46
7/27/2019 @Chapter 7, Cable Testing Fundamental http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 1/46 7-1 Chapter 7 Cable Testing Fundamentals  Contents Testing Twisted Pair Cables Wire Map Test Wire Length Test DC Resistance Test  NEXT, ELFEXT, and Power Sum Test Attenuation Test Return Loss Test Impedance Test Delay and Skew Test Capacitance Test ACR and Power Sum ACR Test Headroom Test Testing and Troubleshooting 10Base-T Cabling Testing and Troubleshooting Nexans Cabling System Testing and Troubleshooting with a Block Connector System Testing and Troubleshooting Coax Cabling Testing and Troubleshooting Fiber Optics

@Chapter 7, Cable Testing Fundamental

Embed Size (px)

Citation preview

Page 1: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 1/46

7-1

Chapter 7Cable Testing Fundamentals

Contents

• Testing Twisted Pair Cables

Wire Map Test

• Wire Length Test

• DC Resistance Test

• NEXT, ELFEXT, and Power Sum Test

• Attenuation Test

• Return Loss Test

• Impedance Test

• Delay and Skew Test

• Capacitance Test

• ACR and Power Sum ACR Test

• Headroom Test

• Testing and Troubleshooting 10Base-T Cabling

• Testing and Troubleshooting Nexans Cabling System

• Testing and Troubleshooting with a Block Connector System

• Testing and Troubleshooting Coax Cabling

• Testing and Troubleshooting Fiber Optics

Page 2: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 2/46

Page 3: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 3/46

Chapter 7Cable Testing Fundamentals

7-3

Table 7-1: Twisted Pair Cable Types and Associated Networks

Cable Type Network ExamplesTIA Cat 3, 5e, 6 UTP or STP and 7 STP Ethernet, Fast Ethernet, ATM, and Gigabit

Ethernet

ISO Class C, D, E and F UTP or STP Ethernet, Fast Ethernet, ATM

TP-PMD / TP-DDI FDDI or ATM on Copper

10Base-T Ethernet

Single Pair Telephone, Apple Local Talk, ISDN

Shielded Two-Pair (1,2,7,8) ATM, Fiber Channel on Copper

Shielded Twisted Pair (STP)

Testing the continuity of the shield is important and requires shielded test leads at

both the Display Handset and Remote Handset. When testing STP, be sure to

select Shielded Cable Type (STP) in the Cable Type menu.

Twists are maintained to within ½”as required per TIA Category 5e insulation guidelines

Screened Category 5e Cable(ScTP)

Metallic shield provides EMIprotection

Figure 7-2: Shield Continuity Test Connections Twisted Pair

USOC Wiring

If a USOC (Universal Service Ordering Code) or other wiring scheme is used, a

special adapter may be required for connection. Refer to Appendix D, Model

Accessories for a full list of available cable adapters.Note: If your testing requirements include connection to something other than an RJ-45

or Tera style jack, refer to Testing with Block Adapters later in this chapter.

Page 4: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 4/46

Chapter 7Cable Testing Fundamentals

7-4

Permanent Link Test Setup

ANSI, EIA, TIA, and ISO all provide two network communication circuit

specifications: permanent link and channel link. A permanent link consists of up

to 90 meters of horizontal network cabling. The permanent link (shown below) is

used to certify the horizontal network cable installation before network

connection and user hookup. It excludes adapters, patchcords, and jumpers.

Display

Handset

RemoteHandset

Channel Link Adapter and2 Meter Patchcord

NetworkPatch Panel

RJ-45Wall Outlet

Horizontal Network Cable(Maximum of 90 Meters)

AUTOTEST

NETMONITOR

WIREMAP

TDR

LENGTH

SHIFT

750 MHz Certifier

F1 F2 F3 F4F5 F6 F7 F8

ENTER

Escape

TALK

ANALYZE

SETUP

HELP

1 2 3

4 5 6

7 8 9

0

ABC D EF G HI

J KL M NO P QR

ST U VWX YZ

SPACE

AUTOTEST

ESCAPE

TONE MODE

TONE

PAGE

TALK

SHIFT

Remote

HAZARD PAS S FAIL ON

750 MHz Certifier

Channel Link Adapter and2 Meter Patchcord

Figure 7-3: Permanent Link Test Connections

Connection Cable Length Limits

Horizontal Network Cable Maximum of 90 meters

Note: Ensure that the Cable Type is set to Twisted Pair Permanent Link.If you exceed the tester length test limits, the tester will fail the link.

