  • Compendium of Background InformationCOBI-CD

    cobas e 411

  • cobas e 411Revision history

    Order numbers

    Edition notice cobas e 411 Compendium of Background Information

    Intended use This CD is provided as an information source for background information regarding the cobas e 411 analyzer The information on this CD is available in PDF-format and requires Adobe Acrobat Reader to be installed.

    Copyrights 2001-2006, Roche Diagnostics GmbH. All rights reserved.

    Trademarks COBAS, COBAS C, COBAS E, ELECSYS, and LIFE NEEDS ANSWERS are trademarks of Roche.

    All other trademarks are the propery of their respective owners.

    Instrument approvals The cobas e 411 analyzer meets the requirements stated in Directive 98/79/EC of the European Parliament and the Council of the European Union (EU) on in vitro diagnostic medical devices. Furthermore, the cobas e 411 analyzer is manufactured and tested according to International Standard IEC 61010-1, 2nd edition, Safety requirements for electrical equipment for measurement, control and laboratory use, Part 1: General requirements. This International Standard is equivalent to the national standards Underwriters Laboratories (UL) 61010-1 2nd edition for the USA, and CAN/CSA C22.2 No. 61010-1:2004 for Canada. Compliance is demonstrated by the following marks:

    Notice to the purchaser The purchase of this product allows the purchaser to use it solely for detection by ECL Technology for human in vitro diagnostic uses. No general patent or other license of any kind other than this specific right of use from purchase is granted hereby. This product may not be used by purchaser to conduct life science research and/or

    Manual Version

    Template Version

    Revision date Changes

    1.0 3.0

    Language Order number

    English 0490 5148 018

    French 0490 5148 080

    German 0490 5148 001

    Italian 0490 5148 050

    Portuguese 0490 5148 046

    Spanish 0490 5148 036

    Complies with the IVD (in vitro diagnostic) directive 98/79/EC.

    Issued by Underwriters Laboratories, Inc. (UL) for Canada and the USA.C USRoche Diagnostics

    2 COBI-CD Version 1.0

  • cobas e 411development, patient self-testing, drug discovery and/or development or in any veterinary, food, water or environmental testing or use.

    US Pat. 5,147,806; US Pat. 5,779,976; US Pat. 6,325,973; US Pat. 5,466,416; US Pat. 5,624,637; US Pat. 5,720,922; US Pat. 5,061,445; US Pat. 5,068,088; US Pat. 5,247,243; US Pat. 5,296,191 and corresponding patents in other countries.

    Contact addresses

    Manufacturer

    Authorized Representative

    Hitachi High-Technologies Corporation

    24-14. Nishi-shimbashi. 1-chome. Minato-ku

    Tokyo. 105-8717 JAPAN

    Roche Diagnostics GmbH

    Sandhofer Strasse 116

    D-68305 Mannheim

    GermanyRoche Diagnostics

    COBI-CD Version 1.0 3

  • cobas e 411Conventions used

    Visual cues are used to help you quickly locate and interpret information in this manual. This section explains formatting conventions used in this manual.

    Symbols The following symbols are used:

    Abbreviations The following abbreviations are used:

    Symbol Used for

    a Procedural step

    o List item

    e Cross-reference

    h Call up of screen

    Note

    Caution

    Warning

    Laser Radiation

    Biohazard

    Disk system specific

    Rack system specific

    Abbreviation Definition

    A

    ANSI American National Standards Institute

    C

    CBT Computer Based Training

    CCITT Comit consultatif international tlphonique et tlgraphique

    (Consultative Committee on International Telegraph and Telephone)

    CE Conformit Europenne

    CLAS 2 Clinical Laboratory Automation System 2

    CLIA Clinical Laboratory Improvement Amendments

    COBI-CD Compendium of Background Information

    CSA Canadian Standards AssociationRoche Diagnostics

    4 COBI-CD Version 1.0

  • cobas e 411D

    dBA decibel weighted against the A-frequency response curve. This curve

    approximates the audible range of the human ear.

    DIL diluent

    E

    EC European Community

    ECL electrochemiluminescence

    EMC electromagnetic compatibility

    EN european standard

    F

    FIFO first in first out

    H

    HCFA Health Care Financing Administration

    I

    IEC International Electrical Commission

    IS internal standard (ISE module)

    IVD in vitro diagnostic directive

    K

    KVA kilovolt-Ampere. Unit for expressing rating of AC electrical

    machinery.

    L

    LDL lower detection limit see analytical sensitivity

    LIS laboratory information system

    LLD liquid level detection

    M

    MSDS material safety data sheet

    N

    NCCLS National Committee for Clinical Laboratory Standards

    P

    PC/CC ProCell/CleanCell

    Q

    QC quality control

    R

    REF reference solution for ISE module

    S

    SD standard deviation

    S/R sample/reagent

    SVGA Super Video Graphics Adapter

    T

    TPA tripropylamine

    U

    UL Underwriters Laboratories Inc.

    Abbreviation DefinitionRoche Diagnostics

    COBI-CD Version 1.0 5

  • cobas e 411V

    VDE Verband Deutscher Elektrotechniker (association of German

    electrical engineers)

    Abbreviation DefinitionRoche Diagnostics

    6 COBI-CD Version 1.0

  • cobas e 411Table of contents

    Revision history 2Contact addresses 3Conventions used 4Table of contents 7

    Mechanical theory Part A

    1 Mechanical theoryIntroduction A-5Preparative operations A-6Test protocols A-7Assay sequence A-8Workflow and throughput A-11Operation flow in analysis A-13Detailed assay sequence A-14Dilution steps A-21Pretreatment steps A-22Analyzer status conditions A-23

    Measurement technology Part B

    2 ECL technologyECL measuring principles B-5Advantages of ECL technology B-10

    Test principles Part C

    3 Test principlesTest principles C-5

    4 Reagent conceptIntroduction C-15Data transfer media C-15Data transfer rules C-16Reagents for cobas e 411 analyzer tests C-16Product labeling C-18Data links C-19Calibration C-21Master calibration C-22Lot calibration C-23Reagent pack calibration C-23Difference between lot and reagent calibration C-24Calibration procedures C-25Calibration stability C-26Calibration validation C-26Calibration assessment C-27Calibration of quantitative assays C-30Calibration of qualitative assays C-33Result calculation for qualitative assays C-33

    Quality control Part D

    5 Quality control conceptControl target value (first) assignment D-5

    Index Part E

    Index E-3Roche Diagnostics

    COBI-CD Version 1.0 7

  • cobas e 411Roche Diagnostics

    8 COBI-CD Version 1.0

  • 1 Mechanical theory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3

    Mechanical theory A

  • cobas e 411 1 Mechanical theoryTable of contentsMechanical theory

    This chapter provides an overview of the mechanical theory of the cobas e 411 analyzer. The assay sequence and operational flow are described, as well as dilution steps.

    Introduction .................................................................................................................. 31

    Preparative operations .................................................................................................. 32

    Test protocols ................................................................................................................. 33

    Assay sequence ............................................................................................................... 34

    Run operation .......................................................................................................... 34

    First incubation at 37 C ................................................................................... 34

    Pipetting of additional reagent ........................................................................ 34

    Second incubation at 37 C ............................................................................... 35

    Pipetting of additional reagent (pretreatment assays) .................................... 35

    Third incubation at 37 C (pretreatment assays) ............................................. 35

    Aspirating the reaction mixture ....................................................................... 35

    Cleaning the measuring cell ............................................................................. 35

    Finalization ........................................................................................................ 35

    Workflow and throughput ............................................................................................ 37

    Effects of test combinations on throughput ......................................................... 37

    9-minute tests only ............................................................................................ 37

    18-minute tests only .......................................................................................... 37

    Combination of 9- and 18-minute tests ........................................................... 37

    27-minute tests only .......................................................................................... 38

    Combination of 18- and 27-minute tests ......................................................... 38

    Typical test durations .............................................................................................. 38

    Operation flow in analysis ............................................................................................ 39

    Detailed assay sequence ................................................................................................ 40

    Preoperational steps ............................................................................................... 40

    Dispensation of reagent 1, reagent 2, and sample (disk system) .......................... 41

    Dispensation of reagent 1, reagent 2, and sample (rack system) .......................... 43

    First incubation ....................................................................................................... 44

    Microbead preparation ........................................................................................... 45

    In this chapter Chapter 1Roche Diagnostics

    COBI-CD Version 1.0 A-3

  • cobas e 411 1 Mechanical theoryTable of contentsMicrobead aspiration and dispensation ................................................................ 45

    Second incubation ................................................................................................... 45

    Preparations for the measurement process ........................................................... 46

    Measurement process .............................................................................................. 46

    Signal detection and conversion ............................................................................. 47

    Automatic analyzer cycles ....................................................................................... 47

    Dilution steps ................................................................................................................ 47

    Assay with one-step dilution ............................................................................. 47

    Assay with two-step dilution ............................................................................. 48

    Pretreatment steps ......................................................................................................... 48

    Pretreatment assay ............................................................................................. 48

    Analyzer status conditions ............................................................................................ 49

