MSA
CALIBRATION TEST SYSTEMMODEL R CHECK KIT USINGFLOW CONTROLPART NO. 459948
instruction manual
IMPORTANT WARNING \
THIS MANUAL MUST BE CAREFULLY READ BY ALL INDIVIDUALSWHO HAVE OR WILL HAVE THE RESPONSIBILITY FOR USING ORSERVICING THE PRODUCT. Like any piece of complex equipment, theCALIBRATION TEST SYSTEM MODEL R CHECK KIT USING FLOWCONTROL PART NO. 459948 will perform as designed only if it Is usedand serviced in accordance with the manufacturer's instructions.OTHERWISE IT COULD FAIL TO PERFORM AS DESIGNED ANDPERSONS WHO RELY ON THIS PRODUCT FOR THEIR SAFETYCOULD SUSTAIN SEVERE BODILY INJURY OR DEATH.The warranties made by Mine Safety Appliances Company with respectto the product are voided if the product is not installed, used and servicedin accordance with the instructions in this manual. Please protectyourself and your employees by following them. We encourage ourcustomers to write or call for a demonstration of this equipment prior touse or for any additional information relative to use or repairs.
©COPYRIGHT, MINE SAFETY APPLIANCES COMPANY, 1985
Manufactured by
MINE SAFETY APPLIANCES COMPANYPITTSBURGH, PENNSYLVANIA, U.S.A. 15230(L) REV. 5 .996388
IMPORTANT WARNING
THIS MANUAL MUST BE CAREFULLY READ BY ALL INDIVIDUALS WHO HAVE ORWILL HAVE THE RESPONSIBILITY FOR USING OR SERVICING THE PRODUCT. Likeany piece of complex equipment, the CALIBRATION TEST SYSTEM MODEL R CHECKKIT USING FLOW CONTROL PART NO. 459948 will perform as designed only if it is usedand serviced in accordance with the manufacturer's instructions. OTHERWISE IT COULDFAIL TO PERFORM AS DESIGNED AND PERSONS WHO RELY ON THIS PRODUCTFOR THEIR SAFETY COULD SUSTAIN SEVERE BODILY INJURY OR DEATH.
The warranties made by Mine Safety Appliances Company with respect to the product are voidedif the product is not installed, used and serviced in accordance with the instructions in thismanual. Please protect yourself and your employees by following them. We encourage ourcustomers to write or call for a demonstration of this equipment prior to use or for any additionalinformation relative to use or repairs.
Table of Contents
Calibration Test System for use withExplosimeter® Models 2,2a, 2b, 2c, 3,4,5 and 6Gascope® Model 53 Calibrated for Natural Gas, 0-100% LEL ScaleCombustible Gas Indicator, Models 40, & 40s, 0-100% LEL ScaleModels 20 and 21,0-100% LEL Pentane Scale 1
Cah'bration Test System for use withCombustible Gas and Oxygen Indicator Model 260 and 261 3
Calibration Test System for use withCarbon Monoxide Indicator Model 70 4
Calibration Test System for use withCombustible Gas and Oxygen Alarm Model 260s, LEL Scale Only 5
Calibration Test System for use withModel 30 Combustible Gas Indicator 6
Calibration Test System for use withTankscope™, Model B Indicator 7
Calibration Test System for use withCarbon Monoxide, Combustible Gas And Oxygen Alarm Model 360 8
Cah'bration Test System for use withGascope® Model 60 [0-5% Natural Gas (methane) Range]Model 62 [0-100% LEL Pentane]Model 62s [0-100% LEL Methane]Model 62 Special [0-100% LEL Methane0-5 or LEL Scales Only 10
Calibration Test System Model R Check Kit Using Flow Control part no. 459948
Calibration Test System for use withExplosimeter® Models 2,2a, 2b, 2c, 3,4, 5 and 6
Gascope Model 53 Calibrated for Natural Gas, 0-100% LEL ScaleCombustible Gas Indicator, Models 40, and 40s, 0-100% LEL Scale
Models 20 and 21,0-100% LEL Pentane Scale
Flow Control 459948
Hose 449401
Calibration gas, 40% LEL Methane (Methane 2.0%) 459945
OR Calibration gas, 50% LEL Methane (Methane 2.5%) 459942
Instructions
Before the calibration can be checked, theinstrument and its aspirator sampling bulbmust be in operating condition as describedin the instrument Instruction Manual.
