Comparison of ASTM D 6730 (DHA) and ASTM D 6839

Embed Size (px)

Citation preview

  • Comparison of ASTM Comparison of ASTM D 6730 (DHA) and ASTM D 6730 (DHA) and ASTM

    D 6839 (PIONA/D 6839 (PIONA/ ReformulyzerReformulyzer))

    25th Anniversary Seminar Series 2006

  • Comparison of D 6730 and D 6839 2

    Number of Possible Number of Possible HydrocarbonsHydrocarbons

    Paraffins

    Naphthenes

    Aromatics

    Olefins

    050

    100150200250300350400450500

    5 6 7 8 9 10 11C-number

    #

    i

    s

    o

    m

    e

    r

    s

  • Comparison of D 6730 and D 6839 3

    Individual ComponentsIndividual Components1961 Packed Columns C3 C9 122 peaks

    1968 Glass Capillary Naphthas 240 peaks

    1979 Fused Silica Naphthas 380 peaks

    1996 Fused Silica Gasoline 900+ peaks

  • Comparison of D 6730 and D 6839 4

    Hydrocarbon Type/Carbon #Hydrocarbon Type/Carbon #

    1965 TCEP Aromatics

    1968 Molsieve 13X Paraffins/Naphthenes

    1970 Molsieve 5A Iso/n-paraffins

    1983 Irreversible adsorption Olefins

    1985 Reversible adsorption Olefins

    1992 Reversible adsorption Oxygenates

  • Comparison of D 6730 and D 6839 5

    ChromatogramsChromatogramsDHA

    Reformulyzer

  • Comparison of D 6730 and D 6839 6

    Reformer Feed (HSR)Reformer Feed (HSR)Composition:n,i-Paraffins, Naphthenes, Aromatics (incl. Benzene)

    0

    5

    10

    15

    20

    25

    30

    35

    nPar af f ins iPar af f ins Naphthenes Ar omatics

    DHA

    REF

    %

  • Comparison of D 6730 and D 6839 7

    Light Straight RunLight Straight RunDHAThe DHA separation andidentification of LSR is highly reliable.

    ReformulyzerPNA, PiPNA and PIANO mode can be used to determine the hydrocarbon types in LSR.

    02468

    10121416

    LSR

    C

    a

    r

    b

    o

    n

    #

    DHAReformulyzer

    A light straight run is a saturatednaphtha with a boiling range of30-100 C consisting of C4 - C8hydrocarbons.

  • Comparison of D 6730 and D 6839 8

    Heavy Straight RunHeavy Straight RunDHADue to the high amount ofhigh boiling isomers, DHAdoes not provide a completeseparation on the capillarycolumn.

    ReformulyzerThe PNA, PiPNA and PIANOmode can very well be usedto determine thehydrocarbon types.

    02468

    10121416

    HSR

    C

    a

    r

    b

    o

    n

    #

    DHAReformulyzer

    A heavy straight run is a saturatednaphtha with a boiling range of60 - 185 C consisting of C5 - C12hydrocarbons, that includes a largenumber of isomers, especiallynaphthenes.

  • Comparison of D 6730 and D 6839 9

    ReformateReformateComposition:n,i-Paraffins, Naphthenes, Aromatics (incl. Benzene), Olefins

    0

    10

    20

    30

    40

    50

    60

    70

    nPar af f ins iPar af f ins Naphthenes Ar omatics

    DHAREF

    %

  • Comparison of D 6730 and D 6839 10

    ReformateReformateDHAThe DHA analysis provides a complete separation and identifies the components

    ReformulyzerThe PNA and PIANO modes provide an accurate hydrocarbon typeclassification per carbonnumber.

    Reformate is a converted naphthathat consists of paraffins, somenapthenes, aromatics and smallamounts of olefins. The carbonnumber range of reformate consistsof C4 - C10 hydrocarbons andsome naphthalenes. The boilingrange is 30 - 185C.

