AU400e™ Main Specifications and System Operations
General CharacteristicsTest Menu Selection: 122On-board Parameters: 99Maximum Tests/Hour: 400 photometric + 400 ISEAnalytical System: Fully selective discrete analyticalMethods: Endpoint
Endpoint with self blankingEndpoint with sample blankingKinetic rateFixed rateISE
Sample Processing Selection: Single, multiple or profileSample Volume: 2 - 50 µL [0.5 µL increments]Sample Type: Serum, plasma, urine, CSFSample Tray Capacity: 80 samples by rack
22 samples by carousel Sample Container: 12 x 75 - 16 x 100 mm tubes
3, 5, 7, 10 mL primary collection tubesand pediatric cups
Sample Transport: 10 position bar-coded rackSTAT Sample Transport: 22 user-definable for STAT, QC, cal, routines,
enclosed and refrigerated moduleSample Quality Analysis: Lipemia, Hemolysis, Icterus
Clot detectionProbe crash prevention
Sample Identification: Bar code reader [Maximum 20 digits; 95 mm]Sample Bar Code Formats: As mixed bar codes:
NW7Code 39Code 1282 of 5 interleaved2 of 5 standardAlone: ISBT code 128
Reagent Identification: Bar code readerReagent Configuration: Concentrated and full strengthReagent Volume: R1 and R2: 25 - 300 µL [1 µL increments]Reagent Delivery: One reagent pipetter dispensingReagent Storage: 4 - 12° CReagent Bottle Capacity: 15, 30, 60 mLReagent Storage Capacity: R1 + R2 = 76 user definedReaction Mixture: Minimum = 150 µL
Maximum = 550 µLReaction Temperature: 37° CReaction Incubation Method: Dry bathReaction Cell: 6 x 8 x 30 mm nondisposable glass cuvetteReaction Time: 8 min. + 37 sec. (maximum)Repeat Run: Automatic repeat run
Increase or decrease sample volumeReflex Testing: User definedSample Predilution: Dilution ratio 5 - 100 by one-fold increments
AU400e™ Main Specifications and System Operations
ISE ModuleAnalytical System: Vertical block flow cell with ISEAnalytical Tests: Sodium [Na+], Potassium [K+], Chloride [Cl -]Sample Throughput: 200 samples/hourMaximum Tests/Hour: 600, if ISE mode only Sample Type: Serum, plasma, urine [undiluted]Sample Volume: 20 µL [serum], 25 µL [urine]Final Dilution Ratio: 1:34.1Measuring Range: Serum and plasma:
Sodium 50 - 200 mEq/LPotassium 1 - 10 mEq/LChloride 50 - 200 mEq/LUrine:Sodium 10 - 400 mEq/LPotassium 2 - 300 mEq/LChloride 25 - 400 mEq/L
Electrode Composition: Sodium Crown-ether membranePotassium Crown-ether membraneChloride Molecular oriented PVC membraneReference Liquid on liquid membrane
Optical CharacteristicsPhotometer: Multi-wavelength diffraction grating spectrophotometerLight Source: Tungsten-halogen lampPhotometric Modes: Monochromatic or bichromaticWavelengths: 340, 380, 410, 450, 480, 520, 540, 570, 600, 660,
700, 750 and 800 nmLight Path: 6 mmPhotodetector: Silicon photo cellPhotometric Range: 0 to 2.5Photometric Sensitivity: 0.0001 OD
Data ProcessorInterface: Standard RS232C; bidirectional
Real time and batch processingProgram Storage: ≥ 10 GB hard drivePatient Storage: 30,000 filesData Input: Alphanumeric keyboard, mouse or interfaceDisplay: 17 in. flat screen monitorPrinter: Dot matrixSoftware: Windows NT®
Graphic user interfaceModem-linked technical supportOnline HELP programUser defined parameter windowReal time QCReaction curve monitorUser defined repeat run criteriaReflex testingDaily test statisticsUser defined calculated testsSample and test oriented data editingCLIA software packageSystem status screensAdvanced calibrationAuto-QC
AU400e™AU400e™
Electrical RequirementsElectrical Consumption: 3.5 KVACurrent [Circuit Breaker Required]: 20 Amp [standard]
30 Amp with UPSCircuit: Dedicated and noise freeVoltage: 208/220 VAC [±10%] single phaseFrequency: 60 [±1 Hz]Ground Requirements: Resistance < 100 ohmsLocation: Power cable length 10 m
Environmental RequirementsAverage Heat Output: 5,000 BTUAmbient Temperature: 18 - 32 ± 2° C during measurementAmbient Humidity: 40 - 80% relative humidityWater Consumption: 20 L/hourWater Type: Deionized CAP Type II, bacteria freeSupply: Continuous flowResistivity: > 1 MegaohmPressure: 7 - 57 PSIControl: Shut off valve within 30 ft.
AU
400e
SP
EC
02
.09-
7.5M
Olympus AU400e Components
ReagentCompartment
CuvetteWasher/Dryer
ISE Module
Reagent Probe Mixer Cuvettes Sample Probe
STAT/QC/Cal/Routines
Rack Loader
AU400e is a trademark of Olympus and Olympus is a registered trademark of Olympus Optical Co., Ltd. Windows NT® is a registered trademark of Microsoft Corp.
Drain RequirementsNon-hazardous Waste: Floor drain required
Maximum height - 4 in. gravityOptional waste pump unitMaximum height - 4 ft.
DimensionsWidth: 57.1 in.Depth: 29.9 in.Height: 47.6 in.Weight: 924 lbs.
Installation Requirements
Olympus America Inc., Diagnostic Systems GroupTwo Corporate Center Drive, Melville, NY 11747-3157 • Tel: 1-800-223-0125
Latin America Division, 6100 Blue Lagoon Drive, Suite 390, Miami, FL 33126 • Tel: 305-260-4240 • Fax: 305-266-9168www.olympus.com
ISO 9001C E R T I F I E DDesign and productionadheres to ISO9001