Page 5: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 5/46

Chapter 7Cable Testing Fundamentals

7-5

Channel Link Test Setup

A channel link includes all aspects of the cabling system. It consists of up to 90

meters of horizontal network cabling, user patchcords, jumpers, and channel

adapters at each end. The channel link (shown below) is used to certify the

network installation, including the horizontal link and user patchcords.

DisplayHandset

RemoteHandset

User Patch Cord

Channel Adapter

Channel Adapter User Patch Cord

NetworkPatch Panel

RJ-45Wall Outlet

Horizontal Network Cable(Maximum of 90 Meters)

AUTOTEST

NETMONITOR

WIREMAP

TDR

LENGTH

SHIFT

750 MHz Certifier

F1 F2 F3 F4

F5 F6 F7 F8

ENTER

Escape

TALK

ANALYZE

SETUP

HELP

1 2 3

4 5 6

7 8 9

0

A BC D EF GHI

J KL M NO P QR

S TU V WX YZ

SPACE

AUTOTEST

ESCAPE

TONE MODE

TONE

PAGE

TALK

SHIFT

Remote

HAZARD PAS S FAIL ON

750 MHz Certifier

Figure 7-4: Channel Link Test Connections

Connection Cable Length Limits

Horizontal Network Cable Maximum of 90 meters

User Patchcords Combined Maximum length of 10 meters

Note: Ensure that the Cable Type is set to Twisted Pair Channel Link when testing with

channel adapters. If you exceed the tester length test limits, the tester will fail thelink.

Page 6: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 6/46

Page 7: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 7/46

Page 8: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 8/46

Chapter 7Cable Testing Fundamentals

7-8

Troubleshooting Wiremap Problems

Problem: One or more open pins

Probable Causes Connector-to-wire punch down not mated

Defective jack or plug.

Broken wire(s).

Other Tests Affected Test Possible ResultDC Resistance Fail.

Attenuation Fail.

NEXT Some false measurements.

Mutual Capacitance 0 reading possible.

Length May be low if the open is near the

Display Handset.

Problem: Shorted pins

Probable Causes Conductors making contact at a connector.

Jack or plug has pin or circuit defect.

Cable damaged.

Other Tests Affected Test Possible ResultDC Resistance Low or zero.

Attenuation Fail.

NEXT Some false measurements.

Capacitance Over limit.

Length Reduced or shorted pairs.

Problem: Miswired pinsProbable Causes Conductors reversed at a connector.

Other Tests Affected Test Possible Result

Usually none Infrequently, one or more tests

may fail.

Page 9: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 9/46

Page 10: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 10/46

Page 11: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 11/46

Chapter 7Cable Testing Fundamentals

7-11

DC Resistance Test

This test measures the loop resistance of each pair of wires. The instrument tests

to make sure total loop resistance does not exceed recommended limits. Results

are displayed with resistance in ohms for each pair, and a comparison limit for the

cable type.

Note: The RH is required to perform this test.

Page 12: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 12/46

Chapter 7Cable Testing Fundamentals

7-12

DC Resistance Errors

All four pairs of a network link should have approximately the same resistance.

Pair resistance that exceeds the limit is indicated as a failure. The maximum limits

in the default tables are based on the maximum length limit of the link or cable

segment.

Troubleshooting DC Resistance Problems

Problem: Excessive Resistance

Probable Causes Mismatched cable types.

Poor punch block connection.

Poor RJ-45 termination connections.

Wire pair has a tap (never done).Cable damage.

Shorted cable.

Other Tests Affected Test Possible Result

Wire Map May fail.

Attenuation May fail.

NEXT May have false readings.

Capacitance May fail.

Problem: One wire pair has a very high DC loop resistance, othersare normal.

Probable Causes Poor connection points.

Cable damage.

Connector blades not fully piercing wire insulation.

Worn Connector

Other Tests Affected Test Possible Result

Wire Map May fail.

Attenuation May fail.

NEXT May have false readings.

Capacitance May fail.

Page 13: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 13/46

Page 14: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 14/46

Ch

Page 15: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 15/46

Chapter 7Cable Testing Fundamentals

7-15

The FEXT test is similar to the NEXT test except that the traffic is generated at

the RH and crosstalk is measured at the DH.

• NEXT measurements are made at each end of the cable for all pair

combinations (pair 1-2 vs. 3-6, etc.), yielding a total of twelve measurements.

• ELFEXT measurements are made with the DH and RH for all possible pair

combinations (1-2 vs. 3-6, 3-6 vs. 1-2, 1-2 etc.) at both ends yielding a total

of twenty-four measurements.

Power Sum NEXT and Power Sum ELFEXT

Power Sum tests measure the crosstalk effects of three transmitting pairs on the

fourth pair in the same cable sheath.

