2
www.julabo.de JULABO GmbH Eisenbahnstraße 45 77960 Seelbach / Germany Tel. +49 (0) 7823 51-0 Objective This case study tests the maximum low temperature of the PRESTO ® A40 with a 20 liters vacuum insulated glass reactor. The A40 is connected to the reactor via 2.0 m metal tubings. The A40 is programmed to cool down from +25 °C to maximum low temperature. Test Results See chart on back page: The A40 cooled the reactor from +25 °C down to maximum low temperature of -32 °C in 4 h 30 min. Case Study JULABO PRESTO ® A40 Cooling a 20 liters reactor from +25 °C to maximum low temperature -32 °C Case Study 2014.01 - 01 EN Page 1 of 2 Tip You can also use the robust Pt100 with PTFE coating. More tips on back page >> Test Conditions JULABO unit JULABO Presto A40 Cooling power +20 °C 1.2 kW 0 °C 0.9 kW -20 °C 0.6 kW Heating capacity 2.7 kW Band limit No Flow pressure 0.40 bar Bath fluid JULABO Thermal HL40 Reactor Triple walled 20 liters glass reactor (Asahi) filled with 18 liter JULABO Thermal HL40 Jacket volume 7.0 l Control External (ICC) Environment Room temperature +20 °C Humidity 45 % Voltage 230 V / 50 Hz

Case Study - JULABO€¦ · JULABO GmbH Eisenbahnstraße 45 77960 Seelbach / Germany Tel. +49 (0) 7823 51-0 Objective This case study tests the maximum low temperature of the

  • Upload
    doandan

  • View
    219

  • Download
    1

Embed Size (px)

Citation preview

Page 1: Case Study - JULABO€¦ · JULABO GmbH Eisenbahnstraße 45 77960 Seelbach / Germany Tel. +49 (0) 7823 51-0 Objective This case study tests the maximum low temperature of the

www.julabo.de

JULABO GmbH Eisenbahnstraße 4577960 Seelbach / Germany Tel. +49 (0) 7823 51-0

ObjectiveThis case study tests the maximum low temperature of the PRESTO® A40 with a 20 liters vacuum insulated glass reactor. The A40 is connected to the reactor via 2.0 m metal tubings. The A40 is programmed to cool down from +25 °C to maximum low temperature.

Test ResultsSee chart on back page: The A40 cooled the reactor from +25 °C down to maximum low temperature of -32 °C in 4 h 30 min.

Case Study

JULABO PRESTO® A40Cooling a 20 liters reactor from

+25 °C to maximum low temperature -32 °C

Case Study 2014.01 - 01 EN Page 1 of 2

TipYou can also use the robust Pt100 with PTFE coating.

More tips on back page >>

Test ConditionsJULABO unit JULABO Presto A40Cooling power +20 °C 1.2 kW 0 °C 0.9 kW -20 °C 0.6 kWHeating capacity 2.7 kWBand limit NoFlow pressure 0.40 barBath fluid JULABO Thermal HL40Reactor Triple walled 20 liters glass reactor (Asahi) filled with 18 liter JULABO Thermal HL40Jacket volume 7.0 lControl External (ICC)

EnvironmentRoom temperature +20 °CHumidity 45 %Voltage 230 V / 50 Hz

Page 2: Case Study - JULABO€¦ · JULABO GmbH Eisenbahnstraße 45 77960 Seelbach / Germany Tel. +49 (0) 7823 51-0 Objective This case study tests the maximum low temperature of the

JULABO GmbH Eisenbahnstraße 4577960 Seelbach / Germany Tel. +49 (0) 7823 51-0

www.julabo.de

0 1 1,50,5

°C

+20

-20

0

0 4 53 h1 2

-32

Cool-down time 4 h 30 min

Measured with EasyTEMP Professional

Case Study 2013.11 - 01 EN Page 2 of 2

Setpoint

Temperature in reactor‘s interior

Temperature in reactor‘s jacket

TipMake use of the option to regulate the pump pressure. You can define the desired pressure in the PRESTO® settings.

TipThe Ethernet inter- face permits full access to all ope-rational functions of the PRESTO®.

lowest temperature