Data management of Argo oxygen data. V. Thierry, D. Gilbert, T. Kobayashi. Profiling floats equipped with O2 sensor. DO concentration data from 426 profiling floats (~ 117 active floats) are distributed through the Argo data stream. DO concentration measurements. DO concentration measurements. - PowerPoint PPT Presentation

Citation preview

Page 1: Data management of Argo oxygen data

Data management of Argo oxygen data

V. Thierry, D. Gilbert, T. Kobayashi

Page 2: Data management of Argo oxygen data

Profiling floats equipped with O2 sensor

DO concentration data from 426 profiling floats (~ 117 active floats) are distributed through the Argo data stream

Page 3: Data management of Argo oxygen data

DO concentration measurements

Page 4: Data management of Argo oxygen data

DO concentration measurements

Page 5: Data management of Argo oxygen data

Data processing

Data processing at DAC level

Unit conversion

Internal processing

Transmission of sensor output

SBE43

*****Unit: Count, Hz or V

DOUnit: ml/L

SBE DOXYUnit: u-mol/kg

Optode

Phase shiftUnit: degree

DOUnit: u-mol/L

DOUnit: u-mol/L

Optode DOXY (1)Unit: u-mol/kg

DOUnit: u-mol/L

DOUnit: u-mol/L

Optode DOXY (2)Unit: u-mol/kg

MeasurementPhase shift Unit: degree

Optode

DOUnit: u-mol/L

*****Unit: Count, Hz or V

Phase shiftUnit: degree

AdjustmentSalinity and pressure

Page 6: Data management of Argo oxygen data

Present parameters and recommendations for the management of DO data

• Unit: mol/kg

• DOXY: DO concentration in mol/kg either deduced from the sensor output or transmitted by the Aanderaa optode at Pres=0 and Psal=S0 (and converted from mol/l to mol/kg ).

• DOXY_ADJUSTED: P, S compensation applied for the Aaanderaa optode when DO is transmitted

• BPHASE_DOXY: BPHASE measured by the Aanderaa optode, when it is transmitted

• TEMP_DOXY: Temperature measured by the Aanderaa otpode, when it is transmitted

Page 7: Data management of Argo oxygen data

Management of the O2 data at the DAC level

• Present parameters do not allow us to manage in a consistent way the O2-related variables (with different units) transmitted to the DAC

• The data processing varies from one PI to the other and the data and metadata are reported in different ways

• Unit conversion not systematically applied

• Salinity and pressure compensation applied in the “ADJUSTED” field

• The available DOXY from the different floats are not comparable and a user cannot use with confidence O2 data from other groups

Need to standardize procedures and to provide clear recommendations

Action item 14 from AST10 : D. Gilbert to work with T. Kobayashi and V. Thierry to ensure DACs are processing oxygen data according to recommendations.

We do not address QC issue

Page 8: Data management of Argo oxygen data

• DOXY = DO concentration estimated from the telemetered variables, calibrations coefficients and CTD values: PRES, TEMP (or TEMP_DOXY) and PSAL

• Pressure and salinity compensations (optode) are taken into account

• Converted in the official Argo unit: mol/kg

• DOXY_ADJUSTED is the DO concentration corrected for any sensor drift and offset. DOXY_ADJUSTED is calculated from the other “ADJUSTED” fields.

• Calibration coefficients, equations and references used to convert the telemetered variables in DOXY must be carefully documented in the metadata

Proposition for what could be DOXY

Page 9: Data management of Argo oxygen data

What variables to store ? 1st solution

• Store each telemetered variables in new parameters with meaningful names:– VOLTAGE_DOXY (Unit = V) for SBE43

– FREQUENCY_DOXY (Unit = Hz) for SBE43

– COUNTS_DOXY (no Unit ?) for SBE43

– BPHASE_DOXY (Unit = degree) for Aanderaa optode

– DPHASE_DOXY (Unit = degree) for Aanderaa optode

– CONCENT_DOXY (Unit = mol/l) for Aanderaa optode

– TEMP_DOXY (Unit = degree Celsius) for Aanderaa optode

– XXX_DOXY for any new variables

• Calculate and store DOXY data from the telemetered variables and converted in in M/kg– DOXY (mol/kg)

Page 10: Data management of Argo oxygen data

• Store only DOXY data converted in a DO concentration value consistent with all other from whatever is transmitted by the floats and converted in M/kg :– DOXY (mol/kg)

• Do not store anymore– BPHASE_DOXY (Unit = degree) for Aanderaa optode

– TEMP_DOXY (Unit = degree Celsius) for Aanderaa optode

• “DOXY_ADJUSTED” fields are corrected for any sensor drift or offset of the DO sensor

What variables to store ? 2nd solution

Page 11: Data management of Argo oxygen data

Any other proposition ?

Page 12: Data management of Argo oxygen data

• Do we cover all cases ?• Is 256 character enough to document the processing of O2

data in the “PREDEPLOYMENT_XXX” fields of the metadata file ?

• Need help to review the different cases and to provide advise on how to fill the metadata

• A proposal has been written to provide all necessary information to the DAC. This proposal will be updated according to the agreed solution and provided to anyone interested.

Questions/remarks

Page 13: Data management of Argo oxygen data

Metadata

• Carefully document the telemetered variables in the metadata

• Example for the aanderaa optode transmitting DO concentration in mol/L

SENSOR_UNITS = “micromole/L”

PARAMETER= “CONCENT_DOXY”

PREDEPLOYMENT_CALIB_EQUATION = “O2=C0+C1.P+C2.P^2+C3.P^3+C4.P^4 with Ci=Ci0+Ci1.t+Ci2.t^2+Ci3.t^3 ; CONCENT_DOXY=O2 exp[ S (B0+B1.Ts+B2.Ts^2+B3.Ts^3)+C0S^2]; Ts=ln[(298.15-t)/(298.15+t)]; with P=DPHASE, t=TEMP_DOXY, S=S0”

PREDEPLOYMENT_CALIB_COEFFICIENT = “S0=35” or “S0=0”

PREDEPLOYMENT_CALIB_COMMENT = “CONCENT_DOXY is calculated by the optode from DPHASE, TEMP_DOXY, PRES=0 and S0; see TD218 operating manual oxygen optode 3830, 3835, 3930, 3975, 4130, 4175”

Page 14: Data management of Argo oxygen data

Exemple: transmission of CONCENT_DOXY = DO concentration from optode at PRES=0 and PSAL=S0

• PARAMETER=”DOXY”

• PREDEPLOYMENT_CALIB_EQUATION = “O2c= O2 exp[ (S-S0) (B0+B1Ts+B2Ts^2+B3Ts^3)+C0(S^2-S0^2)]; DO=O2c [1+ 0.032 * PRES/1000]; O2=CONCENT_DOXY; S=PSAL; Ts=ln[(298.15-t)/(298.15+t)]; t=TEMP (or TEMP_DOXY); DOXY[micromole/kg]=DO [micromole/L] / ρ ; ρ= potential density [kg/L] at zero pressure and potential temperature”

• PREDEPLOYMENT_CALIB_COEFFICIENT = “S0=35 (or any other value); B0= -6.24097e-3; B1=-6.93498e-3; B2=-6.90358e-3; B3=-4.29155e-3 ”

• PREDEPLOYMENT_CALIB_COMMENT = “Pressure and salinity compensation are estimated to convert CONCENT_DOXY in DOXY. DOXY is converted from micromole/L to micromole/kg; see TD218 operating manual oxygen optode 3830, 3835, 3930, 3975, 4130, 4175”