28
Norma Italiana N O R M A I T A L I A N A CEI CNR CONSIGLIO NAZIONALE DELLE RICERCHE • AEI ASSOCIAZIONE ELETTROTECNICA ED ELETTRONICA ITALIANA Data Pubblicazione Edizione Classificazione Fascicolo COMITATO ELETTROTECNICO ITALIANO Titolo Title CEI EN 60814 1998-11 Seconda 10-13 4840E Liquidi isolanti - Carta e cartone impregnati d’olio Determinazione della quantità d’acqua mediante titolazione automatica coulometrica di Karl Fischer Insulating liquids - Oil-impregnated paper and pressboard Determination of water by automatic coulometric Karl Fischer titration ELETTROTECNICA GENERALE E MATERIALI PER USO ELETTRICO NORMA TECNICA

CEI 10-13

Embed Size (px)

Citation preview

Page 1: CEI 10-13

Norma Italiana

N O R M A I T A L I A N A C E I

CNR

CONSIGLIO NAZIONALE DELLE RICERCHE •

AEI

ASSOCIAZIONE ELETTROTECNICA ED ELETTRONICA ITALIANA

Data Pubblicazione

Edizione

Classificazione Fascicolo

COMITATOELETTROTECNICO

ITALIANO

Titolo

Title

CEI EN 60814

1998-11

Seconda

10-13 4840E

Liquidi isolanti - Carta e cartone impregnati d’olio

Determinazione della quantità d’acqua mediante titolazione automatica coulometrica di Karl Fischer

Insulating liquids - Oil-impregnated paper and pressboard

Determination of water by automatic coulometric Karl Fischer titration

ELETTROTECNICA GENERALE E MATERIALI PER USO ELETTRICO

NO

RM

A TE

CNIC

A

Page 2: CEI 10-13

© CEI - Milano 1998. Riproduzione vietata.

Tutti i diritti sono riservati. Nessuna parte del presente Documento può essere riprodotta o diffusa con un mezzo qualsiasi senza il consenso scritto del CEI.Le Norme CEI sono revisionate, quando necessario, con la pubblicazione sia di nuove edizioni sia di varianti. È importante pertanto che gli utenti delle stesse si accertino di essere in possesso dell’ultima edizione o variante.

SOMMARIO

La Norma descrive i metodi per la determinazione del contenuto d’acqua nei liquidi isolanti e nell’isola-mento cellulosico impregnato col reagente generato coulometricamente di Karl Fisher.La presente EN viene pubblicata dal CEI nella sola lingua originale inglese, a causa della limitata utilizza-

zione, particolarmente mirata a settori specialistici.

DESCRITTORI

• DESCRIPTORS

Materiali isolanti elettrici •

Electrical insulating materials;

Materiali isolanti elettrici liquidi •

Liquid electrical insulating materials;

Oli isolanti •

Insulating oils;

Carta •

Paper;

Cartoni •

Paperboards;

Analisi chimica •

Chemical analysis;

Determinazione del contenuto •

Determination of content;

Acqua •

Water;

Metodo Karl Fisher •

Karl

Fisher method;

Metodi coulometrici •

Coulometric methods;

COLLEGAMENTI/RELAZIONI TRA DOCUMENTI

Nazionali

Europei

(IDT) EN 60814:1997-10;

Internazionali

(IDT) IEC 60814:1997-08 ;

Legislativi

INFORMAZIONI EDITORIALI

Norma Italiana

CEI EN 60814

Pubblicazione

Norma Tecnica

Carattere Doc.

Stato Edizione

In vigore

Data validità

1999-1-1

Ambito validità

Europeo e Internazionale

Varianti

Nessuna

Ed. Prec. Fasc.

3361 H:1997-09

Comitato Tecnico

10-Fluidi isolanti

Approvata dal

Presidente del CEI

in Data

1998-11-5

CENELEC

in Data

1997-10-1

Sottoposta a

inchiesta pubblica come Documento originale

Chiusa in data

1997-6-15

Gruppo Abb.

1

Sezioni Abb.

B

ICS

29.035.40;

CDU

LEGENDA

(IDT) La Norma in oggetto è identica alle Norme indicate dopo il riferimento (IDT)

Page 3: CEI 10-13

CENELEC members are bound to comply with theCEN/CENELEC Internal Regulations which stipulatethe conditions for giving this European Standard thestatus of a National Standard without any alteration.Up-to-date lists and bibliographical references con-cerning such National Standards may be obtained onapplication to the Central Secretariat or to anyCENELEC member.This European Standard exists in three official ver-sions (English, French, German).A version in any other language and notified to theCENELEC Central Secretariat has the same status asthe official versions.CENELEC members are the national electrotechnicalcommittees of: Austria, Belgium, Denmark, Finland,France, Germany, Greece, Iceland, Ireland, Italy, Lu-xembourg, Netherlands, Norway, Portugal, Spain,Sweden, Switzerland and United Kingdom.

I Comitati Nazionali membri del CENELEC sono tenu-ti, in accordo col regolamento interno del CEN/CENE-LEC, ad adottare questa Norma Europea, senza alcunamodifica, come Norma Nazionale.Gli elenchi aggiornati e i relativi riferimenti di tali Nor-me Nazionali possono essere ottenuti rivolgendosi alSegretario Centrale del CENELEC o agli uffici di qual-siasi Comitato Nazionale membro.La presente Norma Europea esiste in tre versioni uffi-ciali (inglese, francese, tedesco).Una traduzione effettuata da un altro Paese membro,sotto la sua responsabilità, nella sua lingua nazionalee notificata al CENELEC, ha la medesima validità.I membri del CENELEC sono i Comitati ElettrotecniciNazionali dei seguenti Paesi: Austria, Belgio, Dani-marca, Finlandia, Francia, Germania, Grecia, Irlanda,Islanda, Italia, Lussemburgo, Norvegia, Olanda, Por-togallo, Regno Unito, Spagna, Svezia e Svizzera.

© CENELEC Copyright reserved to all CENELEC members. I diritti di riproduzione di questa Norma Europea sono riservati esclu-sivamente ai membri nazionali del CENELEC.