1. Attach the flow control to the calibrationgas tank.
2. Connect the hose to the flow control andto the instrument inlet fitting.
3. Open the flow control valve.
4. Record the meter reading after itstabilizes.
NOTE: It is not necessary to operate theaspirator bulb to obtain thecalibration sample. Depending onwhich calibration gas is used, theinstrument reading should bewithin one of the followingranges:
2C*
2, 2A, 2B, 40,6, 20 and 21
3
4
5
53
40S
2%METHANE
40%
42 to 60%
30 to 38%
35 to 47%
68 to 84%
40 to 51%
50 to 70%
50%
54 to 76%
40 to 52%
48 to 65%
76 to off scale
53 to 68%
65 to 95%
"This instrument has an internal span control.Adjust the span control to obtain the correct
reading for the gas used; if this cannot be done,replace the detector filament.
If the instrument does not read within theacceptable range, replace the detectorfilament unit (the black filament, part no.11355 in Models 2, 2A, 2B, 2C, 3, 4, 6, 20and 21, 40, 40S and 53 OR part no. 75476in the Model 5) and repeat the calibration
Using Flow Control part no. 459948 Calibration Test System Model R Check Kit
check procedure. See applicable instrumentInstruction Manual for complete servicedetails.
5. Close the flow control valve.
6. Remove the hose from the flow controland the inlet fitting on the instrument.
7. Remove the flow control from thecalibration gas tank.
Calibration Test System Model R Check Kit Using Flow Control part no. 459948
Calibration Test System for use withCombustible Gas and Oxygen Indicator Model 260 and 261
PARTS REQUIREDFlow Control
Hose
Calibration gas, 50% LEL Pentane (Pentane 0.75% and Oxygen 15% in nitrogen)
NOTE: The Model 260 and 261 are normally factory-calibrated on pentane.
459948
449401
476304
Instructions
Test the operation of the pump and sampleflow indicator by momentarily placing afinger over the sample inlet of theinstrument. The flow indicator should be atthe top of the window when the inlet isopen; the indicator should then fall fromview when the inlet is blocked. If not, seethe instrument Instruction Manual forrequired maintenance.
1. Attach the flow control to the calibrationgas tank.
2. Connect the hose to the flow control.
3. Open the flow control valve.
4. Connect the hose fitting to the inlet of theinstrument; after about 15 seconds theLEL meter should be stable and indicatingbetween 47 and 55%. If the indication is
not in the correct range, remove the rightend of the indicator and adjust the SPANor "S" control to obtain 50%.
5. Verify the oxygen meter indication; itshould be between 13 and 17%. (This is aresponse check only. The oxygen sensorshould be calibrated to 20.8% oxygen infresh air before each use.)
6. Disconnect the hose fitting from theinstrument.
7. Close the flow control valve.
8. Remove the hose from the flow control.
9. Remove the flow control from thecalibration gas tank.
Using Flow Control part no. 459948 Calibration Test System Model R Check Kit
Calibration Test System for use withCarbon Monoxide Indicator Model 70
Flow Control
Hose
Calibration gas, Carbon Monoxide 60 ppm
Calibration gas, Carbon Monoxide 300 ppm
M?N*X459948
449401
461768
461769
Instructions
Before the calibration can be checked, theModel 70 must be in operating condition.The instrument must be allowed to stabilizefor at least 1 hour in the STAND-BY mode ifpreviously stored OFF. If a sampling lineand/or moisturizing bottle are to be usedwith the instrument, they must be connectedto the instrument while the calibration ischecked.
1. Attach the flow control to the calibrationgas tank.
2. Connect the hose to the flow control.
3. Connect the hose fitting to the instrumentinlet or to the instrument sampling line, ifone is to be used.
4. Open the regulator valve.
5. Set the MODE switch to the ON positionand the RANGE switch to the range beingchecked.
6. Allow the instrument to reach a stableindication. Adjust the SPAN control, ifnecessary, to obtain the correct indicationfor the calibration gas used.
