76
Brachytherapy Dosimetry in the Clinic Margaret Bidmead Royal Marsden NHS Trust

Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

  • Upload
    hanga

  • View
    220

  • Download
    2

Embed Size (px)

Citation preview

Page 1: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Brachytherapy

Dosimetry in the

ClinicMargaret Bidmead

Royal Marsden NHS Trust

Page 2: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Two Curies

Bequerel

Page 3: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Brachytherapy

What is it?

“Sealed radioactive source(s) placed directly

into, or immediately adjacent to the volume

to be treated.”

Treatment of melanoma, 1910, Sydney, Australia

Page 4: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Uses for Radium

1904

“Guaranteed to

Contain Real

Refined Radium

And to be

Perfectly

Harmless”

Self-

luminous

paint-

stopped

in 1950’s

Radium Girls-

Occupational disease

Labour law

Page 5: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Brachytherapy Statistics:

4 % of all new cancer cases receive brachytherapy (12% in USA)

Gynae - 68%

Head & Neck - 5%

Bronchus - 7%

Rectum - 3%

prostate-25%

Page 6: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Types of Brachytherapy

IntracavitaryCervix

vagina

oesophagus/bronchus

Page 7: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Why do we use it?

localised treatment, i.e. high dose to the target

volume, whilst sparing adjacent normal tissue, and

minimal whole body dose

Page 8: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Types of BrachytherapyInterstitial Prostate

Head and neck

Breast

Rectum

Page 9: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Types of Brachytherapy

Surface Moulds

Angiosarcoma

Melanoma

Basal Cell carcinoma

Recurrent disease

Page 10: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Implantation Rules

System of work for source implantation

How best to cover defined volume

How best to minimize dose to organs at

risk

What is practically achievable

What AKR of source(s) available

Page 11: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Clinical Dosimetry Systems

Intracavitary ……Historically Modified

Manchester System, Point A: now volumetric

Interstitial with Ir….Paris System

Line sources…..prescription at distance from a

plane

Permanent seeds……volumetric isodose

coverage (D90)

Moulds…….different Manchester System,

treatment at a specified distance

Page 12: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Dosimetry Overview

Purpose of source calibration and dosimetry?

To verify source strength independently and improve precision.

To achieve national and international agreement

(DOH requirement: check measured source strength against manufacturer’s certificate)

Source specification:

AKR in air @ 1m - RAKR

Page 13: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Brachytherapy Source Specification

Reference Quantity Definition Units

Reference

Air Kerma

Rate

(RAKR)

Air kerma rate due to

photons of energy

greater than , at a

point located at a

distance of 1m from

the source centre, in

vacuo.

Gyhr-1

at 1m

AAPM TG 43

Update:

Med. Phys. 31

(3), March

2004

Air-kerma

strength

Sk

The air kerma rate,

in vacuo, and due to

photons of energy

greater than , at

distance d,

multiplied by the

square of the

distance.

Gym2hr-1

(U)

Page 14: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Source Strength

Past:

Activity

Effective

Activity

mg Ra

equivalent

Present:

Air Kerma Rate

Under standard

conditions

NB: Link to algorithm

in TPS or MC model

of each source

Uncertainties in conversion factors

Page 15: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Reference Standards & Traceability

Direct calibration method:

source or calibrator has been calibrated at a national standards laboratory (NPL in the UK, NIST in the USA).

Secondary method:

source is calibrated in comparison with a source of the same design and comparable strength, which has direct traceability.

Page 16: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Reference Standards & Traceability

Remote traceability : if the user relies on the manufacturer’s calibration as the only standard. (not recommended)

Page 17: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Calibration Methods

Ion chamber in air: AKR at known

distance from the source

Calibrated re-entrant well chamber

Ion chamber and source inside a

solid phantom. (2nd check?)

Page 18: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Linear response with source strength.

Accurately measured energy response.

Response dependent on geometric

configuration (size, shape, orientation

and position), filtration

and encapsulation of

source.

