Measuring Components Business Unit 1-1-1, Katase, Fujisawa-shi, Kanagawa-ken, 251-8531 Japan FUJISAWA PLANT  TEL +81-466-22-7152 / FAX +81-466-22-1701 http://www.minebea-mcd.com EN 294-1626 MINEBEACO.,LTD. General specifications Thank you very much for your purchasing Minebea’ s Digital Conversion Module, model CSD-892-76. This [start guide] for CSD-892-76 explains the installation method, the connection method of a load cell, calibration procedure, etc. Also, carefully store this [start guide]so that it can be referred to at any time. Please download the instruction manual from our homepage and confirm the further details. http://www.minebea-mcd.com/product/i-amp/csd892.html Short bar SHIELD SHIELD -SIG +SIG +EXC -EXC -SIG +SIG +EXC -EXC Operating temperature range -10 ℃ ~ 50 ℃ Operating humidity : 85 %RH or less (Non condensing.) Storage temperature range : -20 ℃ 60 ℃ Power-supply voltage : DC24 V(Permissible variable range DC20.4 V to DC27.6 V) Power consumption : Approx. 2.4 W (at DC24 V) Outline dimensions (W×H×D) 72 mm × 96 mm × 67.4 mm (excludes protruding parts) Weight : Approx.260 g  The instrument can connect with the strain gage applied transducers, such as load cells, pressure sensors, etc. Please connect the load cell to the upper terminal block. We describe here as an example of connection with load cells, and please process it similally for connecting with other strain gage applied transducers. Please ground the protective ground terminal by D-class with single earth. Otherwise, it may cause an unexpected malfunction because of the influence of noise from the other equipments. Please refer [Safety precautions] about the connection to comply with EMC instruction. Environment not to set up Do not install the Instrument in following places. It may cause a damage to the Instrument. Places explosed to direct sunlight and/or places in the high temperature. Places in a highly humid area. Places where the instrument is directly affected by vibrations or mechanical shocks. Environments with full of dust or coarse particulates. Environments containing of corrosive gas or salt. Environments with rapid change in temperature and/or humidity. Near the devices which generate magnetism or electromagnetic waves. Environments vulnerable to radioactivity or radioactivity rays. Environments where chemical reaction may take place such as a laboratory. Applicable environment Please set it up into the operation control panel when this instrument is used in the environment that includes the drop of water and many dust etc.. EMC directive(EN61326-1:2006) [Electrical equipment for measurement, control, and laboratory use - EMC requirements] [Immunity test requirements for equipment intended for use in industrial locations] Please observe the following conditions strictly when this instrument suits the below, it may not conform to the above standard when these are neglected. ※CE conformity standard is not effective in case of using USB interface. Location of installation:Please set up in the shielded case or control panel where EMC solution is given in case of comforming. Power supply:The power supply by which DC24V is supplied to the instrument must use the conformed product of CE standard, and set the instrument up in the same storage case. Cable:Use the shielded cable other than the power cable. Shield processing : Please connect the shielded cable of load cell with E terminal in upper terminal block. Also, connect the shield of RS-422/RS-485 cable with the F.G. terminal in the lower terminal block. Please make sure to do the shield processing on the cable for RS-422/RS-485 and the external control I/O by conduit piping including the control panel. Grounding:The ground of this instrument shall apply the individual ground by using the F.G. terminal in lower terminal block. Please make the single ground of the instrument, or make the single ground of the storage case after connecting the instrument with the storage case when you store only the DC24 V power circuit for the instrument and the instrument in the storage case. Please give the earth of the instrument as a single earth when you store equipment other than the instrument in the storage case. The shielded cable line is used for the connection of the external control I/O, and connect the shield with F.G terminal of the lower terminal block. Point! Point! Accessories Start guide (in Japanese) 1 piece. Start guide (in English) 1 piece. The external control input is executed by shortening each input and COM.1 with a contact point or open collector after wiring the connector. The external control output is executed by open collector output. (Open collector rated: V CE = DC35 V, I C = DC50 mA at maximum.) Connection with 6-wires cable. Connection with 4-wires cable. Attached connector for RS-422/RS-485 : MSTB2.5/6-ST-5.08 (by PHOENIX CONTACT) As to the termination, connect the 110 Ω after shortening the TRM. terminal and the RDB terminal at the furthest RS-422/RS-485 connecter from the host. The total extension length of the cable is less than 1 km approximately. Point! Short bar 2 pieces RS-422/485 connector 1 piece Please make sure to open the lower terminal cover when you pull down a lower claw with a flat screwdriver, etc. If you don't open it, the cover may be dameged. Please connect the power supply and the ground surely as shown in the figures and also use within the regurated power supply condition. E B D C G F A E B D C G F A FUNC. key… Enter into the function mode, or change each status to the former status set in advance. Shifts to Calibration setting mode.(By pushing 2 seconds or more) CAL-S/▲ key … Set value increment. Shifts to SPAN calibration mode of a simple calibration.