Comitato Europeo di Normalizzazione Elettrotecnica European Committee for Electrotechnical Standardization

Comité Européen de Normalisation ElectrotechniqueEuropäisches Komitee für Elektrotechnische Normung

C E N E L E C

Secrétariat Central: rue de Stassart 35, B - 1050 Bruxelles

E u r o p ä i s c h e N o r m • N o r m e E u r o p é e n n e • E u r o p e a n S t a n d a r d • N o r m a E u r o p e a

EN 60814:1997-10

Sostituisce il documento HD 487 S1 (1987)

Liquidi isolanti - Carta e cartone impregnati d’olio

Determinazione della quantità d’acqua mediante titolazione automatica coulometrica di Karl Fischer

Insulating liquids - Oil-impregnated paper and pressboard

Determination of water by automatic coulometric Karl Fischer titration

Page 4: CEI 10-13

CONTENTS

Rif. Topic Pag

.

NORMA TECNICACEI EN 60814:1998-11

Pagina iv

1

GENERAL

1

1.1

Scope

................................................................................................................................................................................................

1

1.2

Normative references

............................................................................................................................................................

1

2

DIRECT TITRATION FOR LOW VISCOSITY LIQUIDS

1

2.1

Field of application

................................................................................................................................................................

1

2.2

Chemistry

.......................................................................................................................................................................................

1

2.3

Apparatus

.......................................................................................................................................................................................

2

2.4

Reagents and auxiliary materials

..................................................................................................................................

4

2.5

Preparation of the apparatus

...........................................................................................................................................

4

2.6

Sampling methods

...................................................................................................................................................................

4

2.7

Procedure

......................................................................................................................................................................................

5

2.8

Calculation of the result

......................................................................................................................................................

5

2.9

Report

...............................................................................................................................................................................................

6

2.10

Precision

.........................................................................................................................................................................................

6

3

EVAPORATIVE STRIPPING METHOD FOR HIGH VISCOSITY LIQUIDS

6

3.1

Field of application

................................................................................................................................................................

6

3.2

Outline of the method

.........................................................................................................................................................

6

3.3

Apparatus and reagents

......................................................................................................................................................

6

3.4

Procedure

......................................................................................................................................................................................

7

3.5

Calculation of water content

...........................................................................................................................................

7

3.6

Report

...............................................................................................................................................................................................

8

4

DETERMINATION OF WATER IN OIL-IMPREGNATED PAPER AND PRESSBOARD

8

4.1

Field of application

................................................................................................................................................................

8

4.2

Determination of water after previous extraction with methanol

........................................................

8

4.3

Determination by direct titration

................................................................................................................................

10

4.4

Evaporative stripping method

......................................................................................................................................

11

A N N E X

A

informative

METHOD FOR SAMPLING OF OIL-IMPREGNATED PAPER AND PRESSBOARD

17

A N N E X

ZA

normative

Normative references to international publications with their corresponding European publications

20

Page 5: CEI 10-13

NORMA TECNICACEI EN 60814:1998-11

Pagina v

FOREWORD

The text of document 10/406/FDIS, future edition 2 of IEC 60814, prepared byIEC TC 10, Fluids for electrotechnical applications, was submitted to theIEC-CENELEC parallel vote and was approved by CENELEC as EN 60814 on1997/10/01.

This European Standard supersedes HD 487 S1:1987.

The following dates were fixed:

n

latest date by which the EN has to be implemented at national level by publi-cation of an identical national standard or by endorsement

(dop) 1998/07/01

n

latest date by which the national standards conflicting with the EN have to bewithdrawn

(dow) 1998/07/01

Annexes designated “normative” are part of the body of the standard.

Annexes designated “informative” are given for information only.

In this standard, annex ZA is normative and annex A is informative.

Annex ZA has been added by CENELEC.

ENDORSEMENT NOTICEThe text of the International Standard IEC 60814:1997 was approved by CENELECas a European Standard without any modification.

Page 6: CEI 10-13

NORMA TECNICACEI EN 60814:1998-11Pagina vi

Page 7: CEI 10-13

NORMA TECNICACEI EN 60814:1998-11

Pagina 1 di 22

1 GENERAL

1.1 ScopeThis International Standard describes methods for the determination of water ininsulating liquids and in oil-impregnated cellulosic insulation with coulometrical-ly generated Karl Fischer reagent.

The method in clause 2 is applicable to water concentrations above 2 mg/kg inliquids having viscosity of less than 100 mm2/s at 40 °C.

The test method in clause 3, where water is extracted by means of a nitrogenstream, is the preferred method for insulating liquids of viscosity higher than100 mm2/s.

Clause 4 describes methods for the determination of water content in oil-impreg-nated paper and pressboard over the range 0,1% to 20% by mass.

1.2 Normative referencesThe following normative documents contain provisions which, through referencein this text, constitute provisions of this International Standard. At the time ofpublication, the editions indicated were valid. All normative documents are sub-ject to revision, and parties to agreements based on this International Standardare encouraged to investigate the possibility of applying the most recent editionsof the normative documents indicated below(1). Members of IEC and ISO main-tain registers of currently valid International Standards.

2 DIRECT TITRATION FOR LOW VISCOSITY LIQUIDS

2.1 Field of applicationThis method is applicable to water concentrations above 2 mg/kg in liquids hav-ing viscosity up to 100 mm2/s at 40 °C. The precision data given in 2.10 applyonly to new liquids.

Notes: 1 For liquids in service, the accuracy of the method may be affected by the presence of con-taminants and degradation products.

2 The method has been designed to be particularly suitable to hydrocarbon and ester liquids.With other liquids, particularly silicone fluids, methanol free reagents must be used.

2.2 ChemistryThe reactions occurring in a Karl Fischer titration are known to be complex, butare essentially of water with iodine, sulphur dioxide, an organic base and an al-cohol in an organic solvent. The original Karl Fischer reagent used pyridine andmethanol, and the reactions may be expressed as:

(1)

(2)

(1) Editor’s Note: For the list of Publications, see Annex ZA.

