2
Model CM-100 On-Line Colorimeter FEATURES Monitors process colors and, concentrations in liquid and gaseous streams Fiber optic cable connections to remote cell. Easy automatic zero. No moving parts; measurement technique reduces effects of turbidity, dirt, bubbles 4-20mA output; serial output. Bright local display; alarm relays. Class 1, Div 1, Groups B,C,D EEx d IIC T6 Category 3G GL applied research, inc. P.O. Box 187 Grayslake, Illinois 60030 USA Telephone: 847/223-2220 Fax: 847/223-2287 1997 GL applied research, inc. Part Number 914000-000-1

Cm100 Online Colorimeter

  • Upload
    su-tan

  • View
    10

  • Download
    3

Embed Size (px)

DESCRIPTION

Cm100 Online Colorimeter

Citation preview

Page 1: Cm100 Online Colorimeter

Model CM-100On-Line

Colorimeter

FEATURES♦ Monitors process colors and,

concentrations in liquid andgaseous streams

♦ Fiber optic cable connections toremote cell. Easy automaticzero.

♦ No moving parts; measurementtechnique reduces effects ofturbidity, dirt, bubbles

♦ 4-20mA output; serial output.♦ Bright local display; alarm

relays.♦ Class 1, Div 1, Groups B,C,D♦ EEx d IIC T6 Category 3G

GL applied research, inc.P.O. Box 187

Grayslake, Illinois 60030 USATelephone: 847/223-2220

Fax: 847/223-2287

1997 GL applied research, inc.Part Number 914000-000-1

Page 2: Cm100 Online Colorimeter

SPECIFICATIONSCM-100

Performance Specifications

Calibration ASTM color standards in liquid streams(D-1500, D-156, D-1209, D2392). Con-centration of UV/VIS/NIR absorbingcomponents in liquid or gaseousstreams. Refer to Individual InstrumentSpecification Sheet for information on yourparticular instrument.

Repeatability<2% of scale.

Operating Temperature Range

0° to 55°C (32° to 131°F) standard; -20° to55°C (-4° to 131°F) optional).

Sample Pressure10 bar (147 psig) standard cell; 100 bar(1470 psig) at 150°C (302°F) optional.

Sample Temperature

-20° to 150°C (-4° to 302°F).

Signal Outputs

Digital Display3 ½ digit LED.

Analog OutputIsolated 4-20mA self-powered.Linear scale; 4mA equals bottomof measurement range, 20mA equals top ofmeasurement range.

Serial OutputRS232 standard; RS485 or current loop optional.

Alarm RelaysDual-level SPDT alarm relay rated at 6 ampsat 28 VDC or 120/300 VAC. Alarm levelmay be set anywhere within the measuringrange of the instrument.

Electrical Specifications

Power115/230-240 VAC, 50/60 Hz. Single phase,100 watts nominal.

Safety Characteristics

Designed for use in Class I, Division 1,Groups B,C, and D. Enclosure is CENELECand CSA certified, FM approved. NEMA3,4,7,9. CE certifiedATEX Directive EEx d II C T6 Cat. 3G.

Physical Dimensions

Control/Power Panel254 (H) x 381 (W) x 152 (D) mm (10 x 15 x6 inches).

Flow Cell EnclosureTypical: 203 mm (H) x 203 mm (W) x 152mm (D) (8” x 8” x 6”).

NOTE: Flow cell enclosure dimensions mayvary with application. Flanged Cellssupplied without enclosures.

NOTE: Wetted material in flow cell consistsof 316 stainless steel, borosilicate glass,Teflon, or other inert materials as requiredby application.

Due to GL applied research’s commitment tocontinual product improvement, specificationssubject to change without notice.