(By pushing 2 seconds or more) F/ENTER key … Register the setting value. Execute function set in F/ENTER key. (Selectable from None, Hold, Zero set, Zero clear, Tare weight cancellation, Tare weight clear, changeover of gross valu/net value. Shift to the setting mode of comparator, Check the RS-422/485 address.) CAL-Z/  key … Carry up the setting value. Shifts to ZERO calibration mode of a simple calibration. (By pushing 2 seconds or more) Also shifts to calibration lock mode of a simple calibration. (By pushing 2 seconds or more CAL-Z / and CAL-S/▲ at the same time) Digital Conversion Module start guide CSD-892-76 Front installation Upward installation Installation on ceiling Vertical installation When connecting wires, turn off the power supply without fail. Do not supply the electric power until the installation is completed. There is no switch that switches ON/OFF of power supply in the main body. Do not drop or do not give the high impact because the terminal block is made by the resin. Please separate the cable connected with the main body from the noise source such as I/O for the control, power supply line as much as possible. Do not share the wiring with the other lines as a special conduit wiring. Be sure to connect a grounding wire with exclusive use of D-type single grounding. Do not share it with other groundings for power supply system. The tightening torque of terminal screws on the terminal board is 0.6N・m. Please use the solderless terminal with 6.2 mm or less in the width. (O type 1.25-3 or Y type 1.25-3.5) ●  The stripping length of electrical cable tip is 6 mm. ●  The tightening torque of terminal screws of RS-422/RS-485 connector is 0.6 N・m. ●  For the connection of RS-422/RS-485 cable, use a twisted pair cable line with shield and connect the shield with F.G. terminal in the lower terminal block. ●  When you connect with this instrument, please check the direction of upper/lower sides of the connector, and insert until it bumps. In this start guide, the following signs are used to understand the important matter easilly. Be sure to read for safety. Connection with RS-422/485 interface Safety precautions When you install it in the control panel, please set it up in the front or upward so that the front panel that includes the display of this instrument may come forward. Direction of installation Attaching method to a DIN rail Removing method from a DIN rail Please connect the power supply and ground to the lower terminal block as shown in the figures below. Connection with power supply and ground/Connection of external control I/O Notes for connecting wires Each name and functions Connection with Load Cell This instrument adopts the rail mounting type with 35mm width of DIN standard. Hang an upper claw of this instrument on the upper side of DIN rail. Press this instrument onto DIN rail until a click sound is heard. Confirm the firm fixation. Confilm all the cables including the power supply cable have come off from the main body. Pull a lower claw of this instrument downward from the front direction using a screwdriver, and then remove this instrument from DIN rail. Please refer to the instruction manual for the details of an internal circuit etc. Please refer to the instruction manual for the parallel connection of a load cell, and the extension of a cable. ① Unit display section The upper terminal cover ③ Condition display section Load display section ④ Status LED ⑥ CAL-Z/ key ⑦ CAL-S/▲key ⑨ Connecter for USB interface FUNC. key ⑧ F/ENTER key The lower terminal cover Permissible variable range : DC20.4 V DC27.6 V) D-class with single earth DC24V 1 2 3 4 5 6 8 9 10 11 12 13 7 OUT 1 Common 2(For output) Output 1 OUT 2 Output 2 OUT 3 Output 3 OUT 4 Output 4 OUT 5 Output 5 8 9 10 11 12 13 F.G. + 24V Frame ground Power supply DC24V 0V Power supply DC0V COM.1 COM.2 Common 1(For input) I N 1 Input 1 I N 2 Input 2 I N 3 Input 3 Point! CAL-Z CAL-Z FUNC. FUNC. F F ENTER ENTER CAL-S CAL-S In using the 4-wires cable, shorten the load cell terminals A - F and C - G with the attached short bar. The length of the cable at 6-wires type connection is 100 m or less, and 4-wires type connection is 30 m or less. There may become outside the quality assurance for the length more than the regulations. When the tension is applied with the use of tension type or universal (tension/compression) type load cell, and display of (+) direction is required, connect the input (+) of the load cell to the terminal B and the input (-) of the load cell to the terminal D individually. (Excludes CLS-A, CL-B, UL(A), UT(A) of Minebea's load cell.) Point! ※Described colors are Minebea's standard. (Excludes LSM-B of Minebea's load cell.) +Load cell input Load cell input SHIELD B D C G F A E SHIELD +EXC SEN +SEN EXC +SIG SIG +Load cell power supply +Sensing input Sensing input Load cell power supply Red 1 2 3 4 5 6 7 ZERO calibration HOLD function is effective. Key lock is effective. ERROR signal outputs. External control output turns ON. The load display is net value SPAN calibration Display of mV/V value The measured data is steady. Tare weight cancellation is executed. The load display is gross value. Center zero Please refer to the instruction manual about simple calibration. ④ Status LED … The communication status of RS-422/485 interface is monitored with two LED. ⑨ Connecter for USB interface USB interface is used for reading the set data from host and writing the parameter details set for the measurement and the the calibration The subject driver installation is needed. This sign forewarns the presence of hazards that the user may resut in serious injury. This sign describes an attention and a limitation, etc in the operation and the work. ① Unit display section … Display the unit set for measurement. Load display section … Display [Gross/Net value], [Overload] or [Error]. Also display the condition or the setting value in the various setting. Condition display section … Each display is turn on under the follow condition. Instruction manual Instruction manual Instruction manual Instruction manual Instruction manual Instruction manual Please refer the instruction manual for the details of wiring. Instruction manual Instruction manual TRM. RDB RDA SDB SDA S.G. SDB TRM. S.G. RDB RDA SDA Difference output (-) Difference output (+) Difference input (-) Difference input (+) Signal ground Cable end resistance Black White Orange Green Blue Yellow