Properties of Re-entrant Well

Chambers

Page 19: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Properties of Re-entrant Well

Chambers

Reproducibility better than 2%.

Proven long-term stability (+-1%)

Signal-to-noise ratio > 100:1.

Collection efficiency better than

99%.

Easy to set up and use

Page 20: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Issues With Well Chambers

Effect of source type used for standard calibration

Sweet spot of chamber

Saturation effects of well chambers at high currents

Temperature stabilisation

Method of constancy check

Second, independent check of source calibration

Page 21: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

HDR/PDR Afterloading Systems

Nucletron HDR

Micro Selectron

Ir192 (370Gbq)

Ir192 (370GBq)

Co60 (74GBq)

source availableFlexitron

Dual source

Ir192

(Co60/Yb169)

Varian / GammaMed

Single Ir192 source

Page 22: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Currents from HDR sources

The BS EN 60731:1997 standard sets a "maximum input current" for a secondary standard dosemeter at 5 nanoamp

currents of 10-30nA per Citherefore possibly 500nA from a very hot HDR source

Page 23: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

RAKR Stability Analysis

Data collected by Chris Lee

0.980

0.990

1.000

1.010

1.020

0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20

Source number

Ratio of well chamber to Farmer measurement

Reference air kerma rate (RAKR) comparison between

a Standard Imaging 1000 Plus well chamber calibrated using the new COP,

and a Farmer chamber, calibrated at 280 kVp according to the old 1992 BIR/IPSM

recommendations.

Page 24: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

RAKR PTW 22004 Well chamber, calibrated according to the new COP,

and an NE 2571 Farmer Chamber , calibrated according to the IPEM 1992

Recommendations,

manufacturer’s source certificate for a Nucletron microSelectron-v1 (classic) HDR

192Ir source

0.98

0.99

1.00

1.01

1.02

Aug 04 Nov 04 Feb 05 May 05 Nov 05 Feb 06 May 06 Aug 06 Nov 06

Ratio of Well Chamber to Farmer

MeasurementsRatio of Well Chamber Measurement to

Certificate Value

Page 25: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Constancy checks of well-

chambers (long t1/2 source)

Cs 137 source, available but custom made jig for well chamber

required.

Activity and associated current to be selected

Cs137HTSL Cs

0.980

0.985

0.990

0.995

1.000

1.005

1.010

1.015

1.020

01-S

ep-0

7

01-D

ec-0

7

01-M

ar-

08

01-J

un-0

8

31-A

ug-0

8

01-D

ec-0

8

02-M

ar-

09

01-J

un-0

9

01-S

ep-0

9

Date

Measu

rem

en

t vari

ati

on

no

rmali

sed

to

th

e

refe

ren

ce

. Error bars give estimates of the 95% confidence intervals

based on the uncertainties of the temperature, pressure

and fluctuations of the measured charge.

Page 26: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Constancy checks of well-

chambers (2)

Calculate ratio for the two measured currents and

compare with the original reference value. Ratio

should agree with reference to within 2%.

If difference greater than 2% should be investigated

Problem may lie with either chamber

Page 27: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Ir-192 Clinical Source

Calibration

Objective:

Determine reference AKR of an HDR

source to ensure that the agreement

with the RAKR stated on the source

certificate agrees within +/- 5%.

+/- 3% should be possible

Page 28: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Ir-192 Source Calibration

When to Calibrate?

Field chamber should be cross calibrated with

secondary standard chamber immediately after

return from NPL

At machine Commissioning.

Following major repair or modification.

Following Source Replacement.

Page 29: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Associated QC

Timer checks, source position and

room safety checks should also be

done following service, source

replacement.

