CAPNOSTAT ® 5 MAINSTREAM CO 2 SENSOR 21st Century CO 2 Technology for mainstream CO 2 monitoring in critically ill, intubated patients. The CAPNOSTAT 5 Sensor is the ideal capnography solution for your patient monitoring system. It provides technologically advanced measurement of End-Tidal Carbon Dioxide (ETCO 2 ), respiration rate, and a clear, accurate capnogram at all respiratory rates up to 150 breaths per minute. The CAPNOSTAT 5 Sensor is: • Flexible • Compact • Reliable • Affordable The CAPNOSTAT 5 can be combined with other components and software to provide respiratory monitoring capabilities that are comprehensive, yet remarkably flexible. IT’S ALL INSIDE! As the recognized global leader in capnography, Respironics has been providing innovative and cost- effective products to OEM customers for over 18 years. Now, our engineers have developed a new miniature PC board that puts all measurement and analysis functions right inside the sensor head. This keeps costs and space requirements down by elim- inating the need for a PC board inside your monitor. The data output is a simple RS232 signal, so it’s a snap to integrate the new CAPNOSTAT 5 Sensor into your patient monitoring system. We also use the very latest Digital Signal Processor (DSP) for reliable and flexible analog-to-digital conversion. 800.243.3444 OPTION 3 WWW.OEM.RESPIRONICS.COM 203.697.6488

CAPNOSTAT 5 MAINSTREAM CO SENSOR - pennvet.com

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

CAPNOSTAT® 5 MAINSTREAM CO2 SENSOR21st Century CO2 Technology™ for mainstream CO2 monitoring

in critically ill, intubated patients.

The CAPNOSTAT 5 Sensor is the ideal capnography

solution for your patient monitoring system. It

provides technologically advanced measurement of

End-Tidal Carbon Dioxide (ETCO2), respiration rate,

and a clear, accurate capnogram at all respiratory

rates up to 150 breaths per minute.

The CAPNOSTAT 5 Sensor is:

• Flexible • Compact

• Reliable • Affordable

The CAPNOSTAT 5 can be combined with other

components and software to provide respiratory

monitoring capabilities that are comprehensive, yet

remarkably flexible.

IT’S ALL INSIDE!As the recognized global leader in capnography,

Respironics has been providing innovative and cost-

effective products to OEM customers for over

18 years. Now, our engineers have developed a new

miniature PC board that puts all measurement and

analysis functions right inside the sensor head. This

keeps costs and space requirements down by elim-

inating the need for a PC board inside your monitor.

The data output is a simple RS232 signal, so it’s a

snap to integrate the new CAPNOSTAT 5 Sensor

into your patient monitoring system. We also use

the very latest Digital Signal Processor (DSP) for

reliable and flexible analog-to-digital conversion.

8 0 0 . 2 4 3 . 3 4 4 4 O P T I O N 3 W W W . O E M . R E S P I R O N I C S . C O M 2 0 3 . 6 9 7 . 6 4 8 8

Transducer Type Mainstream CO2 Sensor

Principle of Operation Non-dispersive infrared (NDIR) single beam optics, dual wavelength, no moving parts

Initialization Time Capnogram, displayed in less than 15 seconds at an ambient temperature of 25ºC, full specifications within 2 minutes

CO2 Measurement Range 0 to 150 mm Hg, 0 to 19.7%, 0 to 20 kPa (at 760 mm Hg) Barometric Pressure supplied by host

Rise Time Less than 60 ms - Adult Reusable or Single-Patient-Use Airway Adapter Less than 60 ms - Infant Reusable or Single-Patient-Use Airway Adapter

CO2 Resolution 0.1 mm Hg 0 to 69 mm Hg0.25 mm Hg 70 to 150 mm Hg

CO2 Accuracy 0 – 40 mm Hg ±2 mm Hg41 – 70 mm Hg ±5% of reading71 – 100 mm Hg ±8% of reading101 – 150 mm Hg ±10% of reading

CO2 Stability Short Term Drift: Drift over four hours shall not exceed 0.8 mm Hg max.Long Term Drift: Accuracy specification will be maintained over a 120-hour period

CO2 Noise RMS noise of the sensor shall be less than or equal to 0.25 mm Hg at 7.5% CO2

Respiration Rate Range 0 to 150 Breaths Per Minute (BPM)

Respiration Rate Accuracy ±1 breath

Compensations Barometric Pressure 400 mm Hg to 850 mm Hg (Supplied) Operator selectable O2, N2O, HE and Agent compensation.

