FFDM -FCRm QC Requirements-What You REALLY Need to Know
M elissaC. Martin , M.S.,FACR, FAAPM, FACMP
AAPM Annual Meeting July 28,2008
FDA Appr oval for Mammography
• Fuji FCRmSystemapprovedforMammographyon July 10,2006by theFDA asa FFDM DigitalMammographySystem
CR TechnologyCR Technology
ProtectiveLayerProtective LayerPhoto-stimuablephosphor layerPhoto-stimuablephosphor layer
SupportSupport
J.Ed Barnes,Ph.D.- MTMI
FujiFuji MammoMammo CRCR
•Pixel size 50 micron
•Use dual sided reading to DQE
•Imaging Plate has thicker phosphor layer
•Results in 40-50% increase in NEQ (DQE)
•Post processing enhances image
(e.g. filtering to recognize and enhancecalcifications)
J. Ed Barnes,Ph.D.- MTMI
2
FCMFCM
J.Ed Barnes,Ph.D. - MTMI
1mm
1mm
EnlargedImages
Standard System (100 micron pixel)
FCRm (50 micron pixel)Dual side scanJ. Ed Barnes, Ph.D.- MTMI
Fuji SystemComponents
ClearView -CSm
Breast Imagi ng Review Statio n
ClearView -1mFlash Plus IIPm
J.Ed Barnes, Ph.D.- MTMI
Fuji ClearviewCSm Reader
• Four cassettestacker
• Processes20 screeningexams/hour(or 100CR plates/hour)
• Multi -objectiveFrequencyProcessing(MFP)(enhancesdense/peripheral tissue)
• PatternEnhancement ProcessingforMammography (PEM)(enhancescalcifications)
J.Ed Barnes, Ph.D.,- MTMI
3
OtherCR MammographyVenders- World Wide
•Agfa
•Kodak
•Konica
•Canon
J. Ed Barnes,Ph.D.- MTMI
KodakMammography CR•Singlesidedplate•Thinnerphosphorlayer (improvesresolution,reducesDQE)•Phosphorgrain smallerandhomogenous•BaFBr:Eu phospor•48.5micron pixel•Readout speed reduced
•Twice asslow•IncreasesSNR•Useconventional reader(modified)
J.Ed Barnes,Ph.D.- MTMI
Agfa CR Mammography
•Single sidedplate, flexible
•BaSrFBrI:Eu phosphor
•Smaller grain powdered phosphor
•50 micron pixel
•Scanslower thanconventional CR
•Wider light collection guide
•Thinner plateto increaseresolution
J.Ed Barnes, Ph.D.- MTMI
KonicaCR Mammography
•Single sidedplate, rigid
•BaFI:Eu phosphor
•43.75micronpixel
•Utilize “PhasedContrastCR”
•Exposebreaston a mag platformto a 14x17cassette
•Demagnify to a 24x30cm image(air grid)
J. Ed Barnes,Ph.D. - MTMI
4
CR MammographySystems
for AGFA, FUJI,KODAK, KONICA
“ All measurements wereequivalenttoor exceeded theperformanceof a typicalmodernfilm/screensystem”
FromCentre for Evidence-basedPurchasingReport 06047, October2006
J.Ed Barnes, Ph.D.- MTMI
PSPRadiography(CR)
• Currently the majortechnologyavailable for largefield-of-view digital imaging
• Basedupon theprinciplesof photostimulatedluminescence;20+ yearsof experience
• Operationemulatesthescreen-film paradigm in useandhandling.. (flexible but laborintensive)
• Manufacturingtrends:– Smaller,faster, lessexpensive
J. Seibert - UC Davis
PSPDetector
• PhotostimulableStoragePhosphor (PSP)
Phosphor PlatePhosphor PlatePhosphor Plate Cassette HolderCassette HolderCassette Holder
• Coating thickness:– Standardresolution: ~100µm BaFBr
– High resolution: ~50-70 µm BaFBr
• Dual-sideread; structured phosphor – CsBr
LaserLaserstimulationstimulation
2 eV2 eVee
Incident xIncident x--raysrays
8 eV8 eV
F center trapF center trapee--PSLPSL
3 eV3 eVeeeeee
J.Seibert, Ph.D. - UC Davis
Computed RadiographyComputed Radiography ““readerreader””
Information panelInformation panel
PlatePlatestackerstacker
FujiFuji
J. Seibert- UC Davis
5
CRCRReaderReader
CR QCCR QCWorkstationWorkstation
Display / ArchiveDisplay / Archive
Laser film printerLaser film printer
DICOM / PACSDICOM / PACS
Image AcquisitionImage Acquisition
CRCR ““systemsystem””: more than the IP: more than the IP’’s and the reader!!s and the reader!!
