Template Revision D
CASE 8000 DPMFRM-022Preventive Maintenance Inspection Report 30 Jun 1999
Page 1 of 4
Customer______________________________ Customer Number______________________ Date___________
FE ___________________________________ FE ID________ Call Number______________________________
Equipment Serial Number ________________ Software Revision _________
Configuration ()
___ 12SL ___ Remote Viewing ___ Full Disclosure
___ Data Storage on File Server
Tools
1. Leakage tester 3. Multimeter 5. ESD Mat and Wrist Strap
2. Standard hand tools 4. Service Manual (p/n 415047-001)
Visual Inspection ()
Inspect the following for excess wear and/or any visual signs of damage.
1. __Coiled patient cable 3.____AC power cord
2. __Patient leadwires 4.____Chassis hardware
Cleaning ()
1. ___ Clean all internals of dust 2. ___ Clean external surfaces
Calibration ()
1. ___ Writer Tests 2. ___ Key Pad Test
Electrical Safety Checks ()Refer to PM Application Form DPMAPP-001-4 class __ for leakage standards for specific voltages.
1. ___ Wall Receptacle Test
2. ___ Ground Continuity (Impedence of protected connection) Test(.1or .2w/power cord)
3. Leakage Open Closed
Normal Reversed Normal Reversed
Ground(Earth) wire leakage to ground test ___ ___ NA NA
Chassis(Enclosure) leakage to ground test ___ ___ ___ ___
Source Risk(Patient cable leakage to ground) Leakage ___ ___ ___ ___
Sink Risk(Patient cable leakage into patient leads) Leakage NA NA ___ ___
Checkout Procedure ()
A.___ Attach ECG simulator, press the RHYTHM key and check the report for proper data, paper tracking and clarity.
B.___ Select a protocol and start treadmill(or ergometer) to verify speed and grade( or workload for ergometers) are operating
properly.
C.___ Depress the Emergency stop button(if installed) while the treadmill(if installed) is running at 3 m.p.h. to verify the
treadmill stops correctly.
D.___ Depress the Treadmill Stop button on the CASE 8000 stress keyboard while the treadmill(if installed) is running at 3
m.p.h. and verify the treadmill stops correctly.
E.___ Perform the Keyboard Test using the Service Screen.
F.___ Perform Writer Diagnostics from the Service Screen.
8/11/2019 casepm.pdf
2/4
DPMFRM-022
Page 2 of 4
Template Revision D
Comments
Briefly describe all repairs/adjustments made and list all parts replaced:
___________________________________________________________________________________________________
___________________________________________________________________________________________________
8/11/2019 casepm.pdf
3/4
DPMFRM-022
Page 3 of 4
Template Revision D
FRU Checkout Procedure
FRU
Description*
Tools Visual
Inspection
Cleaning Calibration Electrical
Safety Tests
Checkout
Procedure
CASE 8000 Stress Test
Main CPU Mother Pcb
1,3,5 - - 1 3 A thru F
Acquistion I/F Pcb 1,3,5 1,2 - 1 3 A thru F
Network Pcb 1,3,5 - - 1 3 A thru F
Analog I/O Panel 1,3,5 - - 1 3 A thru F
Writer CPU Pcb 1,3,5 - - 1 3 A thru F
Writer Power Supply 1,3,5 - - 1 3 A thru F
Printhead 1,3,5 - - 1 3 A thru F
Hard Drive 1,3,5 - - 1 3 A thru F
Floppy Drive 1,3,5 - - 1 3 A thru F
Video Pcb 1,3,5 - - 1 3 A thru F
Software - - - 1,2 A thru F
Hardware Upgrades
(Including Firmware)
Perform complete PM
Supplies Perform applicable Checkout Procedure(s)
Non-Listed FRUs Perform complete PM
No Parts Replaced Perform applicable Checkout Procedure(s)
* The FRU Checkout procedure for any listed FRU also applies to its internal PCBs and parts.