GE Healthcare Multiparameter hemodynamic module with simultaneous measurement of 3-, 7- or 12-lead ECG in S/5 modular monitors and 3-, 7-, 8- or 12-lead ECG in CARESCAPE* modular monitors, both with ST segment analysis, respiration, pulse oximetry, two temperatures, non-invasive blood pressure and optional two invasive blood pressures (in E-PSMP). ECG features Continuous 3-, 7-, 8-, and 12-lead ECG with 3, 5, 6, or 10 leadwires and ST segment analysis with CARESCAPE modular monitors Continuous 3-, 7- and 12-lead ECG measurement with 3, 5 or 10 leadwires and ST segment analysis with S/5 modular monitors Manual storage of up to six reference QRS complexes Manual or automatic creation of ST snapshots with CARESCAPE modular monitors‘ ST alarms capability Compatible with the wide range of Multi-Link* cables and leadwire sets Lethal arrhythmia analysis is standard with full arrhythmia recognition as a monitor software option Excellent signal quality due to advanced electrical interference filtering QT/QTc values are calculated with 5, 6 or 10 leadwires with CARESCAPE modular monitors 12SL* and 12RL* 12-lead ECG analysis available with CARESCAPE modular monitors SpO 2 features Specified accuracy even with moving and restless patients The scale of plethysmographic waveform can be fixed or set to autoscale according to the user selection Variety of SpO 2 sensors for various applications Non-invasive blood pressure (NIBP) features Direct keys on the module for starting or stopping a single measurement, or initiating or stopping a NIBP cycle Venous stasis for easy cannulation Visual indicator for elapsed time from previous measurement Automatic BP hose detection for adult and infant cuffs Respiration rate features Adjustable trigger level with breath indicator Specific slow waveform speed to see different respiratory patterns Invasive blood pressure features Selectable labels for invasive blood pressures Waveforms displayed with combined or individual scales A dedicated wedge pressure (PCWP) measurement display view Adjustable cursor to detect changes in the invasive pressure waveform Patient Side Module, E-PSM, E-PSMP For continuous hemodynamic monitoring

CARESCAPE Module (e)-psm(p) Specs

  • Upload
    acertys

  • View
    222

  • Download
    3

Embed Size (px)

DESCRIPTION

 

Citation preview

Page 1: CARESCAPE Module (e)-psm(p) Specs

GE Healthcare

Multiparameter hemodynamic module with simultaneous measurement of 3-, 7- or 12-lead ECG in S/5 modular monitors and 3-, 7-, 8- or 12-lead ECG in CARESCAPE* modular monitors, both with ST segment analysis, respiration, pulse oximetry, two temperatures, non-invasive blood pressure and optional two invasive blood pressures (in E-PSMP).

ECG features• Continuous3-,7-,8-,and12-leadECGwith3,5,6,or

10 leadwires and ST segment analysis with CARESCAPE modular monitors

• Continuous3-,7-and12-leadECGmeasurementwith 3, 5 or 10 leadwires and ST segment analysis with S/5 modular monitors

• ManualstorageofuptosixreferenceQRScomplexes

• ManualorautomaticcreationofSTsnapshotswithCARESCAPE modular monitors‘ ST alarms capability

• CompatiblewiththewiderangeofMulti-Link*cablesandleadwire sets

• Lethalarrhythmiaanalysisisstandardwithfullarrhythmia recognition as a monitor software option

• Excellentsignalqualityduetoadvancedelectrical interference filtering

• QT/QTcvaluesarecalculatedwith5,6or10leadwireswith CARESCAPE modular monitors

• 12SL*and12RL*12-leadECGanalysisavailablewithCARESCAPE modular monitors

SpO2 features• Specifiedaccuracyevenwithmovingandrestless

patients

• Thescaleofplethysmographicwaveformcanbefixed or set to autoscale according to the user selection