1

2

3

6

4

5

7

8

2

1

3

6

4

5

7

8

Effects of 3 Pairs on 1 Pair

Page 16: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 16/46

Page 17: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 17/46

Chapter 7

Page 18: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 18/46

Chapter 7Cable Testing Fundamentals

7-18

NEXT and ELFEXT Errors

Crosstalk is usually caused by poor connector termination on the ends of the

cable. The smaller the number, the greater the crosstalk.

Troubleshooting NEXT and ELFEXT Problems

Problem: Low dB test readingsProbable Causes Installed cable or patch cable not correctly rated.

Defective, poor quality cable or too many connectors.

Poor quality installation at the connection points.

Too much insulation has been stripped from the wires at

termination.

A pair of wires has been untwisted too much at termination.

Split-pairs.Poor quality connectors or connectors not rated to desired

category.

Delay skew (ELFEXT).

Other Tests Affected Test Possible Result

Return Loss May be over limit.

NEXT May show same symptoms.

Chapter 7

Page 19: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 19/46

Chapter 7Cable Testing Fundamentals

7-19

Attenuation Test

This test measures the overall signal strength loss in the cable and verifies that it

is within acceptable limits. Low attenuation is essential for error-free

transmission. Attenuation is measured by injecting a signal of known amplitude at

the Remote Handset and reading the amplitude at the Display Handset.

Note: The RH is required to perform this test.

Chapter 7

Page 20: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 20/46

C apteCable Testing Fundamentals

7-20

Attenuation Errors

Attenuation causes a loss of signal strength over a cable. The loss increases with

cable length, signal frequency, and temperature. Attenuation testing can be used

to find problems in the cable, connectors, or connecting hardware. The larger the

number, the greater the attenuation.

Troubleshooting Attenuation Problems

Problem: High Attenuation Reading

Probable Causes Poor connector termination points.

Excessive cable length.

Incorrect or poor quality adapter cable.

Incorrect cable.

Other Tests Affected Test Possible Result

DC Loop Resistance May be high.

Capacitance May be high.

Length May be over limit.

NEXT May be low on pair combinations.

Average Impedance May be low.

Return Loss May be over limit.

Page 21: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 21/46

Page 22: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 22/46

Chapter 7

Page 23: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 23/46

Cable Testing Fundamentals

7-23

Impedance Test

Average impedance is derived from electrical delay and capacitance

measurements. The results of this test are expressed in ohms. Average impedance

testing can help identify physical damage to the cable, connector defects, or cable

segments with incorrect characteristic impedance.

This test uses capacitive measurements; therefore, it is necessary to specify the

correct cable type in order to accurately perform the test.

Note: If a CAT 3 cable is selected (specified as the cable type where PVC is used in the

cable insulation) but a CAT 5 cable (where Teflon is used as the cable insulation)is actually used, the average impedance will be calculated incorrectly. To avoid this problem, be sure to specify the correct cable type.

Note: The RH is not required to perform this test.

Chapter 7C bl T ti F d t l

Page 24: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 24/46

Cable Testing Fundamentals

7-24

Impedance Errors

Impedance errors cause signal reflection and strength reduction. Averageimpedance of each pair should be equal to the LAN system impedance of 100,

120, or 150 Ω, plus or minus 15 Ω.

Troubleshooting Impedance Problems

Problem: High Impedance Readings

Probable Causes Compression, stretching, or excessive bending damage to

the cable.

Defective connectors.

Insulation damage at a connector.

Ground loops created between cable shielding (if used) and

equipment grounding (via RS-232 cable to computer, or auxiliary power).

Improperly chosen cables or patch cords.

Moisture in the cable.

Other Tests Affected Test Possible Result

Length Affected pairs will appear longer.

Average Impedance Change in average impedance isinversely proportional to change in

capacitance.

Page 25: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 25/46

Page 26: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 26/46

Page 27: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 27/46

Chapter 7Cable Testing Fundamentals

Page 28: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 28/46

Cable Testing Fundamentals

7-28

Capacitance Errors

The larger the capacitance, the higher the error rate. Small changes in the

capacitance measurements are normal due to the handling of the cable during

shipping and installation. The addition of connectors and patch cables will also

affect capacitance values.

Troubleshooting Capacitance Problems

Problem: Capacitance Exceeds the Maximum Limit

Probable Causes Compression, stretching, or excessive bending damage to

the cable.

Defective connectors.

Insulation damage at a connector.

Ground loops created between cable shielding (if used) andequipment grounding (via RS-232 cable to computer, or

auxiliary power).

