Upload
vothien
View
216
Download
2
Embed Size (px)
Citation preview
Rapid Programmer™ System, Version 3.7Model 3834
Clinician’s Manual
CAUTION: The Rapid Programmer™ system is part of a regulated
medical device system and any changes to the system may be
performed only with express written authorization from St. Jude
Medical. Changes to the system include the installation of any
software, addition of hardware that is not approved by St. Jude
Medical, or modifications to the operating system of the personal
digital assistant (PDA). The result of these actions may cause the
PDA to malfunction or become inoperable.
CAUTION: Federal (USA) law restricts this device to sale by or on
the order of a physician.
Information in this document is subject to change without notice and does not represent a commitment on the part of St. Jude Medical. The Rapid Programmer system software, excluding any user-input data, is furnished under a license agreement or nondisclosure agreement and may be used or copied only in accordance with the terms and conditions of the applicable agreement.
Bluetooth® is a registered trademark of Bluetooth SIG, Inc.
Unless otherwise noted, ™ indicates that the name is a trademark of, or licensed to, St. Jude Medical or one of its subsidiaries. ST. JUDE MEDICAL and the nine-squares symbol are trademarks and service marks of St. Jude Medical, Inc. and its related companies.
Pat. http://patents.sjm.com
© 2016 St. Jude Medical, Inc. All rights reserved.
Rapid Programmer System Clinician’s Manual i
Contents
Prescription and Safety Information . . . . . . . . . . . . . . . . . . . . . . . . 1Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1Prescription Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2Device Compatibility Guidelines. . . . . . . . . . . . . . . . . . . . . . . . . . 4
Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Contents of Package . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Parts of the Rapid Programmer System . . . . . . . . . . . . . . . . . . . . 7
Setting Up the Rapid Programmer System . . . . . . . . . . . . . . . . . . . . 9Inserting the System Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Charging the System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Turning the System On and Off . . . . . . . . . . . . . . . . . . . . . . . . . . 11Connecting a Patient Programmer . . . . . . . . . . . . . . . . . . . . . . . . 12Connecting an MTS System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13Inserting a Database Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15Removing a Database Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16Replacing the System Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Getting Started . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18Navigating the Main Menu Screen . . . . . . . . . . . . . . . . . . . . . . . . 18Adjusting Backlight Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Adjusting Screen Brightness . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20Setting the Time and Date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Checking the Programmer Battery Status . . . . . . . . . . . . . . . . . . . 22Adjusting the Automatic Shutoff Settings . . . . . . . . . . . . . . . . . . . 23Calibrating the Touch Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . 24Adjusting the Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25Starting the Demo . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Clinician’s Manual
ii Rapid Programmer System Clinician’s Manual
Intraoperative and Postoperative Programming . . . . . . . . . . . . . . . . . 27Buttons and Icons on the Programming Screens . . . . . . . . . . . . . . 27Starting an Intraoperative Testing Session . . . . . . . . . . . . . . . . . . 29Starting a New Programming Session or Resuming the
Previous Session . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32Selecting the Lead Configuration . . . . . . . . . . . . . . . . . . . . . . . . . 34Resetting the Hours Used for the MTS System . . . . . . . . . . . . . . . 34Adding a New Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35Copying a Selected Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35Editing a Selected Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36Deleting a Selected Program, Stim Set, or Burst Pulse . . . . . . . . . 36Switching Between Tonic Mode and
BurstDR™ Stimulation Mode . . . . . . . . . . . . . . . . . . . . . . . . . . 37Configuring Electrode Polarities . . . . . . . . . . . . . . . . . . . . . . . . . . 38Adjusting Pulse Width. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39Adjusting Frequency for Tonic Programs . . . . . . . . . . . . . . . . . . . . 39Adjusting Frequency for BurstDR™ Stimulation Programs . . . . . . . 39Adjusting Amplitude and Step Size . . . . . . . . . . . . . . . . . . . . . . . 40Setting Perception and Comfort Amplitudes for Tonic Programs . . . 40Setting the Target Amplitude for
BurstDR™ Stimulation Programs . . . . . . . . . . . . . . . . . . . . . . . 41Adding a Stim Set or a Burst Pulse to a Program . . . . . . . . . . . . . 41Using the MultiSteering Programming Feature . . . . . . . . . . . . . . . 42Creating a Stimulation Map . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43Using the Dynamic MultiStim Programming Feature . . . . . . . . . . . 45Viewing the Test Log. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47Measuring Impedance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48Modifying IPG Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50Saving a Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52Using the Active Balancing Programming Feature . . . . . . . . . . . . . 52Downloading Programs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54Overriding the Usage Range for Amplitude . . . . . . . . . . . . . . . . . . 55
Managing Patient Records . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57Buttons and Icons on the Patient Database Screens . . . . . . . . . . . 57Accessing the Patient Database . . . . . . . . . . . . . . . . . . . . . . . . . . 58Changing the Password for the Patient Database . . . . . . . . . . . . . . 60Adding a Patient to the Database. . . . . . . . . . . . . . . . . . . . . . . . . 61Editing a Patient’s Information . . . . . . . . . . . . . . . . . . . . . . . . . . 62Deleting a Patient From the Database . . . . . . . . . . . . . . . . . . . . . 63Combining Patient Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64Merging and Copying Patient Databases . . . . . . . . . . . . . . . . . . . . 65Exporting Patient Records . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66Restoring a Database . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Rapid Programmer System Clinician’s Manual iii
Printing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68Types of Printable Reports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68Printing a Neurostimulation Record . . . . . . . . . . . . . . . . . . . . . . . 69Printing the Patient List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70Printing a Lead Diagnostics Report . . . . . . . . . . . . . . . . . . . . . . . 71
Cleaning and Maintaining the Rapid Programmer System . . . . . . . . . 72Handling the Programmer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72Cleaning the Programmer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72Storing and Transporting the Programmer . . . . . . . . . . . . . . . . . . . 73
Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74Troubleshooting Chart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74Performing a Soft Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80Reinstalling the Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81Copying Log Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81Returning the Rapid Programmer System for Service . . . . . . . . . . . 82
Service and Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . 83Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83Ordering Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Appendix A: Printable Reports of the Rapid Programmer System . . . . 85Neurostimulation Record. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85Patient List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86Lead Diagnostics Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Appendix B: Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Appendix C: Regulatory Statements. . . . . . . . . . . . . . . . . . . . . . . . . 88Statement of FCC Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . 88Declaration of Conformity (Industry Canada) Notice to
Users of Radio and Television. . . . . . . . . . . . . . . . . . . . . . . . . . 88Disposal Guidelines for Battery-Powered Devices . . . . . . . . . . . . . . 89
Appendix D: Symbols and Definitions . . . . . . . . . . . . . . . . . . . . . . . 90
Appendix E: Electromagnetic Compatibility Guidelines . . . . . . . . . . . 92
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
Rapid Programmer System Clinician’s Manual 1
Prescription and Safety Information
Read this section for important prescription and safety information.
NOTE: The information contained in this manual is provided solely for operating the Rapid
Programmer™ system and must not be considered all-inclusive with respect to a St. Jude
Medical™ neurostimulation system or other system components. Before using the Rapid
Programmer system to program a patient, you should thoroughly read this manual and
become familiar with using the components of the neurostimulation systems that the Rapid
Programmer system supports.
Intended Use
The Rapid Programmer™ system is intended to help clinicians determine
the best performance of a neurostimulation system by leading the patient
through a series of parameter combinations for the neurostimulation system
and recording the stimulation effects. The system is intended to be used for
programming while it is disconnected from an external power source.
Prescription Information
Refer to the clinician’s manual for the appropriate device to get specifi c
instructions and other important information, including indications for use,
contraindications, warnings, precautions, and adverse effects related to
those components.
Warnings
The following warnings apply to the programming system:
Explosive or fl ammable gases. Do not use programming devices in an
environment where explosive or fl ammable gas vapors are present. The
operation of programming devices could cause these vapors to ignite
resulting in severe burns, other injuries, or death.
Battery warning. This device contains a lithium ion rechargeable battery . To
reduce the risk of fi re or burns, do not disassemble, crush, puncture, short
external contacts on, or dispose of the battery in fi re or water. Do not insert
metal objects into the charging or communications ports of the device.
This can short-circuit the battery and result in danger. Replace the battery
with only an authorized spare. A risk of explosion is possible if the battery
is replaced by an incorrect battery type. Keep battery packs away from
children.
2 Rapid Programmer System Clinician’s Manual
Case damage. If the programmer case is pierced or ruptured, injury could
result from exposure to the battery chemicals or sharp edges. Inspect
programmer integrity before use. Do not use if the device shows signs of
damage. Return to St. Jude Medical for evaluation.
Shock hazards. To prevent electrical shock, (1) Use this device to program a
patient programmer or a trial system only when it is operating under battery
power and is not connected to a power outlet or to any external accessories;
(2) do not expose the system to rain or moisture; and (3) do not open the
outer case of the system or the system’s power supply. No part of this
product is serviceable by the user. Refer servicing to qualifi ed personnel
only.
Device components. The use of components not approved for use by
St. Jude Medical may result in damage to the system and increased risk to
the patient.
Precautions
The following precautions apply to the programming system:
Nonsterile device. The medical device system comprised of the programmer
and other St. Jude Medical devices as specifi ed in this document must
be used in accordance with IEC 60601-1-1 for patient programming .
This device is a nonsterile device and must be kept out of the sterile fi eld
(patient environment).
Clinician training. Clinicians should be familiar with neurostimulation
therapy and be experienced diagnosing and treating the indication for which
the neurostimulation system components are being used.
Device inspection. Before operating the system each time, inspect the
device and all its components for mechanical and electrical integrity. Avoid
using the system if the device or its components are damaged. Return
damaged components to St. Jude Medical for evaluation.
Position the power cord carefully. Position the power adapter so that it will
not be walked on, tripped over, or pinched by items placed upon or against
it. Pay particular attention to the plug, electrical outlet, and the point where
the power adapter exits the product.
Rapid Programmer System Clinician’s Manual 3
Unplug the power cord properly. When unplugging the power adapter from
an electrical outlet, do not pull on cords and cables. Grasp and pull from
the plug.
Avoid hot areas. The product should be placed away from heat sources such
as radiators, heat registers, stoves, and other products that produce heat.
If the product shows noticeable signs of damage due to overheating, do not
use it. Return damaged components to St. Jude Medical for evaluation.
Avoid blocking the vents. Slots and other openings are provided for
ventilation. To prevent overheating, do not obstruct vents. Using the device
on a soft surface, such as a pillow, blanket, rug, or thick clothing may block
airfl ow.
Battery inspection. Periodically before using the programmer, check the
programmer battery for bulging due to aging or other damage. Avoid using
the programmer if the battery is deformed.
Device modifi cation. To prevent damage to the system, do not modify the
device in any way.
Patient contact. Avoid making contact with the patient when using the
device.
Electromagnetic interference (EMI). Certain commercial electrical
equipment (e.g., some computer monitors, arc welders, induction furnaces,
and resistance welders), communication equipment (e.g., microwave
transmitters, linear power amplifi ers, and high-power amateur transmitters),
and high-voltage power lines may generate suffi cient EMI to interfere with
the operation of a neurostimulation system if they are approached too
closely.
