Upload
gavin-cole
View
219
Download
1
Embed Size (px)
Citation preview
Chapter 10
Body Temperature, Heat, Fat, and Body Temperature, Heat, Fat, and MovementMovement
Figure 10.1 The human temperature regulation system can increase or decrease body temperature.
Anteriorhypothalamus
set-point
Control center
Posteriorhypothalmus
Actuators
SweatingSkin vasodilation
Decrease metabolism
Muscle shiveringSkin vasoconstrictionIncrease metabolism
Skin thermalreceptors
Hypothalamic thermalreceptors
Coretemperature
Skintemperature
–
+
Figure 10.2 Examples of resistance–temperature curve for three NTC thermistors, = 3000 , 3500 K, and 4000 K. R0 is the resistance at T0 = 298 K.
00 T
11exp
TTRR
V oltm ete r
C onstan tan
C opperC opper
C o ld junction , T 2
(R e fe rence)
T 1 , H o t junction(M easuring p robe)
Figure 10.3 The J type thermocouple is formed from copper and constantan.
C1 and C2 are constants that depend on the thermocouple pair with T in kelvins
Probe (thermistors)
Wheatstone bridge
Signal conditioner and final temperature
detector ( processor)
Digital display
Audible alarm
Figure 10.4 An electronic thermometer uses a thermistor sensor probe.
Ear
IR
Shutter Ambient sensor
Sensor Amp.
MUX A/D
Shutter switch
Window Waveguide
Micro processor
Digital display
T a
T b
Figure 10.5 The infrared thermometer opens a shutter to expose the sensor to radiation from the ear.
A = effective body (target) area = Stefan-Boltzmann constanta = emissivity of surroundings (sensor)Tb = body temperatureTa = sensor temperature
Net flux of infrared radiation:
Constantan
Copper
Insulation layer
Thermocouple connection
Metal wall Cooling tubes
Figure 10.6 In a gradient layer calorimeter, thermocouples measure the difference in temperature across the wall. Ventilating system and measurements not shown.
air mass flow rate
Air in
Air out
T1
T2
Insulated chamber(polyurethane)
Thermometer
Thermometer
Figure 10.7 An air–flow calorimeter measures inlet and outlet temperatures, flows, and humidity.
heat loss
specific heattemperature change of the ventilating air
T1
T2
Water in
Water out
Inlet thermometer
Outlet thermometerPolyvinyl chloride tubes
Figure 10.8 The water flow calorimeter measures the inlet and outlet water temperature.
AC
Powersource
Powermeter
Thermostat
Heater
Feedback control
Figure 10.9 The compensating heater calorimeter requires less heater power when the subject supplies heat.
From
Inspired gas source
Patient Expired gas from patient
Switch
CO 2
Pressure transducer
Vacuum pump
Flow transducer
Thermistor
Exhaust
Computer control
Mixing chamber
O 2
Figure 10.10 A microcomputer-based open-circuit system includes a mixing chamber, O2 and CO2 analyzers, and the various variables (pressure, flow, temperature) used to calculate .
H um id ifie r
C O 2 and O2ana lyzers
F rom insp ired gas source
E xha la tionva lve
E xp ira to ry lim b
Insp ira to rylim b
R oom a ir
C onstan tflow
genera to r
F IO 2
FE C O 2 ,FE O 2 F *C O 2
M echan ica lven tila to r
M etabo licm on ito r
P a tien t
Figure 10.11 A dilution system used by Deltatrac uses mechanical ventilation. FIO2 is the inspired oxygen concentration, FECO2 is the true expired carbon dioxide concentration, FEO2 is the expired oxygen concentration, and F*CO2 is the diluted carbon dioxide concentration.
C O2 ana lyzer
F rom ventila tor
B ellow s
C O 2S crubber
M ix ingcham ber
S am plegas
A s oxygensupply and
analyzer
P atient
S pirom eter
Figure 10.12 A closed-circuit system uses the volumetric loss principle. The spirometer is used as an oxygen supply and the volume change as a function of time in the spirometer is used to calculate the rate of oxygen consumption.
2H2O deuterium waterH2
18O oxygen-18 enriched water
2H labeled water pool
18O labeled water and bicarbonate pools
2H disappearance (k2) = rH2O
18O disappearance (k18) = rH2O + rCO2
k18 – k2 = rCO2
Figure 10.13 Principle of the doubly labeled water method. r is the production rate, W is the size of total body water, k represents rate constants determined from the experiment.
(a)
Shoulder
Triceps
skinfold
Lange
caliper
(b)
Figure 10.14 (a) Lange skinfold caliper used for assessing thickness of subcutaneous fat. (b) Illustration of an example of skinfold measurement, triceps skinfold taken on the midline posterior surface of the arm over the triceps muscle.
Excitation ac current source (I)
Detected voltage drop (E)
Figure 10.15 In resistance measurement on the ipsilateral side of the body current flows through one arm, the trunk, and one leg.
V out = k
V i
(a)
Plastic ribbons
Connecting rods Parallelogram
linkage
Triaxial goniometer
(b)
Knee joint
Figure 10.16 (a) A goniometer attached to the shank and thigh to measure knee rotation. Vi is the input voltage. Vout is the output voltage that is proportional to the angle of knee rotation. (b) Subject wearing a triaxial goniometer on knee joint.
Seismic
mass
Cantilever
Strain
V a
E
Figure 10.17 Vertical acceleration of the accelerometer frame bends the cantilever beam because the seismic mass remains at rest. Voltage output (V) is proportional to the acceleration (a). E is the supply voltage.
Figure 10.18 An example of a gait analysis setup includes a four-camera kinematic system, two force platforms, and an electromyogram (EMG) telemetry system.
A/D converter
EMG receiver
Camera control
and Interface
Display system
Walk way
Analog interface
Camera 1
Camera 2 Camera 3
Camera 4
EMG transmitter
(option)
Force platforms (option)
Markers
Computer