CSD-892-76-- - minebea-mcd.com … · CSD-892-76. This [start guide] for ... The stripping length of electrical cable tip is 6 mm. ... Confilm all the cables including the power supply

Embed Size (px)

Citation preview

Page 1: CSD-892-76-- - minebea-mcd.com … · CSD-892-76. This [start guide] for ... The stripping length of electrical cable tip is 6 mm. ... Confilm all the cables including the power supply

Measuring Components Business Unit1-1-1, Katase, Fujisawa-shi, Kanagawa-ken, 251-8531 JapanFUJISAWA PLANT  TEL +81-466-22-7152 / FAX +81-466-22-1701

http://www.minebea-mcd.comEN 294-1626

MINEBEACO.,LTD.General specifications

Thank you very much for your purchasing Minebea’s Digital Conversion Module, model CSD-892-76.This [start guide] for CSD-892-76 explains the installation method, the connection method of a load cell, calibration procedure, etc.Also, carefully store this [start guide]so that it can be referred to at any time.Please download the instruction manual from our homepage and confirm the further details.

http://www.minebea-mcd.com/product/i-amp/csd892.html

Short bar

SHIELD SHIELD

-SIG+SIG

+EXC

-EXC

-SIG+SIG

+EXC

-EXC

 ● Operating temperature range : -10 ℃ ~ 50 ℃ ● Operating humidity : 85 %RH or less (Non condensing.) ● Storage temperature range : -20 ℃ ~ 60 ℃ ● Power-supply voltage : DC24 V(Permissible variable range DC20.4 V to DC27.6 V) ● Power consumption : Approx. 2.4 W (at DC24 V) ● Outline dimensions (W×H×D) : 72 mm × 96 mm × 67.4 mm (excludes protruding parts) ● Weight : Approx.260 g 

The instrument can connect with the strain gage applied transducers, such as load cells,

pressure sensors, etc. Please connect the load cell to the upper terminal block.We describe here as an example of connection with load cells, and please process it similally for connecting with other strain gage applied transducers.

