3/2015 23-7739-03
BD, BD Logo and all other trademarks are property of Becton, Dickinson and Company. © 2015 BD
Becton, Dickinson and CompanyBD Biosciences2350 Qume DriveSan Jose, CA 95131 USA
Benex LimitedPottery Road, Dun Laoghaire,Co. Dublin, IrelandTel +353.1.202.5222Fax +353.1.202.5388
BD BiosciencesEuropean Customer SupportTel +32.2.400.98.95Fax [email protected]
Becton Dickinson Pty Ltd,4 Research Park Drive,Macquarie University Research Park,North Ryde NSW 2113, Australia
Becton Dickinson Limited,8 Pacific Rise, Mt. Wellington,Auckland, New Zealand
IVD
BD™ Plasma Count Kit
50 Tests per kit—Catalog No. 338331
23-7739-03-en.book Page i Wednesday, March 25, 2015 5:43 PM
23-7739-03-en.book Page ii Wednesday, March 25, 2015 5:43 PM
iii
CONTENTS
1. INTENDED USE ............................................................................ 5
2. SUMMARY AND EXPLANATION ................................................ 5
3. PRINCIPLES OF THE PROCEDURE................................................ 5
4. REAGENT..................................................................................... 6
Precautions ................................................................................... 7
Storage and Handling ................................................................... 7
5. INSTRUMENT............................................................................... 8
6. SPECIMEN COLLECTION AND PREPARATION............................ 9
Interfering Conditions .................................................................. 9
7. PROCEDURE ................................................................................ 9
Reagents Provided ........................................................................ 9
Reagents and Materials Required but Not Provided ..................... 9
Staining Procedure...................................................................... 10
BD FACSVia Instrument Quality Control................................... 12
BD FACS Brand Instrument Setup .............................................. 12
Acquiring Samples ...................................................................... 13
Data Analysis ............................................................................. 15
8. RESULTS..................................................................................... 18
Calculating Absolute Counts ...................................................... 18
9. LIMITATIONS............................................................................. 19
10. PERFORMANCE CHARACTERISTICS.......................................... 20
23-7739-03-en.book Page iii Wednesday, March 25, 2015 5:43 PM
iv
Precision for BD FACSCalibur flow cytometers.......................... 20
Low Range Study ....................................................................... 24
Linearity for BD FACSCalibur flow cytometers.......................... 27
Stability for BD FACSCalibur flow cytometers ........................... 28
Accuracy for BD FACSCalibur flow cytometers ......................... 28
Precision for BD FACSVia flow cytometers ................................ 30
Linearity for BD FACSVia flow cytometers ................................ 31
Accuracy for BD FACSVia flow cytometers................................ 32
WARRANTY..................................................................................... 35
TROUBLESHOOTING ....................................................................... 36
REFERENCES .................................................................................... 36
23-7739-03-en.book Page iv Wednesday, March 25, 2015 5:43 PM
5
BD™ PLASMA COUNT KIT
1. INTENDED USE
The BD™ Plasma Count kit is for in vitro diagnostic use on the BD FACSVia™ system using the BD FACSVia™ clinical software, and on the BD FACSCalibur™ flow cytometer using BD CellQuest™ Pro and BD FACSComp™ software. The kit is intended for in vitro diagnostic use to identify and enumerate residual leucocytes (rWBCs), red blood cells (rRBCs), and platelets (rPLTs) in fresh human plasma, using flow cytometry.
2. SUMMARY AND EXPLANATION
The quality of fresh frozen plasma (FFP) can be determined by counting rWBCs, rRBCs, and rPLTs prior to freezing.1-3 Flow cytometric applications provide a rapid, quantitative, and reproducible method to identify and enumerate residual cell populations.4-9 The BD Plasma Count assay incorporates BD Trucount™ tubes to determine absolute cell counts of rWBCs, rRBCs, and rPLTs using a single tube.
3. PRINCIPLES OF THE PROCEDURE
A single-tube assay is performed by staining a known volume of plasma sample with BD Plasma Count reagents in an individual BD Trucount tube. Fluorochrome-labeled antibodies bind specifically to cell-surface antigens, while the nucleic acid dye stains the DNA and RNA of all nucleated cells. The lyophilized pellet in the BD Trucount tube dissolves, releasing a known number of fluorescent beads.
During analysis, the absolute number (cells/µL) of residual cells in the sample can be determined by dividing the number of residual cellular events by the number of fluorescent bead events, and then multiplying the result by the BD Trucount bead concentration.4
23-7739-03-en.book Page 5 Wednesday, March 25, 2015 5:43 PM
6
4. REAGENT
The BD Plasma Count kit contains BD Plasma Count reagent and BD Trucount tubes, sufficient for 50 tests. The BD Plasma Count kit contains the following components. Each reagent is provided in phosphate-buffered saline (PBS) with stabilizer and 0.1% sodium azide.
• Reagent Vial A: Thiazole orange in stabilization buffer. • Reagent Vial B: FITC-labeled CD235a, clone GAR-2(HIR-2) and
PerCP-Cy™5.5*-labeled CD41a, clone HIP8, pre-titrated in stabilization buffer.
• BD Trucount tubes, each containing a lyophilized pellet of 4.2-µm fluorescent-dyed beads.
