Basic Measurement1

Embed Size (px)

Citation preview

  • 7/30/2019 Basic Measurement1

    1/22

    P a g e | 1

    Rene Zamarripa

    Basic Measurements4/9- 4/16

    4/23/2012

    Thermo-Fluid Lab MECH 3313

    Dr. Norman Love

  • 7/30/2019 Basic Measurement1

    2/22

    P a g e | 2

    For the pressure gauge the objectives are to practice calibration of a

    measurement device and to practice calculating experimental uncertainty. For the

    viscosity, the objective is to be familiar with experimental methods of viscosity

    measurment and to study the effect of temperature on viscosity variation. For

    temperature, the objectives were to construct and use thermocouples and to

    measure temperature on the surface of a flat plate.

  • 7/30/2019 Basic Measurement1

    3/22

    P a g e | 3

    Table of Contents:

    Introduction.. 7

    Background Theory.7-8

    Experimental Apparatus..9

    Experimental Procedure11

    Data Presentation12

    Discussion.. 17

    Discuss and conclude.12-14

    Results.. 18

    Conclusion19

    Recommendations.2 0

    References..2 1

    Appendices22

  • 7/30/2019 Basic Measurement1

    4/22

    P a g e | 4

    Nomenclature

    -viscosity

    F-Force

    cp-Viscosity Values

    A-area

    x-mean value

    xi-sum of measurements

    sx-standard deviation

    -uncertainty

  • 7/30/2019 Basic Measurement1

    5/22

    P a g e | 5

    List of figure:

    Figure 1-1-Pressure Gauge

    Figure 1-Pressure gauge apparatus

    Figure 2-Viscosity Apparatus

    Figure 3-Temperature apparatus

    Figure 4-calibration curve.

    Figure 5-viscosity versus temperature

    Figure 6- temperature versus time

    Figure 7-temperature versus time with hot plate.

  • 7/30/2019 Basic Measurement1

    6/22

    P a g e | 6

    List of tables

    Table 1- pressure gauge including all of the 5 pressures.

    Table 2- oil numbers along with the rpm and the viscosity values.

  • 7/30/2019 Basic Measurement1

    7/22

    P a g e | 7

    Introduction

    Pressure Gauge

    In the experiment we used a Bourdon tube gauge. The Bourdon pressure gauge

    uses the principle that a flattened tube tends to change to be straightened or

    larger circular cross-section when pressurized. Bourdon tubes measure gauge

    pressure, relative to ambient atmospheric pressure, as opposed to absolute

    pressure; vacuum is sensed as a reverse motion. We also used a dead weight

    tester. A dead weight tester apparatus uses known traceable weights to apply

    pressure to a fluid for checking the accuracy of readings from a pressure gauge. A

    dead weight tester (DWT) is a calibration standard method that uses a piston

    cylinder on which a load is placed to make equilibrium with an applied pressure

    underneath the piston. Deadweight testers take the pressure measured by a

    deadweight tester is defined through other quantities: length, mass and time.

    Typically deadweight testers are used in calibration laboratories to calibrate

    pressure transfer standards like electronic pressure measuring devices.

    Figure 1-1

    Viscosity

    Viscosity is a measure of the resistance of a fluid which is being deformed by

    either shear or tensile stress. In everyday terms (and for fluids only), viscosity is

    "thickness" or "internal friction". Thus, water is "thin", having a lower viscosity,

    while honey is "thick", having a higher viscosity. Put simply, the less viscous the

    fluid is, the greater its ease of movement (fluidity).

  • 7/30/2019 Basic Measurement1

    8/22

    P a g e | 8

    The applied force is proportional to the area and velocity gradient in the fluid:

    Absolute Viscosity

    Temperature

    Temperature is a physical property of matter that quantitatively expresses the

    common notions of hot and cold. Objects of low temperature are cold, while

    various degrees of higher temperatures are referred to as warm or hot. Heat

    spontaneously flows from bodies of a higher temperature to bodies of lower

    temperature; at a rate that increases with the temperature difference and the

    thermal conductivity. In this experiment we used T-Type thermocouple wires

    which were red and blue. The red wire had the conducting metal of Constantan

    and blue had copper. These wires can withstand temperatures between -200 to

    300 Celsius.A thermocouple is a device consisting of two different conductors

    (usually metal alloys) that produce a voltage, proportional to a temperature

    difference, between either ends of the two conductors. We use the voltage of theatmosphere to calculate the temperature.