■ Please ground the protective ground terminal by D-class with single earth. Otherwise, it may cause an unexpected malfunction because of the influence of noise from the other equipments.■ Please refer [Safety precautions] about the connection to comply with EMC instruction.

Environment not to set up

 Do not install the Instrument in following places. It may cause a damage to the Instrument.

  ● Places explosed to direct sunlight and/or places in the high temperature.  ● Places in a highly humid area.  ● Places where the instrument is directly affected by vibrations or mechanical shocks.  ● Environments with full of dust or coarse particulates.  ● Environments containing of corrosive gas or salt.  ● Environments with rapid change in temperature and/or humidity.  ● Near the devices which generate magnetism or electromagnetic waves.

  ● Environments vulnerable to radioactivity or radioactivity rays.  ● Environments where chemical reaction may take place such as a laboratory.

Applicable environment Please set it up into the operation control panel when this instrument is used in the environment that includes the drop of water and many dust etc..

EMC directive(EN61326-1:2006) [Electrical equipment for measurement, control, and laboratory use - EMC requirements] [Immunity test requirements for equipment intended for use in industrial locations] Please observe the following conditions strictly when this instrument suits the below, it may not conform to the above standard when these are neglected. ※CE conformity standard is not effective in case of using USB interface.

 ● Location of installation:Please set up in the shielded case or control panel where EMC solution is given in case of comforming. ● Power supply:The power supply by which DC24V is supplied to the instrument must use the conformed product of CE standard, and set the instrument up in the same storage case. ● Cable:Use the shielded cable other than the power cable. ● Shield processing : Please connect the shielded cable of load cell with E terminal in upper terminal block. Also, connect the shield of RS-422/RS-485 cable with the F.G. terminal in the lower terminal block. Please make sure to do the shield processing on the cable for RS-422/RS-485 and the external control I/O by conduit piping including the control panel. ● Grounding:The ground of this instrument shall apply the individual ground by using the F.G. terminal in lower terminal block. Please make the single ground of the instrument, or make the single ground of the storage case after connecting the instrument with the storage case when you store only the DC24 V power circuit for the instrument and the instrument in the storage case. Please give the earth of the instrument as a single earth when you store equipment other than the instrument in the storage case.

■ The shielded cable line is used for the connection of the external control I/O, and connect the shield with F.G terminal of the lower terminal block.Point!

Point!

Accessories

❶ Start guide (in Japanese) 1 piece. Start guide (in English) 1 piece.

 The external control input is executed by shortening each input and COM.1 with a contact point or open collector after wiring the connector. The external control output is executed by open collector output. (Open collector rated: VCE = DC35 V, IC = DC50 mA at maximum.)

Connection with 6-wires cable. Connection with 4-wires cable.

■ Attached connector for RS-422/RS-485 : MSTB2.5/6-ST-5.08 (by PHOENIX CONTACT) ■ As to the termination, connect the 110 Ω after shortening the TRM. terminal and the RDB terminal at the furthest RS-422/RS-485 connecter from the host.■ The total extension length of the cable is less than 1 km approximately.

Point!❷ Short bar 2 pieces ❸ RS-422/485 connector

1 piece

● Please make sure to open the lower terminal cover when you pull down a lower claw with a flat screwdriver, etc. If you don't open it, the cover may be dameged.

● Please connect the power supply and the ground surely as shown in the figures and also use within the regurated power supply condition.

E B D C G F A E B D C G F A

⑤ FUNC. key… Enter into the function mode, or change each status to the         former status set in advance.        Shifts to Calibration setting mode.(By pushing 2 seconds or more)

⑦ CAL-S/▲ key … Set value increment. Shifts to SPAN calibration mode of a         simple calibration.(By pushing 2 seconds or more)

⑧ F/ENTER key … Register the setting value. Execute function set in F/ENTER key.        (Selectable from None, Hold, Zero set, Zero clear, Tare weight         cancellation, Tare weight clear, changeover of gross valu/net value.          Shift to the setting mode of comparator, Check the RS-422/485 address.)