Wireless use restrictions. In some environments, the use of wireless
functions (e.g., Bluetooth® wireless technology) may be restricted. Such
restrictions may apply aboard airplanes, in hospitals, near explosives, or in
hazardous locations. If you are unsure of the policy that applies to the use
of this device, please ask for authorization to use it before turning it on.
4 Rapid Programmer System Clinician’s Manual
Device Compatibility Guidelines
Only the following St. Jude Medical™ components may be used with the
Rapid Programmer™ system:
Patient Programmer, Models 3850, 3851, 3852, 3853, 3855, and
3856
MTS™ system
Bluetooth® printer, Model 3823
Rapid Programmer system accessories (see the “Ordering Information”
section)
No other components may be connected to the Rapid Programmer system.
Use only St. Jude Medical-approved accessories with the Rapid Programmer
system. Use only the power adapters and power supply that are supplied
with the system or from St. Jude Medical as a replacement.
Rapid Programmer System Clinician’s Manual 5
Product Description
The Rapid Programmer™ system interfaces with St. Jude Medical™
neurostimulation systems and is intended to be used to noninvasively
control device parameters. The Rapid Programmer system software is
installed only on a St. Jude Medical-approved PDA. Users interface with the
software and enter data using a stylus.
Contents of Package
The Rapid Programmer™ system (Model 3834) comes with the following
items.
Table 1: Package contents of the Rapid Programmer system
Item Description
Rapid Programmer
(Model 3834)
Interfaces with St. Jude Medical
neurostimulation systems and
allows you to noninvasively
control device parameters
Power adapter
(included with
Model 3834)
Allows you to use a wall outlet
as a power source to charge the
Rapid Programmer system
MTS communication
cable (Model 1273)
Connects an MTS system to the
CF card cable
IrDA communication
cable (Model 1280)
Connects a patient programmer
to the CF card cable
AAA batteries
(included with
Model 1280)
Provide power for the IrDA
communication cable
6 Rapid Programmer System Clinician’s Manual
Item Description
Compact fl ash to DB9
cable (CF card cable,
Model 1291)
Connects the Rapid
Programmer system to the MTS
communication cable or IrDA
communication cable
Stylus for RP Model
3834 (Model 1626)
Allows you to operate the touch
screen of the Rapid Programmer
system
Extended battery for
RP Model 3834
(Model 1629)
Provides power for the Rapid
Programmer system
Rapid Programmer
database card
(Model 3805)
Stores patient data and Rapid
Programmer system information
Rapid Programmer System Clinician’s Manual 7
Parts of the Rapid Programmer System
The following images provide information about the parts of the Rapid
Programmer™ system:
1
3
2
4
5 6 7 8 9
Figure 1: Front of the Rapid Programmer system
Legend:
1. Power button
2. Light sensor
3. Power and Bluetooth® LEDs
4. Touch screen
5. Inactive button
6. Inactive button
7. Inactive button
8. Inactive button
9. Inactive button
1
2
Figure 2: Back of the Rapid Programmer system
Legend:
1. Back speaker
2. Battery compartment cover
8 Rapid Programmer System Clinician’s Manual
1
3
2
4 5 6
Figure 3: Top and bottom of the Rapid
Programmer system
Legend:
1. CF card slot
2. Database card slot
3. Not used
4. AC 24-pin adapter port
5. Charger port
6. Not used
1
3
2
Figure 4: Sides of the Rapid Programmer system
Legend:
1. Inactive button
2. Stylus
3. Soft reset button
Rapid Programmer System Clinician’s Manual 9
Setting Up the Rapid Programmer System
The following sections provide instructions for setting up the Rapid
Programmer™ system.
Inserting the System Battery
When you receive your system, the programmer’s rechargeable lithium ion
battery is removed and needs to be inserted. To insert the programmer
battery , follow these steps:
1. Using your thumbnail, press in at the notch on the side of the battery
compartment cover.
2. Pull up to remove the cover from the back of the programmer.
1
2
Figure 5: Sequence for removing the battery compartment cover
10 Rapid Programmer System Clinician’s Manual
3. Hold the end of the battery release ribbon outside the battery
compartment, and insert the new battery by aligning the pins on the
bottom of the battery with the housing in the battery compartment.
Ensure that the ribbon is exposed after pressing the battery into place.
Figure 6: Inserting the battery
4. Insert the left side of the battery compartment cover into the notches
in the left side of the battery compartment, and press the cover closed.
Charging the System
Before you use the system for the fi rst time, you need to fully charge the
system. You can charge the system using either the power adapter or the car
charger (sold separately).
NOTE: A solid amber light on the front of the programmer indicates that the battery is
charging. When the battery is fully charged, the light turns solid green.
WARNING: Do not program a patient while the system is being
charged.
Rapid Programmer System Clinician’s Manual 11
Using the Power Adapter
To charge the system using the power adapter, follow these steps:
1. Insert the appropriate plug accessory into the slot on the power adapter.
2. Plug the power adapter into an electrical outlet.
3. Insert the other end of the power adapter into the 24-pin power
adapter port on the bottom of the programmer.
1
3
2
Figure 7: Sequence for connecting the power adapter for charging
Using the Car Charger
To charge the system using the car charger, follow these steps:
1. Assemble the car charger by connecting the cable to the socket plug.
2. Plug the car charger into the automobile’s cigarette lighter socket.
3. Insert the other end of the car charger into the corresponding charger
port on the bottom of the programmer.
Turning the System On and Off
To turn the system on or off, press the power button on the upper right
corner on the front of the programmer.
12 Rapid Programmer System Clinician’s Manual
Connecting a Patient Programmer
To connect a patient programmer to the Rapid Programmer™ system, follow
these steps:
NOTE: To keep the CF card connected to its cable, handle all cables with care while inserting
and connecting them.
CAUTION: To avoid damaging the IrDA communication cable, do
not remove the ribbon from the battery compartment end of the
IrDA communication cable.
1. If necessary, remove the protective plastic card from the CF card slot
on the top of the Rapid Programmer system.
2. Insert the CF card cable into the Rapid Programmer system (see the
following figure, left). Ensure the proper orientation of the card; the
“FRONT” label on the CF card should face the front the programmer.
3. If needed, insert three AAA batteries (included) into the battery
compartment end of the IrDA cable.
4. Connect the 9-pin connection end of IrDA cable to the CF card cable,
and then connect the battery compartment end of the IrDA cable to the
patient programmer (see the following figure, right).
Figure 8: Connecting the CF card cable Figure 9: Connecting the IrDA cable
Rapid Programmer System Clinician’s Manual 13
Connecting an MTS System
To connect an MTS™ system to the Rapid Programmer™ system, follow
these steps:
NOTE: To keep the CF card connected to its cable, handle all cables with care while inserting
and connecting them.
1. If necessary, remove the protective plastic card from the CF card slot
on the top of the Rapid Programmer system.
2. Insert the CF card cable into the Rapid Programmer system (see
the previous figure, left). Ensure the proper orientation of the card;
the “FRONT” label on the CF card should face the front of the
programmer.
3. Open the communication port door on the MTS system.
NOTE: You may need to unlock the communication port door before connecting the
communication cable to the MTS system. To unlock the communication port door, insert
the stylus or a similar object into the release port of the MTS system, and slide the door
until the communication port is exposed. After you are fi nished using the MTS system,
close and lock the communication port door.
4. Insert the MTS communication cable plug into the MTS communication
port.
3
2
1
Figure 10: Plugging the MTS communication
cable into the communication port
Legend:
1. Communication port
2. Communication port door
3. Release port
14 Rapid Programmer System Clinician’s Manual
5. Connect the 9-pin connection end of the MTS communication cable to
the CF card cable.
Figure 11: Connecting the MTS communication cable to the CF card cable
Rapid Programmer System Clinician’s Manual 15
Inserting a Database Card
To insert a database card into the programmer, follow these steps:
1. If necessary, remove the protective plastic card from the database
card slot.
2. With the database card label facing toward the front of the programmer
and the gold pins facing toward the back, insert the card into the
database card slot and press until the card is seated flush with the top
edge of the programmer.
CAUTION: To avoid damaging the system, insert the database card
with the label facing toward the front of the programmer.
Figure 12: Inserting the database card
16 Rapid Programmer System Clinician’s Manual
Removing a Database Card
To remove a database card from the programmer, follow these steps:
1. Ensure that the system is turned off or that you are on the Main Menu
screen.
NOTE: The system database can be corrupted if a user removes the database card while
the programmer is writing to the card. If the database gets corrupted, all data will be lost.
2. Push down slightly on the database card and release to unseat it.
3. When the card pops up, pull it from the database card slot.
Figure 13: Removing the database card
Rapid Programmer System Clinician’s Manual 17
Replacing the System Battery
If the system battery reaches the end of its life, you will need to replace
the battery with a St. Jude Medical-approved spare. To replace the system
battery , follow these steps:
1. Ensure that the system is turned off.
2. Using your thumbnail, press in at the notch on the side of the battery
compartment cover and then pull up to remove the cover from the back
of the programmer.
3. To release the battery , gently pull the ribbon straight out from the
programmer.
Figure 14: Releasing the battery
4. Remove the battery .
5. Hold the end of the battery release ribbon outside the battery
compartment, and insert the new battery by aligning the pins on the
bottom of the battery with the housing in the battery compartment.
Ensure that the ribbon is exposed after pressing the battery into place.
6. Insert the left side of the battery compartment cover into the notches
in the left side of the battery compartment, and press the cover closed.
18 Rapid Programmer System Clinician’s Manual
Getting Started
This section contains information about icons and buttons and
provides instructions for performing basic functions, such as adjusting
programmer settings.
Navigating the Main Menu Screen
When you turn on the Rapid Programmer™ system, the Main Menu screen
appears. From the Main Menu screen, you can access the different user
screens and menus. The following table provides defi nitions of the buttons
and menu options that are available on the Main Menu screen.
Table 2: Buttons and menu options on the Main Menu screen
Button or
Menu Option Definition
New Session enables you to start a new programming
session with a patient.
Last Session allows you to resume the last programming
session.
Intra-Op allows you to perform intraoperative testing.
During intraoperative testing, programs will not be
uploaded from or downloaded to the patient programmer or
MTS system.
Print Last Session allows you to print the last programming
session.
Patient List displays patient records.
View Demo enables you to see a demonstration of the
software without having a device connected to the Rapid
Programmer system.
Tools
Within the Tools menu, you can restore the database, copy
log fi les, reset the Rapid Programmer system, and reinstall
the Rapid Programmer system software.
Settings
Within the Settings menu, you can change preferences for
the following settings: backlight, brightness, clock, power,
regional, screen, and volume.
Rapid Programmer System Clinician’s Manual 19
Adjusting Backlight Settings
While you are working with the programmer, backlighting illuminates the
screen. To help preserve the battery capacity, you can set the backlight
to turn off after a period of inactivity. You can adjust this setting for the
programmer when it operates on battery power or on external power. To
adjust backlight settings, follow these steps:
1. On the Main Menu screen, tap Settings, then Backlight. The Backlight
screen appears.
Figure 15: The Backlight screen
2. Do one of the following:
To change settings for battery power, tap the Battery Power tab.
To change settings for external power, tap the External Power tab.