Thiazole orange is a nucleic acid dye that penetrates cell membranes and binds with RNA and DNA.10,11
Anti-glycophorin A (CD235a) reacts with a major sialoglycoprotein present on the human erythrocyte membrane.12 Anti-glycophorin A is composed of mouse IgG2b heavy chains and kappa light chains.
Glycophorin A is a transmembrane dimeric complex of 31 kilodaltons (kDa) with its carboxyterminal ends extending into the cytoplasm of the red cell.13 Glycophorin A is expressed on human red blood cells, normoblasts, and erythroid precursor cells.14 Mature non-nucleated red blood cells are characteristically glycophorin A positive but CD45 negative.15
* Cy™ is a trademark of GE Healthcare. This product is subject to proprietary rights of GE Healthcare and Carnegie Mellon University, and is made and sold under license from GE Healthcare. This product is licensed for sale only for in vitro diagnostics. It is not licensed for any other use. If you require any additional license to use this product and do not have one, return this material, unopened, to BD Biosciences, 2350 Qume Drive, San Jose, CA 95131, and any money paid for the material will be refunded.
23-7739-03-en.book Page 6 Wednesday, March 25, 2015 5:43 PM
7
CD41a recognizes the calcium-dependent gpIIb/IIIa complex, with a molecular weight of 140 kDa.16 This complex is found on platelets and platelet precursors. It acts as a receptor for fibrinogen, von Willebrand factor (vWf), fibronectin, and vitronectin, and it mediates platelet adhesion and aggregation.17-19 The CD41a antibody is composed of mouse IgG1 heavy chains and kappa light chains.
Precautions
• For In Vitro Diagnostic Use.• The addition of a precise volume of sample is critical. Pipets must
be calibrated to deliver exactly 25 µL and 50 µL of sample. If necessary, perform the reverse pipetting technique according to the pipet manufacturer’s instructions.
• Care should be taken to avoid microbial contamination, which could cause aberrant results.
• Do not use the reagents if you observe any change in appearance. Precipitation or discoloration indicates instability or deterioration.
• Before use, inspect the BD Trucount tube to make sure the pellet is intact and below the retainer.
• Bead count varies by lot of BD Trucount tubes. It is critical to use the bead count shown on the lot of BD Trucount tubes that you are currently using when calculating absolute cell counts. Do not mix multiple lots of tubes in the same assay.
WARNING All biological specimens and materials coming in contact with them are considered biohazards. Handle as if capable of transmitting infection20,21 and dispose of with proper precautions in accordance with federal, state, and local regulations. Never pipette by mouth. Wear suitable protective clothing, eyewear, and gloves.
Storage and Handling
• Store the BD Plasma Count kit at 2°C–8°C. Do not use after the expiration date shown on the label.
• Avoid unnecessary exposure of the reagent to light.
23-7739-03-en.book Page 7 Wednesday, March 25, 2015 5:43 PM
8
• Do not freeze the reagents or expose them to direct light during storage or incubation with cells. Keep the reagent vials dry.
• Store BD Trucount tubes in their original foil pouch at 2°C–25°C. To avoid potential condensation, open the pouch only after it has reached room temperature and carefully reseal the pouch immediately after removing a tube. Examine the desiccant each time you open the pouch. If the desiccant has turned from blue to lavender, discard the remaining tubes. Use tubes within 1 hour after removal from the foil pouch. Once the pouch has been opened, the tubes are stable for 1 month. Do not use beyond the expiration date indicated on the packaging.
5. INSTRUMENT
The BD Plasma Count kit is designed for use on a flow cytometer equipped with appropriate computer hardware and software. The cytometer must have a 488-nm laser capable of detecting forward scatter (FSC) and side scatter (SSC) and three-color fluorescence, and must be able to threshold or discriminate using the FSC channel.
You can use a BD FACS™ brand cytometer, such as the BD FACSCalibur flow cytometer running BD FACSComp software (version 4.0 or later), with the optional BD FACS™ Loader.
We recommend using BD Calibrite™ beads to set photomultiplier tube (PMT) voltages, fluorescence compensation, and to check instrument sensitivity before use.
You can also use the BD FACSVia flow cytometer with the optional BD FACSVia™ Loader. For this cytometer, use BD™ CS&T beads for instrument QC.
Results can be achieved using other platforms. For flow cytometers manufactured by companies other than BD Biosciences, set up the cytometer for three-color acquisition following the manufacturer’s recommendations.
23-7739-03-en.book Page 8 Wednesday, March 25, 2015 5:43 PM
9
6. SPECIMEN COLLECTION AND PREPARATION
A minimum of 200 µL of plasma is required for this procedure. For residual cell enumeration, use plasma samples with additive, or fresh plasma samples with no additive within 18 hours of collection. Store plasma samples with no additive at room temperature (20°C– 25°C) under constant agitation until ready for staining.
Interfering Conditions
• Do not use previously fixed and stored plasma.• Samples without additive that are refrigerated before staining
could give aberrant results.• Hemolyzed, lipemic, or clotted samples should be rejected.• RBC agglutination in stained samples increases during storage.
Therefore, acquire samples within 1 hour of staining.