    A. Stop (analyzer stop) ............................................................................................ 49

    A. Stop/L. Stop (analyzer stop/line stop) ............................................................... 49

    A. Stop/R. Stop (analyzer stop/rack stop) .............................................................. 49

    BC card scan ............................................................................................................ 49

    E. Stop (emergency stop) ........................................................................................ 49

    Finalization .............................................................................................................. 49

    Finalization maint. .................................................................................................. 50

    Initialization ............................................................................................................ 50

    L. & A. reset all (line & analyzer) ............................................................................ 50

    L. Stop (line stop) .................................................................................................... 50

    Liquid flow cleaning ................................................................................................ 50

    M. Cell preparation ................................................................................................. 50

    Operation ................................................................................................................. 50

    P. Stop (partial stop) ................................................................................................ 50

    R. Stop (rack stop) ................................................................................................... 51

    Rack clear ................................................................................................................. 51

    Reagent scan ............................................................................................................ 51

    S/R pipetter prime ................................................................................................... 51

    S/R probe LLD volt. ................................................................................................. 51

    S. Stop (sampling stop) ........................................................................................... 51

    S. Stop-S. Scan ......................................................................................................... 51

    Sample scan ............................................................................................................. 51

    Sipper LLD volt. ...................................................................................................... 52

    Sipper pipet. prime .................................................................................................. 52

    Standby .................................................................................................................... 52

    Stop .......................................................................................................................... 52

    System reset .............................................................................................................. 52Roche Diagnostics

    COBI-CD Version 1.0 A-4

  • cobas e 411 1 Mechanical theoryIntroductionIntroduction

    The cobas e 411 analyzer automates the immunoassay reactions utilizing electrochemiluminescence (ECL). The individual test steps and how the system performs the necessary procedures are described here.

    e For information on ECL immunoassay reaction methods, see Chapter 2 ECL technology. Roche Diagnostics

    COBI-CD Version 1.0 A-5

  • 1 Mechanical theory cobas e 411Preparative operationsPreparative operations

    Once the analyzer is switched on, the initialization process starts. During initialization, the mechanisms are reset to their home positions.

    e Figure A-1 shows the run preparation process for the cobas e 411 hardware.

    Figure A-1 Run preparation process

    Start

    First order?

    Mechanical units reset

    Were reagents exchanged?

    Scan of the reagent barcode

    Counting AssayTips and

    AssayCups

    Have 90 or more minutes passed since

    the last mixing?

    Microbeads mixing

    Is the inventory sufficient?

    Scheduling

    First pipetting

    Pipetting continues

    Clearing the incubator and the AssayTip

    and AssayCup trays

    Volume check for

    ProCell and CleanCell

    Alarm inventory short (item)

    45-xx-01

    Preparation cycle

    First sipping

    Resume cycle

    Sipping continues

    No

    Yes

    Yes

    Short

    Enough

    NoRoche Diagnostics

    A-6 COBI-CD Version 1.0

  • cobas e 411 1 Mechanical theoryTest protocolsTest protocols

    There are 28 test protocols that can be used on the analyzer. These protocols are predefined by Roche Diagnostics for each test and cannot be changed by the operator.

    No. Step 0 Inc 0 Step 1 Inc 1 Step 2 Inc 2 Det.

    0 B R1 R2 S i 1 D

    1 B R1 S i 1 R2 i 2 D

    2 R1 R2 S i 1 B i 2 D

    3 R1 S i 1 B R2 i 2 D

    4 R0 S B R1 R2 DL i 1 D

    5 R0 S B R1 DL i 1 R2 i 2 D

    6 R0 S R1 R2 DL i 1 B i 2 D

    7 R0 S R1 DL i 1 B R2 i 2 D

    8 R0 S -> DL1 R0 B R1 R2 DL i 1 D

    9 R0 S -> DL1 R0 B R1 DL i 1 R2 i 2 D

    10 R0 S -> DL1 R0 R1 R2 DL i 1 B i 2 D

    11 R0 S -> DL1 R0 R1 DL i 1 B R2 i 2 D

    12 PS S i 0 B R1 R2 i 1 D

    13 PS S i 0 B R1 i 1 R2 i 2 D

    14 PS S i 0 R1 R2 i 1 B i 2 D

    15 PS S i 0 R1 i 1 B R2 i 2 D

    16 RM S i 0 B R1 R2 DL i 1 D

    17 RM S i 0 B R1 DL i 1 R2 i 2 D

    18 RM S i 0 R1 R2 DL i 1 B i 2 D

    19 RM S i 0 R1 DL i 1 B R2 i 2 D

    20 RM S -> DL1 RM i 0 B R1 R2 DL i 1 D

    21 RM S -> DL1 RM i 0 B R1 DL i 1 R2 i 2 D

    22 RM S -> DL1 RM i 0 R1 R2 DL i 1 B i 2 D

    23 RM S -> DL1 RM i 0 R1 DL i 1 B R2 i 2 D

    24 R1 R1 i 1 D'

    25 R1 R2 i 1 D'

    26 R2 R2 i 1 D'

    27 PS1 PS2 S i 0 R1 R2 i 1 B i 2 D'

    28 PS1 PS2 S i 0 R1 i 1 B R2 i 2 D'

    29 i 1 D'

    ... (Reserve) i 1 D'

    63 i 1 D'

    Table A-1 Test protocols R1 = Reagent 1

    R2 = Reagent 2

    B = Beads (microbeads in the assay reagent pack)

    RO = Universal diluent (special reagent pack)

    PS = Pretreatment solution (for assays such as B12, Folat, and Anti-HBc)

    RM = Pretreatment for IgM

    S = Sample/calibrator/control

    DL = Diluted sample

    D = Detection with magnet drive

    D' = Detection without magnet drive

    I = IncubationRoche Diagnostics

    COBI-CD Version 1.0 A-7

  • 1 Mechanical theory cobas e 411Assay sequenceAssay sequence

    An immunological ECL test is made up of various pipetting steps, at least one incubation period and a measurement step. Generally at least three test components (sample, reagent and microbeads) are pipetted into an AssayCup. After the appropriate incubation period, the reaction mixture is aspirated into the measuring cell where the measurement process takes place. Each of the required pipetting cycles is performed within a defined period (42 seconds).

    The number of pipetting steps, as well as the make up of the reaction mixture are dependent on the test method (1 or 2 step test). For some methods, predilution with diluent and/or pretreatment with a special reagent is necessary. Thus the number of pipetting steps is increased.

    The following steps apply in principle to all methods. The sequence of the individual processes differ from test to test.

    Run operation

    After the appropriate test selections for patient samples are made in the software, operation is started according to the predetermined test protocol for each assay selected. Initially, at least one reagent (R1 or R2) and the sample or microbeads (M) are aspirated one after another by the S/R probe. After each aspiration, the outside of the S/R probe AssayTip is cleaned at the rinse station. The sample and reagents are dispensed into a new AssayCup and the AssayTip is ejected into the solid waste tray.

    For some tests that require sample dilution or pretreatment, diluent or pretreatment reagent is pipetted together with sample into an AssayCup. An aliquot of the diluted/ pretreated sample is then dispensed with reagent into a second AssayCup. Therefore, certain tests with predilution/pretreatment may require two or more AssayCups.

    e For more information on dilution, see Dilution steps on page A-21.

    First incubation at 37 C

    The incubation times are 4.5 or 9 minutes long, depending on the test. Some tests require only two incubation periods, whereas tests that include pretreatment can require three incubation periods. During the incubation step(s) the immune complex products are formed.

    Pipetting of additional reagent

    Some assays (usually those with more than one incubation step) require additional reagent pipetting. As in the initial reagent pipetting step, a new AssayTip is picked up before reagent aspiration. The S/R probe AssayTip is washed at the rinse station after each liquid aspiration. The liquid is then dispensed into the corresponding AssayCup where the sample and other liquids were dispensed in the first pipetting step. The probe rises while dispensing the reaction mixture back into the AssayCup, thereby mixing the solution and accelerating the reaction in the AssayCup. The AssayTip is discarded into the solid waste tray when pipetting is complete.Roche Diagnostics

    A-8 COBI-CD Version 1.0

  • cobas e 411 1 Mechanical theoryAssay sequenceSecond incubation at 37 C

    If necessary, a second incubation step (4.5 or 9 minutes) occurs.

    If using a pretreatment assay, the second incubation is similar to that described above for First Incubation at 37 C.

    Pipetting of additional reagent (pretreatment assays)

    For pretreatment assays, reagent pipetting is similar to that described above for Pipetting of additional reagent occurs.

    Third incubation at 37 C (pretreatment assays)

    If necessary, a third incubation step (9 minutes) occurs for pretreatment assays.

    Aspirating the reaction mixture

    In this process, the sipper probe first aspirates ProCell (tripropylamine solution, TPA) to prepare the measuring cell. Then, the sipper probe aspirates the reaction mixture from the AssayCup and transfers it to the measuring cell. The sipper probe is washed at the rinse station and ProCell is aspirated again to rinse away the unbound constituents of reagent and sample. Next, the ECL reaction in the measuring cell occurs.

    Cleaning the measuring cell

    Once the measurement is complete, the measuring cell is cleaned with CleanCell and prepared for a new measurement process.