⑥ CAL-Z/  key … Carry up the setting value. Shifts to ZERO calibration mode         of a simple calibration. (By pushing 2 seconds or more)         Also shifts to calibration lock mode of a simple calibration.        (By pushing 2 seconds or more CAL-Z / and CAL-S/▲ at the same time)

Digital Conversion Module

start guideCSD-892-76

Front installation

Upward installation

Installation on ceiling

Vertical installation

● When connecting wires, turn off the power supply without fail.● Do not supply the electric power until the installation is completed. There is no switch that switches ON/OFF of power supply in the main body.● Do not drop or do not give the high impact because the terminal block is made by the resin.● Please separate the cable connected with the main body from the noise source such as I/O for the control, power supply line as much as possible.● Do not share the wiring with the other lines as a special conduit wiring.● Be sure to connect a grounding wire with exclusive use of D-type single grounding. Do not share it with other groundings for power supply system.● The tightening torque of terminal screws on the terminal board is 0.6N・m.● Please use the solderless terminal with 6.2 mm or less in the width. (O type 1.25-3 or Y type 1.25-3.5)

● The stripping length of electrical cable tip is 6 mm.● The tightening torque of terminal screws of RS-422/RS-485 connector is 0.6 N・m.● For the connection of RS-422/RS-485 cable, use a twisted pair cable line with shield and connect the shield with F.G. terminal in the lower terminal block.● When you connect with this instrument, please check the direction of upper/lower sides of the connector, and insert until it bumps.

In this start guide, the following signs are used to understand the important mattereasilly. Be sure to read for safety.

Connection with RS-422/485 interface

Safety precautions

When you install it in the control panel, please set it up in the front or upward so that the front panel that includes the display of this instrument may come forward.

Direction of installation

Attaching method to a DIN rail

Removing method from a DIN rail

Please connect the power supply and ground to the lower terminal block as shown in the figures below.

Connection with power supply and ground/Connection of external control I/O

Notes for connecting wires

Each name and functions

Connection with Load Cell

This instrument adopts the rail mounting type with 35mm width of DIN standard.

❶ Hang an upper claw of this instrument on the upper side of DIN rail.

❷ Press this instrument onto DIN rail until a click sound is heard.

❸ Confirm the firm fixation.

❶ Confilm all the cables including the power supply cable have come off from the main body.

❷ Pull a lower claw of this instrument downward from the front direction using a screwdriver, and then remove this instrument from DIN rail.

Please refer to the instruction manual for the details of an internal circuit etc.

Please refer to the instruction manual for the parallel connection of a load cell, and the extension of a cable.

① Unit display section

The upper terminal cover

③ Condition display section

② Load display  section

④ Status LED

⑥ CAL-Z/ key

⑦ CAL-S/▲key

⑨ Connecter for USB  interface⑤ FUNC. key

⑧ F/ENTER key

The lower terminal cover

Permissible variable range : DC20.4 V ~ DC27.6 V)D-class with single earth DC24V

1 2 3 4 5 6

8 9 10 11 12 13

7

OUT1

Common 2(For output)

Output 1

OUT2 Output 2

OUT3 Output 3

OUT4 Output 4

OUT5 Output 5

8

9

10

11

12

13

F.G.

+24V

Frame ground

Power supply DC24V

0V Power supply DC0V

COM.1

COM.2

Common 1 (For input)

I N 1 Input 1

I N 2 Input 2

I N 3 Input 3

Point!

CAL-ZCAL-Z

FUNC.FUNC.

FF

ENTERENTER

CAL-SCAL-S

■ In using the 4-wires cable, shorten the load cell terminals A - F and C - G with the attached short bar.■ The length of the cable at 6-wires type connection is 100 m or less, and 4-wires type connection is 30 m or less. There may become outside the quality assurance for the length more than the regulations.■ When the tension is applied with the use of tension type or universal (tension/compression) type load cell, and display of (+) direction is required, connect the input (+) of the load cell to the terminal B and the input (-) of the load cell to the terminal D individually. (Excludes CLS-A, CL-B, UL(A), UT(A) of Minebea's load cell.)

Point!

※Described colors are Minebea's standard. (Excludes LSM-B of Minebea's load cell.)

+Load cell input

-Load cell input

SHIELD

B

D

C

G

F

A

E SHIELD

+EXC

-SEN

+SEN

-EXC

+SIG

-SIG

+Load cell power supply

+Sensing input

-Sensing input

-Load cell power supply

Red

1

2

3

4

5

6

7

ZERO calibration

HOLD function is effective.