J.Seibert-UC Davis
Digital ImageNetworking
• Modality Worklist Input (from RIS via HL-7)
• TechnologistQC Workstation– Imagemanipulation processingreconciliation
– ReconciliationandimageQA
• PACS andDICOM– Digital ImagingCommunicationsin Medicine
– Providesstandardfor modality interfaces, storage/retrieval,andprint
• DICOM image output
J. Seibert- UC Davis
Photostimulated Luminescence
Incident LaserIncident LaserBeamBeam
PMTPMT
Protective LayerProtective Layer
Phosphor LayerPhosphor Layer
Base SupportBase Support
LightLightScatteringScattering
Laser Light SpreadLaser Light Spread
PhotostimulatedPhotostimulatedLuminescenceLuminescence
"Effective" readout diameter"Effective" readout diameter
ExposedExposedImagingImaging
PlatePlate
Light guideLight guidePSLPSLSignalSignal
J.Seibert - UC Davis
CR PointCR Point--scan readoutscan readout
PMTPMT
PolygonalPolygonalMirrorMirror
LaserLaserSourceSource
Light channeling guideLight channeling guide
Output SignalOutput Signal
ReferenceReferencedetectordetector
CylindricalCylindricalmirrormirrorff--θθ
lenslens
ADCADC
Laser beam:Laser beam:Scan directionScan direction
Plate translation:Plate translation:SubSub--scan directionscan direction
To imageTo imageprocessorprocessor
ADCADC
x= 1279x= 1279y= 1333y= 1333z= 500z= 500
ERASURE → ReuseJ. Seibert- UC Davis
6
Scan DirectionScan Direction
SubSub--scan Directionscan Direction
Laser beam deflectionLaser beam deflection
Plate translationPlate translation
Typical CR resolution:35 x 43 cm -- 2.5 lp/mm (200 µm)24 x 30 cm -- 3.3 lp/mm (150 µm)18 x 24 cm -- 5.0 lp/mm (100 µm)
Screen/film resolution:7-10 lp/mm (80 µm - 25 µm)
J.Seibert - UC Davis
Phosphor Plate CyclePhosphor Plate Cycle
PSPPSP
Base supportBase support
reusereuse
plate erasure:plate erasure:remove residual signalremove residual signal
light erasurelight erasure
plate exposure:plate exposure:create latent imagecreate latent image
xx--ray exposureray exposure
plate readout:plate readout:extract latent imageextract latent image
laser beam scanlaser beam scan
J. Seibert- UC Davis
CR Innovations
• Dual sidereadout
• Mammographyapplications
• High-speedline scan systems(<10sec)
• StructuredPSP
• Low costtable-top CR readers
J.Seibert - UC Davis
MirrorMirror
To imageTo imageprocessorprocessor
LightErasure
ADC
Light channeling guideLight channeling guide
Cylindrical mirrorCylindrical mirror
LaserLaserSourceSource
ReferenceReferencedetectordetector ff--θθ
lenslens
CRCRcassettecassette
ErasureErasureStageStageTransportTransport
PMTPMTOutput SignalOutput Signal
To imageTo imageprocessorprocessor
ADC
J. Seibert- UC Davis
7
CR Mammography
• Dual-side readout capabilities (increasedDQE)• Fine laser beam spotsize,50µm (10 lp/mm)• Mammography-specificimageprocessing
CR cassette
ConventionalMammo system CR reader
Two cassette sizes
QC workstation
18x2424x30
Scanning laser beam
50 µmspot
2 light guides
Transparent base
Low attenuationcassette, special IP
Fuji CRmSystem
CR PlateReader
CRPlates
SectionF = MedicalPhysicist’s QC Tests
Medical PhysicsAnnual Tests• MammographicUnit AssemblyEvaluation• Collimation Assessment• S-ValueConfirmation• SystemResolution• kVp AccuracyandReproducibility• HVL Measurement• BreastEntranceExposureand GlandularAverage
Dose• RadiationOutputRate• PhantomImageQuality Evaluation
- PrintedImages- Softcopy
8
MedicalPhysicsAnnual Tests• AEC SystemPerformance• SystemArtifact Evaluation• CR ReaderScannerPerformance• Dynamic Range• Primary Erasure
- Additive andMultiplicative Lag Effects• Inter PlateConsistency• Viewing andViewingConditions• ReviewWorkstationTests• Printer Tests• Technologist’s QC
1. Mammographic Unit Assemb ly Evaluatio nPass/Fail/ NA
Free-standing unit is mechanically stable PassAll moving parts move smoothly, without obstructions to motion PassAll locks and detents work properly PassImage receptor holder assembly is free from vibrations PassImage receptor slides smoothly into holder assembly PassImage receptor is held securely by assembly in any orientation PassCompressed breast thickness scale is accurate to +/-0.5 cm, reproducible to +/-2 mm PassPatient or operator is not exposed to sharp or rough edges, or other hazards PassOperator technique control charts are posted PassOperator protected during exposure by adequate radiation shielding PassAll indicator lights working properly PassAuto decompression can be overridden to maintain compression (status displayed) PassManual emergency compression release can be activated in the event of a power failure Pass
Comments: The following four items satisfy MQSA Final Regulations:1.) Upon power interruption, the gantry remains stationary and stable.2.) Under compression, the paddle deflection was: 18x24 cm: 3 mm; 24x30 cm: 4 mm.3.) The collimator light illuminance was 420 lux.4.) The compression paddle indicates the location of the AEC sensor. Compression is set at 31pounds.The technique chart above reflects the results of our measurements and should be posted on the unit.The unit will be used clinically in the Au to-Filter mode.