3. To turn off the backlight after a period of inactivity, ensure that the
first check box is selected, and then tap the list to select the amount
of time before the backlight turns off.
4. Adjust other settings as desired by tapping the appropriate check boxes
and lists.
20 Rapid Programmer System Clinician’s Manual
Adjusting Screen Brightness
You can adjust the level of brightness of the backlighting on the screen for the
programmer when it operates on battery power and external power. A lower
brightness level help preserve battery capacity when operating the programmer
on battery power. To adjust the screen brightness , follow these steps:
1. On the Main Menu screen, tap Settings, then Brightness. The
Brightness Level screen appears.
Figure 16: The Brightness Level screen
2. Tap and drag the desired slider to adjust the brightness level for
battery power or external power.
NOTE: A light sensor on the front of the programmer detects the environmental light level
and automatically adjusts the screen brightness in a small range based on the current
setting.
Rapid Programmer System Clinician’s Manual 21
Setting the Time and Date
The time and date appear on printable reports. To set the time and date for
your home location or a visited location, follow these steps:
1. On the Main Menu screen, tap Settings, then Clock. The Clocks &
Alarms screen appears.
Figure 17: The Clocks & Alarms screen
NOTE: When you fi rst start using the programmer or if you remove the programmer battery ,
you may be prompted to set the time and date, which means that the Clocks & Alarms
screen automatically opens.
2. Do one of the following:
To set the time and date for your home location, tap the Home
option button.
To set the time and date for a visited location, tap the Visiting
option button.
22 Rapid Programmer System Clinician’s Manual
3. To adjust the time zone, tap arrow beside the time zone box and select
the correct zone from the list that appears.
4. To adjust the time, tap the hour, minute, second, or morning or
afternoon designation, and then tap the increase and decrease buttons
to adjust the value.
5. To adjust the date, tap the arrow beside the date box to expand a
calendar, and do any of the following:
To adjust the month, tap the name of the month and select the
correct month from the list that appears.
To adjust the year, tap the year and then tap the increase and
decrease buttons.
To adjust the day, tap the correct day in the calendar.
6. Tap the Home or Visiting option button to select the time that you want
to use.
Checking the Programmer Battery Status
To check the status of the programmer battery , follow these steps:
1. On the Main Menu screen, tap Settings, then Power. The Power screen
appears.
2. Ensure that the Battery tab is selected to view the amount of power
remaining in the battery .
Figure 18: The Power screen showing battery power information
Rapid Programmer System Clinician’s Manual 23
Adjusting the Automatic Shutoff Settings
You can set the programmer to turn off after a period of inactivity, and you
can adjust this setting for the programmer when it operates on battery power
or on external power. To adjust these automatic shutoff settings, follow
these steps:
1. On the Main Menu screen, tap Settings, then Power. The Power screen
appears.
2. Tap the Advanced tab.
Figure 19: The Power screen showing automatic shutoff settings
3. To set the automatic shutoff time on battery power, ensure that the
battery power check box is selected, and then tap the list to select the
amount of time before the programmer turns off.
4. To set the automatic shutoff time for external power, tap the external
power check box, and then tap the list to select the amount of time
before the programmer turns off.
24 Rapid Programmer System Clinician’s Manual
Calibrating the Touch Screen
The touch screen display will occasionally require calibration to increase
the accuracy of screen taps. To calibrate the touch screen, do the following
action:
On the Main Menu screen, tap Settings, then Screen. The align screen
screen appears. Follow all on-screen instructions carefully, and ensure
that the stylus touches the center of the target graphic for proper
calibration.
NOTE: The Main Menu screen appears automatically after you are fi nished.
Figure 20: The align screen screen
Rapid Programmer System Clinician’s Manual 25
Adjusting the Volume
To adjust the volume of the programmer, follow these steps:
1. On the Main Menu screen, tap Settings, then Volume. The Sound
screen appears.
2. Tap and drag the slider to the desired volume.
Figure 21: The Sound screen
26 Rapid Programmer System Clinician’s Manual
Starting the Demo
The demo feature allows you to practice creating a program for systems
without having an actual patient or device. To start a demo session, follow
these steps:
1. From the Main Menu screen, tap . The Demo Session dialog box
appears.
Figure 22: The Demo Session dialog box
2. In the Select a demo list, tap the type of device that will be
represented in the demo.
3. Tap OK. The demo simulates a postoperative programming session.
Refer to sections in “Intraoperative and Postoperative Programming”
for instructions on how to select lead configurations and program in
demo mode.
NOTE: While you can perform the demo with any of the available lead confi gurations, only
specifi c lead confi gurations contain default programs, which have demonstration program
parameters and test log data. If you select a confi guration other than one of these specifi c
lead confi gurations, a Lead Change message appears. When you tap OK in this message,
all the default programs are deleted for the demo session.
Rapid Programmer System Clinician’s Manual 27
Intraoperative and Postoperative Programming
This section provides instructions and information about intraoperative and
postoperative testing and programming . Before programming with a patient,
you may want to view the demo to familiarize yourself with the programmer
functions (refer to “Starting the Demo ”).
NOTE: To keep the CF card connected to its cable, handle all cables with care while
programming .
Buttons and Icons on the Programming Screens
The following table provides defi nitions for buttons and icons that you may
encounter while testing and programming .
NOTE: Not all buttons or icons will be available on all screens. You can also access the
defi nitions of on-screen buttons and icons by tapping the help icon ( ) in the upper right-
hand corner of many of the screens.
Table 3: Buttons and icons on the programming screens
Button or Icon Definition
Access defi nitions of on-screen buttons and icons.
Turn the on-screen keyboard on or off.
Add a new program.
Copy the selected program.
Edit the selected program.
Delete the selected program or stimset on the Programs screen.
Print a report or list.
Balance the selected program.
Automatically return the amplitude to the programmed perception value
while steering stimulation. (When this feature is active, the icon on this
button is blue.)
28 Rapid Programmer System Clinician’s Manual
Button or Icon Definition
Use the MultiSteering programming feature to steer the electrical fi eld up.
Use the MultiSteering programming feature to steer the electrical fi eld down.
Use the MultiSteering programming feature to steer the electrical fi eld left.
Use the MultiSteering programming feature to steer the electrical fi eld right.
Shift the electrode polarities up.
Shift the electrode polarities down.
Increase the value of a parameter (Increase button).
Decrease the value of a parameter (Decrease button).
Change the polarity of the IPG can.
Set the target, comfort, or perception amplitude to the current amplitude
level.
Ramp the current amplitude to the target, comfort, or perception level.
Map where the patient feels stimulation.
Stop stimulation immediately for all active stim sets.
Add a new stimset or burst pulse to the program.
Delete a stimset or burst pulse from the program.
Turn the selected stimset on.
Turn the selected stimset off.
View the test log.
Rapid Programmer System Clinician’s Manual 29
Button or Icon Definition
Check impedance on the lead confi guration.
Edit the timing and magnet mode of the program.
Activate or inactivate BurstDR™ stimulation programming . (When this
feature is active, the icon on this button is blue.)
Starting an Intraoperative Testing Session
To start intraoperative testing, follow these steps:
1. Ensure the system is connected to the patient programmer or MTS™
unit to be programmed.
NOTE: If testing using a patient programmer, ensure that the patient programmer is
communicating with the generator before starting the testing session. For instructions on
communicating with the generator, refer to the clinician’s manual or user’s guide for the
appropriate system.
NOTE: Once the system has successfully established communication with the device to
be programmed, the screen of the patient programmer or MTS unit displays COMPUTER
COMMUNICATIONS.
2. On the Main Menu screen, tap . The Select Lead dialog box
appears.
30 Rapid Programmer System Clinician’s Manual
3. In the Select a lead configuration for intra-operative testing list, tap the
appropriate lead configuration.
Figure 23: The Select Lead dialog box
4. Tap OK. The Testing screen appears.
Figure 24: The intraoperative testing screen
Rapid Programmer System Clinician’s Manual 31
5. Perform intraoperative testing.
To enhance the speed of an intraoperative testing session, the testing
functions are limited to the following. Refer to the corresponding sections in
this manual for instructions on performing these functions.
Setting electrode polarities
Adjusting amplitude and amplitude step size
Adjusting frequency and pulse width
Measuring impedance
Using the MultiSteering™ programming feature
32 Rapid Programmer System Clinician’s Manual
Starting a New Programming Session or Resuming the Previous Session
To start a new programming session or resume the previous session, follow
these steps:
1. Ensure the system is connected to the patient programmer or MTS™
unit to be programmed.
NOTE: Before programming a patient, ensure you are not in demo mode.
NOTE: If programming using a patient programmer, ensure that the patient programmer
is communicating with the generator before starting the programming session. For
instructions on communicating with the generator refer to the clinician’s manual or user’s
guide for the appropriate system.
NOTE: Once the system has successfully established communication with the device to
be programmed, the screen of the patient programmer or MTS unit displays COMPUTER
COMMUNICATIONS.
2. On the Main Menu screen, tap one of the following as appropriate.
Begin programming a new patient or a different device,
or program the same device and discard the unsaved
programming information from the previous session.
Return to the last programming session.
NOTE: Do not use this option when using the system to program
a different patient or if a change has been made to the patient
programmer or MTS unit while disconnected from the system.
3. If a message appears stating, “All information from the prior session
will be deleted. Continue?”, tap OK.
4. When using the MTS system, if a message appears stating, “Is this a
new patient?”, tap Yes to delete all the programs on the MTS system or
tap No to keep the existing programs. The Uploading Programs screen
and then the Programs screen appears. The Programs screen displays
the Device Info tab.
NOTE: When the “Is this a new patient?” message appears, if you tap Yes by mistake,
you can prevent the programs on the MTS system from being deleted by tapping Cancel
Session on the Programs screen.
Rapid Programmer System Clinician’s Manual 33
Figure 25: The Programs screen
5. If needed, in the Device Info tab, tap (edit) next to the patient’s name.
The Patient Details screen appears, which displays the Patient Info and
Comments tabs.
NOTE: Tap in the bottom of the screen to display the on-screen keyboard. Tap the
123 button in the upper left corner of the keyboard to switch between numbers and
letters.
Figure 26: The Patient Details screen
34 Rapid Programmer System Clinician’s Manual
6. In the Patient Info and Comments tabs, use the on-screen keyboard to
enter information in the boxes as needed.
7. When finished, tap Save.
Selecting the Lead Configuration
To select the patient’s lead confi guration for postoperative programming ,
follow these steps:
On the Device Info tab on the Programs screen, select the patient’s
lead configuration from the Lead Configuration list.
NOTE: If single lead is inserted into the top header of a dual-header generator, you should
not select a single lead confi guration. Instead select the dual-lead option (indicated
by “x2”) of the lead you want to program. Program the lead numbered 9 to 16 (for an
Octrode™ lead) or 5 to 8 (for a Quattrode™ lead).
Resetting the Hours Used for the MTS System
When programming the MTS™ system for a new patient, you can reset the
usage counter to zero.
To reset the hours used, follow these steps:
1. On the Device Info tab on the Programs screen, tap (reset).
2. When the Confirm Reset message appears, tap OK.
Figure 27: The Confi rm Reset message
Rapid Programmer System Clinician’s Manual 35
Adding a New Program
To add a new program, follow these steps:
1. On the Programs tab on the Programs screen, tap . The Testing
screen appears.
