24Feb2015EndofDayShutdown
CyTOF2EndoftheDayShutdown(SoftwareVersion6.0.626)
A. Washingsamplelines(Afteryourlastsamplehasbeenrun)1. FollowthefinalwashproceduresoutlinedintheSampleAcquisitionSOP.2. Oncebackgroundlevelsareacceptable,we’rereadytoshutdown.
B. TurningoffthePlasma
1. Clickthe“About”TabontheCyTOFhomewindow
a. Select“Login”intoadministratoruseraccount1. Enablesaccesstothe“ControlPanel”2. Administrator/donuts
2. Clickthe“ControlPanel”TabintheCyTOFMainMenutoopenthecontrolpanel3. Clickthe“Plasma”TabtoopenthePlasmacontrol4. Click“StopPlasma”
a. Shutdowntakesabout5-10minutes
24Feb2015EndofDayShutdown
b. TheSyringePump,ChillerandHeaterwillshutoffaspartoftheshutdownprocedure
5. WaitforthepromptindicatingasuccessfulshutdownbeforeproceedingtoPartC
C. TurnoftheArgonGasflowontheTank1. Ifgasisbeingdrawnfromacylinder,pleaseshutoffgasflowattheCylinder2. IfgasisbeingdrawnfromtheArgondewar,noactionisrequired.
24Feb2015EndofDayShutdown
D. RemovetheNebulizerfromtheNebulizerport
E. Disassemblethenebulizer1. RemovetheSampleCapillaryfromtheNebulizer2. PlacetheFlangelessfixtureinitshomeunderneaththeheater
3. DisconnectthegaslinefromtheNebulizera. UnscrewtheSwagelokfromtheNebulizersidebeingcarefultocatchtheferruleandtheO-ring
1. RecommendthatyouholdtheNebulizerupsidedowntocatchallthreepiecesoftheattachmentassembly
a. WhiteNylonFerrule
24Feb2015EndofDayShutdown
b. BlackO-ringc. WhiteSwageloknut
b. RemovetheNebulizerc. ReassembletheSwagelokfixturewiththeferruleandO-ringembeddedontothegaslineandstorethegaslineonthehousingnexttotheHeatercage
d. FilltheNebulizerwiththemilliQwaterusingasyringeattachedtotheNebulizersidearmandtheNebulizersampleline
e. PlacethenebulizerinitsStorageCasefilledwithMilliQwaterandsoakovernightoruntilnextuse
24Feb2015EndofDayShutdown
F. ClosetheCyTOFsoftwarepackageG. Doublecheckthatyouperformeda“FACSDataCheck-in”H. Logoutofyourwindowsaccounttostopthatday’saccounting
LeaveallothercomponentsontheCyTOFrunningindefinitely