FCRm Collimati on
Testfor Routinely Used:
Collimator,CompressionPaddle,BuckyCombinations
Target Materials
Fuji CRm System
MeasureExposureRate at Surface of Breast Support
S-Value Confir mation
9
FCRmS ValueConfirmation
• Makeexposureat 25 kVp with Mo/Moto giveexposure> 20 mR
• Exposecassetteat this technique- wait 10 minutesandprocess
• RecordS value
• Calculatethe“ corrected S value”
Action Limit:Corrected S value should not exceed
the range of 120 +/- 20% (96 to 144)
1 - S Value Confirmati on
Date:Exposure Unit ID: Lorad M-IVImage Reader ID: Fuji ClearView CSm
Is the 95%uji ClearView CS
kVp: 25mAs: 3.0
Exposure 1 (mR): 37.0Exposure 2 (mR): 37.1Exposure 3 (mR): 37.0
Average Exposure: 37.0 Exposure readings within 5%? Yes
IP Exposure Time 1: 2.0 IP Read Time 1: 2 minIP Exposure Time 2: 2.0 IP Read Time 2: 2 minIP Exposure Time 3: 2.0 IP Read Time 3: 2 min
S Value 1: 66.0S Value 2: 68.0S Value 3: 65.0
Average S Value: 66.3
Corrected S Value: 122.8
Corrected S Value Pass/Fail? PASSCorrective Action Needed? NO
Descriptio n of Correctiv e Action:Cor rective Act io n Taken? (Yes, No?)
Date of Corrective Action:Resu lts of New Test: (Pass, Fail?)
Comments:
November 21, 2007
FCRmSystemResolution
• Up to 10 lp/mm Test Pattern
• 4 cm Acrylic Phantom
• Parallel andPerpendicular to Tubewith
3-5 degreeangle
Systemlimi ting resolutionmustbe8 lp/mm
+/- 2 lp/mm (6 to 10 lp/mm)
2. Evaluation of System Resol uti on
X-ray Tube Manufacturer: Varian (1/02) Model #: B-115/M-113R
Visible black points: No NoAnode material Mo Mo
Nominal kVp setting 27 25
Nominal mA setting 30 100
Density control setting 0 0mAs 57 124S Value (Digital) 58 63Magnification factor 1.80 CONTACTLimitin g resol ution bars parallel to A-C axis 10 7in line -pairs per mm bars perpendicular to A-C axis 10 7
Action Limit: For Fuji FCRm, if the limi ting resolu tion with the bars paralle lto the anode-cathode axis is < 8 +/2 line-pairs/m m orperpendicul ar to the anode -cathode axis,then a more detailed inve stiga tion of the reasonshould be made and corre ct ive action sho uld betaken.
Fuji CRm
10
FCRmDensityControl Function
4 cm Acrylic, Accreditation PhantomTechnique
Repeatat -2, -1, 0, 1, 2, etc. densitysettings
Record mAs
mAs Changesshouldbe5 to 15%perstep
Densi ty Control Function :
Imaging mode: Film/Screen Focal spot: 0.3 mm MOmA: kVp:Phantom thickness: 4 cm BR-12 Cassette ID: QC
Relati ve to NormalDensity Selector % mAs Density
Setting Change Change5 11.0% Pass4 11.5% Pass3 7.0% Pass2 11.2% Pass1 12.0% Pass0 0% Pass-1 14.4% Pass-2 12.8% Pass-3 9.8% Pass-4 12.2% Pass-5 12.6% Pass
Action Lim it: Each step should result in a 5 to 15% change in mAs.If not, seek service.
