Current Meter Maintenance Model OSS-B1. ‘Hydrographic Basics’ Training Course No. 27743. Section 1 Prior To Discharge Measurements. Prior To Discharge Measurements. Remove fan assembly from carry case Turn fan carrier anti-clockwise. Anti-Clockwise. Prior To Discharge Measurements. - PowerPoint PPT Presentation
Citation preview
Current Meter MaintenanceModel OSS-B1
‘Hydrographic Basics’ Training
Course No. 27743
Section 1Prior To Discharge Measurements
Prior To Discharge Measurements
• Remove fan assembly from carry case
• Turn fan carrier anti-clockwise
Anti-Clockwise
Prior To Discharge Measurements
• Separate fan carrier from shaft assembly (keep away from contact with any dirt particles)
• Hold the fan upright and ‘fill’ with approved HS current meter oil
Use Lint Free Tissue
Prior To Discharge Measurements
• Replace axle and fan carrier, excess oil will be forced up through the capillary gap around the axle, ensure fan carrier is tightened
• Wipe the excess oil of
the meter using lint free paper towel
Clockwise
Prior To Discharge Measurements
• Do not forget to close the oil bottle straight after usage to avoid contamination of the oil
• Slide the fan assembly into the body and use the stubby screwdriver to tighten the retaining screw to lock the fan assembly in place
Prior To Discharge Measurements
• Prior to spin test hold the current meter in the horizontal position and spin by hand for approx. one minute to get rid of the hydraulic pressure
• Spin test the meter against time to test the performance of the shaft and bearings
Prior To Discharge Measurements
• Spin test results are dependent on the fan type
• Connect your current meter to the counter using the lead provided (this will test that the lead, reed switch and counter are working prior to use)
• If the current meter fails the spin test, disassemble and clean all components again.
• If the current meter continues to fail the test, replace bearings
OSS-B1 Spin Test
Fan No. Minimum Time (secs)
1 90
A 110
2 90
4 90
Section 2During and After Discharge
Measurements
During and after Discharge Measurements
After every discharge measurement, (or one hour of use), the current
meter should be cleaned and refilled with oil, if continuing a discharge
measurement.
During and after Discharge Measurements
• Undo the cleaning jar cap and ‘¾ fill’ with white spirit
• Remove fan assembly from meter body by undoing the retaining screw and separate the body from the fan
Cleaning Jar
During and after Discharge Measurements
• Remove fan carrier and shaft assembly (do not place in contact with dirt particles)
• Empty the old oil in a recycling container and dispose (do NOT re-use)
During and after Discharge Measurements
• Put the shaft and bearing assembly into the cleaning jar.
• Agitate the jar • Fill up the fan with white
spirit, place your thumb over the hole and shake vigorously for a few seconds.
• Current meter body should be wiped clean with white spirit (there is no need to immerse)
During and after Discharge Measurements
• Blow dry the shaft assembly, fan carrier and fan.
• Re-assemble the fan, coat bearings with a small amount of oil if discharge measurement is complete
During and after Discharge Measurements
• Wipe box clean and return the fan assembly to the case
• If continuing a discharge measurement, refill the fan with oil
During and after Discharge Measurements
• Spin test the current meter against time as shown in section 1
• After the spin test is performed connect your current meter to the counter using the lead, to test the lead, reed switch and counter.
Section 3Office Cleaning
Office Cleaning
Office cleaning should be performed on a monthly basis or whenever the
meter is dirty
Office Cleaning
• Separate fan carrier from shaft assembly
• Current meter body should be wiped clean with soapy water (there is no need to immerse)
Anti-Clockwise
Office Cleaning
• Separate fan carrier (1) from shaft assembly (2)
1 2
Office Cleaning
Remove Nut Remove Bearing
Office Cleaning
Remove Outer Spacer
Remove Inner Spacer
Office Cleaning
Remove Bearing Fan assembly fully dismantled ready to clean
Office Cleaning
Place items in sink, filled with hot soapy water, soak for at least 5 minutes
Clean items with a bottlebrush cleaner
Office Cleaning
Cleaning Fan Cleaning Fan Carrier
Office Cleaning
Cleaning outer spacer
Rinse full assembly with tap water
Office Cleaning
• Place full assembly on lint free tissue
• Ready to dry and re-assemble
• Rinse all items in white spirit prior to re-assembly
Office Cleaning
Dry Fan Carrier Dry Fan
Office Cleaning
Dry Outer Spacer Dry Inner Spacer
Office Cleaning
Blow Dry Bearing & Nut
Office Cleaning
Start Shaft Assembly, Bearing First
Assemble Inner Spacer
Office Cleaning
Assemble Outer Spacer
Assemble Bearing
Office Cleaning
Assemble axle nut Dip shaft assembly in white spirit to lubricate
bearings
Office Cleaning
Blow Excess White Spirit
Assemble Axle With Fan Carrier
Office Cleaning
• Re-assemble fan and place in carry case Clockwise
Section 4Calibration
CalibrationAS 3778 ‘Measurement of water flow in open channels, Part 6.1, Section 7.6 ‘Recalibration’ states:
“Meters shall be recalibrated whenever their performance is suspect. In practice, for
individually rated meters, recalibration is sometimes carried out at yearly intervals or after
300 h of use, whichever is the shorter.”