H2O I2 SO2 3C5H5N 2C5H5N.HI + C5H5N.SO3®+ + +

C5H5N.SO3 CH3OH C5H5NH.SO4.CH3®+

CEI EN 60814:1998-11

28

Page 8: CEI 10-13

NORMA TECNICACEI EN 60814:1998-11Pagina 2 di 22

Other base-alcohol combinations are possible and may be necessary for titrationson some insulating liquids.

In coulometric Karl Fischer titration, the sample is mixed with a base/alcohol so-lution of iodide ion and sulphur dioxide. Iodine is generated electrolytically andreacts with water in a similar way to that shown in reactions (1) and (2). Iodine isgenerated in proportion to the quantity of electricity according to Faraday’s law,as shown by the following reaction:

(3)

One mole of iodine reacts with one mole of water stoichiometrically as shown inreactions (1) so that 1 mg of water is equivalent to 10,72 C (number of cou-lombs). Based on this principle it is possible to determine the amount of waterdirectly from the quantity of electricity (number of coulombs) required for theelectrolysis.

2.3 Apparatus

2.3.1 Principle of operationThe titration vessel has the configuration of an electrolysis cell with two compart-ments connected by a porous diaphragm. The anodic compartment contains themixture of reagent-solvent and sample (anodic solution), the cathodic compart-ment (generator assembly) contains anhydrous reagent (cathodic solution). Onboth sides of the diaphragm are located the electrolysis electrodes.

Note - Titrators without the porous diaphragm may be used.

Iodine generated by the electrolysis, as shown in reaction (3), reacts with the wa-ter in a similar way to the Karl Fischer reactions (1) and (2). The end-point of thereaction is detected by a pair of platinum electrodes immersed in the anodic so-lution. At the end of the titration, excess iodine depolarizes the dual platinumelectrodes, giving a change in the current/voltage ratio which is used to activatethe end-point indicator and to stop the current integrator.

The current integrator integrates the current consumed during the electrolysis,calculates the water equivalent according to Faraday’s law, and finally displays itin micrograms of water.

2.3.2 Description of the apparatusCommercial coulometric Karl Fischer titrators use proprietary circuitry. The fol-lowing description of one suitable form of apparatus is given for illustrative pur-poses only.

The block diagram shown in figure 1 illustrates the apparatus and includes thecomponents detailed below.

2.3.2.1 Titration vessel assemblyAn example of a suitable titration vessel assembly is shown in figure 2. However,the changes in instrument technology are such that radically different designsmay become available which comply with the technical requirements of thisstandard. The exemplified apparatus consists of:n a flanged glass reaction vessel (a) with sample injection plug (b) and drain

cock (c) (optional);n a polytetrafluoroethylene lid (d), flanged to match the titration vessel, with

three holes to receive the electrodes and drying tube;

2 I-

2 e I2®–

Page 9: CEI 10-13

NORMA TECNICACEI EN 60814:1998-11

Pagina 3 di 22

n a generator assembly (combined electrolysis cell) (e) consisting of a glasstube closed at its lower end by a diaphragm and equipped with platinumelectrodes on each side of the diaphragm;

Note The diaphragm may consist of ion exchange membrane, fritted disc, ceramic filter or other sys-tem to prevent diffusion of both solutions, while allowing enough current for electrolysis.

n detector electrodes: dual platinum electrodes for measurement of potential orcurrent (f);

n a polytetrafluoroethylene coated stirrer bar (g);n drying tubes (h) to protect the titration vessel and the generator assembly

from atmospheric moisturen silicone rubber septa to seal the injection port. It is recommended that cross-

cuts should be made in the septa before use, to enable blunt, square-endedneedles to be used for sample injection (see 2.4.2 d)). Septa should be re-placed as required to prevent air leakage as indicated by excessive instrumentdrift.

2.3.2.2 Detection circuitDC constant voltage or a.c. constant current is supplied to the detector electrodes(dual platinum measuring electrodes) so that the end-point may be detected fromthe change of the polarized current or voltage.

2.3.2.3 Current regulator circuitThis circuit controls the electrolysis according to the signal from the detector cir-cuit.

2.3.2.4 DC power supplyDC power supply for electrolysis.

2.3.2.5 End-point indicatorIndicates when the end-point has been reached.

2.3.2.6 Current integratorMeasures the quantity of electricity consumed by the electrolysis cell during thetitration, then calculates and displays the quantity of water, in micrograms, corre-sponding to it.

Note Some instruments have built-in calculation facilities, and display the water concentration for aspecific sample quantity.

2.3.2.7 Electromagnetic stirrerElectromagnetic stirrer, capable of maintaining a constant speed sufficient to en-sure adequate dispersion. (The content of the titration vessel will not in generalbe a single phase, since most insulating liquids are not completely miscible withthe reagent liquids.)

Page 10: CEI 10-13

NORMA TECNICACEI EN 60814:1998-11Pagina 4 di 22

2.4 Reagents and auxiliary materialsWARNING - Certain reagents may be detrimental to health and must be handledwith proper care.

2.4.1 ReagentsPrepared reagents are commercially available, but care is needed that the reagentis suitable for the particular type of instrument used and the insulating liquid un-der test.

Interfering reactions may take place between methanol based reagents and somesilicone compounds. In addition, similar reactions may take place with alde-hydes, some ketones, and conjugated unsaturated organic acids, which may bepresent in oils as degradation products or contaminants. In these cases, reagentsnot based on methanol are recommended.

Note Some insulating liquids may require the use of additional or alternative solvents.

2.4.2 Auxiliary materialsNeutralizing solution, methanol containing approximately 20 mg water/cm3.

Desiccant, for example, anhydrous magnesium perchlorate or self-indicating sili-cagel.

Lubricant grease: Polytetrafluoroethylene based or fluorinated hydrocarbontypes. Commercially supplied greases to this description have been found satis-factory.