75.4
125.0109.098.386.3
201.0187.0166.0146.0
mAs255.0227.0
FCRmReproducibility andImageModeTracking
• 4 cm Acrylic with Clinical Technique
• PositionIon Chamberin Beam
• RecordmAs andExposure
• Repeat3 Times
• Repeatin eachMode(small,large,magandno grid)
Action Limits:
Coefficient of Variationfor Exposuremustnot exceed0.05
No Significant Differencein Exposure betweensmall andlargebucky whenusingsimilar grid
4. Automat ic Exposur e Contro l (AEC) System Perfo rmanceAEC position: #3 Density control: 0Small cassette ID: QC Large cassette ID: 24 x 30
Performance Capabil ity:
CNR PER OBJECT THICKNESSImage Mode: Auto TimeFocal spot: LargeDensity Control 0
Phantom thickness Target/Filter kVpmAs
(Auto) CNR
CNR(relative to
4cm PMMA)Acceptable
Level2 cm Mo/Mo 25 26.5 16.3 1.11 >100%4 cm Mo/Mo 25 131 14.7 1.00 100%6 cm Mo/Rh 30 218 12.5 0.85 >75%8 cm
PASS
Image Mode Trackin gmA: 100Phantom thickness: 4 cm BR-12Anode/Filter: Mo/Mo
Image Mode AEC MODE Target Filter kVp mAs Commentssmall grid Auto Filter Mo Mo 26 146.0 HTC Gridlarge grid Auto Filter Mo Mo 25 113.0 HTC Grid
Magnification/no grid Auto kV Mo Mo 27 57.7 No Grid
11
Fuji CRm System
Dynamic RangeTest
FCRmDynamicRange
• To confirm dynamicrangeof readerandplate
• Use2 and 4 cm acrylic
• Techniquefor 6 cm
• View imageto determineif 3 regionsarevisible anddiscernable
Fuji CRmSystem- CNR Test
PASS
>75%91.6%13.765280.0282.030.0RhMo6 cm
100%100.0%15.030130.0135.025.0MoMo4 cm
>110%114.4%17.19728.027.225.0MoMo2 cm
Acceptablelevel
CNR(relativ e to 4cm
PMMA)CNRmAs
(Manua l)mAs
(Auto )kVpFilterTargetPhantom Thickness
Auto Tim eAEC Mode:
0Density Control:
3Foca l Spot:
Auto Tim eImage Mode:
CNR PER OBJECT THICKNESS
FCRmReader ScannerPerformance
• Establish thatIP Readerandoptics
donot exhibit scanor print ji tter
• Non-Grid Exposure
• Positionrulers in “T” formation
• ExposeandProcess
Action Limit :
Imagemusthavesmoothbordersfree from jaggededges or defects
12
3. CR Reader Scanner Performan ce Test
Date:
Parallel to Chest Wall
Workstation MonitorCorrective Action Needed?Laser Film Printer*Corrective Action Needed?Corrective Action Description?Corrective Action Taken?Date of Corrective ActionResults of New Test
Perpendicular to Chest Wall
Workstation MonitorCorrective Action Needed?Laser Film Printer*Corrective Action Needed?Corrective Action Description?Corrective Action Taken?Date of Corrective ActionResults of New Test
*If hard copy is used for final interpretation.
November 21, 2007
Image Reader ID: Fuji ClearView CSm # 5722213(Required for Annual Surveys; not required for Mammography Equipment Evaluations of new units. However,
Pass
NoNA
NA
NANANA
NA
Acceptance Limits
Acceptance Limits
NANA
NoPass
PassNo
PassNo
FCRmPrimary Erasure
• To assesstheeraseperformanceof reader andplates
Additive: Shoot phantom, process,wait 1 minute,reprocess,changeS value to 10X original, inspectfor visibility of phantomimage
Multiplicative: Shoot& processphantom,shoot 4cm acrylic on samecassette, processand inspectfor visibility of phantomimage
LSG BH: F
Target Filter kVp mAsMo Mo 25 139 Printer:
Accep tabl e?CORRECTIVE ACTIVE NEEDED?CORRECTIVE ACTIVE DESCRIPTCORRECTIVE ACTIVE TAKEN?CORRECTIVE DATE:RESULTS OF NEW TEST:
Target Filter kVp mAsMo Mo 25 139Mo Mo 25 132
Accep tabl e?CORRECTIVE ACTIVE NEEDED?CORRECTIVE ACTIVE DESCRIPTCORRECTIVE ACTIVE TAKEN?CORRECTIVE DATE:RESULTS OF NEW TEST:
The mAs noted on the generator read-out must not change by more than ±15% from the previous test.
