View
223
Download
1
Tags:
Embed Size (px)
Citation preview
Aim
Aim is to suggest ways MICE can fail Look with a critical eye at all MICE systems
From an Analysis/emittance measurement standpoint
A personal view of the associated risk/impact Dependent on my knowledge of the subsystem
Incomplete by definition
What can go wrong?
Detectors Detectors are insufficiently calibrated Detectors have insufficient resolution PID accuracy is insufficient Event rates issues Apertures are too small
Beam Beamline doesn’t fill phase space/match
adequately/consistently Beamline is too impure Event rates issues
Analysis/Optics Beam heating is dominated by optics Beam weighting is too difficult Systematics reduction is too difficult
TOF Resolution
Required to measure bunch length ~ 0.5 ns RMS from RF Bucket size
For 1e-3 emittance measurement resolution of TOF should be <14%*0.5 ns ~ 70 ps
At tracker reference plane Vs RF zero crossing Including materials effects and tracker energy resolution between
TOF and tracker Diffuser between TOF I and upstream tracker
Is this possible? Requirement/consideration also needed for correlations
Between t and x,y,px,py,pz
Tracker Resolution
Pz resolution of tracker should be ~ 2.5 MeV Marginal at low Pt Contingency - weaken the tracker field?
Serious knock on for optics & beamline
Tracker field TRD says 4T/240, 4T/200, 3.4T/170, 2.8T/140 Is this the final word? Knock on for optics & beamline Requires some planning to change these values
Light loss RF Background
Detector Calibration (Hardware)
Requirement For 1e-3 emittance measurement detectors must be
calibrated to 10% of RMS Example
Tracker must be aligned to ~ 10% * 2 MeV/c ~ 0.2 MeV/c 0.2 MeV/c ~ 1 mrad * 200 MeV/c This is comparable to solenoid distortion (!)
Example Timing measurement must be calibrated to 7 ps At the tracker reference plane Including materials effects and tracker pz
I have not seen a full demonstration/plan for this calibration accuracy either for Tracker or TOF
The 10% requirement is historical and may need to be updated
Vessel & Support FEA Results (S. Virostek LBNL CM13)
50 ton, uniform axial load on vessel; fully fixed at support stand base
Max deflection: 1mm
Detector Aperture
Scraping is a significant effect if we want to predict the number of muons in a nufact acceptance
Beta-tapering means scraping continues down the beamline First order?
We should measure it But it is unlikely that we will be able to measure the full
acceptance of the cooling cell
PID Resolution
PID looks capable of achieving <1e-3 impurity downstream
I have no feel for upstream detectors Depends on a high purity muon beam
Impurity/efficiency requirements ignore position of mis-pid in the beam
Mis-pid at high “emittance” will bias the measurement more
Beam Matching Cooling is dependent on a well-matched beam A direct measurement of cooling with minimal beam weighting is highly desirable
Much stronger argument Challenging conditions to work under
Quadrupoles, material, scraping, high emittances Worry about RMS’s and mean Matched beams below 6 are high risk
No/thin diffuser removes a d.o.f. from the optics Requirement for tight focussing is looking highly challenging Collimation will leave a nasty beam, more susceptible to emittance growth
Beam Purity
Given uncertainties in the beamline, beam purity is also uncertain
Although indications are it should be quite good If the upstream PID is not so good this will
become important
Event Rate
Number of /sec depends on several potential limiting factors Number of protons on target Number of /proton on target Max rate at TOF 0 / Ckov I efficiency from TOF 0, Ckov I to tracker Max rate at tracker 600 => 250 good =>
Issues How far can we dip into ISIS? How efficient is the beamline matching section
With collimators for low emittance Assuming the 600 vs 250 at tracker resolved?
Optical heating
Optical heating/cooling is a serious effect If we are to claim cooling, we have to understand it to the
1e-3 level This is poorly understood at present Transverse
Transverse heating can be countered by selecting different beta functions for the beam at different momenta
Longitudinal Longitudinal heating tends to be worse as there is no
longitudinal focussing (RF at 90o) Even with RF at 40o significant heating in the gap between
the TRP and the linac
Beam Weighting
Beam weighting in a 6D phase space is not easy problem
Working in 6D In the presence of detector errors Amplitude Momentum correlation? Momentum dependent beta function?
Required for both bunch emittance and single particle emittance analyses
Important if beamline fails to deliver matched beam
Cooling Channel Measurements
Measuring B-fields and LH2 looks okay RF measurement looks hard
No obvious strategy Particle-based measurements will be difficult
Run the RF Compare with tracking code?
This will be used to calibrate the measurement But not needed directly for the emittance measurement
Systematics Reduction Emittance Measurement
Take calibration Feed it into G4MICE Use it to predict systematic error on experiment from
detector resolution What constraints/requirements does this place on
G4MICE? Accuracy of tracking Accuracy of physics processes
e.g. MuScat in the SciFi Accuracy of geometric model
e.g. material between TOF and SciFi Accuracy of digitisation
For TOF
Personal View - Risk Analysis (1)
Issue Risk Impact
Remedy/Comment
Tracker Calibration
Med High Tracker group -> how accurate is the calibration?
Timing Calibration
High Med Can we accurately measure time at the TRP?
Transverse Resolution
Low High okay
Longitudinal resolution
High Med Longitudinal emittance resolution will probably be poor
PID resolution Med Med No clue about upstream PID
Rate Med Low Run for longer
Apertures High Low Accept poor PID in scraping region?
Beamline match High Med More manpower in beamline or beam weighting
Beamline purity Low Med More manpower in beamline
Personal View - Risk Analysis (2)
Issue Risk Impact
Remedy/Comment
Transverse beam heating
Med High Study non-linear dynamics of cooling channel
Longitudinal beam heating
High Med Study non-linear dynamics of cooling channelShowing 6D cooling looks difficult at present
Beam weighting Med High Crucial if the beamline is poorly matched
Channel calibration
Med Med Demonstrate a method for measuring RF
Resolution Red. Med Med Need constraints on G4MICE accuracy