Key lock is effective.

ERROR signal outputs.

External control output turns ON.

The load display is net value

SPAN calibration Display of mV/V value

The measured data is steady. Tare weight cancellation is executed.

The load display is gross value.

Center zero

Please refer to the instruction manual about simple calibration.

④ Status LED … The communication status of RS-422/485 interface is monitored with two LED.

⑨ Connecter for USB interface … USB interface is used for reading the set data from host         and writing the parameter details set for the measurement and the         the calibration The subject driver installation is needed.

This sign forewarns the presence of hazards that the user may resutin serious injury.

This sign describes an attention and a limitation, etc in the operationand the work.

① Unit display section … Display the unit set for measurement.

② Load display section … Display [Gross/Net value], [Overload] or [Error].         Also display the condition or the setting value in the various setting.

③ Condition display section … Each display is turn on under the follow condition.

InstructionmanualInstructionmanual

InstructionmanualInstructionmanual

InstructionmanualInstructionmanual

Please refer the instruction manual for the details of wiring.

InstructionmanualInstructionmanual

TRM. RDB RDA SDB SDAS.G.

SDB

TRM.S.G.

RDBRDA

SDADifference output (-)Difference output (+)

Difference input (-)Difference input (+)

Signal groundCable end resistance

Black

White

Orange

Green

Blue

Yellow

Page 2: CSD-892-76-- - minebea-mcd.com … · CSD-892-76. This [start guide] for ... The stripping length of electrical cable tip is 6 mm. ... Confilm all the cables including the power supply

▶▶

▶▶

▶▶

▶▶

▶▶

2

3

4

2

2

3

2

3

2

3

4

2

3

2

3

2

3

2

3

2

3

※ Memorized load cell output voltage is displayed..

F

ENTER

F

ENTER

F

ENTER

F

ENTER

STABLE

F

ENTER

F

ENTER

F

ENTER

F

ENTER

FUNC.

STABLE

By pushing   key, displays, and then by

pushing    key, [mV/V] of condition display is turned

on, and       [mV/V] displays.

After setting, push   key to display .

Please refer to the instruction manual for[Calibration at zero

with inputting the numeric value].

Turn on the power supplyAdvance preparations

Switch to the calibration modeStep 1

Switch to the calibration modeStep 1

Step 2

Step 3

Step 4

Step 5

Step 6

Step 7

Measurement mode

Change from the standard measurement mode to the

By pushing key three times, the display changesas follows.

  displays. Push   key here.

When   key is pushed at ,       

       is displayed.

displays and it enters into the calibration mode.

Push   key when      blinks, and the

       sign lights on.

Error display for zero calibration

Error display for setting the weighing capacity or mass of the weight.

※ The weighing capacity memorized now.

selects the changed digit.

increments the value of selected digit.

selects the digit to be changed.

increments the value of selected digit.

The unitmemorized.The unitmemorized.

The positionmemorized.The positionmemorized.

The set-up unit.The set-up unit.

Settingposition.Settingposition.

Weighing capacitymemorized.Weighing capacitymemorized.

Setting weighing capacity.Settingweighing capacity.

*Set the condition that nothing is put on the measuring part.

blinks for about 2 seconds when the output value of the load cell exceeds over ZERO adjustable range at minus side. (at -2.5mV/V or less)

blinks for about 2 seconds when the output value of the load cell exceeds over ZERO adjustable range at plus side. (at 2.5mV/V or more)

blinks for about 2 seconds when the output of the load cell doesn’t reach within the span adjustable range ([Load cell output voltage at SPAN] - [Load cell output voltage at ZERO] ≦ 0.0mV/V)

blinks for about 2 seconds when the output of the load cell exceeds over the SPAN adjustable range. (at 3.1 mV/V or more)

SPAN is memorized, and       displays.

● The load value display is not stabilized.

 ▶ An indicated value may not be stabilized because of the influence of vibration.  You can stabilize an indicated value by three kinds of filter facilities of analog filter, digital filter, and stabilized filter.   Refer to the operation manual for details, such as the setting method of these functions.