Percent Change in mAs:Accep tabl e?
A unif orm 4 cm sheet imageACR MAPP
NA
Multiplicative Lag testExposure con ditio ns
NA
PassNoNANA
ACR MAPP
52
520
Artif act Evaluation
S valu e of the ACR Phanto m Image:
S valu e applied to non-expos ure image (10Times Phantom Image S Value)
Cassette / IP Numb er A30389595C
Additive Lag testExposure con ditio ns
Artif act EvaluationPassNoNA
Pass
NANANA
5.2%
FCRmInter-PlateConsistency
• To confirm x-ray absorption & SNRconsistency
• Exposureplateswith 4 cm acrylic usingclinicaltechnique
• RecordmAs (mustbewithin 10%)
• CalculateSNR
(mustbewithin 15%)
13
Yes69.662Yes133A777, #6
Yes69.521Yes132A111, #5
Yes70.166Yes132A222, #4
Yes70.152Yes132A123, #1
Yes70.952Yes132A234, #2
Yes70.946Yes132A456, #3
Acceptabl eSNRAccept ablemAsCassett e ID
Group: SmallCasset tes
25MoMoAutoTimeExposu re conditions
kVpFil terTargetAEC-mode
Lorad M-IIIExposure Unite ID:
SNR limit is plus or minus15 % from mean.
mAs limit is plus or minus10 % from mean.
Inter PlateConsistency
FCRmCNR perObjectThickness• CNR usingclinical techiquefor 2 cm
• Repeatfor 4 and6 cm
Action Limits:CNR of 2 cm relative to 4 cm must be > 100%
CNR of 6 cm relative to 4 cm must be > 75%
Image Mode:Focal Spot:Density Control:AEC Mode:
Phantom Thickness Target Filt er kVpmAs
(Auto)mAs
(Manual)CNR
CNR(relat ive to
4cm PMMA)
Acceptablelevel
2 cm Mo Mo 25.0 26.2 26.0 13.270 100.8% >100%4 cm Mo Mo 25.0 134.0 135.0 13.165 100.0% 100%6 cm Mo Rh 30.0 227.0 225.0 10.890 82.7% >75%
PASS
CNR PER OBJECT THICKNESSAuto Time
L0
Auto Time
FCRmContrastto NoiseRatio Test(CNR)
• To examineconsistencyof CNR ratiomeasuredover time
• Use4 cm Acrylic and0.2mm Aluminum
• ManualTechnique(Mo/Mo, 26 kVp, 125 mAs)
• CalculateCNR using software
Action Limit:+/- 20% of baseline
14
FCRmViewing Conditions
• HardcopyInterpretation: Sameconditionsasto beusedfor screen-film mammography images.Filmmaskingdevicesandhot lights mustbeavailableto theradiologist.
• Softcopy Interpretations: Follow themonitormanufacturer’s QC procedures.If no monitormanufacturer’s QC proceduresavailable, followproceduresoutlined in Fuji QC Manual.
Fuji CRm - Monitor Tests
Site: Med. Img Ctr of So Cal Date of Survey: 4/19/2007
Room #: Reading Room Date of Installation: Apr-07
DOME©E5
Target Value = 500 cd/sq-m
499.2
Serial Numbers
Luminance Value Setting (cd/sq-m)
Left Monitor Right Monitor
703PNKN00028
500.88
703PNKN00011
Is the CXtra icon present in thetaskbar and shown as a green
check mark?
Manual Conformance report forboth displays attached?
Is DICOM calibration Graph forboth displays attached?
NoYes
Yes
Yes
No
No
LaserPrinter-mustbeabletoprint imagesonmammographylaserfilm
FCRmPrinter AnnualQC
• Follow theprinter manufacturer’s QC programfortestprocedures,frequency,performanceandcorrectiveaction.
• If noprinter manufacturer’s QC program,followtheQC programoutlinedin Fuji QC Manual.
• Testsmustbeperformedinitially prior tointerpretinghardcopymammogramsandaftermajor repairsandthenannually.