7. PROCEDURE
Reagents Provided
• Reagent Vial A • Reagent Vial B• BD Trucount tubes
Reagents and Materials Required but Not Provided
• Falcon®† disposable 12 x 75-mm polypropylene test tubes or equivalent
• Micropipettor with tips• Vortex mixer and roller mixer• BD FACS™ Clean solution (Catalog No. 340345)
For BD FACS brand cytometers running BD FACSComp software:
• BD Calibrite beads (Catalog No. 349502)• BD Calibrite PerCP-Cy5.5 beads (Catalog No. 345036)
† Falcon is a registered trademark of Corning Incorporated.
23-7739-03-en.book Page 9 Wednesday, March 25, 2015 5:43 PM
10
• BD Calibrite™ 3 beads (Catalog No. 340486)• BD FACSFlow™ sheath fluid (Catalog No. 342003) or equivalent
CAUTION Use only BD FACSFlow sheath fluid diluent to dilute BD Calibrite beads.
• Wash buffer of phosphate-buffered saline (PBS) with 0.1% sodium azide
• BD FACS™ Rinse solution (Catalog No. 340346)
For BD FACSVia cytometers:
• BD CS&T beads (Catalog Nos. 656504, 656505)
Staining Procedure
Before beginning the staining procedure, do the following:
• Allow Reagent Vial A to come to room temperature.• Filter the wash buffer (PBS) through a 0.2-µm filter to avoid
increased background signals in the stained plasma sample.
CAUTION Pipette the sample and reagents onto the side of each tube just above the stainless steel retainer. If the sample remains on the side of the tube, it will not be stained with the reagent. Do not touch the bead pellet.
1. For each blood component sample, dispense a minimum of 200 µL of fresh plasma into labeled 12 x 75-mm Falcon tubes.
NOTE Use a new tip for each sample.
2. Remove a BD Trucount tube from the foil pouch, and label the tube.
NOTE Before use, verify that the bead pellet is intact and under the metal retainer at the bottom of each tube. If not, discard the tube and replace it with another.
23-7739-03-en.book Page 10 Wednesday, March 25, 2015 5:43 PM
11
3. Using a calibrated pipet, transfer the appropriate amount of blood component sample (25 or 50 µL [Table 1]) into each labeled BD Trucount tube.
NOTE Counting precision (expressed as the coefficient of variation [CV]) is determined by the number of events counted. Counting fewer events results in a higher CV. In rare event analysis, the CV is influenced by sample concentration and sample volume. Thus, analyzing a higher cell concentration and counting more cells results in a lower CV.22 We recommend that you consider the precision required for your application when choosing a sample volume (Table 1).
4. Add 100 µL of Reagent A to each BD Trucount tube.
5. Add 20 µL of Reagent B to each BD Trucount tube.
6. Cap the tubes and vortex gently to mix.
7. Incubate the tubes for 15 minutes in the dark at room temperature.
8. Add 1 mL of filtered wash buffer (PBS) to each tube to stop the staining process.
9. Cap the tubes and vortex gently.
The cells are now ready to be analyzed on the flow cytometer. Protect stained samples from direct light.
Acquire samples within 1 hour after preparation. If stained samples are not acquired immediately after preparation, store samples in the dark at 2°C–8°C for up to 1 hour. Vortex thoroughly just before acquisition.
Table 1 Volume of sample used to enumerate different ranges of rWBCs in plasma
0.5–50 rWBC/µL
>50–500 rWBC/µL
Sample Volume 50 µL 25 µL
23-7739-03-en.book Page 11 Wednesday, March 25, 2015 5:43 PM
12
If stained samples are to be run on the BD FACS Loader or BD FACSVia Loader, vortex each tube immediately before placing into the BD FACS Loader carousel or BD FACSVia Loader arm.
NOTE Do not load more tubes on the rack than can be run within 1 hour.
BD FACSVia Instrument Quality Control
Instrument QC checks instrument performance, updates compensation values accordingly, and assesses the background instrument noise.
We recommend running a control sample daily to optimize instrument settings and as a quality control check of the system.23
See the cytometer’s Instructions For Use (IFU) for instructions on how to run instrument QC.
BD FACS Brand Instrument Setup
This section provides guidelines for instrument setup.
Setup is not required for the BD FACSVia flow cytometer.
Preparing for Instrument Setup
Before running stained plasma samples on the flow cytometer, do the following:
• Clean the instrument’s flow cell with BD FACS Clean solution followed by BD FACS Rinse solution. Refer to your cytometer IFU.
• Check the noise level of the BD FACSFlow sheath fluid. The overall noise level should not exceed 30–45 events/second when FSC mode is set to Log and FSC threshold is set between 100–150 channels. If the noise level exceeds this recommendation, use a new cubitainer of sheath fluid. If the problem persists, change the sheath filter.
23-7739-03-en.book Page 12 Wednesday, March 25, 2015 5:43 PM
13
Setting Up the Instrument
For BD FACS brand cytometers running BD FACSComp software, we recommend using BD Calibrite beads (FITC, PE, and PerCP-Cy5.5) with BD FACSComp software using the three-color Lyse/No-Wash (LNW) setup. See the BD Calibrite Beads IFU and the BD FACSComp Software User’s Guide for instructions.
For flow cytometers manufactured by companies other than BD, use either BD Calibrite beads and PerCP-Cy5.5 beads or BD Calibrite 3 beads to adjust compensation. Ensure the unlabeled beads are in the lower-left quadrant of the two-dimensional fluorescence plots.
