12
1 DVLS³ (S)imply, (S)mart, multi-(S)ensor “Focus on Hydrogen” Exclusive North American Distributor

Da Vinci DVLS3 Simply, Smart Hydrogen Sensor for Gas Chromatographs

Embed Size (px)

DESCRIPTION

Envantage is the exclusive North American distributor of the Da Vinci DVLS3 hydrogen sensor. Install this sensor in your GC to reduce the risk of explosion due to hydrogen leaks that may develop around your instrument hot zones.

Citation preview

Page 1: Da Vinci DVLS3 Simply, Smart Hydrogen Sensor for Gas Chromatographs

1

DVLS³ (S)imply, (S)mart, multi-(S)ensor “Focus on Hydrogen”

Exclusive North American Distributor

Page 2: Da Vinci DVLS3 Simply, Smart Hydrogen Sensor for Gas Chromatographs

2 2 29 November 2012

4 Reasons for using H2 as a Carrier Gas

1. Fast Analysis: Fast diffusion rate; 4 times faster than

N2

Half as viscous as helium; higher linear gas velocity and shorter retention times

2. High Efficiency: Flattest Van Deemter curve

DVLS³ (S)imply, (S)mart, multi-(S)ensor

Page 3: Da Vinci DVLS3 Simply, Smart Hydrogen Sensor for Gas Chromatographs

3 3 29 November 2012

4 Reasons for using H2 as a Carrier Gas

3. Prolonged Column Life: Some applications have a lower elution temperature, therefore the column life is longer

4. High Efficiency: 3x’s less expensive than its helium equivalent

Bottle or generator

DVLS³ (S)imply, (S)mart, multi-(S)ensor

Page 4: Da Vinci DVLS3 Simply, Smart Hydrogen Sensor for Gas Chromatographs

4 4 29 November 2012

Reluctant to use H2 as a Carrier Gas?

Hydrogen is an Explosive Gas: Undetected gas leaks can lead to

an explosion in the GC Oven

LEL 4% hydrogen in Air

DVLS³ (S)imply, (S)mart, multi-(S)ensor

Page 5: Da Vinci DVLS3 Simply, Smart Hydrogen Sensor for Gas Chromatographs

5 5 29 November 2012

DVLS³ (S)imply, (S)mart, multi-(S)ensor Hydrogen Sensor

Page 6: Da Vinci DVLS3 Simply, Smart Hydrogen Sensor for Gas Chromatographs

6 6 29 November 2012

DVLS³ (S)imply, (S)mart, multi-(S)ensor Sensor Installed in the GC Oven

Page 7: Da Vinci DVLS3 Simply, Smart Hydrogen Sensor for Gas Chromatographs

7 7 29 November 2012

DVLS³ (S)imply, (S)mart, multi-(S)ensor Carrier Gas Switch

Page 8: Da Vinci DVLS3 Simply, Smart Hydrogen Sensor for Gas Chromatographs

8 8 29 November 2012

DVLS³ (S)imply, (S)mart, multi-(S)ensor Sequence Stop after Analysis

Page 9: Da Vinci DVLS3 Simply, Smart Hydrogen Sensor for Gas Chromatographs

9 9 29 November 2012

DVLS³ (S)imply, (S)mart, multi-(S)ensor Calibration

Zero Point Calibrated Using Air

Alarm level Calibrated using Calibration Mixture

Yearly or after maintenance or repair

Page 10: Da Vinci DVLS3 Simply, Smart Hydrogen Sensor for Gas Chromatographs

10 10 29 November 2012

DVLS³ (S)imply, (S)mart, multi-(S)ensor Alarms

Acoustic

Optical (blinking display)

SMS Message

Page 11: Da Vinci DVLS3 Simply, Smart Hydrogen Sensor for Gas Chromatographs

11 11 29 November 2012

DVLS³ (S)imply, (S)mart, multi-(S)ensor Catalystic Pellistor Sensor & Wheatstone Bridge

Page 12: Da Vinci DVLS3 Simply, Smart Hydrogen Sensor for Gas Chromatographs

12 12 29 November 2012

DVLS³ (S)imply, (S)mart, multi-(S)ensor Summary & Benefits

Catalytic Pellistor gas specific sensor, linear range of 0-2% H2 (0-50% LEL)

Unaffected by humidity, stable output for long periods, more resistant to shocks and vibrations.

Expected lifetime: over five years

Long term stability; drift sensitivity: less than 2mV a month

User defined alarm: optical, acoustic and/or SMS up to 50% LEL

Instrument readings: provide real time sensor readings with alarm levels, channel states

User menu for configuration (password protected)

Valve : High pressure 3 way solenoid valve

Oven operating temperature: up to 450ο C

Multiple Sensors: Max 4 sensors individually controlled

Sensor options: temperature, barometer, level, oxygen or hydrogen as

a leak detector.