15
OD Requirements forOD Requirements forHiHi--Resolution Laser ImagersResolution Laser Imagers
Dmax > 3.5 OD
Mid-density > 1.5 OD
10. Breast Entranc e Exposure, AEC Repro duc ibilit y, Average Gland uDose, and Radiat ion Output Rate
Imaging mode: SID (cm):Screen type: Source-detector distance (cm):Film type: Source-bucky distance (cm):Cassette size (cm): Dosimeter used:Field Restriction: Energy correction factor:
Breast thickness (cm)PhantomNominal kVp settingAnode materialFilterAEC modeAEC density control settingMeasured HVL (mm Al)
Breast Entrance Exposur e and AEC Reprodu cibility:
R mAs R mAs R mAs R mAs
Exposure #1 1.581 147.0 1.480 121.0 1.231 89.1 0.705 75.2Exposure #2 1.498 139.0Exposure #3 1.521 142.0Exposure #4 1.501 140.0
Mean values 1.525 142.0 1.480 121.0 1.231 89.1 0.705 75.2Standard deviations (SD) 0.039 3.56
0.025 0.025
Action l imi t: If coe ffic ient of variation for either R or mAs exceeds 0.05, seek service.
Aver age Glandular Dose:
Inv Sq corrected skin exp 1.525 1.480 1.231 0.705
153 159 164 217
233 235 202 153
Action l imi t: If average glandular dose exceeds 300 mrad (3 mGy) for 4.2 cm effec tiv e breas tthickness, seek service or technique adjust ment. Corre cti ve action must be takenbefore furth er examina tions are performed if the test results fail MQSA regulations.
Radiation Output Rate:
3 sec,kVp Anode Filter
SID(cm)
Exp(mR)
mAsTime(sec)
Rate(mR/s)
Kerma(mGy/s)
4.5 cm above 28 Mo Mo 65 4740 300 3 1580 13.8breast support 28 Mo Mo
Air Kerma (mGy/sec) = Exp Rate (mR/s) x 0.00873 mGy/mR
Action l imi t: If output rate is less tha n 800 mR/sec (7.0 mGy air kerma/sec), seek service.
Coeffi cient ofvariati on (CV)
Dose conversion factorfrom Tables 1-3 (mrad/R)Comput ed averagegland ular dose (mrad)
MoAuto Filter
00.3
1.00
DigitalFuji CR
CR18 x 24
656064
Keithley TriadCollimators
RhAuto Time
00.43
4.2RMI 156-15914
28Mo
MoAuto Time
00.32
4.2RMI 156-15914
27Mo
MoAuto Time
00.31
25Mo
4.2RMI 156-15914
26Mo
RMI 156-159144.2
PhantomImageQuality Check
PossibleItems
( 6 )
( 5 )
( 5 )
ProperScoringTechniquesof theACR MammographyAccreditationPhantom
16
Fibers
• Full PointScore
- Full lengthof fiber visible (+/- 1 mm) -usefirst fiber as a referencefor length
- Correct location
- Correct orientation
Fibers
• Half Point Score
- Not all, but > 50%visible
- Correct location
- Correct orientation
- Stop if orientation or bordersnot visible
Fibers- Artifacts
• In WrongLocation of PhantomInsert
– Fiber-like
– At leastasapparent aslastfiber scored
– Diameter andlength> lastscoredfiber
Fibers- Artifacts -Deductions
• Deduct1.0- lastfiber is scoredas wholeandartifact is > 75%of full fiber length
• Deduct0.5- lastfiber is scoredas 1/2 andartifact is > 50%of full fiber length
• Only deductfrom last fiber scored
Action Limits: Minimum of 4 fibersfor all unitsexceptHologic andSiemensFFDM units.Minimumof 5 fibers for Hologic andSiemensFFDM units
17
Speck Groups
Full PointScore
- > 4 specksvisible
- Correctlocations
Half Point Score
- 2 or 3 specks visible
- Correctlocations
SpeckGroups- Artifacts
• Noiseor speck-like
• As apparentaslast specks scored(i.e. - atleastasbright and > diameter)
• Do not subtract bright artifactscausedbydustor emulsion pickoff
Speck Groups- Deductions
• Deduct1-for-1 from scoredspecks
• Only deductfrom thespecks(s) in thelastgroup scored (with eithera 1.0or 0.5score)
SpeckGroups
Action Limit:
Minimumscoreof 3 speck groupsfor all unitsexcept Hologic andSiemensFFDM units
Minimumscoreof 4 speck groupsfor Hologic andSiemensFFDM units
18
MassesScore
• Full Point
- Density differencevisible
- Generally circular(> 75% circumference)
- Correct location
MassesScore
• Half Point
- Density differencevisible
- Not generally circular (< 75%circumference)
- Correctlocation
MassesScore- Arti facts
In Wronglocationof PhantomInsert
- Mass- like
- As apparentaslast mass scored
Deduction
Only deduct from last massscored
MassesScores
Action Limit:
Minimumscoreof 3 massesfor all units exceptHologic andSiemensFFDM units
Minimumscoreof 4 massesfor Hologic andSiemensFFDM units
19
Acceptable PhantomImage
WW = 200 (suggestion)
Description: AcceptableACR Phantom image
Minimum AcceptablePhantom Image Scorefor the Fuji FCRm systemis:4 fibers, 3 speck groups,3 masses
ImageQuality Values
• ThePhantomImageQuality on theFujiCRmsystemwil l reflect very closelywhatwasachievedpreviouslyon thefilm/screensystemin use- expectno significantchangein quality.
