Challenges of implementing JCTLM methods in the routine Clinical Laboratories R.J. Singh, Ph.D. Mayo Clinic

Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

Embed Size (px)

Citation preview

Page 1: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

Challenges of implementing JCTLM

methods in the routine Clinical

Laboratories

R.J. Singh, Ph.D.

Mayo Clinic

Page 2: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

• 100s of Millions of 25-OH Vitamin D

test have been run.

• Total 25-OH Vitamin D test is not a

useful test in itself.

2

Page 3: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

African Americans were Misdiagnosed-NEJM

3

Page 4: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

Medicare Payment for 25-OH-D

4

Page 5: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

Vitamin D Monthly Test Volumes Endocrine Lab Rochester, 2004-2010

0

10,000

20,000

30,000

40,000

50,000

60,000

70,000

80,000

90,000

Jan

04

Aug

04

Feb 0

5

Sep

05

Mar

06

Oct 0

6

Apr

07

Nov

07

Jun

08

Dec

08

Jul 0

9

Jan

10

Aug

10

25-hydroxy vitamin D

1,25-dihydroxy vitamin D

5

Page 6: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

• Other

• Diabetes

• Syndrome X

• Obesity

• Pulmonary

• Wheezing

• FEV1

• Bone and muscle

• Osteoporosis

• Osteomalacia

• Osteoarthritis

• Rickets

• Muscle aches and weakness

Vitamin D Deficiency Sunscreen Melanin Latitude Winter X

Sun

Medications & supplements

• Antiseizure

• Glucocorticoids

• Rifampin

• HAART

• St John’s Wort

• Malabsorption • Crohn • Whipple • Cystic fibrosis • Celiac

• Hepatic failure • Liver disease • Renal failure • Nephrotic syndrome

Consequences

• Psychiatric

• Schizophrenia

• Depression

• Infections

• Urinary

• Tuberculosis

• Circulatory

• High blood pressure

• Coronary heart disease

• Cancer

• Breast

• Colon

• Prostate

• Other

25 OH D 1,25 OH D

6

Page 7: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

Rickets in a Premature Infant

7

Page 8: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

Birth History

• Born through C-section

• Intrauterine growth retardation

• Abnormal umbilical blood flow on ultrasound

• Apgars of 6 and 7

• Infant intubated and given Survanta

• Infant quickly transferred to neonatal intensive care unit for medical management

8

Page 9: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

Ionized Calcium

9

Page 10: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

• Calcium Tests are precise and are well standardized

• Normal values range from 8.5 to 10.2 mg/dL for Total Calcium

• 4.8-5.7 mg/dL for ionized Calcium

• Medically Allowable Error – 0.37 mg/dL

• CLIA Acceptable Performance 1.0 mg/dL

Calcium Testing

10

Page 11: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

Fractures Identified on Routine Radiographic Surveillance for Line Placement

11

Page 12: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

Skin production

Diet: fish, meat

Vitamin D3

25 OH Vitamin D3

Liver

1,25 (OH)2 Vitamin D3

Kidney

HO

OH

OH

CH2

HO

CH2

HO

OH

CH2

12

Page 13: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

Right side of patient Left side of patient

On coronals 13

Page 14: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

Vitamin D Endocrine System

Ca-PO4

1,25 D

Intestine

Absorption

Ca2+

Kidney PTH

Bioactivation 25D 1,25D

+

Bone

PTH

PO4

Reabsorption

Ca2+

PTH

1,25 D

Ca2+

Resorption PO4

+

+

+

+

14

Page 15: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

• 4 month old admitted to PICU

• Calcium 18, 15.5, 15.4

• Phosphorus 2 and 1.8

• PTH <6

• PTHrP-Pending

• What will cause PTH to be low-Any interferences in the PTH assay?

Day 1 Dr.’s Questions to the Lab

15

Page 16: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

• Ultrasound of Abdomen and Kidneys Pending

• Breast Fed

• Some teething Medicine-Homeopathic

• Some Natural Vitamin D-Mom doesn‟t remember the brand

• But If she is giving excess Vitamin D than P should be high as well

• skeletal survey as well.

• ?Jensen‟s chondrodysplasia. Usually though the calcium is not that high

Day 1

PTH Calcium Question

16

Page 17: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

Case: Serum Calcium

7

8

9

10

11

12

13

14

15

16

0 10 20 30 40 50

Se

rum

Ca, m

g/d

L

Days

PICU Transfer,

Calcitonin

Discharge from SMH

Readmitted in MCHS

Pamidronate

17

Page 18: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

15.5

10.4

9.8

10.9

12.3

10.8

11.5

11.1

11.5

12

11.6

8

9

10

11

12

13

14

15

16

0

50

100

150

200

250

300

350

400

450

0 5 10 15 20 25 30 35 40 45

25-OH-D3+3-Epi-25(OH)D3 Total Calcium

Days

To

tal

Calc

ium

mg

/dL

Vit

am

in D

ng

/ml

Vitamin D Toxicity

18

Page 19: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

Day 2 in PICU 4-month-old female History

• Since birth, She has been given supplemental vitamin D.