Acquiring Samples
For sample acquisition instructions using the BD FACSVia flow cytometer, see the BD FACSVia System IFU.
For BD FACS brand cytometers running BD FACSComp software, we recommend using BD CellQuest™ Pro software for residual cell enumeration.
NOTE To ensure thorough resuspension of cells and beads, gently vortex samples immediately before acquisition.
1. In BD CellQuest Pro software, create an Experiment document containing five Acquisition-to-Analysis dot plots with the following parameters: FSC vs SSC, FL1 vs FL2, FL2 vs FL3, FSC vs FL3, FL1 vs FL3.
2. After using BD FACSComp software to set up your cytometer, load the CalibFile.LNW file and modify:
• Set FSC (P1) and SSC (P2) to Log mode.• Set FSC threshold to channel 100–150.
3. Vortex the first sample and place it on the cytometer.
4. Set the fluid control to HI and begin acquiring the sample in Setup mode.
23-7739-03-en.book Page 13 Wednesday, March 25, 2015 5:43 PM
14
If you are using the BD FACS Loader to acquire samples, see the BD FACS Loader User’s Guide for complete instructions.
5. Adjust instrument settings, if necessary.
6. Stop acquisition, and save the optimized instrument settings.
7. Draw the following regions around the bead population in the corresponding plots.
See Figure 1 for R1, R2, and R3 locations.
8. In the Gate List, define the bead gate as R1 and R2 and R3 and change the label from G4 to Beads.
9. Change the following settings in the Acquisition & Storage window:
• Select Beads from the Acquisition stops when 10,000 events are acquired menu.
• Verify that instrument resolution is set to 1,024 and the P1, P2, P3, P4, and P5 checkboxes are selected.
10. Acquire samples, saving the data.
Data Analysis
This section provides data analysis instructions for BD FACS brand cytometers running BD FACSComp software. For data analysis instructions using the BD FACSVia flow cytometer, see the BD FACSVia Plasma Count Assay Guide.
Plot Region
FL1 vs FL2 R1
FL2 vs FL3 R2
FL1 vs FL3 R3
23-7739-03-en.book Page 14 Wednesday, March 25, 2015 5:43 PM
15
Setting rWBC Gates
1. Draw the following regions around the WBC population in the corresponding plots (Figure 1).
2. In the Gate List, change the G9 label to rWBC and combine the following regions:
rWBC = R4 and R5 and R6 and R7
3. Select the Hilite checkbox for rWBC.
Setting rRBC Gates
4. Draw the following regions around the RBC population in the corresponding plots (Figure 1).
5. In the Gate List, change the G12 label to rRBC and combine the following regions:
rRBC = R8 and R9
NOTE All Glycophorin A-FITC positive RBCs appear in R9. Morphologically altered RBCs do not appear in R8, and are therefore excluded from rRBC calculation by the rRBC gate definition.5,7,24
Plot Region
FSC vs SSC R4
FL1 vs FL2 R5
FL2 vs FL3 R6
FL1 vs FL3 R7
Plot Region
FSC vs SSC R8
FL1 vs FL2 R9
23-7739-03-en.book Page 15 Wednesday, March 25, 2015 5:43 PM
16
Setting rPLT Gates
6. Draw the following regions around the PLT population in the corresponding plots (Figure 1).
7. In the Gate List, change the G15 label to rPLT and combine the following regions:
rPLT = R10 and R11
Viewing Analysis Plots
8. Arrange the gates in the Gate List in descending order: Beads, rWBC, rRBC, rPLT.
The analysis plots should look similar to those shown in Figure 1.
9. Display Region and Gate statistics for the sample data.
10. Calculate absolute counts. (See Calculating Absolute Counts in the Results section.)
Plot Region
FSC vs FL3 R10
FL1 vs FL3 R11
23-7739-03-en.book Page 16 Wednesday, March 25, 2015 5:43 PM
17
Figure 1 Representative plots showing gating strategy for BD Plasma Count assay
23-7739-03-en.book Page 17 Wednesday, March 25, 2015 5:43 PM
18
8. RESULTS
Calculating Absolute Counts
NOTE Results are automatically calculated for the BD FACSVia flow cytometer.
23-7739-03-en.book Page 18 Wednesday, March 25, 2015 5:43 PM
19
For other instruments, the absolute number of residual cells is determined by dividing the number of residual cell events acquired by the number of fluorescent bead events acquired, and multiplying the result by the bead concentration.
Use the following equation to calculate the absolute cell count of the residual cell population.
(GX/G4) x (BPT/V) = residual cells/µL, where:
• GX = the number of residual cells in the rWBC, rRBC, or rPLT gate
• G4 = the number of beads in the G4 gate• BPT = beads/tube, which is found on the BD Trucount foil pouch
and can vary from lot to lot• V = sample volume
Figure 2 shows the statistics view for residual cell analysis of a plasma sample and rWBC, rRBC, and rPLT calculations.