Site Name Report DateAddress Survey DateMedic al Physicist's Name Signa tur eX-Ray Unit Manuf acturer ModelDate of Unit Installation Room IDFFDM Image Receptor Mfr FFDM ModelFFDM QC Manual Version: (check one; must use version applicable to system tested; contact mfr if questions)
Accessory Equipment:
Review Workstation*
Laser Film Printer*
Survey Type: Mammo Eqpt Evaluation of new unit (include MQSA Rqmts for Mammo Eqpt checklist) Annual Survey
PASS/FAIL1. S Value Confirmation ( < 120 + 20% [96 < corrected S value < 144])
2. System Resolu tion (8 lp/mm + 2 lp/mm in both directions)
3. CR Reader Scanner Performance4. Imaging Plate (IP) Fog5. Mamm ography Unit Assembly Evaluation6. Collimation Assessmen t Test date if different from above:
Chest wall edge of X-ray field extends to edge of IRDeviation between X-ray field and light field <2% of SIDX-ray field does not extend beyond any side of the IR by more than 2% of SIDPaddle chest wall edge not beyond IR by more than 1% of SID or appear on the image
7. Automatic Exposure Contr ol (AEC) System Perfo rmance Ass essmentAEC density control function meets Fuji performance criteriaReproducibility (CV) for either exposure or mAs is <0.05Image mode tracking meets Fuji performance criteriaCNR per object thickness meets Fuji performance criteria
8. System Artif act Evaluation9. Phantom Image Quality Evaluation Fibers Specks Masses (at least 4 fibers, 3 speck groups & 3 masses)
Phantom IQ (printed images)Phantom IQ (softcopy)Other tests meet Fuji performance criteria (mAs, OD & DD for hardcopy, S value for soft copy)
10. Dynam ic Range11. Primar y Erasure12. Inter -Plate Consistency (variation of mAs within + 10%; SNR within + 15%)
13. kVp Accur acy and Repro ducibil ity Test date if different from above:
Measured average kVp within +5% of indicated kVpkVp coefficient of variation <0.02
14. Dose (average glandular dose for average breast is < 3 mGy [300 mrad]) mrad15. Beam Qualit y Assessment & HVL Measureme nt Test date if different from above:
HVL >kVp/100 mm Al16. Radiation Output Test date if different from above:
Radiation output rate is >800 mR/sec (7.0 mGy/sec) at 28 kVp with Mo/Mo17. Viewing and Viewing Condition s Test date if different from above:
Mammographic viewbox luminance >3000 cd/m^2 (nit)Room illuminance (viewbox surface & seen by observer) <50 lux
18. Review Work station (RWS) Test s** (for all RWS, even if located offsite)
Overall Results ("Pass" means all tests pass; indicate "Fail" if any test fails)
Fuji FCRm
MEDICAL PHYSICIST'S MAMMOGRAPHY QC TEST SUMMARYFull -Field Digital – Fuji
Model QC Manual Version
Medical Phys icist's QC Tests
Manufacturer Location
*FDA recommends that only monitors and printers specifically cleared for FFDM use by FDA’s Office of Device Evaluation (ODE) be used. SeeFDA's Policy Guidance Help System www.fda.gov/CDRH/MAMMOGRAPHY/robohelp/START.HTM
*** YOUR MEDICAL PHYSICIST MUST SUMMARIZE HIS/HER RESULTS ON THIS FORM ***
**FDA requires that all RWS comply with a QC program that is subst antially the same as that recommended by the image receptormanufacture r . If the RWS is FDA-approved, the RWS's QC manual is considered to be "substantially the same" and you may follow it. (Checkwith the RWS manufacturer for their FDA clearance status and QC manual.) If the RWS is not FDA-approved for FFDM, you must follow theQC manual provided by the image receptor manufacturer. (Check with the image receptor manufacturer for their required tests.)