Glass syringes for sample measurement and introduction. 10 cm3 and 5 cm3 sy-ringes in accordance with ISO 595, with needles of suitable length and diameter.Needles of length 100 mm and bore 1 mm have been found satisfactory for gen-eral use. Blunt, square-ended needles are preferred, in order to minimize damageto the septa.

2.5 Preparation of the apparatusPrepare and assemble the apparatus, install the reagents and carry out the stabili-zation procedure in accordance with the manufacturer’s instructions.

2.6 Sampling methodsIf samples taken are intended for additional tests to water content the water anal-ysis shall be carried out first.

2.6.1 Routine samplingFor routine tests, the sampling methods described in clause 2 and 3.1 ofIEC 60475 shall be used. Sample bottles shall be dried by heating in an oven for16 h to 24 h at 115 °C ± 5 °C.

2.6.2 Recommended samplingFor better accuracy, and particularly where the moisture content is very low, theprocedures described in clause 4 of IEC 60567 shall be used. Sample bottles shallbe prepared as specified in 2.7.1. The syringes and needles shall be dismantled,dried for at least 8 h at 115 °C ± 5 °C, cooled in a desiccator with anhydrous sili-ca gel, and kept in the desiccator until required.

Note The accuracy of the determination can be adversely affected by the taking of composite or aver-age samples, which is not recommended.

Page 11: CEI 10-13

NORMA TECNICACEI EN 60814:1998-11

Pagina 5 di 22

2.6.3 StorageSamples for water analysis shall at no time be exposed to direct sunlight. Be-tween the time of collection and analysis, the samples should be kept in thedark, and the time between collection and analysis should be not greater thanseven days.

2.7 Procedurea) With samples which have been collected in bottles

Clean and dry in a well-ventilated oven at 115 °C ± 5 °C a glass syringe andneedle of a suitable size. Allow to cool in a desiccator. Fill the syringe withthe insulating liquid, keeping the tip of the needle well below the surface ofthe liquid. Re-close the glass bottle immediately. Holding the syringe vertical-ly with the needle uppermost, discharge all air bubbles. Discharge the con-tents of the syringe to waste. Refill the syringe and weigh to the nearest 0,1 g.With samples which have been collected in a syringeDischarge approximately 2 cm3 to flush out the needle and weigh the syringeto the nearest 0,1 g.The sample size is governed by the expected range of moisture content.However, the optimum sample size has been found to be 5 cm3 for mosttypes of insulating liquids, new and used, with water contents of between2 mg/kg and 100 mg/kg.

b) Operate the instrument controls to start the electrolysis according to the man-ufacturer’s instructions and quickly inject a suitable quantity of sample intothe titration vessel through the septum. The tip of the needle shall not be al-lowed to touch the surface of the reagent. Re-weigh the syringe and recordthe mass M, in grams, of sample injected.Ensure that the oil is thoroughly mixed with the solvent. The speed of thestirrer shall not be changed after the instrument has self-equilibrated or dur-ing the titration.

c) Read out the quantity of water titrated, m, (expressed in micrograms (mg))from the display when the titration is completed.

d) Carry out a duplicate determination by rinsing the syringe twice in the oilsample and refilling and weighing as in step a). Carry out steps b) and c).

e) After several determinations, a considerable quantity of insulating liquid mayhave accumulated. Remove the excess of liquid in accordance with the man-ufacturer’s instructions, which should also be followed with respect to the rel-ative levels of the anode and cathode reagents and re-stabilization of the in-strument.

After several withdrawals, the titration vessel and generator electrode should berecharged with fresh solutions and the stabilization procedure repeated.

2.8 Calculation of the result

(4)

where

m is the quantity of water titrated in micrograms (mg);

M is the mass of insulating liquid in grams (g).

water content mg/kg( )mM-----=

Page 12: CEI 10-13

NORMA TECNICACEI EN 60814:1998-11Pagina 6 di 22

2.9 ReportThe water content of the insulating liquid sample is the mean of the duplicate deter-minations, expressed to the nearest integer in milligrams per kilogram (mg/kg).

2.10 Precision

Note Precision data have been established for hydrocarbon based insulating liquids

2.10.1 RepeatabilityDuplicate determinations carried out by one operator should be considered sus-pect at the 95% confidence level if they differ by more than ,where x is the average of the duplicate determinations.

2.10.2 ReproducibilityWhen two laboratories carry out tests on identical test material, each shall pro-duce duplicate results and report their mean.

The two means should be considered suspect at the 95% confidence level if they dif-fer by more than mg/kg, where x is the average of the two means.

3 EVAPORATIVE STRIPPING METHOD FOR HIGH VISCOSITY LIQUIDS

3.1 Field of applicationThis method is designed for the determination of water in unused insulating liq-uids having viscosity greater than 100 mm2 /s at 40 °C. This method is mainly ap-plicable to water concentration above 2 mg/kg.

Note Alternatively, high viscosity liquids may also be analysed for water content by the method inclause 2 after dilution with a suitable solvent. Precision for a dilution procedure will be affect-ed by the degree of dilution and the water content and viscosity of the chosen solvent.

3.2 Outline of the methodA known amount of the insulating liquid is heated in a closed vessel located nextto the Karl Fischer apparatus. The water evolved is quantitatively transferred intothe titration vessel by a stream of dry nitrogen gas where coulometric titration isperformed.

3.3 Apparatus and reagentsThe block diagram of the apparatus is illustrated in figure 3 and consists of thefollowing items:n titrator: automatic coulometric Kart Fischer (see 2.3);n evaporator: glass vessel, 100 ml capacity; nitrogen inlet tube, 1,25 mm inside

diameter (see figure 4);n heater: clear, transparent electro-conductive glass heater;n temperature controller: automatic control, accuracy ±2 °C;n carrier gas: nitrogen gas, commercial grade, less than 10 ml/l water;n desiccants for carrier gas: one silica gel column plus two phosphorus pentox-

ide columns;n reagents as listed in 2.4.

Note When interfering compounds are present, the need for suitable reagents in the titrator is ofequal importance as for direct sample injection.