    It takes 42 seconds (one pipetting cycle) from the aspiration of the reaction mixture by the sipper probe until the measuring cell is filled with ProCell and ready for the next sample.

    Finalization

    Thirty minutes after documentation of the last result, the sipper pipetter flushes system water through the sipper probe, and then fills the measuring cell with ProCell before the analyzer returns to Standby mode.

    After this procedure, every 30 minutes the waste pump of the S/R rinse station runs for 2 seconds (waste consumption approximately 12 mL). This procedure stops if the operation switch is switched off.

    e Figure A-2 shows the finalization process for the cobas e 411 hardware. Roche Diagnostics

    COBI-CD Version 1.0 A-9

  • 1 Mechanical theory cobas e 411Assay sequenceFigure A-2 Finalization process

    Last sipping

    Ten cycles waiting for the new order

    Gripper moves to its home position

    Cleaning the sipper flow with system water

    Pipetter prime

    Pipetter end wash

    Finalization

    Sipper prime

    Filling the sipper nozzle with water

    Standby

    Filling the measuring cell with ProCellRoche Diagnostics

    A-10 COBI-CD Version 1.0

  • cobas e 411 1 Mechanical theoryWorkflow and throughputWorkflow and throughput

    The workflow on the cobas e 411 analyzer system is entirely sample-orientated. Owing to the availability of a new disposable AssayTip for each test, there is no risk of contamination. Therefore it is possible to perform assays in any sequence, thus allowing samples to be completed one after the other.

    When all assays on the system are 18-minute assays, the optimal throughput of 88 results per hour can be reached, producing a result every 42 seconds. In combination with 9- or 27-minute assays, or in combination with two-step dilution assays, the instrument will slow down, depending on the percentage and sequence of tests with other incubation times.

    Effects of test combinations on throughput

    The various available tests have different durations. The throughput of the cobas e 411 analyzer depends upon the way in which tests of a given duration are combined, as explained for each of the following combinations. There may be short periods of throughput slow-down on the disk system due to the loading of a new sample disk. Such gaps do not occur when the rack system is used, because the Roche Diagnostics/Hitachi 5-position racks load continuously.

    9-minute tests only

    9-minute tests have two incubation periods, each of 4.5 minutes duration. If only 9-minute tests are performed, the optimal throughput will always be reached regardless of the test mixture.

    All 9-minute tests follow the same time protocol. Therefore, there will be no timing conflicts. In one 42-second cycle, the cobas e 411 will simultaneously perform S1 (first reagent pipetting), S2 (second reagent pipetting), and D (detection).

    18-minute tests only

    18-minute tests have two incubation periods, each of 9 minutes duration. If only 18-minute tests are performed, the optimal throughput will always be reached regardless of the test mixture.

    All 18-minute tests follow the same time protocol. Therefore, there will be no timing conflicts. In one 42-second cycle, the cobas e 411 will simultaneously perform S1 (first reagent pipetting), S2 (second reagent pipetting), and D (detection).

    Combination of 9- and 18-minute tests

    When tests of these two durations are combined, the throughput of the cobas e 411 depends on the percentage and distribution of the 9-minute tests. As a limiting factor, it is not possible to plan the detection of two tests during one 42 second cycle. When scheduling the first pipetting of a 9-minute test, the system has to be sure to have an open cycle for detection 9 minutes later. Depending on the percentage and distribution of the 9-minute assays, throughput may or may not be affected. If the number of requested 9-minute tests is very small, larger throughput gaps will exist.Roche Diagnostics

    COBI-CD Version 1.0 A-11

  • 1 Mechanical theory cobas e 411Workflow and throughput27-minute tests only

    27-minute tests have three incubation periods, each of 9 minutes duration. If only 27-minute assays are performed, the throughput of the cobas e 411 is reduced to 44 results per hour. Every 13 cycles, the cobas e 411 comes into a timing problem. It is not possible to perform a S0 (pretreatment pipetting) together with a S1 (first reagent pipetting) within one 42 second cycle. When this happens, the instrument will stand for 13 cycles (9 minutes) until it can pipette again without conflict.

    Combination of 18- and 27-minute tests

    When 18- and 27-minute assays are combined, the number of gaps created depends on the assay mix and on the exact test sequence. The gaps can vary from 1 to 13 idle cycles (42 seconds to 9 minutes). Limiting factors are that only one detection can take place during one 42-second cycle, and that S0 (the pretreatment step) cannot be combined with S1 (the first reagent pipetting).

    Typical test durations

    e Table A-2 contains details of the duration of some typical tests. This is not a complete list of tests, but is provided as an example.

    Test 9 minutes 18 minutes 27 minutes

    Thyroid TSH, T3, FT3, T4, FT4,

    T-uptake, TG, Anti-TG,

    Anti-TPO

    Fertility hCG LH, FSH, Prolactin, Prog,

    Testo, E2, HCG+beta,

    Cortisol, DHEAS, SHBG

    Cardiac CK-MB, Troponin T,

    Myoglobin

    CK-MB, Troponin T,

    Myoglobin, Digoxin,

    Digitoxin, proBNP

    Oncology PSA, fPSA, CEA, AFP,

    CA 125m CA 15-3 II,

    CA 19-9, CA 72-4,

    Cyfra 21-1, NSE, S100

    Infectious disease HBsAg, anti-HBs,

    anti-HBc IgM(a)

    anti-HAV IgM *

    Anti-HAV, Anti, HBe,

    HBeAg, HIV Antigen,

    HIV combi

    anti-HBc

    Anemia Ferritin Vit B12, Folate

    Diabetes c-Peptide, e-Peptide, Insulin

    Bone B-CrossLaps, Total P1NP, N-MID Osteocalcin, PTH(b)

    Other IgE

    Table A-2 Durations of some typical tests

    (a) 18-minute test with two-step predilution

    (b) Under developmentRoche Diagnostics

    A-12 COBI-CD Version 1.0

  • cobas e 411 1 Mechanical theoryOperation flow in analysisOperation flow in analysis

    e Figure A-3 shows a flow chart of the operational process.

    (a) Short Turn Around Time

    Figure A-3 Operational process

    Rerun

    Assigned

    Pre-start inspection

    Switch on

    (Initialization and Standby)

    Check alarm button

    Routine operation

    Calibration and control

    Routine or STAT(a) sampling

    ------------------------------------------------------

    Results

    Sampling Stop

    (Finalization, Stop, and Standby)

    Switch off

    Maintenance

    Pre-routine operationRoche Diagnostics

    COBI-CD Version 1.0 A-13

  • 1 Mechanical theory cobas e 411Detailed assay sequenceDetailed assay sequence

    The mechanical process of the instrument is described below with a sandwich test, TSH (thyroid-stimulating hormone), used as an example. This example assumes that the reagent pack was already registered by the analyzer and does not need calibration. All results are calculated on the basis of an existing lot calibration.

    Preoperational steps

    When Start is pressed from Standby mode, the following preoperational steps occur:

    1. The analyzer resets all mechanisms to their respective home positions and accesses the data disk. Next, the S/R pipetter primes the S/R probe.

    2. The gripper checks for an AssayTip in position number 1 of the AssayTip trays. If this position is empty, the gripper remembers where it last left off and checks that position. If this position is empty, the gripper considers the whole tray empty and the Inventory screen is updated accordingly.

    3. During the AssayTip check, the S/R probe is checked for the presence of an AssayTip. The probe moves to the AssayTip eject station and performs the movements to eject an AssayTip. If an AssayTip is present, it is ejected.

    4. After the AssayTip check is complete, the AssayCups are checked in the same manner. During the cup check, the analyzer finishes priming the probes.

    5. Next, the gripper checks the last three of the five positions on the pipetting station. If an AssayCup is present, the analyzer goes through the following cup disposal sequence:

    a) The gripper places an AssayTip in position 1 of the pipetting station.

    b) The S/R probe picks up the AssayTip, descends into the AssayCup, and aspirates any liquid.

    c) The AssayCup is discarded, while the S/R probe moves to the rinse station and dispenses any aspirated liquid.

    d) The AssayTip is washed and then discarded.

    6. The gripper moves to the incubator, where it checks all 32 incubator positions. If an AssayCup is present, the gripper moves the AssayCup to position 5 on the pipetting station and uses the same procedure listed in step 5 to discard the AssayCup.

    7. The S/R probe AssayTip is discarded after all the incubator positions are checked.

    If the analyzer is in S. Stop, the gripper remembers where it last left off and checks for an AssayTip

    in that position.Roche Diagnostics

    A-14 COBI-CD Version 1.0

  • cobas e 411 1 Mechanical theoryDetailed assay sequenceDispensation of reagent 1, reagent 2, and sample (disk system)

    5. The S/R probe moves from its Standby position to the R1 aspiration position. While activating liquid level detection, the probe descends until it is 2 mm below the reagent surface and aspirates 50 L of R1.

    While the probe is aspirating R1, the gripper puts another AssayTip in position 1 of the pipetting station.