2. Set the desired parameters.
1
92
5
3
6
4
7 8
10
11
13
15
12
14
16
Figure 28: Features of the Testing screen
Legend:
1. Shifts the electrode polarities or the
electrical fi eld
2. Automatically returns the amplitude to
the programmed perception level when
steering stimulation
3. Adds a new stim set to the program
4. Deletes the current stimset
5. Switches to BurstDR™ stimulation
programming mode
6. Shows the test log
7. Checks impedance on the electrode
confi guration
8. Edits the timing and magnet mode of
the program
9. Selects the stim set for testing.
10. Turns the stim set on or off
11. Sets the comfort amplitude to the
current level.
12. Ramps the current amplitude to the
comfort level
13. Sets the perception amplitude to the
current level
14. Ramps the current amplitude to the
perception level
15. Maps the location of perceived
stimulation
16. Stops stimulation on all stim sets
Copying a Selected Program
To copy a program, follow these steps:
1. On the Programs tab on the Programs screen, select the program from
the list that you want to copy.
2. Tap . The duplicated program appears as the next available program
number.
36 Rapid Programmer System Clinician’s Manual
Editing a Selected Program
To edit a program , follow these steps:
1. On the Programs tab on the Programs screen, select the program from
the list that you want to edit.
2. Tap . The Testing screen appears.
3. Edit the desired parameters.
Deleting a Selected Program, Stim Set, or Burst Pulse
To delete a program or stim set , follow these steps:
1. On the Programs tab of the Programs screen, select the program or
stim set from the list that you want to delete.
2. Tap . The Confirm Delete message appears.
3. Tap OK.
NOTE: If you select and delete the pulses of a BurstDR™ stimulation program on the
Programs screen, you will delete the entire program.
NOTE: You can also delete a stim set or a burst pulse while viewing it on the Testing
screen. While on the Testing screen, tap the number of the stim set or burst pulse you
want to delete, tap , and follow the on-screen prompts.
Figure 29: The Confi rm Delete message
Rapid Programmer System Clinician’s Manual 37
Switching Between Tonic Mode and BurstDR™ Stimulation Mode
Some systems allow you to create and modify two types of programs:
BurstDR™ stimulation program. A program where the stimulation
frequency consists of a group of pulses in rapid succession followed
by a period without pulses before repeating the group. This type
of program is available only on some systems that the Rapid
Programmer™ system supports.
Tonic program . A program where the stimulation frequency consists
of a single, repeating pulse. This program type is available on all
neurostimulation systems that the Rapid Programmer system supports.
If the system you are programming is capable of BurstDR stimulation,
you can switch to BurstDR stimulation programming mode on the Testing
screen.
NOTE: When creating a new program, you start programming in tonic programming mode by
default.
NOTE: When you switch between BurstDR stimulation mode and tonic mode, all programmed
values reset to default except the electrode confi guration.
To switch between tonic mode and BurstDR stimulation mode, follow these
steps:
1. On the Testing screen, tap . When BurstDR stimulation mode is
active, the icon turns blue, the screen background turns blue, and
the Testing screen displays the features for BurstDR stimulation
programming (see the following figure).
2. To return to tonic mode, tap . When BurstDR stimulation mode is
inactive, the icon turns gray, the screen background turns white, and
the Testing screen displays the features for tonic programming .
38 Rapid Programmer System Clinician’s Manual
9
11
7
10
8
12 13
5
1
3
2
6
4
Figure 30: Features of the Testing screen
in BurstDR stimulation mode
Legend:
1. Displays the overall frequency of the
BurstDR stimulation program
2. Displays the frequency of each burst
pulse in a pulse group
3. Sets the target amplitude to the current
level
4. Ramps the target amplitude to the target
level
5. Maps the location of perceived
stimulation
6. Stops stimulation on all stim sets
7. Adds a burst pulse to the pulse group
8. Deletes a burst pulse
9. Displays the number of burst pulses
10. Turns stimulation on or off
11. Inactivates BurstDR stimulation
programming mode
12. Checks impedance on the electrode
confi guration
13. Edits the timing and magnet mode of the
program
Configuring Electrode Polarities
Individual electrodes on a lead can be set to positive (anode), negative
(cathode), or neutral. The IPG (also called a “can”) also may be confi gured
as positive (anode) or neutral.
NOTE: The IPG can be used as an anode only for tonic programs.
NOTE: When the IPG is used as an anode, lead polarities can be only negative or neutral. All
lead polarities that are confi gured as anodes are automatically set to neutral.
NOTE: If you manually change polarities while stimulation is active, stimulation turns off.
If you manually change polarities after you have set values for perception and comfort
amplitude, a Reset Parameters message appears. If you tap OK, the perception and comfort
amplitude values reset to zero and the stimulation map is erased.
To confi gure electrode polarities, follow these steps:
1. On the Testing screen, tap a numbered electrode button on the lead
diagram. The first tap changes a neutral electrode to a negative polarity
(cathode); the second tap changes the electrode to a positive polarity
(anode); and the third tap returns the electrode to a neutral state.
Rapid Programmer System Clinician’s Manual 39
2. To use the IPG as an anode, tap in the lower left of the Testing
screen. This icon turns red and shows a plus sign when activated. To
return the IPG to a neutral state, tap .
Adjusting Pulse Width
During intraoperative testing and postoperative testing for new programs,
the PW (pulse width ) box on the Testing screen displays default values,
which are determined by the system type and programming mode (tonic or
BurstDR™ stimulation).
To change values for pulse width , do either of the following:
Tap the Increase/Decrease buttons next to the PW box.
Tap the PW box, and then tap a value in the drop-down list.
Adjusting Frequency for Tonic Programs
During intraoperative testing and postoperative testing for new tonic
programs, the Freq (frequency ) box on the Testing screen displays system
default values
To change values for frequency , do either of the following:
Tap the Increase/Decrease buttons next to the Freq box.
Tap the Freq box, and then tap a value in the drop-down list.
Adjusting Frequency for BurstDR™ Stimulation Programs
BurstDR™ stimulation programs have two frequency values that you can
set: the overall frequency (Burst Rate) and the frequency of each individual
pulse in a pulse group (Intra-burst). During postoperative testing for new
BurstDR stimulation programs, the Burst Rate and Intra-burst boxes on the
Testing screen display system default values.
To change the frequency values for a BurstDR stimulation program, do
either of the following:
Tap the Increase/Decrease buttons next to the respective box.
Tap the respective box, and then tap a value in the drop-down list.
40 Rapid Programmer System Clinician’s Manual
Adjusting Amplitude and Step Size
On the Testing screen, you can adjust amplitude and step size. Amplitude is
the strength of the electrical pulse that is delivered from the lead. Step size
is the increment at which amplitude will change when you adjust amplitude
during programming. Programs that use a higher amplitude setting deplete
the neurostimulation system battery faster.
NOTE: At any time during testing, you can immediately stop the patient’s stimulation by
tapping .
NOTE: When the generator reaches its energy limit , the Energy Limit dialog box appears. To
reduce the energy output of the generator, tap ok and follow the on-screen suggestions.
NOTE: When the generator reaches its output limit, the parameter boxes will turn yellow. To
reduce the output of the generator, try reducing the amplitude or pulse width or try setting
additional anodes or cathodes on the lead confi guration.
To adjust amplitude, follow these steps:
1. If stimulation is off, tap the Increase button next to the Amp box to
start stimulation.
2. To change the step size, tap the Increase/Decrease buttons next to the
Step box, or tap the Step box and then tap a value in the drop-down
list.
3. Tap the Increase/Decrease buttons next to the Amp box to increase or
decrease stimulation accordingly.
Setting Perception and Comfort Amplitudes for Tonic Programs
To complete a tonic program during postoperative programming , you
must set values for perception (Perc) and comfort (Comf) amplitudes for
each stim set. The perception amplitude is the value at which the patient
fi rst feels stimulation. The comfort amplitude is the value at which the
stimulation feels comfortable to the patient.
NOTE: When the generator reaches its output limit, the parameter boxes will turn yellow. To
reduce the output of the generator, try reducing the amplitude or pulse width or try setting
additional anodes or cathodes on the lead confi guration.
NOTE: During programming , if you change electrode polarities after you have set the values
for perception and comfort amplitude, the perception and comfort amplitude values will reset
to zero.
Rapid Programmer System Clinician’s Manual 41
NOTE: Once you have set the comfort and perception amplitudes, you can tap next to the
Comf and Perc boxes to ramp the amplitude to the respective comfort or perception value.
To set a patient’s perception and comfort amplitudes, follow these steps:
1. Tap the Increase/Decrease buttons next to the Amp box to reach the
patient’s perception amplitude.
2. To set the perception amplitude, tap next to the Perc box.
3. Tap the Increase/Decrease buttons next to the Amp box to reach the
patient’s comfort amplitude.
4. To set the comfort amplitude, tap next to the Comf box.
Setting the Target Amplitude for BurstDR™ Stimulation Programs
To complete a BurstDR™ stimulation program during postoperative
programming, you must set the value for the target amplitude . The target
amplitude is the value at which stimulation feels comfortable to the patient.
NOTE: When the generator reaches its output limit, the parameter boxes will turn yellow. To
reduce the output of the generator, try reducing the amplitude or pulse width or try setting
additional anodes or cathodes on the lead confi guration.
NOTE: During programming , if you change electrode polarities after you have set the value
for target amplitude , the target amplitude value will reset to zero.
NOTE: Once you have set the target amplitude, you can tap next to the Target box to
ramp the amplitude to the target value.
To set a patient’s target amplitude , follow these steps:
1. Tap the Increase/Decrease buttons next to the Amp box to reach the
patient’s target amplitude.
2. Tap next to the Target box.
Adding a Stim Set or a Burst Pulse to a Program
To add another stim set to a tonic program or another burst pulse to a
BurstDR™ stimulaton program , follow these steps:
1. On the Testing screen, tap .
2. Set the electrode polarities and other parameters as needed.
42 Rapid Programmer System Clinician’s Manual
Using the MultiSteering Programming Feature
The MultiSteering ™ programming feature provides a simplifi ed method for
shaping and fi ne-tuning the electrical fi eld in real time. During testing, certain
electrode combinations can be shifted while stimulation remains active.
NOTE: This programming feature is available only for tonic programs.
NOTE: When programming dual Octrode™ leads, you can steer stimulation laterally between
the two leads. Confi rm that the electrodes on the patient’s leads are aligned as they appear
on the Testing screen.
NOTE: Because some nerve fi bers may be more sensitive to electrical fi elds than others, you
should adjust the amplitude to the patient’s perception level (Perc) each time before steering
to the next combination.
NOTE: If you have set the Perc value and the return-to-perception button ( ) is active,
the system automatically reduces the amplitude to the perception level each time you steer
stimulation. If the button is active, the icon is blue.
NOTE: If , , , or appear dimmed, it means that the selected electrode combination
cannot be steered in that direction. Try adding or removing polarities until the MultiSteering
buttons become active.