Possibly some applicator

checks(especially rings)

Page 30: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Quality AssuranceRemote Afterloading Systems

1. Source IntegritySources inaccessibleHigh activityWipe source path or transfer cable at machine serviceCheck catheters for activity

2. Source Position CheckFollowing position calibration or after source change. AutoradiographyUse reference jig viewed by in room camera

Page 31: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Quality AssuranceRemote Afterloading Systems

HDR source stepping checks during a single transfer: 10mm steps 1500mm-1470mm

1500mm

1490mm

1480mm

1470mm

Page 32: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Required Measurements for

Calibration

2 independent measurements:A and B

Measure AKR to derive activity.

Uses measured AKR to deliver a set dose.

2 physicists.

2 dosimeter / chamber combinations in ideal situation.

Page 33: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Ir-192 Source Calibration

How?

–Using a re-entrant chamber traceable to NPL

–Prior to measurement leave well chamber connected to

electrometer in treatment room for at least half an hour.

Page 34: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Published 2010

Recommends use of a well chamber

Defines the factor ksg to account for different source types.

No consensus so set to 1.

Comparison with source cert may be used as the second

check if cert comes from a approved and registered

calibration centre.

Well chamber stability check – ratio or check source

Electrometer calibration in terms of current.

Ir192 HDR Source Calibration in ClinicThe IPEM Code of Practice based on the NPL Air Kerma Standard

Page 35: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Set up well chamber 1m from walls and floor on low scatter

surface

Thermal equilibrium

Connect to calibrated electrometer

300V (collecting electrode is +ve wrt outer electrode)

Find position of max response

Measure stable current from at least 3 source transfers

Measure ion recombination using half voltage technique,

typically around 1.001 for 370GBq source

Correct for temp/pressure, ion recombination, source decay

Ir192 HDR Source Calibration in ClinicThe IPEM Code of Practice based on the NPL Air Kerma Standard

METHOD

Page 36: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Ir-192 Source Calibration

Part A: Measure Reference Air Kerma Rate (RAKR) of new source using the NPL calibrated secondary standard well chamber and electrometer, using chamber calibration coefficient

Connect to the HDR unit using the same transfer tube for sake of consistency

Minimize scatter conditions

RKN

RKN

RKN

Page 37: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

If the water equivalent thickness of the

tabletop (at least 1 m above floor level) is

between 2.5 mm and 15 mm, the scatter

contribution to the measured ionisation

current is within ±0.1% of the scatter

contribution during calibration of the well

chamber at the NPL.

Page 38: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Position of maximum chamber response

(Sweet Spot) of chamber

Each centre must find its own sweet spot for a particular transfer tube

NPL will also test for sweet spot location

“Peak” is broad (15-20mm) so any positional error (<+-1mm) in source is not significant

Min and Max measurement positions should be at least +-10mm from expected SS position

Range in position should be within +- 2.5mm of previously determined value

Page 39: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2

Source

Source Number D36B 1458 7/2/2007

0.5

0.55

0.6

0.65

0.7

0.75

0.8

0.85

0.9

0.95

1

1.05

128

0

128

5

129

0

129

5

130

0

130

5

131

0

131

5

132

0

132

5

133

0

133

5

134

0

134

5

135

0

135

5

136

0

136

5

137

0

137

5

138

0

MicroSelectron Source Position in Channel 3

Cu

rren

t (n

A)

PT

W U

nid

os 2

0525

Position

Postion + 1mm

Postion -1mm

Page 40: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Part A

• Send the source to the position of

maximum chamber response where

the NPL calibration is valid.

• Take at least 3 current readings ,

each from separate source transfers

and correct for background. There

should be no significant trend in the

readings.

• Calculate mean corrected current and

its standard deviation (SD< 0.1%)

Page 41: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Ion Recombination

Recommend Attix two voltage method for determination of recombination factor within the COP for users to determine their own factor. Therefore Part A readings repeated with chamber voltage at 150V

Factor ~ 1.001for 300GBq source but does depend on AKR

F H Attix, AAPM radiotherapy dosimetry protocol, Med. Phys. 11, 714 (1984)).

