2
The TI500 Globe-Trotter™ Neonatal Transport System from Dräger Medical combines all the technology of the TI500 Transport System with additional features for an improved level of care. FEATURES INTEGRATED SUPPORT The TI500 Globe-Trotter Neonatal Trans- port System from Dräger is designed to transport infants in a thermal neutral environment and also provides integrated ventilatory support. Your system can be configured with a selection of ventilators and oxygen analyzers to meet your needs. FRONT-MOUNTED CONTROLS In addition to the incubator controls, all ventilator, tank gauges, and gas-related controls are front mounted for quick access and heightened visibility. INTERNAL TANK STORAGE Gas tanks are housed beneath the incubator in the TI500 Globe-Trotter Neonatal Transport System frame so they can easily be carried together with the incubator unit. Regulators, yoke fittings, and flowmeters are all built in, eliminating external tubing. AIR COMPRESSOR OPTION The TI500 Globe-Trotter Neonatal Trans- port System from Dräger is the only transport system that has the option of an integrated air compressor. You can deliver medical grade air while eliminating the need to carry air tanks. AUXILIARY FLOW METER An auxiliary flowmeter is standard on the TI500 Globe-Trotter Neonatal Transport System with an additional hand circuit to allow manual resuscitation if necessary. This offers the ability to ventilate twins in one incubator. OTHER KEY ADVANTAGES – Facilities for fluid delivery and vital signs monitoring – Strain relief blocks to prevent accidental extubation – Capacity for recharging auxiliary support equipment batteries – Available on a choice of stands TI500 Globetrotter™ D-1591-2009

CEGA Enhances Transfer Capabilities For Neonates

Embed Size (px)

DESCRIPTION

The new incubator, which is mounted on a ParAid sled, offers a complete neonatal transport solution. Among its features is a fully integrated ventilator for intubated neonates. It also offers head access and a slide out mattress to allow easy access for intubation or CPR. It replaces CEGA’s current Drager-5400 and is the first incubator of its kind to be offered by a UK fixed wing air ambulance provider.

Citation preview

Page 1: CEGA Enhances Transfer Capabilities For Neonates

The TI500 Globe-Trotter™ Neonatal Transport System fromDräger Medical combines all the technology of the TI500 Transport System with additional features for an improvedlevel of care.

FEATURES INTEGRATED SUPPORT

The TI500 Globe-Trotter Neonatal Trans-port System from Dräger is designed totransport infants in a thermal neutralenvironment and also provides integratedventilatory support. Your system can beconfigured with a selection of ventilatorsand oxygen analyzers to meet your needs.

FRONT-MOUNTED CONTROLS

In addition to the incubator controls, allventilator, tank gauges, and gas-relatedcontrols are front mounted for quickaccess and heightened visibility.

INTERNAL TANK STORAGE

Gas tanks are housed beneath theincubator in the TI500 Globe-TrotterNeonatal Transport System frame so theycan easily be carried together with theincubator unit. Regulators, yoke fittings,and flowmeters are all built in, eliminatingexternal tubing.

AIR COMPRESSOR OPTION

The TI500 Globe-Trotter Neonatal Trans-port System from Dräger is the onlytransport system that has the option of anintegrated air compressor. You can delivermedical grade air while eliminating theneed to carry air tanks.

AUXILIARY FLOW METER

An auxiliary flowmeter is standard on theTI500 Globe-Trotter Neonatal TransportSystem with an additional hand circuit toallow manual resuscitation if necessary.This offers the ability to ventilate twins inone incubator.

OTHER KEY ADVANTAGES

– Facilities for fluid delivery and vitalsigns monitoring

– Strain relief blocks to prevent accidental extubation

– Capacity for recharging auxiliary support equipment batteries

– Available on a choice of stands

TI500 Globetrotter™

D-1

591-

2009

Page 2: CEGA Enhances Transfer Capabilities For Neonates

TI500 GLOBETROTTER™02 |

TECHNICAL SPECIFICATIONS DRÄGER TI500 GLOBETROTTER™

Physical Attributes

Height 23 in (58.4 cm)

Length 51 in (129.5 cm)

Width 22.5 in (56.5 cm)

Weight without compressor 158 lbs (71.6 kg)1

Weight with compressor 185 lbs (83.9 kg)

Electrical Specifications

AC Power requirements 110/120 V, 50/60/400 Hz

220/240 V, 50/60/400 Hz

DC (incubator only) 12 V or 28 V

Chassis leakage current 120 V < 300 µA + 10%

220 V < 500 µA + 10%

6 AC outlets

AC and incubator DC cable included

Compressor Battery Specifications

Type Rechargeable, 12 V DC,20 AH Sealed gel-type battery

Battery Run Time 3 hours minimum withblender set at 21% oxygen

Life expectancy 200 complete charge/discharge cycles minimum

Charge Time 20 hours from full discharge

Low Battery Alarm when 30 minutes of compressor air remain

Notes: 1) weight includes one battery

USADraeger Medical, Inc.3135 Quarry Road Telford, PA 18969-1042, USATel +1 215 721 5400Toll-free +1 800 437 2437Fax +1 215 723 [email protected]

CANADADraeger Medical Canada Inc.120 East Beaver Creek Road Suite 104Richmond Hill Ontario L4B 4V1, CanadaTel +1 905 763-3702Toll-free +1 866 343 22 73Fax +1 905 [email protected]

EUROPE, MIDDLE EAST, AFRICA,LATIN AMERICA, ASIA, PACIFICDräger Medical GmbHMoislinger Allee 53–5523558 Lübeck, GermanyTel +49 451 882-0Fax +49 451 [email protected]

Manufacturer:Draeger Medical Systems, Inc.Telford, PA 18969, USAThe quality management system at Draeger Medical Systems, Inc. iscertified according to ISO 13485, ISO 9001 and Annex II.3 of Directive93/42/EEC (Medical devices).

90 5

0 46

4 |

09.1

0-4

| M

arke

ting

Com

mun

icat

ions

| L

SL

| LE

| P

rinte

d in

Ger

man

y |

Chl

orin

e-fr

ee –

envi

ronm

enta

lly c

ompa

tible

| S

ubje

ct to

mod

ifica

tions

| ©

201

0 D

räge

rwer

k A

G &

Co.

KG

aA

HEADQUARTERSDrägerwerk AG & Co. KGaAMoislinger Allee 53–5523558 Lübeck, Germany

www.draeger.com