    By Nick Bublitz

    Performing a Linear Scan Sensitivity/

    Velocity/Wedge Delay Calibration-Longitudinal Wave PA Using SDHs at

    Known Depths 2.0 SW

  • 8/12/2019 Calibrations Linear 2 0 SW

    2/24

    2

    Calibrations-Required itemsCalibration block with 2 same size reflectors

    at known depths- (Navships block with

    SDHs)

    Couplant-water works best

    Guide to keep from skewing- use a ruler, set

    the block on its side, or on IHC wedges can

    screw down one sides carbide pins.

  • 8/12/2019 Calibrations Linear 2 0 SW

    3/24

    3

    Velocity Calibration The velocity calibration will fine tune the velocity to

    that of the calibration block. It will require two

    reflectors of the same size at two different knowndepths.

    In many cases the default material velocity and

    wave type set by the operator in the group wizardis sufficient and this calibration can be skippedbecause wedge delay will adjust accordingly to

    allow accurate measurements. If performed the velocity calibration should be

    done before the wedge delay calibration.

  • 8/12/2019 Calibrations Linear 2 0 SW

    4/24

    4

    Select Velocity CalibrationGo to the Wizard-

    Calibration menu

    Select Ultrasound and

    Velocity

    Start the wizard

  • 8/12/2019 Calibrations Linear 2 0 SW

    5/24

    5

    Select A-Scan Put the probe on your

    block over the first

    reflector. Choose a VPA near the

    middle of your probe.

    Adjust range as needed toensure you will see the 2

    reflectors you have

    chosen.

  • 8/12/2019 Calibrations Linear 2 0 SW

    6/24

    6

    Set Depth 1 and 2 Select depth from echo

    type

    Set depth 1 and depth

    2 to your reflector

    depths.

  • 8/12/2019 Calibrations Linear 2 0 SW

    7/24

    7

    Set Gate A on Depth 1 Peak the reflector at

    the chosen VPA.

    Adjust the gate to

    encompass the

    reflector.Get position.

  • 8/12/2019 Calibrations Linear 2 0 SW

    8/24

  • 8/12/2019 Calibrations Linear 2 0 SW

    9/24

    9

    Accept or Restart View the velocity

    determination

    If it is sensible accept

    the calibration, if not

    restart.

  • 8/12/2019 Calibrations Linear 2 0 SW

    10/24

    10

    Wedge Delay-Depth ReflectorWedge Delay will enter compensation for

    the travel of the beams through the wedge

    medium and account for the various exit

    points or wedge wear.

    Wedge delay needs only 1 distinct reflector-a known depth reflector will be used here.

    After wedge calibrations, accuratemeasurements can be taken.

  • 8/12/2019 Calibrations Linear 2 0 SW

    11/24

    11

    Wedge Delay Calibration-Depth Enter the calibration

    menu

    Select Ultrasound and

    Wedge Delay-Start

    Select depth and input

    depth to the reflector

    and a tolerance

  • 8/12/2019 Calibrations Linear 2 0 SW

    12/24

    12

    Set Section

    Decide if you want to

    calibrate the whole

    angular sweep orbreak it up into

    separate calibrations-

    for most setups with a

    good calibration block

    you can calibrate the

    whole sweep. Hit Next

  • 8/12/2019 Calibrations Linear 2 0 SW

    13/24

    13

    Set Gate A on Depth A

    Put the probe on the block so

    you can see the chosen

    depth reflector.Adjust the gate

    start/width/threshold to

    encompass the reflector.Move the probe forward

    and backwards to ensure

    gate is encompassing

    reflector through all angles

    and there are no other

    reflectors that will affect

    calibration.

  • 8/12/2019 Calibrations Linear 2 0 SW

    14/24

    14

    Calibrate and Accept

    Move the probe forwardand backward applyingeven pressure andavoiding skewing right orleft.

    Move in both directions

    until the green line is builtacross the whole angularrange. Adjust gain asneeded.

    Hit calibrate.

  • 8/12/2019 Calibrations Linear 2 0 SW

    15/24

    15

    Accept or Restart

    After hitting calibrate-

    rerun the probe

    through the angularrange verifying you are

    within the set

    tolerance.

    Accept or Restart.

  • 8/12/2019 Calibrations Linear 2 0 SW

    16/24

    16

    Sensitivity Calibrations

    Sensitivity calibrations equalize the

    sensitivity (amplitude response) from a

    given reflector through all the VPAs

    (beams).

    This will insure no matter what VPA thereflector is seen with the % FSH is the same

    for rejection or detection purposes as wellas for the amplitude based color coded

    imaging selections.

  • 8/12/2019 Calibrations Linear 2 0 SW

    17/24

    17

    Getting Started

    Choose a SDH and

    couple your probe

    above it Use a guide system to

    keep the probe from

    skewing side to side.

  • 8/12/2019 Calibrations Linear 2 0 SW

    18/24

    18

    Sensitivity Calibration

    Enter Calibration menu

    and select Sensitivity

    Choose a referencescreen height to

    equalize sensitivity to

    (80% FSH) and atolerance

  • 8/12/2019 Calibrations Linear 2 0 SW

    19/24

    19

    Set Section

    Choose whether to

    calibrate the whole

    angular sweep in onestep or break it up into

    separate calibrations-

    if your calibration blockhas adequate spacing

    between SDHs you

    can usually do the

    whole range.

  • 8/12/2019 Calibrations Linear 2 0 SW

    20/24

    20

    Set Gate A on Echo ASet the gate to encompass the

    SDH reflection. Adjust the

    start/width/threshold of the gate.

    Move the probe forward and

    backward to ensure the gate

    covers the reflector through allthe angles and is not on any

    other reflectors that may inhibit

    the calibration

  • 8/12/2019 Calibrations Linear 2 0 SW

    21/24

    21

    Set Compensation Gain

    Allows you to adjustthe gain as needed to

    ensure the reflector isseen properlythroughout all the

    angles In most cases no

    adjustment or an

    adjustment of theoverall gain (if overreference level FSH%)

    is needed.

  • 8/12/2019 Calibrations Linear 2 0 SW

    22/24

    22

    Calibrate and Accept

    Applying even pressure

    and not skewing right or

    left, move the probeforward and backwards to

    build the green line as the

    reflector is seen by the

    angles.

    Ensure the green line is

    on-screen across the

    whole screen-not to low orover the tolerance lines.

  • 8/12/2019 Calibrations Linear 2 0 SW

    23/24

    23

    Calibrate and Accept

    Press Calibrate

    Verify the calibration

    by repeating theforward and backwardmovement of the probe

    The green line shouldfall within the definedtolerance lines across

    all the angles.Accept or Recalibrate

    as needed.

  • 8/12/2019 Calibrations Linear 2 0 SW