32
Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method

Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method

Embed Size (px)

Citation preview

Page 1: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method

Calibration Techniques

1. Calibration Curve Method

2. Standard Additions Method

3. Internal Standard Method

Page 2: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method

Calibration Curve Method

1. Most convenient when a large number of similar samples are to be analyzed.

2. Most common technique.

3. Facilitates calculation of Figures of Merit.

Page 3: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method

Calibration Curve Procedure

1. Prepare a series of standard solutions (analyte solutions with known concentrations).

2. Plot [analyte] vs. Analytical Signal.

3. Use signal for unknown to find [analyte].

Page 4: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method

Example: Pb in Blood by GFAAS

[Pb] Signal

(ppb) (mAbs)

   

0.50 3.76

1.50 9.16

2.50 15.03

3.50 20.42

4.50 25.33

5.50 31.87

Results of linear regression:

S = mC + b

m = 5.56 mAbs/ppb

b = 0.93 mAbs

Page 5: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method

0

5

10

15

20

25

30

35

0 1 2 3 4 5 6

Pb Concentration (ppb)

mA

bs

y = 5.56x + 0.93

Page 6: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method

A sample containing an unknown amount of Pb gives a signal of 27.5 mAbs. Calculate the Pb concentration.

S = mC + b

C = (S - b) / m

C = (27.5 mAbs – 0.92 mAbs) / 5.56 mAbs / ppb

C = 4.78 ppb

(3 significant figures)

Page 7: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method

Calculate the LOD for Pb

20 blank measurements gives an average signal

0.92 mAbs

with a standard deviation of

σbl = 0.36 mAbs

LOD = 3 σbl/m = 3 x 0.36 mAbs / 5.56 mAbs/ppb

LOD = 0.2 ppb

(1 significant figure)

Page 8: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method

Find the LDR for Pb

Lower end = LOD = 0.2 ppb

(include this point on the calibration curve)

SLOD = 5.56 x 0.2 + 0.93 = 2.0 mAbs

(0.2 ppb , 2.0 mAbs)

Page 9: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method

Find the LDR for Pb

Upper end = collect points beyond the linear region and estimate the 95% point.

Suppose a standard containing 18.5 ppb gives rise to s signal of 98.52 mAbs

This is approximately 5% below the expected value of 103.71 mAbs

(18.50 ppb , 98.52 mAbs)

Page 10: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method

Find the LDR for Pb

LDR = 0.2 ppb to 18.50 ppb

or

LDR = log(18.5) – log(0.2) = 1.97

2.0 orders of magnitude

or

2.0 decades

Page 11: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method

Find the Linearity

Calculate the slope of the log-log plot

log[Pb] log(S)

-0.70 0.30-0.30 0.580.18 0.960.40 1.180.54 1.310.65 1.400.74 1.501.27 1.99

Page 12: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method

0.00

0.50

1.00

1.50

2.00

2.50

-1.00 -0.50 0.00 0.50 1.00 1.50

log(Pb concentration)

log

(Sig

na

l)

y = 0.0865 x + 0.853

Not Linear??

Page 13: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method

Not Linear??

0

20

40

60

80

100

120

0 2 4 6 8 10 12 14 16 18 20

Pb Concentration (ppb)

Sig

na

l (m

Ab

s)

Page 14: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method

Remember

S = mC + b

log(S) = log (mC + b)

b must be ZERO!!

log(S) = log(m) + log(C)

The original curve did not pass through the origin. We must subtract the blank signal from each point.

Page 15: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method

Corrected Data

[Pb] Signal(ppb) (mAbs)0.20 1.070.50 2.831.50 8.232.50 14.103.50 19.494.50 24.405.50 30.94

18.50 97.59

log[Pb] log(S)

-0.70 0.03-0.30 0.450.18 0.920.40 1.150.54 1.290.65 1.390.74 1.491.27 1.99

Page 16: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method

Linear!

y = 0.9965x + 0.7419

0.00

0.50

1.00

1.50

2.00

2.50

-1.00 -0.50 0.00 0.50 1.00 1.50

log(Pb concentration)

log

(sig

na

l)

Page 17: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method

Standard Addition Method

1. Most convenient when a small number of samples are to be analyzed.

2. Useful when the analyte is present in a complicated matrix and no ideal blank is available.

Page 18: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method

Standard Addition Procedure

1. Add one or more increments of a standard solution to sample aliquots of the same size. Each mixture is then diluted to the same volume.

2. Prepare a plot of Analytical Signal versus:a) volume of standard solution added, or

b) concentration of analyte added.

Page 19: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method

Standard Addition Procedure

3. The x-intercept of the standard addition plot corresponds to the amount of analyte that must have been present in the sample (after accounting for dilution).

4. The standard addition method assumes:a) the curve is linear over the concentration range

b) the y-intercept of a calibration curve would be 0

Page 20: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method

Example: Fe in Drinking Water

Sample Volume

(mL)

Standard Volume

(mL) Signal (V)

10 0 0.21510 5 0.42410 10 0.68510 15 0.82610 20 0.967

The concentration of the Fe standard solution is 11.1 ppm

All solutions are diluted to a final volume of 50 mL

Page 21: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method

-0.2

0

0.2

0.4

0.6

0.8

1

1.2

-10 -5 0 5 10 15 20 25

Volume of standard added (mL)

Sig

na

l (V

)

-6.08 mL

Page 22: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method

[Fe] = ?

x-intercept = -6.08 mL

Therefore, 10 mL of sample diluted to 50 mL would give a signal equivalent to 6.08 mL of standard diluted to 50 mL.

Vsam x [Fe]sam = Vstd x [Fe]std

10.0 mL x [Fe] = 6.08 mL x 11.1 ppm

[Fe] = 6.75 ppm

Page 23: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method

Internal Standard Method

1. Most convenient when variations in analytical sample size, position, or matrix limit the precision of a technique.

2. May correct for certain types of noise.

Page 24: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method

Internal Standard Procedure

1. Prepare a set of standard solutions for analyte (A) as with the calibration curve method, but add a constant amount of a second species (B) to each solution.

2. Prepare a plot of SA/SB versus [A].

Page 25: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method

Notes

1. The resulting measurement will be independent of sample size and position.

2. Species A & B must not produce signals that interfere with each other. Usually they are separated by wavelength or time.

Page 26: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method

Example: Pb by ICP EmissionEach Pb solution contains 100 ppm Cu.

[Pb] (ppm) Pb Cu Pb/Cu

20 112 1347 0.08340 243 1527 0.15960 326 1383 0.23680 355 1135 0.313100 558 1440 0.388

Signal

Page 27: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method
Page 28: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method

No Internal Standard Correction

0

100

200

300

400

500

600

0 20 40 60 80 100 120

[Pb] (ppm)

Pb

Em

issi

on

Sig

nal

Page 29: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method

0.000

0.050

0.100

0.150

0.200

0.250

0.300

0.350

0.400

0.450

0 20 40 60 80 100 120

[Pb] (ppm)

Pb

Em

issi

on

Sig

nal

Internal Standard Correction

Page 30: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method

Results for an unknown sample after adding 100 ppm Cu

Run Pb Cu Pb/Cu

1 346 1426 0.2432 297 1229 0.2423 328 1366 0.2404 331 1371 0.2415 324 1356 0.239

mean 325 1350 0.241σ 17.8 72.7 0.00144S/N 18.2 18.6 167

Signal

Page 31: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method
Page 32: Calibration Techniques 1. Calibration Curve Method 2. Standard Additions Method 3. Internal Standard Method