• Mom reports that she has given 0.25 mL (likely 100 units) on a daily basis.

• Labs and clinical presentation at outside ER were consistent with dehydration. She received a fluid bolus and repeat labs > 12 hours later showed improvement in BUN as well as a decline in her total calcium.

19

Page 20: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

In US Vitamin D Supplements are not

Regulated by FDA

20

Page 21: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

How much vitamin D was the baby getting anyway?

• The vitamin D content of the supplement was threefold higher (6000 IU of D/drop) than listed on the label (2000IU).

• Combined with the gross overdosing, this was estimated to have resulted in a daily vitamin D dose of 50,000 IU for two months

• Nephrocalcinosis was noted on ultrasound 21

Page 22: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked
Page 23: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

23

• HPLC assay in 1981

• 50 samples/day

• RIA in March 2000

• 150 samples/day

• LC-MS/MS in November 2004

• 3500 samples/day

History of 25-OH Vitamin D Testing

Page 24: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

24 Hollis-Journal of Steroid Biochemistry & Molecular Biology 103 (2007) 473–476

Deciding on an Assay Platform

“if an RIA or ELISA better suits the

needs of the laboratory, the results

of these assays should correlate

closely with result from the

physical chemical „gold standard‟

methods of HPLC or mass

spectrometry.”

Page 25: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

Laboratory Automation

Picture of the UVA lab

here

Page 26: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

FE500 at UVA

Modular at John’s Hopkins

Workstation and Modular Pre-Analytical Automation

Pathfinders in Sydney, Australia

Page 27: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

Mass Spec Environement

Page 28: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

28

Page 29: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

29

Page 30: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

30

Plasma (3-5 ml) + 3H Metabolites

Diethylether (2X)

Methanol: Methylene Chloride (1:3)

Backwash w/ 0.1m NaPO4 (pH 10.5)

Sephadex LH-20

(1:1:9 Chloroform : Methanol : Hexane)

25(OH)D Dihydroxyvitamin D Vitamin D

History of Vitamin D Testing Tribute to Dr. Deluca

Page 31: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

31

Sephadex LH-20

(50:50 Hexane: CHCl3)

OR

Lipidex 5000

(10:90 CHCl3:Hexane)

hplc on Zorbax Sil

(4:96 IP :Hexane)

Quantitate both D2 & D3

Forms by UV Absorbance

Lipidex 5000

(5:95 CHCl3:Hexane)

Hplc on Zorbax Sil

(11:89 IPA: Hexane)

24, 25-(OH)2D

Complex

25, 26-(OH)2D3 1, 25-(OH)2D

Quantitate Both D2 & D3 Forms Simultaneously by Competitive Protein Binding

Assay

Hplc on Zorabx Sil

(0.25:99.75 IPA:MeCl2)

Hplc on Zorbax ODS

(4:96 H2O:MeOH)

Quantitate both D2 & D3

By UV Absorbance

Quantitate by Competitive Protein

Binding Assay

Hplc on Zorbax Sil

(2:98 IPA:MeCl2

25-OHD3-26, 23 Lactone 24, 25-(OH)2D2 24, 25-(OH)2D3 25, 26-(OH)2D2

Quantitate Individually by competitive protein binding assay

History of Vitamin D Testing

Page 32: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

25-OH-Vitamin D is Sticky

Page 33: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

• Pipet 100 µL serum, add internal standard and incubate 15 minutes

• Precipitate proteins with 250 µL Acetonitrile

• Vortex and centrifuge 15 minutes at 3000 rpm

• Pipet 200 µL of supernatant into autosampler vials

• Initiate analysis on LC-MS/MS

• Results are produced every 3 minutes 15 seconds

LC-MS/MS Method

Page 34: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

34

Page 35: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

Sample Preparation for LC-MS/MS

• Serum samples (1 to 2 g) were weighed into a 50 mL glass centrifuge tube.

• At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked (exact mass known) with an appropriate amount of the internal standard solution (see Supporting Information) to achieve an approximately 1:1 ratio of analyte to internal standard.

• After equilibration for 1 h at room temperature, the pH of each sample was adjusted to pH 9.8 ± 0.2 with carbonate buffer (0.1 g/mL, pH 9.8).

• Analytes were isolated from the serum matrix by liquid–liquid extraction with 8 mL of hexane:ethyl acetate (50:50, volume fractions).

• Samples were extracted twice, and the extracts were combined and dried under nitrogen at 45 °C.

• The residue was reconstituted with methanol (200 to 250 μL).