• VarietyofSpO2 sensors for various applications

Non-invasive blood pressure (NIBP) features• Directkeysonthemoduleforstartingorstoppinga

single measurement, or initiating or stopping a NIBP cycle

• Venousstasisforeasycannulation

• Visualindicatorforelapsedtimefromprevious measurement

• AutomaticBPhosedetectionforadultandinfantcuffs

Respiration rate features• Adjustabletriggerlevelwithbreathindicator

• Specificslowwaveformspeedtoseedifferent respiratory patterns

Invasive blood pressure features• Selectablelabelsforinvasivebloodpressures

• Waveformsdisplayedwithcombinedorindividualscales

• Adedicatedwedgepressure(PCWP)measurement display view

• Adjustablecursortodetectchangesintheinvasive pressure waveform

Patient Side Module, E-PSM, E-PSMPFor continuous hemodynamic monitoring

Page 2: CARESCAPE Module (e)-psm(p) Specs

Technical Data

Directfunctionkeys

AutoOn/Off Starts/stopsNIBPcycle

Start/Cancel Starts/stops single NIBP

Zero P1 Zeros invasive blood pressure P1

Zero P2 Zeros invasive blood pressure P2

ST segment analysis

ST segment measured and displayed simultaneously for up to 12 ECG leads

ST level range

CARESCAPE modular monitors -20 to 20 mm

S/5 modular monitors -9 to 9 mm

Displayresolution 0.1mm

Averaging Calculatedfrom8QRScomplexes

ST trends

CARESCAPE modular monitors up to 72 hours

S/5 modular monitors up to 72 hours (ICU) up to 24 hours (ANE)

QRSTcomplexstorage Onecomplexstoredevery 10 minutes and manual storing of6complexes

Alarms

CARESCAPE modular monitors Adjustablehighandlowalarm

limits for anterior, inferior, lateral and individual leads

S/5modularmonitors Adjustablehighandlowalarm limits for anterior, inferior and lateral leads

Impedance respiration

Measurement method Transthoracic impedance

Measuringrange Impedance,0.2to32Ohm

Maxbaseimpedance 5kOhm

Respiration range 4 to 120 breath/min

Accuracy ±5% or ±5 breath/min

Update 10 s

Displayresolution 1breath/min

Sweepspeed(trace) Slow0.625mm/s,fast6.25mm/s

Alarms Adjustablehighandlowalarmlimits for respiration rate, apnea

Note: Impedance respiration measurement is intended for patients over three years old.

WARNING:CARESCAPE modular monitors: Apnea events -- The monitor may not detect all episodes ofinadequatebreathing,nordoesitdistinguishbetweencentralobstructiveandmixedapnea events.

S/5 monitors: These devices are not apnea monitor systems intended to alarm primarily upon the cessation of breathing. In central apnea it indicates an alarm after a predeterminedtimesincethelastbreathdetection.Donotattempttousethesedevicesfor detecting obstructive or mixed apneas, since respiration movements and impedancevariations may continue in these cases.

ECG

Measurement method

CARESCAPE modular monitors 3, 7, 8 and 12 lead ECG with 3, 5, 6or10leadwires

S/5 modular monitors 3, 7 and 12 lead ECG with 3, 5 or 10 leadwires

ECG Lead selection

It is possible to monitor simultaneously one or more of the following leads:

CARESCAPE modular monitors I,II,III,aVL,aVF,aVR,V1,V2,V3, V4,V5,V6,dV2,dV3,dV4,dV6

S/5modularmonitors I,II,III,aVL,aVF,aVR,V1,V2,V3, V4,V5,V6

NOTE:AllspecificationsarerelevanttoS/5andCARESCAPEmodularmonitors,unlessotherwisenoted.

Page 3: CARESCAPE Module (e)-psm(p) Specs

Heart rate

Measurement range

CARESCAPE modular monitors 30 to 300 bpm

S/5 modular monitors 30 to 250 bpm

Measurement accuracy ±5% or ±5 bpm

Displayresolution 1bpm

Displayaveragingtime

CARESCAPE module monitors 10 s median values

S/5 modular monitors Average of 12 s median values, or 5 s, depending on monitor

Displayupdatetime

CARESCAPE modular monitors 1 s

S/5 modular monitors 5 s or 1 s, depending on monitor

Alarms

CARESCAPE modular monitors

S/5 modular monitors

Asystole Asystole

Leads-offdetection Leads-offdetection,bradycardia, tachycardia

Adjustablehighandlowalarm limits for heart rate

Adjustablehighandlowalarmlimits for heart rate

VFib VFib

VTach VTachy

RapidVT1

Optional: Optional:

VT>2 VRun>3

Couplet VCouplet

Bigeminy VBigeminy

Trigeminy VTrigeminy

R on T RonTPVC

Missing Beat Missing Beat

MultifocalPVCs MultifocalPVCs

FrequentPVCs FrequentPVCs

FrequentSVC1

Idioventricular rhythm 1

LongR-R1

SVTachy1

Atrial fibrillation 1

1Onlysoftwareversion07andlater.

