Upload
abdalla-faris
View
215
Download
0
Embed Size (px)
Citation preview
7/30/2019 Basic Measurement1
1/22
P a g e | 1
Rene Zamarripa
Basic Measurements4/9- 4/16
4/23/2012
Thermo-Fluid Lab MECH 3313
Dr. Norman Love
7/30/2019 Basic Measurement1
2/22
P a g e | 2
For the pressure gauge the objectives are to practice calibration of a
measurement device and to practice calculating experimental uncertainty. For the
viscosity, the objective is to be familiar with experimental methods of viscosity
measurment and to study the effect of temperature on viscosity variation. For
temperature, the objectives were to construct and use thermocouples and to
measure temperature on the surface of a flat plate.
7/30/2019 Basic Measurement1
3/22
P a g e | 3
Table of Contents:
Introduction.. 7
Background Theory.7-8
Experimental Apparatus..9
Experimental Procedure11
Data Presentation12
Discussion.. 17
Discuss and conclude.12-14
Results.. 18
Conclusion19
Recommendations.2 0
References..2 1
Appendices22
7/30/2019 Basic Measurement1
4/22
P a g e | 4
Nomenclature
-viscosity
F-Force
cp-Viscosity Values
A-area
x-mean value
xi-sum of measurements
sx-standard deviation
-uncertainty
7/30/2019 Basic Measurement1
5/22
P a g e | 5
List of figure:
Figure 1-1-Pressure Gauge
Figure 1-Pressure gauge apparatus
Figure 2-Viscosity Apparatus
Figure 3-Temperature apparatus
Figure 4-calibration curve.
Figure 5-viscosity versus temperature
Figure 6- temperature versus time
Figure 7-temperature versus time with hot plate.
7/30/2019 Basic Measurement1
6/22
P a g e | 6
List of tables
Table 1- pressure gauge including all of the 5 pressures.
Table 2- oil numbers along with the rpm and the viscosity values.
7/30/2019 Basic Measurement1
7/22
P a g e | 7
Introduction
Pressure Gauge
In the experiment we used a Bourdon tube gauge. The Bourdon pressure gauge
uses the principle that a flattened tube tends to change to be straightened or
larger circular cross-section when pressurized. Bourdon tubes measure gauge
pressure, relative to ambient atmospheric pressure, as opposed to absolute
pressure; vacuum is sensed as a reverse motion. We also used a dead weight
tester. A dead weight tester apparatus uses known traceable weights to apply
pressure to a fluid for checking the accuracy of readings from a pressure gauge. A
dead weight tester (DWT) is a calibration standard method that uses a piston
cylinder on which a load is placed to make equilibrium with an applied pressure
underneath the piston. Deadweight testers take the pressure measured by a
deadweight tester is defined through other quantities: length, mass and time.
Typically deadweight testers are used in calibration laboratories to calibrate
pressure transfer standards like electronic pressure measuring devices.
Figure 1-1
Viscosity
Viscosity is a measure of the resistance of a fluid which is being deformed by
either shear or tensile stress. In everyday terms (and for fluids only), viscosity is
"thickness" or "internal friction". Thus, water is "thin", having a lower viscosity,
while honey is "thick", having a higher viscosity. Put simply, the less viscous the
fluid is, the greater its ease of movement (fluidity).
7/30/2019 Basic Measurement1
8/22
P a g e | 8
The applied force is proportional to the area and velocity gradient in the fluid:
Absolute Viscosity
Temperature
Temperature is a physical property of matter that quantitatively expresses the
common notions of hot and cold. Objects of low temperature are cold, while
various degrees of higher temperatures are referred to as warm or hot. Heat
spontaneously flows from bodies of a higher temperature to bodies of lower
temperature; at a rate that increases with the temperature difference and the
thermal conductivity. In this experiment we used T-Type thermocouple wires
which were red and blue. The red wire had the conducting metal of Constantan
and blue had copper. These wires can withstand temperatures between -200 to
300 Celsius.A thermocouple is a device consisting of two different conductors
(usually metal alloys) that produce a voltage, proportional to a temperature
difference, between either ends of the two conductors. We use the voltage of theatmosphere to calculate the temperature.