0 60, x mg/kg

1 50, x mg/kg

Page 13: CEI 10-13

NORMA TECNICACEI EN 60814:1998-11

Pagina 7 di 22

3.4 Procedurea) Operate the Karl Fischer titrator as per procedures specified in 2.5.b) Inject a sufficient volume (approximately 10 cm3) of base oil into the evapo-

rator vessel via the septum of the injection plug so as to immerse the tip ofthe nitrogen inlet tube to permit bubbling of the gas. This base oil may be theinsulating liquid to be tested or any other miscible oil which does not reactwith the sample and of similar viscosity at the operating temperature to theoil being analysed.

c) Set the temperature of the evaporator unit to 130 °C ± 5 °C and allow thetemperature to stabilize. Purge the whole system with dry nitrogen gas at aflow rate of 50 cm3/min to 200 cm3/min to thoroughly dry the base oil as in-dicated by a low and stable reading of background current.

d) Turn on the titration switch. Run two blanks of 10 min each on this initialbase oil by following steps f) and g) of the procedure. If the two blanks arewithin 5 mg of water, take the average blank value (m1). If the two blanks dif-fer by more than 5 mg of water, further drying of the base oil may be needed.

e) introduce a suitable quantity of the insulating liquid (M) into the evaporatorvessel using a syringe (minimum needle size: 2 mm inside diameter) or othersuitable sampler. Weigh the syringe before and after liquid injection to obtainthe accurate mass of sample. Use a 10 g sample for water contents greaterthan 10 mg/kg and a 20 g ± 5 g sample for water contents less than10 mg/kg.

f) Turn off titration switch on main unit, or otherwise inhibit titration, for a10 min period to allow moisture released in the evaporator to accumulate inthe titration cell.

g) After 10 min, turn on the titration switch and allow to titrate to the end-point.h) Read the quantity of water titrated (m2) from the Karl Fischer apparatus.

Note When the water contents are higher than 50 mg/kg and/or the mass of sample is 10 g, there isthe possibility that complete evaporation of the water has not taken place. In such cases, theanalysis should be repeated as in steps e) to h) but using an increased titration delay timeand/or a lower amount of oil. With some designs of titrator this purpose can be achieved by set-ting the delay time sufficiently long.

i) It is good practice to confirm stable operating conditions by carrying out an-other blank determination for a 10 min period. If the value determined iswithin 5 mg of the original blank value m1, no further titration is needed.

j) Carry out a duplicate determination.

Note In the case of routine tests steps d) and i) may be omitted and in steps f) and g) the 10 min maybe reduced to 1 min.

3.5 Calculation of water contentFor each individual determination, calculate the water content as follows:

(5)

where

m1 is the average blank reading in micrograms (mg);

m2 is the mass of water titrated during sample analysis in micrograms (mg);

M is the mass of the sample in grams (g).

water content (mg/kg) m2 m1–

M---------------------=

Page 14: CEI 10-13

NORMA TECNICACEI EN 60814:1998-11Pagina 8 di 22

3.6 ReportThe water content of the insulating liquid sample is expressed as the average ofduplicate determinations to the nearest integer in milligrams per kilogram.

4 DETERMINATION OF WATER IN OIL-IMPREGNATED PAPER AND PRESSBOARD

4.1 Field of applicationThese methods apply to the determination of the water content in oil-impregnat-ed cellulosic insulation materials. Three procedures are described:n with the procedure in 4.2, water is first extracted with absolute methanol and

the determination carried out on the extract;n with the procedure in 4.3, water extraction takes place directly in the titration

vessel (see note);

Note This procedure is restricted to cellulosic material of thickness up to approximately 1 mm be-cause an oil-impregnated sample of higher thickness can never be completely extracted withinthe normal duration of a determination.

n with the procedure in 4.4, a specimen of impregnated paper is heated in anoven next to the Karl Fischer apparatus; the water vapour evolved is quantita-tively transferred into the titration vessel by a stream of dry nitrogen gas.

4.2 Determination of water after previous extraction with methanol

4.2.1 Reagents and auxiliary materialsWARNING - Certain reagents may be detrimental to health and the environmentand must be handled and disposed of with proper care.

In addition to the reagents and materials listed in 2.4, the following is required:a) methanol, analytical reagent grade, water content approximately 0,05%;b) chlorine free suitable solvent, technical grade;c) magnesium turnings.

4.2.2 Apparatusa) Coulometric Karl Fischer titrator as indicated in 2.3.b) Methanol distilling apparatus.c) Dried methanol container, to contain the distilled methanol and which can be

protected from the atmosphere by a drying tube. The container is fitted at itslower part with a draining cock and a spherical joint to adapt to the extrac-tion tube (see figure 5).

d) Graduated, gas-tight extraction tubes with an approximate capacity of 50 cm3.e) Metal tweezers.

4.2.3 Preparation of the apparatusAll the glassware and metal tweezers should be washed free from previous sam-ples, either with a suitable detergent solution or soapy water or a chlorine freesolvent, rinsed thoroughly with warm water, then deionized water and finally, af-ter draining, with methanol. The clean material should be dried in an oven at115 °C ± 5 °C overnight, then cooled to room temperature in a desiccator.

Page 15: CEI 10-13

NORMA TECNICACEI EN 60814:1998-11

Pagina 9 di 22

4.2.4 Procedurea) Distil methanol over magnesium turnings in order to lower its water content

below 200 mg/kg. Collect the distilled methanol in the appropriate containerprotected from the atmosphere by a drying tube filled with magnesium per-chlorate. Determine the water content of the methanol and record the value.

b) Take an extraction tube containing a paper sample or put a paper sample inextraction tube.

Notes: 1 Select the mass of the sample so that the amount of water to be measured is within 1 mg to4 mg.