    6. If the S/R probe does not detect liquid as it descends, no reagent aspiration can occur, and an alarm is generated.

    7. After R1 aspiration, the S/R probe rises and moves to the rinse station. To prevent the aspirated R1 from coming into contact with the water in the rinse station, the probe aspirates 10 L of air. The rinse station externally washes the AssayTip.

    8. During step 7, the reagent rotor rotates until the TSH reagent pack is in the R2 position.

    9. The S/R probe moves from the rinse station to the R2 aspiration position while aspirating another 10 L of air. This air layer prevents R1 from mixing with R2. While activating liquid level detection, the probe descends until it is 2 mm below the reagent surface and aspirates 50 L of R2. While the probe is aspirating R2, the gripper moves an AssayCup to position 5 of the pipetting station.

    e See Figure A-4 for the location of the R2 aspiration position.

    A TSH sample is present on position 1 of the sample disk.

    1. After preoperational functions are complete, the gripper takes an AssayTip from the tip tray and transports it to position 1 of the pipetting station. The gripper returns to its Standby position.

    2. The sample disk rotates until position 1 is in the sampling position.

    3. The S/R probe moves to position 1 of the pipetting station, descends to obtain the AssayTip, rises, and returns to its Standby position.

    4. During this time, the reagent rotor rotates until the TSH reagent pack is at the cap open/close mechanism. The mechanism moves forward and opens the caps on the reagent pack. The disk rotates again to move the TSH reagent to the R1 position.

    A R1 aspiration position B R2 aspiration position

    Figure A-4 R1 and R2 aspiration positions

    A

    BRoche Diagnostics

    COBI-CD Version 1.0 A-15

  • 1 Mechanical theory cobas e 411Detailed assay sequence10. Upon completion of R2 aspiration, the S/R probe rises and moves to the rinse station. To prevent the aspirated R2 from coming into contact with the water in the rinse station, the probe aspirates another 10 L of air. The rinse station externally washes the AssayTip.

    11. After R2 aspiration, the reagent rotor rotates until the TSH reagent pack is at the cap open/close mechanism. The mechanism moves out and closes the caps.

    12. The S/R probe moves from the rinse station to the sampling position while aspirating another 10 L of air. While activating liquid level detection, the probe descends until it is 2 mm below the sample surface and aspirates 50 L of sample. During sample aspiration, clot detection is activated.

    13. The S/R probe moves from the sampling position to position 5 of the pipetting station. The probe descends until the tip reaches 2 mm below the calculated level of the reaction mixture surface and dispenses the sample, R2, and R1. The probe's downward displacement is determined by calculating the volume of the reaction mixture for the sample and using downward-displacement tables in the software. The probe does not rise during dispensation.

    e See Figure A-5 for the location of the sampling position for disk systems.

    14. After dispensation, the S/R probe moves to the tip eject position and ejects the AssayTip.

    A Sampling position

    Figure A-5 Sampling position (disk system)

    ARoche Diagnostics

    A-16 COBI-CD Version 1.0

  • cobas e 411 1 Mechanical theoryDetailed assay sequenceDispensation of reagent 1, reagent 2, and sample (rack system)

    2. The pusher arm pushes the racks in the A-Line forward to the B-Line. The arm returns to its home position. The first rack loads on the B-Line.

    e For additional information on the A-Line and B-Line, see the Sample/reagent area components section in the Analyzer components chapter of the cobas e 411 analyzer Operators Manual.

    3. As the rack incrementally moves on the B-Line, the rack barcode reader scans all five rack positions and rack ID. When scanning is complete, position 1 of the rack is in the sampling position.

    4. The S/R probe moves to position 1 of the pipetting station, descends to obtain the AssayTip, rises, and returns to its Standby position.

    5. During this time, the reagent rotor rotates until the TSH reagent pack is at the cap-open/close mechanism. The mechanism moves forward and opens the caps on the reagent pack. The disk rotates again to move the TSH reagent to the R1 position.

    e See Figure A-4 on page A-15 for details of the R1 and R2 aspiration positions.

    6. The S/R probe moves from its Standby position to the R1 aspiration position. While activating liquid level detection, the probe descends until it is 2 mm below the surface of the reagent and aspirates 50 L of R1.

    e See Figure A-4 for the location of the R1 aspiration position.

    While aspirating R1, the gripper puts another AssayTip in position 1 of the pipetting station.

    7. If the S/R probe does not detect liquid during descent, no reagent aspiration can occur, an alarm is generated.

    8. After R1 aspiration, the S/R probe rises and moves to the rinse station. To prevent the aspirated R1 from contacting the water in the rinse station, the probe aspirates 10 L of air. The rinse station externally washes the AssayTip.

    9. During step 8, the reagent rotor rotates until the TSH reagent pack is in the R2 position.

    10. The S/R probe moves from the rinse station to the R2 position while aspirating another 10 L of air. This air layer prevents R1 from mixing with R2. While activating liquid level detection, the probe descends until it is 2 mm below the surface of the reagent and aspirates 50 L of R2. While aspirating R2 the gripper moves an AssayCup to position 5 of the pipetting station.

    e See Figure A-4 for the location of the R2 aspiration position.

    11. Upon completion of R2 aspiration, the S/R probe rises and moves to the rinse station. To prevent the aspirated R2 from coming into contact with the water in the rinse station, the probe aspirates another 10 L of air. The rinse station externally washes the AssayTip.

    12. After R2 aspiration, the reagent rotor rotates until the TSH reagent pack is at the cap-open/close mechanism. The mechanism moves out and closes the caps.

    13. The S/R probe moves from the rinse station to the sampling position while aspirating another 10 L of air. While activating liquid level detection, the probe

    A TSH sample is present on position 1 of the rack.

    1. After preoperational functions are complete, the gripper takes an AssayTip from the tip tray and transports it to position 1 of the pipetting station. The gripper returns to its Standby position. Roche Diagnostics

    COBI-CD Version 1.0 A-17

  • 1 Mechanical theory cobas e 411Detailed assay sequencedescends until it is 2 mm below the surface of the sample and aspirates 50 L of sample. During sample aspiration, clot detection is activated.

    14. The S/R probe moves from the sampling position to position 5 of the pipetting station. The probe descends until the tip reaches 2 mm below where the calculated level of the reaction mixture surface should be and dispenses the sample, R2, and R1. The probe's downward displacement is determined by calculating the volume of the reaction mixture for the sample and using downward-displacement tables in the software. The probe does not rise during dispensation.

    e See Figure A-6 for the location of the sampling position for rack systems.

    15. After dispensation, the S/R probe moves to the tip eject position and ejects the AssayTip.

    First incubation

    1. The gripper picks and transports the cup containing the reaction mixture from the pipetting station to the incubator.

    2. The cup is incubated at 37 C for 9 minutes.

    3. During incubation, the analyzer continues to perform operations for other test(s) or sample(s), if necessary.

    A Sampling position

    Figure A-6 Sampling position (rack system)

    ARoche Diagnostics

    A-18 COBI-CD Version 1.0

  • cobas e 411 1 Mechanical theoryDetailed assay sequenceMicrobead preparation

    Before the first incubation is completed, the TSH microbeads are mixed to facilitate their aspiration and dispensation.

    1. The reagent rotor rotates until the TSH reagent pack is at the reagent cap-open/close mechanism. The mechanism moves out and opens the cap. The disk moves the reagent pack to the mixing position.

    2. The mixer moves over the reagent rotor and descends into the microbeads to 1.4 mm above the bottom of the bottle.

    3. The mixer stirs the microbeads to obtain a homogeneous suspension.

    4. During the mixing, the gripper obtains a fresh AssayTip and transports it to position 2 of the pipetting station.

    5. When mixing is complete, the mixer rises and returns to the rinse station where it descends and rotates in the rinse station for washing.

    6. At the same time, the reagent rotor rotates the TSH reagent pack to the microbead pipetting position.

    Microbead aspiration and dispensation

    1. The gripper grasps the incubating cup and transports it to position 5 of the pipetting station.

    2. The S/R probe moves to the pipetting station and obtains the fresh AssayTip and moves to the microbead pipetting position.

    3. While activating the liquid level detection, the S/R probe descends below the reagent surface and aspirates 40 L of microbeads.

    4. After reagent aspiration, the S/R probe rises, moves to position 5 of the pipetting station and descends to dispense the microbeads.

    5. After dispense, the S/R probe descends further and aspirates the entire volume of reaction mixture. The probe rises while dispensing the reaction mixture back into the cup, thereby mixing the solution and accelerating the reaction in the cup. This mixing takes place only once.

    6. The S/R probe moves to the tip eject position and discards the AssayTip.

    Second incubation

    1. The gripper picks the cup containing the mixed reaction mixture and returns it to the incubator.

    2. The cup is incubated at 37 C for 9 minutes.

    3. During incubation, the analyzer continues to perform operations for other test(s) or sample(s), if necessary.Roche Diagnostics

    COBI-CD Version 1.0 A-19

  • 1 Mechanical theory cobas e 411Detailed assay sequencePreparations for the measurement process

    Before the second incubation is completed, the sipper probe aspirates ProCell into the measuring cell to facilitate measurement.

