TRANQUILITY IISimple and intuitive multi-parameter monitor
The Schiller’s TRANQUILITY II is a pre-configured, flexible and reliable adult, pediatric and neonate patient monitor for all types of settings including Surgery, ICU, PACU, Recovery, and Emergency.
The TRANQUILITY II basic configuration is capable to measure ECG, Respiration, SpO2 (Digital & Masimo options), NIBP, and dual Temperature. The ease to upgrade to EtCO2, dual IBP and Anesthetic Agents depending on the clinical needs make this monitor
adaptable to various situations and workflows in any medical environment.
The TRANQUILITY II Patient Monitor is a comprehensive monitoring system designed for compiling, processing, analyzing, and displaying data from up to nine (9) different parameters. With a color LCD 12.1” touchscreen display and optional printer in one device. Compact, lightweight, portable, and built-in battery facilitates the transportation of the patient.
COMPLETE MONITORING IN A PORTABLE DEVICE
LIMITLESS SUPERVISION
The Tranquility II may function as an independent monitor or connected to SCHILLER’s patient monitor central station TRANQUILITY VIEW with the option to integrate into the hospital’s Electronic Medical Record using HL7.
TRANQUILITY II
TRANQUILITY II
Technical SpecificationsTechnical Specifications
Compact design is convenient for mobile monitoring
12.1” color TFT touchscreen display
Multi-Lead ECG monitoring
4, 6 & 8 waveforms shown
Drug dose calculations
NIBP list & OxyCRG dynamic view display
Optional: Dual IBP, EtCO2, Anesthetic Agent and Built-in Thermal Recorder
Networkable with Central Monitoring System (wireless or wired network, up to 32 beds)
GENERAL SPECIFICATIONS
Display:
Resolution:
Trace:
Waveforms:
Indicator:
Trend time:
Thermal printerRecorder:
Record width:
Recorder paper:
Record speed:
Oxy CRG:
12.1” color TFT, Touch Screen
800 x 600 pixels
4, 6, 8 wave forms
ECG (I, II, III, aVR, aVL, aVF,V1-V6), PLETH, RESP, IBPx2,ETCO2
Alarm indicatorPower indicatorQRS beep and alarm sound
Up to 72 hrs
Built-in, thermal array,2 channels
48 mm
50 mm
25 mm/s
Combines HR, SPO2 andRespiration trends into a singlegraph
PERFORMANCE SPECIFICATIONS
POWERSource:
AC Power:
Battery:
Operating timeunder normalcondition:
Operating timeafter the firstalarm of lowbattery:
TEMPERATUREWorking:
Storage:
RELATIVE HUMIDITYWorking:
Storage:
External AC power and internalbattery
100~240VAC, 50/60 Hz, <150VA
Rechargeable Lithium ion battery
2 hours
15 minutes
41°F to 104°F (5°C to 40°C )
-14°F to 113°F (-10°C to 45°C)
≤80%
≤80%
OPERATION ENVIRONMENT
Technical Specifications
TRANQUILITY II
Input:
Lead Choice:
Gain Choice:
FILTERSDiagnostic mode:
Monitoring mode:
Surgical mode:
ECG Waveforms:
HR Display Range:
Sweep Speed:
Accuracy:
ST Segment Detection:
Measurement Range:
ArrhythmiaAnalysis:
ALARM LIMIT RANGE SETTINGUpper limit:
Lower limit:
5-Lead ECG cable standard 12-Lead ECG cable optional andstandard AAMI line for connection
5-Lead: I, II, III, aVR, aVL, aVF, V12-Lead: I, II, III, aVR, aVL, aVF, V1~V6
x0.25, x0.5, x1.0, x2.0
0.05 Hz~130 Hz
0.5 Hz~40 Hz
1 Hz~20 Hz
3 to 12 leads
30~300 bpm
12.5, 25, 50 & 100 mm/s
±1bpm or ±1%, whichever isgreater
by lead
-2~2 mV
13 types
80-400 bpm
20-150 bpm
ECG
Measuring Technology:
Cuff Inflating:
Measuring Period:
Mode:
Measuring intervalin auto mode:
Pulse Rate Range:
MEASURING RANGESystolic:
Diastolic:
Resolution:
ACCURACYPressure:
Overpressureprotection:
Alarm Limit Settings:
Automatic oscillatingmeasurement
<30s (0~300 mmHg, standardAdult Cuff)
AVE <40s
Manual, Auto, STAT
2 minutes~4 hours
30 ~250 bpm
Adult/Pediatric: 40~250mmHgNeonate: 40~135mmHg
Adult/Pediatric: 15~200mmHgNeonate: 15~100mmHg
1mmHg
Maximum Mean error: ±5mmHgMaximum Standard Deviation:8mmHg
Adult mode: 280 (mmHg)Neonate mode: 150 (mmHg)
SYS 30~240 mmHgDIA 15~180 mmHg
NIBP
Technical Specifications
TRANQUILITY II
Range:
Accuracy:
Display resolution:
Alarm upper andlower limit setting:
Channels:
T1 and T2: 77 °F~122°F (25°C~50°C)Delta T: 0°F~9.9°F (0°C~5.5°C)
±0.2°C (25.0°C~34.9°C) / (77°F~94.8°F)±0.1°C (35.0°C~39.9°C) / (95°F~103.8°F)±0.2°C (40.0°C~44.9°C) / (104°F~112.8°F)
0.2°F (0.1°C)
50°F to 122°F (10°C to 50°C)
2 channels, provide T1, T2, and Delta T
TEMPERATURE
ASpO2:
SpO2 Range:
SpO2 Accuracy:
Pulse Rate Range:
Pulse RateAccuracy:
SpO2 Alarm upperand lower limitsetting:
SpO2 Probe:
Anti-motion SpO2
0~100%
+-2% 70-100%, non-motion+-3% 70-100%, motion
30~250 bpm
±2 bpm (non-motion) ±3 bpm (motion)
50~100%
Red light LED wavelength: 660nm ± 5nmInfrared light LED wavelength: 940nm ± 10nm
SPO2
TRANQUILITY II
Options
SpO2 Masimo SET®Reusable & disposable sensors
ARMS Accuracy for RD
Motion and low perfusion
For patients adult, pediatric, and neonate
EtCO2Mode of Sampling:
Measurement technology:
Apnea Time:
Alarm Limit:
Respiration rate
ANESTHETIC GAS MONITORING(PHASEIN)Infrared Sidestream Analyzer (ISA) from Masimo
Range:
Barometric Pressuresetting range:
Gas Temperaturesetting range:
Anesthetic Agents:
EtCO2
Nitrous oxide (N2O)
Oxygen (O2)
Respiration rate
-100%
Sidestream
Infrared Absorption
10~60 s
Upper Limit: 20~100 mmHgLower Limit: 20~100 mmHg
0.0~20.0%
400~850 mmHg
32°F ~ 122°F (0~50°C)
Sevoflurane, Isoflurane,Halothane, Desflurane, andEnflurane
MASIMO TECHNOLOGY
Measuring Technology:
Measurement Range:
Accuracy:
Channel:
PressureTransducer:
Labels:
ImpedanceRange:
Direct Invasive Measurement
-10~300 mmHg
±1 mmHg or ±2%, whichever isgreater
2 channels
Sensitivity, 5µV/V/mmHg
ART, PA, CVP, LAP, RAP, ICP(MAP)
300~3000 Ω
IBP
Built-in, thermal array, 2 channels
THERMAL PRINTER
Rolling Stand
Wall Mount
MOUNTING OPTIONS