  • 7/30/2019 Basic Measurement1

    9/22

    P a g e | 9

    Figure 1

    Figure 2

    Figure 3

    http://meter.com.my/datalogger/software_clip_image002.gifhttp://meter.com.my/datalogger/software_clip_image002.gifhttp://meter.com.my/datalogger/software_clip_image002.gif
  • 7/30/2019 Basic Measurement1

    10/22

    P a g e | 10

    Apparatus

    Pressure Gauge

    Bourdon-Tube gauge

    Dead weight tester

    Weights

    Viscosity

    Rotary Viscometer

    Thermometer

    1000 ml beaker

    Hot plate

    Stand

    Temperature

    T-Type thermocouple wires

    Thermocouple welder

    Thermometer Hot plate

    Pc data acquisition system

  • 7/30/2019 Basic Measurement1

    11/22

    P a g e | 11

    Pressure gauge:

    1. First, we set up the dead weight making sure the valve on the side was fully

    closed.

    2. Then we made our baseline reading which was the resting place of the gauge inwhich there was no pressure exerted.

    3. Then we placed a weight on top and gave it a few pumps on the lever until the

    weight slightly rose.

    4. Then we took down the pressure at which it was at

    5. We did these steps 5 with the same weight and repeated more times with

    different weights

    Viscosity1. First we set up oil 1 under the rotary viscometer

    2. Then selected an rpm and gathered its viscosity value.

    3. We did this for 3 different rpm settings

    4. Then we proceeded to turn on the hot plate and took down the

    temperature of the oil at room temperature for a desired rpm.

    5. Then we took 5 temperatures randomly along with the viscosity value.

    6. After we took the temperatures, we change the oil to the next one and

    placed it under the viscometer.

    7. We then proceeded to gather its viscosity value for 3 different rpm settings.

    8. Then we turned on the hot plate one last time and took the temperature

    ever increments of 5 degrees Fahrenheit along with the viscosity value

    Temperature

    1. First we cut a thermocouple wire and shaved off the ends to expose thewire.

    2. Using the thermocouple welder, we made a butt-joint at one end of the

    thermocouple wire.

  • 7/30/2019 Basic Measurement1

    12/22

    P a g e | 12

    3. Then we connected the wire into data acquisition software and we each

    took room temperature every one of us taking 30 seconds longer than the

    last.

    4. Then we turned on the hot plate

    5. We used the last thermocouple which was connected to the software and

    took the temperature readings

    6. As soon as the temperature on the software would become constant, we

    would take the time and increase the temperature on the hotplate

    7. We did this for 3 different temperatures

  • 7/30/2019 Basic Measurement1

    13/22

    P a g e | 13

    Pressure Gauge

    Table 1

    L40 L50 L60 L70 L80

    430 PSI 439 PSI 441 PSI 458 PSI 462 PSI431 PSI 440 PSI 441 PSI 458 PSI 462 PSI

    431 PSI 440 PSI 441 PSI 459 PSI 462 PSI

    431 PSI 440 PSI 442 PSI 458 PSI 463 PSI

    430 PSI 439 PSI 442 PSI 458 PSI 463 PSI

    Discuss and Conclude

    1. Pressure is being calibrated in this case because of the piston it has in the

    tester. As more pressure in the fluid increases, the weight slowly rises, thus

    creating the pressure reading.

    2

    figure 4

    4 The deviation between the actual and measured values is rather high. This is

    due to the calibration of the gauge which at normal pressure was 400.