    1. The sipper probe moves from its home position to a ProCell bottle and descends to 2 mm below the solution level and aspirates ProCell into the measuring cell. As the probe descends, liquid level detection is activated. The sipper probe can descend as low as 1.3 mm above the bottom of the ProCell bottle.

    2. The sipper probe rises.

    Measurement process

    1. The gripper picks and transports the cup that has completed its second incubation from the incubator to the aspiration station.

    2. The sipper probe moves to the aspiration station and descends into the cup.

    3. When the sipper probe detects the reaction mixture in the cup, it aspirates 150 L.

    4. After aspiration, the sipper probe rises, aspirates 10 L of air, and moves to the sipper rinse station to descend for rinsing.

    5. The gripper grasps the cup from the aspiration station, transports it to the cup disposal opening, and discards the cup.

    6. The sipper probe is rinsed.

    7. The sipper probe rises and moves to the ProCell position, descends into the bottle and aspirates ProCell in a set aspiration/dispensation sequence. The immune complexes are magnetically captured on the working electrode, but unbound reagent and sample are washed away by ProCell.

    e For additional information on the measuring cell, see Chapter 2 ECL technology.

    8. After the bound-free separation, a voltage is applied between the working electrode and the counter electrode. The ECL reaction is initiated and measured by the photomultiplier.

    e For additional information on binding and bound-free separation, see Chapter 3 Test principles.

    9. The sipper probe rises and moves to the CleanCell position and aspirates 20 L of air. The probe then descends into the CleanCell bottle and aspirates reagent. This procedure is repeated eight times. The alternatating flow of air and cleaning solution washes the measuring cell. During this washing process, a voltage is applied between the electrodes, which aids in the cleaning process.

    10. The sipper probe moves to the sipper rinse station, aspirates 20 L of air, and descends into the rinse station for washing.

    11. Finally, the sipper probe rises and moves to the ProCell bottle. The probe descends into the bottle and aspirates 500 L of ProCell. Next, the probe aspirates 90 L of ProCell and moves to the rinse station. At the rinse station, the probe dispenses 35 L to flush the probe and prepare it for the next sample. During the aspirations of the ProCell, a sequence of voltages is applied three times to prepare the electrodes for the next measurement.

    One cycle of the measurement process consumes approximately 2 mL each of ProCell and CleanCell.Roche Diagnostics

    A-20 COBI-CD Version 1.0

  • cobas e 411 1 Mechanical theoryDilution stepsSignal detection and conversion

    The measuring cell is kept at a constant 28 C throughout the measurement process. The photomultiplier tube detects and converts the ECL signal into an electric signal from which the cobas e 411 analyzer calculates assay results.

    Automatic analyzer cycles

    There are certain analyzer functions that occur automatically while the analyzer is switched on.

    o While the analyzer is in operation, the solid waste tray periodically shakes for 1.5 seconds.

    o While the analyzer is in Standby, the reagent rotor turns 90 every 30 minutes.

    o While the analyzer is in Standby, the rinse stations for the S/R probe and sipper probe are switched on for 2 seconds every 30 minutes.

    o Microbeads undergo a long mix when starting from Standby and then every 90 minutes until pipetting starts.

    o Microbeads undergo a short mix and then a short mix every 60 minutes for each reagent pack.

    Dilution steps

    The following is a description of how an assay with a dilution is performed, including the number of AssayTips and AssayCups used in the process.

    Assay with one-step dilution

    (1:2, 1:5, 1:10) AssayTip 1 ~ diluent (wash)* + sample

    AssayTip 1 Diluent (wash)* + sample AssayCup 1AssayTip 2 R1 (wash)* + R2 (wash)* AssayCup 2 (1st incubation)AssayTip 3 M (wash)* AssayCup 2 (2nd incubation)Detection

    Table A-3 Dilution steps for an assay with one-step dilution (1:2, 1:5, 1:10)* (wash) = the outside of the AssayTip is washed.

    R1 = Reagent 1

    R2 = Reagent 2

    M = MicrobeadsRoche Diagnostics

    COBI-CD Version 1.0 A-21

  • 1 Mechanical theory cobas e 411Pretreatment stepsAssay with two-step dilution

    (1:50, 1:100)

    Pretreatment steps

    In certain test protocols, pretreatment reagent is added before R1, R2, or M, as summarized in the following table.

    Pretreatment assay

    AssayTip 1 Diluent (wash)* + sample AssayCup 1AssayTip 2 Diluent (wash)*

    + diluted sample from AssayCup 1

    AssayCup 2

    AssayTip 3 R1 (wash)* + R2 (wash)*+ diluted sample from AssayCup 2

    AssayCup 3 (1st incubation)

    AssayTip 4 M (wash)* AssayCup 3 (2nd incubation)

    Detection

    Table A-4 Dilution steps for an assay with two-step dilution (1:50, 1:100)* (wash) = the outside of the AssayTip is washed.

    R1 = Reagent 1

    R2 = Reagent 2

    M = Microbeads

    AssayTip 1 PT1 (wash)* + PT2 (wash)*+ sample

    AssayCup 1 (1st incubation)

    AssayTip 2 R1 + pretreated sample in AssayCup 1 AssayCup 1 (2nd incubation)

    AssayTip 3 M (wash)* + R2+ reaction mixture in AssayCup 1

    AssayCup 1 (3rd incubation)

    Detection

    Table A-5 Pretreatment steps for an assay* (wash) = the outside of the AssayTip is washed.

    PT1 = Pretreatment 1

    PT2 = Pretreatment 2

    R1 = Reagent 1

    R2 = Reagent 2

    M = MicrobeadsRoche Diagnostics

    A-22 COBI-CD Version 1.0

  • cobas e 411 1 Mechanical theoryAnalyzer status conditionsAnalyzer status conditions

    The cobas e 411 analyzer has a number of status conditions. The status conditions usually seen during routine operation or maintenance procedures are listed below. Several other status conditions, most them seen during various adjustment or maintenance procedures performed by a Roche Diagnostics service representative, are not included below.

    e Refer to the Alarm screen for further information about instrument alarms.

    A. Stop (analyzer stop)

    A. Stop/L. Stop (analyzer stop/line stop)

    A. Stop/R. Stop (analyzer stop/rack stop)

    BC card scan

    This status is seen when a barcode card scan is initiated from the Control Definition or Calibration Data screens.

    E. Stop (emergency stop)

    An emergency stop condition exists. An alarm was issued. Take the appropriate measures to resolve the problem.

    Finalization

    This is the status of the analyzer when it is between the status conditions S. Stop and Standby.

    The analyzer is no longer able to continue operation. An alarm was issued. Take the appropriate measures to resolve the problem.

    The analyzer is already in A. Stop status when the lines stop operation.

    The analyzer is already in A. Stop status when the A-Line stops supplying racks to the B-Line. Roche Diagnostics

    COBI-CD Version 1.0 A-23

  • 1 Mechanical theory cobas e 411Analyzer status conditionsFinalization maint.

    This status occurs when Finalization Maintenance is initiated from the Maintenance screen.

    Initialization

    This status is seen when the cobas e 411 analyzer is switched on, or when Start is pressed from Standby.

    L. & A. reset all (line & analyzer)

    L. Stop (line stop)

    Liquid flow cleaning

    Liquid flow cleaning occurs when this function is initiated from the Maintenance screen.

    M. Cell preparation

    Measuring cell (M. Cell) preparation occurs when this function is initiated from the Maintenance screen.

    Operation

    This is the status during which the cobas e 411 analyzer performs its routine operations.

    P. Stop (partial stop)

    A partial stop condition exists. An alarm was issued. Take the appropriate measures to resolve the problem.

    L. and A. reset all status occurs when the corresponding function is initiated from the Maintenance screen. This function resets the analyzer and the lines.

    All lines stop operation. An alarm was issued. Take the appropriate measures to resolve the problem. Roche Diagnostics

    A-24 COBI-CD Version 1.0

  • cobas e 411 1 Mechanical theoryAnalyzer status conditionsR. Stop (rack stop)

    Rack clear

    Reagent scan

    This status is seen when a reagent scan is initiated from the Inventory screen.

    S/R pipetter prime

    This status occurs when the S/R (sample/reagent) pipetter prime is initiated from the Maintenance screen.

    S/R probe LLD volt.

    This status is seen when the analyzer is monitoring the liquid level detection voltage of the S/R (sample/reagent) probe. The check is initiated from the Voltage Monitor screen (Utility) folder.

    S. Stop (sampling stop)

    S. Stop-S. Scan

    Sample scan

    This status occurs when there are no more racks to process on the A-Line or B-Line.

    Rack clear status occurs when the corresponding function is initiated from the Maintenance screen. This function clears any remaining racks on the A-, B- or C-Lines.

    This status occurs when S. Stop is pressed or when sampling is complete.

    The analyzer is in S. Stop and a sample scan is requested from the Status screen, or S is pressed while the analyzer is in S. Stop.

    This status occurs when a sample scan is initiated from the Status screen.Roche Diagnostics

    COBI-CD Version 1.0 A-25

  • 1 Mechanical theory cobas e 411Analyzer status conditionsSipper LLD volt.

    The analyzer is monitoring the liquid level detection voltage of the sipper probe. The check is initiated from the Voltage Monitor screen (Utility) folder.