Figure 2 Statistics view and resulting calculation for rWBC, rRBC, and rPLT analysis of a plasma sample
(140/10,000) x (48,554/25 µL) = 27 rWBC/µL plasma
23-7739-03-en.book Page 19 Wednesday, March 25, 2015 5:43 PM
20
(2,086/10,000) x (48,554/25 µL) = 405 rWBC/µL plasma
(129,720/10,000) x (48,554/25 µL) = 25,194 rWBC/µL plasma
9. LIMITATIONS
• Performance has not been established on stream-in-air cytometer systems.
• The addition of a precise volume of blood is critical when BD Trucount tubes are used. The pipet used should be calibrated to deliver 25 µL or 50 µL of sample.
• Gently vortex samples immediately prior to running them on the flow cytometer to ensure thorough resuspension of cells and beads.
• Nucleated red cells contain nucleic acid and could be detected as rWBCs in this assay. However, nucleated red cells are not present in detectable quantities in blood from normal individuals. Their expected frequency in fresh plasma will be even lower and thus less interfering.6
• As for any assay that involves counting statistics, the number of events counted could influence the result.23 We recommend collecting at least 100,000 total events when analyzing leucoreduced plasma, or at least 10,000 BD Trucount beads for leucocyte-depleted plasma.
• Samples containing >38,000 rPLTs have not been validated with this assay.
23-7739-03-en.book Page 20 Wednesday, March 25, 2015 5:43 PM
21
10. PERFORMANCE CHARACTERISTICS
Precision for BD FACSCalibur Flow Cytometers
Based on combined dilution series, following ICH and CLSI guidelines, five different concentration levels were tested in triplicate on five different days.25,26 Target values for the five concentration levels were obtained from dilutions of one cell stock solution that was measured on a Sysmex®‡ F-820 automatic cell counter. Purified WBC, RBC, and PLT samples were spiked into cell-free plasma. Cell-free plasma was used as blank. Fresh samples were prepared on each of the five days.
Prepared samples were stained according to the Cell Staining procedure in the Procedure section. See Table 2 for results and Figure 3 for mean values.
‡ Sysmex is a registered trademark of Sysmex Corporation
Table 2 Precision of BD Plasma Count assay for each concentration level (n=15)
Cell Type Level (cells/µL) Mean (cells/µL) SDa (cells/µL) %CVb
WBC 50 51 3.9 7.5
275 291 15.0 5.2
500 508 43.8 8.6
725 758 40.5 5.3
950 967 58.6 6.1
RBC 600 606 55.6 9.2
3,300 2,892 306.8 10.6
6,000 4,737 572.5 12.1
23-7739-03-en.book Page 21 Wednesday, March 25, 2015 5:43 PM
22
The following table is a legend for all plots of Figure 3.
PLT 2,000 1,995 215.7 10.8
11,000 10,858 952.8 8.8
20,000 19,566 1,807.7 9.2
29,000 28,446 2,425.6 8.5
38,000 36,546 3,143.6 8.6
a. SD = standard deviationb. CV = coefficient of variation
Item Description
• Mean value (n=15)
— Linear regression of mean value
- - - Expected value
Table 2 Precision of BD Plasma Count assay for each concentration level (n=15)
Cell Type Level (cells/µL) Mean (cells/µL) SDa (cells/µL) %CVb
23-7739-03-en.book Page 22 Wednesday, March 25, 2015 5:43 PM
23
Figure 3 Linearity of BD Plasma Count assay (Mean values for each cell type at the five different levels are taken from Table 2.)
Statistic Value
r2 0.999
Slope 1.023
Intercept 3.26
23-7739-03-en.book Page 23 Wednesday, March 25, 2015 5:43 PM
24
Statistic Value
r2 0.996
Slope 0.789
Intercept 106.41
23-7739-03-en.book Page 24 Wednesday, March 25, 2015 5:43 PM
25
Low Range Study
In a separate study, accuracy of WBC and RBC low range results from the BD Plasma Count assay were compared to estimated cell counts obtained from dilutions of one cell stock solution that was measured on a Sysmex F-820 cell counter. Samples were prepared, diluted, and analyzed using the same methods as in the general comparison study.
Statistic Value
r2 0.999
Slope 0.966
Intercept 146.81
23-7739-03-en.book Page 25 Wednesday, March 25, 2015 5:43 PM
26
See Table 3 and Table 4 for results and Figure 4 for plots.Table 3 Precision of BD Plasma Count assay for detection of WBCs and RBCs in the
low cell range (n=15)
Cell Type Level (cells/µL) Mean (cells/µL) SDa (cells/µL)
a. SD = standard deviation
%CVb
b. CV = coefficient of variation
WBC 0.6 0.7 0.28 40.4
1.95 2.4 0.35 14.5
3.30 3.7 0.69 18.8
4.65 5.7 0.93 16.3
6.00 6.0 1.13 18.7
RBC 600 597 48.80 8.2
1,950 1,625 155.14 9.6
3,300 2,707 203.46 7.5
4,650 3,215 226.27 7.0
6,000 4,472 563.16 12.6
Table 4 Accuracy of BD Plasma Count assay vs expected low cell rangea of each cell type (n=90)
a. Concentration levels were obtained from dilutions of one cell stock solution that was measured on a Sysmex F-820 cell counter.
Cell Type Range (cells/µL) Slope Intercept Correlation (r2)
WBC 0–6.0 1.052 0.199 0.979
RBC 0–6,000 0.716 133.47 0.990
23-7739-03-en.book Page 26 Wednesday, March 25, 2015 5:43 PM
27
The following table is a legend for all plots of Figure 4.