The calibration is an operation that matches the load of the sensor to the display of this instrument to display the electrical signal from the sensor (load cell) as an accurate load.

Here, we explain the calibration procedure from a factory-shipments state based on the condition at the factory-shipment. As an example, we explains by the case that the weight of 4 ton can be prepared, the weighing value is 6.000 ton and a scale interval is 0.002 ton.We describe the example that zero is registered by [Calibration at zero by the weight], span is registered by [Calibration at span by the weight] or [Calibration at span with inputting the numeric value].

Set the unit

Step 2Set the unit

Set the decimal point display position

Step 3Set the decimal point display position

Set the scale interval

Step 4Set the scale interval

Set the weighing capacity

Step 5

Step 6

Set the weighing capacity

Set the mass of the weight

After turning on the power supply of DC24 V, CSD-892-76 is energized for about 10 minutes to stabilize CSD-892-76 and weighing section (load cell).

Switch from measurement mode to the calibration mode.

Set the unit. Selectable unit : [OFF(non)], [g], [kg], [t], [lb], [N] [kN], [Pa], [kPa], [MPa]

Set the decimal point display positionSelectable position : [OFF(non)], [0.0], [0.00], [0.000] [0.0000]

Set the scale intervalSelectable interval : [1],[2],[5],[10],[20],[50]

Set the weighing capacity. (The maximum load that can be measured.)

Set the mass of the weight put on the measuring section.Use the weight by 2/3 or more of the weighing capacity to reduse the calibration error. Otherwise, input the same value as weighing capacity when there is no weight.

ZERO calibration by the measurement value. ZERO calibration by numeric input.

SPAN calibratin by numeric input.

Electric calibration

Electric Calibration

SPAN calibration by the weight.

(CCAL)

(UNIT)

(D.P.)

(SCAL)

(DISP)

(LOAD)

(ZERO) (Z MV)

(END)

Point!

Point!■ The load display changes with 0.002t → 0.004t → 0.006t, when the scale interval is set to [02]. (When you set the decimal point as [0.000] and the unit as [t].)

(SPAN) (S MV)

ZERO is registered by reading the load cell output value in the condition that nothing is put on the weighing section. (Initial condition including tare.)

When CSD-892-76 used is replaced with another unit and only when all the below requirements are accepted, ZERO calibration by the value input can be applied. ・When ZERO calibration data before the replacement remains.  ・When the condition that nothing is put on the measuring section (such as an initial load) is impossible.

SPAN is registered by inputting the value in which the load cell output value at zero is subtracted from the load cell output value assumed when loaded for rating weight.

SPAN is registered by reading the load cell output value in the condition to put the weight set in step 6 on measuring part.

FUNC.FUNC.

■ When   key pushed since step 2, the set value before is cancelled, and

returns to     . The set value is not preserved when canceling on the way.Point!

■ Adjust the mass of the weight used for the calibration to 2/3 or more of the weighing capacity.■ When the weight of the same mass as weighing capacity can be prepared, set as [Weighing capacity = Mass of weight].■ When the span calibration in step 8 is registered by the value input, set as [Weighing capacity = Mass of weight].

display blinks for about 2 seconds when the calibration is executed, and [Weighing capacity]<[Mass of weight] is set.* Set the each value to become as [Weighing capacity ≧ Mass of weight].

Point!

Point!

t

t

t

t

t

kg

F

ENTER

F

ENTER

F

ENTER

F

ENTER

F

ENTER

※ Weight value put on measurement section.

F

ENTER

※ The unit memorized now is displayed.

※ The position of decimal point memorized now is displayed.

When   key is pushed at ,

is displayed.

Select the position of decimal point.

F

ENTER

※ The scale interval memorized now.

After setting, push   key to display .F

ENTER

F

ENTER

ZERO is registered by reading the load cell output value in the condition that cothing is put on the measuring part(Initial condition including tare).

ZERO is registered by reading the load cell output value in the condition that nothing is put on the measuring part (Initial condition including tare).

※ Please check the rated capacity and the tare weight value,etc., so that ZERO should not exceed the adjustable range.