NOTE: When the generator reaches its output limit, the parameter boxes will turn yellow
and MultiSteering programming will be temporarily unavailable. To use MultiSteering
programming, reduce the output by adjusting pulse width or amplitude.
Rapid Programmer System Clinician’s Manual 43
To use the MultiSteering programming feature
Tap , , , or to steer the stimulation through a set of optimized
electrode combinations designed to incrementally shift the stimulation.
2 3
4
5
1
Figure 31: MultiSteering programming
buttons available on the Testing screen
Legend:
1. Shift up
2. Shift left (dimmed)
3. Shift right
4. Shift down
5. Return the amplitude to the
programmed perception value
(when active)
Creating a Stimulation Map
During testing, you can create a stimulation map that shows where the
patient feels stimulation with a specifi c stimulation set for tonic programs
or with a pulse group for BurstDR™ stimulation programs. The stimulation
map appears as an icon on the user’s patient programmer or MTS™ unit.
44 Rapid Programmer System Clinician’s Manual
To create a stimulation map , follow these steps:
1. On the Testing screen, tap . A screen appears with a graphic
representation of the human body and the Draw option selected.
Figure 32: The stimulation map screen
2. Tap the areas on the graphic that correspond to the areas where the
patient feels stimulation. The tapped areas are highlighted in green.
3. To remove a highlighted area, tap Erase and then the area on the
graphic.
4. To save the stimulation map , tap Done.
Rapid Programmer System Clinician’s Manual 45
Using the Dynamic MultiStim Programming Feature
The Dynamic MultiStim™ programming feature allows you to actively turn
a stim set on or off to assess how well the stim sets work separately and
together as a program.
NOTE: This programming feature is available only for tonic stimulation programs.
NOTE: During Dynamic MultiStim program ming, all stim sets must have the same frequency .
If you change frequency in one stim set, frequency automatically changes in the program’s
other stim sets.
NOTE: When you add a stim set , the amplitude of all the stim sets returns to zero.
The following list describes the stim set buttons used for Dynamic
MultiStim programming :
1When the number of a stim set is blue, you are viewing that
stim set. Tap the number to select the stim set and adjust
that stim set’s settings.
When the amplitude is more than zero (active) for that stim
set, the Stim Set button is “on” (green). If stimulation is
off, tap the Stim Set button to ramp the amplitude to the
last value that was reached for the stim set during the same
testing session.
When amplitude is zero for that stim set, the Stim Set button
is “off.” If stimulation is more than zero (active) for a stim
set, tap the Stim Set button to turn off stimulation.
46 Rapid Programmer System Clinician’s Manual
2 3
1
Figure 33: Buttons for Dynamic MultiStim
programming
Legend:
1. Stim set selected
2. Stim set button on
3. Stim set button off
Rapid Programmer System Clinician’s Manual 47
Viewing the Test Log
The test log allows you to view stim sets that have been tested on the
patient and to add or replace them within the current program.
NOTE: The test log is available only for tonic stimulation programs.
NOTE: Test log data is only kept within a single programming session. If you start a new
session, the test log contains only the stim sets that were uploaded from the device.
NOTE: On the Test Log screen (see the following fi gure), if you select the Show valid stimsets
only check box, the test log will show only the stim sets that are compatible with the present
program. For program compatibility, see the clinician’s manual for the appropriate device.
To view the test log , do either of the following:
From the Testing screen, tap .
From the Programs tab on the Programs screen, select the desired stim
set from the list, and tap .
Figure 34: The Test Log screen
48 Rapid Programmer System Clinician’s Manual
To Add a Stim Set From the Test Log
You can select a stim set from the Test Log screen to add to the current
program.
NOTE: To be able to add a stim set to an existing program, the parameters of the program
must match the device’s capabilities and the program must not contain its maximum number
of stim sets, which is determined by the device type. If your parameters are outside of
acceptable ranges for the device, an error message appears.
To add a stim set from the test log, follow these steps:
1. On the Test Log screen, tap the Add tab.
2. Select a stim set from the list.
3. Tap Add.
To Replace a Stim Set From the Test Log
You can also replace the current stim set with a selected stim set from the
Test Log screen.
To replace a stim set from the test log, follow these steps:
1. On the Test Log screen, tap the Replace tab.
2. Select a stim set from list.
3. Tap Replace.
Measuring Impedance
During testing, you may want to periodically measure the impedance on the
electrode array as a diagnostic tool to ensure that current is fl owing through
the system properly. Impedance measurements may help you troubleshoot
issues with the neurostimulation system.
NOTE: If the parameter boxes on the Testing screen turn yellow, your impedance
measurement will probably be high. See the “Troubleshooting” section for more information.
To measure impedance for the programmed electrode array, follow these
steps:
1. On the lower left corner of the Testing screen, tap . The Impedance
Results dialog box appears (see the following figure, left).
Rapid Programmer System Clinician’s Manual 49
2. If the option is available, tap Diagnostics in the Impedance Results
dialog box to get diagnostic details for the entire lead configuration.
The Diagnostic Summary details the condition and the impedance
values for each electrode by number (see the following figure, right). If
desired, tap Print to print a lead diagnostics report (see “Printing ”).
Figure 35: The Impedance Results dialog box Figure 36: The Diagnostic Summary screen
3. Refer to the applicable clinician’s manual for information on
interpreting the results of the impedance measurement.
4. When you are finished, do one of the following to return to the Testing
screen:
In the Impedance Results dialog box, tap Done.
On the Diagnostic Summary screen, tap OK.
50 Rapid Programmer System Clinician’s Manual
Modifying IPG Modes
From the Edit Modes screen, you can modify IPG modes by performing the
following functions:
Set the magnet mode
Set the program mode
Set the on and off times for cycle and bolus mode s
Set the ramp time for continuous and bolus modes
To modify IPG modes, follow these steps:
1. On the Testing screen, tap . The Edit Modes screen appears.
2. Modify the settings as desired. The following subsections provide more
information about using these settings.
3. When you are finished, tap Save.
Figure 37: The Edit Modes screen
To Set the Magnet Mode
The magnet mode settings control how the IPG responds when it senses a
magnet. To set the magnet mode , tap the desired option button:
Off, which means the IPG will turn off when it senses a magnet
On/Off, which means the IPG will toggle between on and off each time
it senses a magnet
Rapid Programmer System Clinician’s Manual 51
To Set the Ramp Time
Ramp time is the approximate number of seconds that the IPG takes when
starting stimulation to arrive at the perception amplitude for tonic programs
or the target amplitude for BurstDR™ stimulation programs. Ramp time is
saved to the IPG; it is not the ramp time that you use while programming .
You can adjust the ramp time for the IPG when it is set to continuous
or bolus mode (see the “To Set the Program Mode” section for more
information). The default ramp time is 4 seconds.
To change the ramp time, tap the Increase/Decrease buttons next to
the Ramp Time box.
To Set the Program Mode
You can set the stimulation mode of a program. The following table provides
defi nitions of the three program modes that may be available.
Table 4: Defi nitions of IPG Program Modes
Mode Definition
Continuous Stimulation is delivered continuously while the IPG is on and the
program is running.
Cycle Stimulation is on for a programmed amount of time (on time) and off for
a programmed amount of time (off time). The cycle alternates between
on and off time until the IPG is turned off or the program is stopped.
Bolus Stimulation is on for a programmed amount of time (on time) and then
enters a phase of off time. During the off time phase, stimulation cannot
be restarted. After the off time has passed, stimulation must be restarted
manually.
To set the program mode , follow these steps:
1. Under Program Mode, tap the desired check box. If you tap Bolus or
Cycle , the On Time and Off Time timer boxes become available.
2. Tap the Increase/Decrease buttons beside the timer boxes to set the on
and off phase times.
NOTE: You can set the on and off cycle times to a maximum of 24 hours and a minimum of
5 seconds.
NOTE: You can set the on and off bolus times to a maximum of 18 hours and a minimum of
1 second.
52 Rapid Programmer System Clinician’s Manual
Saving a Program
During postoperative programming , you must save a program before
downloading it.
To save the program, tap Save on the Testing screen.
NOTE: During intraoperative testing, you do not need to save the program. To exit an
intraoperative testing session and return to the Main Menu screen, simply tap Done on the
Testing screen.
NOTE: If any parameters of a stim set are incomplete for a program, a Stimset Error
message appears when you try to save the program. To correct this issue, ensure that you
have confi gured valid electrode polarity combinations and have set values for comfort and
perception amplitude on all stim sets.
Using the Active Balancing Programming Feature
The Active Balancing ™ programming feature allows you to individually
adjust the amplitudes of each stim set within a program, which helps
balance the feeling of stimulation that is delivered to a patient.
NOTE: This programming feature is available only for tonic stimulation programs.
To balance a program, follow these steps:
1. On the Programs tab on the Programs screen, select a program from
the list.
2. Tap . The Balancing screen appears.
Rapid Programmer System Clinician’s Manual 53
Figure 38: The Balancing screen
3. Tap the Increase button beside the master amplitude box. The
amplitude ramps up to the perception value (Perc).
NOTE: To stop stimulation, tap Stop. This button turns off the master amplitude but does
not change the slider values.
4. Tap the Increase/Decrease buttons to adjust the master amplitude until
the patient indicates that stimulation is comfortable.
5. Tap above or below the slider to increase or decrease the amplitude
level of a stim set.
NOTE: To reset all sliders to zero, tap Reset.
6. If the program contains more than four stim sets, tap More to display
the remaining stim sets, and adjust the stim sets if necessary.
7. Tap Save.
54 Rapid Programmer System Clinician’s Manual
Downloading Programs
After completing a programming session, you can download programs to the
patient programmer or MTS™ unit that is communicating with the system.
All programs are downloaded at once.
To download programs, follow these steps:
1. On the Programs tab of the Programs screen, select the program from
the list that will be the active program on the patient programmer or
MTS unit.
Figure 39: A program selected on the Programs screen
Rapid Programmer System Clinician’s Manual 55
2. Tap Download, and then tap OK in the Replace Programs message that
appears (see the following figure). The Downloading Programs screen
appears; and after the programs have been downloaded to the patient
programmer or MTS unit, the Main Menu screen appears.
NOTE: If downloading programs to an IPG, ensure that the programs have fi nished
downloading before moving the programming wand from over the IPG site.
Figure 40: The Replace Programs message
3. To ensure that the device has completed the download process, turn the
patient programmer or MTS unit off and then on again.
Overriding the Usage Range for Amplitude
Before downloading programs, you can override the usage range for
amplitude. The Override Usage Range screen allows you to override the
tested amplitude range by increasing the maximum tolerable amplitude or
decreasing the perception amplitude setting by up to 25 percent without
testing stimulation on the patient.
NOTE: The feature to override the usage range is available only for tonic stimulation
programs.
NOTE: If you override the amplitude usage range of a MultiStim ™ program that has been
balanced using the Active Balancing™ programming feature, you may need to rebalance the
program (see “Using the Active Balancing Programming Feature”).
56 Rapid Programmer System Clinician’s Manual
To override the usage range
1. On the Programs tab of the Programs screen, tap and hold Download
for 5 seconds until the Override Usage Range screen appears.
Figure 41: The Override Usage Range screen
2. Tap the Increase/Decrease buttons to change the maximum tolerable
amplitude and perception amplitude percentages as desired.