Figure 4 Linearity of BD Plasma Count assay vs expected low cell range (mean values for the five different concentration levels of each cell type are taken from Table 3).
Item Description
• Mean value (n=15)
— Linear regression
- - - Expected value
Statistic Value
r2 0.979
Slope 1.052
Intercept 0.199
23-7739-03-en.book Page 27 Wednesday, March 25, 2015 5:43 PM
28
Linearity for BD FACSCalibur Flow Cytometers
Based on combined dilution series, following ICH and CLSI guidelines, five different concentration levels were tested in triplicate on five different days.25,26 Target values for the five concentration levels were obtained from dilutions of one cell stock solution that was measured on a Sysmex F-820 automatic cell counter. Purified WBC, RBC, and PLT samples were spiked into cell-free plasma. Cell-free plasma was used as blank. Fresh samples were prepared on each of the five days. Prepared samples were stained according to the Cell Staining procedure described in the Procedure section. See Figure 4 for linearity results.
NOTE The rRBC values shown indicate a gradual agglutination effect with increasing concentration, leading to the underestimation of rRBCs at concentrations over 3,000 cells/µL. See statistics in Table 2.
Statistic Value
r2 0.990
Slope 0.716
Intercept 133.47
23-7739-03-en.book Page 28 Wednesday, March 25, 2015 5:43 PM
29
Stability for BD FACSCalibur Flow Cytometers
Samples stained with BD Plasma Count reagents for the enumeration of the three cell types in FFP were stable for up to 1 hour after staining when stored at 4°C in the dark.
Accuracy for BD FACSCalibur Flow Cytometers
General Comparison vs Expected Values
In one study, the cell concentration results from the BD Plasma Count assay were compared for accuracy with expected values of red cell, white cell, and platelet counting.5 The samples were obtained from dilutions of one stock solution that was measured on the blood cell counter (Sysmex F-820 model). The stock solution was prepared from spiked RBCs, WBCs, and PLTs into cell-free plasma. The mean value results are shown in Table 5.
Both WBC and PLT results from the flow cytometer compare directly to the expected values. The RBC values shown indicate a gradual agglutination effect with increasing concentration, leading to the underestimation of RBCs at concentrations over 3,300 cells/µL (bias >10%).5
Table 5 Accuracy of BD Plasma Count assay vs expected cell rangea of each cell type (n=90)
a. Concentration levels were obtained from dilutions of one cell stock solution that was measured on a Sysmex F-820 cell counter
Cell Type Range (cells/µL) Slope Intercept Correlation (r2)
WBC 0–950 1.023 3.26 0.999
RBC 0–6,000 0.789 106.41 0.996
PLT 0–38,000 0.966 146.81 0.999
23-7739-03-en.book Page 29 Wednesday, March 25, 2015 5:43 PM
30
BD Plasma Count Assay vs Microscopic Reading
The BD Plasma Count assay and the Fuchs-Rosenthal assay were compared for accuracy of residual red blood cell enumeration. A total of 79 samples of healthy donors were stained according to the Cell Staining procedure described in the Procedure section. Absolute rRBC counts were determined by manual counting on the microscope using a Fuchs-Rosenthal counting chamber and by flow cytometry. Study results show excellent comparison between the flow cytometric determination and the microscopy method. Results are shown in Figure 5.
The following table is a legend for all plots of Figure 5.
Item Description
+ Value (n=79)
— Linear regression
- - - Expected value
23-7739-03-en.book Page 30 Wednesday, March 25, 2015 5:43 PM
31
Figure 5 Accuracy of BD Plasma Count assay vs Fuchs-Rosenthal manual counting method
Precision for BD FACSVia Flow Cytometers
The system precision study for the enumeration of rWBC, rPLT, and rRBC absolute counts was determined across two BD FACSVia flow cytometers (one manual and one with an automated loader). Two operators acquired two concentration levels of Plasma Count test samples stained in duplicate using the BD Plasma Count Kit and one donor per day over five days.
Precision study samples were created at two concentration levels (high and low) by spiking concentrated RBCs, WBCs, and PLTs from autologous whole blood into cell-free plasma. Concentration levels (WBC/µL, RBC/µL, and PLT/µL) of the test samples were established using the BD FACSCalibur flow cytometer.
Statistic Value
r2 0.960
Slope 0.927
Intercept 41.65
23-7739-03-en.book Page 31 Wednesday, March 25, 2015 5:43 PM
32
NOTE The precision measurement includes the following variance components: Instrument, Operator, Between Run, Within Run.
Linearity for BD FACSVia Flow Cytometers
System linearity of the BD FACSVia flow cytometer when using BD Plasma Count Kit was determined using triplicate measurements of 11 equally spaced concentrations of autologous rWBCs, rPLTs, and rRBCs spiked into cell-free plasma. Linearity was assessed with the following ranges:
Table 6 Precision for BD FACSVia flow cytometers
Cell Type Mean SD CV
WBC Low 102.1 7.0 6.9%
WBC High 286.5 17.6 6.2%
RBC Low 1705.5 76.4 4.5%
RBC High 9880.5 816.0 8.3%
PLT Low 10,329.3 550.0 5.3%
PLT High 75,909.5 3323.1 4.4%
Table 7 Linearity for BD FACSVia flow cytometers
Cell Type Range
WBC 4–1,133
RBC 5–8,608
PLT 172–134,507
23-7739-03-en.book Page 32 Wednesday, March 25, 2015 5:43 PM
33
Accuracy for BD FACSVia Flow Cytometers
Deming regression analysis was performed for the BD FACSVia manual and BD FACSCalibur manual. Plasma samples were prepared by centrifugation or apheresis method and autologous rWBC, rPLT, and rRBC subsets were added to achieve various concentrations for measurement.
