BT30 Instructions for use Biological indicator for Dry Heat sterilization

Embed Size (px)

Citation preview

Page 1: BT30 Instructions for use Biological indicator for Dry Heat sterilization

BT30 Instructions for useBiological indicator for Dry Heat sterilization

Page 2: BT30 Instructions for use Biological indicator for Dry Heat sterilization

TUBE B

TUBE A

Page 3: BT30 Instructions for use Biological indicator for Dry Heat sterilization

Identify TUBE A writing the sterilizer Nº, load Nº and processing date on the indicator label.

Page 4: BT30 Instructions for use Biological indicator for Dry Heat sterilization

Pack the biological indicator along with the material for sterilization.

Page 5: BT30 Instructions for use Biological indicator for Dry Heat sterilization

Do not sterilize TUBE B.

Page 6: BT30 Instructions for use Biological indicator for Dry Heat sterilization

Sterilize as usual.

Page 7: BT30 Instructions for use Biological indicator for Dry Heat sterilization

Once the sterilization processhas finished, verify that IB labelcolor has changed to brown.Let the IB cool down.

Page 8: BT30 Instructions for use Biological indicator for Dry Heat sterilization

Break TUBE B ampoule with the crusher in BIONOVA® IC10/20 incubator or with an ampoule crusher.

Page 9: BT30 Instructions for use Biological indicator for Dry Heat sterilization

Remove and discard the cap of TUBE B under sterile environment or laminar flow.

STERILITYconditions

Page 10: BT30 Instructions for use Biological indicator for Dry Heat sterilization

Remove the cap of TUBE Aand discard the tube.

STERILITYconditions

Page 11: BT30 Instructions for use Biological indicator for Dry Heat sterilization

Put TUBE A cap inside TUBE Bin such a way that the sporespaper carrier gets in contactwith the culture medium.IMPORTANT! Make sure thecarrier does not get in contactwith the exterior side of thetube or any other surface.

STERILITYconditions

Page 12: BT30 Instructions for use Biological indicator for Dry Heat sterilization

Close the cap tightly.

STERILITYconditions

Page 13: BT30 Instructions for use Biological indicator for Dry Heat sterilization

Place the processed IB in the incubator.IMPORTANT! Use a non-sterilized biological indicator as a positive control each time aprocessed indicator is incubated.

Page 14: BT30 Instructions for use Biological indicator for Dry Heat sterilization

Incubate at 37±2 ºC for 48 hs. maximum.

Page 15: BT30 Instructions for use Biological indicator for Dry Heat sterilization

If after incubation the color of the growth indicator medium has not changed, the result is NEGATIVE, thus, the sterilization process is EFFICIENT.

Page 16: BT30 Instructions for use Biological indicator for Dry Heat sterilization

If after incubation the color of the growth indicator medium has changed to yellow,the result is POSITIVE, thus, the sterilization process HAS FAILED.

Page 17: BT30 Instructions for use Biological indicator for Dry Heat sterilization

Record the results and discard the tube immediately according to your country's healthcare and safety regulations.

Page 18: BT30 Instructions for use Biological indicator for Dry Heat sterilization