Filter modes

CARESCAPE modular monitors 50Hzpowersupplyfrequency: Monitoring filter: 0.05 to 32 Hz Diagnosticfilter: 0.05to150Hz

60Hzpowersupplyfrequency: Monitoring filter: 0.05 to 40 Hz Diagnosticfilter:0.05to150Hz

Moderate filter: 0.05 to 22 Hz Maximum filter: 5 to 25 Hz

S/5modularmonitors 50Hzpowersupplyfrequency: Monitoring filter: 0.5 to 30 Hz ST filter: 0.05 to 30 Hz Diagnosticfilter: 0.05to150Hz

60Hzpowersupplyfrequency: Monitoring filter: 0.5 to 40 Hz ST filter: 0.05 to 40 Hz Diagnosticfilter: 0.05to150Hz

Defibrillationprotection 5000V,360J

Sweep speed

CARESCAPE modular monitors 6.25,12.5,25or50mm/s

S/5 modular monitors 12.5, 25 or 50 mm/s (15“ screen)

Gain

CARESCAPE modular monitors 0.5x=5mm/mV 1x=10mm/mV 2x=20mm/mV 4x=40mm/mV

S/5modularmonitors 0.2to5.0cm/mV

Pacemakerpulsedetection

CARESCAPE modular monitors Detectionlevel ±2mVto±700mV Pulse duration 0.1 to 2 ms

S/5modularmonitors Detectionlevel ±2mVto±700mV Pulse duration 0.5 to 2 ms

Allowableoffset ±1Vdc

CMRR

CARESCAPE modular monitors ≥ 90 dB minimum at 50 Hz

S/5 modular monitors ≥ 95 dB

Pacemakerdetectormaynotoperatecorrectlyduringtheuseofhigh-frequency(HF)surgicalequipment.ThedisturbancesofHFsurgicalequipmenttypicallycausefalsepositive pacer detection.

NOTE:AllspecificationsarerelevanttoS/5andCARESCAPEmodularmonitors,unlessotherwisenoted.

Page 4: CARESCAPE Module (e)-psm(p) Specs

Invasive blood pressure

Measurement method Invasive blood pressure is converted to an electrical signal by a pressure transducer. The signal is continuously displayed as a waveform and numeric value. The invasive pressure setup consisting of connecting tubing, pressure transducer, an intravenous bag of normal saline all connected together bystopcocks,isattachedtothecatheter. The pressure transducer is placed at the same level with the heart and electrically zeroed

Measurement range -40 to 320 mmHg

Measurement accuracy ±5% or ± 2 mmHg

Waveform display

Range -30 to 300 mmHg

Scales Adjustablein10mmHgincrements

Numerical display

Resolution 1 mmHg

Alarms Adjustablehighandlowalarm limits for systolic, diastolic and mean pressures

Pulse rate

CARESCAPE modular monitors HR from ART, ABP

S/5 modular monitors HR from ART

Measurement range 30 to 250 bpm

Displayresolution 1bpm

Accuracy ±5% or ± 5 bpm

Transducersensitivity 5µV/V/mmHg

Pressurefilter 0to22Hz(-3dB)adjustableupper limit 4 to 22 Hz

Temperature

2 temperature measurement channels

Measurement method The temperature is measured by a probe whose resistance varies when the temperature changes

Temperature units °C or °F

Numerical display 2 temperatures, differential T2-T1

Measurement range 10 to 45°C (50 to 113°F)

Measurement accuracy with sensors ±0.2°C (25 to 45°C) / ±0.4°F (77 to 113°F)

±0.3°C (10 to 24.9°C) / ±0.5°F(50to76.9°F)

Displayresolution ±0.1°C(±0.1°F)

Probes Use only GE Healthcare temperature probes or defibrillator-proof YSI 400 series probes

NIBP

Numerical display Systolic, diastolic and mean arterial pressure

Measurementprinciple Oscillometricwithlineardeflation

Measurementrange Adult: 25to260mmHg Child: 25 to 190 mmHg Infant: 15 to 140 mmHg

Measurement intervals

CARESCAPE modular monitors 1, 2, 2.5, 3, 4, 5, 10, 15, 20, 30 min, 1 h, 2 h, and 4 h or continous STAT mode

S/5 modular monitors 1, 2.5, 3, 5, 10, 15, 30 min, 1 h, 2 h, 4 h or continuous STAT mode

Alarms Adjustablehighandlowalarmlimits for systolic, diastolic, mean, cuff loose, cuff occlusion

NOTE:AllspecificationsarerelevanttoS/5andCARESCAPEmodularmonitors,unlessotherwisenoted.