3. Select one or more program check boxes to which you want to apply
the changes.
-OR-
Select the Apply to all programs check box to select all check boxes.
4. Tap Download, and then tap OK in the Replace Programs message that
appears. The programs are downloaded to the patient’s device.
Rapid Programmer System Clinician’s Manual 57
Managing Patient Records
The Rapid Programmer™ system allows you to manage patient records by
adding, editing, or deleting patient data. To manage patient data, ensure
that a database card that is compatible with the Rapid Programmer system
is inserted.
Buttons and Icons on the Patient Database Screens
The following table provides defi nitions for buttons and icons that you may
encounter while managing patient records.
NOTE: Not all buttons or icons will be available on all screens. You can also access the
defi nitions of on-screen buttons and icons by tapping the help icon ( ) in the upper right-
hand corner of many of the screens.
Table 5: Buttons and icons on the patient database screens
Button or Icon Definition
Access defi nitions of on-screen buttons and icons.
Turn the on-screen keyboard on or off.
Add a new patient.
Edit a selected (highlighted) patient.
Combine two checked patients into one record.
Delete checked patients.
Print the patient list (on the Patients screen).
Print the programs for the patient (on the Patient Details screen).
Export checked patient records to the database card .
58 Rapid Programmer System Clinician’s Manual
Accessing the Patient Database
To access the patient database, you must enter a password . Contact
Technical Support for the default password related to your software version.
Once you have the password , follow these steps:
1. From the Main Menu screen, tap . The Enter Password screen
appears.
Figure 42: The Enter Password screen displaying the on-screen keyboard
2. If necessary, tap at the bottom of the screen to display the
on-screen keyboard. Tap the 123 button in the upper left corner of the
keyboard to switch between symbols and letters.
3. Enter the password in the Enter Password box. Passwords can be up
to 11 characters. See “Changing the Password for Viewing Patients” to
change the password .
NOTE: The password is case sensitive.
Rapid Programmer System Clinician’s Manual 59
4. Tap OK. The Patients screen appears. You can sort list items by
tapping the column heading.
5. If desired, tap to print the patient list (see “Printing ”).
Figure 43: The Patients screen
60 Rapid Programmer System Clinician’s Manual
Changing the Password for the Patient Database
To change the password for the patient database, follow these steps:
1. On the Patients screen, tap Database, and then tap Change Password.
The Change Password screen appears.
2. Use the on-screen keyboard to fill in the boxes, and tap OK when
finished.
Figure 44: The Change Password screen
Rapid Programmer System Clinician’s Manual 61
Adding a Patient to the Database
To add a patient to the database, follow these steps:
1. From the Patients screen, tap . The Patient Details screen appears.
Figure 45: The Patient Details screen
2. On the Patient Info and Comments tabs, tap a box you want to
complete and enter the information using the on-screen keyboard.
3. If desired, tap to print the selected patient’s neurostimulation record
(see “Printing” ).
4. Tap Save to save the patient’s information and return to the Patients
screen.
62 Rapid Programmer System Clinician’s Manual
Editing a Patient’s Information
To edit a patient’s information in the database , follow these steps:
1. From the list on the Patients screen, tap the name of the patient you
want to edit. The selected name is highlighted.
Figure 46: Highlighting a patient on the Patients screen
2. Tap . The Patient Details screen appears.
Figure 47: The Patient Details screen
Rapid Programmer System Clinician’s Manual 63
3. Using the on-screen keyboard, edit the information on the Patient Info
and Comments tabs.
4. Tap Save to save the patient’s information and return to the Patients
screen.
Deleting a Patient From the Database
To delete a patient from the database, follow these steps:
1. On the Patients screen, select the check boxes that correspond with
the patients whom you want to delete (see the following figure, left). To
select all patients, tap in the header. Tap again to deselect all
patients.
2. Tap , and then tap OK when the Confirm Delete message appears
(see the following figure, right).
Figure 48: Selecting patient check boxes
on the Patients screen
Figure 49: The Confi rm Delete message
64 Rapid Programmer System Clinician’s Manual
Combining Patient Data
The combining patient data feature allows you to combine information from
two patient records into one record. The patient name from only one patient
is kept, and only the most recent session of the two patients is retained. In
order for the two patient record s to be combined, either must be true:
The generator in each record must share the same model number and
serial number
The generator in one of the patient records must not have a model
number and serial number assigned
To combine patient data, follow these steps:
1. On the Patients screen, select the check boxes from the list for the two
records you want to combine (see the following figure, left).
2. Tap . The Combine Patients screen appears (see the following figure,
right).
Figure 50: Selecting patient check boxes
on the Patients screen
Figure 51: The Combine Patients screen
3. Select which patient’s record will store the combined information by
tapping the button next to that patient’s name.
NOTE: The patient information from the most recent session will be saved.
4. Tap OK. The records are combined and the unselected patient record is
deleted.
Rapid Programmer System Clinician’s Manual 65
Merging and Copying Patient Database s
The merge feature allows you to combine two patient databases from two
separate database cards. The copy feature allows you to copy the patient
database from one database card to another.
To merge or copy data, follow these steps:
1. In the lower left corner of the Patients screen, tap Database, and then
tap either Merge or Copy.
2. When the Confirm Merge or Confirm Copy message appears, tap OK,
and follow the on-screen instructions.
Figure 52: The Database menu on the Patients screen
66 Rapid Programmer System Clinician’s Manual
Exporting Patient Records
On the Patients screen, you can export patient records to a database card as
a fi le (PatientExport.txt) that can be opened on a computer using a database
card reader.
CAUTION: When viewing files on the database card with a database
card reader, view only the PatientExport.txt file and do not edit or
delete any other files on the card. Otherwise, you may inadvertently
delete the database from the card.
To export patient records, follow these steps:
1. Ensure that the database card containing the records is inserted in the
programmer.
2. From the list on the Patients screen, select the check boxes of the
patient records to export.
3. Tap . The Confirm Export message appears (see the following figure,
left).
4. Tap OK.
5. When the Export Complete message appears, tap ok (see the following
figure, right).
Figure 53: The Confi rm Export message Figure 54: The Export Complete message
Rapid Programmer System Clinician’s Manual 67
Restoring a Database
The programmer automatically backs up the database before every major
database operation. The programmer stores up to seven daily backup fi les
that are captured at the start of the fi rst session of each day, as well as
up to three monthly backup fi les that are captured at the start of the fi rst
session of the month. If something unwanted happens to the database, you
can try to restore it from one of these backup fi les.
To restore a database , follow these steps:
1. Ensure that the database card containing the patient database is
inserted.
2. On the Main Menu screen, tap the Tools menu, and then tap Restore
Database. The Restore Database screen appears.
Figure 55: The Restore Database screen
3. From the list of available files, tap the date of the file you want to
restore.
4. Tap OK. When the Restore Complete message appears, tap ok to return
to the Main Menu screen.
68 Rapid Programmer System Clinician’s Manual
Printing
You may print programming and database data from the Rapid
Programmer™ system using the St. Jude Medical Bluetooth® printer (Model
3823).
NOTE: The Rapid Programmer system can print to the Bluetooth printer from a distance up
to 10 m (33 ft).
NOTE: It may take up to a minute to start printing after you tap Print.
Types of Printable Reports
The following table lists the types of reports you can print and the locations
from which you can print them. To see full examples and to learn more
about the information contained on these reports, see Appendix A.
Table 6: Types of printable reports
Report Type Description Print From...
Neurostimulation
record
Contains a record of all programs,
as well as generator and lead
information, for a selected patient
The Main Menu screen (prints
the record from the last testing
session only).
The Programs tab on the
Testing screen.
The Patient Details screen
for a selected patient in the
database.
Patient list Lists patient names from the
database and the dates of their
last revisions
The Patient History Details screen
in the patient records database.
Lead diagnostics
report
Shows the results of the diagnostic
testing performed on a generator
and leads for a specifi c program
NOTE: This report is available only
for the Eon, Prodigy, Protégé, and
MTS systems.
The Diagnostic Summary screen
after measuring impedance .
Rapid Programmer System Clinician’s Manual 69
Printing a Neurostimulation Record
To print a neurostimulation record , follow these steps:
1. Do one of the following:
If printing from the Main Menu screen, tap .
If printing from the Testing screen or Patient Details screen, tap .
2. Ensure that the printer is turned on.
3. On the Print Preview screen, tap Print.
4. If a Multiple Printers Found message appears, tap the box in the
message to open a list of nearby printers, select a printer from the list,
and then tap OK.
5. After you have finished printing, tap Done and turn off the printer.
Figure 56: The Print Preview screen for a neurostimulation record
70 Rapid Programmer System Clinician’s Manual
Printing the Patient List
To print the patient list , follow these steps:
1. On the Patients screen, tap .
2. Ensure the printer is turned on.
3. On the Print Preview screen, tap Print.
4. If a Multiple Printers Found message appears, tap the box in the
message to open a list of nearby printers, select a printer from the list,
and then tap OK.
5. After you have finished printing , tap Done and turn off the printer.
Figure 57: The Print Preview screen for a patient list
Rapid Programmer System Clinician’s Manual 71
Printing a Lead Diagnostics Report
To print a lead diagnostics report , follow these steps:
1. Ensure the printer is turned on.
2. On the Diagnostic Summary screen, tap Print.
3. If a Multiple Printers Found message appears, tap the box in the
message to open a list of nearby printers, select a printer from the list,
and then tap OK.
4. When you have finished printing, tap OK and turn off the printer.
Figure 58: The Diagnostic Summary screen
72 Rapid Programmer System Clinician’s Manual
Cleaning and Maintaining the Rapid Programmer System
This section provides tips and other information to help you clean and
maintain the Rapid Programmer™ system and keep it running its best.
Handling the Programmer
To properly handle the programmer, follow these tips:
Handle the programmer with care. It is a sensitive electronic device
that can be damaged by rough handling (e.g., dropping it on the
ground).
Do not expose the programmer to prolonged direct sunlight.
Use the programmer within a safe range of environmental conditions
(see Appendix B).
Avoid exposing the programmer to sand, dust, and other environmental
hazards.
Keep the programmer away from food and beverages, and avoid doing
things that might get the programmer wet. The programmer is not
waterproof.
If the programmer does get wet, do not use it. Contact Technical
Support for help.
Cleaning the Programmer
To properly clean programmer, follow these tips:
When cleaning or disinfecting the programmer, turn off the programmer
and unplug the power adapter.
Use a soft, dry cloth to remove dust from the screen. Never use glass
cleaners.
Clean the programmer by wiping off the outer surface using a damp,
lint-free cloth and a small amount of mild soap.
Do not submerge the programmer or use a cloth that is soaking wet.
Do not use alcohol, cleaning solutions, or solvents to clean the
programmer. They may damage the surface finishes of the programmer.
Rapid Programmer System Clinician’s Manual 73
Storing and Transporting the Programmer
To properly store and transport the programmer, follow these tips:
If storing the programmer for a month or longer, turn the programmer
off, fully charge the programmer battery , and remove and store the
programmer battery separately to maximize its life.
Store the programmer and the programmer battery in a cool, dry place
(see Appendix B).
Transport the programmer in a carrying case.