    02468

    10121416

    Reformate

    C

    a

    r

    b

    o

    n

    #

    DHA

    Reformulyzer

  • Comparison of D 6730 and D 6839 11

    FCC FCC NaphthaNaphthaComposition:Paraffins, Naphthenes, Aromatics (incl. Benzene), Olefins

    0

    10

    2 0

    3 0

    4 0

    5 0

    Par af f i ns Ol ef i ns Naphthenes Ar omat i cs

    DHA

    REF

    %

  • Comparison of D 6730 and D 6839 12

    FCC NaphthaFCC NaphthaDHAThe capillary column is notable to separate all individual components; theidentification is not reliablefor components abovetoluene.

    ReformulyzerThe PHONA mode provides correct carbon number/ hydrocarbon type information.

    02468

    10121416

    FCC Naphtha

    C

    a

    r

    b

    o

    n

    #

    DHAReformulyzer

    The FCC naphtha is unsaturatedand consists of paraffins,naphthenes, olefins and aromaticswith a carbon number range ofC4 -C12. The boiling range is30 - 200 C.

  • Comparison of D 6730 and D 6839 13

    Gasoline + EthanolGasoline + EthanolBlended from:Reformate, Isomerate, Alkylate, FCC, Oxygenate, Reformer Feed

    0

    10

    2 0

    3 0

    4 0

    5 0

    6 0

    Par af f in s Olef in s Napht hen es Ar omat ics Oxygen at es

    DHA

    REF%

  • Comparison of D 6730 and D 6839 14

    FinishedFinished GasolineGasolineDHA AnalysisIf the gasoline is blendedfrom saturates and one ortwo oxygenates, DHAprovides sufficient data.If the blend contains olefins, the DHA data is only reliable up to toluene.

    ReformulyzerASTM D 6893 provides therequired data for bothenvironmental regulationsand process control.

    0 2 4 6 8

    10 12 14 16

    GasolineC

    arbon #

    DHAReformulyzer

    Finished gasoline is a hydrocarbonblend of paraffins, naphthenes,olefins, aromatics and possiblysome oxygenates. The boilingrange is 30 -185C and the carbonnumber range is C4 - C10.

  • Comparison of D 6730 and D 6839 15

    SampleSample ScopeScope

    DHA Multi-DimensionalStraight Run + +FCC -- +

    Reformate + +

    Isomerate + +

    Alkylate + +

    Oxygenates + +

    Gasoline -- +

  • Comparison of D 6730 and D 6839 16

    ApplicationApplication ComparisonComparisonDHA Multi-Dimensional

    Standard Method

    D 6729, D 6730, D 6733, D 5134

    D 6839, EN 14517, D 6293

    Analysis Time 120 150 min28 min on fast DHA

    75 min for Winterspec mode

    Columns 1 or 2 7 (including traps)

    User Group Not planned Yes

    Crude Oil Yes, light end in crude analysis (combi design)

    No, FBP is 275C/527F

  • Comparison of D 6730 and D 6839 17

    ApplicationApplication ComparisonComparisonDHA Multi-Dimensional

    Individual Components Benzene, Toluene, Oxygenates

    Group-type Group-type

    RON RON

    Mol.Weight Mol.Weight

    Specific Gravity Density

    TBP C:H:O ratio Cal.Value

  • Comparison of D 6730 and D 6839 18

    ConclusionsConclusions

    Both DHA & Reformulyzer are suitable foranalyzing refinery streams

    DHA is recommended when mainly detailedinformation is needed at component level

    Reformulyzer is required when samples containing higher amounts of Olefins and/orNaphthenes above C7 8 are present in the sample

  • Comparison of D 6730 and D 6839 19

    QuestionsQuestions

    ?

    Comparison of ASTM D 6730 (DHA) and ASTM D 6839 (PIONA/ Reformulyzer)25th Anniversary Seminar Series 2006Number of Possible HydrocarbonsChromatogramsReformer Feed (HSR)Light Straight RunHeavy Straight RunReformateReformateFCC NaphthaFCC NaphthaGasoline + EthanolFinished GasolineSample ScopeApplication ComparisonApplication ComparisonConclusionsQuestions