Cell concentrations of test samples were established on the BD FACSCalibur prior to inclusion in the BD FACSVia accuracy study.
Figure 6 Accuracy of the BD Plasma Count kit on BD FACSVia vs BD FACSCalibur for rWBC in plasma
Variable N R2 Slope Intercept
rWBC 86 0.99 0.98 (0.97, 1.0) 1.6 (-1.8, 5.0)
1000
900
800
700
600
500
400
300
200
100
0
BD
FA
CS
Via
rWB
C (c
ells
/µL)
BD FACSCalibur rWBC (cells/µL)
0 100 200 300 400 500 600 700 800 900 1000
23-7739-03-en.book Page 33 Wednesday, March 25, 2015 5:43 PM
34
Figure 7 Accuracy of the BD Plasma Count kit on BD FACSVia vs BD FACSCalibur for rRBC in plasma
Variable N R2 Slope Intercept
rRBC 86 0.99 0.98 (0.96, 1.01) -26.3 , 38.3
11,000
10,000
9000
8000
7000
6000
5000
4000
3000
2000
1000
0
BD
FA
CS
Via
rRB
C (c
ells
/µL)
BD FACSCalibur rRBC (cells/µL)
0 1000 2000 3000 4000 5000 6000 7000 8000 9000
23-7739-03-en.book Page 34 Wednesday, March 25, 2015 5:43 PM
35
Figure 8 Accuracy of the BD Plasma Count kit on BD FACSVia vs BD FACSCalibur for rPLT in plasma
WARRANTY
Unless otherwise indicated in any applicable BD general conditions of sale for non-US customers, the following warranty applies to the purchase of these products.
THE PRODUCTS SOLD HEREUNDER ARE WARRANTED ONLY TO CONFORM TO THE QUANTITY AND CONTENTS STATED ON THE LABEL OR IN THE PRODUCT LABELING AT THE TIME OF DELIVERY TO THE CUSTOMER. BD DISCLAIMS HEREBY ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY AND FITNESS FOR ANY PARTICULAR PURPOSE AND NONINFRINGEMENT. BD’S SOLE LIABILITY IS LIMITED TO EITHER REPLACEMENT OF
Variable N R2 Slope Intercept
rRBC 86 1.0 0.97 (0.95, 0.98) 45.8 (-329.9, 421.6)
110,000
100,000
90,000
80,000
70,000
60,000
50,000
40,000
30,000
20,000
10,000
0
BD
FA
CS
Via
rPLT
(cel
ls/µ
L)
BD FACSCalibur rPLT (cells/µL)
0 20,000 40,000 60,000 80,000 100,000 120,000
23-7739-03-en.book Page 35 Wednesday, March 25, 2015 5:43 PM
36
THE PRODUCTS OR REFUND OF THE PURCHASE PRICE. BD IS NOT LIABLE FOR PROPERTY DAMAGE OR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES, INCLUDING PERSONAL INJURY, OR ECONOMIC LOSS, CAUSED BY THE PRODUCT.
TROUBLESHOOTING
REFERENCES1 Guide to the preparation, use and quality assurance of blood components. (10th ed.)
Council of Europe Publishing; 2004:260.
2 Guidelines for the production of blood and blood components and for the use of blood products (Haemotherapie). Bundesgesundheitsbl-Gesundheitsforsch-Gesundheitsschutz; 2000;43:555-589.
3 Heiden M, Seitz R. Quality of therapeutic plasma-requirements for marketing authorization. Thromb Res. 2002;107 Suppl 1:47-51.
Problem Possible Cause Solution
Poor resolution between debris and residual cells
Inappropriate instrument settings
Follow proper instrument setup procedures; optimize instrument settings as required.
Staining dim or fading, or no residual cells
Reagents out of specification
Store BD Plasma Count reagents at 2°C–8°C and do not use after expiration date; repeat staining using new reagent.
Insufficient reagent Repeat staining with fresh sample; use recommended volumes; properly mix sample with reagents.
Cells not analyzed within 1 hour of staining
Repeat staining with fresh sample; analyze promptly or store samples in the dark at 4°C up to 1 hour.
Improper medium preparation (sodium azide omitted)
Use sodium azide in PBS if used as sample dilution medium.
Few or no BD Trucount beads detected
Bead pellet damaged or not present
Before use, inspect BD Trucount tube; use tubes within 1 hour after removal from foil pouch; do not use beyond expiration date.
Cytometer malfunctioning
Troubleshoot instrument.
23-7739-03-en.book Page 36 Wednesday, March 25, 2015 5:43 PM
37
4 Vowells S, Cadden M, Wagner C, Nuñez R. Detection and Enumeration of Residual White Blood Cells in Leucoreduced Red Blood Cell and Platelet Products Using the LeucoCOUNT Kit. Becton, Dickinson, and Company; 1998.