Page 42: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source
Page 43: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Part B

• Use RAKR measured in Part A to

calculate the time required to deliver a

specified air kerma at 1m. Set time on

HDR unit and measure the delivered air

kerma with an independent chamber

and electrometer

• The time set must be corrected for the

additional air kerma recorded during

source transit

Page 44: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Second “independent” check

Is the AKR from the source certificate a

second check?

Yes in Austria, the Netherlands, Belgium and

Norway! (unless the measurement is >5%

different from the certificate)

AAPM TG-56 (Nath et al 1997) for a

confirmatory check of source strength

Page 45: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Phantom end-to-end check

Small phantom drilled to take a tertiary

instrument (e.g. Farmer chamber) at a

fixed distance of 1cm-5cm from a drilled

channel which carries the source

catheter would be suitable for a “second

check”

(3-5cm probably best for HDR)

This would allow “end-to-end” check

Page 46: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Calibration Outcome

Source accepted if:the activity calculated by AKR measurement is

within +/-5% of that stated on the source certificate.

dose delivery measurement calculated from the RAKR measurement is also within +/-3%.

If source is from a nationally accredited lab then we can use the source certificate supplied by the manufacturer as a “second check”.

Page 47: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

However it is suggested that the AKR

measured at the Hospital is used in the

treatment planning system, rather than

the AKR from the source certificate.

Generally do not know the measurement

conditions/parameters of source

certifying institute

Page 48: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Error Frequency

Page 49: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

When calculating dosimetry - it is

important to consider:

Method of dose calculation used and standardisation of input data

Page 50: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

If activity is used in the TPS or for “convenience” then dose rate is usually calculated via:

Where k is a specific gamma ray constant or an air kerma rate constant

If A is determined from the RAKR, then k is a dummy variable which must be identical to that in the TPS.

How does a TPS calculate dose?

2r

AkD

Page 51: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Update of AAPM Task Group No.43 Report: A

revised AAPM protocol for brachytherapy

dose calculations

Rivard M J, Coursey B M, DeWerd L A,

Hanson W F, Huq M S, Ibbott G S,

Mitch M G, Nath R and Williamson J F

2004 Med. Phys. 31 633-74

Page 52: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

the dose rate constant (λ)

radial dose function g(r,) (absorption and

scattering)

geometry factors or source data (spatial

resolution of radioactivity within source)

G (r,Ө)

anisotropy function & factors F (r, Ө)

,],/,, 00 rFrgrGrGSrD k

Page 53: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Schematic diagram illustrating the geometry considered by the Task Group 43

dose calculation formalism for a linear radioactive source (adapted from AAPM

1995).

TG43: Medical Physics 22 (2) 209-235 Feb 1995

Med Phys. 2007 Jun;34(6):2187-205

Page 54: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Cylindrically symmetric source

,],/,, 00 rFrgrGrGSrD k

Air Kerma Strength Dose rate constant: dose rate to water at 1cm

Page 55: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Website for validated MC source data (TG-

43)

Currently accessible through Valencia

University

New data continuously validated by a

BRAPHYQS/AAPM subgroup

Easy access through ESTRO website:

www.estro.org

Page 56: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

“Insurance”

Good QA procedures

Second checks

Work instructions

Thorough familiarity with TPS

Communication between multi-

disciplinary staff groups

Page 57: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

MRI at time of

Brachytherapy

AFTER:

45Gy EXTB and

200mg/m2 cis-platin

GTV measured

2cm w, 3cmt, 3cmh

Page 58: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

6 cm tube

+ medium ovoid

2cm 15

10

10

20 20

AR AL

1.5

1

12 2

HDR

But R < 65Gy

or < 60 % dose to A

Page 59: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Standard loading pattern

Point A

Point A

45 Gy EBRT + 4 x BT:

85 Gyαβ10

60 Gyαβ10

15 mg10 mg10 mg

20

mg

20

mg

Courtesy : Kirisits, Vienna

Page 60: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Standard loading pattern

Courtesy : Kirisits, Vienna

Page 61: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

E-lfh website

Page 62: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Patient Results

Before Treatment After Treatment

?