7/30/2019 Basic Measurement1
9/22
P a g e | 9
Figure 1
Figure 2
Figure 3
http://meter.com.my/datalogger/software_clip_image002.gifhttp://meter.com.my/datalogger/software_clip_image002.gifhttp://meter.com.my/datalogger/software_clip_image002.gif7/30/2019 Basic Measurement1
10/22
P a g e | 10
Apparatus
Pressure Gauge
Bourdon-Tube gauge
Dead weight tester
Weights
Viscosity
Rotary Viscometer
Thermometer
1000 ml beaker
Hot plate
Stand
Temperature
T-Type thermocouple wires
Thermocouple welder
Thermometer Hot plate
Pc data acquisition system
7/30/2019 Basic Measurement1
11/22
P a g e | 11
Pressure gauge:
1. First, we set up the dead weight making sure the valve on the side was fully
closed.
2. Then we made our baseline reading which was the resting place of the gauge inwhich there was no pressure exerted.
3. Then we placed a weight on top and gave it a few pumps on the lever until the
weight slightly rose.
4. Then we took down the pressure at which it was at
5. We did these steps 5 with the same weight and repeated more times with
different weights
Viscosity1. First we set up oil 1 under the rotary viscometer
2. Then selected an rpm and gathered its viscosity value.
3. We did this for 3 different rpm settings
4. Then we proceeded to turn on the hot plate and took down the
temperature of the oil at room temperature for a desired rpm.
5. Then we took 5 temperatures randomly along with the viscosity value.
6. After we took the temperatures, we change the oil to the next one and
placed it under the viscometer.
7. We then proceeded to gather its viscosity value for 3 different rpm settings.
8. Then we turned on the hot plate one last time and took the temperature
ever increments of 5 degrees Fahrenheit along with the viscosity value
Temperature
1. First we cut a thermocouple wire and shaved off the ends to expose thewire.
2. Using the thermocouple welder, we made a butt-joint at one end of the
thermocouple wire.
7/30/2019 Basic Measurement1
12/22
P a g e | 12
3. Then we connected the wire into data acquisition software and we each
took room temperature every one of us taking 30 seconds longer than the
last.
4. Then we turned on the hot plate
5. We used the last thermocouple which was connected to the software and
took the temperature readings
6. As soon as the temperature on the software would become constant, we
would take the time and increase the temperature on the hotplate
7. We did this for 3 different temperatures
7/30/2019 Basic Measurement1
13/22
P a g e | 13
Pressure Gauge
Table 1
L40 L50 L60 L70 L80
430 PSI 439 PSI 441 PSI 458 PSI 462 PSI431 PSI 440 PSI 441 PSI 458 PSI 462 PSI
431 PSI 440 PSI 441 PSI 459 PSI 462 PSI
431 PSI 440 PSI 442 PSI 458 PSI 463 PSI
430 PSI 439 PSI 442 PSI 458 PSI 463 PSI
Discuss and Conclude
1. Pressure is being calibrated in this case because of the piston it has in the
tester. As more pressure in the fluid increases, the weight slowly rises, thus
creating the pressure reading.
2
figure 4
4 The deviation between the actual and measured values is rather high. This is
due to the calibration of the gauge which at normal pressure was 400.