If you are carrying the programmer in a briefcase, purse, or any other
carrying case, make sure no objects in the case are pressing on the
device.
Do not put heavy or sharp objects on the programmer.
74 Rapid Programmer System Clinician’s Manual
Troubleshooting
This section provides troubleshooting suggestions for issues you might
have with the Rapid Programmer™ system. For additional information,
call Technical Support. When you call, be ready to give the representative
information about what you were doing when the error occurred and what
error messages appeared.
Troubleshooting Chart
Typically, the Rapid Programmer™ system displays a message for issues
associated with programming patients and managing records. If you are
experiencing issues, check the programmer screen for a message and follow
any instructions it gives to correct the issue. If you still experience the issue
or if you experience an issue without receiving an on-screen message, refer
to the following table for possible causes and solutions.
NOTE: In the following table, the term “programmer” refers to the Rapid Programmer system.
The term “patient programmer” refers to the patient programmer for the neurostimulation
system.
Table 7: Possible causes and solutions for potential issues
Problem Possible Cause Possible Solution
Programmer has
no power or has
lost power.
Programmer’s battery is drained. Recharge the battery using the
power adapter or the car charger.
Programmer’s battery is depleted
or defective.
Replace the battery .
Programmer is damaged or
malfunctioning.
Replace the programmer.
Programmer will
not charge.
Power adapter or car charger
is disconnected from the
programmer.
Connect the power adapter or car
charger to the programmer.
Correct plug adapter (voltage
converter) is not connected to the
power adapter’s plug.
Connect the appropriate plug
adapter (voltage converter) to the
power adapter’s plug.
Power adapter or car charger is
defective.
Replace the power adapter or car
charger.
Programmer’s battery is depleted
or defective.
Replace the battery .
Programmer is damaged or
malfunctioning.
Replace the programmer.
Rapid Programmer System Clinician’s Manual 75
Problem Possible Cause Possible Solution
Nothing is
displayed on the
screen.
Programmer is off or has timed
out.
Turn on the programmer.
Programmer’s battery is drained. Recharge the battery using the
power adapter or the car charger.
Programmer’s battery is depleted
or defective.
Replace the battery .
Screen is damaged or
malfunctioning.
If the programmer appears to be
powered on but without display,
the screen may be defective.
Contact Technical Support.
Programmer will
not respond to
input.
Programmer has locked up. Perform a soft reset .
Touch-screen interface is damaged
or malfunctioning.
Replace the programmer.
Report will not
print from the
printer (Model
3823).
Printer is off. Turn on the printer.
Printer is out of paper. Add a paper roll to the printer.
Interference is disrupting
Bluetooth communication.
Turn off or remove the source of
the interference.
-OR-
Move away from the interference.
Printer is damaged or
malfunctioning.
Replace the printer.
76 Rapid Programmer System Clinician’s Manual
Problem Possible Cause Possible Solution
Programmer is not
communicating
with the IPG.
Programmer is in demo mode. Exit the programming session and
return to the Main Menu screen.
Restart the programming session
by tapping the New Session button
or the Intra-Op button as desired.
Patient programmer is off or has
timed out and is in standby mode.
Turn on the patient programmer.
Ensure that the screen displays
COMPUTER COMMUNICATIONS.
Batteries in the patient
programmer are depleted.
Replace the batteries in the
patient programmer.
Programmer battery is drained. Charge the programmer battery .
IrDA cable is disconnected. Connect the IrDA cable to the
CF card cable and the patient
programmer.
CF card cable is disconnected
or improperly seated in the
programmer
Connect the CF card cable to the
IrDA cable, and properly seat the
CF card cable in the programmer.
NOTE: When removing the CF
card cable, wait about 3 seconds
before reinserting it into the
programmer.
Communication wand is not
connected to the patient
programmer.
Connect the wand to the patient
programmer.
Communication wand is not
directly over the IPG site.
Center the wand directly over the
IPG site.
IrDA cable is damaged or
malfunctioning.
Replace the IrDA cable.
CF card cable is damaged or
malfunctioning.
Replace the CF card cable.
Communication wand is damaged
or malfunctioning.
Replace the wand.
Patient programmer is damaged or
malfunctioning.
Replace the patient programmer.
Programmer is damaged or
malfunctioning.
Replace the programmer.
Rapid Programmer System Clinician’s Manual 77
Problem Possible Cause Possible Solution
Programmer is not
communicating
with the MTS
system.
Programmer is in demo mode. Exit the programming session and
return to the Main Menu screen.
Restart the programming session
by tapping the New Session button
or the Intra-Op button as desired.
MTS system is off or has timed out
and is in standby mode
Turn on the MTS system.
Ensure that the screen displays
COMPUTER COMMUNICATIONS.
Batteries in the MTS system are
depleted.
Replace the batteries in the MTS
system.
Programmer battery is drained. Charge the programmer battery .
MTS communication cable is
disconnected.
Connect the MTS communication
cable to the programmer and the
MTS system.
CF card cable is disconnected
or improperly seated in the
programmer
Connect the CF card cable to the
MTS communication cable, and
properly seat the CF card cable in
the programmer.
NOTE: When removing the CF
card cable, wait about 3 seconds
before reinserting it into the
programmer.
MTS communication cable is
damaged or malfunctioning.
Replace the MTS communication
cable.
CF card cable is damaged or
malfunctioning.
Replace the CF card cable.
MTS system is damaged or
malfunctioning.
Replace the MTS system.
Programmer is damaged or
malfunctioning.
Replace the programmer.
78 Rapid Programmer System Clinician’s Manual
Problem Possible Cause Possible Solution
Patient is
not receiving
stimulation.
Programmer is in demo mode Exit the programming session and
return to the Main Menu screen.
Restart the programming session
by tapping the New Session button
or the Intra-Op button as desired.
Stimulation is off. Turn on stimulation.
IPG battery is low and stimulation
shut off (rechargeable IPGs).
Recharge the IPG.
Lead is damaged or has become
disconnected.
Check impedance on the
lead using the programmer. If
impedance is high, then the lead
may be damaged or disconnected.
Check the connections or consider
replacing the lead.
Refer to “Measuring Impedance”
for more information.
IPG battery is depleted
(nonrechargeable) or unable to
hold a charge (rechargeable).
Replace the IPG. Contact
Technical Support.
IPG is damaged or malfunctioning. Replace the IPG. Contact
Technical Support.
Parameter boxes
on the Testing
screen turn
yellow.
Generator has reached its output
limit
Reduce the amplitude or pulse
width.
-OR-
Set additional polarities (anode or
cathode) on the lead.
Lead is damaged or has become
disconnected.
Check impedance on the
lead using the programmer. If
impedance remains high despite
changes to program settings,
then the lead may be damaged
or disconnected. Check the
connections or consider replacing
the lead.
Refer to “Measuring Impedance”
for more information.
Rapid Programmer System Clinician’s Manual 79
Problem Possible Cause Possible Solution
Patient record
will not export to
database card .
Database card is not seated
properly in the card slot on the
programmer.
Remove and reseat the database
card to ensure it is inserted
properly.
Database card is full. Use another database card .
Database card is locked from
editing.
Exit the patient database and
return to the Main Menu screen,
remove the database card and
slide the lock on the side of the
card to a down (unlocked) position,
and reinsert the database card .
Database card is damaged or
corrupt.
Replace the database card .
Database is not
restored from the
database card .
Database card is not seated
properly in the card slot on the
programmer.
Remove and reseat the database
card to ensure it is inserted
properly.
Programmer will not read the data
stored on the database card .
Try to view fi les on the database
card using another programmer.
If the fi les are viewable,
then the programmer may be
malfunctioning. Contact Technical
Support.
Database card is locked from
editing.
Exit the patient database and
return to the Main Menu screen,
remove the database card and
slide the lock on the side of the
card to a down (unlocked) position,
and reinsert the database card .
Database card is damaged or
corrupt.
Replace the database card . (You
will not be able to restore your
database.)
80 Rapid Programmer System Clinician’s Manual
Performing a Soft Reset
If the programmer is locked and will not respond to input, you can use
the soft reset feature to restart the system. This is similar to rebooting
your computer. When you soft reset the system, stored patient records and
confi gured settings are unaffected. Note that even if the programmer is
locked, you can still turn off stimulation to the patient using the power key
on the patient programmer or MTS™ system.
CAUTION: When performing a soft reset , you will lose any unsaved
data.
If the Device Is Locked
Using the stylus, press the small recessed button located on the right
side of the programmer. The system will reset and return to the Main
Menu screen.
Figure 59: Using the stylus to perform a soft reset
If the System Responds to Input
1. On the Main Menu screen, tap the Tools menu, and then tap Reset
PDA. The Confirm Reset message appears.
2. Tap OK. The programmer resets.
Rapid Programmer System Clinician’s Manual 81
Reinstalling the Software
If you are having problems operating your system after performing a soft
reset , you may try reinstalling the software to reinitialize the system. When
you reinstall the software, stored patient records and confi gured settings
except system volume are unaffected.
To reinstall the software, follow these steps:
1. On the Main Menu screen, tap Tools, then tap Reinstall Software. The
Confirm Reinstall message appears.
Figure 60: The Confi rm Reinstall message
2. Tap OK. The “Installing Rapid Programmer™ 3.7” screen appears and
the PDA reinstalls the programmer software.
3. When the Installation Complete screen appears, tap Finish. The
programmer restarts.
Copying Log Files
If an error occurs and you are unable to operate the programmer, you can
copy the log fi les to a database card and then send the card to St. Jude
Medical for inspection or call Technical Support for further instructions.
To copy log fi les to a database card , follow these steps:
1. Insert a database card in the card slot on top of the programmer.
82 Rapid Programmer System Clinician’s Manual
2. On the Main Menu screen, tap the Tools menu, and then tap Copy Log
Files.
3. When the Copy Complete message appears, tap OK.
Returning the Rapid Programmer System for Service
If you must return the Rapid Programmer™ system for service, keep your
database card and erase the database from the programmer before you
return it. To do this, follow these steps:
1. Remove the database card containing the patient data, and store the
database card in a safe place to maintain data integrity (see Appendix
B for storage conditions).
2. From the Main Menu screen, tap Settings, then Brightness. The
Brightness Level screen appears.
3. If the on-screen keyboard is not visible, tap to open it.
4. Type clear and then tap the ENTER key ( ).
5. In the Clear All Patient Data message that appears, tap OK. All backup
databases are deleted from the Rapid Programmer system.
Figure 61: The Clear All Patient Data message
6. When the message appears confirming that the patient data has been
deleted, tap OK.
7. Return the programmer to St. Jude Medical (see “Technical Support”).
Rapid Programmer System Clinician’s Manual 83
Service and Ordering Information
This section provides information for contacting Technical Support and for
ordering parts and accessories.
Technical Support
For technical questions and support for you St. Jude Medical™
neuromodulation product, use the following information:
+1 972 309 8000
+1 800 727 7846 (toll-free within North America)
For additional assistance, call your local St. Jude Medical representative.
84 Rapid Programmer System Clinician’s Manual
Ordering Information
The following table lists model numbers for ordering the kit, kit
components, and accessories for the Rapid Programmer™ system.