    Sipper pipet. prime

    This status occurs when the sipper pipetter prime is initiated from the Maintenance screen.

    Standby

    The analyzer is not performing any operations.

    Stop

    This status occurs when Stop is pressed or when a Stop alarm condition exists. If an alarm exists, take the appropriate measures to resolve the problem.

    System reset

    A system reset is initiated from the Maintenance screen. Roche Diagnostics

    A-26 COBI-CD Version 1.0

  • 2 ECL technology . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-3

    Measurement technology B

  • cobas e 411 2 ECL technologyTable of contentsECL technology

    This chapter provides an overview of the electrochemiluminescent technology in the cobas e 411 analyzer system. The use of the ruthenium complex and the measuring cell in which the reaction occurs are described.

    ECL measuring principles .............................................................................................. 5

    Use of the ruthenium complex ................................................................................ 5

    The ECL reaction at the electrode surface ............................................................... 6

    ECL signal generation ............................................................................................... 8

    ECL measuring cell .................................................................................................... 9

    Advantages of ECL technology ..................................................................................... 10

    In this chapter Chapter 2Roche Diagnostics

    COBI-CD Version 1.0 B-3

  • cobas e 411 2 ECL technologyTable of contentsRoche Diagnostics

    COBI-CD Version 1.0 B-4

  • cobas e 411 2 ECL technologyECL measuring principlesECL measuring principles

    Electrochemiluminescent (ECL) processes are known to occur with numerous molecules, including compounds of ruthenium, osmium, rhenium, and other elements.

    ECL is a process in which highly reactive species are generated from stable precursors at the surface of an electrode. These highly reactive species react with one another, producing light.

    The development of ECL/Origen immunoassays is based on the use of a ruthenium(II)-tris(bipyridyl) [Ru(bpy)3]

    2+ complex and tripropylamine (TPA). The final chemiluminescent product is formed during the detection step.

    e For further information on the ruthenium complex, refer to Figure B-1.

    The chemiluminescent reactions that lead to the emission of light from the ruthenium complex are triggered electrically, rather than chemically. This is achieved by applying a voltage to the immunological complexes (including the ruthenium complex) that are attached to streptavidin-coated microbeads. The advantage of electrically initiating the chemiluminescent reaction is that the entire reaction can be precisely controlled.

    Use of the ruthenium complex

    ECL technology uses a ruthenium chelate as the complex for the development of light. Salts of ruthenium-tris(bipyridyl) are stable, water-soluble compounds. The bipyridyl ligands can be readily modified with reactive groups to form activated chemiluminescent compounds.

    For the development of ECL immunoassays, a N-hydroxysuccinimide (NHS) ester of a modified Ru(bpy)3 complex is used because it can be easily coupled with amino groups of proteins, haptens, and nucleic acids. This allows the detection technology to be applied to a wide variety of analytes.

    Figure B-1 The ruthenium complex

    Ru

    N

    N

    N O

    O

    O

    N

    2+

    N

    N

    N

    ORoche Diagnostics

    COBI-CD Version 1.0 B-5

  • 2 ECL technology cobas e 411ECL measuring principlesThe ECL reaction at the electrode surface

    Two electrochemically active substances, the ruthenium complex and tripropylamine (TPA), are involved in the reactions that lead to the emission of light. Both substances remain stable as long as a voltage is not applied.

    The ECL reaction of ruthenium-tris(bipyridyl)2+ and TPA occurs at the surface of a platinum electrode. The applied voltage creates an electrical field, which causes all the materials in this field to react. TPA is oxidized at the electrode, releases an electron and forms an intermediate TPA radical-cation, which further reacts by releasing a proton (H+) to form a TPA radical (TPAo).

    e For further information on the detection of a ruthenium-labeled immune complex, refer to Figure B-2.

    In turn, the ruthenium complex also releases an electron at the surface of the electrode thus oxidizing to form the Ru(bpy)3

    3+ cation. This ruthenium cation is the second reaction component for the following chemiluminescent reaction with the TPA radical.

    e For further information on the ECL reaction at the electrode surface, refer to Figure B-3.

    Figure B-2 Detection of a ruthenium-labeled immune complex

    TPA

    TPA

    TPA+

    -H+

    e-

    e-

    Electrode

    Diffusion

    Photon

    Magnetic microbeadRoche Diagnostics

    B-6 COBI-CD Version 1.0

  • cobas e 411 2 ECL technologyECL measuring principlesTPAo and Ru(bpy)33+ react with one another, whereby Ru(bpy)3

    3+ is reduced to Ru(bpy)3

    2+ and at the same time forms an excited state through energy transfer. This excited state is unstable and decays, with emission of a photon at 620 nm to its original state. The reaction cycle then starts again. The tripropylamine radical reduces to by-products that do not affect the chemiluminescence process. TPA is used up and therefore must be present in excess. The reaction is controlled by diffusion of the TPA and the amount of ruthenium complex present. As TPA in the electrical field is depleted, the signal strength (light) is slowly reduced once the maximum is reached.

    Although TPA is depleted during measurement, the ruthenium ground-state complex is continually regenerated. This means that the ruthenium complex can perform many light-generating cycles during the measurement process. This has an inherent amplification effect that contributes to the sensitivity of the technology. Many photons can be created from one antigen-antibody complex.

    Figure B-3 The ECL reaction at the electrode surface

    TPA+

    TPA

    -H+

    Ru(bpy)33+

    e-

    e-

    e-

    Ru(bpy)32+

    Ru(bpy)32+

    TPA

    Photon (620 nm)

    Electrode surface

    excited stateground

    stateRoche Diagnostics

    COBI-CD Version 1.0 B-7

  • 2 ECL technology cobas e 411ECL measuring principlesECL signal generation

    The following figure illustrates a typical ECL signal generation. Viewed from an electrical perspective, the reaction can be explained as follows: When a voltage is applied to the electrode of the measuring cell, a brief peak of light emission occurs, which can be detected as the resulting ECL signal. A defined area under the curve is measured around the intensity maximum.

    The dotted line indicates the voltage at the electrode used to generate the ECL signal. The solid line is the actual light output measured by the photomultiplier detector.

    Figure B-4 ECL signal generation

    0.000

    50,000

    100,000

    150,000

    200,000

    250,000

    300,000

    350,000

    0.40 0.60 0.80 1.00 1.200.20

    0

    300

    600

    900

    1200

    1500

    ECL intensity (counts)

    Time [s]

    applied voltage [mV]Roche Diagnostics

    B-8 COBI-CD Version 1.0

  • cobas e 411 2 ECL technologyECL measuring principlesECL measuring cell

    The core of the detection unit is the ECL measuring cell, which is designed as a flow-through cell. The following figure shows the main components of the measuring cell:

    The temperature is maintained at 28C . Three operating steps are performed in the measuring cell:

    o Bound/free separation

    Using a magnet, the streptavidin microbeads that are coated with antigen-antibody complexes are uniformly deposited on the working electrode. A system buffer (ProCell) is used to wash the particles on the working electrode and to flush out the excess reagent and sample materials from the measuring cell.

    o ECL reaction

    To initiate the ECL reaction, the magnet is removed and a voltage is applied to the electrode. The microbeads that are coated with antigen-antibody complexes are deposited onto the electrode. The light emission is measured with a

    A Screw B Counter electrode C Optical window

    D Distance washer E Top cell F Cell gap

    G Gasket H O-ring I Diaphram

    J Reference electrode K Outlet fitting L Working electrode

    M Movable magnet N Inlet fitting O Cell body

    Figure B-5 Measuring cell of the detection unit

    O

    N M L K

    J

    I

    H

    G

    F

    EDCBARoche Diagnostics

    COBI-CD Version 1.0 B-9

  • 2 ECL technology cobas e 411Advantages of ECL technologyphotomultiplier. The system then uses the corresponding signals for the calculation of results.

    o Release of microbeads and cell cleaning

    Once the measurement is completed, the paramagnetic microbeads are washed away from the electrode surface with a special cleaning solution (CleanCell). The surface of the measuring cell is regenerated by varying the potential on the electrode. The cell is then ready for another measurement.

    Advantages of ECL technology

    ECL (electrochemiluminescence) is an innovative technology that offers distinct advantages over other detection techniques:

    o The extremely stable nonisotopic label means that you can use convenient liquid reagents.

    o The combination of enhanced sensitivity and short incubation times leads to high-quality assays and rapid results.

    o The large measuring range, encompassing five orders of magnitude, minimizes the need for dilutions and repeats, reducing handling time and reagent consumption.

    o The applicability of the technique to detect all analytes provides a solid platform for menu expansion.

    A Magnetic microbeads with bound antigen-antibody complex

    B Photomultiplier

    C Counter electrode D Unbound antibody (ruthenium-labeled)

    E Flow channel F Magnet

    G Working electrode

    Figure B-6 ECL measuring cell

    F

    D

    B

    A

    G

    C

    ERoche Diagnostics

    B-10 COBI-CD Version 1.0

  • 3 Test principles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-3

    4 Reagent concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-13

    Test principles C

  • cobas e 411 3 Test principlesTable of contentsTest principles

    This chapter provides an overview of the immunology test principles used by the cobas e 411 analyzer.