    410

    420

    430

    440

    450

    460

    470

    40 50 60 70 80

    Measuredvalue

    True Value

    Calibration Curve

    L40

    L50

    L60

    L70

    L80

  • 7/30/2019 Basic Measurement1

    14/22

    P a g e | 14

    Viscosity

    Table 2

    Oil 1

    SPEED

    20

    RPM

    30

    RPM

    50

    RPM

    OIL 2

    SPEED

    12

    RPM

    20

    RPM

    30

    RPM

    CP 101.1 106.4 109.7 CP 116 122.7 127.4

    30

    RPM

    TEMP

    F

    CP 30

    RPM

    TEMP

    F

    CP

    75 106.4 75 130

    78 86 80 125.6

    82 78.4 85 124

    84 70.4 90 95

    90 61.2 95 75.6

    Figure 5

    0

    20

    40

    60

    80

    100

    120

    140

    75 80 85 90 95

    V

    iscosity

    Temperature in F

    Viscosity vs Temp

    Viscosity oil 1

    Viscosity Oil 2

  • 7/30/2019 Basic Measurement1

    15/22

    P a g e | 15

    Temperature

    Figure 6

    Figure 4

    Figure 7

    Discuss and conclude

    2. It takes 1:48 for the first setting and 4:14 for the second and 4:50 for the last.

    This was because the temperature of the hot plate was not yet hot enough.

    22.2

    22.4

    22.6

    22.8

    23

    23.2

    23.4

    23.6

    23.8

    24

    143

    85

    127

    169

    211

    253

    295

    337

    379

    421

    463

    505

    547

    589

    631

    673

    715

    757

    TemperatureinC

    Time

    Temerature vs Time

    Test 1

    Test 2

    Test 3

    Test 4

    0

    20

    40

    60

    80

    100

    120

    140

    160

    180

    198

    195

    292

    389

    486

    583

    680

    777

    874

    971

    1068

    1165

    1262

    1359

    1456

    1553

    1650

    1747

    Temp

    Time

    Temp vs time with hot plate

    Test 1

    Test 2

    Test 3

  • 7/30/2019 Basic Measurement1

    16/22

    P a g e | 16

    In table 1, we have all of the data acquired for the pressure gauge including all of

    the 5 pressures.

    In table 2, we have all of the data including the oil numbers along with the rpm

    and the viscosity values.

    In figure 4 its the calibration curve.

    In figure 5 it is the viscosity versus temperature in the viscosity lab

    In figure 6 it is temperature versus time in the temperature lab

    In figure 7 its the temperature versus time but with the hot plate on.

  • 7/30/2019 Basic Measurement1

    17/22

    P a g e | 17

    Pressure gauge Discussion:

    The data we got for the pressure gauge experiment was in accordance to the

    procedure because as we put the dead weights on the tester and pumped the

    lever to increase the pressure, we would slowly levitate the weight and thepressure would be recorded on the gauge. But since the gauge was not calibrated

    perfectly at zero, we had to start the experiment and calibrate it at 400 psi. This

    however did not change the experiment outcomes; we could still get the actual

    measurement of these weights by subtracting the 400.

    Viscosity Discussion:

    The data that we acquired for this experiment was according to the procedure

    because as the rotary viscometer would begin to spin at a certain rpm with oil 1,

    the viscosity values would begin to show on the screen. This however would vary

    as we would increase the rpm of the viscometer, steadily increasing the values. As

    we changed the oils to oil 2, we automatically began to see the changes in the

    viscosity because the rpm was at low revolutions and the value of viscosity was

    rather higher than the oil 1.

    Temperature.

    The data that we acquired for this experiment was according to the procedure

    because as we would connect the thermocouples that we welded together into

    the data acquisition software, the temperature of the surrounding air would show

    up on the data acquisition program. As we would wait several seconds holding the

    thermocouple in the empty space, the temperature would fluctuate, thus not

    being very accurate. So we then took turns connecting our own thermocouples

    into the software and increasing the time we would hold it by 30 seconds, we

    would began to get more accurate as the time would go on.

  • 7/30/2019 Basic Measurement1

    18/22

    P a g e | 18

    Pressure gauge

    In table #1, we can see all the major data that was collected from the

    experiment, including, the dead weight and the pressures we got with our

    calibrated gauge. Since the gauge was not at zero, we calibrated it to 400 andthen just subtract the 400 by the calibrated results and compare them to the

    dead weights results. In figure 4, the calibration curve, we can see the different

    pressures compared to the dead weight pressure. As the dead weights pressure

    rose by 10, the calibrated pressure rose, but not at a steady increasing rate.