2 It may be advisable to cut up cellulosic material if it is too thick to ensure better extractiontaking care to avoid any exchange of moisture with the surroundings during the opera-tion.

c) Fit the extraction tube to the dry methanol container, open the upper cock,connect to the vacuum line for a very short time and, using an appropriateburette, admit 1 cm3 to 10 cm3 of methanol. Close the upper cock and dis-connect the tube.

d) Repeat step c) with another tube that does not contain any paper. This will beused for the blank test.

e) Shake the sample and blank tubes for 2 h.f) Operate the Karl Fisher titrator as per procedures specified in 2.5.g) Fit the sample tube to the titration vessel, transfer the methanol to the solvent

and titrate to the end-point. Read the mass of water titrated (m2) from the dis-play on the titrator.

h) Repeat step g) with the blank tube and record the display reading (m1).i) Remove the paper from the titration tube. Degrease the paper with chlorine

free solvent and dry at 110 °C ± 5 °C for at least 2 h. Let the paper sample coolin a desiccator and then weigh. Record the mass M, in grams, of the paper.

j) Carry out a duplicate determination.

4.2.5 Calculation of the results

(6)

where

m1 is the mass of water measured for blank titration, in micrograms (mg);

m2 is the mass of water measured in the sample titration, in micrograms (mg);

M is the mass of the paper sample, in grams (g).

4.2.6 ReportThe water content of the insulation is expressed as the average of duplicatemeasurements to the nearest 0,01%. The water content of the methanol usedshall be quoted.

water content (% by mass) m2 m1–

M---------------------è øæ ö 10

4–´=

Page 16: CEI 10-13

NORMA TECNICACEI EN 60814:1998-11Pagina 10 di 22

4.3 Determination by direct titration

4.3.1 Reagents and auxiliary materialsIn addition to the reagents and materials listed in 2.4, a suitable chlorine-free sol-vent is required for degreasing the sample material.

4.3.2 Apparatusa) Coulometric Karl Fischer titrator as described in 2.3.b) Metal tweezers.

4.3.3 Procedurea) Operate the Karl Fischer titrator as per procedures specified in 2.5.b) Introduce, as quickly as possible, the paper sample by means of dried metal

tweezers into the titration vessel.

Note See notes 1 and 2 item b) of 4.2.4.

c) Turn off the titration switch on the main unit, or otherwise inhibit titration, fora time period ranging from 15 min to 25 min (the exact time shall be deter-mined by the operator) to allow transfer of moisture from the paper to thesolvent. At the end of the extraction time, turn on the titration and titrate tothe end-point.

d) Read the quantity of water titrated (m2) from the Karl Fischer apparatus.e) Carry out a blank test of the same duration in order to evaluate the amount of

moisture that enters the system during the test. Record the mass (m1) titrated.f) In some designs of titrator, the drift is continuously monitored and automati-

cally subtracted from the water reading. If so the blank test may not be need-ed though care should be used to ensure stable operating throughout the ti-tration.

Note It is important to limit to a minimum the duration of the determination, which can be fixed bythe operator on the basis of experience and the type of sample examined. If necessary, the opti-mum period can be determined by repeatedly carrying out the procedure using increasing ex-traction times until a constant result is obtained. To obtain reproducible results over a series oftests, the same extraction period should be used throughout the series.

g) Remove the paper from the titration tube. Degrease the paper with the sol-vent and dry at 110 °C for at least 2 h. Let the paper sample cool in a desicca-tor and then weigh it. Record the mass M, in grams, of the paper.

h) Carry out a duplicate determination.

4.3.4 Calculation of results

(7)

where

m1 is the mass of water measured for blank titration in micrograms (mg);

m2 is the mass of water measured in the sample titration in micrograms (mg);

M is the mass of the paper sample in grams (g).

4.3.5 ReportThe water content of the insulation is expressed as the average of duplicatemeasurements to the nearest 0,01%. The time of extraction shall be quoted.

water content (% by mass) m2 m1–

M---------------------è øæ ö 10

4–´=

Page 17: CEI 10-13

NORMA TECNICACEI EN 60814:1998-11

Pagina 11 di 22

4.4 Evaporative stripping method

4.4.1 ApparatusThe block diagram of the apparatus is illustrated in figure 3 with components aslisted in 3.3.

The boat type of evaporator using a gas-solid extraction is also suitable.

4.4.2 Procedurea) Operate the Karl Fischer apparatus as per procedures specified in 2.7.b) Heat the evaporator to a suitable temperature (130 °C for low viscosity oil im-

pregnated paper or pressboard, 140 °C for paper or pressboard impregnatedwith viscous oils or compounds).

c) Adjust the carrier gas flow so as to ensure a steady flow rate of 50 cm3/min to100 cm3/min. Insert the tip of the carrier gas outlet tube into the titration ves-sel and allow the system to stabilize as indicated by a low and stable readingof background current (drift).

d) Quickly introduce the paper sample taken from a storage flask into the evap-orator vessel. For papers with very low moisture content a paper sample ofabout 0,5 g is adequate.

e) Turn off titration switch on main unit, or otherwise inhibit titration, for a20 min period to allow moisture released in the evaporator to accumulate inthe titration vessel. At the end of the extraction time, switch on titration andallow to titrate to the end-point. Read the quantity of water titrated (m2) fromthe Karl Fischer apparatus.

Note Extraction time will depend on sample characteristics as well as system parameters. Although a20 min period was found adequate in most cases, the operator will have to determine optimumoperating conditions.

f) Carry out a blank test of the same duration in order to evaluate the amount ofmoisture that enters the system during the test. Record the mass titrated (m1).In some designs of titrator the drift is continuously monitored and automati-cally subtracted from the water reading. If so the blank test may not be need-ed though care should be taken to ensure stable operating throughout the ti-tration.

g) Remove the paper from the titration tube. Degrease the paper with the sol-vent and dry at 110 °C ± 5 °C for at least 2 h. Let the paper sample cool in adesiccator and then weigh it. Record the mass M, in grams, of the paper.

h) Carry out a duplicate determination.

4.4.3 Calculation of results

(8)

where

m1 is the mass of water measured for blank titration in micrograms (mg);

m2 is the mass of water measured in the sample titration in micrograms (mg);

M is the mass of the paper sample in grams (g).