1st edition, 07.2006 897N0602, 2006
On-site Off-site
On-site Off-site
OTHER (write in):
Click in boxes to use drop-down lists
Evaluation of Site's Technolog ist QC Program
PASS/FAILMQSA Regs
1. CNR Weekly Check Weekly2. Phantom Image Weekly3. Visual Checklist Monthly4. Repeat Analysis Quarterly5. Compression Semi-annually6. Printer QC* See FDA guidance7. Review Workstation QC-Overall* See FDA guidance
Medical Physic ist' s Recomm endations for Quality Improvement
MEDICAL PHYSICIST'S MAMMOGRAPHY QC TEST SUMMARY
(Required for Annual Surveys; not required for Mammography Equipment Evaluations of new units. However, medical physicists must reviewthe site's technologist QC program within 45 days and complete this section so that the facility may submit this form along with the entire
Mammography Equipment Evaluation report with their phantom and clinical images to the ACR.)
(Fuji, cont inued)
20
Fuji Digital Mammography QCTechnologist’s Tests
TESTCNR-Contrast/NoisePhantom ImagePrinter/Moni torVisual ChecklistRepeatAnalysisCompressionIP Fog
FREQUENCYWeeklyWeekly
Per ManufacturerM onthly
QuarterlySemi-annualSemi-annual
Fuji Digital Mammography QCTechnologist’s Tests
Printer TestsNo film in cassetteImaging Plate should be erased after 8
hours of non-useImaging Plate should be processedor
erased after eachexposurePrin ters are monitored according to the
manufacturer ’s approved QC program
Fuji Digital Mammography QCTechnologist’s Tests
Phantom ImagesTestConfi rms imagequality using ACR PhantomToolsrequired - QC Cassette& ACR Phantom
Method - AEC Exposure,processIP using thePhysics,ACR MAPP Menu
Visually inspect the image. Samescoring methodasMQSA
Correction Period - Before any fur therexaminationsare performed
Fuji Digital Mammography QCTechnologist’s Tests
CNR (Contrast/NoiseRatio) TestEvaluatesNoiseand Contrast using a fixed x-
ray exposureTools needed- 4 cm acrylic block & 0.2mm AlM ethod - Fixed manual techniqueCNR Measurementmust be within +/- 20%Correction Period - Before any further
examinations are performed
21
Fuji Digital Mammography QCTechnologist’s Tests
Testsin Accordancewith ACR/MQSAVisual Checklist Monthl yRepeatAnalysis Quarterl y
Compression Test Semi-Annual
Same Action Lim its and Corrective Action asMQSA
IP Fog replacesDarkroom Fog test - SemiAnnual
Review WorkstationQC
(See FDA guidance)
Compression(25-45 lb)
(semi-annually)
Repeat Analysis(≤2% change)
(quarterly)
Visual Checklist(monthly)
DECNOVOCTSEPAUGJULJUNMAYAPRMARFEBJANMonth
Year
Visual Checklist - Technologist QC
FULL-FIELD DIGITAL MAMMOGRAPHY QUALIT Y CONTROL CHECKLIST –FUJI FCRmDaily and Weekly Tests
ReviewWorkstation QC
(See FDAguidance)
Printer QC(See FDAguidance)
Phantom Image(weekly)
CNR WeeklyCheck (weekly)
Initial s
Date
Month
Year
Fuji Digital Mammography QCTechnologist’s Tests
Imaging Plate FogConfirm that the storageconditions of the exposure
room protects the IP cassettes fr om scatter
Tools required - IP CassetteMethod - Tape coin to front of cassetteand place in
storagelocation dur ing test exposures
Visually inspectimagefor visibil ity of coinCorrection Period - Before any fur ther examinations
are performed
22
Fuji Digital Mammography QCTechnologist’s Tests
Inter -Plate Consistency- As NeededThe Inter -plate Consistency Test replaces the
ScreenSpeedTest
Must be performedwhen new IPs and/orcassettesare introduced
Can be performed by the QC technologist underQualified Medical Physicist oversight
Fuji Digital Mammography QCTechnologist’s Tests
Inter-Plate Consistency- As NeededConfirm that Imaging Platesusedin the facility
are similar in sensitivity and imagequality
Tools Required - new and existing IPs andCassettes,4 cm acrylic block
Method - ExposecassettesusingAEC (time)Mode
Acceptable results: mAs +/- 10%, SNR +/- 15%
Correction Period - Wi thin 30 daysof the test
Thank You !! !