Improperly chosen cables or patch cords.

Moisture in the cable.

Poor connections at punch downs and wall plates

Other Tests Affected Test Possible Result

Length Affected pairs will appear longer.Average Impedance Change in average impedance is

inversely proportional to change in

capacitance.

Chapter 7Cable Testing Fundamentals

Page 29: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 29/46

g

7-29

ACR and Power Sum ACR Test

The ACR (Attenuation-to-Crosstalk Ratio) test performs a mathematical

comparison (difference calculation) between the results of the Attenuation and

NEXT tests. The difference reading between each pair gives an indication of how

problem-free the cable pair will be for transmissions.

The ACR measurements are calculated pair-to-pair.

Note: The RH is required to perform these tests.

Page 30: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 30/46

Chapter 7Cable Testing Fundamentals

Page 31: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 31/46

7-31

Headroom Test

The Headroom measurement is a mathematical analysis of the data already

existing from previous tests. The calculated value is the sum of the Power Sum

ACR test (Power Sum ACR of the worst pair after the attenuation for that pair has

been normalized to 100 meters or 328 feet) and the additional margin between the

worst case PS NEXT and the limit for PS NEXT.

Headroom provides a simplified means of reporting the margin available in asingle cable run which will support an application with error-free performance. It

also gives an indication of additional margin which may be achieved through the

utilization of “enhanced” cable and connectors and careful installation practices.

Note: The RH is required to perform this test.

Chapter 7Cable Testing Fundamentals

Page 32: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 32/46

7-32

Headroom Errors

The Headroom number, reported in dB, characterizes the worst-case margin

found in a single cable run. A large number is desirable, since it indicates a

strong signal and little noise interference. The pass/fail limit for Headroom is the

same as Power Sum ACR.

Testing and Troubleshooting 10BASE-T Cabling

10BASE-T Ethernet systems use twisted pair cabling for transmission of network

data frames. Both the cable and connecting hardware must meet minimum

standards as specified in the IEEE 802.3 standard. The default settings for

10BASE-T network links in the LANTEK tester reflect these standards.

10BASE-T systems use the 1 and 2 pins for transmit and the 3 and 6 pins for receive, as shown in Figure 7-5. The instrument passes or fails the Wire Map

based on this pin configuration. If your system does not use the IEEE 802.3

wiring standard, a custom adapter is required to align nonstandard transmit and

receive pairs.

Figure 7-5: 10BASE-T Connector

Note: Other pairs may be wired, but 10BASE-T uses only the pairs shown.

Chapter 7Cable Testing Fundamentals

Page 33: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 33/46

7-33

Testing and Troubleshooting Nexans Cabling System

A field calibration should be performed prior to implementing a test of the

Nexans cabling system. This process will ensure (1) synchronizing of the units,

(2) qualifying (testing) of the patchcords and (3) gathering of loss data regarding

the patchcords and mated connections.

The field calibration is a 4-step process. Process 1 and 2 are performed with the

Nexans Calibration Adapters connected to the Handsets. Process 3 and 4 are performed with the Nexans Permanent Link Adapters open-ended patchcords and

then the Nexans Calibration Load Terminator attached.

The Nexans testing kit comprises a Category 7 connector product which is

backward compatible with Category 6 – RJ45s.

The equipment required for a Nexans calibration procedure are:

• Display Handset

• Remote Handset

• Nexans Calibration Adapters

(Two adapters joined by a short segment of Category 7 cable)

• Nexans Calibration Load Terminator

(100Ω Jack)

• Nexans Permanent Link Adapters

(A set of two adapters, each with a patchcord (approximately 2 meters)soldered to the adapter at one end and a Nexans Category 7 plug at the

opposite end)

Page 34: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 34/46

Page 35: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 35/46

Chapter 7Cable Testing Fundamentals

Page 36: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 36/46

7-36

Testing a Nexans Cabling System

Figure 7-6: Typical Configuration for Nexans Cable Testing

Typical configuration for testing has the DH unit and RH unit connected to the

Nexans Permanent Link Adapters. Each Nexans Permanent Link Adapter has a

patchcord (approximately 2 meters) soldered to the adapter at one end and a

Nexans Category 7 plug at the opposite end. The cable under test is connected at

the Category 7 plug end of both the DH and RH ends.

Chapter 7Cable Testing Fundamentals

Page 37: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 37/46

7-37

Testing and Troubleshooting with a Block Connector

SystemAt times, it is necessary to test directly from a connecting block to either a patch

panel or office outlet.

A field calibration should be performed prior to implementing a test of the block

system. This process will ensure (1) synchronizing of the units, (2) qualifying

(testing) of the patchcords and (3) gathering of loss data regarding the patchcordsand mated connections.