7. Remove the hose from the regulator andinstrument.
8. Close the regulator valve.
9. Remove the regulator from the calibrationgas tank.
Calibration Test System Model R Check Kit Using Flow Control part no. 459948
Calibration Test System for use withCombustible Gas and Oxygen Alarm Model 260s, LEL Scale Only
Flow Control
Hose
Calibration gas, 28% LEL Propane (Propane 0.6%)
PABT NO.
459948
449401
459943
NOTE: The Model 260S is factory-calibrated on Propane.
Instructions
Test the operation of the pump and sampleflow indicator by momentarily placing afinger over the sample inlet of theinstrument. The flow indicator should be atthe top of the window when the inlet isopen; the indicator should then fall fromview when the inlet is blocked. If not, seethe instrument Instruction Manual forrequired maintenance.
1. Attach the flow control to the calibrationgas tank.
2. Connect the hose to the flow control.
3. Open the flow control valve.
4. Connect the hose fitting to the inlet of theinstrument; after about 15 seconds the
LEL indication should be stable andindicating between 24 and 33% LEL (Ifthe indication is not in the correct range,remove the right-hand side cover andadjust the SPAN or "S" control to obtain areading of 28).
5. Disconnect the hose fitting from theinstrument.
6. Close the flow control valve.
7. Remove the hose from the flow control.
8. Remove the flow control from thecalibration gas tank.
Using Flow Control part no. 459948 Calibration Test System Model R Check Kit
Calibration Test System for use withModel 30 Combustible Gas Indicator
Flow Control
Hose
Calibration Gas, 40% LEL Methane (2.0% methane )
AND Calibration Gas, 28% LEL Propane (0.60% Propane)
PART NO.
459948
449401
459945
459943
Instructions
Before the calibration can be checked, theinstrument and its aspirator sampling bulbmust be in operating condition as describedin the instrument Instruction Manual.
1. Attach the flow control to the calibrationgas tank.
2. Connect the hose to the flow control andinstrument inlet fitting.
3. Open the flow control valve.
4. Record the meter reading after itstabilizes.
NOTE: It is not necessary to operate theaspirator bulb to obtain thecalibration sample. Depending onwhich calibration gas is used, theinstrument reading should bewithin one of the followingranges:
CALIBRATIONGAS
40% LEL Methane
28% LEL Propane
„
48-60%
31-39%
j
0
18-24%
5. Close the flow control valve.
6. Remove the hose from the flow controland instrument inlet fitting.
7. Remove the flow control from thecalibration gas tank.
8. Repeat the steps 1 through 7 using theother calibration gas.
If the instrument does not read within theacceptable range, replace the detectorfilament unit, part no. 11355, (in the wellmarked WHITE). Repeat the calibrationcheck procedure; see instrumentInstruction Manual for complete servicedetails.
Calibration Test System Model R Check Kit Using Flow Control part no. 459948
Calibration Test System for use withTankscope™ Model B Indicator
Flow Control
Hose
Calibration gas, Butane, 8% in Nitrogen
t f-u>* i»vL-
459948
449401
460345
NOTE: Calibration gas can is not accurate when used at temperatures below SOT.
Instructions
Before the calibration can be checked, theinstrument and its aspirator sampling bulbmust be in operating condition as describedin the instrument Instruction Manual.
NOTE: Calibration should be done at atemperature between 50° and 90T.
1. Attach the flow control to the calibrationgas tank.
2. Connect the hose to the flow control andto the instrument inlet fitting.
3. Open the flow control valve and allow thecalibration gas to flow through theinstrument for 10 to 20 seconds.
4. Close the flow control valve.
5. The instrument reading should be withinthe range of 7.6 to 8.4% gas. If theinstrument does not read within thisrange, use a small screwdriver to adjustthe panel SPAN control until a reading of8.0% is obtained.