    Test principles ............................................................................................................... 13

    Competitive principle ............................................................................................. 14

    Sandwich principle .................................................................................................. 16

    Bridging principle ................................................................................................... 18

    In this chapter Chapter 3Roche Diagnostics

    COBI-CD Version 1.0 C-3

  • cobas e 411 3 Test principlesTable of contentsRoche Diagnostics

    COBI-CD Version 1.0 C-4

  • cobas e 411 3 Test principlesTest principlesTest principles

    Three test principles are available on the cobas e 411 analyzer:

    o Competitive principle for extremely small analytes

    o Sandwich principle (one or two steps) for larger analytes

    o Bridging principle to detect antibodies in the sample

    The following diagram illustrates the three available test principles:

    e For detailed descriptions of these principles, see:Competitive principle on page C-6

    Sandwich principle on page C-8

    Bridging principle on page C-10

    Figure C-1 ECL assay principles

    Sandwich principle for high molecular weight analysis

    Bridging principle to determine IgG and IgM

    Competitive principle for low molecular weight haptens

    Surface of para-magnetic microbead

    Streptavidin-biotin binding

    Analyte

    Antibody

    ECL label Roche Diagnostics

    COBI-CD Version 1.0 C-5

  • 3 Test principles cobas e 411Test principlesCompetitive principle

    This principle is applied to analytes of low molecular weight, such as free triiodothyronine (FT3).

    e Refer to Figure C-2 on page C-7 for an illustration of the competitive principle.

    o In the first step, sample and a specific anti-T3 antibody labeled with a ruthenium complex are combined in an assay cup.

    o After the first incubation, biotinylated T3 and streptavidin-coated paramagnetic microbeads are added. The still-free binding sites of the labeled antibody become occupied, with formation of an antibody-hapten complex. The entire complex is bound to the microbeads through the interaction of biotin and streptavidin.

    o After the second incubation, the reaction mixture containing the immune complexes is transported into the measuring cell. The immune complexes are magnetically captured on the working electrode, but unbound reagent and sample are washed away by ProCell.

    o In the ECL reaction, the conjugate is a ruthenium-based derivative and the chemiluminescent reaction is electrically stimulated to produce light. The amount of light produced is indirectly proportional to the amount of antigen in the patient sample.

    The concentration of the antigen is evaluated and calculated by means of a calibration curve that was established using standards of known antigen concentration.Roche Diagnostics

    C-6 COBI-CD Version 1.0

  • cobas e 411 3 Test principlesTest principlesFigure C-2 Competitive principle

    TPA ECL

    TPA

    COMPETITIVE PRINCIPLE

    FIRST REACTION

    Magnetic force and

    electrical potential

    Signal (light)

    Concentration

    SECOND REACTION

    LIGHT REACTION

    Antigen

    Biotinylated

    antigen

    Ruthenium-labeled

    antibody

    Streptavidin-coated

    microbead

    TripropylamineRoche Diagnostics

    COBI-CD Version 1.0 C-7

  • 3 Test principles cobas e 411Test principlesSandwich principle

    The sandwich principle is applied to higher molecular weight analytes, such as thyroid-stimulating hormone (TSH).

    e Refer to Figure C-3 on page C-9 for an illustration of the sandwich principle.

    o In the first step, the patient sample is combined in an AssayCup with a reagent containing biotinylated TSH antibody and a ruthenium-labeled TSH-specific antibody in an assay cup. During a 9-minute incubation step, antibodies capture the TSH present in the sample.

    o In the second step, streptavidin-coated paramagnetic microbeads are added. During a second 9-minute incubation, the biotinylated antibody attaches to the streptavidin-coated surface of the microbeads.

    o After the second incubation, the reaction mixture containing the immune complexes is transported into the measuring cell; the immune complexes are magnetically entrapped on the working electrode, and the unbound reagent and sample are washed away by ProCell.

    o In the ECL reaction, the conjugate is a ruthenium-based derivative and the chemiluminescent reaction is electrically stimulated to produce light. The amount of light produced is directly proportional to the amount of TSH in the sample.

    The concentration of the antigen or analyte is evaluated and calculated by means of a calibration curve using standards of known antigen concentration. Roche Diagnostics

    C-8 COBI-CD Version 1.0

  • cobas e 411 3 Test principlesTest principlesFigure C-3 Sandwich principle

    TPA ECL

    TPA

    SANDWICH PRINCIPLE

    FIRST REACTION

    Magnetic force and

    electrical potential

    Signal (light)

    Concentration

    SECOND REACTION

    LIGHT REACTION

    Antigen

    Biotinylated

    antibody

    Ruthenium-labeled

    antibody

    Streptavidin-coated

    microbead

    Tripropylamine

    Serum constituentsRoche Diagnostics

    COBI-CD Version 1.0 C-9

  • 3 Test principles cobas e 411Test principlesBridging principle

    The bridging principle is similar to the sandwich principle, except that the assay is designed to detect antibodies (for example, IgG, IgM, and IgA), not antigens. This is accomplished by including biotinylated and ruthenium-labeled antigens in the reagents for which the targeted antibody has affinity.

    e Refer to Figure C-4 on page C-11 for an illustration of the bridging principle.

    o In the first step, serum antibodies bind with the biotinylated and ruthenium-labeled antigens to form an immune complex.

    o The immune complex then reacts with streptavidin-coated microbeads through the action of the biotinylated antigen.

    o After the second incubation, the reaction mixture containing the immune complexes is transported into the measuring cell; the immune complexes are magnetically entrapped on the working electrode, and the unbound reagent and sample are washed away by ProCell.

    o In the ECL reaction, the conjugate is a ruthenium based derivative and the chemiluminescent reaction is electrically stimulated to produce light. The amount of light produced is directly proportional to the amount of analyte in the sample.

    The concentration of the antibody is evaluated and calculated by means of a calibration curve that was established using standards of known antibody concentrations. Roche Diagnostics

    C-10 COBI-CD Version 1.0

  • cobas e 411 3 Test principlesTest principlesFigure C-4 Bridging principle

    TPA

    TPA

    ECL

    BRIDGING PRINCIPLE

    FIRST REACTION

    Magnetic force and

    electrical potential

    Signal (light)

    Concentration

    SECOND REACTION

    LIGHT REACTION

    Biotinylated

    antigen Serum

    antibodies

    Tripropylamine

    Serum

    constituents

    Streptavidin-coated

    microbeadsRuthenium-

    labeled antigenRoche Diagnostics

    COBI-CD Version 1.0 C-11

  • 3 Test principles cobas e 411Test principlesRoche Diagnostics

    C-12 COBI-CD Version 1.0

  • cobas e 411 4 Reagent conceptTable of contentsReagent concept

    This chapter provides an overview of all types of reagents used on the cobas e 411 analyzer system. It describes the various reagent containers used, and also provides an overview of the system-related reagent management, explaining processes such as how the system registers new reagents, and how it monitors reagent consumption.

    Introduction ................................................................................................................. 15

    Data transfer media ....................................................................................................... 15

    Data transfer rules ......................................................................................................... 16

    Reagents for cobas e 411 analyzer tests ........................................................................ 16

    Diluents .................................................................................................................... 16

    System reagents ........................................................................................................ 17

    Calibrators and controls .......................................................................................... 17

    Reagent packs .......................................................................................................... 17

    Product labeling ............................................................................................................ 18

    Data links ...................................................................................................................... 19

    Calibration .................................................................................................................... 21

    Master calibration ......................................................................................................... 22

    Lot calibration ............................................................................................................... 23

    Reagent pack calibration ............................................................................................... 23

    Difference between lot and reagent calibration .......................................................... 24

    Calibration procedures ................................................................................................ 25

    Calibration stability ....................................................................................................... 26

    Calibration validation ................................................................................................... 26

    Calibration assessment .................................................................................................. 27

    Missing values .......................................................................................................... 27

    Monotony of curve (quantitative assays only) ....................................................... 27

    Slope (qualitative assays only) ................................................................................ 27

    Calibration factor (quantitative assays only) ......................................................... 28

    Minimum signal ...................................................................................................... 29

    Minimum/maximum signal (qualitative assays only) .......................................... 29

    Minimum difference (quantitative assays only) ................................................... 29

    Minimum acceptable difference (qualitative assays only) .................................... 29

    In this chapter Chapter 4Roche Diagnostics

    COBI-CD Version 1.0 C-13

  • cobas e 411 4 Reagent conceptTable of contentsDeviation of duplication ......................................................................................... 29

    System errors ........................................................................................................... 29

    Cal. .......................................................................................................................... 30

    Target (quantitative assays only) ............................................................................ 30

    Cutoff (qualitative assays only) ............................................................................... 30

    Borderline (qualitative assays only) ........................................................................ 30

    Calibration of quantitative assays ................................................................................. 30

    Rodbard function .................................................................................................... 31

    Linear calibration function ..................................................................................... 32

    Linear reciprocal calibration function ................................................................... 32

    Calibration of qualitative assays ................................................................................... 33

    Result calculation for qualitative assays ....................................................................... 33Roche Diagnostics

    COBI-CD Version 1.0 C-14

  • cobas e 411 4 Reagent conceptIntroductionIntroduction

    Elecsys reagent packs (cobas e packs) have a special 2D (two-dimensional) barcode. This allows fully automatic registration and management of reagent information. The advantage of this barcode is that manual entry or additional monitoring is not necessary. The ready-to-use liquid reagents are loaded into one of the 18 positions on the reagent rotor. Reagents are available for analysis after their barcodes are scanned.