Page 5: CARESCAPE Module (e)-psm(p) Specs

Typical measurement time Adult 23 seconds Infants 20 seconds

Safety features Adult Child Infant

Initialinflationpressure (mmHg) 170 150 120

Max.inflationpressure (mmHg) 280 200 145

Overpressurelimit 320 220 160

Max. measurement time 120 120 75

Note:NIBPmeasurementisintendedforpatientsweighingover5kg/11lb.

Pulseoximetry(SpO2)

Measurement method Red and infrared light absorption SpO2

Measurement and display range 0 to 100%

Calibration range 70 to 100%, calibrated against functional oxygen saturation

Measurement accuracy2 (Arms)3 70to100%:±2digits(%SpO2) ±3 digits during clinical patient motion < 70% unspecified

Displayresolution 1digit(%SpO2)

Displayaveragingtime

CARESCAPE modular monitors 3 s, 12 s

S/5modularmonitors Adjustablebeat-to-beat, 10 s, 20 s

Alarms AdjustablehighandlowalarmlimitsforSpO2 and pulse rate, probe off, no probe, no pulse

Pulse rate

Measurement and display range 30 to 250 bpm

Displayresolution 1bpm

Measurement accuracy 4 ±5% or ±5 bpm, whichever is greater

Plethysmographic pulse waveform

Scales 2, 5, 10, 20, 50 mod% or automatic scaling

2 Accuracyisbasedondeephypoxiastudieswithvolunteeredsubjectsduringnon-motion conditions over a wide range of arterial blood oxygen saturations as com-paredtoarterialbloodCO-Oximetry.

3 About two-thirds of pulse oximeter measurements can be expected to fall within ±ArmsofthevaluemeasuredbyaCO-oximeter.

4 MeasurementmodulesusingGEOhmedatechnologywithOxyTip+*sensorshavebeen validated for pulse rate accuracy over the specified range in bench top testing against a patient simulator.

Note:ForeachSpO2 accessory, refer to the instructions for use in the accessory packageforpatientweightlimits.

Environment specifications

Operatingconditions

Temperature 10 to 40°C (50 to 104°F)

Relative humidity 10 to 90% non-condensing

Storage conditions

Temperature -25 to 70°C (-13 to 158°F)

Relative humidity 10 to 90% non-condensing

Physical specifications

Dimensions(HxWxD) 14.0x5.1x13.2(17.1)cm (5.5x2.0x5.2(6.7)in)

Weight 0.6kg(1.3lb)

Warranty

Oneyear

NOTE:AllspecificationsarerelevanttoS/5andCARESCAPEmodularmonitors,unlessotherwisenoted.

Page 6: CARESCAPE Module (e)-psm(p) Specs

EMEA DOC1133430 4/12 (GlobalversionDOC1085720)

About GE HealthcareGE Healthcare provides transformational medical technologies and services that areshapinganewageofpatientcare.Ourbroadexpertiseinmedicalimagingandinformation technologies, medical diagnostics, patient monitoring systems, drug discovery, biopharmaceutical manufacturing technologies, performance improvement and performance solutions services help our customers to deliver better care to more people around the world at a lower cost. In addition, we partner with healthcare leaders, striving to leverage the global policy change necessary to implement a successful shift to sustainable healthcare systems.

Our“healthymagination”visionforthefutureinvitestheworldtojoinusonour journeyaswecontinuouslydevelopinnovationsfocusedonreducingcosts,increasingaccessandimprovingqualityaroundtheworld.HeadquarteredintheUnitedKingdom,GEHealthcareisaunitofGeneralElectricCompany(NYSE:GE).Worldwide,GEHealthcareemployees are committed to serving healthcare professionals and their patients in more than 100 countries. For more information about GE Healthcare, visit our website at www.gehealthcare.com.

© 2012 General Electric Company – All rights reserved. *GE,GEMonogram,12RL,12SL,CARESCAPEandOxyTip+aretrademarksofGeneralElectricCompany.

GeneralElectricCompanyreservestherighttomakechanges in specification and features shown herein, or discontinue the product described at any time without notice or obligation. Contact your GE representative for the most current information

GEHealthcareFinlandOy,aGeneralElectriccompany, doing business as GE Healthcare.

GE Healthcare, a division of General Electric Company

GE HealthcareP.O.Box900,FIN-00031GE,FinlandGEDirectUnitedKingdom:+44(0)8000329201 www.gehealthcare.com