410
420
430
440
450
460
470
40 50 60 70 80
Measuredvalue
True Value
Calibration Curve
L40
L50
L60
L70
L80
7/30/2019 Basic Measurement1
14/22
P a g e | 14
Viscosity
Table 2
Oil 1
SPEED
20
RPM
30
RPM
50
RPM
OIL 2
SPEED
12
RPM
20
RPM
30
RPM
CP 101.1 106.4 109.7 CP 116 122.7 127.4
30
RPM
TEMP
F
CP 30
RPM
TEMP
F
CP
75 106.4 75 130
78 86 80 125.6
82 78.4 85 124
84 70.4 90 95
90 61.2 95 75.6
Figure 5
0
20
40
60
80
100
120
140
75 80 85 90 95
V
iscosity
Temperature in F
Viscosity vs Temp
Viscosity oil 1
Viscosity Oil 2
7/30/2019 Basic Measurement1
15/22
P a g e | 15
Temperature
Figure 6
Figure 4
Figure 7
Discuss and conclude
2. It takes 1:48 for the first setting and 4:14 for the second and 4:50 for the last.
This was because the temperature of the hot plate was not yet hot enough.
22.2
22.4
22.6
22.8
23
23.2
23.4
23.6
23.8
24
143
85
127
169
211
253
295
337
379
421
463
505
547
589
631
673
715
757
TemperatureinC
Time
Temerature vs Time
Test 1
Test 2
Test 3
Test 4
0
20
40
60
80
100
120
140
160
180
198
195
292
389
486
583
680
777
874
971
1068
1165
1262
1359
1456
1553
1650
1747
Temp
Time
Temp vs time with hot plate
Test 1
Test 2
Test 3
7/30/2019 Basic Measurement1
16/22
P a g e | 16
In table 1, we have all of the data acquired for the pressure gauge including all of
the 5 pressures.
In table 2, we have all of the data including the oil numbers along with the rpm
and the viscosity values.
In figure 4 its the calibration curve.
In figure 5 it is the viscosity versus temperature in the viscosity lab
In figure 6 it is temperature versus time in the temperature lab
In figure 7 its the temperature versus time but with the hot plate on.
7/30/2019 Basic Measurement1
17/22
P a g e | 17
Pressure gauge Discussion:
The data we got for the pressure gauge experiment was in accordance to the
procedure because as we put the dead weights on the tester and pumped the
lever to increase the pressure, we would slowly levitate the weight and thepressure would be recorded on the gauge. But since the gauge was not calibrated
perfectly at zero, we had to start the experiment and calibrate it at 400 psi. This
however did not change the experiment outcomes; we could still get the actual
measurement of these weights by subtracting the 400.
Viscosity Discussion:
The data that we acquired for this experiment was according to the procedure
because as the rotary viscometer would begin to spin at a certain rpm with oil 1,
the viscosity values would begin to show on the screen. This however would vary
as we would increase the rpm of the viscometer, steadily increasing the values. As
we changed the oils to oil 2, we automatically began to see the changes in the
viscosity because the rpm was at low revolutions and the value of viscosity was
rather higher than the oil 1.
Temperature.
The data that we acquired for this experiment was according to the procedure
because as we would connect the thermocouples that we welded together into
the data acquisition software, the temperature of the surrounding air would show
up on the data acquisition program. As we would wait several seconds holding the
thermocouple in the empty space, the temperature would fluctuate, thus not
being very accurate. So we then took turns connecting our own thermocouples
into the software and increasing the time we would hold it by 30 seconds, we
would began to get more accurate as the time would go on.
7/30/2019 Basic Measurement1
18/22
P a g e | 18
Pressure gauge
In table #1, we can see all the major data that was collected from the
experiment, including, the dead weight and the pressures we got with our
calibrated gauge. Since the gauge was not at zero, we calibrated it to 400 andthen just subtract the 400 by the calibrated results and compare them to the
dead weights results. In figure 4, the calibration curve, we can see the different
pressures compared to the dead weight pressure. As the dead weights pressure
rose by 10, the calibrated pressure rose, but not at a steady increasing rate.