Calibration No routine user calibration required. An airway adapter zero is required when changing to a different style of airway adapter.

Airway Adapters Single-patient-use or reusable, < 5 cc dead space (adult), < 1 cc dead space infantAdapter taper meets ISO 5356-1

Voltage Requirements +5.0 VDC, 1.125 W Typical, 1.5 W Maximum.

Power Rating Rated input: 1.1 Watts typical. Steady StateUp to 1.5 Watts maximum on power up (warm up)

Interconnection Standard – Lemo Redel 8-pin plastic, Material PSUSensor Plug: PAB.M0.8GL.AC39GZBend Relief: GMA.1B.030.DJHost Receptacle: PKB.M0.8GL.LJ

Note: LoFlo and Pin out: CAPNOSTAT are 1 VA 5.0Vinterchangeable with 2 Shield Shieldthe host monitor. Either 3 DGND Digital returndevice may be connected 4 VSRC 5.0Vand used via the RS232 5 TxD Serial data from CAPNOSTATconnector. 6 RxD Serial data from Host

7 AGND Analog return8 SYNC Waveform Synchronization

Temperature and Humidity Operating: 0 to 45ºC, 10 to 90% RH, non-condensingStorage: -40 to 70ºC, <90% RH, non-condensing

Water Resistance IPX4 – Splash-proof (sensor head only)

Shock Impact EN60068-2-6 Sinusoidal VibrationEN60068-2-27 ShockEN60068-2-64 Random VibrationAble to withstand repeated 6-foot drops onto tiled floor while operating

Data Interface RS232 compatible, bi-directional, 19200 Baud, Standard N-8-1.

Data Output CO2 gas concentration (mm Hg), End-tidal CO2, Inspired CO2, Respiratory RateGas and barometric pressure compensated when supplied by host

Regulatory Designed to meet IEC 60601-1-2, EN55011 – CISPIR 11 Class B (Radiated and Conductive Emissions), IEC 61000-4-2 Electrostatic Discharge Immunity, IEC 61000-4-3 Radiated Immunity, Designed to comply with 93/42/EEC (MDD CE Marking), FDA Standards, ASTM F1456-01 Minimum Performance and Safety Requirements for Capnometers and ISO21647. Medical Electrical Equipment performance requirements for the basic safety and essential performance of respiratory gas monitors.

Specifications are subject to change without notice.

CAPNOSTAT® 5 MAINSTREAM CO2 SENSOR – SPECIFICATIONS

GO MAINSTREAM

The CAPNOSTAT® 5 Sensor is designed for

mainstream measurements using sophisti-

cated infrared absorption spectroscopy

to measure ETCO2. With mainstream,

measurements are taken directly at the

patient’s airway, so response is faster and

there is less chance of erroneous, artifact

data. The analyzer portion of the sensor

cannot be contaminated with patient

secretions, and there are no pumping

or pneumatic components to replace.

Maintenance requirements and overall

cost-of-ownership are minimized. Our

exclusive family of airway adapters are

easy to use and durable enough even for

long-term ventilator patients. Yet, they are

inexpensive and disposable.

The CAPNOSTAT 5 Sensor can be inte-

grated with other components, software

packages and accessories, making it

easy for you to configure your patient

monitoring systems with state-of-the-art

respiratory management and analysis

technology now and into the future.

COMPLETE OEM SOLUTIONS

The CAPNOSTAT 5 is just one of many

solutions we offer to our OEM customers.

Respironic’s comprehensive engineering

team customizes innovative products to

meet customer requirements and provides

clinical and marketing support to help

meet the growing needs of your business.

Customer Service: 1-800-243-3444, option 3 or 203-697-6488www.oem.respironics.com

Respironics, CAPNOSTAT and LoFlo are registered trademarks and 21st Century CO2

Technology is a trademark of Respironics, Inc. and its affiliates.

CAUTION: U.S. federal law restricts this device to sale by or on the order of a physician.

©2007 Respironics, Inc. and its affiliates. All rights reserved.

Geyer KB 7/10/07 MCI 4101194