    The handling of calibrators and Roche Diagnostics controls is similar to that of reagents. Some calibrators are supplied ready to use. Lyophilized controls and some calibrators must be prepared and transferred into the appropriate container. For quantitative assays, calibrator and control information is stored on 2D barcode cards. For qualitative assays, all information necessary for calibration is encoded on the barcoded labels.

    Data transfer media

    The following data sources are available for cobas e 411 analyzer applications:

    1. Barcodes on reagent packs (2D matrix barcode)

    o Reagent pack

    o Diluent reagent pack

    o Pretreatment reagent pack

    o BlankCell

    2. Barcodes on calibrator and controls vials (1D barcode)

    o Calibrator primary vial

    o Control primary vial

    3. cobas Link and system database (created during installation)

    o Assay

    o Calibrator and barcode card

    o Control and barcode card

    Barcodes on calibrator and controls vials carry information such as calibrator or control identification, level number, and lot ID.

    Reagent pack barcodes and downloaded data encode a lot more information, including application codes, calibration validation criteria, and expiry dates.

    e For further information of information encoded onto the barcode labels, see Data links on page C-19. Roche Diagnostics

    COBI-CD Version 1.0 C-15

  • 4 Reagent concept cobas e 411Data transfer rulesData transfer rules

    Calibrator data source Calibrator data are encoded on the 2D barcode of the reagent pack.

    o If the reagent pack was produced after the CalSet (calibrator set), the target values from the reagent pack have priority and are used for generating the calibration curve.

    o If the reagent pack was produced before the CalSet, data followed by target values from the calibrator card are used for generating the calibration curve.

    Control target values fromdifferent data sources

    Control target values are provided by different data sources.

    If a target value for a control for a specific PreciControl lot/reagent lot combination has been manually, the analyzer uses this value rather than the value read from the control barcode card or reagent barcode. When determining which control target value to use, the analyzer applies the following priority rules:

    Priority 1: Target values entered manually for a specific reagent lot

    Priority 2: Target values read from the reagent barcode

    Priority 3: Target vaules read from the control barcode card (or data downloaded from cobas Link - under development)

    If a new reagent lot or control lot is then placed on the analyzer, it uses the control values encoded in the barcodes for these new lots.

    Reagents for cobas e 411 analyzer tests

    This section describes all reagents necessary to run the cobas e 411 analyzer, and the reagents that are specific for each available test. The available tests are classified into different groups:

    o Thyroid

    o Fertility

    o Cardiac

    o Oncology

    o Infectious disease

    o Anemia

    o Diabetes

    o Bone

    o Othere For more information, see User access levels in the Software description section of the

    online Help.

    Diluents

    For most tests where dilution may be necessary, use Universal Diluent or MultiAssay as diluent. However, some testsincluding Anti-HAV, Estradiol, Progesterone, and NSE (neuron-specific enolase )require specific diluents. Roche Diagnostics

    C-16 COBI-CD Version 1.0

  • cobas e 411 4 Reagent conceptReagents for cobas e 411 analyzer testse For information on required diluents and recommended dilution factors, refer to the package inserts of the particular assay reagents.

    System reagents

    The cobas e 411 analyzer uses the following system reagents:

    Calibrators and controls

    There are specific calibrators for each test. As for quality controls, there are both controls covering multiple tests and controls that are specific for only a single test.

    e For information on required calibrators and controls, refer to the respective package insert.

    To get information about what calibrator and controls are currently needed for calibration or QC, print a Calibration/QC Load List from the software.

    Reagent packs

    The principal reagent container for the cobas e 411 is the cobas e pack.

    e For additional information on the packaging and barcoding of reagents, see the Product labeling section in the System overview chapter of the cobas e 411 analyzer Operators Manual.

    Reagent Use Bottle size

    ProCell o Conditioning of the electrodeso Transport of the assay reaction mixtureo Washing of the streptavidin-coated microbeadso Signal generation

    380 mL

    CleanCell o Cleaning of the tubing system and of the measuring cell after every measuremento Conditioning of the electrodes

    380 mL

    SysClean o Sodium hypochlorite solution used for cleaning of the measuring cells (every two weeks).o SysClean is not stored on the instrument.

    100 mL

    SysWash o Increasing rinsing efficiency between pipetting steps.o Avoiding reagent carryover. o Preventing bacterial growth.o SysWash is added to the distilled water container, with a dilution 1+100.

    500 mL

    Table C-1 System reagents for the cobas e 411 analyzer.Roche Diagnostics

    COBI-CD Version 1.0 C-17

  • 4 Reagent concept cobas e 411Product labelingA cobas e pack is a reagent pack that consists of three separate, capped reagent containers. The cobas e 411 can open and close these caps automatically. There is an individual reagent pack available for each test.

    Product labeling

    Each reagent pack is equipped with a barcode label. The barcode label contains reagent, control, and calibration information. The reagent barcode labels are in a unique format. The symbology uses portable data files (PDF) and is called PDF417. Traditional linear barcodes serve as a link to relevant information stored in a database. However a PDF417 is a two-dimensional, stacked barcode encoded to contain an entire data record. The large amount of data that can be encoded allows all instrument settings to be included, as well as the master calibration curve and additional information for the assay. From this master calibration curve and from the operator 2-point calibration, the analyzer derives the update of the master calibration curve.

    Every PDF417 symbol (barcode) contains two error detection codewords that are used like the check digit in linear barcode symbologies to detect decode errors and verify that all data has been read and decoded accurately. Additionally, PDF417 provides error correction in the event that portions of the symbol have been damaged, destroyed or are unreadable. (a)

    It is a combination of this error detection and error correction that ensures a reliable barcode. There should only be a small number of exceptional cases when barcodes are so badly damaged that the analyzer cannot read them. If the barcode cannot be read

    Figure C-5 Reagent pack

    (a) 1. Itkin S, Martell J. A PDF417 Primer: A Guide to Understanding Second Generation Barcodes and

    Portable Data Files. Bohemia, NY: Symbol Technologies, Inc; 1992:17-18.Roche Diagnostics

    C-18 COBI-CD Version 1.0

  • cobas e 411 4 Reagent conceptData linksand the reagent lot has previously been used by the analyzer, you can manually enter the 15-digit number found on the reagent barcode label into the software.

    Data links

    This following table illustrates the information that may be encoded on the barcode labels. Background colors are used to indicate calibration links that exist between information on separate barcodes.

    The exact information encoded onto the barcode labels is not detailed here, as this information is propriety to Roche Diagnostics. Items that are not directly linked are mapped within the cobas e 411 software or user interface. Roche Diagnostics

    COBI-CD Version 1.0 C-19

  • 4 Reagent concept cobas e 411Data linkse For further information about reagent checklists, see Reagents, calibrators, and controls in the Troubleshooting chapter of the cobas e 411 analyzer Operators Manual.

    Reagent Pack barcode

    Calibrator barcode card

    Calibrator vial barcode

    Control barcode card

    Control vial barcode

    Diluentbarcode

    Test number Test number Test number Test number

    (space for 28

    different tests (test

    number, target

    values and ranges

    in %)

    Test number

    Lot number:

    calibrator (space

    for 5 different

    calibrator control

    values)

    Lot number:

    calibrator

    Lot number:

    calibrator

    Lot number:

    reagent pack

    Lot number:

    reagent pack (space

    for 10 different

    reagent pack lots

    and calibrator

    target values)

    Reagent pack:

    bottle number

    Calibrator vial

    number

    Control vial

    number

    Test lot ID

    (only for ID-assays

    or assays where

    calibrators are

    within the reagent

    pack package)

    Test lot ID Test lot ID Test lot ID Test lot ID

    Calibrator levels Calibrator level

    number

    Control level

    number

    Lot number:

    control (space for

    10 different control

    lot target values)

    Lot number:

    control

    Lot number:

    control

    Rodbard

    parameters

    Calibration

    validation criteria

    Calibrator

    identification

    Control

    identification

    Control number Control number

    Expiry date Expiration date Expiration date Expiration date

    Table C-2 Calibration barcode informationRoche Diagnostics

    C-20 COBI-CD Version 1.0

  • cobas e 411 4 Reagent conceptCalibrationCalibration

    Calibration is required to determine the concentration of an unknown substance as accurately as possible. For this, a master calibration curve is generated at Roche Diagnostics during production of the reagent and is encoded in the 2D barcode of the appropriate reagent pack. This information is then transferred to the analyzer. At the customer site, the analyzer generates an update of the master curve by measuring two calibrators under routine laboratory conditions.

    The calibration curve produced from the barcoded master calibration and the measured calibration is specific to each reagent lot and, in some cases, to an individual reagent pack.