Error display of SPAN calibration

※ Please check the rated capacity of load cell and mass of the weight so that SPAN should not

exceed the adjustable range.

lights on when the load display exceeds over [(+Weighing capacity) + 9D] or[+110 % of weighing capacity]. (This error display depends on the setting.)

lights on when the load display falls below [(-Weighing capacity) - 9D], [-110 % of weighing capacity], or [-20D]. (This error display depends on the setting.)

*Check the setting value or retry the calibration along with the trouble shooting in instruction manual.

Please set the value in which the output value equal to ZEROis subtracted from the load cell output value equal to the setweighing capacity in each mV/V.

In such a case

Error display during the measurement after the calibration.

FUNC.FUNC.

The valuememorizedThe valuememorized

The set-up valueThe set-up value

mV/V

mV/V

Finish the calibration mode.

Please set the scale interval.

Unit display section changes [OFF(non)],[g],[kg],[t][lb],[N],[kN],[Pa],[kPa],[MPa]

Set the weighing capacity.

Set the mass of the weight.

■ The set value is registered temporary until fixing it by pressing [F/ENTER] key on display.■ Execute the calibration with actual load when you can prepare the weight of   2/3 or more of weighing capacity. The accuracy over 1/1000 can be expected.  Because it is a calibration in an actual measurement system. ■ SPAN calibration by the value input is possible even when there is no weight.

Step 7 Actual weight calibration

Step 9

Step 8 Actual weight calibration Step 8

Finish the calibration

(END)

→→ →

Example of display changes for setting scale interval.

Set the mass of the weight

ZERO calibration by measurement value.Step 7 Actual weight calibration

kg

SPAN calibration by the weight.Step 8 Actual weight calibration

Step 8

Step 9End of calibration

■ The set value is registering temporary until fixing by pushing the F/ENTER key on   display. The set value is not preserved when canceling the calibration on  the way.

CAL-ZCAL-Z

CAL-ZCAL-Z

CAL-ZCAL-Z

CAL-ZCAL-Z

CAL-ZCAL-Z

Selects the unit by   or   key.CAL-ZCAL-Z CAL-SCAL-S

Selects the position by   or   key.CAL-ZCAL-Z CAL-SCAL-S

CAL-ZCAL-Z CAL-SCAL-S

CAL-SCAL-SCAL-SCAL-S

CAL-SCAL-S

The guide of acalibration prosedure

The guide of acalibration prosedure

The guide of acalibration prosedure

What is the calibration.What is the calibration.

Prosedure for calibration.Prosedure for calibration.

The detail of calibration prosedure.The detail of

calibration prosedure.

When there is a weight. When there is not a weight.

Electric calibration

display by pushing    key.

When   key is pushed at ,

       is displayed.

Select the unit.

After setting, push   key to display .

F

ENTER

The position of decimal point changes as[0 (non)], [0.0], [0.00], [0.000] and [0.0000].

When   key is pushed at       display,

      is displayed.

The scale interval memorized.

The scale interval memorized.

Settingscale interval.Settingscale interval.

Selects the scale interval by   or   key.

Lower 2 digits changes as [01], [02], [05], [10], [20], and [50].

When   key is pushed at       display,

      is displayed.

After setting, push   key to display .

After setting, push   key to display .

When   key is pushed at       display,

      is displayed.

selects the changed digit.

increments the value of selected digit.

Weight value equal to weighing capacity. Weight value equal to weighing capacity.

Weight value put on measurement section.Weight value put on measurement section.

Push   key when displays.

ZERO is memorized, and        is displayed.

Push   key with displayed.

*Please put the weight equal to the mass set to the measuring part by step 6 [Set the mass of weight].

Push thes   key after       blinks, and

      sign lights on.

By pushing key twice, return to the measurement mode.

SPAN is registered by inputting the value in whichthe load cell output value of ZERO is subtractedfrom the output value of the load cell expected whenthe load for measuring is put.

SPAN is registered by inputting the value in which the load cell output value of ZERO is subtracted from the output value of the load cell expected whenthe load for measuring is put.

SPAN is registered by reading the output value in teh condition to put the weight set according to step 6.SPAN is registered by reading the output value in the condition to put the weight set according to step 6.

F

ENTER

The display becoms , and the set data is memorized internally.

Push   key to complete the calibration mode.

SPAN calibration by numeric input.

InstructionmanualInstructionmanual

After the setting, push   key to display .

Measurement mode