cause significant inaccuracies, especially in samples containing high dissolved solids or high acid concentrations.

Salt buildup at the tip of the nebulizer, affecting aerosol flow rate and nebulization.

Reductionby diluting the sample or by using a peristaltic pump, by using an internal standard or by using a high solids nebulizer.

Page 71: Atomic Absorption & Atomic Emission Spectroscopy

Interferences of ICPChemical interferences:

include molecular compound formation, ionization effects, and solute vaporization effects.

Normally, these effects are not significant with the ICP technique.

Chemical interferences are highly dependent on matrix type and the specific analyte element.

Page 72: Atomic Absorption & Atomic Emission Spectroscopy

Memory interferences: When analytes in a previous sample contribute

to the signals measured in a new sample. Memory effects can result

from sample deposition on the uptake tubing to the nebulizer

from the build up of sample material in the plasma torch and spray chamber.

The site where these effects occur is dependent on the element and can be minimized by flushing the system with a rinse blank between

samples.High salt concentrations can cause analyte

signal suppressions and confuse interference tests.

Page 73: Atomic Absorption & Atomic Emission Spectroscopy

Typical Calibration ICP curves

Page 74: Atomic Absorption & Atomic Emission Spectroscopy

Calibration curves of ICP-AES

Page 75: Atomic Absorption & Atomic Emission Spectroscopy

10B. Arc and Spark AESArc and Spark Excitation Sources:

Limited to semi-quantitative/qualitative analysis (arc flicker)

Usually performed on solidsLargely displaced by plasma-AES

Electric current flowing between two C electrodes

Page 76: Atomic Absorption & Atomic Emission Spectroscopy

Carbon electrodesSample pressed into

electrode or mixed with Cu powder and pressed - Briquetting (pelleting)

Cyanogen bands (CN) 350-420 nm occur with C electrodes in air -He, Ar atmosphere

Arc/spark unstable each line measured >20 sneeds multichannel

detection

Page 77: Atomic Absorption & Atomic Emission Spectroscopy

Arc and Spark spectrograph

Page 78: Atomic Absorption & Atomic Emission Spectroscopy

spectrographBeginning 1930sphotographic film

CheapLong integration timesDifficult to develop/analyzeNon-linearity of line "darkness“

Gamma functionPlate calibration

Page 79: Atomic Absorption & Atomic Emission Spectroscopy

Multichannel photoelectric spectrometermultichannel PMT instruments

for rapid determinations (<20 lines) but not versatile

For routine analysis of solidsmetals, alloys, ores, rocks, soils

portable instrumentsMultichannel charge transfer devices

Recently on the marketOrignally developed for plasma sources

Page 80: Atomic Absorption & Atomic Emission Spectroscopy

Comparison Between Atomic Absorption and Emission Spectroscopy Absorption

- Measure trace metal concentrations in complex matrices .

- Atomic absorption depends upon the number of ground state

atoms .

Emission

- Measure trace metal concentrations in complex matrices .

- Atomic emission depends upon the number of excited atoms .

Page 81: Atomic Absorption & Atomic Emission Spectroscopy

- It measures the radiation absorbed by the ground state atoms.

- Presence of a light source ( HCL ) .

- The temperature in the atomizer is adjusted to atomize the analyte atoms in the ground state only.

- It measures the radiation emitted by the excited atoms .

- Absence of the light source .

- The temperature in the atomizer is big enough to atomize the analyte atoms and excite them to a higher energy level.