4.4.4 ReportThe water content of the insulation is expressed as the average of duplicatemeasurements to the nearest 0,01 %. The time of extraction shall be quoted.

water content (% by mass) m2 m1–

M---------------------è øæ ö 10

4–´=

Page 18: CEI 10-13

NORMA TECNICACEI EN 60814:1998-11Pagina 12 di 22

Fig. 1 Block diagram of automatic titratorCAPTION

1 Titration cell assembly2 Detection circuit3 current regulator circuit4 Power circuit (electrolysis)5 End-point displayer6 current integrator7 Electromagnetic stirrer

Page 19: CEI 10-13

NORMA TECNICACEI EN 60814:1998-11

Pagina 13 di 22

Fig. 2 Suitable titration vessel assemblyCAPTION

a Titration vessel (anodic compartment)b Sample injection plugc Drain cockd Polytetrafluoroethylene lide Generator assembly (cathodic compartment)f Dual platinum electrodes for potential measurementg Polytetrafluoroethylene coated stirrer barh Drying tubes

Page 20: CEI 10-13

NORMA TECNICACEI EN 60814:1998-11Pagina 14 di 22

Fig. 3 Block diagram of automatic titrator and evaporation stripping unitCAPTION

1 Flowmeter2 Silica-gel drying tube3 Phosphorus pentoxide drying tube4 Phosphorus pentoxide drying tube5 Evaporator glass vessel with heater6 Temperature controller7 Titration cell8 Control unit

Page 21: CEI 10-13

NORMA TECNICACEI EN 60814:1998-11

Pagina 15 di 22

Fig. 4 Evaporator glass vessel with heaterCAPTION

1 Nitrogen inlet tube2 Thermocouple3 Sample injection plug4 Nitrogen and evaporated water outlet tube5 Heater6 Drain cock

Page 22: CEI 10-13

NORMA TECNICACEI EN 60814:1998-11Pagina 16 di 22

Fig. 5 Methanol container and extraction tubeCAPTION

1 Lower part of the methanol container2 To vacuum pump3 Three way cock4 S 19/9 spherical ground joint (ball)5 S 19/9 spherical ground joint (cup)6 Upper cock7 Extraction tube, capacity 125 ml8 Lower cock9 Conical ground joint 14/23

Page 23: CEI 10-13

NORMA TECNICACEI EN 60814:1998-11

Pagina 17 di 22

A N N E XA informative METHOD FOR SAMPLING OF OIL-IMPREGNATED PAPER

AND PRESSBOARD

A.1 GeneralThis annex describes a few among the various techniques which have beenfound suitable to take samples of oil-impregnated paper and pressboard fromoil-filled electrical equipment. They are intended to provide guidelines for lessexperienced persons and set out the main precautions which should be taken.

The sampling method depends upon the kind of insulation concerned (for exam-ple cable or transformer). However, in every case it is important to avoid any riskof moisture exchange between the zone of insulation from which the sample istaken and the surrounding medium. For this reason, it is recommended that sam-ples be taken while this part of the insulation is still immersed in the impregnantor as soon as it has been removed from the impregnant.

Where it is practicable, it is recommended to take the paper samples with atrepan, if not, razor blades, scalpels or scissors are used.

A.2 Sampling from cable

A.2.1 Sampling a part of cableThe cable is raised at the position where it is to be cut and the metal sheathtinned at two positions not less than 50 cm apart to facilitate plumbing on metalend caps when the sample is removed.

The first cut is made at the appropriate position within one of the tinned areasusing a power or hand saw or abrasive cutting disc and the end of the proposedsample closed immediately with a metal end cap which should be provided witha nut and nipple to relieve oil pressure during plumbing.

The second cut is made within the second of the tinned areas not less than 50 cmfrom the first cut and the capping procedure repeated.

The two cut ends of the remaining cable should also be similarly closed.

If metal end caps are not available, heat shrinkable or other plastic caps may beused provided they are securely taped onto the cable. In this case it is advisableto place the sample in a tin or other closed container with minimum residual airspace until the paper sample is to be taken.

A.2.2 Dissection of cables and sampling of papersA part of cable approximately 30 cm is cut if possible after the sample cable is re-ceived in the laboratory. For this purpose, the sample cable is put in a glove boxwhich is swept by dry nitrogen. The nitrogen is obtained from a commercial gasbottle and passed through a drying tube filled with magnesium perchlorate usedas desiccant to lower the dew-point of the nitrogen to -56 °C. The desiccantshould be regularly replaced. The sample cable is cut up in the glove box.

Paper samples should be taken with a trepan as described in A.3.2 from at leasttwo places at different distances from the conductor for each core of the cable.

The first two or three layers are discarded. Samples are taken from the next five orsix layers and then from the eight layers closest to the conductor. Paper samplesshould not be taken near the ends of the cable sample because they may have ac-cidently absorbed humidity at the time the cable sample was taken. The papersamples are put in storage flasks which are sealed hermetically (see clause A.4).

Page 24: CEI 10-13

NORMA TECNICACEI EN 60814:1998-11Pagina 18 di 22

All these operations should be carried out as quickly as possible. The flasks con-taining the paper samples are removed from the glove box and the moisture de-terminations carried out with the minimum delay according to clause 4.

A.3 Sampling from transformers

A.3.1 Sampling of oil-impregnated paper and pressboard from thin insulation (thickness less than 3 mm)Samples of oil-impregnated paper or pressboard are cut by means of a sharp in-strument (razor blade, scalpel or scissors). All manipulations are carried out usingpreviously dried metal instruments. The samples are then put in storage flaskswhich are sealed hermetically. Then the water determinations are carried out ac-cording to clause 4.

A.3.2 Sampling oil-impregnated paper and pressboard from thick insulation(greater than 3 mm)When the thickness of the insulation is greater than 3 mm, the samples are besttaken with a trepan. The trepan consists of a tube with a cutting edge at one end,and a piston inside the tube. The trepan fitted to a brace is driven into the insula-tion by rotating it. The paper disks so obtained are stored inside the tube and re-main there when the trepan is withdrawn. The paper samples can be pushed outof the tube by means of the piston and put directly in a storage flask as describedin clause A.4.

