The F-10 is used to measure currents on 50Hz, 60Hz, and 400 Hz power lines. The F-10 current probe has a transfer impedance of .25Ω from 2kHz to 2MHz. The F- 10 has a usable frequency range overlapping the fundamental powerline frequencies. It can handle 350 amperes of primary power line currents from DC to 400 Hz without saturation. F-10 Current Probe Fischer Custom Communications, Inc. F-10 Specifications Frequency: 100 Hz – 3 MHz Connector: Type-N Internal diameter: 32mm DC to 60 Hz: 350 amperes External diameter: 94mm 400 MHz: 350 amperes Height: 64mm RF(CW): 100 amperes Z t Ω 1 : 0.25 Pulse 2 : 100 amperes dBΩ 1 : −12 1: Probe calibrated with 50Ω+j0Ω Load Impedance 2: Depends upon the pulse width and pulse rep. Rate. 20603 Earl St., Torrance, CA 90503 · Phone: 310-303-3300 Fax 310-371-6268 · email: [email protected] © Fischer Custom Communications, Inc. 072803DS-F-10 All rights reserved Printed in USA -43 -38 -33 -28 -23 -18 -13 10Hz 100Hz 1kHz 10kHz 100kHz 1MHz 10MHz Frequency Tranfer Impedance (dB ohms)

Current Probe - · PDF fileThe F-10 is used to measure currents on 50Hz, 60Hz, and 400 Hz power lines. The F-10 current probe has a transfer impedance of .25Ω from 2kHz to 2MHz

  • Upload
    vohuong

  • View
    221

  • Download
    2

Embed Size (px)

Citation preview

Page 1: Current Probe -  · PDF fileThe F-10 is used to measure currents on 50Hz, 60Hz, and 400 Hz power lines. The F-10 current probe has a transfer impedance of .25Ω from 2kHz to 2MHz

The F-10 is used to measure currents on 50Hz, 60Hz, and 400 Hz power lines. The F-10 current probe has a transfer impedance of .25Ω from 2kHz to 2MHz. The F-10 has a usable frequency range overlapping thefundamental powerline frequencies. It can handle 350 amperes of primary power line currents from DC to 400Hz without saturation.

F-10Current Probe

Fischer Custom Communications, Inc.

F-10 Specifications Frequency: 100 Hz – 3 MHz Connector: Type-N Internal diameter: 32mm DC to 60 Hz: 350 amperes External diameter: 94mm 400 MHz: 350 amperes Height: 64mm RF(CW): 100 amperes ZtΩ1: 0.25 Pulse2: 100 amperes dBΩ1: −12 1: Probe calibrated with 50Ω+j0Ω Load Impedance 2: Depends upon the pulse width and pulse rep. Rate.

20603 Earl St., Torrance, CA 90503 · Phone: 310-303-3300 Fax 310-371-6268 · email: [email protected]

© Fischer Custom Communications, Inc. 072803DS-F-10