casepm.pdf

Embed Size (px)

Citation preview

  • 8/11/2019 casepm.pdf

    1/4

    Template Revision D

    CASE 8000 DPMFRM-022Preventive Maintenance Inspection Report 30 Jun 1999

    Page 1 of 4

    Customer______________________________ Customer Number______________________ Date___________

    FE ___________________________________ FE ID________ Call Number______________________________

    Equipment Serial Number ________________ Software Revision _________

    Configuration ()

    ___ 12SL ___ Remote Viewing ___ Full Disclosure

    ___ Data Storage on File Server

    Tools

    1. Leakage tester 3. Multimeter 5. ESD Mat and Wrist Strap

    2. Standard hand tools 4. Service Manual (p/n 415047-001)

    Visual Inspection ()

    Inspect the following for excess wear and/or any visual signs of damage.

    1. __Coiled patient cable 3.____AC power cord

    2. __Patient leadwires 4.____Chassis hardware

    Cleaning ()

    1. ___ Clean all internals of dust 2. ___ Clean external surfaces

    Calibration ()

    1. ___ Writer Tests 2. ___ Key Pad Test

    Electrical Safety Checks ()Refer to PM Application Form DPMAPP-001-4 class __ for leakage standards for specific voltages.

    1. ___ Wall Receptacle Test

    2. ___ Ground Continuity (Impedence of protected connection) Test(.1or .2w/power cord)

    3. Leakage Open Closed

    Normal Reversed Normal Reversed

    Ground(Earth) wire leakage to ground test ___ ___ NA NA

    Chassis(Enclosure) leakage to ground test ___ ___ ___ ___

    Source Risk(Patient cable leakage to ground) Leakage ___ ___ ___ ___

    Sink Risk(Patient cable leakage into patient leads) Leakage NA NA ___ ___

    Checkout Procedure ()

    A.___ Attach ECG simulator, press the RHYTHM key and check the report for proper data, paper tracking and clarity.

    B.___ Select a protocol and start treadmill(or ergometer) to verify speed and grade( or workload for ergometers) are operating

    properly.

    C.___ Depress the Emergency stop button(if installed) while the treadmill(if installed) is running at 3 m.p.h. to verify the

    treadmill stops correctly.

    D.___ Depress the Treadmill Stop button on the CASE 8000 stress keyboard while the treadmill(if installed) is running at 3

    m.p.h. and verify the treadmill stops correctly.

    E.___ Perform the Keyboard Test using the Service Screen.

    F.___ Perform Writer Diagnostics from the Service Screen.

  • 8/11/2019 casepm.pdf

    2/4

    DPMFRM-022

    Page 2 of 4

    Template Revision D

    Comments

    Briefly describe all repairs/adjustments made and list all parts replaced:

    ___________________________________________________________________________________________________

    ___________________________________________________________________________________________________

  • 8/11/2019 casepm.pdf

    3/4

    DPMFRM-022

    Page 3 of 4

    Template Revision D

    FRU Checkout Procedure

    FRU

    Description*

    Tools Visual

    Inspection

    Cleaning Calibration Electrical

    Safety Tests

    Checkout

    Procedure

    CASE 8000 Stress Test

    Main CPU Mother Pcb

    1,3,5 - - 1 3 A thru F

    Acquistion I/F Pcb 1,3,5 1,2 - 1 3 A thru F

    Network Pcb 1,3,5 - - 1 3 A thru F

    Analog I/O Panel 1,3,5 - - 1 3 A thru F

    Writer CPU Pcb 1,3,5 - - 1 3 A thru F

    Writer Power Supply 1,3,5 - - 1 3 A thru F

    Printhead 1,3,5 - - 1 3 A thru F

    Hard Drive 1,3,5 - - 1 3 A thru F

    Floppy Drive 1,3,5 - - 1 3 A thru F

    Video Pcb 1,3,5 - - 1 3 A thru F

    Software - - - 1,2 A thru F

    Hardware Upgrades

    (Including Firmware)

    Perform complete PM

    Supplies Perform applicable Checkout Procedure(s)

    Non-Listed FRUs Perform complete PM

    No Parts Replaced Perform applicable Checkout Procedure(s)

    * The FRU Checkout procedure for any listed FRU also applies to its internal PCBs and parts.

  • 8/11/2019 casepm.pdf

    4/4

    DPMFRM-022

    Page 4 of 4

    Template Revision D