The field calibration is a 4-step process. Process 1 and 2 are performed with the

patchcords connected to the Handsets. Process 3 and 4 are performed with open-

ended patchcords (Only one end connected to the Handsets).

Note: The following process describes the calibration and testing procedures for theblock connector system. These procedures can be used for either the 110, 210,BIX, or 66 Block systems.

The equipment required for a block calibration procedure are:

• Display Handset

• Remote Handset

• RJ45 to RJ45 Patchcord

• RJ45 to Block Plug Patchcord

• A Block Calibration Adapter (T568A or T568B) depending on the

connection system that will be scheduled for testing.

Page 38: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 38/46

Chapter 7Cable Testing Fundamentals

Page 39: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 39/46

7-39

Field Calibration using Block Adapters

Note: A Block Adapter Kit can be ordered from IDEAL INDUSTRIES. The kit containsadapters for the following block systems: 110, 210, BIX and 66.

Step 1

Connect the channel adapters to the Display Handset (DH) and Remote Handset

(RH).

Power both units on.

Connect the RJ45 to RJ45 Patchcord that you plan to use as the RH Patchcord to

the adapters of the DH and RH units.

From the DH Ready screen, select Field Calibration .

The Calibration screen appears.

Chapter 7Cable Testing Fundamentals

Page 40: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 40/46

7-40

From the DH Field Calibration

screen, select to begin the

calibration process on the first (RH)Patchcord. This first process takes

about 30 seconds to complete.

At the completion of the first calibration process, tag the RH end of the first

patchcord. Disconnect the first patchcord from the DH and RH unit adapters. This

tag will remind you which end to reinsert into the RH for Step 4.

Step 2

Insert the block calibration adapter either T568A or T568B into the RH unitadapter. Insert the second (DH) RJ45 to Block Plug patchcord into both the DH

and RH adapters. The RJ45 end into the DH unit adapter, the block plug into the

block calibration adapter connected to the RH.

Note: If field testing a TIA-568A connection system, use a T568A calibration adapter. If field testing a TIA-568B connection system, use a T568B calibration adapter.

Page 41: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 41/46

Chapter 7Cable Testing Fundamentals

Page 42: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 42/46

7-42

Step 4

From on the RH, press to begin the fourth calibration process.

If calibration is successful, the DH will briefly Display “Calibration Complete”

and the RH will briefly display the PASS light. The handsets and the patchcords

are ready for testing procedures.

If calibration is unsuccessful, the DH will briefly display either a Warning screen

displaying “No Remote Handset” or a Calibration Failure screen.

Page 43: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 43/46

Chapter 7Cable Testing Fundamentals

Page 44: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 44/46

7-44

Testing and Troubleshooting Coax Cabling

Field Calibration using Coax Adapters

Field Calibration using Coax adapters uses a modification of the LANTEK 4-step

process. Since Coax testing is done for low frequencies, the additional data

obtained during calibration processes is essentially ignored therefore the

LANTEK will request one calibration process step to be performed.

To calibrate the tester, perform the following:

Connect the Coax adapters to the

Display Handset (DH) and Remote

Handset (RH).

Power both units on.

Insert the short calibration cable into

both the DH and RH adapters.

From the DH Ready screen, select

Field Calibration. The Calibration

screen appears.

From the DH Field Calibration screen,

select to begin the calibration

process.

When calibration process is completed, the LANTEK is ready for testing.

Chapter 7Cable Testing Fundamentals

Page 45: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 45/46

7-45

Testing and Troubleshooting Fiber Optics

FIBERTEK TM allows you to perform optical power loss measurements for both

Singlemode and Multimode fiber optic cables on either the LANTEK ®

6 or 7

Cable Certifiers.

The fiber testing performed makes use of laser sources for all wavelengths,

permitting certification of Gigabit Ethernet applications on the fiber optic cable.

TRACETEK TM is an advanced troubleshooting tool designed to quickly identify

and provide assistance in diagnosing common cabling problems.

Fiber Optics testing and troubleshooting kit(s) are available and contain the

FIBERTEK TM

and/or TRACETEK TM

products as well as detailed information

within a manual structure regarding their function.

For further details and/or procurements of these products call your local IDEAL

INDUSTRIES representative.

Chapter 7Cable Testing Fundamentals

Page 46: @Chapter 7, Cable Testing Fundamental

7/27/2019 @Chapter 7, Cable Testing Fundamental

http://slidepdf.com/reader/full/chapter-7-cable-testing-fundamental 46/46

7-46