Table 8: Accessories and kit components for the Rapid Programmer system
Model Number Description
1273 MTS communication cable
1280 IrDA communication cable
1291 Compact fl ash to DB9 cable
1300 Printer paper
1303 Battery for Bluetooth printer
1304 Car charger, RP Model 3834
1321 Power adapter for Bluetooth printer
1601 Rapid Programmer case
1622 Power adapter for RP Model 3834
1626 Stylus for RP Model 3834
1627 MTS joining sleeve for RP Model 3834
1629 Extended battery for RP Model 3834
3805 Rapid Programmer database card
3823 Bluetooth printer
3834 Rapid Programmer
Rapid Programmer System Clinician’s Manual 85
Appendix A: Printable Reports of the Rapid Programmer System
This section provides examples of the types of reports that you can print
from the Rapid Programmer™ system and provides information about what
information is available on each report. For more information on printing
reports , see the “Printing” section. The Rapid Programmer system can print
the following types of reports :
Neurostimulation record
Patient list
Lead diagnostics report
Neurostimulation Record
A neurostimulation record contains a record of all programs, as well as lead
and generator information, for a selected patient. The following fi gure shows
the location of information on the report.
1
3
4
5
7
6
2
Figure 62: Parts of a neurostimulation
record
Legend:
1. Patient information.
2. Generator and lead information,
including the initial session date and
the serial number of the generator.
3. Information about the device the
patient uses to control stimulation,
including the serial number of the
device and the number of hours that
stimulation has been on.
4. Date of the programming session.
5. Program number for the stim set
settings that follow.
6. Stimulation map of the associated
program.
7. Stim set settings of the associated
program. Stim set are listed in
ascending order (1, 2, 3, etc.).
86 Rapid Programmer System Clinician’s Manual
Patient List
The patient list shows patient names from the database and provides the
dates of their last revisions. The following fi gure shows the location of
information on the report.
1 2
Figure 63: Parts of a patient list
Legend:
1. Patient name.
2. Date of the patient’s last programming
session.
Lead Diagnostics Report
A lead diagnostics report shows the results of the diagnostic testing
(impedance testing) performed on a generator and leads for a program on an
Eon™, Prodigy™, Protégé™, or MTS™ system. The following fi gure shows
the location of information on the report.
1
3 4
2
Figure 64: Parts of a lead
diagnostics report
Legend:
1. Patient name and information about
the system being calibrated, including
stimulator and lead information.
2. Session date of the diagnostic test.
3. Number of the electrode that was
tested.
4. Impedance reading of the associated
electrode.
Rapid Programmer System Clinician’s Manual 87
Appendix B: Specifications
The following tables show hardware specifi cations for the Rapid
Programmer™ system (Model 3834).
Table 9: Rapid Programmer system specifi cations
System Specifications
OEM Model HP iPAQ 200 series
Memory 128 MB SDRAM
256 MB fl ash ROM
Display 4.0-in. VGA TFT, resolution 640 × 480 px
Communications Bluetooth® wireless technology 2.0, 10 meter range (approximately
33 feet), high-speed, low-power, short-range wireless
communication with other Bluetooth devices, support collaborative
coexistence with IEEE 802.11b/g
Indicator Lights Solid amber: battery charging
Solid green: battery charge complete
Solid blue: Bluetooth wireless technology is on
Battery 3800 mAh, lithium ion, removable, rechargeable battery
Table 10: Rapid Programmer system physical specifi cations
Physical Specifications
Metric U.S.
Height 13.38 cm 5.27 in
Width 7.54 cm 2.97 in
Depth 2.41 cm 0.95 in
Weight 238.00 g 8.40 oz
Table 11: Rapid Programmer system storage and operating conditions
Storage and Operating Conditions
Metric U.S.
Temperature Range Operating 10°C–40°C 50°F–104°F
Storage -10°C–55°C 14°F–131°F
Humidity Range (Noncondensing) Operating 30%–75% 30%–75%
Storage 30%–90% 30%–90%
Pressure Range Operating 70–101 kPa 10.2–14.6 psi
Storage 70–101 kPa 10.2–14.6 psi
88 Rapid Programmer System Clinician’s Manual
Appendix C: Regulatory Statements
This section contains regulatory statements about the system or its
accessories.
Statement of FCC Compliance
This equipment has been tested and found to comply with the limits
for a Class B digital device, pursuant to part 15 of the FCC rules. These
limits are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates, uses,
and can radiate radiofrequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will not
occur in a particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined by
turning the equipment off and on, the user is encouraged to try to correct
the interference by one or more of the following measures:
Reorient or relocate the receiving antenna.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit different from that to
which the receiver is connected.
Consult the dealer or an experienced radio/TV technician for help.
Operation is subject to the following two conditions:
This device may not cause harmful interference.
This device must accept any interference received, including
interference that may cause undesired operation.
Modifi cations not expressly approved by the manufacturer could void the
user’s authority to operate the equipment under FCC rules.
Declaration of Conformity (Industry Canada) Notice to Users of Radio and Television
This Class B digital apparatus meets all the requirements of the Canadian
interference-causing equipment regulations.
Rapid Programmer System Clinician’s Manual 89
Disposal Guidelines for Battery-Powered Devices
This device contains a battery and the label is affi xed to this device in
accordance with European Council directives 2002/96/EC and 2006/66/
EC. These directives call for separate collection and disposal of electrical
and electronic equipment and batteries. Sorting such waste and removing
it from other forms of waste lessens the contribution of potentially toxic
substances into municipal disposal systems and into the larger ecosystem.
Return the device to St. Jude Medical at the end of its operating life.
90 Rapid Programmer System Clinician’s Manual
Appendix D: Symbols and Definitions
The following symbols may be used in this document and on some of the
products and packaging.
Table 12: Symbols and defi nitions
Symbol Description
Caution, consult accompanying documents
Consult instructions for use
Follow instructions for use on this website
Magnetic Resonance (MR) Unsafe, an item poses unacceptable
risks to the patient, medical staff, or other persons within an
MR environment
Expiration date
Date of manufacture
Manufacuring facility
Temperature limits for storage conditions
Humidity limits
Pressure limits
Catalog number
Manufacturer
Contents quantity
Programmer
Accessories
SN Serial number
Rapid Programmer System Clinician’s Manual 91
Symbol Description
Batch code
Prescription use only
This product shall not be treated as household waste. Instead
it is the user’s responsibility to return this product to St. Jude
Medical for reprocessing.
By ensuring that this product is disposed of properly, you
will help prevent potential negative consequences for the
environment and human health, which could be caused by
inappropriate waste handling of this product. The recycling of
materials will help to conserve natural resources.
For more information about how to return this product for recycling,
please contact St. Jude Medical.
Complies with AS/NZS standards: AS/NZS 4268:2003
This device is listed by Underwriters Laboratories (UL) Inc. as
certifi ed
This equipment has been tested and found to comply with the
limits for a Class B digital device, pursuant to Part 15 of the
FCC rules.
92 Rapid Programmer System Clinician’s Manual
Appendix E: Electromagnetic Compatibility Guidelines
The Rapid Programmer™ system (Model 3834), hereafter the system, is
part of a medical equipment system and should be used with the following
guidance:
Table 13: Guidance and Manufacturer’s Declaration—Electromagnetic Emissions
The system is intended for use in the electromagnetic environment specifi ed below. The
customer or user of the system should ensure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment Guidance
RF emissions
CISPR 11
Group 1 The system uses RF energy only for its internal
function.
Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class A The system is suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fl uctuations/
fl icker emissions
IEC 61000-3-3
Not applicable
Rapid Programmer System Clinician’s Manual 93
Index
A
Active Balancing, 52, 55
Amplitude
Comfort, 38, 40, 52
Maximum tolerable, 55
Overriding usage range for, 55, 56
Perception, 27, 28, 38, 40, 41,
42, 43, 51, 52, 53, 55
Target, 41
B
Battery, 1, 2, 3, 9, 10, 17, 19, 20,
21, 22, 23, 40, 73, 74, 75, 76,
77, 84, 87
Charging, 10
Disposal, 89
Status, 22
Bolus mode, 50, 51
Burst program
Adding a burst pulse, 38, 41
Defi nition, 37
Deleting a burst pulse, 36, 38
Setting frequency, 39
Switching to, 29, 35, 37
C
Calibrating
Touch screen, 24
Card, 8, 15, 16, 57, 65, 66, 67,
79, 81, 82, 84
Inserting, 15
Removing, 16
Connecting to devices, 12, 13
Continuous mode, 50, 51
Cycle mode, 50, 51
D
Database
Accessing, 58
Adding a patient, 61
Backup, 67
Buttons and icons, 57
Combining patient records, 64
Deleting, 82
Deleting a patient, 63
Editing a patient, 62
Exporting, 66, 79
Merging and copying, 65
Password, 58, 60
Restore, 18
Restoring, 67, 79
Demo, 18, 26, 27, 76, 77, 78
Dynamic MultiStim programming,
45, 46, 55
E
Electrodes, 42
Polarity, 38, 42
F
Frequency, 31, 38, 39, 45
I
Impedance testing, 29, 31, 48, 49,
68, 78
IPG
As anode, 38, 39
Modes, 50
L
Lead, 78
Diagnostics. See Impedance
testing
Lead diagnostics report, 49, 71
Leads
Selecting, 26, 30, 34
94 Rapid Programmer System Clinician’s Manual
M
Magnet mode, 29, 50
Maximum output. See Output limit
MultiSteering programming, 28, 31,
42, 43
MultiStim. See Dynamic MultiStim
O
Ordering information, 84
Output limit, 40
P
Printing, 68, 69, 70
Lead diagnostics report, 49
Neurostimulation record, 61
Patient list, 59
Program
Editing, 36
Saving, 52
Programming, 1, 2, 26, 27, 28, 31,
32, 34, 40, 41, 42, 43, 45, 46,
47, 51, 52, 54, 55, 74, 76, 77,
78
Burst, 29, 37, 39, 41
Tonic, 37, 39, 40, 42, 45, 47,
52, 55
Program mode, 50, 51
Programs
Adding, 35
Copying, 35
Deleting, 36
Downloading, 54, 55
Pulse width, 31, 39
R
Reports, 85
Lead diagnostics report, 68, 86
Neurostimulation record, 61, 68,
69, 85
Patient list, 57, 59, 68, 70
Types of, 68, 85
Resetting the MTS usage counter,
34
S
SD card. See Database card
Settings
Automatic shutoff, 23
Backlight, 19
Brightness, 20
Time and date, 21
Volume, 25
Soft reset, 75, 80, 81
Step size, 40
Stim sets
Adding, 41, 45, 48
Deleting, 36
Replacing, 47, 48
Selecting, 45
Stimulation map, 38, 43, 44
Symbols, 90
T
Technical support, 83
Test log, 26, 47
Tonic program
Defi nition, 37
Switching to, 37
Troubleshooting, 74
*100144565* 2016-09ARTEN100144565 A
St. Jude Medical
6901 Preston Road
Plano, Texas 75024
USA
+1 972 309 8000
St. Jude Medical
Coordination Center BVBA
The Corporate Village
Da VinciIaan 11 Box F1
1935 Zaventem
Belgium
+32 2 774 68 11
sjm.com