Anal. Chem. 2017, 89, 4907−4913

Page 36: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

36

Challenge Throughput

Page 37: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

25-OH Vit D3-d6

internal standard

25-OH Vitamin D3

25-OH Vitamin D2

Time (min)

Pe

ak

In

ten

sit

y

LC-MS/MS VITAMIN D

Page 38: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

Chromatograms

Lensmeyer GL et al: Clin Chem 52:2305, 2006

Page 39: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

Lensmeyer GL et al: Clin Chem 52:1120, 2006

Comparison of the Proposed HPLC Method with LC-MS/MS

Page 40: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

40

Performance of LC-MS/MS Method at Two Different Sites

y = 0.976x - 0.031

R2 = 0.990

0

10

20

30

40

50

60

70

80

0 20 40 60 80

25 OHD ng/ml (Site 1)

25 O

HD

ng

/ml (S

ite 2

)

Page 41: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

41

LC-MS/MS-LC/MS/MS

-25

-20

-15

-10

-5

0

5

10

15

20

25

0 10 20 30 40 50 60 70 80

Mean 25 OH D ng/ml

Difference

ng/ml

38-42 ng/ml

Page 42: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

©2011 MFMER | 3081836-42

Methods for Testing of 25-OH-Vitamin D

1. Competitive protein-binding assay (CPB)

2. High-performance liquid chromatography (HPLC)

3. Radioimmunoassay RIA and chemiluminescence immunoassay CLIA

4. Liquid chromatography mass spectrometry (LC-MS, LC-MS/MS)

Page 43: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

Immunoassays in Clinical Laboratories and the way of the Dodo

Tandem Mass-spectrometry - the quiet revolution in endocrine-, drug- and

metabolic laboratory testing

Page 44: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

Immunoassays – central to current lab testing

• Immunoassays allow sensitive detection of a wide range of analytes

• Immunoassays can be very specific

• Immunoassays come in a wide range of formats and can be automated

• Today’s reference ranges for a wide range of analytes have been established with Immunoassays

• Immunoassay problems and limitations are well understood

Page 45: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

Methods Participating in DEQAS

45 DEQAS REVIEW 2016/2017

Page 46: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

Accreditation Agencies look for only Precision among

Peers and not Necessarily Accuracy

Page 47: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

47

Laboratory proficiency testing should

be based on the ability to measure

accurately and precisely samples

containing known concentrations of

testosterone, not only on agreement

with others using the same method

An Endocrine Society Position Statement: Precision and

Accuracy Matters

J Clin Endocrinol Metab 92:405-413, 2007

Page 48: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

48 DEQAS REVIEW 2016/2017

No of Labs Participating in DEQAS

Page 49: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

Mean %CV HPLC and LC-MS/MS

49 DEQAS REVIEW 2016/2017

Page 50: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

50 DEQAS REVIEW 2016/2017

Mean %CV HPLC and LC-MS/MS

Page 51: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

Mean CV% for Major Automated Assays

51 DEQAS REVIEW 2016/2017

Page 52: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

Mean % Bias from NIST Major Automated Assays

52 DEQAS REVIEW 2016/2017

Page 53: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

Competitive Immunoassays

(conc. limited)

www.aapsj.org/view.asp?art=aapsj0902029

Page 54: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

Characteristics of 25-OH-D Methods

Clinical Chemistry 58-531–542 (2012) 54

Page 55: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

25-OH-Vitamin D is Sticky

Page 56: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

© Copyright 2009 by the American Association for Clinical Chemistry

Accuracy of 6 Routine 25-OH-vitamin D assays: Influence of Vitamin D Binding Protein Concentration

March 2012-www.clinchem.org/content/article/58/3/543.full

Page 57: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

Clinical Chemistry May 2012 :58, 948

Page 58: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

Platform

HPLC

GC/MS

MS/MS

Competitive

Immunoassays

Impact of MS/MS in Laboratory Medicine (Mayo)

1998

>400

>50

0

Many

2009

<100

<30

58

Few

Page 59: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

MS/MS Presence in Mayo-DLMP

Page 60: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

Precision Required to See Change in a Single Patient

60

Page 61: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

The way out of the chaos Vitamin D measurement standardization:

• The lack of standardized laboratory measurement of serum 25-hydroxyvitamin D [25-OHD] in vitamin D impedes development of consensus to define stages of vitamin D status.

• 60,000 vitamin D papers have been published with nearly all the values from non-standardized assays. We cannot just ignore this old data and focus on standardizing 25-OHD in current and future research.

• Retrospective standardization results can vary widely depending on if initial 25-OHD measurements were determined using an assay that was positively or negatively biased and the level of bias in the old assay.

61 The Journal of Steroid Biochemistry and Molecular Biology Volume 173, October 2017, Pages 117-121

Page 62: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

The way out of the chaos Vitamin D measurement standardization

• An International effort is needed to identify key prior studies with stored samples for re-analysis and standardization, initially, to define the 25-OHD level associated with D deficiency (rickets/osteomalacia).

• Subsequent work could focus on defining inadequacy.

• Suspending publication of meta-analyses based on unstandardized 25(OH)D results.

62 The Journal of Steroid Biochemistry and Molecular Biology Volume 173, October 2017, Pages 117-121

Page 63: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

LC-MS/MS Vitamin 25(OH) D test results can be

“analytically accurate” but

not

“clinically relevant”

Vitamin D Test Expert Discusses Mass Spectrometry

http://www.darkreport.com/dark/12_22_2008.htm

Page 64: Challenges of implementing JCTLM methods in teh routine ... · • At least one level of SRM 972 was used as a quality control material with each set of samples. Each sample was spiked

Thanks