Viscosity:
In table #2, we can see all the major data that was collected from the experiment,
including the rpm settings in both oil 1 and 2, the temperatures and the viscosity
values. We saw that in oil 1 the rpm setting would be slightly higher than that of
oil 2 meaning that the viscosity of the oil 2 was greater. Then as we would turn on
the hot plate and stir the oils, we would see that the viscosity values would
sharply decrease. This occurred in both oils as seen in figure 5. Both oils were put
to the hot plate experiment and as the graph show that oil 2 temperature would
increase, viscosity sharply drops almost constantly. To maintain the constant
pressure throughout the oil we had to stir.
Temperature
In figures 6 we can see just how accurately our data can be with more time. As
the thermocouple was exposed in increasing time intervals of 30 seconds more
each time, the temperature would become a more accurate reading. The
thermocouple that was held for 2 minutes got the most accurate results as
shown. In figure 7 we have the graph of the 3 different hot plat settings recorded
until they reached steady state. We can see that the second setting took a long
time. An educated guess would be that the hot plate was the culprit becauseitselectrical and takes a long time to heat up the plate.
7/30/2019 Basic Measurement1
19/22
P a g e | 19
Pressure Gauge:
The main results we got in the pressure gauge experiment was closely related to
the theory because as the fluid would get compressed in the piston, there was
just enough pressure to lift the dead weight just enough to keep it suspended inthe air. We would read the results in the Bourdon-Tube gauge. But since the
gauge wasnt functioning properly, we had to calibrate it at 400 which was theoriginal place the needle was at resting place. The deviation between the actual
and measured values was rather large because of the calibration of the gauge.
Viscosity:
The results that we got clearly resembled what lies in the theory of viscosity.Meaning that when the viscometer would began to spin with a certain rpm, the
force that the viscometer puts out to create those rpm is the viscosity value. This
value differs between rpm. For example in oil 1 the viscosity values were between
100 at 20 rpm. However in oil 2 for 20 rpm the viscosity values were around 120
meaning that the more viscous of the two oils was oil 2. But as we began to
increase and add heat to the oils, the viscosity of the oils began to decrease
meaning that as the oil heats up the viscosity decreases.
Temperature
The results that we obtained in the temperature experiment helped us to
construct thermocouples and measure temperature with them. This clearly
resembles the underlying theory. As the thermocouple sits for a long time at
steady temperature, the results are more accurate than a thermocouple that sits
for 30 seconds. This same theory applies to thermometers because we need to
leave the thermometer at the temperature we want to measure as to getaccurate results. But as we began applying heat to the thermocouple the time
interval between that moment and steady state would take several moments,
because of the electric hot plate that takes its time to accomplish the desired heat
setting.
7/30/2019 Basic Measurement1
20/22
P a g e | 20
Recommendation
Pressure gauge
A recommendation for the pressure gauge would be to have computerized
software that accurately measures the pressure.
Viscosity
A recommendation for the viscosity experiment would be for the viscometer to
give steady cp results.
Temperature
A recommendation for the temperature experiment would be for the software to
have an immediate reading of the temperature instead of waiting for steady
state.
7/30/2019 Basic Measurement1
21/22
P a g e | 21
References
1) fluidengineering.co.nr/Manometer.htm
2) http://www.omega.com/literature/transactions/volume3/pressure3.html
3)J. P. Holman Heat Transfer, McGraw-Hill, 2002.
4) "Thermocouple temperature sensors". Temperatures.com. Retrieved 2007-11-04.
5) Kondepudi, D. (2008). Introduction to Modern Thermodynamics, Wiley, Chichester, Section32. pages 106-108.
7/30/2019 Basic Measurement1
22/22
P a g e | 22
Appendices
Uncertainty analysis
Pressure Gauge
Xi
sx
=30.684.94 =39.6109.9 =41.4114 =58.2161.5 =62.4173.2
Viscosity
Oil 1 Oil 2
xi
402.4 550.2
x
80.48 110.04
sx
160.96 220.08
223.41248 305.471
=80.48223.41 =110.04305.474
153 198 207 291 312
30.6 39.6 41.4 58.2 62.4
61.2 79.2 82.8 116.4 124.8
84.9456 109.9296 114.9264 161.5632 173.2224