With the same trepan it is possible to obtain in one operation samples down to adepth of 2 mm to 3 mm and in a second operation down to a depth of 4 mm toS mm. To obtain samples at greater depths it is necessary to continue by usingsuccessively tubes of decreasing diameter. The water determinations are carriedout according to clause 4.

A.4 Storage of paper or pressboard samplesDifferent storage procedures may be considered, according to whether water de-termination is carried out immediately after sampling or later. When moisture canbe determined immediately after sampling the paper samples may be put inflasks cleaned and dried as described in 2.6.1. When the determination is carriedout after previous extraction with methanol, the paper samples are stored directlyin the extraction vessel (see figure 5) and extracted with methanol.

When moisture cannot be determined immediately after sampling it is recom-mended that the samples of paper or pressboard are placed in storage flasksfilled with the insulation impregnant from which the samples were removed oralternatively with dry oil of water content less than 5 mg/kg. The insulation im-pregnant is obtained using techniques described in IEC 60567.

The storage vessels shall be washed and dried as described in 2.6.1. The flasksshall be as small as possible so as to limit the risks of moisture exchange be-tween the impregnant and the paper (e.g. a volume of 50 cm3 for a few grams ofpaper).

The flasks will be fitted with screwed caps and suitable rubber washers. A smallvolume of air should be left above the liquid to compensate for expansion of thelatter with changes in temperature. In some cases, syringes may conveniently beused as storage and transport vessels.

Page 25: CEI 10-13

NORMA TECNICACEI EN 60814:1998-11

Pagina 19 di 22

When syringes are used, the following procedure is recommended:n with a needle fitted, fill the syringe with oil until the piston emerges from the

body of the syringe;n block the needle end by means of a suitable cap;n introduce the sample;n refit the piston into the body of the syringe, remove the cap from the needle,

and depress the piston until the required volume of oil is drained;n block the needle end again by means of the cap. The filled syringe is now

ready for transportation.

To remove the paper samples from the syringe proceed as follows:n hold the syringe with the needle pointing upwards, remove the cap and pull

out the piston until it nearly emerges from the body of the syringe;n block the needle end again by means of the cap;n turn the syringe upside down;n remove the piston from the body of the syringe and take out the paper sam-

ples with previously dried metal tweezers.

Page 26: CEI 10-13

NORMA TECNICACEI EN 60814:1998-11Pagina 20 di 22

A N N E XZA normative Normative references to international publications

with their corresponding European publicationsThis European Standard incorporates by dated or undated reference, provisionsfrom other publications. These normative references are cited at the appropriateplaces in the text and the publications are listed hereafter. For dated references,subsequent amendments to or revisions of any of these publications apply to thisEuropean Standard only when incorporated in it by amendment or revision. Forundated references the latest edition of the publication referred to applies (in-cluding amendments).

Note When an international publication has been modified by common modifications, indicated by(mod), the relevant EN/HD applies.

Publication Year Title EN/HD Year

IEC 60475 1974 Method of sampling liquid dielectrics — —

EN 60567 1992 Guide for the sampling of gases and of oil from oil-filled electrical equipment and for the analysis of free and dissolved gases

EN 60567 1992

ISO 595-1 1986 Reusable all-glass or metal-and-glass syringes for medical use Part 1: Dimensions

— —

ISO 595-2 1987 Part 2: Design, performance requirements and tests EN ISO 595-2 1994

Fine Documento

Page 27: CEI 10-13

NORMA TECNICACEI EN 60814:1998-11

Pagina 21 di 22

Page 28: CEI 10-13

NORMA TECNICACEI EN 60814:1998-11Totale Pagine 28

La presente Norma è stata compilata dal Comitato Elettrotecnico Italiano e beneficia del riconoscimento di cui alla legge 1º Marzo 1968, n. 186.

Editore CEI, Comitato Elettrotecnico Italiano, Milano - Stampa in proprioAutorizzazione del Tribunale di Milano N. 4093 del 24 luglio 1956

Responsabile: Ing. E. Camagni

Sede del Punto di Vendita e di Consultazione 20126 Milano - Viale Monza, 261tel. 02/25773.1 • fax 02/25773.222 • E-MAIL [email protected]

Lire 84.000

10 – Fluidi isolanti

CEI 10-1 Oli minerali isolanti per trasformatori e per apparecchiatureelettriche

CEI 10-10 Interpretazione delle analisi dei gas disciolti nell’olio di trasfor-matori ed altre apparecchiature elettriche

CEI 10-11 Caratteristiche di gassing dei liquidi isolanti per effetto di cam-pi elettrici ad azione ionizzante

CEI 10-14 Rivelazione e determinazione di additivi antiossidanti specificinegli oli isolanti

CEI 10-15 Determinazione dell’acqua negli oli isolanti, nella carta e neicartoni impregnati d’olio

CEI EN 60465 (CEI 10-18) Specifiche per oli minerali isolanti nuovi per cavi a circolazioned’olio

CEI EN 61100 (CEI 10-22) Classificazione dei liquidi isolanti in base al punto di combu-stione ed al potere calorifico inferiore

CEI EN 60567 (CEI 10-23) Guida al campionamento di gas e olio prelevati da apparecchi-ature elettriche riempite d’olio e all’analisi dei gas liberi e di-sciolti

CEI 10-24 Classificazione generale dei liquidi isolanti

CEI EN 61065 (CEI 10-25) Metodo di valutazione delle caratteristiche di fluidità a bassatemperatura di oli minerali isolanti dopo l’invecchiamento

CEI EN 61144 (CEI 10-26) Metodo di prova per la determinazione dell’indice di ossigenodei liquidi isolanti

CEI EN 61125 (CEI 10-27) Liquidi isolanti nuovi a base di idrocarburi Metodi di prova perla valutazione della stabilità all’ossidazione

CEI EN 61181 (CEI 10-28) Materiali isolanti impregnati Applicazione dell’analisi dei gasdisciolti (DGA) alle prove eseguite in fabbrica su apparecchia-ture elettriche

CEI EN 61198 (CEI 10-29) Oli minerali isolanti Metodi per la determinazione di furfurolo2 e suoi derivati