Chapter 5
UV/visible Absorption spectroscopy-
Part 2
1
Apparatus and Instruments used for
Uv/Vis measurements
2
3
4
5
6
7
8
9
10
11
12
EMR
13
14
15
Deuterium Lamps
Used to obtain ultraviolet absorption spectra. Above 350 nm its
output is too weak for accurate measurements of absorption. The
irradiance is too low for fluorescence measurements. This is a low
pressure discharge lamp that produces light by the reaction,
D2 + e D2* 2D + hn
16
Quartz Halogen Lamps
Used to obtain visible absorption spectra. Its output below 300 nm is too weak for accurate measurements of absorption. The irradiance is too low for most fluorescence measurements. Power supply: 12 V dc at 4 A, and highly regulated.
17
Xenon Arc Lamps
it is difficult to obtain high quality spectra in those regions. This is a high Used to obtain fluorescence excitation, emission spectra, and make quantitative measurements. Below 300 nm the output is too weak to make accurate measurements. The lines between 450 and 500 nm and above 650 nm make pressure discharge lamp.
18
Mercury Arc Lamps
Used to perform quantitative fluorescence measurements, or obtain
emission spectra. The line output precludes obtaining excitation
spectra. Note how the 2537 line is inverted! The UV, blue and green
lines emit 10-100 times more light than a xenon arc lamp. This is a
high pressure discharge lamp.
19
Calibration Pen Lamps
Pen lamps can be purchased with a wide variety of gases that emit
sharp line spectra. The line output is commonly used to calibrate
wavelength separators, such as grating monochromators.
A mercury pen lamp emits intense lines in the ultraviolet through the
blue at 2537, 3130, 3650, 4047, and 4358 . To calibrate in the red use
a neon lamp.
20
21
22
23
24
Absorption Filters
The figure shows the transmission curve for three bandpass
filters. 51715 passes everything except the deep UV. It would
be used to reject the mercury 253.7 nm line. 51670 passes a
band from the near UV to the green, while 51660 passes a band
from ~280 to 400 nm. These would be used with fluorescence
excitation.
25
Absorption filters
The figure shows the transmission curve for two high pass filters.
51294 rejects wavelengths below 500 nm and passes those above.
5131 performs the same function with a cutoff wavelength of ~580 nm.
Note that it is nearly impossible to make a low pass colored glass
filter.
26
An interference filter is created when light is multiply reflected between two parallel surfaces. The beams exiting the filter must be in phase. Only specific wavelength can pass through
The pass band varies with the order, N. To restrict transmission to only one order, the exit surface is most often a colored glass filter.
Interference Filter
27
28
Interference Filter
Center wavelengths can be from the UV to the infrared. Pass bands
are narrower than absorption filters, but broader than grating
monochromators. The filter above is centered at 415 nm with a 10 nm
FWHM and a transmission ~33%. As the bandwidth narrows the
transmission drops.
29
30
31
32
33
34
Prisms
Because of a complicated mathematical relationship between wavelength and bend angle, prisms are seldom used in modern instruments. For maximum dispersion (angular resolution), visible light is separated with a glass prism and ultraviolet is separated with a quartz prism. One common design is the constant deviation, or Pellen-Broca, prism shown in the figure. Constructed from one piece of material, it can be viewed as three separate prisms. For a given orientation, only one wavelength exits at a 90 deviation. To select a different wavelength the prism is rotated about the juncture of the dotted lines, P.
30 45
45
60
30
60
P
35
36
37
38
39
Diffraction Grating
m = 0
+1
+2
-1
-2
+1
a
+1
+2
-1
-2
+1
a i
r
A diffraction grating is created by a spatial modulation in phase or amplitude of an incoming plane wave. When exiting the grating, the input is separated into several plane waves traveling at angle to the original direction of propagation - these are called diffraction orders. The number and intensity of the orders depends upon the functional form of the modulation. The figure at the top is a transmission grating, while that at the bottom is a reflection grating. The diffraction angle depends upon order, m, wavelength, l, and spatial modulation period, a.
sin m m a l
40
Blazed Gratings
+1
+2
-1
-2
+1
a
i
r
By forming the bottom of the grooves into a saw tooth shape, the angle of reflection can be made to correspond to one of the diffraction orders. One order then blazes, that is, it has most of the incoming optical power. Show below are three gratings blazed at 300, 400 and 500 nm. Note how the efficiency drops on either side of the blaze wavelength.
41
Grating Resolution
7.3 : 9
The angular distribution of a monochromatic beam of light reflecting off a diffraction grating is a sinc function (sinx/x). The node spacing of the sinc function is called the angular resolution, D, where N is the number of grating grooves and a is the groove spacing. The goal is to minimize D, which is accomplished by maximizing the grating width, Na. You can also increase the order, which increases m. Grating spectral resolution is defined as, where Dl is the range of wavelengths occurring over D, and l is the center wavelength. High resolution means small Dl. The angle at which light leaves a grating leads to an ambiguous wavelength, e.g. 600 nm, m = 1; 300 nm, m = 2; 200 nm, m = 3.