5 Lambrecht B, Spengler HP, Bauerfeind U, Mohr H. Simultaneous quantitation of contaminating leukocytes, erythrocytes, and platelets in fresh frozen plasma in a single tube assay by flow cytometry. Infusion Therapy and Transfusion Medicine. 2001;28 Suppl 1:52-53.
6 Jilma-Stohlawetz P, Marsik C, Horvath M, Siegmeth H, Hocker P, Jilma B. A new flow cytometric method for simultaneous measurement of residual platelets and RBCs in plasma: validation and application for QC. Transfusion. 2001;41:87-92.
7 Krailadsiri P, Seghatchian J. Residual red cell and platelet content in WBC-reduced plasma measured by a novel flow cytometric method. Transfusion and Apheresis Science. 2001;24:279-286.
8 Frey B, Furrer M, Wettstein M, Minder U. Quantification of residual red blood cells in platelet concentrates and fresh frozen plasma by flow cytometry. Infusion Therapy and Transfusion Medicine. 2001;28 Suppl 1:53.
9 Pichler J, Printz D, Scharner D, Trbojevic D, Siekmann J, Fritsch G. Improved flow cytometric method to enumerate residual cells: minimal linear detection limits for platelets, erythrocytes, and leukocytes. Cytometry. 2002;50:231-237.
10 Lee LG, Chen CH, Chiu LA. Thiazole orange: a new dye for reticulocyte analysis. Cytometry. 1986;7:508-517.
11 Terstappen LW, Loken MR. Five-dimensional flow cytometry as a new approach for blood and bone marrow differentials. Cytometry. 1988;9(6):548-556.
12 Gahmberg CG, Jokinen M, Andersson LC. Expression of the major sialoglycoprotein (glycophorin) on erythroid cells in human bone marrow. Blood. 1978;52:379-387.
13 Wise GE, Oakford LX, Dzandu JK. Ultrastructure of a transmembrane glycoprotein, glycophorin A. Tissue & Cell. 1988;20:219-227.
14 Rogers CE, Bradley MS, Palsson BO, Koller MR. Flow cytometric analysis of human bone marrow perfusion cultures: erythroid development and relationship with burst-forming units-erythroid. Exp Hematol. 1996;24:597-604.
15 de Vries E, De Bruin-Versteeg S, Comans-Bitter WM, et al. Correction for erythroid cell contamination in microassay for immunophenotyping of neonatal lymphocytes. Arch Dis Child Fetal Neonatal Ed. 1999;80:226-229.
16 von dem Borne AE, Modderman PW, Admiraal LG, Nieuwenhuis HK. Platelet antibodies, the overall results. In: Knapp W, Dörken B, Gilks WR, et al, eds. Leucocyte Typing IV: White Cell Differentiation Antigens. New York, NY: Oxford University Press; 1989:951-977.
17 de Haas M, von dem Borne A. CD41/CD61 Workshop Panel report. In: Kishimoto T, Kikutani H, von dem Borne, et al, eds. Leucocyte Typing VI: White Cell Differentiation Antigens. New York, NY: Garland Publishing, Inc.; 1997:643-645.
23-7739-03-en.book Page 37 Wednesday, March 25, 2015 5:43 PM
38
18 Feng R, Shimazaki C, Inaba T, et al. CD34+/CD41a+ cells best predict platelet recovery after autologous peripheral blood stem cell transplantation. Bone Marrow Transplant. 1998;21:1217-1222.
19 Jennings LK, Ashmun RA, Wang WC, Dockter ME. Analysis of human platelet glycoproteins IIb-IIIa and Glanzmann's thrombasthenia in whole blood by flow cytometry. Blood. 1986;68:173-179.
20 Protection of Laboratory Workers from Occupationally Acquired Infections; Approved Guideline —Third Edition. Wayne, PA: Clinical and Laboratory Standards Institute; 2005. CLSI document M29-A3.
21 Centers for Disease Control. Perspectives in disease prevention and health promotion update: universal precautions for prevention of transmission of human immunodeficiency virus, hepatitis B virus, and other bloodborne pathogens in health-care settings. MMWR. 1988;37:377-388.
22 Dzik S. Counting low numbers of leukocytes in leukoreduced blood components. Infusion Therapy and Transfusion Medicine. 1999;26:62-65.
23 Enumeration of Immunologically Defined Cell Populations by Flow Cytometry; Approved Guideline—Second Edition. Wayne, PA: Clinical and Laboratory Standards Institute; 2007. CLSI document H42-A2.
24 Linneweber J, Chow TW, Takano T, et al. Direct detection of red blood cell fragments: a new flow cytometric method to evaluate hemolysis in blood pumps. Asaio J. 2001;47:533-536.
25 Preliminary Evaluation of Quantitative Clinical Laboratory Measurement Procedures; Approved Guideline—Third Edition. Wayne, PA: Clinical Laboratory Standards Institute; 2002. CLSI document EP10-A3 AMD.
26 ICH Q2(R1). Validation of Analytical Procedures: Text and Methodology. 1996.
23-7739-03-en.book Page 38 Wednesday, March 25, 2015 5:43 PM