    Viscosity:

    In table #2, we can see all the major data that was collected from the experiment,

    including the rpm settings in both oil 1 and 2, the temperatures and the viscosity

    values. We saw that in oil 1 the rpm setting would be slightly higher than that of

    oil 2 meaning that the viscosity of the oil 2 was greater. Then as we would turn on

    the hot plate and stir the oils, we would see that the viscosity values would

    sharply decrease. This occurred in both oils as seen in figure 5. Both oils were put

    to the hot plate experiment and as the graph show that oil 2 temperature would

    increase, viscosity sharply drops almost constantly. To maintain the constant

    pressure throughout the oil we had to stir.

    Temperature

    In figures 6 we can see just how accurately our data can be with more time. As

    the thermocouple was exposed in increasing time intervals of 30 seconds more

    each time, the temperature would become a more accurate reading. The

    thermocouple that was held for 2 minutes got the most accurate results as

    shown. In figure 7 we have the graph of the 3 different hot plat settings recorded

    until they reached steady state. We can see that the second setting took a long

    time. An educated guess would be that the hot plate was the culprit becauseitselectrical and takes a long time to heat up the plate.

  • 7/30/2019 Basic Measurement1

    19/22

    P a g e | 19

    Pressure Gauge:

    The main results we got in the pressure gauge experiment was closely related to

    the theory because as the fluid would get compressed in the piston, there was

    just enough pressure to lift the dead weight just enough to keep it suspended inthe air. We would read the results in the Bourdon-Tube gauge. But since the

    gauge wasnt functioning properly, we had to calibrate it at 400 which was theoriginal place the needle was at resting place. The deviation between the actual

    and measured values was rather large because of the calibration of the gauge.

    Viscosity:

    The results that we got clearly resembled what lies in the theory of viscosity.Meaning that when the viscometer would began to spin with a certain rpm, the

    force that the viscometer puts out to create those rpm is the viscosity value. This

    value differs between rpm. For example in oil 1 the viscosity values were between

    100 at 20 rpm. However in oil 2 for 20 rpm the viscosity values were around 120

    meaning that the more viscous of the two oils was oil 2. But as we began to

    increase and add heat to the oils, the viscosity of the oils began to decrease

    meaning that as the oil heats up the viscosity decreases.

    Temperature

    The results that we obtained in the temperature experiment helped us to

    construct thermocouples and measure temperature with them. This clearly

    resembles the underlying theory. As the thermocouple sits for a long time at

    steady temperature, the results are more accurate than a thermocouple that sits

    for 30 seconds. This same theory applies to thermometers because we need to

    leave the thermometer at the temperature we want to measure as to getaccurate results. But as we began applying heat to the thermocouple the time

    interval between that moment and steady state would take several moments,

    because of the electric hot plate that takes its time to accomplish the desired heat

    setting.

  • 7/30/2019 Basic Measurement1

    20/22

    P a g e | 20

    Recommendation

    Pressure gauge

    A recommendation for the pressure gauge would be to have computerized

    software that accurately measures the pressure.

    Viscosity

    A recommendation for the viscosity experiment would be for the viscometer to

    give steady cp results.

    Temperature

    A recommendation for the temperature experiment would be for the software to

    have an immediate reading of the temperature instead of waiting for steady

    state.

  • 7/30/2019 Basic Measurement1

    21/22

    P a g e | 21

    References

    1) fluidengineering.co.nr/Manometer.htm

    2) http://www.omega.com/literature/transactions/volume3/pressure3.html

    3)J. P. Holman Heat Transfer, McGraw-Hill, 2002.

    4) "Thermocouple temperature sensors". Temperatures.com. Retrieved 2007-11-04.

    5) Kondepudi, D. (2008). Introduction to Modern Thermodynamics, Wiley, Chichester, Section32. pages 106-108.

  • 7/30/2019 Basic Measurement1

    22/22

    P a g e | 22

    Appendices

    Uncertainty analysis

    Pressure Gauge

    Xi

    sx

    =30.684.94 =39.6109.9 =41.4114 =58.2161.5 =62.4173.2

    Viscosity

    Oil 1 Oil 2

    xi

    402.4 550.2

    x

    80.48 110.04

    sx

    160.96 220.08

    223.41248 305.471

    =80.48223.41 =110.04305.474

    153 198 207 291 312

    30.6 39.6 41.4 58.2 62.4

    61.2 79.2 82.8 116.4 124.8

    84.9456 109.9296 114.9264 161.5632 173.2224