2mths after treatmentPrior to any treatment

40cc HR-CTV

Courtesy S Aldridge and A Winship

Page 63: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Clinical dosimetry issues

Problem; organ motion, brachy applicator motion Multiple imaging sequences and parameters (fusion)

Varying dose rates (complex biologically equivalent doses)

Application of more complex algorithms (TG 186) Acuros and ACE are two commercial planning systems that implement this now

Page 64: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Prostate Brachytherapy

Page 65: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source
Page 66: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

HDR needles (Ir 192)I 125 seeds

HDR with u/s guided needles and remote afterloading or

LDR Iodine seeds

Page 67: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Test the whole process

Plan analysis

(DVH)

Isodoses

Dose prescription

Applicator and sources

identified

Phantom imaged

Digitised

Page 68: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Aspects that need checking

include:

Input of patient and applicator data

source and applicator localization

dose calculation around sources

use of multi-modality imaging,

evaluation and visualization of 3D dose

distributions

optimization routines

output of patient plan and associated

information

Page 69: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Useful references

http://www-naweb.iaea.org/nahu/dmrp/pdf_files/TECDOC1274.pdf

Page 70: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Use of phantoms and external audit

ESTRO ESQUIRE PROJECTEducation, Science and QUality Assurance

for RadiotherapyBrachytherapy Physics Quality assurance

System

5 cm

Source TLD

Reconstruction Dosimetry

UK: Alanine audit and Gaf chromic

Applicators and line sources

Page 71: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source
Page 72: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source
Page 73: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

- radiation absorption- signal emission

- signal transmission

- signal collection- dose rate

measurement

Important requirementson in-vivo dosimeter:• small size• real-time signal• dose linearity• large dynamic range• bendable

Real time luminescence

dosimetry

Courtesy Gustavo Kertzscher, DTU, Denmark

Page 74: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

References

Wilkinson JM (2001) Source specification and dosimetry. In: Principles and practice of Brachytherapy using afterloading systems. Chapter 2 (Joslin CAF, Flynn A, Hall EJ eds.) Arnold, London

The IPEM code of practice for determination of the reference air

kerma rate for HDR (192)Ir brachytherapy sources based on the

NPL air kerma standard.(2010)A M Bidmead, T Sander, S M Locks,

C D Lee, E G A Aird, R F Nutbrown, A Flynn Physics in Medicine

and Biology 06/2010; 55(11):3145-59.

www.e-lfh.org.uk

Page 75: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

TG43:

Medical Physics 22 (2) 209-235 Feb 1995

Update of AAPM Task Group No. 43 Report: A revised AAPM protocol for brachytherapy dose calculations.

Rivard MJ, Coursey BM, DeWerd LA, Hanson WF, Huq MS, Ibbott GS, Mitch MG, Nath R, Williamson JF.

Med Phys. 2004 Mar;31(3):633-74. Erratum in: Med Phys. 2004 Dec;31(12):3532-3.

Supplement to the 2004 update of the AAPM Task Group No. 43 Report.

Rivard MJ, Butler WM, DeWerd LA, Huq MS, Ibbott GS, Meigooni AS, Melhus CS, Mitch MG, Nath R, Williamson JF; American Association of Physicists in Medicine.

Med Phys. 2007 Jun;34(6):2187-205

Page 76: Brachytherapy Dosimetry in the Clinic - National Physical ... · PDF fileMicro Selectron Ir192 ... Positional Response of Nucletron SDS Well Chamber 25039 with HDR V2 Source Source

Report of the Task Group 186 on model-

based dose calculation methods

in brachytherapy beyond the TG-43

formalism: Current status and

recommendations for clinical

implementation: Med. Phys. 39 (10),

October 2012

Luc Beaulieua,Åsa Carlsson Tedgren,

Jean-François Carrier, Stephen D. Davis,

Firas Mourtada, Mark J. Rivard, Rowan M.

Thomson, Frank Verhaegen, Todd A.

Wareing, Jeffrey F. Williamson