2
cos mNa
l
D
R mNl
l D
42
Commercially Available Gratings
The vast majority of commercially available gratings are plastic replicas. The master grating is coated with an epoxy resin to form a negative image. A thin layer of plastic coats the negative and cured. When removed from the negative, the replica is mounted on glass for rigidity and coated with aluminum. Ruled gratings are made with a spacing from 25 mm-1 to 1,800 mm-1. The upper size is 1515 cm, with 55 cm being typical. Concave ruled gratings can be manufactured with great difficulty. Ruled gratings have a scattered light figure of 10-3. Holographic gratings are made by coating the surface of glass with a photoresist. Two laser beams irradiate the surface at an angle to each other creating interference fringes. Bright fringes polymerize the photoresist. The surface is washed with an organic solvent to dissolve remaining monomer, then coated with aluminum. Spacings up to 2,400 mm-1 are possible, as well as concave gratings. Holographic gratings have a scattered light figure of 10-4.
43
44
45
46
47
48
49
50
• Reverse bias creates a depletion
layer that reduces the
conductance of the junction
nearly to zero
• When radiation impinges on the
chip holes and electrons are
formed in the depletion layer and
those provide a current that is
proportional to radiant power
51
52
53 IC = Integrated Circuit
54
55
56
57
58
General Instrument Designs
59
60
Spectronic 20 optical diagram
61
62
1. Double- beam-in-space configuration
This requires two detectors that must be
perfectly matched
2. Double- beam-in-time configuration
Sample and reference measurements are
separated in time. Rapidly rotated mirror or
“chopper” is used. Only one detector is used
63
General instrument designs double beam-in-space
64
General instrument designs duoble beam-in-time
65
66
67
68
Multichannel photon detector
It consists of an array of tiny photosensitive
detectors that are arranged in a pattern that
all elements of a beam of radiation that has
been dispersed by a grating can be
measured simultaneously
69
Multichannel diode array spectrometer
(Multichannel photon detector)
70
Characteristics of UV/Vis Methods:
Wide applicability to organic and inorganic systems
Sensitivities to 10-4 to 10-7 M
Moderate to high selectivity
Good accuracy, about 1-3% relative uncertainty
Ease and convenient data acquisition
71
Derivative and Dual wavelength Spectrohotometery
• The use of dual dispersing systems are arranged in
such a way that two beams of slightly different
wavelengths (typically 1 or 2 nm) fall alternatively
onto a sample cell and its detector; no reference
beam is used.
• The ordinate parameter is the difference between the
alternate signals, which provides a good
approximation of the derivative of absorbances as a
function of wavelength (DA/ Dl).
72
Dual Wavelength spectrophotometer
•Two beams of differing wavelengths having same
intensity are passed alternatively through a single
sample cell (previous types used single- wavelength)
•Two monochromators are used. One monochromator
may be used but light from the monochromator is
chopped and the monochromator is shifted between
the two wavelengths
• A single covette is used thus, scattering and
instrumental stray light are cancelled
73
Dual-Wavelength Spectrophotometer using two monochromators
74
75
Derivative spectroscopy has been mostly
used in the ultraviolet and visible regions for
qualitative identification of species, in which
the enhanced detail of a derivative spectrum
makes it possible to distinguish among
compounds having overlapping spectra.
Dual-wavelength spectrophotometry has
proven particularly useful for extracting
ultraviolet/visible absorption spectra of
analytes present in turbid solutions, where
light scattering obliterates the details of an
absorption spectrum.
76
Derivative Spectroscopy
77
78
79
Photoacoustic Spectroscopy
• Photoacaustic spectroscopy is based on light absorption effect.
• This effect is observed when a gas in a closed cell is irradiated with a chopped beam of radiation of a wavelength that is absorbed by the gas.
• The absorbed radiation causes periodic heating of the gas, which in turn results in regular pressure fluctuations within the chamber.
• If the chopping rate lies in the acoustical frequency range, these pulses of pressure can be detected by a sensitive microphone.
• Photoacoustic or optoacoustic spectroscopy, which was developed in the early 1970s, provides a means for obtaining ultraviolet and visible absorption spectra of solids, semisolids, or turbid liquids.
• Acquisition of spectra for these kinds of samples by ordinary methods is usually difficult at best and often impossible because of light scattering and reflection.
80
• In photoacoustic studies of solids, the sample
is placed in a closed cell containing air or
some other nonabsorbing gas and a sensitive
microphone.
• The solid is then irradiated with a chopped
beam from a monochromator.
• The photoacoustic effect is observed provided the radiation is absorbed by the solid; the
power of the resulting sound is directly related
to the extent of absorption.
• Radiation reflected or scattered by the sample
has no effect on the microphone and thus does
not interfere. This latter property is perhaps the
most important characteristic of the method.
81
82
Photoacoustic Spectrometer
• Light source is modulated at an audio frequency
• The sample absorbs the radiation and becomes a
heat source producing alternating regions of
compressions and refractions in the enclosed gas
that is acoustic or sound wave
• The acoustic signal is converted into electrical
signal by microphone
• Solid sample is directly irradiated with the
modulated source resulting in production of
acoustic waves in the surrounding gas
83
84
85
86