6. Remove the hose from the flow controland the instrument inlet fitting.
7. Remove the flow control from thecalibration gas tank.
Using Flow Control part no. 459948 Calibration Test System Model R Check Kit
Calibration Test System for use withCarbon Monoxide, Combustible Gas and Oxygen Alarm Model 360
PARTS REQUIREDFlow Control
Hose
Calibration Gas, 50% LEL Pentane (0.75% Pentane 15% Oxygen in Nitrogen)
Calibration Gas, 60 ppm Carbon Monoxide
PAJtT NO-459948
449401
476304
461768
Instructions
Test the operation of the pump and sampleflow indicator by momentarily placing afinger over the sample inlet of theinstrument. The flow indicator should be atthe top of the window when the inlet isopen; the indicator should then fall fromview when the inlet is blocked. If not, seethe instrument Instruction Manual forrequired maintenance.
1. Attach the flow control to the pentanecalibration gas tank.
2. Connect the hose to the flow control.
3. Open the flow control valve.
4. Connect the hose fitting to the instrumentinlet; after approximately 15 seconds, theLEL readout should stabilize and indicatebetween 47 and 55%. If the indication isnot in the correct range, remove the rightend of the instrument and adjust the LELSPAN control to obtain 50%.
5. Verify the oxygen reading; it should bebetween 13 and 17%. (This is a response
check only. The oxygen sensor should becalibrated to 20.8% oxygen in fresh airbefore each use.)
6. Disconnect the hose fitting from theinstrument.
7. Close the flow control valve.
8. Remove the flow control from thecalibration gas tank.
9. Attach the flow control to the carbonmonoxide calibration gas tank.
10. Open the flow control valve.
11. Connect the hose fitting to the instrumentinlet; after approximately 1 minute, theTOX readout should stabilize and indicatebetween 54 and 66 ppm. If the indicationis not in the correct range, remove theright end of the indicator and adjust theTOX SPAN control to obtain 60 ppm.
12. Disconnect the hose fitting from theinstrument.
8
Calibration Test System Model R Check Kit Using Flow Control part no. 459948
13. Close the flow control valve.
14. Remove the hose from the flow control.
15. Remove the flow control from thecalibration gas tank.
Using Flow Control part no. 459948 Calibration Test System Model R Check Kit
Calibration Test System for use withGascope® Model 60 [0-5% Natural Gas (methane) Range]
Model 62 [0-100% LEL Pentane]Model 62s [0-100% LEL Methane]
Model 62 Special [0-100% LEL Methane0-5 or LEL Scales Only
Row Control 459948
Hose 449401
Calibration Gas, 40% LEL Methane (2% Methane) for Model 60, 62S and 62 Special 459945
Calibration Gas, 50% LEL Pentane (0.75% Pentane) for Model 62 with standard Pentane 466193calibration
Model 60 Instruction Manual 466520
Model 62 Instruction Manual 465919
Model 62S and 62 Special Instruction Manual 468453
Instructions
Before the calibration can be checked, theinstrument and its aspirator sampling bulbmust be in operating condition as describedin the Maintenance section of theinstrument Instruction Manual.
The Gascope RANGE switch must be in the0-5 position (Model 60) or the LELposition (Models 62, 62S and 62 Special).
1. Attach the flow control to the calibrationgas tank.
2. Connect the hose to the flow control andto the instrument inlet fitting.
3. Open the flow control valve.
4. Record meter reading after it stabilizes.
NOTE: It is not necessary to operate theaspirator bulb to obtain thecalibration sample. Depending onthe instrument type and thecalibration gas used, theinstrument reading should bewithin one of the followingranges:
'62
62S
62 SPECIAL
1.9 to 2.1%
38 to 42%
38 to 42%
47 to 53%
10
Calibration Test System Model R Check Kit Using Flow Control part no. 459948
5. Disconnect the hose fitting from theinstrument.
6. Close the flow control valve.
7. Remove the hose from the flow control.
8. Remove the flow control from thecalibration gas tank.
NOTE: The above procedure will calibratethe 0-5 or LEL ranges on the
Gascope Models 60 or 62,respectively. The 0-100 or GAS rangeon the Gascopes normally do notneed to be recalibrated unless theinstrument is damaged, usually bydrawing liquids into the sample inlet.If the Gascope is damaged andrepaired, the 0-100 or gas range mustbe calibrated using the additionalequipment and steps stated in theinstrument Instruction Manual.
11
I
M a n u f a c t u r e d by
MSA MINE SAFETY APPLIANCES COMPANYPITTSBURGH, PENNSYLVANIA, U.S.A. 15230